ivivc in pediatric oips - ipac-rs...ipac-rs, ivivc in pediatric oips, h. wachtel, march 20, 2014 15...
TRANSCRIPT
IPAC-RS/UF Orlando Conference 2014 March 20, 2014
IVIVC in Pediatric OIPs
Herbert Wachtel
Declaration of Conflicts of Interest
H. Wachtel is employee of Boehringer Ingelheim Pharma GmbH & Co. KG, Germany.
2 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Contents
Success factors for pediatric inhalation: • the inhalation device, • the pharmaceutical formulation • the patient (adherence…)
Impactor measurements acc. to USP/Ph.Eur. vs. replica • flow profiles and electronic lung
Comparison with in vivo deposition data …… literature „validation“
Checking ped. M-T replica using monodisperse particles
– 3 micrometers (and below?) are our target size Applications to device and formulation development
Summary
introduction
3 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
4
Motivation : Success factors for pediatric inhalation... (1) The inhalation device
SALE – Devices– SALE – Devices – SALE – Devices– SALE – Devices – Devices– SALE –
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
5
Motivation : We are prepared for any challenge ... (2) The pharmaceutical formulation
SALE – Powders for inhalation – SALE – Powders for inhalation – SALE -
toxicologically acceptable force control agents? IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
6
Motivation : We are prepared for any challenge ... (3) The patient
Attention – Customer approaching – Attention – Customer approaching – Attention
Improper use of devices is common
Design inhaler / accessory for the use with children
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Summary of motivation section: - what we want to do:
Make better medicines for children (~>EuPFI, US-PFI) How?
• Demonstrate performance in clinical trials • Establish laboratory tests ( draft: <1602> Spacers and valved holding chambers ) • mix Lab.- Methods and flow profile measurements
-> Handling Studies • low risk • low cost
7 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
The aerodynamic diameter is important, but there are additional influencing factors (inhalation, airway dim. …)
H. Wachtel - PCP ad board 2014 8
ICRP-Modell
0.0
0.2
0.4
0.6
0.8
1.0
0.0 2.0 4.0 6.0 8.0 10.0 12.0
depo
sitio
n pr
obab
ility
aerodynamic diameter [µm]
alveolarbronchialextrathoracictotal
Source: ICRP66; 1994 International Commission of Radiological Protection + later updates
shift expected for younger
The pharmacopeial aerodynamic fine particle assessment: - lots of work but it is far away from the patient
Apparatus A (EP only) Q = 60 L/min
Apparatus C, 4 Q = 60 L/min (and calcul.)
Appa
ratu
s 2 (U
SP o
nly)
Q
= 6
0 L/
min
(and
cal
cul.)
Apparatus E, 5, 6 (with presep.) Q = 15 – 100 L/min
Apparatus D, 1, 3 (with presep.) Q = 28.3 L/min (preferred)
9 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
10
Air flow profiles for testing: important, they co-define the delivered dose, site, …
Compendial testing Duration T (s) = 4 L * 60 (s/min) / Q (L/min)
T flow rate to achieve pressure drop = 4 kPa
alternative testing (averaged ?) flow profile corresponding to flow resistance of device
volume = 4 L
volume = 1.7 L
COPD (fitted avg.)
Example HandiHaler with flow resistance 0.16 Sqrt(mbar)*min/L
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
11
Performance tests closer to reality: Finlay‘s idealized throat models the patient.
The simplistic USP throat used for release testing is complemented by
more realistic throat models for early development.
~ ~
Alberta throat child‘s throat USP-inlet
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Realistic models of children‘s extrathoracic airways for in vitro inhaler testing.
V=0.0398 L V=0.0245 L V=0.0148 L V=0.0112 L
for comparison: adult model V=0.079 L
Age 3-4 yrs Age 2-3 yrs Age 1-2 yrs Age 4-5 yrs
12 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
The in vitro patient:
13
Inhalation device
Mixing inlet
Upper airway model
Electronic lung
To pressurized air
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Minimalistic set-up for realistic inhaler tests using filters and a lung simulator
e.g. Eklira Genuair
Filter
Throat model
Breathing pattern Lung simulator ASL 5000
14 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Contents
Success factors for pediatric inhalation: • the inhalation device, • the pharmaceutical formulation • the patient
Impactor measurments acc. to USP/Ph.Eur. vs. replica • flow profiles and electronic lung
Comparison with in vivo deposition data …… literature „validation“
Checking ped. replica using monodisperse particles
- 3 micrometers (and below?) are our target size Applications to device and formulation development
Summary
introduction
15 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Aerosol deposition studies using our realistic pediatric models are close to published in vivo data (1)!
