Liquid based cytologyLiquid based Cytology

Page 2: Laser scanning cytometry and liquid based cytology

Background

• The conventional pap smear has been the most successful screening test

• Screening every 3-5years has resulted in a 70% reduction in incidence

Page 3: Laser scanning cytometry and liquid based cytology

Why LBC (Liquid based cytology) has been introduced ?

• Continuing improvement to the Cervical screening Programme

• Limitations of conventional cytology• Modernisation of the technique• Future benefits

– Extra tests – HPV, chlamydia, Neiseria Gonorrhoea.

Page 4: Laser scanning cytometry and liquid based cytology

Limitations Of Conventional Smear (From UK studies)

i False Negative Rate of up to 55%1

Sampling and interpretive errors

i Ambiguous reports of 6.4%2

70% are truly negative

30% represent more severe abnormality

i Inadequate specimens of 9.7% 2

1. Hutchinson et al., AJCP, Vol 101-2; 215-219, 2. DOH Statistical Bulletin 2000/2001

Page 5: Laser scanning cytometry and liquid based cytology

Sources Of False Negatives

• Sampling issues (70%)

– cells not collected on the sampling device– cells collected, but not transferred to the slide

• Thicker and thinner areas• Nuclear feathering artifact

• Interpretive issues (30%)

– abnormal cells present on slide but either not seen or misinterpreted• Blood/mucus• Air drying artifact

Page 6: Laser scanning cytometry and liquid based cytology

What does 'Liquid Based Cytology' mean?

• Literally it means

“cytology (the study of cells) through a liquid medium.”

Page 7: Laser scanning cytometry and liquid based cytology

Liquid-based cytology techniques

Page 8: Laser scanning cytometry and liquid based cytology

Technique of LBCStep : I Collection of sample

Page 9: Laser scanning cytometry and liquid based cytology
Page 10: Laser scanning cytometry and liquid based cytology
Page 11: Laser scanning cytometry and liquid based cytology

Specimen Collection

• PreservCyt Solution.• Capped, labeled, and sent to the

laboratory equipped with a ThinPrep 2000 Processor .

Composition• Buffered methanol• No active ingredient

Storage• 15 to 30 C for 6 weeks

Page 12: Laser scanning cytometry and liquid based cytology

Thinprep Processor

Cell dispersion

Cell Collection

Cell Transfer

Page 13: Laser scanning cytometry and liquid based cytology

Thin Prep processor (1)Cell dispersion

• Swirling the sampling device in the preservation solution

• Strong enough to separate debris and disperse mucus, but gentle enough to have no adverse effect on cell appearance.

Page 14: Laser scanning cytometry and liquid based cytology

Thin Prep processor(2) Cell Collection

• A gentle vacuum is created within the ThinPrep Pap Test Filter, which collects cells on the exterior surface of the membrane.

• Cell collection is controlled by the ThinPrep 2000 Processor’s software that monitors the rate of flow through the ThinPrep Pap Test Filter.

Page 15: Laser scanning cytometry and liquid based cytology

• After the cells are collected on the membrane, the ThinPrep Pap Test Filter is inverted and gently pressed against the ThinPrep Microscope Slide.

• Natural attraction and slight positive air pressure cause the cells to adhere to the ThinPrep Microscope Slide resulting in an even distribution of cells in a defined circular area.

Thin Prep processor(3) Cell Transfer

Page 16: Laser scanning cytometry and liquid based cytology

Summary

Page 17: Laser scanning cytometry and liquid based cytology

Advantages of LBC

Page 18: Laser scanning cytometry and liquid based cytology
Page 19: Laser scanning cytometry and liquid based cytology
Page 20: Laser scanning cytometry and liquid based cytology
Page 21: Laser scanning cytometry and liquid based cytology

Conventional smear

Page 22: Laser scanning cytometry and liquid based cytology

The prepared slide with the new ThinPrep

Page 23: Laser scanning cytometry and liquid based cytology

What we see under the microscope conventional smear

Page 24: Laser scanning cytometry and liquid based cytology

What we see under the microscope. Notice the clean back ground and how well the cells are dispersed rendering easier to interpretation.

