kit de cloro lamotte

Upload: marlon2370

Post on 07-Apr-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 Kit de Cloro LaMotte

    1/4

    DPD CHLORINE TEST KIT

    MODEL LP-1-NMC CODE 6901-NMC

    QUANTITY CONTENTS CODE

    50 *Chlorine DPD #1R Tablets (*6999A) FMC6905A-6999ABOX

    50 Chlorine DPD #3R Tablets (*6905A)

    50 *Chlorine DPD #2R Tablets (6904A) FMC6904A-6906ABOX

    50 *Chlorine DPD #4 Instrument Grade Tablets(*6906A)

    3 Test Tubes, plastic, w/caps 0106

    1 Chlorine DPD Comparator, 0.1-1.0 ppm 6978

    *WARNING: Reagents marked with an * are considered to be potential health

    hazards. To view or print a Material Safety Data Sheet (MSDS) for thesereagents see MSDS CD or www.lamotte.com. To obtain a printed copy, contactLamotte by e-mail, phone or fax.

    To order individual reagents or test kit components, use the specified codenumber.

    USE OF THE OCTET COMPARATOR

    The Octet Comparator contains eight permanent color standards. A testsample is inserted into the openings in the top ofthe comparator. The sample can then be

    compared to four color standards at once, and thevalue read off the comparator. For optimum colorcomparison, the comparator should be positionedbetween the operator and a light source, so thatthe light enters through the special light-diffusingscreen in the back of the comparator. Avoidviewing the comparator against direct sunlight oran irregularly lighted background.

    L Mott

    L Mott

  • 8/3/2019 Kit de Cloro LaMotte

    2/4

    PROCEDURE

    FREE AVAILABLE CHLORINE1. Rinse the test tube (0106) with sample water. Fill to 10 mL line with

    sample water.

    2. Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix untiltablet disintegrates.

    3. Immediately insert test tube into Chlorine DPD Comparator (6978).Match sample color to a color standard. Record as Reading #1. Donot discard this sample if testing for Monochloramine, Dichloramineor Total Residual Chlorine.

    NOTE: Reading should be made within one minute of addingreagent.

    Reading #1 = ppm Free Available Chlorine

    MONOCHLORAMINE

    4. Add one Chlorine DPD #2R Tablet (6904A) to sample from Step 3.Cap and mix until tablet disintegrates.

    5. Insert test tube into Chlorine DPD Comparator (6978). Match

    sample color to a color standard. Record as Reading #2.Reading #2 - Reading #1 = ppm Monochloramine

    DICHLORAMINE AND TOTAL RESIDUAL CHLORINE

    6. Add one *Chlorine DPD #3R Tablet (6905A) to sample from Step 5.Cap and mix until tablet disintegrates.

    7. Insert test tube into Chlorine DPD Comparator (6978). Matchsample color to a color standard. Record as Reading #3.

    Reading #3 - Reading #2 = ppm Dichloramine

    Reading #3 = ppm Total Residual Chlorine

  • 8/3/2019 Kit de Cloro LaMotte

    3/4

    TOTAL RESIDUAL CHLORINE AND COMBINED CHLORINE

    1. Rinse test tube (0106) with sample water. Fill to 10 mL line withsample water.

    2. Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix untiltablet disintegrates.

    3. Immediately insert test tube into Chlorine DPD Comparator (6978).Match sample color to a color standard. Record as ppm Free

    Available Chlorine.

    4. Add one *Chlorine DPD #3R Tablet (6905A). Cap and mix untiltablet disintegrates.

    5. Insert test tube into Chlorine DPD Comparator (6978). Matchsample color to a color standard. Record as ppm Total ResidualChlorine.

    Combined Chlorine, ppm =Total Residual Chlorine - Free Available Chlorine

    TOTAL RESIDUAL CHLORINE

    *Chlorine DPD #4 Tablet determines Total Residual Chlorine and is usedwhere a differentiation in the type of chlorine is not required.1. Rinse test tube (0106) with sample water. Fill to 10 mL line with

    sample water.

    2. Add one *Chlorine DPD #4 Instrument Grade Tablet (6906A). Capand mix until tablet disintegrates.

    3. Insert test tube into Chlorine DPD Comparator (6978). Matchsample color to a color standard. Record as ppm Total ResidualChlorine.

  • 8/3/2019 Kit de Cloro LaMotte

    4/4

    LaMOTTE COMPANY

    Helping People Solve Analytical Challenges

    PO Box 329 Chestertown Maryland 21620 USA

    800-344-3100 410-778-3100 (Outside U.S.A.) Fax 410-778-6394Visit us on the web at www.lamotte.com

    05.08