intro to electroanalytical chemistry

Upload: arun231187

Post on 07-Mar-2016

218 views

Category:

Documents


0 download

DESCRIPTION

Electroanalytical Chemistry

TRANSCRIPT

  • Nov 16, 2004Introduction to Electroanalytical ChemistryLecture Date: April 27h, 2008

  • Reading Material Skoog, Holler and Crouch: Ch. 22 (An Introduction to Electroanalytical Chemisty)

    See also Skoog et al. Chapters 23-25.

    Cazes: Chapters 16-19

    For those using electroanalytical chemistry in their work, the following reference is recommended:A. J. Bard and L. R. Faulkner, Electrochemical Methods, 2nd Ed., Wiley, 2001.

  • Advantages of Electroanalytical MethodsMatched against a wide range of spectroscopic and chromatographic techniques, the techniques of electroanalytical chemistry find an important role for several reasons:Electroanalytical methods are often specific for a particular oxidation state of an elementElectrochemical instrumentation is relatively inexpensive and can be miniaturizedElectroanalytical methods provide information about activities (rather than concentration)

  • History of Electroanalytical MethodsMichael Faraday: the law of electrolysisthe amount of a substance deposited from an electrolyte by the action of a current is proportional to the chemical equivalent weight of the substance.Walter Nernst: the Nernst equation (Nobel Prize 1920)Jaroslav Heyrovsky: the invention of polarography: (Nobel Prize 1959) Walter Nernst(1864-1941)Michael Faraday (1791-1867)Jaroslav Heyrovsky(1890-1967)

  • Main Branches of Electroanalytical ChemistryKey to measured quantity: I = current, E = potential, R = resistance, G = conductance, Q = quantity of charge, t = time, vol = volume of a standard solution, m = mass of an electrodispensed speciesInterfacial methodsBulk methodsStatic methods (I = 0)Dynamic methods (I > 0)Potentiometry (E)Conductometry(G = 1/R)Controlled potential Constant current Voltammetry (I = f(E))Amperometric titrations (I = f(E))Based on Figure 22-9 in Skoog, Holler and Crouch, 6th ed.Electro-gravimetry(m)Coulometric titrations(Q = It)

  • Main Branches of Electroanalytical ChemistryPotentiometry: measure the potential of electrochemical cells without drawing substantial currentExamples: pH measurements, ion-selective electrodes, titrations (e.g. KF endpoint determination)

    Coulometry: measures the electricity required to drive an electrolytic oxidation/reduction to completionExamples: titrations (KF titrant generation), chloridometers (AgCl)

    Voltammetry: measures current as a function of applied potential under conditions that keep a working electrode polarizedExamples: cyclic voltammetry, many biosensors

  • Electrochemical CellsZinc (Zn) wants to ionize more than copper (Cu).We can use this behavior to construct a cell:

  • e-e-Potentiometry: Measures equilibrium EAmperometry: Control E, measures I as function of timeCoulometry: Control E, measure total Q over a period of timecontrol measurementElectrochemical Cells and Analytical Methods

  • Electrochemical CellsGalvanic cell: a cell that produces electrical energy

    Electrolytic cell: a cell that consumes electrical energy

    Chemically-reversible cell: a cell in which reversing the direction of the current reverses the reactions at the two electrodes

  • Conduction in an Electrochemical CellElectrons serve as carriers (e.g. moving from Zn through the conductor to the Cu)In the solution, electricity involves the movement of cations and anionsIn the salt bridge both chloride and potassium ions moveAt the electrode surface: an oxidation or a reduction occursCathode: the electrode at which reduction occursAnode: the electrode at which oxidation occurs

  • Oxidation occurs when a chemical species loses an electron.LEO = lose electron is oxidation

    Reduction is when a species gains an electron.GER = gain an electron is reduction

    For example, the chemical reactioncan be decomposed into two half reactions:Leo the Lion Says Ger

  • Faradaic and Non-Faradaic CurrentsFigure 22-2Mass Transfer occurs by:ConvectionMigrationDiffusionFaradaic (governed by Faradays law): direct transfer of electrons, i.e. oxidation at one and reduction at the other electrodeNon-Faradaic: increasing charge of the double layer

  • FundamentalsElectrical charge, q, is measured in coulombs (C). The charge associated with chemical species is related to the number of moles through the Faraday constant, F=96,485.3 (~96,500) C/mole.

    Electrical current, I, is measured in Amperes (A). Current is the amount of charge that passes in a unit time interval (seconds).

