lab7 final project egrb 307
TRANSCRIPT
![Page 1: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/1.jpg)
BIOMEDICAL ENGINEERING
VCU School of Engineering
EGRB 307: BIOMEDICAL INSTRUMENTATION
Laboratory No./ Date #7. 11/12/2015Experiment Title Final Project
Name Khade GrantLab Partner’s Name Sarah Tracy, Kelsey Hideshima
Honor Pledge:
"On my honor, I have neither given nor received unauthorized aid on this assignment. Both authors have contributed equally to this work"
[Sign here]
For Official Use Only
Comments: Grade / Score:
1
![Page 2: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/2.jpg)
I. INTRODUCTION
The purpose of this lab was to design an ECG circuit and use it to measure the ECG signal from the members of the lab group. Furthermore, the signal was to be amplified to a sufficient level for display and analysis. Through the human ECG signal, the average heart rate in beats per minute (bpm) can be calculated. The lab used the principles explored in previous labs, such as high-pass, and low-pass filters, voltage dividers, and amplifiers.
II. METHODS
Part 1
The purpose of this section was to determine the components of the main amplifier using the equations for the low and high cutoff frequencies, as well as the equation for gain. The main amplifier was designed with a passive high-pass filter and an active low-pass filter. It was designed to have a gain of 70. A low-pass filter attenuates high frequency signals above a cutoff frequency but does not attenuate the frequency response of low-frequency signals below the cutoff frequency. A high-pass filter attenuates low frequency signals below a cutoff frequency but does not attenuate the frequency response of high-frequency signals above the cutoff frequency. The circuit was designed to have a low corner frequency of 0.5 Hz and a high-corner frequency of 100 Hz.
The R1 resistor was calculated with the low corner frequency using the equation: fc1 = 1/(2πR1C1).
The feedback (R3) resistor was calculated with the high corner frequency using the equation: fc2 = 1/(2πR3C2).
The R2 resistor was calculated with a theoretical gain of 70, and using: G = (R3/R2) + 1.
The bias resistor (R4) was calculated using: R4 = R1 – (R2R3)/(R2+R3) ≈ R1.
Part 2
The purpose of this part was to build the main amplifier and test its features. Once the values of the components were determined from part 1, the main amplifier was designed on the breadboard using an AD 741 op amp. (Figure 1.)
2
![Page 3: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/3.jpg)
Figure 1. Main Amplifier
The power source was used to supply +4.5V and -4.5V for the circuit. The gain was determined by inputting a signal of 10 Hz with an amplitude of 100mVPP, measuring the output voltage, and using: G = Vo/Vi.
Next, the frequency response for the high-pass filter was determined by slowly decreasing the frequency from 10 Hz to 0.5 Hz and monitoring the output voltage and gain at various intermediate frequencies. The low corner frequency was determined from the resulting frequency response. The frequency response for the low-pass filter was determined by slowly increasing the frequency from 10 Hz to 100 Hz and monitoring the output voltage at various intermediate frequencies. The high corner frequency was determined from the resulting frequency response.
Part 3
The purpose of this part was to determine the components of the pre-amplifier. The RG resistor was calculated with a theoretical gain of 12 and using the equation:
G = (49.4kΩ/RG) + 1.
Both Ra resistors were selected to be 33kΩ. The R0 resistor was determined by using: RG = 2Ra || R0.
Part 4
The purpose of this part was to build the pre-amplifier and test its features. Once the values of the components were determined from part 3, the pre-amplifier was designed on the breadboard using an AD 620 op amp. (Figure 2.)
3
![Page 4: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/4.jpg)
Part 5
The purpose of this part was to construct the entire ECG circuit using both the pre-amplifier and the main amplifier. A 100:1 voltage divider was built to reduce the input signal so that a smaller signal could be input into the main amplifier. This way, the output voltage from the main amplifier would not be too high. The reduction ratio was calculated using Vi/Vo:1.
