international quality assurance programme (iqap ... · sample 1 sm-1 was a blank test sample of...
TRANSCRIPT
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)
INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Summary ReportSEIZED MATERIALS
2014/2
Samples Substances Concentrations Comments on substancesSample 1 [blank sample]
Sample 2 3,4-Methylenedioxymetamfetamine (MDMA) 34.9 %
Lactose -
Sample 3 Metamfetamine 7.7 %
Lactose - %
Sample 4 Cocaine 7.1 %
Lactose - %
Test Samples Information
Samples Comments on samplesSample 1 SM-1 was a blank test sample of glucose and contained no substances in the ICE menu.
Sample 2 SM-2 was prepared from a seizure containing 34.9 % (w/w) MDMA base. The test sample also contained lactose.
Sample 3 SM-3 was prepared from a seizure containing 7.7 % (w/w) metamfetamine base. The test sample also contained lactose.
Sample 4 SM-4 was prepared from a seizure containing 7.1 % (w/w) cocaine base. The test sample also contained lactose. Upon concentration, trace amounts of cinnamoylcocaine isomers and benzoylecgonine were detected.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
2
Table of contents Introduction Page 4 Comments from the International Panel of Forensic Experts Page 4 NPS reported by ICE participants Page 5
Codes and Abbreviations Page 8 Sample 1 Analysis Page 9 Identified substances Page 9
Statement of findings Page 10 Identification methods Page 15
Summary Page 20 Sample 2 Analysis Page 21 Identified substances Page 21
Statement of findings Page 25 Identification methods Page 34
Summary Page 39 Z-Scores Page 40 Sample 3 Analysis Page 42
Identified substances Page 42 Statement of findings Page 46 Identification methods Page 54 Summary Page 59
Z-Scores Page 60 Sample 4 Analysis Page 62
Identified substances Page 62 Statement of findings Page 66 Identification methods Page 74 Summary Page 79
Z-Scores Page 80
This report contains the data received from laboratories participating in the current exercise. The results compiled in this report are not intended to be an overview of the quality of work and cannot be interpreted as such. These comments do not reflect the general state of the art within the profession.
Participant results are reported using a randomly assigned "WebCode". This code maintains participant's anonymity, provides linking of the various report sections, and will change with every report.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
3
Introduction An important element of the UNODC International Quality Assurance Programme (IQAP) is the implementation of the International Collaborative Exercises (ICE). The exercises allow laboratories, from both developing and developed countries, to continuously monitor their performance in drug testing on a truly global scale. This report provides information on analytical results of laboratories participating in the Seized Materials (SM) group. In order to maintain confidentiality, the participating laboratories have been assigned random “Web Codes”, which change every round. The analytical results returned by laboratories participating in ICE are evaluated by UNODC and a confidential report is provided to each laboratory on its own performance. The overall analytical results are reviewed by the UNODC’s International Panel of Forensic Experts which oversees the implementation of these exercises, and offers guidance and support in addressing relevant quality issues. The exercises provide an overview of the performance and capacity of participating laboratories and enable UNODC to tailor technical support in the laboratory sector for greatest impact.
Comments from the International Panel of Forensic Experts
Participation of Laboratories In the 2014/2 round of the ICE programme, results were submitted within both the Seized materials (SM) and Biological Specimens (BS) test groups by 190 laboratories in 61 countries. Within the SM test group, there were 157 participating laboratories from 57 countries and within the BS test group, results were submitted by 71 laboratories from 39 countries. Qualitative Analysis The analytical technique most commonly used for screening of test samples in the SM test group was the marquis reagent (45% of participants), while GC-MS (91%) was the most commonly used technique for identification/confirmation of the components in the test samples followed by FTIR (51%). The results for the qualitative identification of the controlled substances were excellent for all test samples in the SM group as shown in the table 1.
Table 1. Qualitative performance of participants in the 2014/2 round of ICE.
Test sample SM-1
(Blank) SM-2
(MDMA) SM-3
(Metamfetamine) SM-4
(Cocaine) Correct identification
by participants N/A 99% 98% 99%
With regard to the false positive results for substances in the ICE menu there were a total of ten, (two each for SM-1 and SM-3, five for SM-2 and one for SM-4. There were a total of four false negative results in the three SM test samples containing controlled substances. While the overall number of false positive/negative results reported by ICE participants is encouragingly low, Laboratories reporting false positive or false negative results should investigate the reasons for this and corrective actions should be taken in order to continuously improve performance. Quantitative Analysis The proportion of participants who carried out quantitative analysis, 97 (64%) for SM-2, 89 (57%) for SM-3 and 111 (71%) for SM-4 continues to be encouraging, particularly the high percentage of participants who quantified the cocaine in SM-4. It is also encouraging to note that 115 (73%) of participants in the SM test group performed quantitation and of these, 85 (74%) quantified all three test samples. It is understood that quantitative analysis of controlled substances is often governed by local legislation and the requirements of prosecution. With regard to the analytical techniques used by participants performing quantitation, 54% used GC-FID, 25% used HPLC and 17% used GC-MS. z-scores obtained by participants in quantification are shown in the table 2.
4
Table 2. Quantitative performance of participants in the 2014/2 round of ICE.
z-score Test sample SM-2 (MDMA) SM-3 (Metamfetamine) SM-4 (Cocaine)
|z| < 2, satisfactory 88% 83% 79% 2 ≤ |z| < 3, questionable 5% 4% 8%
|z| ≥ 3, unsatisfactory 7% 12% 13% According to the recommendations in ISO 13528:2005, an unsatisfactory z-score is considered to give an action signal and a questionable z-score is considered to give a warning signal. A single action signal or warning signals in two successive rounds shall be taken that an anomaly has occurred that requires investigation. Participants with z-scores outside acceptable limits should review their quantification procedures. In the 2014/1 round of ICE, eighteen laboratories obtained questionable z-scores with 2 ≤ |z| < 3. Of these laboratories, fifteen (83%) obtained acceptable z-scores with |z| < 2 in the current 2014/2 round of ICE. Twenty laboratories obtained unsatisfactory z-scores in ICE 2014/1 with |z| ≥ 3, and of these, ten (50%) improved their performance in quantification in the current 2014/2 ICE round. Laboratories are encouraged to carry out quantification as it can improve the quality of laboratory analyses and can be helpful in assessing the significance of the results. z-score values obtained assist participating laboratories in evaluating the quality of their results and should be monitored over time as part of each laboratory’s quality assurance programme.. Laboratories reporting false positive or false negative results should investigate the reasons for this and corrective actions should be taken in order to continuously improve performance. Participation in the ICE programme also helps in monitoring the effect of corrective actions. UNODC would like to acknowledge the valuable contribution of the Chemical Metrology Laboratory of the Health Sciences Authority, Singapore, for the provision of specific software used for the quantitative statistical calculations in the ICE programme. New Psychoactive Substances (NPS) reported by laboratories participating in the 2014/2 round of the ICE programme During the 2014/2 round of ICE, participating laboratories provided information on NPS that had been identified in their laboratories. In total there were 231 reports of 109 substances. The most commonly reported substances were synthetic cannabinoids (31%) followed by synthetic cathinones (27%) and phenethylamine group substances (22%) as shown in figure 1. This is the same trend that was reported in the 2014/1 round of ICE.
5
Figure 1. NPS reported by ICE laboratories during the 2014/2 round of ICE
NPS most commonly reported by laboratories participating in the 2014/2 round of the ICE programme The substances most often reported by ICE participants for the second round in succession were the phenethylamines, 25I-NBOMe and 25C-NBOMe. These were followed by the synthetic cathinone 3-methylmethcathinone (3-MMC) and the two synthetic cannabinoids, 5F-AKB-48 and BB-22 as shown in table 3.
Table 2. NPS most commonly reported by ICE participants in the 2014/2 round of ICE
Substance IUPAC name Structure
25I-NBOMe 2-(4-iodo-2,5-dimethoxyphenyl)-N-(2-methoxybenzyl)ethan-1-amine
25C-NBOMe 2-(4-chloro-2,5-dimethoxyphenyl)-N-(2-methoxybenzyl)ethan-1-amine
3-MMC 2-(methylamino)-1-(m-tolyl)propan-1-one
31%
22%
27%
3%
13%
0% 3% 1%
Synthetic cannabinoids
Phenethylamines
Synthetic cathinones
Tryptamines
Others
Piperazines
Ketamine and phencyclidine typesubstances
Aminoindanes
6
5F-AKB-48 N-(adamantan-1-yl)-1-(5-fluoropentyl)-1H-indazole-3-carboxamide
BB-22 quinolin-8-yl 1-(cyclohexylmethyl)-1H-indole-3-carboxylate
ICE participants can obtain further details of all the NPS reported to UNODC by logging on to the UNODC ICE portal and selecting the icon for the Early Warning Advisory on New Psychoactive Substances
7
Codes and Abbreviations
Tech Code Name140 Colorimetric reactions
141 Marquis reagent (sulphuric acid, formaldehyde)
142 Cobalt thiocyanate
150 Thin Layer Chromatography
160 High/Ultra High Performance Liquid Chromatography
161 High Performance Liquid Chromatography with diode array detection
170 Gas Chromatography NPD
171 Gas Chromatography FID
172 Gas Chromatography ECD
180 Gas Chromatography/Mass Spectrometry
190 Fourier Transform Infrared Spectrometry
200 Spectrophotometry (visible, UV)
210 Others (specify)
211 NMR
220 Microcrystal test
230 Liquid chromatography/mass spectrometry
231 Liquid chromatography/tandem mass spectrometry
(+) “positive”: Indicates that the analyte is identified; for presumptive tests (e.g. colour reactions), indicates that a positive reaction was obtained.
(-) “negative”: Indicates that the analyte is not identified.
(ANP): Analysis not performed.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
8
Identified substances for Test Sample 1
Code Identified Substances
1P11K1 No Narcotics
424PPI none
43N3MI no drugs
5MKKKW no drugs detected
9OXPXM No
9UJ6IU Lactose
A35PZP DEXTROSE
APAA1A Sample containing no drugs
BMNPI1 2,3-Dihydroxy Propanal
BZZH9I No
C77GE8 No
CKSNPG None
ESOODT NONE
H2UF1T Negative
HQPTHH NONE
KNRKIJ -
OKLOIL Lactose
OMI8O8 NONE
PGOVBE No controlled drugs detected.
PRAXXW STARCH
PYGYE0 no
QLPFFF NONE
QLRWCG NIL
QVV1O8 NO DRUGS
RBJEMN no
ROOGOL psychotropic substance
TVD1KB SM1 was a blanck teste sample, containing no substances in the ICE menu.
UHIEE2 -
VBAPZ2 There is no drug in the sample
VVANL2 NO
WZEYRI None
XE8WIX Nothing
Y00O0U no
YMW8P3 none
ZZF97Z -
Sample 1 Analysis
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
9
Code [blank sample]0D8I70 We found glucose and traces of MDA
1P11K1 The white powder were found to contain No Narcotics.
1WVSJ1 Only glucose is present in this sample.
25A55I
2NQOPT
3OT6ZK not detected substances by method GC
424PPI glucose (identified by ATR and XRD)
43N3MI
47X44P Lactosa
4S9S9W The powder was analysed for the presence of drugs controlled within the provisions of the Misuse of Drugs Act 1971 with a negative result.
555QL5 De los análisis realizados a la muestra SM-1 se concluye que no se detectó presencia de droga
5MKKKW glucose by FT-IR, no findings by GC-MS
6LMHPM
6TMAIV white powder identified as dextrose
75C556 No dangerous drug detected in sample SM-1
7AAFAT A representative sample of the powder was analysed in order to detect the presence of a range of controlled drugs but none were found.
7CBVXP dextrose
7FFRIU
7ZJLKU In the test sample 2014/2/SM-1 was detected dextrose
8SI888 D-Glucose
8SZGPS The white powder was analyzed for controlled substances with a negative result.
9OXPXM On analysis, I found the substance to contain glucose.
9UJ6IU The sample contains lactose. No controlled drugs detected.
A35PZP No Controlled Drugs Detected, indicated the presence of dextrose
ADCNJN This sample was analysed and found to be comprised essentially of glucose
AFX7LD
ANZFXA Controlled substances as Cocaine, Opiates, Amphetamines nor Benzodiazepines were not detected in the sample of white powder analyzed. It was detected to contains Dextrose.
APAA1A
APXA8I
AWAXYQ
BDBQZB The sample contains dextrose
BFDEKH Glucose
BMNPI1 No reaction with Marquis test,Simon Test,Chan Kao's Test. Analysis by GCMS shows the similarity as 2,3-Dihydroxy Propanal which is not notified in our country.
BTSAZE We find in this sample glucose.
BVFV4R GLUCOSE
Statement of findings for Test Sample 1
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
10
Code [blank sample]BYCO9B Sample do not contain illegal or controlled drugs
BZZH9I Glucose
C77GE8 Preliminary screening tests with Marquis, Simon, Fast Blue B, Wagner Test and Cobalt Thiocyanate tests are negative.
C8CHVI Dextrose confirmed in one unit tested of one unit received.
CI8BFC
CKSNPG Negative white powder.
CRY93L
CZPRCC FTIR spectra analysis indicated the presence of glucose (dextrose). Gas chromatography indicated only glucose side products. Sample totally soluble in water.
D2EDTC
DE1ABE
DGKC1G The sample 2014/2/SM-1 does not contain any drug in the menu of ICE programme detected by GC/MS and FTIR.
DPJL4D Glucose and lactose have been detected.
E5598C SM1 contains Glucose.
EIJCK6 GLUCOSE
EQY9D9 no substance was found in this sample
ERARRR FT-IR analysis identificates the white crystalline powder as dextrose monohydrate; it doesn't contain common active principles
ESOODT
EXD7VB No controlled drugs were detected
FAYSEJ
FLXW44 The sample was found to contain glucose.
FS2S09 Sample is a sugar (glucose)
G7SO8D sugars-succrose
GCSVZA Test sample 2014/2/SM-1 does not contain any drugs.
GQTNTT The test sample 2014/2/SM-1 do not found to contain any controlled drugs. The sample was found to contain lactose.
GRF2Z2
GRVGEG No controlled substances were found.
H2UF1T Negative
HPSK6E SM-1 consists of glucose
HQPTHH We have not found any controlled drug in the SM 1 (6)
HXHRHH blank
HYFRMY
IWUUS3
JGWYWW The sample was found to contain glucose.
JPMMF7
JQQQQZ
JY6WCO negative
K1TOYR In the sample was identified glucose.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
11
Code [blank sample]KCFEK5 No controlled drugs detected
KCKKUB DEXTROSE
KF2FY2
KFFFFP Negative for Controlled Drugs
KKCYOY the sample contains only glucose
KNRKIJ -
KR3ZKK The sample does not include any psychoactive substances.
KU2BUC The sample contains glucose (cutting agent). It's not scheduled as illicit drug in France.
LGA9TV
LYKKAN NO NARCOTIC DRUGS OR PSYCHOTROPIC SUBSTANCES DETECTED IN THE SAMPLE SM-1
M9HNLP Based on the above analysis, dangerous drugs were not detected in the sample. The sample was found to contain glucose.
MKKFFY Negative result on Color tests, no drugs peaks on GC and GCMS.With FTIR Analysis, confirmed that the sample contains Dextrose. No Quantification analysis has been made.
MSHOBP was detected Glucose
MUQPYK Dextrose monohydrate
MWBGCG The sample contains glucose.
N2TFCN
NH4NBU The sample doesn't contain any drug.
NI5F66 Se identificó dextrosa monohidratada
O5RDHB
OE3YFV no controlled drugs were detected
OEIEEC Glucose
OKLOIL The sample 2014/2/SM-1 was analysed by chemical test and TLC technique and found to contain lactose
OMI8O8 Sample 2014/2/SM-1 was found to contain no drugs
ON4A8J No controlled drugs detected.
OOYONL The sample contains fructose
OTSLPP Presence of glucose was detected in the sample.
PCGCQB
PFTCFW glucose
PGOVBE No controlled drugs were detected in this powder which was found to contain dextrose.
PRAXXW As a result of the controllet Laboratory tests, we conclude that the test sample 2014/2/SM-1, which is a white substance in dust form is starch. This substance is not a narchotic drug and not illegal in Turkey under the Turkish Drug Law known as the law 2313.
PY333P Test sample 2014/2/SM-1 contains glucose monohydrate.
PYGYE0 no detected substances by method GC/MS
QHSMDY Sample 2014/2/SM-1 consists of white powder. No prohibited substances are detected in this sample.
QLEZZZ
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
12
Code [blank sample]QLPFFF White substance that contains NO DRUGS used as reference samples. It represent some sugar
that we couldn't identify because out FTIR instrument is out of function.QLRWCG The sample 2014/2/SM-1was found to contain No Narcotic Drug.
