introduction and operation of sem simplified
TRANSCRIPT
-
8/10/2019 Introduction and Operation of SEM Simplified
1/42
Scanning Electron Microscopy(SEM)
-
8/10/2019 Introduction and Operation of SEM Simplified
2/42
Outline What can we use a SEM for?
How do we get an image?
Electron beam-sample interactions
Signals that can be used to characterize the
microstructure Secondary electrons
Backscattered electrons
X-rays
Components of the SEM Some comments on resolution
Summary
-
8/10/2019 Introduction and Operation of SEM Simplified
3/42
The most versatile instrument for a
materials scientist?
What can we study in a SEM?
Topography and morphology
Chemistry Crystallography
Orientation of grains
In-situ experiments: Reactions with atmosphere
Effects of temperature
Easy samplepreparation!!
Big samples!
-
8/10/2019 Introduction and Operation of SEM Simplified
4/42
Topography and morphology
High depth offocus
-
8/10/2019 Introduction and Operation of SEM Simplified
5/42
Depth of focus
Optical microscopy vs SEM
A SEM typically has orders of magnitude betterdepth of focus than a optical microscope making
SEM suitable for studying rough surfaces
The higher magnification, the lower depth of focus
Screw length: ~ 0.6 cm
Images: the A to Z of Materials
-
8/10/2019 Introduction and Operation of SEM Simplified
6/42
Ce
Fe Sr
Chemistry
-
8/10/2019 Introduction and Operation of SEM Simplified
7/42
In-situ imaging A modern SEM can be equipped with various
accessories, e.g. a hot stage
-
8/10/2019 Introduction and Operation of SEM Simplified
8/42
In-situ imaging: oxidation of steel at
high temperatures
800 C, pH2O = 667 Pa Formation of Cr2O3
Images: Anders W. B. Skilbred, UiO
2 min 10 min 90 min
-
8/10/2019 Introduction and Operation of SEM Simplified
9/42
How do we get an image?
In brief: we shoot high-energy electrons and
analyze the outcoming electrons/x-rays
Electrons inElectrons out
or: x-rays out
-
8/10/2019 Introduction and Operation of SEM Simplified
10/42
The instrument in brief
-
8/10/2019 Introduction and Operation of SEM Simplified
11/42
How do we get an image?
156 electrons!
Image
Detector
Electron gun
288 electrons!
-
8/10/2019 Introduction and Operation of SEM Simplified
12/42
Electron beam-sample interactions The incident electron beam is scattered in the sample,
both elastically and inelastically
This gives rise to various signals that we can detect(more on that on next slide)
Interaction volume increases with increasing accelerationvoltage and decreases with increasing atomic number
-
8/10/2019 Introduction and Operation of SEM Simplified
13/42
Signals from the sample
Incoming electrons
Secondary electrons
Backscattered
electrons
Auger electrons
X-rays
Cathodo-
luminescence (light)
Sample
-
8/10/2019 Introduction and Operation of SEM Simplified
14/42
Where does the signals come from?
Diameter of the interaction
volume is larger than the
electron spot
resolution is poorer than the
size of the electron spot
-
8/10/2019 Introduction and Operation of SEM Simplified
15/42
Secondary electrons (SE) Generated from the collision
between the incoming electronsand the loosely bonded outerelectrons
Low energy electrons (~10-50 eV)
Only SE generated close tosurface escape (topographicinformation is obtained)
Number of SE is greater than the
number of incoming electrons We differentiate between SE1 and
SE2
-
8/10/2019 Introduction and Operation of SEM Simplified
16/42
SE1
The secondary electrons that are
generated by the incoming electron beam
as they enter the surface
High resolution signal with a resolution
which is only limited by the electron beam
diameter
-
8/10/2019 Introduction and Operation of SEM Simplified
17/42
SE2 The secondary electrons that are generated by the
backscattered electrons that have returned to the surface
after several inelastic scattering events SE2 come from a surface area that is bigger than the spotfrom the incoming electrons resolution is poorer than forSE1 exclusively
