iu22 qc procedure

Upload: thebossplayer20029630

Post on 10-Apr-2018

226 views

Category:

Documents


1 download

TRANSCRIPT

  • 8/8/2019 Iu22 Qc Procedure

    1/11

    1

    Philips iU22 Annual QC Test Procedure

    Equipment Gammex RMI grayscale ultrasound phantom

    Ultrasound gel

    Philips iU22 console

    19 20

    21

    22

    23

  • 8/8/2019 Iu22 Qc Procedure

    2/11

    2

    1 Res/Spd Select between Resolution and Speed. Changes line density

    2 Gray Map Select the gray map to be used for image display

    3 Sector Width Change the number of sectors used in the transducer

    4 Compress Change the amount of image compression

    5 Protocol Select a pre-set protocol to determine scanning parameters

    6 Review Review images that have been obtained during the current exam7 Annotate Annotate the current image

    8 End Exam End the current exam

    9 Transducer Select a transducer to activate

    10 Patient Data Switch to a screen to input patient data

    11 Zoom

    12 Caliper Use the caliper tool to make measurements on the image

    13 Focus* Change the number of foci / change the focal depth

    14 Depth Change the depth

    15 Pointer Show a cursor

    16 Freeze Freeze on the current image / unfreeze the current image

    17 Print Save the current image18 iScan Return to the image acquisition screen

    19 XRes

    20 SonoCT

    21 2D Opt Transducer frequency setting (Pen, Gen or Res)

    22 2D Adjusts the image gain

    23 Dual display mode

    * This button has dual functions. Toggle between the two functions by pressing the button.

  • 8/8/2019 Iu22 Qc Procedure

    3/11

    3

    Test Procedure: Philips iU22

    (1) Power up machine. Verify accuracy of displayed time and date.

    (2) Perform mechanical inspection (wheel locks, etc). Visually inspect each transducer face(cracks, cuts, dents, other damage), connector pins (dust, damage), transducer and power

    cables (cuts, defects, other damage). Check the air filter for cleanliness*.(3) Start new patient using the ZPAF convention specified in the

    Ultrasound_annual_test_schedule.xls spreadsheet.(4) Select transducer and preset (usually Gen/Gen). (Button 9) The active transducer is the one

    shown in green.(5) Remember: NEVER CONNECT/DISCONNECT AN ACTIVE TRANSDUCER

    (To de-activate the current transducer, attach another transducer and select that secondtransducer. You will hear a clicking sound when the second transducer has been activated.)

    (6) Set TGC to be centered.(7) Add the transducer serial number to image annotation. (Button 7)

    (8) Add the phantom serial number to image annotation.(9) For each transducer, turn off SONO-CT, X-RES and harmonics if activated. (These buttons

    are located on the touch-screen. These options are OFF when buttons are blue and ON if buttons are green).

    (10) All tests are performed at the highest frequency available for each transducer. This is doneby setting the 2D Opt to Res. (Button 21)

    (11) Verify all settings each time you select a different transducer. (12) To make sure you are not capturing cine images, freeze each image before saving. Make

    sure the image is annotated properly before it is saved. (Buttons 16 (Freeze) and 17 (Print))

    *Air FilterThe air filter is located on the lower left-hand side of the ultrasound unit.Turn off power before you remove the air filter. Do not turn power on without an air filter installed.

  • 8/8/2019 Iu22 Qc Procedure

    4/11

    4

    (1) Uniformitya. Position the probe in most homogeneous area of phantom (parallel to short axis of the

    phantom, if possible).

    b. Set depth at maximum achievable, but no deeper than just below reflection from thebottom of phantom. (Button 14)

    c. Use only one focus (Button 13), set at the deepest depth or just above the base ofphantom.

    d. 2D Opt: Res (Button 21)e. Set gain to demonstrate uniformity. (Button 22)

    f. Adjust TGC if necessary to eliminate horizontal bandingg. Evaluate for artifacts, dropouts, or excessive banding that cannot be corrected with

    TGC adjustments. If artifacts are present, store additional images to show theseartifacts.

    For curvilinear transducers, display the full FOV. Roll the probe left and right to evaluate for

    dropout across transducer face. Store 2-3 images to demonstrate image quality across the entiretransducer face.

  • 8/8/2019 Iu22 Qc Procedure

    5/11

    5

    For curvilinear transducers, the FOV (sector width) can be reduced(Button 3) for the

    remaining tests.

    (2) Depth of penetrationa. Set depth to the maximum achievable depth, but no deeper than just below the

    reflection from the bottom of the phantom. b. Set the focus at the deepest depth or just above the base of phantom.

    c. Make sure Gain is set to 100% for this test. (If setting the gain to 100% over-saturates the image, a lower gain setting can be used).

    d. If the beam does not penetrate to the bottom of phantom, display images in dual mode by selectingDual mode. (Button 23) The second image should be obtained in air.

