laser scanning system testing—errors and improvements

11
ELSEVIER Measurement 16 (1995) 91 101 Measurement Laser scanning system testing Errors and improvements William Xinzuo Li, Larry D. Mitchell * Virginia Polytechnic Institute and State University, Department of Mechanical Engineering, Blacksburg, VA 24061-0238. USA Abstract The Scanning Laser Doppler Vibrometry (SLDV) technique has brought modal testing into a new era. A galvanometer-based laser scanning system for SLDV provides the position accuracy, speed, and flexibility for data acquisition. A novel parallel-shift method has been developed for testing and calibration of the scanning system to meet the precision requirements of modal testing. This parallel-shift method can provide a cost-effective means for a systematic laser scanning accuracy test. However, a number of measurement errors could occur during the scanning accuracy test. These errors could severally affect the accuracy of the test itself. Quantitative determination of the effects of these errors is necessary to evaluate and to improve the accuracy of the test. This paper gives a detailed analysis for all the errors involved in the galvanometer-based laser scanning accuracy test using the parallel-shift method. Improvements of the test setup and procedure are also proposed. Keywords: Laser scanner; Galvanometer; Parallel-shift method; Scanner test; Calibration List of Symbols d distance from the center of the mirror surface to the input laser beam dop optimal distance from the center of the mirror surface to the input laser beam e mirror surface offset error k thickness of the: mirror z percentile of the: standard normal distribution Z distance between the X-mirror and the scanned X- Y plane AX laser beam parallel shift in the X direction AY laser beam parallel shift in the Ydirection AZ laser-scanner unit movement in the Z direction laser beam scanning angle (optical) * Corresponding author. 0263-2241/95/$09.50 © 1995Elsevier ScienceB.V. All rights reserved SSDI 0263-224l(95)00025-9 ~h angle between the Z-axis and the laser beam at home position displacement of the pinhole ,~: change of the variance that corresponds to the pinhole displacement ,~ ex error in the measurement of AX e~x~ angular error of non-orthogonality of the X and Z stages e= angular error in the scanning angle ~ angular error in the scanning angle ~ due to calcualtion 0 nominal angle of the input laser beam and the mirror surface at home position ~r sample standard deviation 1. Introduction One of the most advanced modal testing tech- niques is to use a scanning Laser Doppler

Upload: william-xinzuo-li

Post on 21-Jun-2016

224 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Laser scanning system testing—Errors and improvements

E L S E V I E R Measurement 16 (1995) 91 101

Measurement

Laser scanning system testing Errors and improvements

William Xinzuo Li, Larry D. Mitchell *

Virginia Polytechnic Institute and State University, Department of Mechanical Engineering, Blacksburg, VA 24061-0238. USA

Abstract

The Scanning Laser Doppler Vibrometry (SLDV) technique has brought modal testing into a new era. A galvanometer-based laser scanning system for SLDV provides the position accuracy, speed, and flexibility for data acquisition. A novel parallel-shift method has been developed for testing and calibration of the scanning system to meet the precision requirements of modal testing. This parallel-shift method can provide a cost-effective means for a systematic laser scanning accuracy test. However, a number of measurement errors could occur during the scanning accuracy test. These errors could severally affect the accuracy of the test itself. Quantitative determination of the effects of these errors is necessary to evaluate and to improve the accuracy of the test. This paper gives a detailed analysis for all the errors involved in the galvanometer-based laser scanning accuracy test using the parallel-shift method. Improvements of the test setup and procedure are also proposed.

Keywords: Laser scanner; Galvanometer; Parallel-shift method; Scanner test; Calibration

List of Symbols

d distance from the center of the mirror surface to the input laser beam

dop optimal distance from the center of the mirror surface to the input laser beam

e mirror surface offset error k thickness of the: mirror z percentile of the: standard normal distribution Z distance between the X-mirror and the

scanned X- Y plane AX laser beam parallel shift in the X direction AY laser beam parallel shift in the Ydirection AZ laser-scanner unit movement in the Z

direction laser beam scanning angle (optical)

* Corresponding author.