    Ohm's law relates current to potential (E) through the resistance (R) of a circuit by E=IR. The potential is measured in Volts (V) and the resistance in Ohms ().

  • Power (P) is measured in Watts (W = J/s) and is related to the current and potential by P= IE.

    The work is measured in Joules (J) and is related to the potential and the amount of charge by work=q E.

    The relationship between the standard Gibb's free energy change, G (J/mole), and the standard electromotive force (EMF), E (V), is given by

    G=-n F E

    where n is the number of electrons transferred and superscript on E0 refers to standard state. Fundamentals

  • Fundamentals: The Nernst EquationThe Nernst equation gives the cell potential E (in volts):Q (the activity quotient) is the ratio of products over reactants as in equilibrium calculations. For the generic reaction:Q is given by:The As are activities. For low-concentration solutions (low ionic strengths):F = faraday (constant)n = # moles electrons in processE0 = standard potential for cell

  • Electrode PotentialsThe reactions in an electrochemical cell can be thought of as two half-cell reactions, each with its own characteristic electrode potentialThese measure the driving force for the reactionBy convention, always written as reductions

    Standard electrode potential (E0): the measure of individual potential of an electrode at standard ambient conditions (298K, solutes at a concentration of 1 M, and gas pressure at 1 bar).

  • Some Standard Electrode Potentials See appendix 3 in Skoog et al. for a more complete list

    ReactionE0 at 298K (Volts)Cl2(g) + 2e- 2 Cl2+1.359O2 (g) + 4H+ + 4e- 2 H2O+1.229Ag+ + e- Ag(s)+0.799Cu2+ + 2e- Cu(s)+0.337Hg2Cl2 + 2e- 2Hg(l) + 2 Cl2+0.2682H+ + 2e- H2 (g)0.000AgI(s) + e- Ag(s) + I2-0.151Cd2+ + 2e- Cd(s)-0.403Zn2+ + 2e- Zn(s)-0.763

  • The Standard Hydrogen Electrode (SHE)A universal reference, but is really a hypothetical electrode (not used in practice)Uses a platinum electrode, which at its surface oxidizes 2H+ to H2 gas.Very sensitive to temperature, pressure, and H+ ion activityBecause the SHE is difficult to make, the saturated calomel electrode (SCE) is used instead. Calomel = mercury (I) chloride

  • Q: What is the electrode potential for the electrode Ag/AgI(s)/I-(0.01 M) ?The overall reaction for this electrode isThis reaction cannot be found in tables of reduction potentials. But the reaction is comprised of two componentsElectrode Potentials

  • We can initially ignore the fact that the electrode contains AgI and find E for the silver ion reduction.Electrode Potentials

  • The Glass pH ElectrodeOne of the most common potentiometric measurements is pH (a so-called p-Ion measurement).

    The common glass pH electrode makes use of junction potentials to determine the hydronium ion concentration in a sample solution.

    A typical glass pH electrode is configured as shown here:

  • The glass pH electrode is used with a Ag/AgCl reference electrode. For most modern pH electrodes the reference electrode is incorporated with the pH indicator electrode.A small frit or hole connects the reference electrode and the sample solutionsThe Glass pH Electrode

  • pH MeasurementsA combination pH electrode combines the indicator and reference into a single unit.

    The potential of this cell is:

    where Eij and Eoj are the junction potentials at the inner and outer layers of the glass membrane.

    Junction potential: occurs at the interface of two electrolytes, caused by unequal diffusion rates of cation and anions across the boundary (e.g the frit in a salt bridge)

  • More About pH MeasurementsThe surface of the glass is hydrated, which allows exchange of hydronium ions for the cation in the glass (sodium or lithium). There are four interface regions, the external solution and hydrated glass, hydrated glass and dry glass on the outside, dry glass and hydrated glass on the inside, and hydrated glass and the internal solution. If the glass is uniform, the two hydrated glass/dry glass interfaces should be identical and should have the same junction potential. Since the glass interface junction potentials then cancel each other, the junction potential is then the difference between the internal and external solutions.