Next, a 100mVpp input signal was input to the full amplifier, and the output voltage was measured. The gain of the entire ECG amplifier was calculated using: G = Vo/Vi-main amp. The reduction ratio was calculated using
After this, the frequency response of the entire ECG amplifier was recorded. The frequency response for the high-pass filter was determined by slowly decreasing the frequency from 10 Hz to 0.5 Hz and monitoring the output voltage and gain at various intermediate frequencies. The low corner frequency was determined from the resulting frequency response. The frequency response for the low-pass filter was determined by slowly increasing the frequency from 10 Hz to 100 Hz and monitoring the output voltage at various intermediate frequencies. The high corner frequency was determined from the resulting frequency response.
Lastly, a 1 Hz cardiac signal was input to the full amplifier with varying input amplitudes of 100mVpp, 200mVpp, and 300mVpp. The output voltage, gain, and calibration parameter from each input signal was measured or determined. The gain was calculated using: G = Vo/Vi-main amp. The calibration parameter was determined using: Calibration parameter = Vo-main-amp/Vi-volt. div.
Part 6
4
Vi (+)
Vo
Vi (-)
AD620
+4.5V
-4.5V
1
2
3
4
7
8
6R
_
+
G
Ref
5
Figure 2. Pre-Amplifier
![Page 5: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/5.jpg)
The purpose of this part was to construct a 9V battery power supply. The battery was used to replace the electric power supply during the human experiment. The power supply consisted of the 9V battery, a 4.5V zener diode, 1N5229B, and a resistor Rp of 460 Ω. The power supply circuit provided the full amplifier with a +4.5V, -4.5V supply. The following battery power supply was built on the breadboard. (Figure 3.)
Part 7
The purpose of this part was to build a driven-right-leg circuit using an AD741 op amp, and add it to the full amplifier. The R4 and R5 resistor values were selected to be equivalent to each other and equal to 5.4 MΩ. The full ECG amplifier with the driven-right-leg circuit is shown below in figure 4.
5
+
-
9V Battery
+4.5 V
-4.5 V
0V or ground
Rp
D
1N5229B
Figure 3. Battery Power Supply
![Page 6: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/6.jpg)
Figure 4. Full ECG amplifier with driven-right-leg circuit
Part 8
The purpose of this part was measure the human ECG signal and determine the heart rate. First, a 1 Hz cardiac signal was input to the full amplifier with varying input amplitudes of 100mVpp, 200mVpp, and 300mVpp. The output voltage, gain, and calibration parameter from each input signal was measured or determined. The gain was calculated using: G = Vo/Vi-main amp. The calibration parameter was determined using: Calibration parameter = Vo-main-amp/Vi-volt. div.
Next, three clamps were used for the electrodes. The cardiac signal of each of the three lab members were evaluated. One electrode was placed on the right hand (RR), another was placed on the left hand (LR), and the third one was placed on the left leg (LL). Using the human ECG signal as the input signal, the full ECG amplifier circuit amplified the input so that the signal could be observed. The ECG output signal was observed for 20 seconds in order to find any possible irregularities. The output ECG signal amplitude was measured.
Next, four successive time intervals for the resulting heart beats were measured and the variability was analyzed. The average heart in bpm, as well as the maximum and minimum heart rates were also determined from the signal. Lastly, the ECG signal was observed after the driven-right-leg circuit was disconnected.
6
![Page 7: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/7.jpg)
III. RESULTS
Main Amplifier
Component Theoretical/Calculated Measured %Difference/%Error
C1 1µF 1.016 µF 1.60%
R1 318.31 kΩ 313.23 kΩ 1.60%
C2 0.01 µF 0.0099 µF 1%
R2 2.3 kΩ 2.27 kΩ 1.30%
R3 159 kΩ 158 kΩ 0.19%
R4 313 kΩ 312.8 kΩ 0.06%
Table 1. Components of Main Amplifier
All Components were within 5% tolerance.