0263-2241/95/$09.50 © 1995 Elsevier Science B.V. All rights reserved SSDI 0263-224l(95)00025-9

~h angle between the Z-axis and the laser beam at home position displacement of the pinhole

,~: change of the variance that corresponds to the pinhole displacement ,~

ex error in the measurement of AX e~x~ angular error of non-orthogonality of the X

and Z stages e= angular error in the scanning angle ~ angular error in the scanning angle ~ due to

calcualtion 0 nominal angle of the input laser beam and

the mirror surface at home position ~r sample standard deviation

1. Introduction

One of the most advanced modal testing tech- niques is to use a scanning Laser Doppler

Page 2: Laser scanning system testing—Errors and improvements

92 W.X. Li, L.D. Mitchell/Measurement 16 (1995) 91-101

Vibrometer (LDV) to obtain dynamic information of a structure [1]. The accuracy of the laser scanning system of the LDV has a direct effect on the accuracy of the modal testing. A parallel-shift method [2] was developed for a systematic laser scanning accuracy test of a galvanometer-based laser scanning system [3] (DE2488/G3B laser scanning system, manufactured by General Scanning Inc.). Fig. 1 shows the laser scanning accuracy test setup that was based on the parallel- shift method. The scanning angle ~ (with respect to the Z-direction) is determined by the following equation:

,fAX~ : : tan- t~--Z-7, (1)

where AZ is the distance that the laser-scanner unit is moved along the Z-axis using the Z transla- tion stage, and AX is the corresponding parallel shift of the laser beam in the X-direction as deter- mined by a photodetector and the X translation stage. The detailed descriptions of the test setup and test procedures can be found in the work by Li [4].

Although this parallel-shift method provides a cost-effective way to test an ultra-precision laser scanning system, it is inevitable that the scanning accuracy test itself involves a number of errors. It is very important to identify these errors and their effects before the results of the scanning accuracy test can be correctly interpreted.

2. Scanning mirror surface offset error

For the purpose of dynamic balance, the mirror of the scanner is mounted symmetrically with respect to the axis of rotation of the scanner. This minimizes the mirror's mass moment of inertia. The mirror surface is off the axis of rotation. Since the incident point of the laser beam cannot be on the axis of rotation, it travels on the mirror surface as the mirror rotates. Fig. 2 shows the situation. The mirror position A is the home position that forms an angle 0 with the input laser beam. When the mirror rotates an angle ~/2 about the axis of rotation to the position B, the output laser beam scans an angle ~. However, since the intersecting point of the input laser beam and the mirror's front surface changes position as the mirror rotates, the output laser beam scans and is also shifted a distance, e, in the direction parallel to the home- position input laser beam. This offset error, e, of the scanned laser beam can be found to be

ds in~ +~sinO 1-cos

e - , (2)

sin ( 0 - 2 ) s i n 0

where d is the distance from the center of the mirror surface to the input laser beam and k is the mirror thickness which is 3 mm (see Fig. 2). The nominal angle between the input laser beam and the mirror surface at the home position, 0, is 45 ° for the X scanner and 53.5 ° for the Y scanner.

Fig. 1. Top view of the setup for the scanning accuracy test.

Page 3: Laser scanning system testing—Errors and improvements

W.X. Li, L.D. Mitchell~Measurement 16 (1995) 91-101 93

Home position of the mirror \

Center of the mirror s u f f a c e ~

Axis of rotation \

V

Mirror front surface /

Output laser beam (at home position) Virtual home position

Fig. 2. The offset error of the scanned laser beam.

Eq. (2) is plotted in Fig. 3 for the X scanner (the plot for the Y scanner is very similar). Three cases of d are plotted. The first case is that the input laser beam is aimed at the center of the mirror surface (d = 0). The other two are of the input laser beam 0.5 m m above and 0.5 m m below the center of the mirror surface (d = 0.5 m m and d = - 0 . 5 ram, respectively). For 0 = 4 5 ° and d = 0 . 5 m m , the output laser beam is shifted laterally 0.2538 m m when it scans + 2 0 ° , which corresponds to an angular error of 224 ~trad for a test object which is 1 m away or 22.4 ~trad when 10 m away from

the scanner. This error certainly affects the scan- ning accuracy, especially when the test object is close to the scanner. To minimize this error, we should find the optimal location of the laser relative to the center of the mirror surface, dop. F rom Fig. 3 we can see that the offset error function (laser beam parallel shift), e, must cont inuously rotates as we change d from +0.5 m m to - 0 . 5 mm. Notice that at d = 0 . 0 m m (the laser beam is aimed at the center of the mirror surface), the offset error func- tion approaches zero, but the error, e, at ~ = + 20 ° is larger than the offset error at ct = - 20 '~. Thus,