QUD1AR with glucose
QVV1O8 LACTOSA, ALOSA
QWEMWS
RBJEMN dextrose
RDJ3M3 Nil
ROOGOL
RSPZ60 No substances controlled by the Misuse of Drugs Act 1971 (as amended) were detected in this sample.
RU2VD6 Any drug of abuse has been detected
S12Q5E One glass bottle containing 0.22 grammes of a powder which did not contain any substances controlled under the Dangerous Drugs Ordinance.
SFUVOJ The analysis perfomed didn't highlight any controled substance.
T60TMT the powder is glucose
TMARDF The analysis did not reveal any controlled drug according to french order, decree of February 22nd 1990.
TTBYYX THE SAMPLE CONTAINS GLUCOSE
TTEC2L
TVD1KB The following tests were conducted: two colorimetric tests, SCOTT and MARQUIS, failed to detect in the case of SM1 sample the presence of any substances present in the ICE. A Thin-Layer chromatography TLC, they were used as a solvent and eluent respectively, Dichloromethane and Methanol, using a silica plate of 20 × 20 cm, on the plate were placed the following samples of references (MDMA; Ketamine; Amfetamine; Mephedrone; Metamfetamine; MDEA; Cafeine; Cocaine; Procaine Penicillin; Lidocaine; Nimetazepam; m-Cpp HCl. After Elution, UV light was used to 265 nm in order to show what substances were separated. In the case of SM1 sample failed to show any substance present in the ICE. A qualitative analysis was performed with the GC/FID, using a column TRACSIL TRB-50, a Flow rate of 1.5 ml/min, FID Detector, the pressure of 14.7 psi, the oven temperature of 250ºc, the temperature for the injector and detector were of 300 C and 330 C respectively. The chromatograms were obtained using the KONIFROM PLUS Software and For the case of SM1 sample nor couldn't identify any peak on the menu of the ICE, concluding that the sample SM1 can be a blank, does not have any substance ICE menu.
U6UXCE white powder not containing any common active principles of drugs
UHIEE2 We not found contain controlled drugs in SM-1.But we found Dextrose.
UJJQFA
UP37JV
UV6OBH Sample one containes dextrose
VBAPZ2 In the sample 1 there is a glucose which is not on the list of drugs
VIKHII D-GLUCOSE HYDRATE WAS IDENTIFIED
VMCYPO
VNZMCG On this sample, Thin Layer Chromatography (TLC), Gas Chromatography / Mass Spectrometry (GC/MS), Fourier Transform Infrared Spectrometry (FTIR), Gas Chromatography (GC/FID) analitycal techniques was applied. We found that this sample does not include drugs.
VVANL2 SIGN IN 2014/2 / SM-1 NOT DETECTED DRUGS.
VXMWHA
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
13
Code [blank sample]W88A8P There are no drugs on powder.
WRGGF7
WRHBIA White powder containing probably glucose. Glucose has not been confirmed due to the lack of glucose standard
WZEYRI *Results of screening tests were negatives for amfetamines, cocaine and opiates. * Spectrocopic parameters obtained with sample are equivalent to those obtained with dextrose.
X5PP2A colorimetric test resulted negative for controlled durgs. GC-FID identified no retention time for controled drugs.
X9GLI9 test sample 14/2/SM-1 contains: Sucrose, L-glucose
XBBWXW tha sample contains glucose
XBNTS9 We didnt find any drug that GC/MS can detect.
XE8WIX Sample SM1: is negative
XOOOLO No controlled drugs were detected. Results indicate the presence of glucose
XRRRRJ contains controlled substances
XTXX6H
XVCTKI probably sugar
XVXXXX Dextrose
Y00O0U no drugs
Y2PTPQ The sample SM-1 contains Dextrose.
YLTLG9 This sample does not contain drugs.
YMW8P3 marquis Reagent (MR), Cobalt thiocyanate (CT) negative. GC/MS undetected. FTIR absotion bands coincidents with the substances used as standard or pattern (Glucose)
YQXIF0
ZB002K white powder not containing any controlled drug
ZIIUIH
ZX2A9A no controlled drugs were identified in the test sample
ZZF97Z recommended color tests show negative.Melting point is 99.6'C.It's confirmed by TLC & GCMS shows also negative for notified drugs and not match with ICE 2/2014 reference standards.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
14
Identification methods for Test Sample 1
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
0D8I70 SI I
S S I I
1P11K1 I I S
1WVSJ1 I I
25A55I SI
2NQOPT SI
3OT6ZK S S SI I I
424PPI I SI ATR
43N3MI S S SI SI
47X44P S I I
4S9S9W S S I
555QL5 S S I I IMS
5MKKKW I SI SI
6LMHPM SI
6TMAIV SI SI
75C556 S S I SI
7AAFAT S S I I
7CBVXP S S S I I
7FFRIU S S I
7ZJLKU I I
8SI888 S S S I SI I
8SZGPS S S I I
9OXPXM S I I
9UJ6IU S S S I I
A35PZP S S I I S SIMONS
ADCNJN SI SI I I I GC/MS derivitisation
AFX7LD S I I
ANZFXA S S S SI I
APAA1A I SI
APXA8I S S S I
AWAXYQ S S S SI I SI I
BDBQZB S S I I
BFDEKH S S S I
BMNPI1 S S S I I Mayer test
Analytical Technique Codes
Legend: S ... used for Screening I ... used for Identification SI ... used for both
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
15
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
BTSAZE S S S I I
BVFV4R S I
BYCO9B SI ASAP-TOF
BZZH9I S SI SI
C77GE8 S S I I S Fast Blue B
C8CHVI SI I I
CI8BFC SI I
CKSNPG S I SI SI
CRY93L S S S I
CZPRCC I I S
D2EDTC SI I
DE1ABE S I
DGKC1G I I SI I
DPJL4D S I I I I Fehling
E5598C I S I I I
EIJCK6 S S S I I I
EQY9D9 I I I I ion scan
ERARRR I SI
ESOODT S S I
EXD7VB SI
FAYSEJ S I I
FLXW44 SI I LC/ELSD
FS2S09 S I S
G7SO8D S I I
GCSVZA S S S I I S Chen-Kao Reagent
GQTNTT S S S I I
GRF2Z2 S S S I I SI
GRVGEG S S I
H2UF1T SI I
HPSK6E S S S I I I SI I Tollens reagent
HQPTHH S I
HXHRHH S SI I I SI
HYFRMY I SI I
IWUUS3 S S I I
JGWYWW
S I S
JPMMF7 S SI SI I
JQQQQZ S S I I S Tanred
JY6WCO I SI
K1TOYR I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
16
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
KCFEK5 S SI
KCKKUB SI I
KF2FY2 I SI
KFFFFP S S S I
KKCYOY I SI I
KNRKIJ SI
KR3ZKK SI S I S
KU2BUC S S I I
LGA9TV SI I
LYKKAN S S S I I
M9HNLP S I S Simon's Reagent
MKKFFY S S S I I I
MSHOBP SI SI I I I RAMAN
MUQPYK I SI SI
MWBGCG
SI I
N2TFCN S SI SI I
NH4NBU S I SI I
NI5F66 SI S S I I
O5RDHB SI I
OE3YFV I S I SI
OEIEEC S I I I D-RX
OKLOIL SI I SI I lead acetate+ammonia test
OMI8O8 S I I
ON4A8J S S I I
OOYONL I I
OTSLPP SI I I I
PCGCQB I SI SI
PFTCFW I I S ion mobility spectrometry
PGOVBE S S I I
PRAXXW S I I
PY333P I S I HPLC/RI
PYGYE0 I SI I
QHSMDY S S I I I
QLEZZZ I SI
QLPFFF S S SI I
QLRWCG S S S I I
QUD1AR SI SI
QVV1O8 I SI
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
17
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
QWEMWS
S S S I I
RBJEMN SI SI
RDJ3M3 S S S I I
ROOGOL S I I
RSPZ60 S S S I I
RU2VD6 SI I I
S12Q5E S S SI
SFUVOJ SI I
T60TMT I I SI I
TMARDF I I
TTBYYX S I I I
TTEC2L S S I
TVD1KB S S S I I
U6UXCE I SI
UHIEE2 S I S
UJJQFA S S I I I
UP37JV SI I
UV6OBH S S SI I I
VBAPZ2 I SI SI
VIKHII I SI I I XRD
VMCYPO I I S I dérivatisation+GCMS
VNZMCG S I I
VVANL2 S S I
VXMWHA I SI SI
W88A8P S S I I S Simon
WRGGF7 SI
WRHBIA SI
WZEYRI S S I I
X5PP2A SI SI SI I
X9GLI9 SI
XBBWXW S I I I
XBNTS9 S S I S Dunqueis levin
XE8WIX I I S immunochromatography
XOOOLO S S I I
XRRRRJ S S I I
XTXX6H S I I
XVCTKI SI I GC-IRD
XVXXXX I SI SI
Y00O0U SI SI SI S I I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
18
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
Y2PTPQ S S I I S TANRED
YLTLG9 S I I
YMW8P3 S S I I
YQXIF0 I SI SI
ZB002K I SI
ZIIUIH I SI
ZX2A9A SI S I I SI mecke
ZZF97Z S S S I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
19
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Frequency of use of screening and identification techniques:
Summary for test sample 1The response summary of all 157 participating laboratories for the identification of SM-1, frequency of use of screening andidentification techniques and false positives and negatives:
Identification of test sample 1:
False positives for Test Sample 1
Code Name Technique
Response SummaryParticipants: 157
Blank test sample
Response Summary Participants: 130
0D8I70 MDA 150/180
ROOGOL MDEA 150/190
Section 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231
Screening 49 71 58 19 0 2 0 3 56 21 2 0 0 0 0
31.0% 44.9% 36.7% 12.0% 0.0% 1.3% 0.0% 1.9% 35.4% 13.3% 1.3% 0.0% 0.0% 0.0% 0.0%
Identification 12 5 4 41 6 12 1 29 143 74 2 3 1 3 1
7.6% 3.2% 2.5% 25.9% 3.8% 7.6% 0.6% 18.4% 90.5% 46.8% 1.3% 1.9% 0.6% 1.9% 0.6%
# labs using this technique
55 71 59 49 6 13 1 30 144 80 3 3 1 3 134.8% 44.9% 37.3% 31.0% 3.8% 8.2% 0.6% 19.0% 91.1% 50.6% 1.9% 1.9% 0.6% 1.9% 0.6%
Response Summary Participants: 157
210
13
8.2%
17
10.8%
2717.1%
20
Identified substances for Test Sample 2
Code Identified Substances
0D8I70 MDMA - lactose
1P11K1 3,4-Methylenedioxymetamfetamine (MDMA)
1WVSJ1 MDMA,HCl
25A55I MDMA
2NQOPT MDMA
3OT6ZK MDMA
424PPI MDMA HCl
43N3MI 3,4-Methylenedioxymetamfetamine (MDMA)
47X44P MDMA
4S9S9W MDMA
555QL5 MDMA
5MKKKW MDMA
6LMHPM mdma
6TMAIV MDMA
75C556 3,4-Methylenedioxy-metamfetamine (MDMA)
7AAFAT MDMA
7CBVXP 3,4 Methyllenedioxymethamphetamine
7FFRIU MDMA
7ZJLKU MDMA
8SI888 MDMA
8SZGPS Methylenedioxymethylamphetamine (MDMA)
9OXPXM 3,4-methylenedioxymethamphetamine (MDMA)
9UJ6IU MDMA
A35PZP METHYLENEDIOXYMETHAMPHETAMINE (MDMA)
ADCNJN MDMA
AFX7LD 3,4-methylenedioxy-metamfetamine (MDMA)
ANZFXA 3,4-Methylenedioxy-metamfetamine (MDMA)
APAA1A MDMA
APXA8I MDMA
AWAXYQ Trace MDA and METHYL EPHEDRINE
BDBQZB MDMA
BFDEKH MDMA
BMNPI1 MDMA (Ecstasy)
BTSAZE MDMA
BVFV4R 3,4-Methylenedioxymetamfetamine (MDMA)
BYCO9B MDMA
BZZH9I MDMA
C77GE8 MDMA
C8CHVI 3,4-MDMA
Sample 2 Analysis
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
21
Code Identified SubstancesCI8BFC MDMA
CKSNPG MDMA
CRY93L 3,4 - Methylenedioxymetamfetamine (MDMA)
CZPRCC MDMA
D2EDTC MDMA
DE1ABE MDMA
DGKC1G MDMA and Lactose
DPJL4D MDMA
E5598C MDMA
EIJCK6 MDMA
EQY9D9 MDMA
ERARRR D,L-3,4-MDMA
ESOODT 3,4-Methylenedioxymetamfetamine (MDMA)
EXD7VB MDMA
FAYSEJ MDMA
FLXW44 MDMA
FS2S09 MDMA
G7SO8D MDMA
GCSVZA 3,4-Methylenedioxymetamfetamine (MDMA)
GQTNTT MDMA
GRF2Z2 3,4-Methylenedioxy-metamfetamine (MDMA) 32,4 %
GRVGEG MDMA
H2UF1T MDMA
HPSK6E MDMA
HQPTHH MDMA3,4-Methylenedioxymetamfetamine
HXHRHH MDMA
HYFRMY MDMA
IWUUS3 3,4 Methylenedioxy-metamfetamine (MDMA)
JGWYWW 3,4-Methylenedioxy-metamfetamine (MDMA)
JPMMF7 MDMA
JQQQQZ 3,4-Methylenedioxymetamfetamine (MDMA)
JY6WCO 3,4-Methylenedioxymetamfetamine (MDMA)
K1TOYR MDMA
KCFEK5 MDMA
KCKKUB MDMA
KF2FY2 MDMA
KFFFFP MDMA
KKCYOY MDMA,HCl
KNRKIJ MDMA
KR3ZKK MDMA
KU2BUC 3,4-Methylenedioxymetamfetamine (MDMA), lactose
LGA9TV MDMA
LYKKAN 3,4-Methylenedioxymetamfetamine (MDMA)
M9HNLP 3,4-Methylenedioxymethamphetamine (MDMA)
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
22
Code Identified SubstancesMKKFFY 3,4-Methylenedioxymethamphetamine Hydrochloride (MDMA HCl)
MSHOBP 3,4-Methylenedioxymetamfetamine (MDMA)
MUQPYK 3,4-Methylenedioxymetamfetamine (MDMA)
MWBGCG MDMA
N2TFCN MDMA
NH4NBU 3,4-MDMA
NI5F66 3,4-Metilenedioximetanfetamina (MDMA)
O5RDHB mdma
OE3YFV MDMA
OEIEEC MDMA
OKLOIL MDMA
OMI8O8 MDMA
ON4A8J MDMA
OOYONL MDMA
OTSLPP 3,4-methylenedioxy-metamfetamine (MDMA)
PCGCQB MDMA
PFTCFW MDMA
PGOVBE Methylenedioxymethylamphetamine Hydrochloride
PRAXXW 3,4-Methylenedioxymetamfetamine (MDMA)
PY333P MDMA
PYGYE0 MDMA
QHSMDY 3,4-Methylenedioxy-metamfetamine (MDMA)
QLEZZZ MDMA
QLPFFF MDMA
QLRWCG 3,4-Methlenedioxy-metamfetamine (MDMA)
QUD1AR MDMA Chlorhydrate
QVV1O8 3,4- METHYLENEDIOXY-METANFETAMINE (MDMA)
QWEMWS MDMA, MDEA
RBJEMN MDMA
RDJ3M3 MDMA
ROOGOL psychotropic substance
RSPZ60 MDMA HCl
RU2VD6 MDMA
S12Q5E MDMA
SFUVOJ MDMA
T60TMT MDMA
TMARDF MDMA
TTBYYX 3,4 METHYLENE DIOXYMETAMPHETAMINE
TTEC2L MDMA
TVD1KB MDMA
U6UXCE MDMA
UHIEE2 3,4-Methylenedioxymetamfetamine (MDMA)
UJJQFA MDMA hydrochloride
UP37JV MDMA
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
23
Code Identified SubstancesUV6OBH MDMA
VBAPZ2 MDMA
VIKHII MDMA
VMCYPO MDMA
VNZMCG MDMA
VVANL2 MDMA AND COCAINE.
