introduction to spectroscopic methods chapter 6 instrumental analysis

27
Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Upload: oswald-harmon

Post on 16-Jan-2016

304 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Introduction to Spectroscopic Methods

Chapter 6

Instrumental Analysis

Page 2: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Introduction

An introduction into any field requires that one learn the terms and symbols associated with work in that field. Unfortunately, the terms used in spectroscopy and Spectrophotometry are somewhat confusing.

The common terms and alternative names/symbols employed in spectroscopy are listed in the table below. The recommended terms and symbols are listed under the column labeled Term and Symbol.

Page 3: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Term SymbolsTerm and Symbol Alternative Name and Symbol

Radiant Power, P, P0 Radiation Intensity, I, I0

Absorbance, A Optical Density, D

Transmittance, T Transmission, T

Path Length, b l, d

Absorptivity, a Extinction Coefficient, k

Molar Absorptivity, Molar Extinction Coefficient

Page 4: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Transmittance

Absorbance, like the previous table shows, can be defined as the base-ten logarithm of the reciprocal of the transmittance :

A = log 1/T = -log I/Io = -log P/Po

Page 5: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Reflection and Scattering Losses

Page 6: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Reflection and Scattering Losses (cont)

Reflection and scattering losses are significant and to compensate for these effects, the power of the beam transmitted by the analyte solution is ordinarily compared with the power of the beam transmitted by an identical cell containing only the solvent. An experimental absorbance that closely approximates the true absorbance is then obtained with the equation:

A = log Psolvent/Psolution x log Po/P

Page 7: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Beer’s LawBouguer, and later Lambert, observed that the fraction of the energy, or the intensity, of radiation absorbed in a thin layer of material depends on the absorbing substance and on the frequency of the incident radiation, and is proportional to the thickness of the layer. At a given concentration of the absorbing species, summation over a series of thin layers, or integration over a finite thickness, leads to an exponential relationship between transmitted intensity and thickness.

Page 8: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Beer’s Law (cont)Beer showed that, at a given thickness, the absorption coefficient introduced by Lambert’s law was directly proportional to the concentration of the absorbing substance in a solution. Combination of these two results gives the relationship now commonly known as Beer’s law. This law states that the amount of radiation absorbed or transmitted by a solution or medium is an exponential function of the concentration of the absorbing substance present and of the length of the path of the radiation through the sample.

Page 9: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Consider This:A parallel beam of monochromatic radiation with power Po strikes the block perpendicular to a

surface after passing through a length b of the material , which contains n absorbing particles , the beam’s power is decreased to P as a result of absorption.

Page 10: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Deviations from Beer’s Law

Beer’s law states that a plot of absorbance versus concentration should give a straight line passing through the origin with a slope equal to ab. However, deviations from direct proportionality between absorbance and concentration are sometimes encountered.

These deviations are a result of one or more of the following three things ; real limitations, instrumental factors or chemical factors.

Page 11: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Real LimitationsBeer’s law is successful in describing the absorption behavior of dilute solutions only ; in this sense it is a limiting law. At high concentrations ( > 0.01M ),the average distance between the species responsible for absorption is diminished to the point where each affects the charge distribution of its neighbors. This interaction, in turn, can alter the species’ ability to absorb at a given wavelength of radiation thus leading to a deviation from Beer’s law.

Page 12: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Limitations (cont)

Deviations also arise because e is dependent upon the refractive index of the solution. Thus, if concentration changes cause significant alterations in the refractive index h of a solution, departures from Beer’s law are observed.

It is not e which is constant and independent of concentration, but the expression:

a = atrue x a/( a² + 2)²

Page 13: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Chemical Deviations

Chemical deviations from Beer’s law are caused by shifts in the position of a chemical or physical equilibrium involving the absorbing species.

A common example of this behavior is found with acid/base indicators.

Deviations arising from chemical factors can only be observed when concentrations are changed.