  • 0.05916pH loglogE thencomposition, fixed a hassolution internal the ifloga If the two solutions are identicallogloglog,',,',mem,',,',,,,,H,,,',,',,,,,,',,',,,++=+-=-==-=--=+++++++++++++kaFRTaFRTaaFRTEaaaaaFRTaaFRTaaFRTEinnsolHoutnsolHinnsolHoutnsolHmemoutglassHinglassoutglassHinnsolHoutnsolHinglassHoutglassHoutnsolHinnsolHinglassHmempH Measurements

  • pH MeasurementsFor a real electrode, the two surfaces will not be identical and the constant k needs to be determined experimentally. The constant k is termed the asymmetry potential. The constant is termed the electromotive efficiency.

  • pH Measurements

  • Q: Why does the pH change the interfacial potential of the glass/aqueous interface?A: The motion of the sodium ions leave behind a negatively charged glass layer that is neutralized to a lesser or greater extent according to the pH.More explanation about how a pH meter really works: The sodium ions must move through the dry part of the membrane and this process is slow. For this reason, the membrane is made very thin. Also, a nonperturbing (low-current) voltmeter is used to read the cell voltage so that only a few sodium ions must move through the dry glass in a given time period.pH Measurements

  • Errors in pH measurements with glass electrodes arise from the following effects:Calibration problems (e.g. drift, or error in the calibration)Junction potentialHigh [Na+] interacting with electrodeHigh acid concentrationEquilibration timeTemperature control

    Typical electrodes have the following performance:Accuracy = +/- 0.02 pH unitsPrecision = +/- 0.002 pH unitspH Electrodes: Errors, Accuracy and Precision

  • Modern pH electrodes are usually of the "combination" type, meaning that a single cylinder contains both the reference electrode, and a glass membrane electrode.Schematically, the total cell may be expressed as

    SCE//test solution ([H3O+]=a1)/glass membrane/[H3O+]=a2, Cl-/AgCl(s)/AgThe Combination pH Electrode

  • A Modern Combination pH Electrode

  • Electrochemical pH Measurements ConcludedConsider a typical problem related to the use of the combination pH electrode. Recall that Ecell = L - 0.0592 V pHQUESTION: If Ecell = -0.115 V at a pH of 4.00, what is the pH of a solution for which Ecell is -0.352 V?

  • ANSWER: First, find L from the measurement of the standard:-0.115 V = L -0.0592 x pH-0.115 V = L -0.0592 x 4.00Therefore, L = 0.122 VSecond, use this value of L to find pH:-0.352 V = 0.122 V - 0.0592 V x pHpH = (0.122 V -(-0.352 V))/0.0592pH = 7.84

  • QUESTION: What does the pH meter read if the pH is 7.00 in a 1 M salt solution having 1 M Na+ ions present?ANSWER: [H+]obs = 1 x 10-7 + 1 x 10-12 Conclusion -- the pH meter reads the true pH under these conditions.

  • The Ion Selective Electrode (ISE)An ISE generally consists of the ion-selective membrane, an internal reference electrode, an external reference electrode, and a voltmeter.Example: an ISE for fluoride (F-)

  • Automatic pKa and log P DeterminationpKa (ionization constant) and log P (octanol/water partition) are important physical parameters that play critical roles in determining how compounds behave in physiological environments and how they interact with enzymes, receptors and cell membranesThe Sirius GLpKa system:combination pH electrodesample trayreagentsliquid dispensors

  • ConductometryConductometry: Detection of electrical conductivityKey analytical applications: conductometric detection in ion-exchange chromatography (IEC or IC) and capillary electrophoresis (CE)

    Used to detect titration endpoints

  • Homework Problems (for Study Only)Chapter 22:22-1

    Chapter 23:23-11

    Potentiometric titration endpoint determination has several advantages, including applicability to turbid solutions. KF uses bipotentiometric endpoint determination with Pt electrodes.

    See figure 22-1 pg 229 in Skoog et al. for more discussion of this cell.

    The salt bridge is a tube filled with an electrolyte (e.g. KCl). The salt bridge isolates the contents of the two cells but permits electrical contact between them.Potentiometry, amperometry, and coulometry can be understood via the basic electrochemical cell shown here (voltammetry requires a more in-depth discussion).Note that Q is not a constant but changes continuously until equilibrium is reached (and Q becomes K, the equilibrium constant). The activity values used to calculate Q are instantaneous.The overall potential of a cell can be measured as we have shown early in this lecture. However, there is no simple way to measure individual isolated electrode potentials. Complications also arise since the potential varies with temperature, concentration and pressure.

    The oxidation potential of a half-reaction is just the negative of the reduction potential in a redox reaction, so it is sufficient to calculate either one of the potentials. The standard electrode potential is thus commonly written as standard reduction potential (by convention).