Sample surface
Incoming electronsSE2
-
8/10/2019 Introduction and Operation of SEM Simplified
18/42
Factors that affect SE emission1. Work function of the surface
2. Beam energy and beam current Electron yield goes through a maximum at low acc.
voltage, then decreases with increasing acc. voltage
Incident electron energy / kV
Se
condary
ele
ctron
yield
-
8/10/2019 Introduction and Operation of SEM Simplified
19/42
Factors that affect SE emission
3. Atomic number (Z)
More SE2 are createdwith increasing Z
The Z-dependence is
more pronounced at
lower beam energies4. The local curvature of the
surface (the most
important factor)
-
8/10/2019 Introduction and Operation of SEM Simplified
20/42
High resolution image is possible
By placing thesecondary electron
detector inside the
lens, mainly SE1
are detected
Resolution of 1 2
nm is possible
-
8/10/2019 Introduction and Operation of SEM Simplified
21/42
Backscattered electrons (BSE)
A fraction of the incidentelectrons is retarded by
the electro-magnetic field
of the nucleus and if the
scattering angle is greaterthan 180 the electron can
escape from the surface
-
8/10/2019 Introduction and Operation of SEM Simplified
22/42
Backscattered electrons (BSE) High energy electrons (elastic scattering)
Fewer BSE than SE We differentiate between BSE1 and BSE2
-
8/10/2019 Introduction and Operation of SEM Simplified
23/42
BSE2
Sample surface
Incoming electrons
BSE2
Most BSE are of BSE2 type
-
8/10/2019 Introduction and Operation of SEM Simplified
24/42
BSE as a function of atomic number
For phases containing more than one element, it is the average
atomic number that determines the backscatter coefficient
-
8/10/2019 Introduction and Operation of SEM Simplified
25/42
Factors that affect BSE emission
Direction of the irritated surface more electrons will hit the BSE detector when
the surface is aligned towards the BSEdetector
Average atomic number
When you want to study differences in
atomic numbers the sample should be aslevelled as possible (sample preparation isan issue!)
-
8/10/2019 Introduction and Operation of SEM Simplified
26/42
BSE vs SE
-
8/10/2019 Introduction and Operation of SEM Simplified
27/42
X-rays
Photons not electrons
Each element has afingerprint X-ray signal
Poorer spatial resolution
than BSE and SE Relatively few X-ray signals
are emitted and the
detector is inefficient relatively long signalcollecting times are needed
-
8/10/2019 Introduction and Operation of SEM Simplified
28/42
X-rays
Most common spectrometer: EDS(energy-dispersive spectrometer)
Signal overlap can be a problem
We can analyze our sample in differentmodes
spot analysis
line scan chemical concentration map (elemental
mapping)
-
8/10/2019 Introduction and Operation of SEM Simplified
29/42
Considerations when using EDS
Dead time some twenty-thirty percent is ok
Statistics
Signal-to-noise ratio
Drift in electron beam with time
Build-up of a carbonaceous contamination
film after extended periods of electronprobe irradiation
-
8/10/2019 Introduction and Operation of SEM Simplified
30/42
Components of the instrument
electron gun (filament)
electromagnetic optics
scan coils
sample stage
detectors
vacuum system computer hardware and
software
-
8/10/2019 Introduction and Operation of SEM Simplified
31/42
Electron guns
We want many electrons per
time unit per area (high currentdensity) and as small electronspot as possible
Traditional guns: thermionic
electron gun (electrons areemitted when a solid is heated) W-wire, LaB6-crystal
Modern: field emission guns(FEG) (cold guns, a strongelectric field is used to extractelectrons) Single crystal of W, etched to a
thin tip
-
8/10/2019 Introduction and Operation of SEM Simplified
32/42
Electron guns With field emission guns we get a smaller spot
and higher current densities compared tothermionic guns
Vacuum requirements are tougher for a fieldemission guns
Single crystal of LaB6 Tungsten wire Field emission tip
-
8/10/2019 Introduction and Operation of SEM Simplified
33/42
Detectors
Secondary electron detector:
(Everhart-Thornley)
Backscattered electrondetector:
(Solid-State Detector)
-
8/10/2019 Introduction and Operation of SEM Simplified
34/42
Our traditional detectors
Secondary electrons: Everhart-ThornleyDetector
Backscattered electrons: Solid State
Detector X-rays: Energy dispersive spectrometer
(EDS)
-
8/10/2019 Introduction and Operation of SEM Simplified
35/42
Why do we need vacuum?