Page 25: Laser scanning cytometry and liquid based cytology
Page 26: Laser scanning cytometry and liquid based cytology
Page 27: Laser scanning cytometry and liquid based cytology
Page 28: Laser scanning cytometry and liquid based cytology

Disadvantages of LBC

Page 29: Laser scanning cytometry and liquid based cytology

Disadvantages of LBC

Page 30: Laser scanning cytometry and liquid based cytology
Page 31: Laser scanning cytometry and liquid based cytology

Manual membrane based method of LBC

• Relatively inexpensive method• Equipments required- vortexer and

laboratory centrifuge

Page 32: Laser scanning cytometry and liquid based cytology

• Solution required are;

Fixative solution• Surepath preservation fluid• Lab prepared- water, NaCl, Na

Citrate, 10% formalin,alcohol

Polymer solution• Agarose, polyethylene glycol• Alcohol, poly-L-lysene

Page 33: Laser scanning cytometry and liquid based cytology

Procedure

Collection of sample

Specimen is vortex mixed

Centrifuge at 800 g for 10 min Vortex mix

Add 1-2 ml of polymer solution

to tube

Decant fixative and blot

excessive fixative

Allow 3-6 drops of suspension to

glass slide

Allow to dry

Stain with conv. Pap

Page 34: Laser scanning cytometry and liquid based cytology

Laser Scanning Cytometry

Page 35: Laser scanning cytometry and liquid based cytology

What is the need?

• Our limited ability to undertake accurate, quantitative measurement of cellular and subcellular factors

• Established technologies in clinical pathology, including conventional microscope-based histopathology and histochemistry, fluorescence microscopy, flow cytometry and computer-based image cytometry, all have limitations.

Page 36: Laser scanning cytometry and liquid based cytology

Introduction

• Imaging + cytometric analysis • Not random, but event-based.• It is closely related to conventional flow

cytometry, which also analyzes individual cells that meet certain characteristics (and is also event-based). Both are therefore cytometric techniques

Page 37: Laser scanning cytometry and liquid based cytology

Limitations of Flow cytometry

1. time-resolved events such as enzyme kinetics cannot be analyzed.

2. Simultaneous study of Morphology of the measured cell is not possible.

3. Cell analysis is at zero spatial resolution.

4. The cell once measured cannot be re-analyzed with another probe(s)

Page 38: Laser scanning cytometry and liquid based cytology

5. Analysis of solid tissue requires cell or nucleus isolation, a procedure that may produce artifacts and loss of the information on tissue architecture.

6.size samples such as fine needle aspirates, spinal fluid, thus, are seldom analyzed by FC.

7. The measured sample cannot be stored for archival preservation.

Page 39: Laser scanning cytometry and liquid based cytology

Introduction

• 2 manufacturers:– CompuCyte Corp. (Cambridge, MA) – Olympus Optical Co. (Tokyo), offers many

advantages of flow cytometry but has no limitations listed above.

• The analytical capabilities of LSC, therefore, complement these of FC, and extend the use of cytometry in many applications

Page 40: Laser scanning cytometry and liquid based cytology
Page 41: Laser scanning cytometry and liquid based cytology

Principle

Page 42: Laser scanning cytometry and liquid based cytology

Lasers

Lasers

Photomultipliers

Page 43: Laser scanning cytometry and liquid based cytology
Page 44: Laser scanning cytometry and liquid based cytology

Interpretation

Page 45: Laser scanning cytometry and liquid based cytology

Thresholding on flow cytometer

• Setting the threshold (or discrimination) on the flow cytometer allows us to eliminate unwanted cells (like erythrocytes) from the saved analysis

Page 46: Laser scanning cytometry and liquid based cytology

Thresholding (or contouring) on the laser scanning cytometer

• On the laser scanning cytometer, any event that is above the threshold (usually DNA fluorescence or sometimes forward scatter) is also considered a “cell” and is also displayed for all parameters.