0.3

0.2 q~ .J~

0.1 e,)

o

~-0.1

~ -0.2

-0.3

~ d , , O * d,,O.Smm . ,

- - - d ~ - O , 6 m m . ~"

-20-18-16-14-12-I0-8 -6 -4 -2 0 2 4 6 8 I0 12 14 16 18 20

Scanning Angle (degree optical)

Fig. 3. Theoretical laser beam offset error for the X scanner.

Page 4: Laser scanning system testing—Errors and improvements

94 W.X. Li, L.D. Mitchell/Measurement 16 (1995) 91 101

the minimal value of the offset error over the scan angle range can be obtained when the offset error at c~= +20°and ~= - 2 0 ° are equal. Using Eq. (2) we can equate e-2o = e+2o and solve for dop:

1 k (1 coslOO), (3) d o p : ~ COS 0

from which, dop=-0.016 mm for the X scanner and d o p : -0.014 mm for the Y scanner. However, since the test method for the scanning accuracy measures the true scanning angle only, the laser beam offset error does not affect the determination of scanning angles, no matter how large it is.

3. Uncertainty of determining the position of the laser spot

The position of the focused laser spot is deter- mined by the photodetector, along with the X and Y stages (refer to Fig. 1). A 25 Ixm pinhole is set in front of the detector to allow only partial laser radiation passing through. The output voltage of the photodetector reaches the maximum when the center portion of the laser beam passes through the pinhole and strikes on the detecting area. The multimeter (Hewlett-Packard 3478A) used to dis- play the output of the photodetector is set to have a resolution of 0.0001 VDC. Although the maxi- mum resolution of the multimeter is 0.00001 VDC, it is too sensitive to the noise of the detecting circuit for this maximum resolution.

When the scanner is about 1 m away from the photodetector, the maximum output of the detector is about 0.4700VDC without the pinhole or 0.3600 VDC with the pinhole for the fine focused laser beam. This implies that about 76.6% of the total laser energy passes through the pinhole when the center of the laser beam coincides with the pinhole's center. Since 99% of the He-Ne laser beam, which was used in the setup, is of the Gaussian TEM00 mode, the cross-section intensity profile of the laser beam is a bi-normal distribution function [5]. Fig. 4 represents the change of the laser energy that passes through the pinhole due to the displacement of the pinhole as a projection into a single flat plane normal distribution. When

mt "si pro lo

i i ~ l ~ e d pinhole

z =-1.19 . 0.025 mm ~ z= 1.19 ,

~ ] Displaced pinhole

/ z= ~z - 1.19 Z£~z + 1.19

Fig. 4. Change of laser energy due to the displacement of the pinhole.

the pinhole is centered with the laser beam, the laser energy passing through the pinhole corres- ponds to the shaded area between z = - 1.19 and z= 1.19, where z is defined as the percentile of the standard normal distribution. If the pinhole is laterally displaced by a distance 6, the shaded area that corresponds to the passing laser energy will be between z=6~-1.19 and z=6~+ 1.19, where 6_, is the change of the variance that corresponds to the pinhole displacement 6. The minimal change of the laser energy that can be detected by the photodetector depends on the sensitivity of the multimeter. For this test setup, the 0.0001VDC multimeter resolution over the maximum output of 0.3600 VDC gives a sensitivity of 0.0278%. This sensitivity corresponds to an area change of 0.000278 for the normal distribution function. The change of the variance, 6z, for the area change then is 0.034. This gives the minimal pinhole displacement, 6, of 0.4 Jam. This is only an approxi- mation due to the fact that both the pinhole and the cross-section of the laser beam are circles, not flat planes. Besides, the intensity of the laser beam usually is not a perfect bi-normal distribution function due to the laser noise and the distortion caused by the beam expander and the mirrors. In the experiment for determining the laser beam's position, the photodetector was slowly moved

