DESCRIPTION
Shin-Nippon Co, Ltd, Japan Measuring capacity Corneal Radius of Curvature ○ 6.50 ~ 9.40 mm Corneal Keratometric Power ○ 36 ~ 52 D Axis Corneal Astigmatism ○ 0 ~ 180° 3TRANSCRIPT
![Page 1: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/1.jpg)
Jae Lim Chung MD, MBAJae Lim Chung MD, MBA1,31,3; Jin Pyo Hong MD; Jin Pyo Hong MD2,32,3; Kyoung Yul Seo MD, ; Kyoung Yul Seo MD, PhDPhD33; Eung Kweon Kim MD, PhD; Eung Kweon Kim MD, PhD33; Tae-im Kim MD, PhD; Tae-im Kim MD, PhD33
11Noo Ne Eye Hospital, Seoul, Korea22Sae Yan Eye Clinic, Seoul, Korea
33Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.
None of the authors have any commercial or financial interests
![Page 2: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/2.jpg)
For Successful Outcome Of Toric IOL Precise Measurement Of Preoperative
Corneal Astigmatism Crucial Point !
To Determine Accuracy & Correlation of 6 Different Keratometers & Topographers
2
![Page 3: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/3.jpg)
Shin-Nippon Co, Ltd, Japan
Measuring capacityCorneal Radius of
Curvature ○ 6.50 ~ 9.40 mm
Corneal Keratometric Power ○ 36 ~ 52 D
Axis Corneal Astigmatism ○ 0 ~ 180°
3
![Page 4: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/4.jpg)
Canon, Japan Measuring capacity
Corneal Radius of Curvature ○ 5.5 ~ 10 mm
Corneal Refractive Power ○ 33.75 ~ 61.25 D
Axis Corneal Astigmatism ○ 0 ~ 180°
4
![Page 5: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/5.jpg)
• Carl Zeiss Meditec, Germany
• 2.5 mm measurement diameter• Cf. Manual keratometers ~ 3 mm
• 30 data points per measurement• 6 spots x 5 samples
5 images automatically captured in 0.5 sec
Displayed result is average of 5 captured images
5
![Page 6: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/6.jpg)
6
Bausch & Lomb, USA Topography-based Simulated K-readings
(=Sim K’s) c Axis 3 & 5 mm zone
statisticsAverage & SD of Mean &
Astigmatic Power Steep & Flat Axis
![Page 7: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/7.jpg)
Oculus, Germany Scheimpflug image-
based Topographic analysis of Corneal Astigmatism
7
![Page 8: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/8.jpg)
Tracey Technologies, USA Dual Acquisition Unit
EysSys Corneal Topographer○ Placido ring-based
Wavefront analyzer
8
![Page 9: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/9.jpg)
Study PopulationPreoperative visit for Cataract Surgery257 eyes of 141 patients○ Age : 54.5 ± 20.5 yrs○ Sex : Male - 54, Female - 87
Manual Keratometric MeasurementDuplicated examination by 2 skillful
examinersConsidered as Standard
Compared with 5 other Keratometers9
![Page 10: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/10.jpg)
Pearson Correlation analysis Between Manual Keratometery Vs Other Methods
10
Vs. Manual Keratometer
Pearson Correlation Coefficient
iTrace 0.947
ORBscan 0.921AutoKeratometer 0.919IOL master 0.857
Pentacam 0.847
![Page 11: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/11.jpg)
Pearson Correlation analysis Between Manual Keratometery Vs Other Methods
11
Vs. Manual Keratometer
Pearson Correlation Coefficient
iTrace 0.856
AutoKeratometer 0.852IOL master 0.751ORBscan 0.646
Pentacam 0.633
![Page 12: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,](https://reader036.vdocuments.us/reader036/viewer/2022062600/5a4d1b327f8b9ab05999bc0f/html5/thumbnails/12.jpg)
Duplicated Manual Keratometric Measurements By 2 Experienced Operators Taken As Standard
iTrace Approximated This Standard The Most
Auto Keratometer, ORB & IOL Master Showed To Be Reasonably Accurate.
Trend For ORB Fail To Read Accurately In Steep Axis
12