king s research portal - core a thorax overlay, and a large female by adding breasts on the thorax...
TRANSCRIPT
King’s Research Portal
DOI:10.1007/s12350-015-0323-0
Document VersionPublisher's PDF, also known as Version of record
Link to publication record in King's Research Portal
Citation for published version (APA):Polycarpou, I., Chrysanthou-Baustert, I., Demetriadou, O., Parpottas, Y., Panagidis, C., Marsden, P. K., &Livieratos, L. (2015). Impact of respiratory motion correction on SPECT myocardial perfusion imaging using amechanically moving phantom assembly with variable cardiac defects. Journal of Nuclear Cardiology.10.1007/s12350-015-0323-0
Citing this paperPlease note that where the full-text provided on King's Research Portal is the Author Accepted Manuscript or Post-Print version this maydiffer from the final Published version. If citing, it is advised that you check and use the publisher's definitive version for pagination,volume/issue, and date of publication details. And where the final published version is provided on the Research Portal, if citing you areagain advised to check the publisher's website for any subsequent corrections.
General rightsCopyright and moral rights for the publications made accessible in the Research Portal are retained by the authors and/or other copyrightowners and it is a condition of accessing publications that users recognize and abide by the legal requirements associated with these rights.
•Users may download and print one copy of any publication from the Research Portal for the purpose of private study or research.•You may not further distribute the material or use it for any profit-making activity or commercial gain•You may freely distribute the URL identifying the publication in the Research Portal
Take down policyIf you believe that this document breaches copyright please contact [email protected] providing details, and we will remove access tothe work immediately and investigate your claim.
Download date: 18. Feb. 2017
ORIGINAL ARTICLE
Impact of respiratory motion correction onSPECT myocardial perfusion imaging using amechanically moving phantom assembly withvariable cardiac defects
Irene Polycarpou, PhD,a,b Isabelle Chrysanthou-Baustert, PhD,c
Ourania Demetriadou, MD,d Yiannis Parpottas, PhD,c,e Christoforos Panagidis,
MD,f Paul K. Marsden, PhD,a and Lefteris Livieratos, PhDa,g
a Division of Imaging Sciences and Biomedical Engineering, King’s College London, London,
United Kingdomb Department of Health Sciences, European University Cyprus, Nicosia, Cyrpusc Frederick Research Center, Nicosia, Cyprusd Department of Nuclear Medicine, Limassol General Hospital, Limassol, Cypruse General Department (Physics-Mathematics), Frederick University, Nicosia, Cyprusf Department of Nuclear Medicine, Nicosia General Hospital, Nicosia, Cyprusg Guy’s and St Thomas’ Hospitals NHS Foundation Trust, London, United Kingdom
Received Jul 27, 2015; accepted Oct 26, 2015
doi:10.1007/s12350-015-0323-0
Background. The aim of this study was to determine the impact of respiratory motioncorrection on SPECT MPI and on defect detection using a phantom assembly.
Methods. SPECT/CT data were acquired using an anthropomorphic phantom withinflatable lungs and with an ECG beating and moving cardiac compartment. The heart motionfollowed the respiratory pattern in the cranio-caudal direction to simulate normal or deepbreathing. Small or large transmural defects were inserted into the myocardial wall of the leftventricle. SPECT/CT images were acquired for each of the four respiratory phases, from exhaleto inhale. A respiratory motion correction was applied using an image-based method withtransformation parameters derived from the SPECT data by a non-rigid registration algo-rithm. A report on defect detection from two physicians and a quantitative analysis on MPIdata were performed before and after applying motion correction.
Results. Respiratory motion correction eliminated artifacts present in the images, resultingin a uniform uptake and reduction of motion blurring, especially in the inferior and anteriorregions of the LV myocardial walls. The physicians’ report after motion correction showed thatimages were corrected for motion.