Link to scintigraphic deposition data of Wildhaber et al. (1999) in six 2 year-old children using the same pressurized Metered Dose Inhaler (albuterol) and valved holding chamber (AeroChamberPlus with facemask).
16
Age 2-3 yrs
0
20
40
60
80
100
Lung deposition Actuator VHC
% d
rug
depo
sitio
n re
late
d to
the l
abel
clai
m
In vivo (Wildhaber et al.*) In vitro (our model)
* J.H. Wildhaber et al. (1999), The Journal of Pediatrics, Vol. 135, No. 1, pp 28-33
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Aerosol deposition studies using our realistic pediatric models are close to published in vivo data (2)!
17
0
20
40
60
80
100
Lung deposition Face Face mask%
of d
rug
depo
sitio
n re
late
d to
the
labe
l cla
im
in vitro in vivo *
* Erzinger et al. (2007), Journal of Aerosol Medicine, Vol. 20, S1, S78-84
0
20
40
60
80
100
Lungdeposition
Face Face mask
% o
f dru
g de
posi
tion
rela
ted
to th
e la
bel c
laim
in vitro in vivo *
2-3 yrs old
1-2 yrs old
Link to scintigraphik deposition data of Erzinger et al. (2007) in children aged 1-3 yrs using a pMDI (albuterol) and a valved holding chamber.
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
don‘t forget the nose!
Aerosol deposition studies using realistic pediatric models are close to published in vivo data .
0
20
40
60
80
100
Lungdeposition
Face Face mask
% o
f dru
g de
posi
tion
rela
ted
to th
e la
bel c
laim
in vitro - oral inhalationin vitro - nasal inhalationin vivo
Link to scintigraphic deposition data of Erzinger et al. (2007)* in children aged 1-3 yrs using a pMDI (albuterol) and a valved holding chamber.
* Erzinger et al. (2007), Journal of Aerosol Medicine, Vol. 20, S1, S78-84
*
1-2 yrs old Oral inhalation:
Nasal inhalation:
18 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Summary of the experimental part (1)
The set up has been shown and compared to: a) pharmacopeial procedures (dose tubes, impactors) b) literature -> ( ~ Validation?) limitation: ethic considerations in children our way forward: refer to existing deposition studies clear gap: Best practice in adults: see e.g. study by Bo Olsson et al. J Aerosol Med and Pulm. Drug Delivery 26 (0), 2013, 1–15 - Validation of a General In Vitro Approach … Recommendation: Please consider recording inhalation flow profiles when planning your next pivotal pediatric study.
19 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Contents
Success factors for pediatric inhalation: • the inhalation device, • the pharmaceutical formulation • the patient
Impactor measurments acc. to USP/Ph.Eur. vs. replica • flow profiles and electronic lung
Comparison with in vivo deposition data …… literature „validation“
Checking ped. replica using monodisperse particles
– 3 micrometers (and below?) are our target size Applications to device and formulation development
Summary
introduction
20 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Where do our <<inhaled>> particles go?
methylene blue Inner coating: Brij + glycerol + water
21 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Checking pediatric replica using monodisperse particles
Vibrating Orifice
Generator
manifold &
Aerosizer
particles
SEM image of methylene blue particle
22 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Example: In-vitro performance with methylene blue
0%
20%
40%
60%
80%
100%
15 L/min 30 L/min 60 L/min
Dose
to L
ung
[%]
1-2 yrs mouth
3µm5µm7µm
0%
20%
40%
60%
80%
100%
15 L/min 30 L/min 60 L/min
Dose
to L
ung
[%]
1-2 yrs nose
3µm
5µm
7µm
Relevant, as children < 18 months will inhale through their nose
n.b.: typical mouth, typical nose, not from the same subject!
23 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Example: In-vitro performance with methylene blue
Partly relevant, older children will inhale through their mouth
0%
20%
40%
60%
80%
100%
15 L/min 30 L/min 60 L/min
Dose
to L
ung
[%]
4-5 yrs mouth
3µm5µm7µm
0%
20%
40%
60%
80%
100%
15 L/min 30 L/min 60 L/min
Dose
to L
ung
[%]
4-5 yrs nose
3µm5µm7µm
n.b.: typical mouth, typical nose, not from the same subject!