• The instrument marks such

events with a red “contour”,

which contains the amount

of fluorescent signal above

the threshold.• Thresholding on the LSC is

therefore referred to as

“contouring”

Page 47: Laser scanning cytometry and liquid based cytology

Event-based contouring on the iCys

The red region is the event or thresholding contour. It is analogous to “thresholding” or “discrimination” on the flow cytometer. It defines the minimum signal intensity that defines a cell. Everything else below it is ignored. Like in cytometry, you need a universal parameter (like scatter or DNA luorescence) as the trigger for threshold contouring

The green region is the inte grating contour. You can set this any number of pixels out from the thresholding contour and measure the brightest pixel (Max Pixel), or the total fluorescence (Integral) within the green region.

The blue regions are the background contours. You can also set these any number of pixels out from the integrating contour, away from the cell. The signal between them is interpreted as the autofluorescence background, which is subtracted from the other signals.

Page 48: Laser scanning cytometry and liquid based cytology
Page 49: Laser scanning cytometry and liquid based cytology

Parameters studied by LSC

1. Integrated fluorescence intensity

2. The maximal intensity

3. The Integration area

4. The perimeter of the integration contour (in micrometers).

5. The fluorescence intensity integrated over the area of a torus of desired width defined by the peripheral contour located around (outside) of the primary integration contour.

Page 50: Laser scanning cytometry and liquid based cytology

6. The X-Y coordinates of maximal pixel locating the measured object on of the microscope stage.

7. The computer clock time at the moment of measurement.

Page 51: Laser scanning cytometry and liquid based cytology

Applications in Cytology and Pathology

Page 52: Laser scanning cytometry and liquid based cytology

In Cytology

• It negates disadvantages of

Requires sufficient amount

Verification /reproducibility not possible

Not for adherent cells

Flow cytmetry

Experience of cytologist needed

Image analysis

Page 53: Laser scanning cytometry and liquid based cytology

• Circulating tumor cells– Quantification monitoring for potential

metastasis/recurrence– Molecular studies therapeutic purposes

Page 54: Laser scanning cytometry and liquid based cytology

Apoptosis studies

• Characterized by certain morphological features e.g. nuclear fragmentation, nuclear condensation

• Methods e.g. annexin V, loss of transmembrane potential in mitochondria, analysis of nuclear fragmentation, alteration of DNA condensation.

Page 55: Laser scanning cytometry and liquid based cytology
Page 56: Laser scanning cytometry and liquid based cytology

Immunophenotyping of leucocyte

• Especially useful when amount of obtained sample is minimum e.g. neonate, critically ill patients

• Upto 5 flurochromes in < 15 microlit of peripheral blood.

• Analysis by 2 methods – Specific Immunostaining e.g. CD45– DNA staining with 7-AAD- difference in

intensity with different leucocytes

Page 57: Laser scanning cytometry and liquid based cytology
Page 58: Laser scanning cytometry and liquid based cytology

Ploidy and DNA index

• Anueploid number characteristic of a malignant cell e.g. Gastric, colon, kidney, head and neck etc.

• Prognostic marker in several tumors• LSC measures amount of DNA in cell

Page 59: Laser scanning cytometry and liquid based cytology

Tumor cells are identified and gated based upon cytokeratin staining.

To verify the morphology of the two populations, cells can be restained with Wright-Giemsa or H & E. Cells from each region can be relocated and visualized by the CompuSort™ process.

Page 60: Laser scanning cytometry and liquid based cytology

Bacterial detection

• detect live and dead bacteria,

• estimate cell numbers, and

• calculate live/dead ratios.

• The methods are easier and more accurate than traditional, manual counts.

propidium iodide (PI), unable to permeate the intact cell membrane of a living cell, but does label dead cells with red fluorescence. SYTO® 16(Molecular Probes) will label the nucleic acids of living cells with green fluorescence. Shown here are E. coli bacteria captured on a membrane and stained with PI and SYTO 16

Page 61: Laser scanning cytometry and liquid based cytology

Disadvantages

• Expensive instrumentation• Exact DNA content of paraffin block is

difficult to determine.

Page 62: Laser scanning cytometry and liquid based cytology