Chemical (corrosion!!) and thermal stabilityis necessary for a well-functioning
filament (gun pressure)
A field emission gun requires ~ 10-10
Torr LaB6: ~ 10
-6 Torr
The signal electrons must travel from the
sample to the detector (chamber pressure) Vacuum requirements is dependant of the
type of detector
-
8/10/2019 Introduction and Operation of SEM Simplified
36/42
Environmental SEM: ESEM
Traditional SEM chamber pressure:~ 10-6 Torr
ESEM: 0.08 30 Torr
Various gases can be used Requires different SE detector
-
8/10/2019 Introduction and Operation of SEM Simplified
37/42
Why ESEM?
To image challenging samples such as: insulating samples
vacuum-sensitive samples (e.g. biological samples)
irradiation-sensitive samples (e.g. thin organic films)
wet samples (oily, dirty, greasy)
To study and image chemical and physical
processes in-situ such as:
mechanical stress-testing oxidation of metals
hydration/dehydration (e.g. watching paint dry)
-
8/10/2019 Introduction and Operation of SEM Simplified
38/42
Quanta 200, FEI
Field emission gun, but no in-lens detector ESEM
Can be equipped with various accessories
to perform in-situ experiments (more onthis on next slide)
Images: FEI
-
8/10/2019 Introduction and Operation of SEM Simplified
39/42
Accessories on our Quanta 200:
GAD Gaseous Analytical Detector
for X-ray analysis in gaseous environments
GSED Gaseous Secondary Electron Detector 500 m aperture, allowing 20 Torr
chamber pressure
Hot stage GSED
Must be used at temperatures above 500 C
EBSD Electron Backscatter Diffraction
Grain orientation, grain and subgrainstructures, phase identification, micro textures
Hot stages 1000C and 1500C
ETD Everhart-Thornley Detector
Secondary electron detector
LFD Large Field Detector used in low vacuum and ESEM mode (SE)
SSD-BSD Solid State Backscattered Detector
Backscatter electrons
EDS Energy dispersive spectroscopy
X-ray analysis
Images: FEI
-
8/10/2019 Introduction and Operation of SEM Simplified
40/42
Some comments on resolution
Best resolution that can be obtained: size of the
electron spot on the sample surface The introduction of FEG has dramatically improved
the resolution of SEMs
The volume from which the signal electrons are
formed defines the resolution SE image has higher resolution than a BSE image
Scanning speed: a weak signal requires slow speed to improve signal-
to-noise ratio
when doing a slow scan drift in the electron beam canaffect the accuracy of the analysis
-
8/10/2019 Introduction and Operation of SEM Simplified
41/42
Summary
The scanning electron microscope is a versatileinstrument that can be used for many purposesand can be equipped with various accessories
An electron probe is scanned across the surface
of the sample and detectors interpret the signalas a function of time
A resolution of 1 2 nm can be obtained whenoperated in a high resolution setup
The introduction of ESEM and the field emissiongun have simplified the imaging of challengingsamples
-
8/10/2019 Introduction and Operation of SEM Simplified
42/42
Summary
Signals:
Secondary electrons (SE): mainlytopography Low energy electrons, high resolution
Surface signal dependent on curvature Backscattered electrons (BSE): mainly
chemistry High energy electrons
Bulk signal dependent on atomic number X-rays: chemistry
Longer recording times are needed