Conclusions. A combination of motion correction with attenuation correction reducesartifacts in SPECT MPI. AC-SPECT images with and without motion correction should besimultaneously inspected to report on small defects. (J Nucl Cardiol 2015)
Key Words: SPECT/CT Æ myocardial perfusion defects Æ motion correction Æ cranio-caudalcardiac motion
Reprint requests: Irene Polycarpou, PhD, Division of Imaging Sciences
and Biomedical Engineering, King’s College London, London,
United Kingdom; [email protected]
1071-3581/$34.00
Copyright � 2015 American Society of Nuclear Cardiology.
AbbreviationsSPECT Single-photon emission computed
tomography
CT Computed tomography
AC Attenuation correction
MPI Myocardial perfusion imaging
PLC Programmable logic controller
MC Motion correction
LV Left ventricle
INTRODUCTION
In clinical SPECT studies, respiratory motion limits
the performance of MPI studies. Heart motion induced
by respiration affects SPECT acquisition in two ways.1
First, spatial misalignment between sequentially
acquired SPECT and CT may affect attenuation correc-
tion of the emission data since SPECT images are
acquired for substantially longer time than the CT
images. Various studies have shown that such a
misalignment may cause image artifacts and under- or
over-estimation of tracer activity.2 Furthermore, respi-
ration may lead to image blurring along the direction of
motion, primarily in the cranio-caudal direction, due to
the fact that SPECT data are averaged over many
respiratory cycles. This can result in misleading dis-
placement of uptake affecting the perceived regional
localization in the myocardial walls.3,4 Motion-induced
artifacts can be a source of false-positive findings and
may be interpreted as ischemia affecting the diagnosis of
coronary artery disease.5
A non-uniform blurring of MPI due to motion
induced by respiration4 may affect the ability to detect
defects.6 The respiratory motion of the heart has been
previously investigated using computational4 and phys-
ical phantoms.3,8 It was simulated on physical phantoms
with a static cardiac compartment8 being on an oscil-
lating plate or with static cardiac compartments3,9,10
inserted within torso phantoms and then being on
oscillating plates. The effects of respiratory motion on
MPI non-uniformity and on the resulting false-positive
rates have also been reported.3,4 A common method to
compensate for motion effects is data gating which
divides the motion cycle into a number of gates based on
either time or amplitude.7,11 Further, the respiratory
motion of the heart on physical phantoms8,10 was
corrected using motion correction algorithms. However,
limited studies, only with computational phantoms6
using the channelized Hotelling observer algorithm,
reported on the effects of respiratory motion on myocar-
dial defect detection.
In this study, the impact of respiratory motion
correction on MPI and on defect detection was inves-
tigated for heart motion amplitudes of normal and deep
breathing. An anthropomorphic phantom, with inflat-
able lungs, and an ECG beating and moving cardiac
compartment, was used to simulate the heart respiration
motion. Motion correction was applied using an image-
based method with transformation parameters derived
from the SPECT data by a non-rigid registration
algorithm. SPECT/CT images were acquired for each
of the four respiratory gates, from exhale to inhale, in
normal and deep breathing modes of the phantom
assembly. Defects could be inserted within the myocar-
dial wall of the LV. Two physicians reported on defect
present or absent before and after correcting for motion
of the AC-SPECT images. A comparison of the quan-
titative analysis and physicians’ reports before and after
motion correction is presented.
MATERIALS AND METHODS
Phantom Assembly
A cardiac phantom of an ECG beating and moving left
ventricle with variable defects, and a respiratory phantom of
inflatable lungs, constructed at Frederick Research Center,
were inserted into an anthropomorphic thorax phantom12 and
used to simulate the heart motion during respiration using
routine imaging conditions.