10 Hz Signal
Name Theoretical/Calculated Measured %Error
Vi 100 mV 105mV 5%
Vo 7.44 V 7.12 V 4.30%
Gain 70.86 67.81 4.30%
Low fc 0.53 Hz 0.48 Hz 9.43%
High fc 100.29 Hz 92 Hz 8.27%
Table 2. Input, Output, and low and high corner frequencies of the Main Amplifier
Low-Pass Filter: Vi = 105mVpp 7
![Page 8: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/8.jpg)
f(hz) Vo Gain
10 7.12 67.80952
20 6.96 66.28571
30 6.8 64.7619
40 6.48 61.71429
50 6.24 59.42857
60 5.84 55.61905
70 5.6 53.33333
80 5.28 50.28571
90 5.12 48.7619
92 5.04 48
94 4.96 47.2381
96 4.88 46.47619
98 4.8 45.71429
100 4.72 44.95238
102 4.6 43.80952
104 4.6 43.80952
106 4.56 43.42857
108 4.52 43.04762
110 4.44 42.28571Table 3. Frequency response of Low-Pass Filter for Main-Amplifier
8
![Page 9: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/9.jpg)
Figure 5. Frequency response of low-pass filter
High-Pass Filter: Vi = 105mVpp
Frequency (hz) Vo (V) Gain10 7.12 67.80952381
9 7.12 67.80952381
8 7.12 67.80952381
7 7.12 67.80952381
6 7.12 67.80952381
5 7.12 67.80952381
4 7.12 67.80952381
3 7.04 67.04761905
2 6.96 66.28571429
1.75 6.88 65.52380952
1.5 6.88 65.52380952
1 6.4 60.95238095
9
![Page 10: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/10.jpg)
0.8 6.16 58.66666667
0.6 5.52 52.57142857
0.5 5.12 48.76190476
0.53 5.2 49.52380952
0.48 5.04 48
0.4 4.48 42.66666667
Table 4. Frequency response of High-Pass Filter for Main-Amplifier
Figure 6. Frequency response of high-pass filter
10
![Page 11: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/11.jpg)
Pre-Amplifier
Component Theoretical Measured Percent Error
RG 4.491 kΩ 4.485 kΩ 0.13%
Ra-1 33 kΩ 33.02 kΩ 0.06%
Ra-2 33 kΩ 32.79 kΩ 0.64%
2Ra 66 kΩ 65.81 kΩ 0.29%
Ro 4.813 kΩ 4.741 kΩ 1.50%
Gain 12 12.01 0.083%Table 5. Components of Pre-Amplifier
All Components were within 5% tolerance.
10 Hz Signal
Name Theoretical Measured Percent Error
Vi 100 mV 104 mV 4%
Vo 1.27 V 1.22 V 3.94%
Gain 12.21 11.73 3.93%Table 6. Waveform measurements for 10 Hz sinusoidal signal
Signal Vi (%error) Vo (%error) Gain (%error)
0.5 Hz 104 mV (4%) 1.22 V (3.94%) 11.73 (3.93%)
10 Hz 104 mV (4%) 1.22 V (3.94%) 11.73 (3.93%)
20 Hz 104 mV (4%) 1.22 V (3.94%) 11.73 (3.93%)
100 Hz 104 mV (4%) 1.22 V (3.94%) 11.73 (3.93%)
200 Hz 104 mV (4%) 1.22 V (3.94%) 11.73 (3.93%)Table 7. Waveform measurements for sinusoidal signals at varying frequencies
11
![Page 12: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/12.jpg)
Cardiac Signal
Name Theoretical Measured Percent Error
Vi 100 mV 108 mV 8%
Vo 1.267 V 1.22 V 3.71%
Gain 11.73 11.30 3.67%Table 8. Waveform measurements for cardiac signal
Figure 7. Output of Cardiac Signal with 1 Hz and 100 mVpp
Full Amplifier
Voltage Divider
Name Theoretical Measured Percent Error
R100k 100 kΩ 101.96 kΩ 1.96%
R10M 10 MΩ 9.98 MΩ 0.2%
Vin 10V 10.4V 4%
Vout 0.1052V 0.108V 2.78%
Reduction Ratio 98.86:1 96.30:1 2.59%Table 9. Components and Measurements of Voltage Divider
12
![Page 13: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/13.jpg)
Vin-voltage divider = 112mV => Vout-voltage divider = 1.16 mV = Vin-main
Vin = 1.16 mV
Name Theoretical Measured Percent Error
Vout 923 mV 976 mV 5.74%
Gain 795.4 841.4 5.78%Table 10. Output signal and gain of voltage divider
Low-Pass Filter
Frequency (hz) Vo (V) Gain
10 0.976 841.3793
20 0.932 803.4483
30 0.88 758.6207
40 0.88 758.6207
50 0.86 741.3793
60 0.86 741.3793
70 0.82 706.8966
80 0.8 689.6552
90 0.74 637.931