Page 14: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Instrumental Factors

Unsatisfactory performance of an instrument may be caused by fluctuations in the power-supply voltage, an unstable light source, or a non-linear response of the detector-amplifier system.

Page 15: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Polychromatic Radiation

Strict adherence to Beer’s law is observed only with truly monochromatic radiation. This sort of radiation is only approached in specialized line emission sources.

All monochromators, regardless of quality and size, have a finite resolving power and therefore minimum instrumental bandwidth.

Page 16: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Polychromatic Radiation (cont)

A good picture of the effect of polychromatic radiation can be presented as follows. When radiation consists of two wavelengths, l and l1, and

assuming that Beer’s law applies at each of these individually the absorbance at l is given by:

log ( Po/P ) = A = abc

a Po/P = 10ebc

Page 17: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Stray Radiation

Stray light affects absorption measurements because stray radiation often differs in wavelength from that of the principal radiation and, in addition, may not have passed through the sample.

Page 18: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Stray Radiation (cont)

When measurements are made in the presence of stray radiation, the observed absorbance is given by:

A¢ = log( Po + Ps)/(P + Ps)

where Ps is the power of nonabsorbed stray

radiation.

Page 19: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Wave Properties of Electromagnetic Radiation

Visible light is a complex phenomenon that is classically explained with a simple model based on propagating rays and wavefronts, a concept first proposed in the late 1600s by Dutch physicist Christiaan Huygens. Electromagnetic radiation, the larger family of wave-like phenomena to which visible light belongs (also known as radiant energy), is the primary vehicle transporting energy through the vast reaches of the universe.

Page 20: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Electromagnetic RadiationThe term electromagnetic radiation, is derived from the characteristic electric and magnetic properties common to all forms of this wave-like energy, as manifested by the generation of both electrical and magnetic oscillating fields as the waves propagate through space.

Electromagnetic radiation is characterized by a broad range of wavelengths and frequencies, each associated with a specific intensity (or amplitude) and quantity of energy. An electromagnetic wave moves or propagates in a direction that is at right angles to the vibrations of both the electric and magnetic oscillating field vectors, carrying energy from its radiation source to undetermined final

destination.

Page 21: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Electromagnetic Wave

Page 22: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Photoelectric EffectThis effect shows how electrons with energy eVo, greater than a particular energy will cause emission of radiation from a metal. The work function is characteristic of the metal and the amount of excess energy determines the frequency of the radiation. For example, a freshly polished, negatively charged zinc plate looses its charge if it is exposed to ultraviolet light. This phenomenon is called the photoelectric effect.

Page 23: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Photoelectric Effect (cont)

Page 24: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Einstein's Explanation Light consists of particles (photons), and the energy of such a particle is proportional to the frequency of the light. There is a certain minimum amount of energy (dependent on the material) which is necessary to remove an electron from the surface of a zinc plate or another solid body (work function). If the energy of a photon is bigger than this value, the electron can be emitted. From this explanation the following equation results:

Ekin   =   h f   –   W

Page 25: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Blackbody Radiation

"Blackbody radiation" or "cavity radiation" refers to an object or system which absorbs all radiation incidents upon it and re-radiates energy which is characteristic of this radiating system only, not dependent upon the type of radiation which is incident upon it.

The radiated energy can be considered to be produced by standing wave or resonant modes of the cavity which is radiating.

Page 26: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Classical Prediction of Blackbody Radiation

Page 27: Introduction to Spectroscopic Methods Chapter 6 Instrumental Analysis

Referenceshttp://micro.magnet.fsu.edu/primer/java/wavebasics/

http://www.olympusmicro.com/primer/lightandcolor/electromaghome.html

http://www.people.vcu.edu/~srutan/chem409/pp22_40/tsld018.htm

http://www.chembio.uoguelph.ca/educmat/chm386/rudiment/tourexp/photelec.htm

http://hyperphysics.phy-astr.gsu.edu/hbase/mod6.html

http://theory.uwinnipeg.ca/physics/quant/node2.html