VXMWHA 3,4-methylenedioxy-metamphetamine
W88A8P MDMA
WRGGF7 MDMA
WRHBIA MDMA
WZEYRI 3,4-Methylenedioxymetamfetamine (MDMA)
X5PP2A MDMA
X9GLI9 MDMA
XBBWXW MDMA
XBNTS9 MDMA
XE8WIX MDMA
XOOOLO MDMA (methylenedioxymethylamphetamine)
XRRRRJ MDMA
XTXX6H 3,4-Methylenedioxymetamfetamine (MDMA)
XVCTKI MDMA (3,4-metylendioximethamfetamine)
XVXXXX MDMA
Y00O0U MDMA
Y2PTPQ 3,4-Methylenedioxymetamphetamine
YLTLG9 MDMA
YMW8P3 N-Methyl-3,4-methylenedioxiamphetamine
YQXIF0 MDMA
ZB002K MDMA
ZIIUIH MDMA
ZX2A9A Heroin and MDEA
ZZF97Z N-Methyl-3,4-methylenedioxyamphetamine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
24
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
0D8I70 + +
1P11K1 + +The white powder were found to contain 29.40% of 3,4-Methylenedioxymetamfetamine (MDMA).
1WVSJ1 + +MDMA,HCl is present (35.3% as base) with lactose
25A55I + ANP
2NQOPT + -MDMA
3OT6ZK + +containet MDMA and lactose
424PPI + +additional substance: lactose
43N3MI + ANP
47X44P + -
4S9S9W + ANPThe powder was analysed and found to contain Methylenedioxymethylamphetamine (also known as MDMA).Methylenedioxymethylamphetamine (MDMA) is a compound structurally derived from an N-Alkyl-alpha-methylphenethylamine by substitution in the ring with an alkylenedioxy substituent, and as such is a class A controlled drug within the provisions of the Misuse of Drugs Act 1971.
555QL5 + ANPDe los análisis realizados a la muestra SM-2 se concluye que hay presencia de MDMA
5MKKKW + -No more findings by GC-MS
6LMHPM + ANPMDMA was detected by GC/MS screening and it was quantified by GC/FID using calibration curve
6TMAIV + +white powder - mixture of the MDMA and lactose
75C556 + -38.65% +/- 3.09% 3,4-Methylenedioxy-metamfetamine (MDMA) detected in sample SM-2
7AAFAT + ANPOn analysis of a representative sample of the powder MDMA (Ecstasy) was detected with a purity of approximately 30%calculated as base.
7CBVXP + +WAS FOUND MDMA AND LACTOSE
7FFRIU + -In analysis, the substance found to contain 38.45% MDMA
7ZJLKU + -In the test sample 2014/2/SM-2 was detected MDMA
Statement of findings for Test Sample 2
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
25
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
8SI888 + +Lactose
8SZGPS + -The white powder was analyzed and found to contain Methylenedioxymethyamphetamine (MDMA).
9OXPXM + +On analysis, I found the substance to contain 36.4 % (w/w) of 3,4-Methylenedioxymethamphetamine (MDMA) base.
9UJ6IU + -The sample contains MDMA (31.9%).
A35PZP + +METHYLDIOXYMETHAMPHETMINE (MDMA) DETECTED, ALSO INDICATED THE PRESENCE OF LACTOSE. AVERAGE PURITY OF DRUG: 33% (CALCULATED AS BASE)
ADCNJN + +This sample was analysed and foudn to contain MDMA at a purity of 36%. This sample was also found to be comprised of lactose and glucose
AFX7LD + +
ANZFXA + +Presence of 3,4-Methylenedioxy-metamfetamine (MDMA) and Lactose, was detected and confirmed in the white powder sample analyzed.
APAA1A + ANPPowder containing MDMA
APXA8I + ANP
AWAXYQ + -
BDBQZB + +The sample contains 3,4-methylenedioxy-metamfetamine and lactose
BFDEKH + -In the sample analyzed 3,4-methylenedioxy-methamphetamine was found
BMNPI1 + -Black color with Marquis Test, Blue color with Simon test and violet color changing to green indicates possible presence of MDMA.Analysis by GCMS shows similarity as N-Methyl-3,4-Methylenedioxyamphetamine ( Ecstasy).
BTSAZE + +In this sample contain two substances MDMA and lactose
BVFV4R + -MDMA 32.68%
BYCO9B + ANPSample contain MDMA (3,4-methylenedioxy-N-methylamphetamine)
BZZH9I + ANPN-Methyl-3,4-methylenedioxyamphetamine
C77GE8 + ANPScreening test with Simon gives Royal Blue. Marquis gives dark blue. TLC shows RF and color are the same as those MDMA standard. GC MS shows m/z and RT are the same as those of MDMA standard. Therefore, the compound is MDMA.
C8CHVI + +3,4-MDMA hydrochloride and lactose confirmed in one unit tested of one unit received.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
26
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
CI8BFC + -
CKSNPG + ANPWhite powder containing MDMA.
CRY93L + ANPThe Sample obtains with the positive result with Marquis reagent.TLC: The Rf sample obtain with the same Rf MDMA Standard.GC: The Rt sample obtain with the same Rt MDMA Standard.
CZPRCC + +Retention time and MS spectra in GC analysis (FID and MS) compared with reference material. Sample totally soluble in water. MDMA as hydrochloride.
D2EDTC + ANPMDMA
DE1ABE + -
DGKC1G + +The sample 2014/2/SM-2 contains MDMA and Lactose detected by GC/MS and FTIR. The percentage by weight of MDMA is 30.3%.
DPJL4D + -Substance commonly detected in our laboratory.
E5598C + +SM2 contains MDMA HCl with 34.54% MDMA base and Lactose.
EIJCK6 + +MDMA 36.1% AS BASE, LACTOSE
EQY9D9 + ANPthis sample was spiked with MDMA
ERARRR + ANPWhite powder containing MDMA (35.5%) which is controlled by national law
ESOODT + ANP
EXD7VB + ANPThe sample was found to contain methylenedioxymethylamphetamine.
FAYSEJ + ANPMDMA existed in salt form.
FLXW44 + +The sample was found to contain 33.1% of MDMA (as base)
FS2S09 + ANPMDMA is controlled by law (KB 22/1/1998)
G7SO8D + -
GCSVZA + ANPTest sample 2014/2/SM-2 contains 3,4-Methylenedioxymetamfetamine (MDMA).
GQTNTT + -The test sample 2014/2/SM-2 was found to contain the controlled drug 3,4-Methylenedioxymetamfetamine (MDMA).
GRF2Z2 + -
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
27
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
GRVGEG + +Sample contains MDMA hydrochloride (20.3% content as base)
H2UF1T + -MDMA
HPSK6E + -SM-2 contains 3,4-methylendioxymetamfetamine (MDMA)
HQPTHH + -The result of the test of the SM 2 (5)received confirmed that the mentioned sample contains 3,4-Methylenedioxymetamfetamine but we have not been able to give a cuantification result becaus we have not got the pattern of the mentioned controlled drug.
HXHRHH + +MDMA + lactose
HYFRMY + -
IWUUS3 + +The chemical test and the coincidence in the time of retencion and mass fragmentogram with the material reference used in the same conditions that the analyzed samples, they indicate in this one the 3,4 Methylenedioxy-metamfetamine (MDMA)presence.
JGWYWW + -The sample was found to contain 37.99% 3,4-Methylenedioxy-metamfetamine (MDMA).
JPMMF7 + +lactose
JQQQQZ + +The sample contains MDMA
JY6WCO + ANP
K1TOYR + +The sample SM2 contained 34.5 % MDMA calculated as a base. In the sample was identified lactose.
KCFEK5 + ANPExamination revealed the presence of MDMA
KCKKUB + +MDMA and LACTOSE
KF2FY2 + +
KFFFFP + ANPThe white powder contains MDMA (3,4-Methylenedioxymetamfetamine)
KKCYOY + +the sample contains MDMA chlorhydrate, a controlled drug
KNRKIJ + ANP-
KR3ZKK + +The sample contains MDMA (3,4-methylenedioxymethamfetamine) at 40% purity. Also it includes lactose as cutting agent.
KU2BUC + +The sample contains 3,4-Methylenedioxymetamfetamine (MDMA) which is scheduled as illicit drug in France. The lactose is not scheduled as illicit drug in France (cutting agent).
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
28
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
LGA9TV + -
LYKKAN + ANPTHE SAMPLE SM-2 WAS ANALYSED AND FOUND TO TO CONTAIN 3,4-Methylenedioxymetamfetamine (MDMA) A PSYCHOTROPIC SUBSTANCES LISTED IN SECOND SCHEDULE OF THE NARCOTIC DRUGS AND PSYCHOTROPIC SUBSTANCES CONTROL ACT CAP 245
M9HNLP + -Based on the above analysis, the sample was found to contain 3,4-Methylenedioxymethamphetamine (MDMA).
MKKFFY + ANPOn Marquis test, gives dark blue color and turn to black. Confirmed MDMA Positive on TLC and GCMS. Chloride test positive with AgNO3. Quantification with GC, the sample contains MDMA base 30.63%
MSHOBP + +3,4-Methylenedioxymetamfetamine (MDMA) and Lactose
MUQPYK + +lactose
MWBGCG + +The sample contains MDMA and lactose.
N2TFCN + +the sample also contain lactose
NH4NBU + +The sample contains 30,9±1.4 % percentage by weight of 3,4-MDMA. The sample also contains lactose.
NI5F66 + +La muestra contiene 3,4-Metilenedioximetanfetamina y lactosa
O5RDHB + +
OE3YFV + +Screening by FTIR and identification/confirmation by GC/MS
OEIEEC + +BASE MDMA 33.63% + LACTOSE
OKLOIL + -The sample 2014/2/SM-2 was analysed by colour test and TLC technique and found to contain drugs MDMA
OMI8O8 + -Sample 2014/2/SM-2 was found to contain MDMA
ON4A8J + -MDMA is the common name for Methylenedioxymethylamphetamine. It is a Class A controlled drug subject to the provisions of the Misuse of Drugs Act 1971 as it is structurally derived from an alpha-methylphenethylamine by substitution in the ring with an alkoxy substituent.
OOYONL + +
OTSLPP + +Presence of 3,4-methylenedioxy-metamfetamine hydrochloride and lactose was detected in the sample.
PCGCQB + +
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
29
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
PFTCFW + +powder with 30,7 % of MDMA base and lactose
PGOVBE + +This powder contained an average of 39% methylenedioxmethylamphetamine hydrochloride.
PRAXXW + +As a result of the controllet Laboratory tests, we conclude that the test sample 2014/2/SM-2, which is a white substance in dust form is a mixture of two different substances.The first ingrediant of the mixture is 3,4-Methylenedioxymetamfetamine (MDMA) and the second ingrediant of the mixture is Lactose. The first ingrediant is a narchotic drug and is illegal in Turkey under the Turkish Drug Law known as the law 2313 but the second ingrediant of the subtance is not a narchotic drug and is not illegal in Turkey.
PY333P + +Test sample 2014/2/SM-2 contains lactose and MDMA 34.40 %, m/m.
PYGYE0 + +contained MDMA and lactose
QHSMDY + +Sample 2014/2/SM-2 consists of off-white powder. 3,4-Methylenedioxy-metamfetamine (MDMA), a prohibited substance included in the National Law for Narcotics, is found to be present in this sample.
QLEZZZ + -
QLPFFF + ANPWhite substance that contains drug MDMA.
QLRWCG + ANPThe sample 2014/2/SM-2 was found to contain 3,4-Methlenedioxy-metamfetamine (MDMA).
QUD1AR + +lactse
QVV1O8 + -THE SAMPLE No. 2014/2/SM2 WAS FOUND METHYLENEDIOXY-METANFETAMINE (MDMA)
QWEMWS + +
RBJEMN + -3,4-Methylenedioxymethamfetamine
RDJ3M3 + -On analysis, i found the sample to "2014/2/SM-2" contain MDMA.
ROOGOL + ANP
RSPZ60 + -The powder contains an average of 33% by weight of MDMA HCl calculated as the base.
RU2VD6 + +It has been detected the presence of 3,4-methylendioxymetamphetamine in the sample
S12Q5E + ANPOne glass bottle containing 0.22 grammes of a powder containing 0.08 grammes of 3,4-methylenedioxymethamphetamine.
SFUVOJ + +The sample contains MDMA, also known as ecstasy
T60TMT + +the powder contains MDMA with a concentration in the sample of 34.4% by weight. It also contains lactose
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
30
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
TMARDF + +The sample contains MDMA which is listed as a controlled drug according to french order, appendix III of February 22nd 1990.
TTBYYX + +THE SAMPLE CONTAINS MDMA HYDOCHLORIDE AND LACTOSE
TTEC2L + ANP
TVD1KB + ANPThe following tests were conducted: two colorimetric tests, SCOTT and MARQUIS, were able to detect if the SM2 sample the presence of MDMA. A Thin-Layer chromatography TLC, they were used as a solvent and eluent respectively, Dichloromethane and Methanol, using a silica plate of 20 × 20 cm, on the plate were placed the following samples of references (MDMA; Ketamine; Amfetamine; Mephedrone; Metamfetamine; MDEA; Cafeine; Cocaine; Procaine Penicillin; Lidocaine; Nimetazepam; m-Cpp HCl. After Elution, UV light was used to 265 nm in order to show what substances were separated. In the case of SM2 sample managed to detect the presence of MDMA. A qualitative analysis was performed with the GC/FID, using a column TRACSIL TRB-50, a Flow rate of 1.5 ml/min, FID Detector, the pressure of 14.7 psi, the oven temperature of 250ºc, the temperature for the injector and detector were of 300 C and 330 C respectively. The chromatograms were obtained using the KONIFROM PLUS Software and For the case of SM2 sample managed to identify a peak of MDMA. In sumula for all three tests have been applied in the case of SM2 sample results were all identical or is all they gave positive for MDMA.
U6UXCE + ANPwhite powder containing 32.2%w/w of MDMA
UHIEE2 + ANPWe found 3,4-Methylenedioxymetamfetamine (MDMA) in SM-2 as controlled drugs
UJJQFA + +the presence of drug confirmed by GC-MS and FTIR analysis, the MDMA hydrochloride confirmed by the anion analysis and FTIR
UP37JV + -Found MDMA with 29.0%w/w
UV6OBH + +MDMA hidrochloride and lactose
VBAPZ2 + +In the sample 2 we have found MDMA.
VIKHII + +MDMA IS PRESENT IN HYDROCHLORIDE FORM. LACTOSE WAS ALSO IDENTIFIED.
VMCYPO + +The powder 2014/2/SM2 contains 31% MDMA base lactose and glucose. MDMA is a controlled substance in France.
VNZMCG + -On this sample, Thin Layer Chromatography (TLC), Gas Chromatography / Mass Spectrometry (GC/MS), Fourier Transform Infrared Spectrometry (FTIR), Gas Chromatography (GC/FID) analitycal techniques was applied. We found that this sample contains 25% MDMA.
VVANL2 + ANPSIGN IN 2014/2 / SM-2 MDMA DETECTED AND COCAINE.
VXMWHA + +
W88A8P + +There is 32% of MDMA in powder (calculated as MDMA base).
WRGGF7 + ANP
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
31
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
WRHBIA + +White powder containing MDMA and probably lactose. Lactose has not been confirmed due to the lack of lactose
WZEYRI + ANP* Results of screening tests were positive for secondary amfetamines. * Retention time and fragmentation pattern obtained with sample are equivalent to those obtained with 3,4-methylenedioxymetamfetamine (MDMA).
X5PP2A + ANPMarquis reagent turned into black and Simon reagent for secundaria anfetamine derivates, turned into blue, so both tests were found positive. GC-FID identified MDMA retention time.
X9GLI9 + ANPMDMA WAS AS HYDROCHLORIDE FORM.
XBBWXW + +the sample contains mixture of MDMA and lactose
XBNTS9 + ANP
XE8WIX + ANPSample no SM2 contains 56.35% of MDMA.
XOOOLO + ANPThe powder was found to contain MDMA at a concentration of 34 percent. MDMA is controlled by the Misuse of Drugs Act 1971 (as amended) as a Class A drug, by virtue of the fact that it is structurally derived from an N-alkyl-α-methylphenethylamine by substitution in the ring with an alkylenedioxy substituent
XRRRRJ + +
XTXX6H + +Also contain lactose
XVCTKI + ANPNarcotics: MDMA (3,4-metylendioximethamfetamine), approximately 33% as free base
XVXXXX + +Lactose
Y00O0U + -MDMA
Y2PTPQ + +The sample SM-2 contains 3,4-Methylenedioxymetamphetamine and Lactose.
