July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector candidates perBelle open proto-collaboration meeting, KEK, July 3-4, 200

Upload: letitia-perkins

Post on 13-Dec-2015

219 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Peter KrižanUniversity of Ljubljana and J. Stefan

Institute

Burle MCP PMTs as photon detector candidates

SuperBelle open proto-collaboration meeting, KEK, July 3-4, 2008

Page 2: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Contents

MCP PMTs

Beam tests

Cross talk sources and modeling, impact on timing

Sensitivity and timing – position dependence

Ageing data

Summary

Page 3: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

BURLE 85011 microchannel plate (MCP) PMT: multi-anode PMT with two MCP stages

excellent performance in beam and bench tests very fast (=50ps for single photons)

Photon detector candidate for 1.5 T: MCP-PMT

Page 4: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

multi-anode PMT with two MCP steps bialkali photocathode gain ~ 0.6 x 106

collection efficiency ~ 60% box dimensions 71x71mm2

active area fraction ~ 52% 2mm quartz window

BURLE 85011 MCP-PMT 64 (8x8) anode pads pitch ~ 6.5mm, gap ~ 0.5mm 25 m pores

Basic parameters of BURLE MCP-PMTs

BURLE 85001 MCP-PMT 4 (2x2) anode pads pitch ~ 25mm, gap ~ 1mm10 m pores

Page 5: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Results: ~13 mrad (single cluster) number of clusters per track N~ 4.5 ~ 6 mrad (per track)→ ~ 4 /K separation at 4 GeV/c

MCP-PMT

multi-anode PMTs

Tested in pion beam combination with multi-anode PMTs.Stable operation, very good performance

To do list:improve collection efficiency and active area fraction higher number of det. photons doneaging study

NIM A567 (2006) 124Beam tests of Burle MCP PMT

Page 6: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Bench tests with pico-second laser: amplifier ORTEC FTA820A discriminator PHILIPS 308 CAMAC TDC Kaizu works KC3781A, 25ps LSB CAMAC charge sensitive ADC

Time resolution as a function of the number of detected photons.

MCP-PMT timing properties

Page 7: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

With a fast photon detector (MCP PMT), a proximity focusing RICH counter can be used also as a time-of-flight counter.

Time difference between and K

2GeV/c /K: t ~ 180ps

4GeV/c /K: t ~ 45ps

TOF capability of a RICH

Cherenkov photonsfrom PMT window

Cherenkov photonsfrom aerogel

MCP-PMTaerogel

track

STARTFor time of flight: use Cherenkov photons emitted in the PMT window

Page 8: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

TOF test with pions and protons at 2 GeV/c.Distance between start counter and MCP-PMT is 65cm

In the real detector ~2m 3x better separation

Expected number of detected Cherenkov photons emitted in the PMT window (2mm) is ~15 Expected resolution ~35 ps

TOF capability: window photons

NIM A572 (2007) 432

Page 9: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Time-of-flight with photons from the PMT window

Cherenkov angle aerogel, n=1.05

Aerogel: kaons (protons) have no signal below 1.6 GeV (3.1 GeV): identification in the veto mode.

Benefits: Čerenkov threshold in glass (or quartz) is much lower than in aerogel.

Window: threshold for kaons (protons) is at ~0.5 GeV (~0.9 GeV): positive identification possible.

Threshold in the windowK p

Page 10: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Burle MCP PMT has excellent timing properties, a promising photon detector also for very precise time measurements.

Additional bench tests needed: study detailed timing properties and cross-talk.

Determine their influence on the position resolution and time resolution

New bench tests: cross-talk and timing properties

Page 11: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Outside dark box: PiLas diode laser system EIG1000D (ALS) 404nm laser head (ALS) filters (0.3%, 12.5%, 25%) optical fiber coupler (focusing)

optical fiber (single mode,~4m core)

Inside dark box mounted on 3D stage: optical fiber coupler (expanding) semitransparent plate reference PMT (Hamamatsu H5783P) focusing lens (spot size ~ 10m)

Scanning setup: optical system

Page 12: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

amplifierORTEC

FTA820A

signal splitterpassive 3-way

discriminatorPhilips

model 806

QDCCAENV965VME

TDCKaizu worksKC3781ACAMAC

NIM

PCLabWindows

CVI

ALSPiLas

controller

laser rate 2kHz (~DAQ rate) amplifier: 350MHz (<1ns rise time) discriminator: leading edge, 300MHz TDC: 25ps LSB(~11ps) QDC: dual range 800pC, 200pC HV 2400V

Scanning setup: read-out

Page 13: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

TDC vs. ADC correlation is fitted with

and used for TDC correction

ADC

raw TDC

Time walk correction

3

21

PADC

PPTDC

ADC

raw TDC

Page 14: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

= 40ps = 37ps

= 38ps = 39ps70%

20%

10%

Corrected TDC distributions for all pads

Response: prompt signal ~ 70% short delay ~ 20% ~ 10% uniform distribution

Corrected TDC

Page 15: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Photo-electron: d

0,max ~ 0.8 mm

t0 ~ 1.4 ns

Δt0 ~ 100 ps

Parameters used: U = 200 V l = 6 mm E

0 = 1 eV

me = 511 keV/c2

e0 = 1.6 10-19 As

Backscattering: d

1,max ~ 12 mm

t1,max

~ 2.8 ns

Charge sharing

Photon electron detection: modeling

e−

d1

l

e−

d0

l

Page 16: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Distributions assuming that photo-electron is emitted at angle uniformly over the solid angle

Maximum variation of photo-electron travel time.