Page 5: Laser scanning system testing—Errors and improvements

W.X. Li, L.D. Mitchell~Measurement 16 (1995) 91-101 95

across laser beam by driving the X stage. When the output of the photodetector reached the maxi- mum value, the position of the photodetector was recorded. The same procedure was repeated 10 times for each laser beam position. The 3a value of the measurements about the mean position of the laser beam is about 5 gm, where ~ is the sample standard deviation. This uncertainty involves the multimeter sensitivity, resolution of the stage's micrometer, backlash of the stage, drift of the laser and scanner, and vibration of the workbench and the test setup.

For the thermal drift and repeatability tests, the angular error in the test that is caused by the uncertainty error is small if the distance between the scanner and the photodetector is large. For instance, the uncertainty error of 5 gm for the determination of the laser beam's position gives only 5 grad angular error for a distance of 1 m between the scanner and the detector. However, the uncertainty error has a large impact on the setup for the scanning angle test. This is because the accuracy of the scanning angle test depends on the travel ranges of the X and Z translation stages, not on the distance between the scanner and the detector (refer to Eq. (1)). The maximum travel range is 25 mm for the X stage and 100 mm for the Z stage. The uncertainty error of 5 lain in AX can cause an angular error of 64 grad in the scanning angle computat ion (using Eq. (1)), if the full travel range of the X stage is used. The angular error can be even higher if only part of the travel range is used in the scanning angle test.

4. Laser beam drift error

The error due to the laser beam drift perhaps is the most serious problem for this test setup. The angular drift of the laser beam with respect to the photodetector comes from several sources.

The first source of drift is the laser itself. This drift should be less than 10 grad according to the manufacturer 's specification [6] . The actual laser drift is confirmed to be within 5 grad for a 30 minutes period by using the test setup. It seems that the laser drift is small and random. Because of the uncertainty error of determining the laser

beam's position, the test acc-;racy for the laser drift cannot be better than 5 grad. Also, the laser drift test with a longer period is not reliable because it would involve the thermal drift due to other com- ponents of the setup since there is little control of the room temperature.

The second cause of the laser beam drift is from the drift of the scanners. The scanner's position transducer and the scanner controller are sensitive to the change of the temperature. According to General Scanning's specifications [6] , the minimal drift of the G3B scanner (with temperature control) is 10 ppm/"C for gain drift and 10 g r a d / C for zero drift, which gives a total drift of 15 grad/ (7 at 2 0 scanning angle. There is no temperature control for the scanner controller ( DE2488 ). Also, General Scanning does not specify the drift on the output of the scanner controller. The output of the control- ler is defined as the digital-to-analog converter (DAC) output. Fig. 5 shows the laser beam angular drift that corresponds to the measured DAC output drift of the X servo and the position output drift of the X scanner. Fig. 6 shows the same similar drifts for the Y servo and the Y scanner. The slope of each fitted line is the gain drift in the unit of microradians per optical degree. The intercept of each fitted line is the zero drift in the unit of microradian. From the fitted lines of the position sensor output for the X and Y scanners, the total drift at 20 ~' scanning angle was about 112.1 and 2.5 grad/~C, respectively.

It can be seen that part of the drift of the position sensor output is caused by the drift of the DAC output. The measured position sensor drift can be used as an approximation of the scanner's drift, but it might not represent the actual angular drift of the scanner. The actual drift of the scanner can be detected by measuring the drift of the laser beam. This presents some difficulties in the current test setup because of the other drifts involved, such as the laser drift and the thermal effect of the work-bench. The laser drift appears small and random as discussed previously. The thermal effect of the work-bench is much more complicated. It may be the most serious cause of the drift of the laser beam during the scanning angle test. The work-bench has a 45 mm thick wooden top. The frame of the work-bench is made of structure steel

Page 6: Laser scanning system testing—Errors and improvements

96 W.X. Li, L.D. Mitchell~Measurement 16 (1995) 91-101

+ J ]

80

60 ~..