The cardiac phantom consisted of an inner–outer mem-
brane system where the inner membrane represented an LV and
the closed cavity within the inner and outer membranes
represented the myocardial wall of the LV. Each 2-mm thick
ellipsoidal membrane was made of transparent silicone with a
density of 1.06 g/cm2. The end systolic volume was 42 mL and
the LV myocardial wall volume was 90 mL. Diastole and
systole could be achieved by pumping water into and out of the
inner membrane using a stepper motor controlled by the PLC
and a piston. Since the myocardial wall cavity was filled with
water which is non-compressive, the pressure was transferred
from the inner to the outer membrane (Figure 1A). The beating
rate (60 beats/minute) and the ejection fraction (45%) followed
the Wigger diagram. A 5% deviation was found using the
Xeleris Myovation (GE Healthcare) compared to the set
ejection fraction. Radionuclide could be injected within the
myocardial wall cavity. Defects could be glued on the outer part
of the inner membrane (Figure 2A). The ECG signal from a
simulator was simultaneously propagated to the PLC to initiate
diastole–systole and to the gating signal circuit which fed into
the gamma-camera system to trigger gating of data acquisition.
The respiratory phantom consisted of human-sized sili-
cone lungs (Figure 2B). The left lung air volume was 510 mL
and the right lung air volume was 565 mL. Each lung was
covered with a 1-mm-thick thermoplastic structures forming
the lung shape at the inhalation phase allowing inflation in an
anatomically correct manner. The breathing cycle and tidal
Polycarpou et al Journal of Nuclear Cardiology�SPECT myocardial perfusion imaging
volume followed a sinusoidal exponential pattern.13 Two
electro-proportional valves, controlled by the respiratory pat-
tern equation in the PLC, regulated the air pressure with
respect to time, entering in the lungs for inhalation and
escaping from the lungs for exhalation, respectively (Fig-
ure 1B). This was achieved using two small pumps with
suction and blowing ports that could be used as vacuum pumps
and small compressors. The tidal volume was 450 and 550 mL
for the normal and deep breathing, respectively. In both cases,
the breathing cycle duration was 6 seconds. A 2% deviation
was found using OsiriX14 compared to the set lung volumes.
The cardiac compartment could follow the respiratory
pattern in the cranio-caudal direction at the level of the
diaphragm. The mechanical system for this motion, a
controlled motor by the PLC and a screwed rod, was attached
to the outer base of the anthropomorphic phantom (Figures 1C,
3). The beating cardiac compartment was set to perform a
maximum displacement of 1.5 and 2.7 cm for the normal and
deep breathing conditions, respectively.15 All three motions
could be independently or simultaneously controlled using a
PLC.
The respiratory pattern was divided into four isochronous
phases (from exhale to inhale). Thus, the breathing lungs and
the moving heart could be mechanically stopped at any
respiratory phase, as shown in Table 1, while the heart was
continuously beating.
The Alderson tissue-equivalent anthropomorphic phan-
tom for nuclear medicine (Radiology Support Devices, Inc.)
Figure 1. Phantom assembly motions: (A) systole and diastole of the cardiac phantom,(B) breathing of the respiratory phantom, and (C) motion of the cardiac phantom during respiration.
Figure 2. Photos of (A) inner and outer membranes of the cardiac compartment with a defect,(B) cardiac compartment in systole and lung compartment at the exhale phase, and (C) cardiaccompartment with defect in diastole and lung compartment at the inhale phase.
Journal of Nuclear Cardiology� Polycarpou et al
SPECT myocardial perfusion imaging
was used to simulate an average male and a large male by
adding a thorax overlay, and a large female by adding breasts
on the thorax overlay (Figure 3). The cardiac and lung
compartments were anatomically hold within the thorax using
a shaped thin solid-water slab. The thorax was hermetically
closed and the rest of the cavity was filled with water to
simulate the soft tissue.