92 0.74 637.931
94 0.74 637.931
96 0.72 620.6897
98 0.72 620.6897
100 0.72 620.6897
102 0.712 613.7931
104 0.704 606.8966
106 0.696 600
108 0.68 586.2069
110 0.68 586.2069Table 11. Frequency response of low-pass filter for full amplifier
13
![Page 14: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/14.jpg)
Figure 8. Frequency response of low-pass filter for Full Amplifier
High-Pass Filter
Frequency (hz) Vo (V) Gain10 0.976 841.37939 1 862.0698 1.01 870.68977 1.01 870.68976 1 862.0695 1 862.0694 0.992 855.17243 0.984 848.27592 0.984 848.2759
1.75 0.976 841.37931.5 0.96 827.5862
1 0.92 793.10340.8 0.88 758.62070.6 0.816 703.44830.5 0.768 662.069
0.53 0.776 668.96550.48 0.744 641.37930.46 0.736 634.48280.44 0.712 613.79310.4 0.688 593.1034
Table 12. Frequency response of high-pass filter for full amplifier14
![Page 15: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/15.jpg)
Figure 9. Frequency response of high-pass filter for Full Amplifier
Low fc 0.48 Hz 0.44 Hz 8.33%
High fc 92 Hz 106 Hz 15.2%
Table 13. High and low corner frequencies of Full Amplifier
1 Hz Signal; Vi = 100mV
Name Theoretical Measured Percent Error
Vi 100 mV 109 mV 9%
Vo 952 mV 920 mV 3.36%Calibration Parameter (mV/mV) 8.734 8.440 3.36%
Table 14. Waveform measurements and Calibration Parameter for 1 Hz Cardiac input signal of 109mV
15
![Page 16: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/16.jpg)
1 Hz Signal; Vi = 200mV
Name Theoretical Measured Percent Error
Vi 200 mV 210 mV 5%
Vo 1835 mV 1700 mV 7.36%Calibration Parameter (mV/mV) 8.738 8.095 7.36%
Table 15. Waveform measurements and Calibration Parameter for 1 Hz Cardiac input signal of 210mV
1 Hz Signal; Vi = 300mV
Name Theoretical Measured Percent Error
Vi 300 mV 312 mV 4%
Vo 2726 mV 2480 mV 9.02%Calibration Parameter (mV/mV) 8.737 7.949 9.02%
Table 16. Waveform measurements and Calibration Parameter for 1 Hz Cardiac input signal 0f 312mV
Average Calibration Parameter: 8.161 mV/mV.
Power Supply
Component Theoretical ActualPercent Error
Rp 460Ω 454.41Ω 1.22%
Battery 9V 8.73V 3%Table 17. Components of Power Supply
16
![Page 17: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/17.jpg)
Driven-Right-Leg Circuit
Component Theoretical ActualPercent Error
R4 5.4 5.443 0.79%
R5 5.4 5.455 1.01%Table 18. Components of Driven-Right-Leg Circuit
Human ECG Signal
1 Hz Signal; Vi = 100mV
Name Theoretical Measured Percent Error
Vi 100 mV 110 mV 10%
Vo 961 mV 960 mV 0.10%
Gain 841.4 840.4 0.12%Calibration Parameter (mV/mV) 8.736 8.727 0.10%
Table 19. Waveform measurements and Calibration Parameter for 1 Hz sinusoidal input signal of 110mV with Driven-Right-Leg Circuit
1 Hz Signal; Vi = 200mV
Name Theoretical Measured Percent Error
Vi 200 mV 210 mV 5%
Vo 1835 mV 1800 mV 1.91%
Gain 841.4 825.4 1.91%Calibration Parameter (mV/mV) 8.738 8.571 1.91%
Table 20. Waveform measurements and Calibration Parameter for 1 Hz sinusoidal input signal of 210mV with Driven-Right-Leg Circuit
17
![Page 18: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/18.jpg)
1 Hz Signal; Vi = 300mV
Name Theoretical Measured Percent Error
Vi 300 mV 310 mV 4%
Vo 2709 mV 2640 mV 2.54%
Gain 841.4 820.1 2.54%Calibration Parameter (mV/mV) 8.739 8.516 2.54%
Table 21. Waveform measurements and Calibration Parameter for 1 Hz sinusoidal input signal of 310mV with Driven-Right-Leg Circuit
Average Gain: 828.6 (1.52% error from 841.4)
Average Calibration Parameter: 8.605 mV/mV.