40 L..q L- ~ ~.

20 ' """ "Ik"Jk 0 - ' "lk. .

-20 "" L-. -40 • DAC maimt daft - 6 0 , • Pos i t i on output drift

- - - DAC ~ line -80 y ,, - 1 . 2 t t x + 3 . 9 4

bJIZ IWI

' J , . . j "'d

" " k , -100 " . . . . Pos i t ion ~ line -120 Y " -4.063 x - 31.767

- 2 0 - 1 8 - 1 6 - 1 4 - 1 2 - 1 0 - 8 -6 -4 -2 0 2 4 6 8 10 12 14 16 18 20

Scanning Angle (degree optical)

Fig. 5. Measu red the rma l drift of the X scanner ' s servo and pos i t ion sensor (from 19°C to 27.5°C r o o m temperature) .

. ~ 25

2o

5

}o -10

-15

¢ ' , k - , L . J

• DAC output drift ) ~q b ~. • Pos i t ion output drift

- DAG ~ line y - -0.Mdlx + 1. t38

Position fittml line y a .0.344x + 11.1rr$

- ~ " k--,L..,~. ,..J .A ....

d

- 2 0 - 1 8 - 1 6 - 1 4 - 1 2 - 1 0 - 8 -6 -4 -2 0 2 4

Scanning Angle (degree optical)

6 8 10 12 14 16 18 20

Fig. 6. M e a s u r e d the rma l drift of the Y scanner ' s servo and pos i t ion sensor (from 20°C to 27.5°C r o o m temperature) .

and sheet metal. The work-bench top is firmly mounted on the frame along all four edges. There is no support under the middle of the work-bench top. The Z rail is made of aluminum and mounted on the center of the work-bench top (refer to Fig. 1). Since the thermal conductivity of alumi- num is almost 1,200 times higher than that of wood, the response to the change of room temper- ature is much greater for the rail than for the bench top. The difference in thermal expansion of the rail and the bench top causes the bench top to bend, which has the same effect on the laser beam as if it were drifting due to temperature change. To qualitatively determine the thermal effect of the work-bench, two experiments were conduced. The laser-scanner unit was set at horizontal position in one experiment and at vertical position in the other experiment. In order to eliminate the effect

of the scanner's drift, a front-surface plane mirror was firmly set between the laser and the scanner to deflect the laser beam directly to the photodetec- tor. Prior to the experiments, the room was cooled down by the use of an air conditioner. After the air conditioner was turned off, the room temper- ature rose gradually and was monitored. The drift of the focused laser spot relative to the detector's position was measured by moving the X and Y translation stages (refer to Fig. 1). The distance between the laser unit and the photodetector was about 1.4 m which gave an angular accuracy of 3.6 grad for the laser beam position error of 5 [am. The data were taken approximately every half hour. Figs. 7 and 8 show the results of these two experiments. The ranges and rate of the changes of room temperature for two experiments were not the same for lack of effective means to control the

Page 7: Laser scanning system testing—Errors and improvements

W.X. Li, L.D. Mitchell~Measurement 16 (1995) 91 I01 97

80

70

6O

"~ 50

40

} 30

20

to

,-1

-10

• Drift In X direction J I • Drift in Y direction --

- - - Fi#ld in X direction y ,. | . Ix - 13t -- ~ " Fitlsd in Y direction y m 8 1 . 1 x - 1194

~ _ _ 0 _ _ _ b - - - I r ' - - - ~

23.4 23.6 23.8 24 24.2 24.4 24.6 24.8 25

Room Temperature (degree C)

Fig. 7. Measured thermal drift of the laser beam without the scanners (laser-scanner unit at vertical position).