Data Acquisition
Acquisitions were performed for three body types with no
cardiac defects, with two small (0.5 cm thickness, 1 9 1 cm in
extent) and with two large (1 cm thickness, 1 9 1 cm in
extent) transmural defects. The defects were positioned at the
anterior and inferior regions of the LV myocardial wall
(Table 2). An activity of 15.5 MBq of Tc99m was injected into
the myocardium wall of the LV based on an estimate of 1.2%
uptake of a clinically relevant administered activity.16
SPECT data were acquired in a GE Millennium VG
gamma-camera with a single-slice Hawkeye CT at Nicosia
General Hospital. Acquisitions were performed using a routine
clinical protocol with Low-Energy High-Resolution collima-
tors in 90� (L-mode) orientation. SPECT data were acquired in
60 projections over 180� of rotation on 64 9 64 matrix
following the default clinical protocol at 20 seconds per
projection. Data were acquired at the required respiratory
phase while the heart was continuously beating. A 20% energy
window was centered over the 140 keV photopeak of Tc99m.
CT images were acquired in the heart region with 10 mm slice
thickness and a matrix size of 128 9 128. SPECT and CT
images were acquired in each of the four respiratory phases for
the normal and deep breathing, respectively (Table 2).
SPECT data were reconstructed using the ordered-subset
expectation-maximization algorithm with three iterations and
16 subsets, as commonly used in clinical practice. The image
matrix size was 64 9 64 9 19 with a voxel size of
6.9 9 6.9 9 6.9 mm. The Butterworth filter (cutoff: 0.52
cycles/cm, power: 5) was applied to the reconstructed images.
Attenuation correction was applied based on CT-derived
attenuation maps. SPECT data were generated with and
without motion correction. All data were re-oriented to form
short-axis images along the long axis of the LV processed by
the same operator in the same way for all reconstructions with
and without motion correction.
Figure 3. Photos of (A) average male, (B) large male, and (C) large female body types of theanthropomorphic phantom.
Table 1. Lung volumes and cardiac displacement at each of the four respiratory phases, from exhale(phase-1) to inhale (phase-4), for the normal and deep breathing modes
Phase
Normal breath Deep breath
Volume (mL) Displacement (mm) Volume (mL) Displacement (mm)
1 1075 0 1075 0
2 1201 5 1201 5
3 1373 10 1433 20
4 1538 15 1635 27
Table 2. SPECT/CT acquisitions for the threebody types, defect types, and breathing modes
Body type Defects Breathing mode
Average male None Static, normal, deep
Two small Static, normal, deep
Two large Static, normal, deep
Large male None Static, normal
Two small Static, normal, deep
Two large Static, normal, deep
Large female None Static, normal, deep
Two large Static, normal, deep
Polycarpou et al Journal of Nuclear Cardiology�SPECT myocardial perfusion imaging
Motion Correction
Motion correction was applied using an image-based
procedure. A non-linear registration algorithm17 was applied
to register each of the reconstructed SPECT respiratory
phases (gates) to the reference gate (lungs at the exhale phase
and no heart motion) allowing for possible non-rigid motion
patterns in the clinical setting in line with previous studies.18
In this way, transformation fields (i.e., displacement vectors
between the reference gate and each of the other gates)
describing the motion were obtained. The registration algo-
rithm is fully automated voxelwise which estimates non-rigid
motion by a combination of multiple local affine components
based on an adaptive hierarchical structure.17 After deriving
the motion fields, each gate was transformed to the reference
gate and these were averaged to create the motion-corrected
image.
A motion-corrected image was generated from the four
SPECT/CT acquisitions obtained when the phantom assembly
was mechanically stopped in each of the four respiratory
phases, for each breathing mode, defect type, and body type
(Table 2). In total, fifteen motion-corrected AC-SPECT
images (8 for the normal and 7 for the deep breathing modes)
were generated. These images were compared with the
corresponding AC-SPECT images (SPECT with breathing
lungs and moving heart, and AC with CT at the reference
gate). Figure 4 shows an example of co-registration of SPECT
with CT.
Image Assessment
The motion-corrected AC-SPECT images were qualita-
tively and quantitatively compared to the AC-SPECT images
to assess the effectiveness of motion correction in reducing
motion artifacts. For the quantitative comparison, the percent-
age increase of the myocardial perfusion uptake values after
applying motion correction was calculated for each of the 17
segments of the polar maps.
