labelqualplan
TRANSCRIPT
-
8/2/2019 LabelQualplan
1/10
Package Label Qualification A Sample Plan
IOPP Medical Device Technical CommitteeLabeling Task Group:
Dave Olson, chair Web LabelMarilyn Heckley Smiths MedicalMark Andersen Boston ScientificGerry Gunderson Quality Tech ServicesBrent Lother Medtronic
-
8/2/2019 LabelQualplan
2/10
Considerations in designing a packagelabel testing/qualification procedure
Face stockStock finish; gloss or matteStock strength; paper or syntheticPrint surface; laser, thermal transfer, inkjet, etc.
-
8/2/2019 LabelQualplan
3/10
Considerations in designing a label testing
/ qualification procedure
Print quality and durabilityPrintability of the designComputer printer selectionAbrasion resistance; preprint and imprintChemical and solvent resistance
-
8/2/2019 LabelQualplan
4/10
Considerations in designing a label testing
/ qualification procedureAdhesive properties
Adhesive composition, e.g. hot melt, emulsionacrylic, etc.Adhesive initial tack and ultimate bondSubstrates adhered to and their shapeApplication temperature
Operating temperatureSterilization environment
-
8/2/2019 LabelQualplan
5/10
Considerations in designing a label testing
/ qualification procedure
Environmental conditionsPackage compositionShipping method and conditionsStorage conditions and length of storageChemical resistance
-
8/2/2019 LabelQualplan
6/10
A Sample Testing/Qualification Plan
Adhere samples to appropriate substrates at
an accepted sample size.Peel tests after 24-72 hours
Temperature and humidity conditioningVisual inspection and peel testsAbrasion testingSterilization and final package testsShake, rattle, and roll tests
-
8/2/2019 LabelQualplan
7/10
Common tests and standards
ASTM D3330, Peel adhesion of PS material
ASTM D5264, Sutherland abrasion and smudgeresistance testASTM F1319, Crockmeter abrasion and smudgeresistance testASTM F2252, Ink adhesion tape testASTM F 2250, Chemical exposure, inks & coatings
ASTM D4169, Distribution testing, Shake, rattle, & roll.ASTM F1980, Accelerated aging
-
8/2/2019 LabelQualplan
8/10
Some common label adhesive
conditioning cyclesLow temp cycle, 48hrs, -30F, ambient RHHumid cycle, 48hrs, 90F, 85% RHHigh temp cycle, 48 hrs, 130F, 25% RH
-
8/2/2019 LabelQualplan
9/10
Sample label stock 90 degree peel values on
Tyvek lid stock, lbs/linear inch
Label stock 20 minutes 24 hoursDevice Co A .7 .95Device Co B .74 .98
Device Co C 1.5 1.8Device Co D 1.2 fiber tear (label)Device Co E 1.2 1.6
-
8/2/2019 LabelQualplan
10/10
Summary and conclusions
There are no standardized protocols for
qualifying package labels.Design your protocol based on customerrequirements, your process, package design,and environmental conditions.Develop rationale for sample size and test
method.Develop your pass/fail protocols beforetesting begins.