35

i,o 25

"~ 20

15

5

-5

-10

• Drift in X direction • Drift in Y direction ~ A

. . . . FittM in X direction . y - 7.5x - 191 FittIKI in Y direction

l _ . . . . . . . . ,

I I I I I I

25.3 25.4 25.5 25.6 25.7 25.8 25.9 26

Room Temperature (degree C)

Fig. 8. Measured thermal drift of the laser beam without the scanners (laser-scanner unit at horizontal positiont.

room temperature. Although the scanner drift was not included in the test, the laser drift was unable to be singled out from the test results. However, the results of the experiments still showed some consistencies of the measured laser beam drift in the X and Y directions regardless of the position of the laser-scanner unit. From Figs. 7 and Fig. 8, the laser beam drift in the X-direction was small, 5.6 ~trad/°C drift at the vertical position (Fig. 7, X- direction) and 7.5 p.rad/°C drift at the horizontal position (Fig. 8, X-direction) of the laser-scanner unit. The main part of the X-direction laser beam drift is most likely contributed by the laser drift. The laser beam drift in the Y-direction was much larger and obviously related to the change of room

temperature that produced the thermal effect of the bench top. The measured laser beam drift in the Y-direction was about 51.1 i.trad/°C at the vertical position (Fig. 7, Y-direction) and 45.5 I.trad/°C at the horizontal position (Fig. 8, Y- direction) of the laser-scanner unit.

The laser beam drift can give a serious problem to the scanning angle test, especially the drift in the X-direction. Fig. 9 illustrates the effect of the laser beam drift in the scanning angle test. Suppose the scanning angle to be tested is e when the scanner is at position A. When the scanner travels a distance, AZ, to position A', the focused laser spot at the X - Y plane should move from position B to B'. However, if the laser beam drifts an angle,

Page 8: Laser scanning system testing—Errors and improvements

98 W.X. Li, L.D. Mitchell/Measurement 16 (1995) 91 101

z 1 i AZ _l~

Fig. 9. The effect of the laser beam's thermal drift in the scanning angle test.

~,, in the X-direction, the laser spot will be at position B" instead of B', and the measured move- ment of the laser spot in the X-direction will be ( ~ + AX). If the distance between the scanner and the X - Y plane is Z, then ~x ~ Z . The angular error in the calculated scanning angle, ~,c, accord- ing to Eq. (1), will be

e~c=tan -1 \ AZ } - ~ (4)

When the scanning angle, ~, is small (c~ = 0 is the worst case) and when the laser beam angular drift, e~, is small (the usual case), Eq. (4) becomes

Z e ~ AZ (5)

That is, the laser beam angular drift error is magnified in the scanning angle test by the ratio of the distance, Z, between the scanner and the photodetector to the distance, AZ, traveled by the scanner unit! If in the test setup, Z is 1,000 mm, AZ is 50 mm, and by the laser beam angular drift, e,, in the X-direction (10~trad), the calculated error, ~,~, in the scanning angle can be as large as 200 grad. Unfortunately, the minimal ratio of the distance between the laser-scanner unit and the photodetector over the travel range of the laser- scanner unit is around 20 for the current test setup. This dilemma is caused by the limited maximum travel of the stage that moves the laser-scanner unit and by the minimal distance between the laser-scanner unit and the photodetector which is limited by the minimal focus length of the beam expander.

5. Non-orthogonality of the X and Z translation stages

In all the equations for computing the scanning angle, the movements of the X and Z translation stages, AX and AZ, are assumed to be orthogonal to each other. In reality, it is impossible to set these two stages perfectly orthogonal. The effect of the non-orthogonality error of the X and Z stages in the scanning angle test certainly cannot be neglected. Fig. 10 shows the effect of this error in the scanning angle test. The measured laser spot movement, AX', instead of AX, is used in Eq. (1) to compute the scanning angle. If one lets ex~ be the non-orthogonality error of the X and Z stages and c~ be the true scanning angle to be tested, then the angular error, e~c, in the calculated scanning angle can be expressed as

e~c--tan 1 cos(e+ex:)j_c~" (6)

Clearly, the error in the calculated scanning angle caused by the non-orthogonality error of the stages is independent of the scanner unit travel distance and the distance between the scanner and the photodetector. For a non-orthogonality error of 0.1 ~ ( 1,745 prad) and a scanning angle of 20 °, the error in the calculated scanning angle is about 205 larad. Therefore, it is crucial to reduce the non- orthogonality error for the scanning angle test. One way to reduce the non-orthogonality error of the stages is to use a reference block (refer to Fig. 1). The two sides of the reference block that are along the X and Z rails were precisely

i AZ £

x

Fig. 10. The effect of the non-orthogonality error of the X and Z stages.