Two experienced nuclear medicine physicians reported on
images without defects and with one or two defects. The
physicians did not know about the phantom parameters or
image correction (AC, MC). The physicians reported on the
defect absence or presence. The physicians’ report was correct
if the defect position was correctly identified. The results from
the report on each defect were evaluated as True Positive (TP),
True Negative (TN), False Positive (FP), and False Negative
(FN).
RESULTS AND DISCUSSION
Figure 5 shows SPECT images of the four respira-
tory phases obtained for the large female with no defects
performing normal breathing. Figures 6 and 7 show AC-
SPECT images without motion and performing normal
and deep breathing before and after MC, obtained for the
average male with no defects and the large female with
large defects, respectively. Motion-induced artifacts
cause seemingly reduced anterior and inferior wall
uptake introducing false-positive findings. The corre-
sponding polar maps of Figure 7 are shown in Figure 8
together with the percentage increase of the myocardial
perfusion uptake for each LV segment after applying
motion correction. A visual inspection of the AC-
SPECT images shows that motion due to respiration
causes blurring and defect smearing implying a much
extensive disease compared to the motion-corrected
images. In particular, heart motion due to respiration
causes under-estimation of the tracer uptake mainly in
the inferior and anterior regions of the LV myocardial
wall. This was also reported in previous studies.3,4,7 The
apical artifact observed in some figures is due to a solid-
water square cross (0.5 cm side, 1 mm thickness)
attached at the apex of the cardiac compartment to
prevent the inner membrane touching the outer mem-
brane during diastole.
Motion correction substantially reduces the motion-
induced artifacts due to respiration observed in both
SPECT slices and polar maps. In particular, images after
motion correction result in more uniform uptake and
reduction of motion blurring, a thinner appearance of the
heart wall, especially in the inferior and anterior regions
of the LV myocardial wall. An increase of the uptake
values was also observed in the septal and lateral regions
after motion correction but not as severe as that
observed in the inferior and anterior regions. After
motion correction, the percentage increase of the MPI
uptake values in the anterior and inferior regions was
18% for normal and 35% for deep breathing, while the
corresponding increase in the apical, septal, and lateral
regions was 7% for normal and 9% for deep breathing.
The quantitative analysis shows that after motion cor-
rection, there is major improvement in the inferior and
anterior regions which are expected to be more affected
by motion.
Figures 9 and 10 show SPECT images with (A)
non-AC and non-MC, (B) AC and non-MC, (C) non-AC
and MC, and (D) AC and MC, for the average male with
small defects performing normal and deep breathing,
respectively. Motion correction does not recover the
uptake without applying attenuation correction. Apply-
ing only attenuation correction does not eliminate
blurring in the anterior and inferior regions due to
Figure 4. Example of transverse, coronal, and sagittal planesof the registered SPECT and CT for the average male with nodefects.
Journal of Nuclear Cardiology� Polycarpou et al
SPECT myocardial perfusion imaging
motion. The extent of improvement depends on the
amplitude of the heart motion.
Two physicians reported on the presence or absence
of cardiac defects in AC-SPECT images, with non-MC
and MC, for the normal and deep breathing modes. The
physicians’ results were considered to be in very good
agreement (Cohen’s Kappa coefficient: 0.86). Motion
led to FP due to artifacts. For example, when no motion
was present in phantoms with no defects, the physicians
reported no defects, which led to a TP evaluation.
However, when the phantoms with no defects performed
normal and deep breathing, the physicians reported false
Figure 5. SPECT slices of short, horizontal long, and vertical long axis, obtained for the largefemale with no defects, performing normal breathing: (A) gate-1; (B) gate-2; (C) gate-3; (D) gate-4.