YLTLG9 + +This sample is MDMA
YMW8P3 + +Marquis Reagent (MR) positive, GC/MS. Retention time and mass charge ratio (M/Z) similar Characteristics with the substances used as Standard or pattern. FTIR absortion bands coincidents with the substances used as standard or pattern (Lactose)
YQXIF0 + +lactose
ZB002K + ANPwhite powder containing MDMA
ZIIUIH + -33.28%
ZX2A9A - ANPThe test sample contained amixture of heroin and MDEA
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
32
Code 3,4-Methylenedioxymetamfetamine (MDMA)
Lactose
ZZF97Z + -Simon's reagent test shows blue color and Marquis Reagent test show blackish grey. Melting Point is 123.5'C. It's confirmed by TLC & GCMS.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
33
Identification methods for Test Sample 2
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
0D8I70 SI I
S S I I
1P11K1 I I S
1WVSJ1 I I
25A55I SI
2NQOPT SI
3OT6ZK S S SI I I
424PPI I SI ATR
43N3MI S S SI SI
47X44P S I I
4S9S9W S S I
555QL5 S S I I IMS
5MKKKW I SI SI
6LMHPM SI
6TMAIV SI SI
75C556 S S I SI
7AAFAT S S I I
7CBVXP S S S I I
7FFRIU S S I
7ZJLKU I I
8SI888 S S S I SI I
8SZGPS S S I I
9OXPXM S I I
9UJ6IU S S S I I
A35PZP S S I I S SIMONS
ADCNJN SI SI I I I GC/MS derivitisation
AFX7LD S I I
ANZFXA S S S SI I
APAA1A I SI
APXA8I S S S I
AWAXYQ S S S SI I SI I
BDBQZB S S I I
BFDEKH S S S I
BMNPI1 S S S I I Mayer test
Analytical Technique Codes
Legend: S ... used for Screening I ... used for Identification SI ... used for both
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
34
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
BTSAZE S S S I I
BVFV4R S I
BYCO9B SI ASAP-TOF
BZZH9I S SI SI
C77GE8 S S I I S Fast Blue B
C8CHVI SI I I
CI8BFC SI I
CKSNPG S I SI SI
CRY93L S S S I
CZPRCC I I S
D2EDTC SI I
DE1ABE S I
DGKC1G I I SI I
DPJL4D S I I I I Fehling
E5598C I S I I I
EIJCK6 S S S I I I
EQY9D9 I I I I ion scan
ERARRR I SI
ESOODT S S I
EXD7VB SI
FAYSEJ S I I
FLXW44 SI I LC/ELSD
FS2S09 S I S
G7SO8D S I I
GCSVZA S S S I I S Chen-Kao Reagent
GQTNTT S S S I I
GRF2Z2 S S S I I SI
GRVGEG S S I
H2UF1T SI I
HPSK6E S S S I I I SI I Tollens reagent
HQPTHH S I
HXHRHH S SI I I SI
HYFRMY I SI I
IWUUS3 S S I I
JGWYWW
S I S
JPMMF7 S SI SI I
JQQQQZ S S I I S Tanred
JY6WCO I SI
K1TOYR I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
35
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
KCFEK5 S SI
KCKKUB SI I
KF2FY2 I SI
KFFFFP S S S I
KKCYOY I SI I
KNRKIJ SI
KR3ZKK SI S I S
KU2BUC S S I I
LGA9TV SI I
LYKKAN S S S I I
M9HNLP S I S Simon's Reagent
MKKFFY S S S I I I
MSHOBP SI SI I I I RAMAN
MUQPYK I SI SI
MWBGCG
SI I
N2TFCN S SI SI I
NH4NBU S I SI I
NI5F66 SI S S I I
O5RDHB SI I
OE3YFV I S I SI
OEIEEC S I I I D-RX
OKLOIL SI I SI I lead acetate+ammonia test
OMI8O8 S I I
ON4A8J S S I I
OOYONL I I
OTSLPP SI I I I
PCGCQB I SI SI
PFTCFW I I S ion mobility spectrometry
PGOVBE S S I I
PRAXXW S I I
PY333P I S I HPLC/RI
PYGYE0 I SI I
QHSMDY S S I I I
QLEZZZ I SI
QLPFFF S S SI I
QLRWCG S S S I I
QUD1AR SI SI
QVV1O8 I SI
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
36
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
QWEMWS
S S S I I
RBJEMN SI SI
RDJ3M3 S S S I I
ROOGOL S I I
RSPZ60 S S S I I
RU2VD6 SI I I
S12Q5E S S SI
SFUVOJ SI I
T60TMT I I SI I
TMARDF I I
TTBYYX S I I I
TTEC2L S S I
TVD1KB S S S I I
U6UXCE I SI
UHIEE2 S I S
UJJQFA S S I I I
UP37JV SI I
UV6OBH S S SI I I
VBAPZ2 I SI SI
VIKHII I SI I I XRD
VMCYPO I I S I dérivatisation+GCMS
VNZMCG S I I
VVANL2 S S I
VXMWHA I SI SI
W88A8P S S I I S Simon
WRGGF7 SI
WRHBIA SI
WZEYRI S S I I
X5PP2A SI SI SI I
X9GLI9 SI
XBBWXW S I I I
XBNTS9 S S I S Dunqueis levin
XE8WIX I I S immunochromatography
XOOOLO S S I I
XRRRRJ S S I I
XTXX6H S I I
XVCTKI SI I GC-IRD
XVXXXX I SI SI
Y00O0U SI SI SI S I I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
37
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
Y2PTPQ S S I I S TANRED
YLTLG9 S I I
YMW8P3 S S I I
YQXIF0 I SI SI
ZB002K I SI
ZIIUIH I SI
ZX2A9A SI S I I SI mecke
ZZF97Z S S S I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
38
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Frequency of use of screening and identification techniques:
Summary for test sample 2The response summary of all 157 participating laboratories for the identification of SM-2, frequency of use of screening andidentification techniques and false positives and negatives:
Identification of test sample 2:
False positives for Test Sample 2
Code Name Technique
VVANL2 Cocaine 180
ZX2A9A Heroin, MDEA 150/210 (Mecke)
Response Summary Participants: 131 Section 141 150
Response Summary Participants: 130
AWAXTQ MDA 161/180
QWEMWS MDEA 150/180
False negatives for SM-2 (techinque(s) used or ANP)
Code: ZX2A9A (141/150)
Section 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231
Screening 49 71 58 19 0 2 0 3 56 21 2 0 0 0 0
31.0% 44.9% 36.7% 12.0% 0.0% 1.3% 0.0% 1.9% 35.4% 13.3% 1.3% 0.0% 0.0% 0.0% 0.0%
Identification 12 5 4 41 6 12 1 29 143 74 2 3 1 3 1
7.6% 3.2% 2.5% 25.9% 3.8% 7.6% 0.6% 18.4% 90.5% 46.8% 1.3% 1.9% 0.6% 1.9% 0.6%
# labs using this technique
55 71 59 49 6 13 1 30 144 80 3 3 1 3 134.8% 44.9% 37.3% 31.0% 3.8% 8.2% 0.6% 19.0% 91.1% 50.6% 1.9% 1.9% 0.6% 1.9% 0.6%
Response Summary Participants: 157
210
13
8.2%
17
10.8%
2717.1%
Response SummaryParticipants: 157
3,4-Methylenedioxymetamfetamine (MDMA):
156
Lactose: 73
39
Z-Scores for Test Sample 2
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Concentration
(mass%)
Deviation
from robust
average
z-score
XBBWXW 2.3 -30.7 -9.0
LYKKAN 14.3 -18.7 -5.5
QWEMWS 19.6 -13.4 -3.9
GRVGEG 20.3 -12.7 -3.7
25A55I 23.5 -9.5 -2.8
VNZMCG 25.0 -8.0 -2.3
KCKKUB 25.7 -7.3 -2.1
PRAXXW 27.0 -6.0 -1.7
ROOGOL 27.3 -5.7 -1.7
0D8I70 28.4 -4.6 -1.3
OMI8O8 28.7 -4.3 -1.2
VXMWHA 28.7 -4.3 -1.2
UP37JV 29.0 -4.0 -1.2
DE1ABE 29.0 -4.0 -1.1
RU2VD6 29.3 -3.7 -1.1
1P11K1 29.4 -3.6 -1.0
43N3MI 29.8 -3.2 -0.9
7AAFAT 30.0 -3.0 -0.9
KU2BUC 30.0 -3.0 -0.9
XVXXXX 30.0 -3.0 -0.9
WRHBIA 30.2 -2.9 -0.8
2NQOPT 30.3 -2.7 -0.8
DGKC1G 30.3 -2.7 -0.8
MKKFFY 30.6 -2.4 -0.7
PFTCFW 30.7 -2.3 -0.7
NH4NBU 30.9 -2.1 -0.6
TTEC2L 30.9 -2.1 -0.6
VMCYPO 31.0 -2.0 -0.6
6LMHPM 31.2 -1.8 -0.5
7ZJLKU 31.2 -1.8 -0.5
H2UF1T 31.5 -1.5 -0.4
KNRKIJ 31.7 -1.3 -0.4
9UJ6IU 31.9 -1.1 -0.3
PYGYE0 31.9 -1.1 -0.3
8SI888 32.0 -1.0 -0.3
MWBGCG 32.0 -1.0 -0.3
Y00O0U 32.0 -1.0 -0.3
C8CHVI 32.2 -0.8 -0.2
U6UXCE 32.2 -0.8 -0.2
3OT6ZK 32.2 -0.8 -0.2
MUQPYK 32.4 -0.6 -0.2
W88A8P 32.4 -0.6 -0.2
GRF2Z2 32.4 -0.6 -0.2
BVFV4R 32.7 -0.3 -0.1
CZPRCC 32.8 -0.2 0.0
APXA8I 32.9 -0.1 0.0
VIKHII 32.9 -0.1 0.0
A35PZP 33.0 0.0 0.0
PCGCQB 33.0 0.0 0.0
Code
MDMA
Concentration (mass%)
Deviation from robust
averagez-score
PGOVBE 33.0 0.0 0.0RSPZ60 33.0 0.0 0.0FLXW44 33.1 0.1 0.0FS2S09 33.1 0.1 0.0XVCTKI 33.2 0.2 0.1ZIIUIH 33.3 0.3 0.1OEIEEC 33.6 0.6 0.2HYFRMY 33.7 0.7 0.2DPJL4D 33.8 0.8 0.3JY6WCO 33.9 0.9 0.3XOOOLO 34.0 1.0 0.3APAA1A 34.1 1.1 0.3QLEZZZ 34.2 1.2 0.4X5PP2A 34.4 1.4 0.4PY333P 34.4 1.4 0.4T60TMT 34.4 1.4 0.4K1TOYR 34.5 1.5 0.5E5598C 34.5 1.5 0.5SFUVOJ 34.7 1.7 0.56TMAIV 35.0 2.0 0.6KKCYOY 35.0 2.0 0.6O5RDHB 35.0 2.0 0.6XBNTS9 35.0 2.0 0.6424PPI 35.2 2.2 0.71WVSJ1 35.3 2.3 0.7UV6OBH 35.4 2.4 0.7ERARRR 35.5 2.5 0.7ADCNJN 35.6 2.6 0.8UJJQFA 35.7 2.7 0.8YLTLG9 35.7 2.7 0.8HPSK6E 35.8 2.8 0.8FAYSEJ 35.9 2.9 0.9OTSLPP 35.9 2.9 0.9EIJCK6 36.1 3.1 0.9S12Q5E 36.2 3.2 1.09OXPXM 36.4 3.4 1.0LGA9TV 36.5 3.5 1.0OOYONL 36.9 3.9 1.2JGWYWW 38.0 5.0 1.5KCFEK5 38.0 5.0 1.57FFRIU 38.5 5.5 1.675C556 38.7 5.7 1.7TMARDF 39.2 6.2 1.8KR3ZKK 40.0 7.0 2.1X9GLI9 41.9 8.9 2.6CRY93L 49.7 16.7 4.9XE8WIX 56.4 23.4 6.8BMNPI1 65.0 32.0 9.4
Code
MDMA
40
Substance: SM-2/3,4-Methylenedioxymetamfetamine (MDMA)
Round: 2014/2
Z-Score Report per Sample per Substance
‐3.7
‐2.8
‐2.3‐2.1
‐1.7 ‐1.7
‐1.3 ‐1.2
‐1.2 ‐1.2
‐1.1 ‐1.1 ‐1.0 ‐0.9 ‐0.9
‐0.9
‐0.9 ‐0.8 ‐0.8
‐0.8 ‐0.7
‐0.7 ‐0.6
‐0.6 ‐0.6 ‐0.5
‐0.5 ‐0.4 ‐0.4 ‐0.3
‐0.3 ‐0.3
‐0.3
‐0.3 ‐0.2
‐0.2
‐0.2 ‐0.2
‐0.2
‐0.2 ‐0.1 0.0 0.0
0.0
0.0
0.0
0.0
0.0 0.0
0.0 0.1 0.1 0.2
0.2 0.3 0.3 0.3 0.3 0.4 0.4
0.4
0.4 0.5
0.5 0.5 0.6
0.6
0.6
0.6 0.7 0.7 0.7 0.7 0.8 0.8
0.8 0.8 0.9
0.9 0.9 1.0 1.0 1.0 1.2
1.5
1.5 1.6 1.7 1.
82.1
2.6
‐4.0
‐3.0
‐2.0
‐1.0
0.0
1.0
2.0
3.0
4.0
XBBW
XWLYKK
ANQWEM
WS
GRVG
EG25
A55I
VNZM
CGKC
KKUB
PRAX
XWRO
OGOL
0D8I70
OMI8O8
VXMWHA
UP3
7JV
DE1A
BERU
2VD6
1P11K1
43N3M
I7A
AFAT
KU2B
UC
XVXX
XXWRH
BIA
2NQOPT
DGKC
1GMKK
FFY
PFTC
FWNH
4NBU
TTEC
2LVM
CYPO
6LMHP
M7ZJLKU
H2UF1T
KNRK
IJ9U
J6IU
PYGY
E08SI888
MWBG
CGY00O
0UC8
CHVI
U6U
XCE
3OT6ZK
MUQPYK
W88A8
PGR
F2Z2
BVFV
4RCZPR
CCAP
XA8I
VIKH
IIA3
5PZP
PCGCQ
BPG
OVB
ERSPZ60
FLXW
44FS2S09
XVCTKI
ZIIUIH
OEIEEC
HYFRM
YDP
JL4D
JY6W
COXO
OOLO
APAA
1AQLEZZZ
X5PP
2APY33
3PT60TMT
K1TO
YRE559
8CSFUVO
J6TMAIV
KKCYOY
O5R
DHB
XBNTS9
424P
PI1W
VSJ1
UV6
OBH
ERAR
RRAD
CNJN
UJJQ
FAYLTLG9
HPSK6E
FAYSEJ
OTSLPP
EIJCK6
S12Q
5E9O
XPXM
LGA9
TVOOYO
NL
JGWYW
WKC
FEK5
7FFRIU
75C5
56TM
ARDF
KR3ZKK
X9GL
I9CR
Y93L
XE8W
IXBM
NPI1
‐9.0 ‐5.5
‐3.9
9.4
6.8
4.9
z-scores can be interpreted by participants in line with ISO 13528:2005, section 7.4.2 and ISO/IEC guide43-1:1997(E) as follows:
|z| < 2 = satisfactory 2 ≤ |z| ≤ 3 = questionable
|z| > 3 = unsatisfactory
According to the recommendations in ISO 13528:2005, an unsatisfactory z-score is considered to give an action signal and a questionable z-score is considered to give a warning signal. A single action signal or warning signals in two successive rounds shall be taken that an anomaly has occurred that requires investigation.