I

Page 9: Laser scanning system testing—Errors and improvements

W.X. Li, L.D. MitchelUMeasurement 16 (1995) 91-.101 99

machined to be perpendicular. According to the certificate of the milling machine upon which the aluminum reference block was manufactured, the perpendicularity of the two machined sides should be within 0.0003"/12 '~, which results in an error, ex~,Of about 0.00143 °. For a scanning angle of 20 °, this perpendicularity error using Eq. (6) gives only 3 grad error, ~ , in the calculated scanning angle. To set the movements of the X and Z stages parallel to the machined edges of the reference block, a dial indicator was mounted on each stage. Obviously, the ability to adjust the stages parallel to the reference block will increase this error.

6. Other errors

In the previous analyses, it is always assumed that the homeposition of the laser beam is parallel to the movement of the Z stage, AZ, which is the Z-axis in the scanning angle test. In reality, accu- rate alignment of the home-position of the laser beam with the Z stage is difficult and is not needed. In Fig. I1, the home position of the laser beam forms an angle, ~h, with the Z-axis in the X-Z plane. This angle, ~h, can be treated as a scanning angle and can be determined by exactly the same method as any other scanning angle. The actual tested scanning angle, ~, is then

, ( A x e + ~=tan \AZ]_~h . (7)

The sign of ~h in Eq. (7) depends on the sign of the tested scanning angle, cc Since this home-

Home-position of the laser beam

x

~Xh

position angle is very small and the time needed to measure this single angle is short, the expected error in determining this angle should be smaller than the error in determining a general scanning angle. This is because the effects of some of the errors discussed previously, such as the thermal drift and the non-orthogonality error, will be small in determining this home-position angle. However, the error in this calculated home-position angle will have the same effect on each tested scanning angle. The entire scanner calibration curve will be shifted by the error in :~h.

There is another assumption made throughout the analyses, that is the scanned laser beam is assumed parallel to the X-Z plane that is defined by the movements of the X and Z stages. Again, it is hard to accurately set the scanned laser beam parallel to the X-Z plane. The Y translation stage of the detector unit is used to solve this problem (refer to Fig. 1). If the scanned laser beam is not parallel to the X-Z plane, the displacement of the scanner unit, AZ, causes the focused laser spot to shift not only in the X-direction, AX, but also in the Y-direction (refer to Fig. 1). This laser beam shift in the Y-direction, AY, is measured by the Y stage. To accommodate the laser beam shift in the Y direction, Eq. ( 1 ) can be modified as

, (,/IAxl2+aYt2 = tan \ AZ ]" ( 8 t

So this AY will not affect the accuracy of the scanning angle test. By carefully setting up the laser-scanner unit, A Y can be very small comparing to AX. For example, the largest measured A Y was 0.066 mm for AZ = 50 mm and AX--- 17.968 mm. The calculated scanning angles, c~, were 19.766408' and 19.766531 ° by Eqs. (1) and (8), respectively. The difference of these two calculated angles was only about 2 grad. Thus, the measurement error of A Y is insignificant and can be ignored. For the same reason, the non-orthogonality error of the Y stage with respect to the X Z plane can also be ignored.

7. Improvements

Fig. 11. The home position of the laser beam is not parallel to the Z-axis.

This test setup can provide the test for a single scanning angle with an accuracy of 20 grad (+ la)

Page 10: Laser scanning system testing—Errors and improvements

100 w.x. Li, L.D. Mitchell/Measurement 16 (1995) 91-101

if a large number of measurements are made for that scanning angle. If multiple angles are mea- sured, then significant time will elapse so that temperature change could cause the various sources of thermal drift to increase the measure- ment error. However, if a computer correction of the nonlinearity error is needed, the test setup needs some improvements to dramatically reduce the errors in the scanning accuracy test.