Figure 6. AC-SPECT slices of short, horizontal long, and vertical long axis, obtained for theaverage male with no defects, without motion (A) and performing normal (B, C) and deep (D, E)breathing before (B, D) and after MC (C, E).
Polycarpou et al Journal of Nuclear Cardiology�SPECT myocardial perfusion imaging
defects in 4 out of 6 images and in 3 out of 4 images,
respectively, which led to an FP evaluation.
The 12 large defects were correctly reported in all
AC-SPECT images, that is, when the phantoms were in
static, normal, and deep breathing modes. Only 4 out of
the 8 small defects were reported in the AC-SPECT
images, when motion was not present (Table 3). How-
ever, when the phantoms performed normal and deep
breathing, all 8 small defects were reported in the AC-
SPECT images. This is because the defects were located
15%
24%
11%
4%
10%
9%
8%
3%
-5%
10%
1%
2%
5% 3%
7%
7% 9%
35%
32%
34%
13%
40%
45%
32%
5%
-4%
10%
5%
8%
3% 5%
12%
13% 6%
A B C
FD E
Figure 8. Quantitative analysis of the AC-SPECT polar maps obtained before (A, D) and after (B,E) applying MC and the percentage increase of the myocardial perfusion uptake (C, F) in each LVsegment after applying MC, for the large female with large defects performing normal (A, B, C)and deep breathing (D, E, F).
Figure 7. AC-SPECT slices of short, horizontal long, and vertical long axis, obtained for the largefemale with large defects, without motion (A) and performing normal (B, C) and deep (D, E)breathing before (B, D) and after MC (C, E).
Journal of Nuclear Cardiology� Polycarpou et al
SPECT myocardial perfusion imaging
Figure 9. Slices of short, horizontal long, and vertical long axis obtained for the average male withsmall defects performing a normal breathing: (A) non-AC and non-MC; (B) AC and non-MC; (C)Non-AC and MC; and (D) AC and MC.
Figure 10. Slices of short, horizontal long, and vertical long axis obtained for the average malewith small defects performing a deep breathing: (A) non-AC and non-MC; (B) AC and non-MC;(C) Non-AC and MC; and (D) AC and MC.
Polycarpou et al Journal of Nuclear Cardiology�SPECT myocardial perfusion imaging
in the anterior and inferior regions in which the uptake
values were mainly reduced due to motion-induced
artifacts. This was also reported in a similar study with
the NCAT phantoms.6
A combination of motion correction with attenua-
tion correction enhanced the reduction of artifacts.
Attenuation correction improved the uptake values in
the basal segments19, while motion correction improved
the uptake values in the anterior and inferior regions and
to a lesser extent in the lateral and septal regions.
Applying motion correction on AC-SPECT images, the
uptake values were increased in the anterior and inferior
regions, in which the small defects were located, with
the result 50% of the defects to be reported and
evaluated as FN, as in the case of static phantoms
(Table 3). Even though the images were corrected for
motion, to report on the presence or absence of small
defects, AC-SPECT images with and without motion
correction should be simultaneously inspected. The
study assumed the absence of misalignment of emission
and transmission (CT) data due to patient motion, and
therefore it represents an ideal case scenario with respect
to attenuation correction. The impact of attenuation
correction in the presence of motion-related misalign-
ment between the SPECT and CT has been highlighted
previously1, and the phantom of this study may be used
in future to investigate such effects.
NEW KNOWLEDGE GAINED
A combination of motion correction with attenua-
tion correction reduces artifacts in SPECT MPI. In AC-
SPECT images, small defects located on the anterior and
inferior regions of the myocardial wall of the left
ventricle become more visible due to motion-induced
artifacts. AC-SPECT images with and without motion
correction should be simultaneously inspected to be able
to report on small defects. The defects smeared by the
respiratory motion implying a much extensive disease
compared to the motion-corrected images leading to
misleading interpretation. Therefore, motion correction
is important for better diagnosis as it decreases the
smearing effect. Motion correction does not fully
recover the uptake without applying attenuation correc-
tion, provided that no misalignment between emission
and transmission data is present or has been corrected
for.