Robust average = 33.9%UNODC value = 34.9%
41
Identified substances for Test Sample 3
Code Identified Substances0D8I70 Metamphetamine - lactose
1P11K1 Methamphetamine
1WVSJ1 MA,HCl
2NQOPT Metamfetamine
3OT6ZK metamfetamine
424PPI metamphetamine HCL
43N3MI Metamfetamine
47X44P Metanfetamina
4S9S9W Methylamphetamine
555QL5 Cocaína, metanfetamina
5MKKKW METAMFETAMINE
6LMHPM metamphetamine
6TMAIV methamphetamine
75C556 Metamfetamine
7AAFAT Methylamphetamine
7CBVXP METHAMPHETAMINE
7FFRIU METHAMPHETAMINE
7ZJLKU Metamfetamine
8SI888 Metamfetamine
8SZGPS Methylamphetamine
9OXPXM Methamphetamine
9UJ6IU metamfetamine
A35PZP METHYLAMPHETAMINE
ADCNJN Methamphetamine
AFX7LD Metamfetamine
ANZFXA Metamfetamine
APAA1A Metamfetamine
APXA8I metanfetamina
BDBQZB metamfetamine
BFDEKH Methamphetamine
BMNPI1 Methamphetamine
BTSAZE Metamfetamine
BVFV4R Metamfetamine
BYCO9B Methamphetamine
BZZH9I Methamphetamine
C77GE8 Methamphetamine
C8CHVI Methamphetamine
CI8BFC METANFETAMINE
CKSNPG Methamphetamine
CRY93L Metamphetamine
Sample 3 Analysis
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
42
Code Identified SubstancesCZPRCC Metamfetamine
D2EDTC methamphetamine
DE1ABE METANFETAMINE
DGKC1G MA and Lactose
DPJL4D Metamfetamine
E5598C Metamfetamine
EIJCK6 METHAMPHETAMINE
EQY9D9 METAMFETAMINE
ERARRR d,l-metamphetamine
ESOODT Metamfetamine
EXD7VB Methylamphetamine
FAYSEJ Metamphetamine
FLXW44 methamphetamine
FS2S09 metamfetamine
G7SO8D Methamphetamine
GCSVZA Methamphetamine
GQTNTT Metamfetamine
GRF2Z2 Metamfetamine 6,6 %
GRVGEG Metamphetamine
H2UF1T METHAMPHETAMINE
HPSK6E Metamfetamine
HQPTHH Metamfetamine
HXHRHH metamfetamine
HYFRMY Methamphetamine
IWUUS3 Metamfetamine
JGWYWW Metamfetamine
JPMMF7 metamfetamine
JQQQQZ Metamfetamine
JY6WCO Metamfetamine
K1TOYR Methamphetamine
KCFEK5 Metamphetamine
KCKKUB METAMPHETAMINE
KF2FY2 Metamfetamine
KFFFFP Metampfetamine
KKCYOY Methamphétamine chlorhydrate
KNRKIJ Metamfetamine
KR3ZKK Methamfetamine
KU2BUC Metamfetamine, lactose
LGA9TV MA
LYKKAN METAMFETAMINE
M9HNLP Methamphetamine
MKKFFY Methamphetamine HCl
MSHOBP Metamfetamine
MUQPYK metamfetamine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
43
Code Identified SubstancesMWBGCG Metamfetamine
N2TFCN methamphetamine
NH4NBU metamfetamin
NI5F66 Metanfetamina
O5RDHB metamphetamine
OE3YFV Metamfetamine
OEIEEC METHANPHETAMINE
OKLOIL Metamfetamine
OMI8O8 Methamphetamine
ON4A8J Methamphetamine
OOYONL metamphetamine
OTSLPP Metamfetamine
PCGCQB Metamfetamine
PFTCFW metamfetamine
PGOVBE Methlyamphetamine Hydrochloride
PRAXXW Lactose
PY333P Metamfetamine
PYGYE0 metamfetamine
QHSMDY Metamfetamine
QLPFFF METHAMPHETAMINE
QLRWCG Metamfetamine
QUD1AR metamfetamine Chlorhydrate
QVV1O8 METANFETAMINE
QWEMWS Methamphetamine
RBJEMN Metamfetamine
RDJ3M3 Metamfetamine
ROOGOL psychotropic substance
RSPZ60 Methylamphetamine HCl
RU2VD6 metamphetamine
S12Q5E Methamphetamine
SFUVOJ Métamphétamine
T60TMT methamphetamine
TMARDF Metamfetamine
TTBYYX METAMFETAMINE
TTEC2L Metanfetamine
TVD1KB METAMFETAMINE
U6UXCE D,L-metamphetamine
UHIEE2 Metamfetamine
UJJQFA Metamfetamine hydrochloride
UP37JV Metamphetamine
UV6OBH methamphetamine
VBAPZ2 metamfetamine
VIKHII METAMFETAMINE
VMCYPO Metamfetamine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
44
Code Identified SubstancesVNZMCG Metamfetamine
VVANL2 METANFETAMINE
W88A8P Metamfetamine
WRGGF7 Methamphetamine
WRHBIA methamphetamine
WZEYRI Metamfetamine
X5PP2A METANFETAMINE
X9GLI9 Metamphetamine
XBBWXW methamphetamine
XBNTS9 methamphetamine
XE8WIX Methamphetamine
XOOOLO Methylamphetamine
XRRRRJ METANPHETAMINE
XTXX6H Metamfetamine
XVCTKI Methamphetamine
XVXXXX Metamphetamine
Y00O0U metamphetamine
Y2PTPQ Methamphetamine
YLTLG9 Methamfetamine
YMW8P3 Methamphetamine
YQXIF0 metamphetamine
ZB002K metamphetamine
ZIIUIH Metamfetamine
ZX2A9A Metamfetamine
ZZF97Z Metamfetamine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
45
Code Lactose Metamfetamine
0D8I70 + +
1P11K1 + +The white powder were found to contain 6.56% of Methamphetamine.
1WVSJ1 + +metamfetamine,HCl is present (7.3% as base) with lactose
25A55I ANP +METAMPHETAMINE
2NQOPT - +Metamfetamine
3OT6ZK + +containet metamfetamine and lactose
424PPI + +additional substance: lactose
43N3MI ANP +
47X44P - +
4S9S9W ANP +The powder was analysed and found to contain Methylamphetamine. Methylamphetamine is a class A controlled drug within the provisions of the Misuse of Drugs Act 1971.
555QL5 ANP +De los análisis realizados a la muestra SM-3 se concluye que hay presencia de cocaína y metanfetamina
5MKKKW - +No more findings by GC-MS
6LMHPM ANP +metamphetamine was detected by GC/MS screening and it was quantified by HPLC using calibration curve
6TMAIV + +white powder - mixture of the methamphetamine and lactose
75C556 - +8.57% +/- 0.51% Metamfetamine detected in sample SM-3
7AAFAT ANP +On analysis of a representative sample of the powder Methylamephetamine was detected with a purity of approximately 12% calculated as base.
7CBVXP + +WAS FOUN METHANFETAMINE AND LACTOSE
7FFRIU - +In analysis, the substance found to contain 73.62% Methamphetamine
7ZJLKU - +In the test sample 2014/2/SM-3 was detected Metamfetamine
8SI888 + +Lactose
8SZGPS - +The white powder was analyzed and found to contain Methylamphetamine.
Statement of findings for Test Sample 3
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
46
Code Lactose Metamfetamine 9OXPXM + +
On analysis, I found the substance to contain 7.6 % (w/w) of Methamphetamine base.
9UJ6IU - +The sample contains metamfetamine (7.4%).
A35PZP + +METHYLAMPHETAMINE DETECTED, ALSO INDICATED THE PRESENCE OF LACTOSE. AVERAGE PURITY OF DRUG: 7% (CALCULATED AS BASE)
ADCNJN + +This sample was anlsyed and found to contain methamphetamine. This sample was also foudn to contain lactose.
AFX7LD + +
ANZFXA + +Presence of Metamfetamine and Lactose, was detected and confirmed in the white powder sample analyzed.
APAA1A ANP +Powder containing metamfetamine
APXA8I ANP +
AWAXYQ - +
BDBQZB + +The sample contains metamfetamine and lactose
BFDEKH + +
BMNPI1 - +Orange color changing to brown with Marquis reagent,blue color with Simon reagent indicates the possible presence of Methamphetamine.Analysis by GCMS shows the similarity as Methamphetamine.
BTSAZE + +In this sample excess two substances Metamfetamine and Lactose
BVFV4R - +Metamfetamine 8.08%
BYCO9B ANP +Sample contain methamphetamine
BZZH9I + +Methamphetamine, lactose
C77GE8 ANP +Screening with Simon Test gives blue color. Marquis gives orange. TLC shows RF and color are the same as those Methamphetamine standard. GCMS shows m/z and RT are the same as those of Methamphetamine Standard. Therefore the compound is Methamphetamine.
C8CHVI + +Methamphetamine hydrochloride and lactose confirmed in one unit of one unit received.
CI8BFC - +
CKSNPG ANP +White powder containing methamphetamine.
CRY93L ANP +The Sample obtains with the positive result with Marquis reagent.TLC: The Rf sample obtain with the same Rf Metamphetamine Standard.GC: The Rt sample obtain with the same Rt Metamphetamine Standard.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
47
Code Lactose Metamfetamine CZPRCC + +
Retention time and MS spectra in GC analysis (FID and MS) compared with reference material. Sample totally soluble in water. Metamfetamine as hydrochloride.
D2EDTC ANP +methamphetamine
DE1ABE - +
DGKC1G + +The sample 2014/2/SM-3 contains MA and Lactose detected by GC/MS and FTIR. The percentage by weight of MDMA is 7.7%.
DPJL4D - +Substance which detection has been increased lately.
E5598C + +SM3 contains Metamfetamine HCl with 8.61% Metamfetamine base and Lactose.
EIJCK6 + +METHAMPHETAMINE 7.46% AS BASE, LACTOSE
EQY9D9 ANP +this sample was spiked with METAMFETAMINE
ERARRR ANP +White powder containing metamphetamine (8.5%w/w) which is controlled by national law
ESOODT ANP +(Metamfetamine hydrochloride)
EXD7VB ANP +The sample was found to contain methylamphetamine.
FAYSEJ ANP +Methamphetamine existed in salt form
FLXW44 + +The sample was found to contain 7.6% of methamphetamine (as base)
FS2S09 ANP +Metamfetamine is controlled by law (KB 22/1/1998)
G7SO8D - +
GCSVZA ANP +Test sample 2014/2/SM-3 contains Methamphetamine.
GQTNTT + +The test sample 2014/2/SM-3 was found to contain the controlled drug metamfetamine. The sample was also found to contain lactose.
GRF2Z2 - +
GRVGEG - +Sample contains metamphetamine hydrochloride (7.4% content as base)
H2UF1T - +METHAMPHETAMINE
HPSK6E + +The sample SM-3 contains metamfetamine
HQPTHH - +The result of the test of the SM 3 (91)received confirmed that the mentioned sample contains Metamfetamine but we have not been able to give a cuantification result becaus we have not got the pattern of the mentioned controlled drug.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
48
Code Lactose Metamfetamine HXHRHH + +
metamfetamine + lactose
HYFRMY - +
IWUUS3 + +The chemical test and the coincidence in the time of retencion and mass fragmentogram with the material reference used in the same conditions that the analyzed samples, they indicate in this one the Metamfetamine presence.
JGWYWW - +The sample was found to contain 8.63% metamfetamine.
JPMMF7 + +lactose
JQQQQZ + +The sample contains Metamfetamine
JY6WCO ANP +
K1TOYR + +The sample SM3 contained 7.6 % methamphetamine calculated as a base. In the sample was identified lactose.
KCFEK5 ANP +Examination revealed the presence of Metamphetamine
KCKKUB + +METAMPHETAMINE and LACTOSE
KF2FY2 + +
KFFFFP ANP +The white powder contains Metampfetamine
KKCYOY + +the sample contains méthamphétamine, a controlled drug
KNRKIJ ANP +-
KR3ZKK + +The sample contains Methamfetamine at 9.2% purity. Also it includes lactose as cutting agent.
KU2BUC + +The sample contains Metamfetamine which is scheduled as illicit drug in France. The lactose is not scheduled as illicit drug in France (cutting agent).
LGA9TV - +
LYKKAN ANP +THE SAMPLE SM- 3 WAS ANALYSED AND FOUND TO TO CONTAIN METAMFETAMINE A PSYCHOTROPIC SUBSTANCES LISTED IN SECOND SCHEDULE OF THE NARCOTIC DRUGS AND PSYCHOTROPIC SUBSTANCES CONTROL ACT CAP 245
M9HNLP - +Based on the above analysis, the sample was found to contain Methamphetamine.
MKKFFY ANP +On Marquis test, gives orange color and turn to brown. Confirmed Methamphetamine Positive on TLC and GCMS. Chloride test positive with AgNO3. Quantification with GC, the sample contains Methamphetamine base 7.35%
MSHOBP + +Metamfetamine and Lactose
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
49
Code Lactose Metamfetamine MUQPYK + +
lactose
MWBGCG + +The sample contains Metamfetamine and lactose.
N2TFCN + +the sample also contain lactose
NH4NBU + +The sample contains 7.32±0.45 % percentage by weight of metamfetamine. The sample also contains lactose.
NI5F66 + +La muestra contiene metanfetamina y lactosa
O5RDHB + +
OE3YFV + +Screening by FTIR and identification/confirmation by GC/MS
OEIEEC + +BASE METHANPHETAMINE 7.18% + LACTOSE
OKLOIL - +The Sample 2014/2/SM-3 was analysed by colour test and TLC technique and found to contain Metamfetamine
OMI8O8 - +Sample 2014/2/SM-3 was found to contain Methamphetamine
ON4A8J - +Methamphetamine is a Class A controlled drug subject to the provisions of the Misuse of Drugs Act 1971.
OOYONL + +
OTSLPP + +Presence of metamfetamine hydrochloride and lactose was detected in the sample.
PCGCQB + +
PFTCFW + +powder with 7,0 % of metamfetamine base and lactose
PGOVBE + +This powder contained an average of 10% methylamphetamine hydrochloride.
PRAXXW + -As a result of the controllet Laboratory tests, we conclude that the test sample 2014/2/SM-3, which is a white substance in dust form is Lactose. This substance is not a narchotic drug and not illegal in Turkey under the Turkish Drug Law known as the law 2313.
PY333P + +Test sample 2014/2/SM-3 contains lactose and metamfetamine 8.46 %, m/m.
PYGYE0 + +contained metamfetamine and lactose
QHSMDY + +Sample 2014/2/SM-3 consists of white powder. Metamfetamine, a prohibited substance included in the National Law for Narcotics, is found to be present in this sample.
QLEZZZ ANP ANP
QLPFFF ANP +White substance that contains drug METHAMPHETAMINE. It contains also sugar that couldn't be identified as our FTIR instrument is out of function.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
50
Code Lactose Metamfetamine QLRWCG ANP +
The sample 2014/2/SM-3 was found to contain Metamfetamine.
QUD1AR + +lactose
QVV1O8 - +THE SAMPLE No. 2014/2/SM3 WAS FOUND METANFETAMINE
QWEMWS + +
RBJEMN + +Metamfetamine + lactose
RDJ3M3 - +On analysis, i found the sample to "2014/2/SM-3" contain Metamfetamine.
ROOGOL ANP +
RSPZ60 + +The powder contained an average of 7% by weight of Methylamphetamine HCl calculated as the base.
RU2VD6 + +It has been detected the presence of metamphetamine in the sample
S12Q5E ANP +One glass bottle containing 0.21 grammes of a powder containing 16 milligrammes of methamphetamine.
SFUVOJ + +The sample contains métamphétamine
T60TMT + +The powder contains methamphetamine with a concentration in the sample of 6.9% by weight. It also contains lactose.
TMARDF + +The sample contains metamfetamine which is listed as a controlled drug according to french order, appendix III of February 22nd 1990.
TTBYYX + +THE SAMPLE CONTAINS METAMFETAMINE HYDOCHLORIDE AND LACTOSE
TTEC2L ANP +
TVD1KB ANP +The following tests were conducted: two colorimetric tests, SCOTT and MARQUIS, were able to detect in the case of the sample SM3 METAMFETAMINE's presence. A Thin-Layer chromatography TLC, they were used as a solvent and eluent respectively, Dichloromethane and Methanol, using a silica plate of 20 × 20 cm, on the plate were placed the following samples of references (MDMA; Ketamine; Amfetamine; Mephedrone; Metamfetamine; MDEA; Cafeine; Cocaine; Procaine Penicillin; Lidocaine; Nimetazepam; m-Cpp HCl. After Elution, UV light was used to 265 nm in order to show what substances were separated. In the case of the sample SM3 was able to detect the presence of METAMFETAMINE. A qualitative analysis was performed with the GC/FID, using a column TRACSIL TRB-50, a Flow rate of 1.5 ml/min, FID Detector, the pressure of 14.7 psi, the oven temperature of 250ºc, the temperature for the injector and detector were of 300 C and 330 C respectively. The chromatograms were obtained using the KONIFROM PLUS Software and For the case of the sample SM3 managed to identify a METAMFETAMINE peak. In sumula for all three tests have been applied in the case of the sample SM3 the results were all identical or is all they gave positive for METAMFETAMINE.
U6UXCE ANP +white powder containing 8.0%w/w of metamphetamine
UHIEE2 ANP +We found Metamfetamine in SM-3 as controlled drugs
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
51
Code Lactose Metamfetamine UJJQFA + +
the presence of drug confirmed by GC-MS and FTIR analysis, the Metamfetamine hydrochloride confirmed by the anion analysis and FTIR
UP37JV + +Found Metamphetamine, quantification not performed
UV6OBH + +metamphetamine hidrochloride and lactose
VBAPZ2 + +In the sample 3 we have found metamfetamine
VIKHII + +METAMFETAMINE IS PRESENT IN HYDROCHLORIDE FORM. LACTOSE WAS ALSO IDENTIFIED.