It is clear that the uncertainty error of determin- ing the position of the scanned laser beam and the laser beam drift error have the most effect on the test accuracy. These two errors are not totally unrelated. For instance, the uncertainty error may be reduced statistically by increasing the number of repetitive measurements, but increasing the number of the measurements means increasing the time, which increases the laser beam drift error. The best improvement to these problems would be to read out the position of the deflected laser spot faster and more precisely. A high-spatial-resolution beam profiler or analyzer would be the first choice. Two commercial products that are claimed having sub-micron position resolution are the Optical Profiler made by the Oriel Corporation and the BeamScan made by Photon Inc. Oriel's Optical Profiler can take up to 250,000 samples for a beam profile with a spatial resolution of 0.1 jam. The aperture size of the slit is 2 jam. The scan range is 20 mm for a single axis or 4 mm for dual axes ( X Y ) with a position resolution of 0.05 jam and a scan rate up to 10Hz. Photon's BeamScan has several models that can determine the intensity profile of a laser beam with a diameter of from 5 jam to 8 ram. The spatial resolution of the profile is 0.14 jam for a dual axes profile if a 2 jam slit is used. The maximum update rate is 10 Hz. Photon also manufactures an even higher-resolution beam profiler called SpotScan. SpotScan measures the profile of a laser beam with a diameter of less than 5 jam to 25 jam and a spatial resolution of 5 nm.

If Oriel's Optical Profiler and Photon's BeamScan have the claimed performance, they will be the ideal devices to determine the position of the focused laser beam. The uncertainty error can be expected to be reduced by up to 50 times. An even greater improvement is that the test time for each scanning angle can be reduced from a half

hour down to a few seconds. This certainly mini- mizes the laser beam drift error. Another advantage of using such a beam profiler is that the quality of the laser beam can be readily visualized, so the necessity of laser and optics improvements can be determined. It is possible that even the beam expander is not needed for the test setup. The cost for this type of beam profiler is about $6,000 to $10,000, which is surprisingly lower than the cost of the CCD type of beam profiler.

To improve the quality of the focused laser beam, one would use a high-quality beam expander with a spatial filter, To reduce thermal drift and provide a stable environment for a consistent scan- ning accuracy test, an optical bench would be a good investment. A set of long-range motor-driven linear stages with high-accuracy encoders, used as the X and Z stages (refer to Fig. 1), could not only greatly reduce the test time, intensity of work, and measurement error (refer to Eq. (5)), but also be used to automate the scanning system test and calibration procedure.

8. Conclusions

The parallel-shift method is a very promising method to systematically test and calibrate a galvanometer-based, ultra-precision laser scanning system for modal testing. However, the current cost-constrained test setup requires some improve- ments to reduce the measurement errors. The most important improvement would be to use a high- resolution laser beam profiler, which can not only reduce the uncertainty error of determining the position of the laser beam but also greatly shorten the time of measurement, thus reduce the laser beam drift errors. If all the necessary improvements are properly made, the accuracy of the parallel method for scanning accuracy test can be expected within the order of microradians.

References

[1] D.E. Oliver, Scanning laser vibrometers as tools for vibration measurement and analysis, Test Eng. Manage. (June/July 1991) 18-21.

Page 11: Laser scanning system testing—Errors and improvements

W.X. Li, L.D. Mitchell~Measurement 16 (1995) 9I 101 101

[2] W.X Li and L.D. Mitchell, A test rig for determination of the position characteristics of a galvanometer-based laser scanning system, Proc. IMAC 12, Hawaii, January 1994, Vol. 2, pp. 1211 1217.

[3] J. Montagu, Update of scanner selection, performances and multi-axis configurations, SPIE 1454 (Beam Deflection and Scanning Technologies) (1991) 160-173.

[4] W.X. Li, A precision laser scanning system for experimental modal analysis: Its test and calibration, M.S. Thesis.

Department of Mechanical Engineering, Virginia Polytechnic Institute and State University, Blacksburg, VA, October 1992.

[5] J.C.A. Chaimowicz, Lightwave Technology: An Introduction, Butterworths, London, 1989.

[6] General Scanning Inc.. G3B Series Optical Scanner (Specifications and S y,stem Performance ~, General Scanning Inc., Watertown, MA, 1990.