CONCLUSIONS
This study investigated the impact of respiratory
motion correction on AC-SPECT MPI and on defect
detection using a moving phantom assembly with
cardiac defects.
The inflatable lungs, and the ECG beating and
moving LV during respiration, simulated normal and
deep breathing. SPECT/CT data were acquired for each
of the four respiratory phases, from exhale to inhale, in
normal and deep breathing modes of the phantom
assembly. These SPECT/CT images were used to
generate the motion-corrected images. A non-linear
registration algorithm was applied to register each of the
four reconstructed respiratory phases to a reference gate
to obtain the transformation fields and generate motion-
corrected images. The quantitative analysis on AC-
SPECT images and the physicians’ report on defect
presence or absence, before and after applying motion
correction, were evaluated.
Motion correction results in more uniform
uptake and reduction of motion blurring, especially
in the inferior and anterior regions. The extent of
improvement depends on the amplitude of the heart
motion displacement. A combination of motion
correction with attenuation correction enhanced the
reduction of artifacts. Respiration motion has more
significant effects on the detection of small defects.
In AC-SPECT images, small defects located on
anterior and inferior regions become more visible
due to motion-induced artifacts. AC-SPECT images
with and without motion correction should be
simultaneously inspected to be able to report on
small defects.
This study is limited by the fact that it simulates
only the cranio-caudal displacement of the heart during
respiration. Also, defect detectability in the lateral and
septal wall was not investigated. The respiratory phases
were acquired sequentially and not continuously as
observed in real patients. Further limitations may be due
to phantom assembly (e.g., chest wall expansion, liver
and heart twisting).
Future studies will investigate more realistic condi-
tions of respiratory patterns and cardiac motion
amplitudes as well as defects in other regions of the
LV. Ultimately, such effects shall be studied in future by
investigating real patient cohorts with respiratory-gated
acquisition.
Table 3. Evaluation of physicians’ report (TP, FN)for the 8 small defects
ACstatic
ACnormal
AC1MCnormal
ACdeep
AC1MCdeep
FN 4 0 4 0 4
TP 4 8 4 8 4
AC and MC were applied on SPECT MPI data for phantoms instatic, normal, and deep breathing modes
Journal of Nuclear Cardiology� Polycarpou et al
SPECT myocardial perfusion imaging
Acknowledgments
We gratefully acknowledge the Nuclear Medicine
Department of the Nicosia General Hospital (Andreas
Mormoris, Chariklia Christodoulou, Lambros Lambrou,
Charoula Charalambous) for helping in acquisitions. We
also thank Guy’s and St Thomas Hospitals NHS Foundation
Trust (Steve Turner, Bill Stevens) for optimizing the cardiac
holder. The Project TCEIA/DTCEIA/0311(BIE)/27, with
Host Organization the Frederick Research Center and
Partner Organizations the Cyprus Ministry of Health and the
King’s College London, was co-financed by the European
Regional Development Fund and the Republic of Cyprus
through the Research Promotion Foundation.
Disclosure
The authors declared that they have no conflict of
interest.
References
1. Goetze S, Brown TL, Lavely WC, Zhang Z, Bengel FM. Atten-
uation correction in myocardial perfusion SPECT/CT: Effects of
misregistration and value of reregistration. J Nucl Med
2007;48(7):1090-5.
2. Fricke H, Fricek E, Weise R, Kammeier A, Lindner O, Burchert
W. A method to remove artifacts in attenuation-corrected
myocardial perfusion SPECT introduced by misalignment
between emission scan and CT-derived attenuation maps. J Nucl
Med 2004;45:1619-25.
3. Pitman AG, Kalff V, Van-Every B, Risa B, Barnden LR, Kelly
MJ. Effect of mechanically simulated diaphragmatic respiratory
motion on myocardial SPECT processed with and without atten-
uation correction. J Nucl Med 2002;43:1259-67.