Figure 10. ECG signal for Khade Grant
Vo = 1.86V
Interval #1 Interval #2 Interval #3 Interval #4
720 ms 840 ms 680 ms 660 msTable 22. Heartbeat intervals for Khade Grant
Average heart rate: 82.8 bpm
Maximum heart rate: 90.9 bpm
18
![Page 19: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/19.jpg)
Minimum heart rate: 71.4 bpm
Figure 11. ECG signal for Sarah Tracy
Vo = 0.508V
Interval #1 Interval #2 Interval #3 Interval #4
780 ms 860 ms 840 ms 780 msTable 23. Heartbeat intervals for Sarah Tracy
Average heart rate: 73.6 bpm
Maximum heart rate: 76.9 bpm
Minimum heart rate: 69.8 bpm
19
![Page 20: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/20.jpg)
Figure 12. ECG signal for Kelsey Hideshima
Vo = 0.720V
Interval #1 Interval #2 Interval #3 Interval #4
664 ms 680 ms 680 ms 660 msTable 24. Heartbeat intervals for Kelsey Hideshima
Average heart rate: 89.4 bpm
Maximum heart rate: 90.9 bpm
Minimum heart rate: 88.2 bpm
Figure 13. ECG signal for Khade Grant w/ disconnected driven-right-leg circuit
20
![Page 21: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/21.jpg)
IV. DISCUSSION
An electrocardiogram (ECG) is a device used to record the electrical activity of the heart. It uses electrodes placed on the patient’s body and obtains measurements over a period of time. Depolarization of the heart with each beat induces small electrical changes on the skin. The ECG electrodes detect these waves which are then amplified into an observable signal.
Overall, the circuit receives an input signal, which is then amplified into an observable output signal using three amplifiers. The circuit was designed with an experimental gain of 828.6. It consisted of two stages. The first stage, consisting of a pre-amplifier and a driven-right leg circuit was designed to produce a gain of 12. The second stage, consisting of the main amplifier was designed to obtain a gain of 70. The full amplifier was therefore designed to obtain a gain with a value over 800.
The pre-amplifier took the initial waveform as the input signal and amplified it. It then sent the amplified signal as the input signal for the main amplifier. The driven-right leg circuit provided the input from the leg on the electrode, whereas the pre-amplifier provided the input from the electrode on the right arm and the left arm. The main amplifier then took the output signal from the pre-amplifier and used this as its input signal. It amplified this signal to produce the final output cardiac waveform.
All the components used to design the main amplifier were well within a 5% tolerance range. The main amplifier produced a gain of 67.81. This produced a 4.30% error from the theoretical gain value of 70.86. The main amplifier had a low corner frequency, low fc, of 0.48 Hz. This produced a 9.43% error from the theoretical low corner frequency of 0.53 Hz. It had a high corner frequency, high fc, of 92 Hz. This produced an 8.27% error from the theoretical high corner frequency of 100.29 Hz. The main amplifier functioned as expected; there were no major problems encountered in the design of the main amplifier.
All the components used to design the pre-amplifier were well within a 5% tolerance range. The pre-amplifier produced a gain of 11.73. This produced a 3.93% error from the theoretical gain value of 12.21. As the frequency of the input signal for the pre-amplifier was varied from 10 Hz to 200 Hz, the output signal of the pre-amplifier remained at a constant value of 1.22V. This showed the robustness of the pre-amplifier. Since heart signals can vary in frequency the pre-amplifier could, therefore, be used to accurately measure signal amplitude regardless of frequency. The pre-amplifier functioned as expected; there were no major problems encountered in the design of the pre-amplifier.