~ 90ps~ 0.8mm

Photo-electron: simple estimates

00

2

Ue

mlt e cos2

0

00 Ue

Eld

000

000 2 Em

Ue

l

Ue

Ett e

t0

d0

Page 17: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Laser: 15ps (rms)Electronics: 12ps (rms)TTS of photo-electron (blue): 26ps (rms)Sum in squares: 32ps very close to 37-40ps

Timing resolution, contributions

= 40ps = 37ps

= 38ps = 39ps

Time resolution of the main peak seems to be dominated by the photo-electron time spread

Page 18: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Distributions assuming that back-scattering by angle is uniform over the solid angle

Travel time t1 vs. travel distance d1

~ 2.8ns ~ 12mm

Elastic back-scattering

2sin21 ld sin2 01 tt

e−

d1

l

~ 12mm

2.8ns

Page 19: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

70%

20%

10% Elastically scattered photo-electrons

Understanding time-of-arrival distribution

Inelastically scattered photo-electrons?

Normal photo-electrons

Page 20: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Number of detected events at different positions of light spot – sum of all 4 channels double counting at pad boundaries due to charge sharing

Photon detection uniformity

Page 21: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

number of all detected events with maximum signal detected by the pad

number of events with maximum signal detected by other pads

number of delayed events with maximum signal detected by the pad

Photon detection response – single pad

back-scattered

charge sharing

Page 22: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Fraction of the signal detected on channel 1 vs. x position of light spot

sizable charge sharing in ~2mm wide boundary area can be used to improve position resolution

Charge sharing

Page 23: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Comparison of the charge sharing effect for red (635 nm) and blue (405 nm) laser

Charge sharingred blue

bluered

As expected: more photo-electron initial energy for blue photons

Page 24: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

all events with maximum signal on channels 1 and 2

delayed (>1.1ns) events with maximum signal on channels 1 and 2

Detailed 1D scan

1 2

2

2

2x12 mm

1

2x12 mm

1

= range of back-scatterd photo-electrons

Page 25: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Time of arrival vs. x

good uniformity over most of the surface

large deviation at active area edge small deviation at pad boundaries

Timing uniformity

1

1 2

Page 26: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Number of detected signals vs. x Small variation over central part

MCP with 8x8 pads: detection vs. x

Response similar to 2x2 MCP PMT: charge sharing and long tails due to photo-electron back-scattering.

Page 27: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

ch. 1

ch. 4

ch. 8

TDC vs. x correlation of single pads: same features as for the 2x2 tube uniform for central pads large variation for pads at the outer edges of the tube

8x8: Timing uniformity for single pads

Page 28: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Conclusions

Back-scattering range and spread in timing depend on the

photocathode-MCP plate distance photocathode-MCP plate voltage

The distance should be as small as possible, ~1mm (in the tested tube 6mm)

The voltage should be as high as possible, 500V max. allowed (in the tested tube fixed to 200V)

Some of the effects will be reduced (or disappear) in high B field, some will remain (timing)

e−

d1

l

Page 29: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

New version of the Burle/Photonis MCP

New version of the BURLE/Photonis (higher active area ratio)

Original plan for the June beam test: test this tube to see if there are any differences in performances the old one .

Unfortunately, poor beam conditions, too many other tests Next beam test (KEK or GSI?)

Page 30: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Page 31: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Ageing – data from Burle/Photonis

Several discussions with Emile Schyns, Group Product Manager, Micro Channel Plates (final a good lead to the company)

Current performance (no Al protection layer): 50% drop of efficiency after 10-15C/tube = 350-540mC/cm2

Expect ~10mC/cm2/year on ARICH (scaling the TOP estimate)

Summer 08: move production to Europe, expect to improve the ageing by a factor >5 (use a different scrubbbing technique, deep UV electrons)

Ageing most probably not a problem any more (homework for us to check our rate estimates)

Page 32: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Cost estimate, production capabilities

Discussion with Emile Schyns

Cost will depend on specs (min. gain, uniformity, pore size, , so we have to finalize our requirements

Expect a rough estimate in two months

Production of ~1000 pieces until 2011 not a problem if the decision is taken early in 2009; could probably do it even faster if really needed.

Page 33: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Summary

Burle multianode MCP PMTs have been tested in beam and bench tests stable operation, very good performance

They have excellent timing properties a promising photon detector also for very precise time measurements.

Additional bench tests were carried out: study detailed timing properties and cross-talk

Next: determine their influence on the position resolution and time resolution

Still some work to do...• Read-out electronics (wave-form sampling, G. Varner?)• Ageing studies• Cost estimate...

Page 34: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

Back-up slides

Page 35: July 4, 2008 SuperBelle Meeting, KEKPeter Križan, Ljubljana Peter Križan University of Ljubljana and J. Stefan Institute Burle MCP PMTs as photon detector

July 4, 2008 SuperBelle Meeting, KEK Peter Križan, Ljubljana

TDC vs. x for channels 1 and 2 large deviation at active area edge small deviation at pad boundaries

Timing uniformity1

2

1 2