VMCYPO + +The powder 2014/2/SM3 contains 7% metamfetamine base, lactose and glucose. metamfetamine base is a controlled substance in France.
VNZMCG - +On this sample, Thin Layer Chromatography (TLC), Gas Chromatography / Mass Spectrometry (GC/MS), Fourier Transform Infrared Spectrometry (FTIR), Gas Chromatography (GC/FID) analitycal techniques was applied. We found that this sample contains 8% Metamfetamine.
VVANL2 ANP +SIGN IN 2014/2 / SM-3 METANFETAMINE DETECTED.
VXMWHA + -
W88A8P + +There is 7.5% of metamfetamine in powder (calculated as metamfetamine base).
WRGGF7 ANP +
WRHBIA + +White powder containing methamphetamine and probably lactose. Lactose has not been confirmed due to the lack of lactose standard
WZEYRI ANP +* Results of screening tests were positive for metamfetamine. * Retention time and fragmentation pattern obtained with sample are equivalent to those obtained with metamfetamine.
X5PP2A ANP +Marquis reagent turned into orange-yelow, and Simon reagent turned into blue color, so both tests were found positive to presuntive metanfetamine. GC.FID confirmed Anfetamine retention time.
X9GLI9 ANP +METAMPHETAMINE WAS AS HUDROCHLORIDE FORM.
XBBWXW - +
XBNTS9 ANP +
XE8WIX ANP +Sample no SM3 contains 26.38% of methamphetamine.
XOOOLO ANP +The powder was found to contain Methylamphetamine at a concentration of 6 percent. Methylamphetamine is controlled by the Misuse of Drugs Act 1971 (as amended) as a Class A drug.
XRRRRJ + +
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
52
Code Lactose Metamfetamine XTXX6H + +
Also contain lactose
XVCTKI ANP +Narcotics: methamfetamine, approximately 7% as free base
XVXXXX + +Lactose
Y00O0U - +metamphetamine
Y2PTPQ + +The sample SM-3 contains metamphetamine and Lactose.
YLTLG9 + +This sample is methamfetamine.
YMW8P3 + +Marquis Reagent (MR) positive, GC/MS. Retention time and mass charge ratio (M/Z) similar Characteristics with the substances used as Standard or pattern. FTIR absortion bands coincidents with the substances used as standard or pattern (Lactose)
YQXIF0 + +lactose
ZB002K ANP +white powder containing metamphetamine
ZIIUIH - +7.47%
ZX2A9A ANP +The test sample was analyses and found to contain metamfetamine
ZZF97Z - +Simon's reagent test shows dark blue @ Marquis reagent test shows orange to brown. Melting point 184.5'C. It's confirmed by TLC & GCMS.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
53
Identification methods for Test Sample 3
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
0D8I70 SI I
S S I I
1P11K1 I I S
1WVSJ1 I I
25A55I SI
2NQOPT SI
3OT6ZK S S SI I I
424PPI I SI ATR
43N3MI S S SI SI
47X44P S I I
4S9S9W S S I
555QL5 S S I I IMS
5MKKKW I SI SI
6LMHPM SI
6TMAIV SI SI
75C556 S S I SI
7AAFAT S S I I
7CBVXP S S S I I
7FFRIU S S I
7ZJLKU I I
8SI888 S S S I SI I
8SZGPS S S I I
9OXPXM S I I
9UJ6IU S S S I I
A35PZP S S I I S SIMONS
ADCNJN SI SI I I I GC/MS derivitisation
AFX7LD S I I
ANZFXA S S S SI I
APAA1A I SI
APXA8I S S S I
AWAXYQ S S S SI I SI I
BDBQZB S S I I
BFDEKH S S S I
BMNPI1 S S S I I Mayer test
Analytical Technique Codes
Legend: S ... used for Screening I ... used for Identification SI ... used for both
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
54
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
BTSAZE S S S I I
BVFV4R S I
BYCO9B SI ASAP-TOF
BZZH9I S SI SI
C77GE8 S S I I S Fast Blue B
C8CHVI SI I I
CI8BFC SI I
CKSNPG S I SI SI
CRY93L S S S I
CZPRCC I I S
D2EDTC SI I
DE1ABE S I
DGKC1G I I SI I
DPJL4D S I I I I Fehling
E5598C I S I I I
EIJCK6 S S S I I I
EQY9D9 I I I I ion scan
ERARRR I SI
ESOODT S S I
EXD7VB SI
FAYSEJ S I I
FLXW44 SI I LC/ELSD
FS2S09 S I S
G7SO8D S I I
GCSVZA S S S I I S Chen-Kao Reagent
GQTNTT S S S I I
GRF2Z2 S S S I I SI
GRVGEG S S I
H2UF1T SI I
HPSK6E S S S I I I SI I Tollens reagent
HQPTHH S I
HXHRHH S SI I I SI
HYFRMY I SI I
IWUUS3 S S I I
JGWYWW
S I S
JPMMF7 S SI SI I
JQQQQZ S S I I S Tanred
JY6WCO I SI
K1TOYR I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
55
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
KCFEK5 S SI
KCKKUB SI I
KF2FY2 I SI
KFFFFP S S S I
KKCYOY I SI I
KNRKIJ SI
KR3ZKK SI S I S
KU2BUC S S I I
LGA9TV SI I
LYKKAN S S S I I
M9HNLP S I S Simon's Reagent
MKKFFY S S S I I I
MSHOBP SI SI I I I RAMAN
MUQPYK I SI SI
MWBGCG
SI I
N2TFCN S SI SI I
NH4NBU S I SI I
NI5F66 SI S S I I
O5RDHB SI I
OE3YFV I S I SI
OEIEEC S I I I D-RX
OKLOIL SI I SI I lead acetate+ammonia test
OMI8O8 S I I
ON4A8J S S I I
OOYONL I I
OTSLPP SI I I I
PCGCQB I SI SI
PFTCFW I I S ion mobility spectrometry
PGOVBE S S I I
PRAXXW S I I
PY333P I S I HPLC/RI
PYGYE0 I SI I
QHSMDY S S I I I
QLEZZZ I SI
QLPFFF S S SI I
QLRWCG S S S I I
QUD1AR SI SI
QVV1O8 I SI
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
56
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
QWEMWS
S S S I I
RBJEMN SI SI
RDJ3M3 S S S I I
ROOGOL S I I
RSPZ60 S S S I I
RU2VD6 SI I I
S12Q5E S S SI
SFUVOJ SI I
T60TMT I I SI I
TMARDF I I
TTBYYX S I I I
TTEC2L S S I
TVD1KB S S S I I
U6UXCE I SI
UHIEE2 S I S
UJJQFA S S I I I
UP37JV SI I
UV6OBH S S SI I I
VBAPZ2 I SI SI
VIKHII I SI I I XRD
VMCYPO I I S I dérivatisation+GCMS
VNZMCG S I I
VVANL2 S S I
VXMWHA I SI SI
W88A8P S S I I S Simon
WRGGF7 SI
WRHBIA SI
WZEYRI S S I I
X5PP2A SI SI SI I
X9GLI9 SI
XBBWXW S I I I
XBNTS9 S S I S Dunqueis levin
XE8WIX I I S immunochromatography
XOOOLO S S I I
XRRRRJ S S I I
XTXX6H S I I
XVCTKI SI I GC-IRD
XVXXXX I SI SI
Y00O0U SI SI SI S I I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
57
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
Y2PTPQ S S I I S TANRED
YLTLG9 S I I
YMW8P3 S S I I
YQXIF0 I SI SI
ZB002K I SI
ZIIUIH I SI
ZX2A9A SI S I I SI mecke
ZZF97Z S S S I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
58
2013/1-SM Copyright (c) 2013 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Frequency of use of screening and identification techniques:
Summary for test sample 3The response summary of all 157 participating laboratories for the identification of SM-3, frequency of use of screening andidentification techniques and false positives and negatives:
Identification of test sample 3:
False positives for Test Sample 3
Code Name Technique
Response SummaryParticipants: 157
Metamfetamine 154
Response Summary Participants: 131 Section 141 150
Response Summary Participants: 130 555QL5
555QL5 Cocaine 190/Ion mobility spec
RSPZ60 Amfetamine 180/190
False negatives for SM-3 (techinque(s) used or ANP)
Code: VXMHWA (170/190) PRAXXW (171/190) QLEZZZ (ANP)
Section 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231
Screening 49 71 58 19 0 2 0 3 56 21 2 0 0 0 0
31.0% 44.9% 36.7% 12.0% 0.0% 1.3% 0.0% 1.9% 35.4% 13.3% 1.3% 0.0% 0.0% 0.0% 0.0%
Identification 12 5 4 41 6 12 1 29 143 74 2 3 1 3 1
7.6% 3.2% 2.5% 25.9% 3.8% 7.6% 0.6% 18.4% 90.5% 46.8% 1.3% 1.9% 0.6% 1.9% 0.6%
# labs using this technique
55 71 59 49 6 13 1 30 144 80 3 3 1 3 134.8% 44.9% 37.3% 31.0% 3.8% 8.2% 0.6% 19.0% 91.1% 50.6% 1.9% 1.9% 0.6% 1.9% 0.6%
Response Summary Participants: 157
210
13
8.2%
17
10.8%
2717.1%
59
Z-Scores for Test Sample 3
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Concentration (mass%)
Deviation from robust
averagez-score
DE1ABE 2.1 -5.5 -7.0BMNPI1 2.6 -5.0 -6.3LYKKAN 3.8 -3.8 -4.8CRY93L 3.9 -3.8 -4.7FAYSEJ 5.8 -1.8 -2.38SI888 6.0 -1.6 -2.0
XOOOLO 6.0 -1.6 -2.0UJJQFA 6.3 -1.3 -1.6RU2VD6 6.4 -1.2 -1.51P11K1 6.6 -1.0 -1.3GRF2Z2 6.6 -1.0 -1.2JY6WCO 6.6 -1.0 -1.2HYFRMY 6.8 -0.8 -1.0XBBWXW 6.8 -0.8 -1.0T60TMT 6.9 -0.7 -0.9TTEC2L 6.9 -0.7 -0.9A35PZP 7.0 -0.6 -0.7PFTCFW 7.0 -0.6 -0.7RSPZ60 7.0 -0.6 -0.7VMCYPO 7.0 -0.6 -0.7XVCTKI 7.0 -0.6 -0.7OOYONL 7.1 -0.5 -0.72NQOPT 7.1 -0.5 -0.6OEIEEC 7.2 -0.4 -0.57ZJLKU 7.2 -0.4 -0.5PCGCQB 7.2 -0.4 -0.51WVSJ1 7.3 -0.3 -0.4APAA1A 7.3 -0.3 -0.4C8CHVI 7.3 -0.3 -0.4
MWBGCG 7.3 -0.3 -0.4X9GLI9 7.3 -0.3 -0.4NH4NBU 7.3 -0.3 -0.3SFUVOJ 7.3 -0.3 -0.3MKKFFY 7.4 -0.3 -0.39UJ6IU 7.4 -0.2 -0.2GRVGEG 7.4 -0.2 -0.2KNRKIJ 7.4 -0.2 -0.2OMI8O8 7.4 -0.2 -0.2PYGYE0 7.4 -0.2 -0.2EIJCK6 7.5 -0.1 -0.2YLTLG9 7.5 -0.1 -0.1ZIIUIH 7.5 -0.1 -0.1
W88A8P 7.5 -0.1 -0.1424PPI 7.5 -0.1 -0.16TMAIV 7.5 -0.1 -0.1
Code
Metamfetamine
Concentration (mass%)
Deviation from robust
averagez-score
3OT6ZK 7.5 -0.1 -0.19OXPXM 7.6 0.0 0.0CZPRCC 7.6 0.0 0.0FLXW44 7.6 0.0 0.0FS2S09 7.6 0.0 0.0K1TOYR 7.6 0.0 0.0OTSLPP 7.6 0.0 0.0S12Q5E 7.6 0.0 0.0VIKHII 7.7 0.1 0.1
DGKC1G 7.7 0.1 0.2Y00O0U 7.7 0.1 0.243N3MI 7.9 0.3 0.46LMHPM 7.9 0.3 0.40D8I70 8.0 0.4 0.5KU2BUC 8.0 0.4 0.5O5RDHB 8.0 0.4 0.5PGOVBE 8.0 0.4 0.5ROOGOL 8.0 0.4 0.5U6UXCE 8.0 0.4 0.5VNZMCG 8.0 0.4 0.5XBNTS9 8.0 0.4 0.5XVXXXX 8.0 0.4 0.5BVFV4R 8.1 0.5 0.6
QWEMWS 8.1 0.5 0.7DPJL4D 8.3 0.7 0.9H2UF1T 8.3 0.7 0.9MUQPYK 8.4 0.8 1.0PY333P 8.5 0.9 1.1ERARRR 8.5 0.9 1.2X5PP2A 8.5 0.9 1.275C556 8.6 1.0 1.3E5598C 8.6 1.0 1.3
JGWYWW 8.6 1.0 1.3KCFEK5 9.0 1.4 1.8TMARDF 9.0 1.4 1.825A55I 9.1 1.5 1.9KR3ZKK 9.2 1.6 2.1UV6OBH 10.2 2.6 3.3KCKKUB 10.3 2.7 3.4HPSK6E 10.3 2.7 3.57AAFAT 12.0 4.4 5.6LGA9TV 14.0 6.4 8.2XE8WIX 26.4 18.8 23.97FFRIU 73.6 66.0 84.0
Code
metamfetamine
60
Substance: SM-3/Metamfetamine
Round: 2014/2
Z-Score Report per Sample per Substance
‐2.3
‐2.0
‐2.0
‐1.6 ‐1.5‐1.3 ‐1.2
‐1.2‐1.0
‐1.0 ‐0.9
‐0.9 ‐0.7
‐0.7
‐0.7
‐0.7
‐0.7 ‐0.7 ‐0.6 ‐0.5
‐0.5
‐0.5 ‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.3
‐0.3
‐0.3 ‐0.2
‐0.2
‐0.2
‐0.2 ‐0.2 ‐0.2
‐0.1
‐0.1
‐0.1
‐0.1
‐0.1 ‐0.1
0.0
0.0
0.0
0.0
0.0
0.0
0.0 0.1 0.2
0.2
0.4
0.4 0.5 0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5 0.6 0.7 0.
9 0.9 1.0 1.1 1.2 1.2 1.3 1.3 1.3
1.8
1.8 1.9 2.1
3.3 3.4 3.5
‐4.0
‐3.0
‐2.0
‐1.0
0.0
1.0
2.0
3.0
4.0
DE1A
BEBM
NPI1
LYKK
ANCR
Y93L
FAYSEJ
8SI888
XOOOLO
UJJQFA
RU2V
D6
1P11
K1GRF2Z2
JY6W
COHY
FRMY
XBBW
XWT60TMT
TTEC2L
A35P
ZPPFTCFW
RSPZ60
VMCY
POXV
CTKI
OOYO
NL
2NQOPT
OEIEEC
7ZJLKU
PCGCQ
B1W
VSJ1
APAA
1AC8
CHVI
MWBG
CGX9
GLI9
NH4N
BUSFUVO
JMKK
FFY
9UJ6IU
GRV
GEG
KNRK
IJOMI8O8
PYGYE0
EIJCK6
YLTLG9
ZIIUIH
W88
A8P
424P
PI6TMAIV
3OT6ZK
9OXP
XMCZPR
CCFLXW
44FS2S09
K1TO
YROTSLPP
S12Q
5EVIKH
IIDG
KC1G
Y00O
0U43N3M
I6LMHPM
0D8I70
KU2B
UC
O5R
DHB
PGOVB
ERO
OGOL
U6U
XCE
VNZM
CGXB
NTS9
XVXX
XXBV
FV4R
QWEM
WS
DPJL4D
H2U
F1T
MUQPY
KPY
333P
ERAR
RRX5
PP2A
75C5
56E5598C
JGWYW
WKC
FEK5
TMAR
DF25
A55I
KR3ZKK
UV6
OBH
KCKK
UB
HPSK6
E7A
AFAT
LGA9
TVXE
8WIX
7FFRIU
‐6.3
‐4.8
‐4.7
84.0
23.9
8.2
5.6
‐7.0
Robust average = 7.6%UNODC value = 7.7%
z-scores can be interpreted by participants in line with ISO 13528:2005, section 7.4.2 and ISO/IEC guide43-1:1997(E) as follows:
|z| < 2 = satisfactory 2 ≤ |z| ≤ 3 = questionable
|z| > 3 = unsatisfactory
According to the recommendations in ISO 13528:2005, an unsatisfactory z-score is considered to give an action signal and a questionable z-score is considered to give a warning signal. A single action signal or warning signals in two successive rounds shall be taken that an anomaly has occurred that requires investigation.