All the components used to design the full amplifier were within a 5% tolerance range. The full amplifier produced an average gain of 828.6. This produced a 1.52% error from the theoretical gain value of 841.4. The full amplifier had a low corner frequency, fc, of 0.44 Hz. This produced an 8.33% error from the theoretical low corner frequency of 0.48 Hz. It had a high corner frequency, fc, of 106 Hz. This
21
![Page 22: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/22.jpg)
produced a 15.2% error from the theoretical high corner frequency of 92 Hz. The frequency response was from 0.44 Hz to 106 Hz. It also had an average calibration parameter of 8.605 mV/mV. The full amplifier functioned as expected; there were no major problems encountered in the design of the full amplifier.
The purpose of the voltage divider was to test the gain of the circuit using a small input signal. The input signal needed to be small enough so that the full amplifier did not amplify the signal out of the range that the oscilloscope could detect. The function generator could not input signals small enough, so a 100:1 voltage divider had to be designed to divide the voltage from the function generator and input the small signal to the full amplifier. The voltage divider had a reduction ratio of 96:30:1. This produced a 2.59% error from the theoretical reduction ratio of 98.86:1. The voltage divider functioned as expected; there were no major problems encountered in the design of the voltage divider.
Three ECG electrodes were used in this experiment. One was placed on the right hand, another was placed on the left arm, and the third was placed on the left leg. These produced three leads or vectors along which the relative depolarization of the heart could be measured. The first lead, Lead I, goes from the left arm (LA) to the right arm (RA). The second lead, Lead II, goes from the left leg (LL) to RA. The third lead, Lead III, goes from LL to LA. The depolarization along each lead produces a resultant vector that represents the depolarization of the heart. The black lead was attached to the left arm and was input into pin 3 of the pre-amplifier (AD620). The white lead was attached to the right arm and was input into pin 2 of the pre-amplifier (AD620). The green lead was attached to the left leg and was input between the two Ra resistors of the pre-amplifier (AD620).
The heart rate of normal healthy humans is between 60 and 100 beats per minute (bpm). For Khade Grant, the average heart rate obtained was 82.8 bpm. For Sarah Tracy, the average heart rate obtained was 73.6 bpm. For Kelsey Hideshima the average heart rate obtained was 89.4 bpm. The average heart rates of all three lab members fell within the range of normal healthy heart rates. The heart rate for Khade Grant had the most variability in terms of beat-to-beat intervals. His heart beat ranged from 660 ms to 840ms. The heart rate for Kelsey Hideshima had the least amount of variability in terms of beat-to-beat intervals. Her heart beat ranged from 660 ms to 680ms. The ECG signal of all three lab members was enlarged to over 500 mV. The ECG amplitudes in normal healthy hearts range from 0.5mV to around 2.5 mV.
The calculations for each part are shown on Appendix I.
V. CONCLUSION
22
![Page 23: Lab7 Final Project EGRB 307](https://reader035.vdocuments.us/reader035/viewer/2022062302/587897e91a28ab375f8b6e85/html5/thumbnails/23.jpg)
Overall, the lab went as expected. The lab provided a sufficient understanding of how ECGs worked and how to design one. Specifically, the lab provided knowledge of how to design an ECG circuit integrating various components such as a main amplifier, pre-amplifier, and driven-right-leg circuit. When problems are encountered, it is important to recheck the circuit to see if everything is wired correctly, and reread the instructions to see if any crucial steps were omitted. The main amplifier, pre-amplifier, driven-right leg circuit, and other components of the full amplifier worked as expected and provided sufficient results. Furthermore, the full amplifier was designed without any major problems and the results obtained met the requirements that were stated in the lab. Specifically, the full amplifier had a gain over 800, a frequency response approximately between 0.5 Hz and 100 Hz, and was able to acquire the ECG signals from all three members of the lab and enlarge their output signals over 500 mV.
23