    Measure the depth where there are no more usable echoes, or where the electronicnoise is equivalent in both images. Store the dual image.

    e. If the beam penetrates to bottom of phantom, a single image can be stored.Reminder: Measure the depth of penetration on the system before the image is stored.

    The depth of penetration is less than the depth of the phantom (i.e., a reflection off the

    bottom of the phantom is not visible).

  • 8/8/2019 Iu22 Qc Procedure

    6/11

    6

    Example where the beam penetrates to the bottom of the

    phantom, shown by the reflection.

  • 8/8/2019 Iu22 Qc Procedure

    7/11

    7

    (3) Caliper accuracy(Axial and Lateral)a. Set the depth at 5-6 cm

    b. Set the focus between the sets of axial and lateral caliper accuracy pins

    c. 2D Opt: Resd. Set the gain so that the pins are highly visible

    e. Adjust the angle of the transducer on the phantom until the pins are as circular aspossible. (This will be easier with the linear arrays than with the curvilinear arrays).

    f. Make 2 measurements: For lateral caliper accuracy, position the cursor above thepins. For vertical caliper accuracy, position the cursor beside the pins (i.e. the cursors

    should not obscure the pins).Reminder: Make the caliper measurements on the system before the image is stored.

  • 8/8/2019 Iu22 Qc Procedure

    8/11

    8

    (4) Resolution (Axial and Lateral)a. Set the depth to ~ 4cm (the depth of the resolution pattern).

    b. Focus at the level of the resolution pattern.c. 2D Opt: Res

    d. Set the gain to minimize background and visualize pins.e. Adjust the angle of the transducer on the phantom until the pins are as circular as

    possible. (This will be easier with the linear arrays than with the curvilinear arrays).f. Freeze the image.

    g. Zoom in on the pins. (Button 11)h. Make a measurement of FWHM. Decrease the gain so that the background is not

    visible. Areas of reverb beneath the pins shouldnotbe included in the FWHMmeasurement.

    i. Freeze the image displaying the FWHM measurements. j. When filling out the QC report, note the minimum distance between pins that is fully

    resolvable.Reminders: The FWHM measurement should agree with the resolution limits shown by

    the resolution pattern.

  • 8/8/2019 Iu22 Qc Procedure

    9/11

    9

    (5) Dead zonea. Set the depth at approximately 2.0 cm

    b. Focus at the shallowest level.c. 2D Opt: Res

    d. Set the gain to visualize pins and the presence or absence of reverb near surface.e. Record the number of pins that are visible.

  • 8/8/2019 Iu22 Qc Procedure

    10/11

    10

    (6) Anechoic cysta. Reduce the depth to ~5.0 cm

    b. Focus at the depth that is just below the largest anechoic object.

    c. 2D Opt: Resd. Turn up the gain to look for reverberations inside the cysts. (Storing this image is

    optional unless there are reverbs/artifacts to demonstrate)e. Turn the gain down to demonstrate a lack of internal reflections/noise. All three cysts

    should be round (non-distorted) with distal enhancement. (Store this image.)

    A zoomed image of the anechoic objects in the phantom.

    A diagram of the anechoic objects, adapted from the Gammex manual.

  • 8/8/2019 Iu22 Qc Procedure

    11/11

    11

    (8) Testing an Additional Transducer

    Test remaining transducers within the same exam (if possible).

    Connect the next transducer to be tested and select the new transducer. (Button 9) Beforedisconnecting a transducer, be sure to de-activate that transducer.

    Repeat the set-up described on p.3 to ensure that all of the proper settings are selected. Modify the image annotation with the serial number of the next transducer. Verify X-Res and

    SonoCT are offfor each transducer and that the resolution is set at the highest available.

    (9) After all Acceptance Testing is Finished

    Review the images (Button 6) to ensure that all of the necessary images were captured. End exam. (Button 8)

    Send the images to Stentor. Shut down the machine

    Clean the gel from the transducers and the phantom with a soft, clean towel.

    (10) Additional Reminders / Some Common Mistakes

    Make sure that for the depth of penetration measurements, the focal depth is set appropriately.(It should be set to the maximum focal depth, or, if the bottom of the phantom is visible in the

    image, to just above the bottom of the phantom. Record distance measurements (depth of penetration, caliper accuracy, FWHM) on the

    ultrasound image. Make sure to make measurements before capturing images, so that thesemeasurements are recorded on the image sent to Stentor.

    If you use the zoom option, make sure you zoom the frozen image, NOT the live image. When testing multiple transducers on a single unit, be sure to update the transducer serial

    number on the images before testing a new transducer.

    Deleting Individual Images To delete individual images, review the images ( Button 6) and select the image number next to

    the images you want to delete. The selected images will have image numbers highlighted inorange. Then select the delete selected images option. Warning: This permanently deletes

    the selected images from the ultrasound hard drive.