61
Identified substances for Test Sample 4
Code Identified Substances0D8I70 cocaine - lactose1P11K1 Cocaine
1WVSJ1 cocaine,HCl
25A55I COCAINE
2NQOPT Cocaine
3OT6ZK cocaine
424PPI cocaine HCl
43N3MI Cocaine
47X44P Cocaina
4S9S9W Cocaine
555QL5 cocaína
5MKKKW COCAINE
6LMHPM cocaine
6TMAIV cocaine
75C556 Cocaine
7AAFAT Cocaine
7CBVXP COCAINE
7FFRIU Cocaine
7ZJLKU Cocaine
8SI888 Cocaine
8SZGPS Cocaine Hydrochloride
9OXPXM Cocaine
9UJ6IU Cocaine
A35PZP COCAINE
ADCNJN Cocaine (HCL)
AFX7LD Cocaine
ANZFXA Cocaine
APAA1A Cocaine
APXA8I cocaina
AWAXYQ ECGONINE METHYL ESTER, CINNAMOYL COCAINE , BENZOIC ACID
BDBQZB cocaine
BFDEKH Cocaíne
BMNPI1 Pseudococaine
BTSAZE Cocaine
BVFV4R Cocaine
BYCO9B Cocaine
BZZH9I Cocaine
C77GE8 Cocaine
C8CHVI Cocaine
Sample 4 Analysis
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
62
Code Identified SubstancesCI8BFC COCAINE
CKSNPG Cocaine
CRY93L Cocaine
CZPRCC Cocaine
D2EDTC cocaine
DE1ABE COCAINE
DGKC1G Cocaine and Lactose
DPJL4D Cocaine
E5598C Cocaine
EIJCK6 COCAINE
EQY9D9 COCAINE
ERARRR cocaine
ESOODT Cocaine
EXD7VB cocaine
FAYSEJ Cocaine
FLXW44 cocaine
FS2S09 cocaine
G7SO8D cocaine
GCSVZA Cocaine
GQTNTT Cocaine
GRF2Z2 Cocaine 6,9 %
GRVGEG Cocaine
H2UF1T COCAINE
HPSK6E Cocaine
HQPTHH COcaíne
HXHRHH cocaine
HYFRMY Cocaine
IWUUS3 Cocaine
JGWYWW Cocaine
JPMMF7 cocaine
JQQQQZ Cocaine
JY6WCO Cocaine
K1TOYR Cocaine
KCFEK5 Cocaine
KCKKUB COCAINE
KF2FY2 Cocaïne
KFFFFP Cocaine
KKCYOY cocaine chlorhydrate
KNRKIJ Cocaine
KR3ZKK Cocaine
KU2BUC cocaine, lactose
LGA9TV cocaine
LYKKAN COCAINE
M9HNLP Cocaine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
63
Code Identified SubstancesMKKFFY Cocaine HCl
MSHOBP Cocaine
MUQPYK cocaine
MWBGCG Cocaine
N2TFCN cocaine
NH4NBU cocaine
NI5F66 Cocaina
O5RDHB cocaine
OE3YFV Cocaine
OEIEEC COCAINE
OKLOIL Cocaine
OMI8O8 Cocaine
ON4A8J Cocaine
OOYONL cocaine
OTSLPP Cocaine
PCGCQB Cocaïne
PFTCFW cocaine
PGOVBE Cocaine Hydrochloride
PRAXXW Cocaine
PY333P Cocaine
PYGYE0 cocaine
QHSMDY Cocaine
QLEZZZ Cocaine
QLPFFF COCAINE
QLRWCG Cocaine
QUD1AR Cocaine Chlorhydrate
QVV1O8 COCAINE
QWEMWS Cocaine
RBJEMN Cocaine
RDJ3M3 Cocaine
ROOGOL narcotic drug
RSPZ60 Cocaine HCl
RU2VD6 Cocaine
S12Q5E Cocaine
SFUVOJ cocaine
T60TMT cocaine
TMARDF Cocaine
TTBYYX COCAIN
TTEC2L Cocaine
TVD1KB COCAINE
U6UXCE cocaine
UHIEE2 Cocaine
UJJQFA Cocaine hydrochloride
UP37JV cocaine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
64
Code Identified SubstancesUV6OBH cocaine
VBAPZ2 Cocaine
VIKHII COCAINE
VMCYPO cocaïne
VNZMCG Cocaine
VVANL2 COCAINE
VXMWHA cocaine
W88A8P Cocaine
WRGGF7 Cocaine
WRHBIA cocaine
WZEYRI Cocaine
X5PP2A COCAINE
X9GLI9 Cocaine
XBBWXW cocaine
XBNTS9 cocaine
XE8WIX cocaine
XOOOLO Cocaine
XRRRRJ COCAINE
XTXX6H Cocaine
XVCTKI Cocaine
XVXXXX Cocaine
Y00O0U cocaine
Y2PTPQ Cocaine
YLTLG9 Cocaine
YMW8P3 Cocaine
YQXIF0 cocaine
ZB002K cocaine
ZIIUIH Cocaine
ZX2A9A cocaine and cannabinol
ZZF97Z Cocaine
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
65
Code Cocaine Lactose 0D8I70 + +
We also found cynnamoilcocaine, methylecgonine and benzoylecgonine
1P11K1 + +The white powder were qualitatively analysed and found to contain cocaine.
1WVSJ1 + +cocaine,HCl is present (7.1% as base) with lactose
25A55I + ANP
2NQOPT + ANPcontained cocaine and trace of methylecgonine, methylecgonidine, cinnamoilcocaine
3OT6ZK + +containet cocaine and lactose
424PPI + +additional substance: lactose
43N3MI + ANP
47X44P + -
4S9S9W + ANPThe powder was analysed and found to contain Cocaine. Cocaine is a class A controlled drug within the provisions of the Misuse of Drugs Act 1971.A portion of the powder was further analysed and found to contain approximately 6% (weight for weight) Cocaine.
555QL5 + ANPDe los análisis realizados a la muestra SM-4 se concluye que hay presencia de cocaína
5MKKKW + +7.2 % Cocaine +/- 0.7%LACTOSE by FT-IR
6LMHPM + ANPcocaine was detected by GC/MS screening and it was quantified by GC/FID using calibration curve
6TMAIV + +white powder - mixture of the cocaine and lactose
75C556 + -Cocaine detected in sample SM-4 but quantitation not performed.
7AAFAT + ANPOn analysis of a representative sample of the powder cocaine was detected with a purity of approximately 7% calculated as base.
7CBVXP + +WAS FOUND COCAINE AND LACTOSE
7FFRIU + -In analysis, the substance founfd to contain Cocaine
7ZJLKU + -In the test sample 2014/2/SM-4 was detected Cocaine
8SI888 + +Lactose
Statement of findings for Test Sample 4
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
66
Code Cocaine Lactose 8SZGPS + +
The white powder was analyzed and found to contain Cocaine Hydrochloride.
9OXPXM + +On analysis, I found the substance to contain 7.2 % (w/w) of Cocaine base.
9UJ6IU + -The sample contains cocaine (7.2%).
A35PZP + +COCAINE DETECTED, ALSO INDICATED THE PRESENCE OF LACTOSE. AVERAGE PURITY OF DRUG: 7% (CALCULATED AS BASE)
ADCNJN + +This sample was found to contain ccaine at a purity of 7%. This sample was also found to contain lactose.
AFX7LD + +
ANZFXA + +Presence of Cocaine and Lactose, was detected and confirmed in the white powder sample analyzed.
APAA1A + ANPPowder containing cocaine
APXA8I + ANP
AWAXYQ + -
BDBQZB + +The sample contanins cocaine and lactose
BFDEKH + +
BMNPI1 + -blue color with Cobalt Thiocyanate test and blue color reappear in the lower chloroform layer by modified Cobalt Thiocynate Test indicates possible presence of Cocaine.Analysis by GCMS shows the similarity as Pseudococaine.
BTSAZE + +In this sample contain two substances Cocaine and Lactose
BVFV4R + -Cocaine 6.84%
BYCO9B + ANPSample contain cocaine
BZZH9I + ANPCocaine
C77GE8 + ANPScreening with Wagner Test gives white PPT. TLC results shows the RF and color are the same as those of Cocaine Standard. GCMS data gives that m/z , RT are the same as those of cocaine standard. Therefore , the compound is Cocaine.
C8CHVI + +Cocaine hydrochloride and lactose confirmed in one unit tested of one unit received.
CI8BFC + -
CKSNPG + ANPWhite powder containing cocaine.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
67
Code Cocaine Lactose CRY93L + ANP
The Sample obtains with the positive result with Cobalt thiocyanate reagent.TLC: The Rf sample obtain with the same Rf Cocaine Standard.GC: The Rt sample obtain with the same Rt Cocaine Standard.
CZPRCC + +Retention time and MS spectra in GC analysis (FID and MS) compared with reference material. Sample totally soluble in water. Cocaine as hydrochloride.
D2EDTC + ANPcocaine
DE1ABE + -
DGKC1G + +The sample 2014/2/SM-4 contains Cocaine and Lactose detected by GC/MS and FTIR. The percentage by weight of MDMA is 6.9%.
DPJL4D + -Substance commonly detected in our laboratory.
E5598C + +SM4 contains Cocaine HCl with 6.90% Cocaine base and Lactose.
EIJCK6 + +COCAINE 6.50% AS BASE, LACTOSE
EQY9D9 + ANPthis sample was spiked with COCAINE
ERARRR + ANPWhite powder containing cocaine(8.0%w/w) which is controlled by national law
ESOODT + ANP
EXD7VB + ANPThe sample was found to contain cocaine
FAYSEJ + ANPCocaine existed in salt form. were found Ergonidine-methyl-ester and Benzoylergonine in the sample.
FLXW44 + +The sample was found to contain 7.1% of cocaine (as base)
FS2S09 + ANPCocaine is controlled by law (KB 8/10/1963)
G7SO8D + -
GCSVZA + ANPTest sample 2014/2/SM-4 contains Cocaine.
GQTNTT + +The test sample 2014/2/SM-4 was found to contain the controlled drug cocaine. The sample was also found to contain lactose.
GRF2Z2 + -
GRVGEG + +Sample contains cocaine hydrochloride (8.4% content as base)
H2UF1T + -COCAINE
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
68
Code Cocaine Lactose HPSK6E + +
The sample SM-4 contains a drug- cocaine
HQPTHH + -The result of the test of the SM 4 (96)received confirmed that the mentioned sample contains Cocaíne with 9.2 percentage by weight (9.2%)
HXHRHH + +cocaine + lactose
HYFRMY + -
IWUUS3 + +The chemical test and the coincidence in the time of retencion and mass fragmentogram with the material reference used in the same conditions that the analyzed samples, they indicate in this one the cocaine presence.
JGWYWW + -The sample was found to contain cocaine.
JPMMF7 + +lactose
JQQQQZ + +The sample contains Cocaine
JY6WCO + ANP
K1TOYR + +The sample SM4 contained 7.5 % cocaine calculated as a base.In the sample was identified lactose.
KCFEK5 + ANPExamination revealed the presence of Cocaine
KCKKUB + +COCAINE and LACTOSE
KF2FY2 + +
KFFFFP + ANPThe white powder contains Cocaine.
KKCYOY + +the sample contains cocaine, a controlled drug
KNRKIJ + ANP-
KR3ZKK + +The sample contains Cocaine at 9% purity. Also it includes lactose as cutting agent.
KU2BUC + +The sample contains cocaine which is scheduled as illicit drug in France. The lactose is not scheduled as illicit drug in France (cutting agent).
LGA9TV + -
LYKKAN + ANPTHE SAMPLE SM-4 WAS ANALYSED AND FOUND TO TO CONTAIN COCAINE A NARCOTIC SUBSTANCES LISTED IN SECOND SCHEDULE OF THE NARCOTIC DRUGS AND PSYCHOTROPIC SUBSTANCES CONTROL ACT CAP 245
M9HNLP + -Based on the above analysis, the sample was found to contain Cocaine.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
69
Code Cocaine Lactose MKKFFY + ANP
Give positive blue color with Cobalt thiocyanate. Confirmed Cocaine Positive on TLC and GCMS. Chloride test positive with AgNO3. Quantification with GC, the sample contains Cocaine base 7.20%
MSHOBP + +Cocaine Hcl and Lactose
MUQPYK + +lactose
MWBGCG + +The sample contains Cocaine and lactose.
N2TFCN + +the sample also contain lactose
NH4NBU + +The sample contains 7.38±0.34 % percentage by weight of cocaine. The sample also contains lactose.
NI5F66 + +La muestra contiene cocaina y lactosa
O5RDHB + +
OE3YFV + +Screening for cobalt thiocyanate (test scott), identification FTIR and confirmation by GC/MS
OEIEEC + +BASE COCAINE 7.45% + LACTOSE
OKLOIL + -The sample 2014/2/SM-4 was analysed by colour test and TLC technique and found to contain Cocaine
OMI8O8 + -Sample 2014/2/SM-3 was found to contain Cocaine
ON4A8J + -Cocaine is a Class A controlled drug subject to the provisions of the Misuse of Drugs Act 1971. The purity was determined as 7%.
OOYONL + +
OTSLPP + +Presence of cocaine hydrochloride and lactose was detected in the sample. The seizure also contained trace-level alkaloidal impurities.
PCGCQB + +
PFTCFW + +powder with 6,7 % of cocaine base and lactose
PGOVBE + +This powder contained an average of 7% cocaine hydrochloride.
PRAXXW + +As a result of the controllet Laboratory tests, we conclude that the test sample 2014/2/SM-4, which is a white substance in dust form is a mixture of two different substances.The first ingrediant of the mixture is Cocaine and the second ingrediant of the mixture is Lactose. The first ingrediant is a narchotic drug and is illegal in Turkey under the Turkish Drug Law known as the law 2313 but the second ingrediant of the subtance is not a narchotic drug and is not illegal in Turkey.
PY333P + +Test sample 2014/2/SM-4 contains lactose and cocaine 6.96 %, m/m.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
70
Code Cocaine Lactose PYGYE0 + +
contained cocaine and lactose
QHSMDY + +Sample 2014/2/SM-4 consists of white powder. Cocaine, a prohibited substance included in the National Law for Narcotics, is found to be present in this sample.
QLEZZZ + -
QLPFFF + ANPWhite substance that contains drug COCAINE. It contains also sugar that couldn't be identified as our FTIR instrument is out of function.
QLRWCG + ANPThe sample 2014/2/SM-3 was found to contain Cocaine.
QUD1AR + +lactose
QVV1O8 + -THE SAMPLE No. 2014/2/SM4 WAS FOUND COCAINE
QWEMWS + +
RBJEMN + +Cocaine and lactose
RDJ3M3 + -On analysis, i found the sample to "2014/2/SM-4" contain Cocaine.
ROOGOL + ANP
RSPZ60 + +The powder contained an average of 7% by weight of Cocaine HCl calculated as the base.
RU2VD6 + +It has been detected the presence of cocaine in the sample
S12Q5E + ANPOne glass bottle containing 0.22 grammes of a powder containing 0.01 grammes of cocaine.
SFUVOJ + +The sample contains cocaine
T60TMT + +the powder is a mixture of lactose and cocaine. The concentration of cocaine in the sample is 6.8% by weight.
TMARDF + +The sample contains cocaine which is listed as a controlled drug according to french order, appendix I of February 22nd 1990.
TTBYYX + +THE SAMPLE CONTAINS 7% OF COCAIN ( EQUIVALENT BASE ) AND LACTOSE
TTEC2L + ANP
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
71
Code Cocaine Lactose TVD1KB + ANP
The following tests were conducted: two colorimetric tests, SCOTT and MARQUIS, were able to detect in the case of the sample the presence of SM4 COCAINE. A Thin-Layer chromatography TLC, they were used as a solvent and eluent respectively, Dichloromethane and Methanol, using a silica plate of 20 × 20 cm, on the plate were placed the following samples of references (MDMA; Ketamine; Amfetamine; Mephedrone; Metamfetamine; MDEA; Cafeine; Cocaine; Procaine Penicillin; Lidocaine; Nimetazepam; m-Cpp HCl. After Elution, UV light was used to 265 nm in order to show what substances were separated. In the case of SM4 sample managed to detect the presence of Cocaine. A qualitative analysis was performed with the GC/FID, using a column TRACSIL TRB-50, a Flow rate of 1.5 ml/min, FID Detector, the pressure of 14.7 psi, the oven temperature of 250ºc, the temperature for the injector and detector were of 300 C and 330 C respectively. The chromatograms were obtained using the KONIFROM PLUS Software and For the case of SM4 sample managed to identify a peak of cocaine. In sumula for all three tests have been applied in the case of SM4 sample results were all identical or is all they gave positive for COCAINE.