4. Tsui BMW, Segars WP, Lalush DS. Effects of upward creep and
respiratory motion in myocardial SPECT. IEEE Trans Nucl Sci
2000;47:1192-5.
5. Wheat JM, Currie GM. Impact of patient motion on myocardial
perfusion SPECT diagnostic integrity: Part 2. J Nucl Med Technol
2004;32:158-63.
6. Yang YW, Chen JC, He X, Wang SJ, Tsui BM. Evaluation of
respiratory motion effect on defect detection in myocardial per-
fusion SPECT: A simulation study. IEEE Trans Nucl Sci
2009;56:671-6.
7. Segars WP, Tsui BMW. Study of the efficacy of respiratory gating
in myocardial SPECT using the new 4DNCAT phantom. IEEE
Nucl Sci 2002;49:675-9.
8. Kovalski G, Israel O, Keidar Z, Frenkel A, Sachs J, Azhari H.
Correction of heart motion due to respiration in clinical myocar-
dial perfusion SPECT scans using respiratory gating. J Nucl Med
2007;48:630-6.
9. Konik A, Kikut J, Lew R, Johnson K, King MA. Comparison of
methods of acquiring attenuation maps for cardiac SPECT in the
presence of respiratory motion. J Nucl Cardiol 2013;20:1093-107.
10. Ko C, Wu Y, Cheng M, Yen R, Wu W, Tzen K. Data-driven
respiratory motion tracking and compensation in CZT cameras: A
comprehensive analysis of phantom and human images. J Nucl
Cardiol 2015;22:308-18.
11. Livieratos L, Rajappan K, Stegger L, Schafers K, Bailey DL,
Camici PG. Respiratory gating of cardiac PET data in list-mode
acquisition. Eur J Nucl Med Mol Imaging 2006;33:584-8.
12. Chrysanthou-Baustert I, Parpottas Y, Demetriadou O, Christofides
S, Yiannakkaras C, Kaolis D, et al. Diagnostic sensitivity of
SPECT myocardial perfusion imaging using a pumping cardiac
phantom with inserted variable defects. J Nucl Cardiol
2013;20(4):609-15.
13. Cook LB. Rebreathing in the Mapleson A, C, and D breathing
systems with sinusoidal and exponential wavefronts. Anaeshesia
1997;52:1182-94.
14. Rosset A, Spadola L, Ratib O. OsiriX: An open-source software
for navigating in multidimensional DICOM images. J Digit
Imaging 2004;17(3):205-16.
15. McLeish K, Hill DL, Atkinson D, Blackall JM, Razavi R. A study
of the motion and deformation of the heart due to respiration.
IEEE Trans Med Imaging 2002;21(9):1142-50.
16. Hesse B, Kristina-Tagil A, Cuocolo C, Anagnostopoulos M,
Bardies J, Bax F, et al. EANM/ESC procedural guidelines for
myocardial perfusion imaging in nuclear cardiology. Eur J Nucl
Med Mol Imaging 2005;32(7):855-97.
17. Buerger C, Schaeffter T, King AP. Hierarchical adaptive local
affine registration for fast and robust respiratory motion estima-
tion. Med Image Anal 2011;15:551-64.
18. Axel M, Zikic D, Botnar R, Bundschuh R, Howe W, Ziegler S,
et al. Dual cardiac-respiratory gated PET: Implementation and
results from a feasibility study. Eur J Nucl Med Mol Imaging
2007;34(9):1447-54.
19. Masood Y, Liu Y-H, DePuey G, Taillefer R, Araujo LI, Allen S,
et al. Clinical validation of SPECT attenuation correction using x-
ray computed tomography-derived attenuation maps: Multicenter
clinical trial with angiographic correlation. J Nucl Cardiol
2005;12:676-86.
Polycarpou et al Journal of Nuclear Cardiology�SPECT myocardial perfusion imaging