24 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Contents
Success factors for pediatric inhalation: • the inhalation device, • the pharmaceutical formulation • the patient
Impactor measurments acc. to USP/Ph.Eur. vs. replica • flow profiles and electronic lung
Comparison with in vivo deposition data …… literature „validation“
Checking ped. replica using monodisperse particles
– 3 micrometers (and below?) are our target size Applications to device and formulation development
Summary
introduction
25 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Application to dry powder inhalers: Easyhaler - Novolizer & idealized child model (4-5 years)
constant flow flow profile
DD throat DTL DD throat DTL
DD throat DTL DD throat DTL
0 20 40 60
drug
mas
s (%
ND
)
80 100
0 20 40 60
drug
mas
s (%
ND
)
80 100
0 20 40 60
drug
mas
s (%
ND
)
80 100
0
20
40
60
drug
mas
s (%
ND
)
80
flow = 28, 41, 60 L/min
flow = 45, 60, 75 L/min
PIF = 9 L/min
PIF = 24, 42, 51 L/min
Easyhaler
Novolizer
albuterol sulfate lactose blend
26 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Formulation development: Test with lactose (34µm or 19µm) blend containing albuterol sulfate
Lactose blend 1 blend 2 34µm 19µm
A. Below, thesis Univ. Düsseldorf (2013)
0 20
40
60
0
20
40
60
DTL
(% D
D)
DTL
(% D
D)
Easyhaler: • no influence of carrier size • no throat effect
Testing @ 4 kPa
Novolizer: • cyclone+impact+ carrier size • throat effect
tube-like
cyclone
Easyhaler
Novolizer
27 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Formulation dependence: Soft pellets do not work everywhere
Am
ount
of d
rug
[% D
D]
Throat
DTL
<5 µm
A. Below, thesis Univ. Düsseldorf (2013)
Alberta Throat Const. Flow 4 kPa 4 L
60 µm
28 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
29
Different types of spacers / valved holding chambers lead to different throat deposition and dose to lung.
0
20
40
60
80
100
Respimat Funhaler Vortex AerochamberPlus
% o
f lab
el c
laim
Mouthpiece Respimat Spacer Throat DTL
DTL33%
DTL22%
DTL33%
ThroatThroat
DTL51%
Throat
RH=50%
Results using throat model and flow profiles a)single breath Respimat and b)5 breaths with spacers of a 5 year-old child.
Can Pediatric Throat Models and Air Flow Profiles Improve Our Dose Finding Stratety? Herbert Wachtel, Deborah Bickmann, Jorg Breitkreutz, Peter Langguth RDD 2010, Vol 1 (2010): pp 195-204
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Focus on the patient: Very young children (below 5 years)
The young child‘s different „hardware“requires adaptation!
0 5 10 15 20-0.1
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
air f
low
rate
(L/s
)
time (s)
0.0
0.5
1.0
1.5
2.0
1/f
T_pu
lse
Tin
release
VA
Delta
Delta
2
VCin
PIFA
inhaled volume (L)
below 5 years
flow resistance
30 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Handling study investigating children below 5 years: stepwise approach
What is checked?
With help by caregiver
Child alone
Valved holding chamber with help by caregiver
below 5 years
31 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
In-vitro results using typical flow profiles of children -- valved holding chamber with face mask --
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
years years years years r.h.
effect of rel. humidity
32
Dose prediction knowing the inhalation profile (and the inhaler/spacer+mask combination)
Typical patient data:
Resulting dosing prediction:
Weight (kg) Two puffs (µg) µg/kg
10 2 x 0.13 0.026
13.1 2 x 0.39 0.060
16.5 2 x 0.46 0.056
22.4 2 x 0.96 0.086
70 (adult) 2 x 1.6 0.046
Existing theory can be applied In-vitro calibration data
33 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Summary & Conclusion
Deposition studies in vitro require: – Throat models – Inhalation flow profiles
and a common level of acceptance / standardization. -> CSA Z264 1-02 (Spacers + … ) There is an urgent need for these tools in order to enable studies representing children of all age groups and even sub-groups immediately when devices / formulations are created. The in vitro studies contribute to a better understanding of device–patient interaction and help e.g. to extend the range of applications a device might face by simplifying tests with accessories, e.g. spacers.
34 IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014
Many thanks to …
- D. Bickmann, A.-M. Ciciliani A. Jung, M. Metzger, R. Winkler: Boehringer Ingelheim - Professor Dr. Jörg Breitkreutz, A. Below: Heinrich Heine-University, Düsseldorf - Professor Dr. Peter Langguth: Johannes Guttenberg University, Mainz
35
Herbert Wachtel Boehringer Ingelheim Pharma GmbH & Co. KG
Respiratory Drug Delivery Binger Str. 173
55216 Ingelheim am Rhein
+49 (0) 6132 – 7798552 [email protected]
IPAC-RS, IVIVC in Pediatric OIPs, H. Wachtel, March 20, 2014