U6UXCE + ANPwhite powder containing 7.2% w/w of cocaine
UHIEE2 + ANPWe found Cocaine in SM-4 as controlled drugs
UJJQFA + +the presence of drug confirmed by GC-MS and FTIR analysis, the Cocaine hydrochloride confirmed by the anion analysis and FTIR
UP37JV + -Found Cocaine with 5.6%w/w
UV6OBH + +cocaine hidrochloride and lactose
VBAPZ2 + +In the sample 4 we have found cocaine
VIKHII + +COCAINE IS PRESENT IN HYDROCHLORIDE FORM. LACTOSE WAS ALSO IDENTIFIED.
VMCYPO + +The powder 2014/2/SM4 contains 8% cocaïne base, lactose and glucose. cocaïne base is a controlled substance in France.
VNZMCG + -On this sample, Thin Layer Chromatography (TLC), Gas Chromatography / Mass Spectrometry (GC/MS), Fourier Transform Infrared Spectrometry (FTIR), Gas Chromatography (GC/FID) analitycal techniques was applied. We found that this sample contains 6% Cocaine.
VVANL2 + ANPSIGN IN 2014/2 / SM-4 COCAINE DETECTED.
VXMWHA + +
W88A8P + +There is 7.3% of cocaine in powder (calculated as cocaine base).
WRGGF7 + ANP
WRHBIA + +White powder containing cocaine and probably lactose. Lactose has not been confirmed due to the lack of lactose standard
WZEYRI + ANP* Results of screening tests were positive for cocaine. * Retention time and fragmentation pattern obtained with sample are equivalent to those obtained with cocaine.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
72
Code Cocaine Lactose X5PP2A + ANP
Cobalt tiocyonate turned into blue color, so this test identified presumptive cocaine. CG-FID confirmed cocaine retention time.
X9GLI9 + ANPCOCAINE WAS AS HYDROCHLORIDE FORM
XBBWXW + +the sample contains mixture of cocaine and lactose
XBNTS9 + ANP
XE8WIX - ANPSample no4 contains 6.42% of cocaine.
XOOOLO + ANPThe powder was found to contain Cocaine at a concentration of 7 percent. Cocaine is controlled by the Misuse of Drugs Act 1971 (as amended) as a Class A drug.
XRRRRJ + -
XTXX6H + +Also contain lactose
XVCTKI + ANPNarcotics: Cocaine, approximately 8% as free base
XVXXXX + +Lactose
Y00O0U + -cocaine
Y2PTPQ + +The sample SM4- contains cocaine and Lactose.
YLTLG9 + +This sample is cocaine.
YMW8P3 + +Cobalt Thiocyanate (CT) positive, GC/MS. Retention time and mass charge ratio (M/Z) similar Characteristics with the substances used as Standard or pattern. FTIR absortion bands coincidents with the substances used as standard or pattern (Lactose)
YQXIF0 + +lactose
ZB002K + ANPwhite powder containing 7.4%w/w of cocaine (mean value of two determinations).
ZIIUIH + -7.43%
ZX2A9A + ANPThe sample contained cannabinol and traces of cocaine
ZZF97Z + -Cobalt thiocyanate test shows blue. Melting point 195.3'C. It's confirmed by TLC & GCMS.
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
73
Identification methods for Test Sample 4
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
0D8I70 SI I
S S I I
1P11K1 I I S
1WVSJ1 I I
25A55I SI
2NQOPT SI
3OT6ZK S S SI I I
424PPI I SI ATR
43N3MI S S SI SI
47X44P S I I
4S9S9W S S I
555QL5 S S I I IMS
5MKKKW I SI SI
6LMHPM SI
6TMAIV SI SI
75C556 S S I SI
7AAFAT S S I I
7CBVXP S S S I I
7FFRIU S S I
7ZJLKU I I
8SI888 S S S I SI I
8SZGPS S S I I
9OXPXM S I I
9UJ6IU S S S I I
A35PZP S S I I S SIMONS
ADCNJN SI SI I I I GC/MS derivitisation
AFX7LD S I I
ANZFXA S S S SI I
APAA1A I SI
APXA8I S S S I
AWAXYQ S S S SI I SI I
BDBQZB S S I I
BFDEKH S S S I
BMNPI1 S S S I I Mayer test
Analytical Technique Codes
Legend: S ... used for Screening I ... used for Identification SI ... used for both
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
74
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
BTSAZE S S S I I
BVFV4R S I
BYCO9B SI ASAP-TOF
BZZH9I S SI SI
C77GE8 S S I I S Fast Blue B
C8CHVI SI I I
CI8BFC SI I
CKSNPG S I SI SI
CRY93L S S S I
CZPRCC I I S
D2EDTC SI I
DE1ABE S I
DGKC1G I I SI I
DPJL4D S I I I I Fehling
E5598C I S I I I
EIJCK6 S S S I I I
EQY9D9 I I I I ion scan
ERARRR I SI
ESOODT S S I
EXD7VB SI
FAYSEJ S I I
FLXW44 SI I LC/ELSD
FS2S09 S I S
G7SO8D S I I
GCSVZA S S S I I S Chen-Kao Reagent
GQTNTT S S S I I
GRF2Z2 S S S I I SI
GRVGEG S S I
H2UF1T SI I
HPSK6E S S S I I I SI I Tollens reagent
HQPTHH S I
HXHRHH S SI I I SI
HYFRMY I SI I
IWUUS3 S S I I
JGWYWW
S I S
JPMMF7 S SI SI I
JQQQQZ S S I I S Tanred
JY6WCO I SI
K1TOYR I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
75
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
KCFEK5 S SI
KCKKUB SI I
KF2FY2 I SI
KFFFFP S S S I
KKCYOY I SI I
KNRKIJ SI
KR3ZKK SI S I S
KU2BUC S S I I
LGA9TV SI I
LYKKAN S S S I I
M9HNLP S I S Simon's Reagent
MKKFFY S S S I I I
MSHOBP SI SI I I I RAMAN
MUQPYK I SI SI
MWBGCG
SI I
N2TFCN S SI SI I
NH4NBU S I SI I
NI5F66 SI S S I I
O5RDHB SI I
OE3YFV I S I SI
OEIEEC S I I I D-RX
OKLOIL SI I SI I lead acetate+ammonia test
OMI8O8 S I I
ON4A8J S S I I
OOYONL I I
OTSLPP SI I I I
PCGCQB I SI SI
PFTCFW I I S ion mobility spectrometry
PGOVBE S S I I
PRAXXW S I I
PY333P I S I HPLC/RI
PYGYE0 I SI I
QHSMDY S S I I I
QLEZZZ I SI
QLPFFF S S SI I
QLRWCG S S S I I
QUD1AR SI SI
QVV1O8 I SI
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
76
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
QWEMWS
S S S I I
RBJEMN SI SI
RDJ3M3 S S S I I
ROOGOL S I I
RSPZ60 S S S I I
RU2VD6 SI I I
S12Q5E S S SI
SFUVOJ SI I
T60TMT I I SI I
TMARDF I I
TTBYYX S I I I
TTEC2L S S I
TVD1KB S S S I I
U6UXCE I SI
UHIEE2 S I S
UJJQFA S S I I I
UP37JV SI I
UV6OBH S S SI I I
VBAPZ2 I SI SI
VIKHII I SI I I XRD
VMCYPO I I S I dérivatisation+GCMS
VNZMCG S I I
VVANL2 S S I
VXMWHA I SI SI
W88A8P S S I I S Simon
WRGGF7 SI
WRHBIA SI
WZEYRI S S I I
X5PP2A SI SI SI I
X9GLI9 SI
XBBWXW S I I I
XBNTS9 S S I S Dunqueis levin
XE8WIX I I S immunochromatography
XOOOLO S S I I
XRRRRJ S S I I
XTXX6H S I I
XVCTKI SI I GC-IRD
XVXXXX I SI SI
Y00O0U SI SI SI S I I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
77
Code 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231 210
Y2PTPQ S S I I S TANRED
YLTLG9 S I I
YMW8P3 S S I I
YQXIF0 I SI SI
ZB002K I SI
ZIIUIH I SI
ZX2A9A SI S I I SI mecke
ZZF97Z S S S I I
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
78
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Frequency of use of screening and identification techniques:
Summary for test sample 4The response summary of all 157 participating laboratories for the identification of SM-4, frequency of use of screening andidentification techniques and false positives and negatives:
Identification of test sample 4:
False positives for Test Sample 1
Code Name Technique
Response SummaryParticipants: 157
Cocaine: 156
Response Summary Participants: 130
ZX2A9A Cannabinol 150/171
False negatives for SM-4 (techinque(s) used or ANP)
140 141 142 150 160 161 170 171 180Response Summary Participants: 131 Section 141 150Section 140 141 142 150 160 161 170 171 180 190 200 211 220 230 231
Screening 49 71 58 19 0 2 0 3 56 21 2 0 0 0 0
31.0% 44.9% 36.7% 12.0% 0.0% 1.3% 0.0% 1.9% 35.4% 13.3% 1.3% 0.0% 0.0% 0.0% 0.0%
Identification 12 5 4 41 6 12 1 29 143 74 2 3 1 3 1
7.6% 3.2% 2.5% 25.9% 3.8% 7.6% 0.6% 18.4% 90.5% 46.8% 1.3% 1.9% 0.6% 1.9% 0.6%
# labs using this technique
55 71 59 49 6 13 1 30 144 80 3 3 1 3 134.8% 44.9% 37.3% 31.0% 3.8% 8.2% 0.6% 19.0% 91.1% 50.6% 1.9% 1.9% 0.6% 1.9% 0.6%
Response Summary Participants: 157
210
13
8.2%
17
10.8%
2717.1%
79
Z-Scores for Test Sample 4
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
Concentration (mass%)
Deviation from robust
averagez-score
CRY93L 2.8 -4.4 -7.3LYKKAN 2.9 -4.3 -7.0KF2FY2 4.0 -3.2 -5.3KCKKUB 4.7 -2.5 -4.1BZZH9I 5.0 -2.2 -3.6MSHOBP 5.3 -1.9 -3.1UP37JV 5.6 -1.6 -2.74S9S9W 6.0 -1.2 -2.0PRAXXW 6.0 -1.2 -2.0QUD1AR 6.0 -1.2 -2.0VNZMCG 6.0 -1.2 -2.0XE8WIX 6.4 -0.8 -1.3EIJCK6 6.5 -0.7 -1.2EXD7VB 6.5 -0.7 -1.2FAYSEJ 6.5 -0.7 -1.2JY6WCO 6.5 -0.7 -1.2Y00O0U 6.6 -0.6 -1.0OOYONL 6.7 -0.5 -0.9PFTCFW 6.7 -0.5 -0.8KNRKIJ 6.7 -0.5 -0.8
QWEMWS 6.8 -0.4 -0.8XBBWXW 6.8 -0.4 -0.8TTBYYX 6.8 -0.4 -0.72NQOPT 6.8 -0.4 -0.7T60TMT 6.8 -0.4 -0.7BVFV4R 6.8 -0.4 -0.6ADCNJN 6.9 -0.3 -0.5DGKC1G 6.9 -0.3 -0.5E5598C 6.9 -0.3 -0.5GRF2Z2 6.9 -0.3 -0.5OMI8O8 6.9 -0.3 -0.5PCGCQB 6.9 -0.3 -0.5PY333P 7.0 -0.2 -0.47AAFAT 7.0 -0.2 -0.48SI888 7.0 -0.2 -0.4A35PZP 7.0 -0.2 -0.4BMNPI1 7.0 -0.2 -0.4CKSNPG 7.0 -0.2 -0.4G7SO8D 7.0 -0.2 -0.4KU2BUC 7.0 -0.2 -0.4O5RDHB 7.0 -0.2 -0.4ON4A8J 7.0 -0.2 -0.4PGOVBE 7.0 -0.2 -0.4RSPZ60 7.0 -0.2 -0.4TMARDF 7.0 -0.2 -0.4XOOOLO 7.0 -0.2 -0.4XVXXXX 7.0 -0.2 -0.4MUQPYK 7.0 -0.2 -0.3
Code
Cocaine
Concentration (mass%)
Deviation from robust
averagez-score
1WVSJ1 7.1 -0.1 -0.27ZJLKU 7.1 -0.1 -0.2APAA1A 7.1 -0.1 -0.2FLXW44 7.1 -0.1 -0.2JQQQQZ 7.1 -0.1 -0.27CBVXP 7.2 0.0 -0.15MKKKW 7.2 0.0 0.09OXPXM 7.2 0.0 0.09UJ6IU 7.2 0.0 0.0CZPRCC 7.2 0.0 0.0H2UF1T 7.2 0.0 0.0MKKFFY 7.2 0.0 0.0
MWBGCG 7.2 0.0 0.0S12Q5E 7.2 0.0 0.0U6UXCE 7.2 0.0 0.0Y2PTPQ 7.2 0.0 0.0UJJQFA 7.2 0.0 0.0W88A8P 7.3 0.1 0.1KKCYOY 7.3 0.1 0.1QLEZZZ 7.3 0.1 0.1XTXX6H 7.3 0.1 0.1NH4NBU 7.4 0.2 0.36TMAIV 7.4 0.2 0.3C8CHVI 7.4 0.2 0.3FS2S09 7.4 0.2 0.3ZB002K 7.4 0.2 0.3VIKHII 7.4 0.2 0.3ZIIUIH 7.4 0.2 0.3OEIEEC 7.5 0.3 0.4K1TOYR 7.5 0.3 0.5XBNTS9 7.5 0.3 0.5YLTLG9 7.6 0.4 0.6SFUVOJ 7.6 0.4 0.6RU2VD6 7.6 0.4 0.6XVCTKI 7.6 0.4 0.6DPJL4D 7.6 0.4 0.7WRHBIA 7.7 0.5 0.73OT6ZK 7.7 0.5 0.843N3MI 7.7 0.5 0.8APXA8I 7.7 0.5 0.8PYGYE0 7.8 0.6 0.9HPSK6E 7.8 0.6 1.06LMHPM 7.8 0.6 1.0DE1ABE 8.0 0.8 1.2ERARRR 8.0 0.8 1.3HYFRMY 8.0 0.8 1.3OTSLPP 8.0 0.8 1.3TTEC2L 8.0 0.8 1.3
Code
Cocaine
80
Z-Scores for Test Sample 4
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)
2014/2-SM Copyright (c) 2015 UNODC
Concentration (mass%)
Deviation from robust
averagez-score
VMCYPO 8.0 0.8 1.3VXMWHA 8.1 0.9 1.5GRVGEG 8.4 1.2 1.9UV6OBH 8.4 1.2 2.0OE3YFV 9.0 1.8 2.9KCFEK5 9.0 1.8 2.9KR3ZKK 9.0 1.8 2.90D8I70 9.2 2.0 3.2
HQPTHH 9.4 2.2 3.6LGA9TV 9.5 2.3 3.7424PPI 9.9 2.7 4.4X5PP2A 10.0 2.8 4.5X9GLI9 10.2 3.0 4.9
ROOGOL 11.1 3.9 6.425A55I 11.4 4.2 6.8
Cocaine
Code
81
Substance: SM-4/Cocaine
Round: 2014/2
Z-Score Report per Sample per Substance
‐3.6
‐3.1
‐2.7
‐2.0
‐2.0
‐2.0
‐2.0
‐1.3 ‐1
.2‐1.2
‐1.2
‐1.2 ‐1
.0 ‐0.9 ‐0.8 ‐0.8 ‐0.8
‐0.8 ‐0.7 ‐0.7
‐0.7 ‐0.6 ‐0.5
‐0.5
‐0.5
‐0.5
‐0.5
‐0.5 ‐0.4 ‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4
‐0.4 ‐0.3 ‐0.2
‐0.2
‐0.2
‐0.2
‐0.2 ‐0.1 0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0 0.0
0.1 0.1
0.1
0.1 0.
3 0.3
0.3
0.3
0.3 0.3
0.3 0.4 0.5
0.5 0.
60.6 0.6
0.6 0.7 0.7 0.8 0.8
0.8 0.9 1.0 1.0
1.2 1.3
1.3
1.3
1.3
1.3
1.5
1.9 2.0
2.9
2.9
2.9
3.2
3.6 3.
7
‐4.0
‐3.0
‐2.0
‐1.0
0.0
1.0
2.0
3.0
4.0
CRY9
3LLYKK
ANKF2FY2
KCKK
UB
BZZH
9IMSH
OBP
UP3
7JV
4S9S9W
PRAX
XWQUD1
ARVN
ZMCG
XE8W
IXEIJCK6
EXD7
VBFA
YSEJ
JY6W
COY0
0O0U
OOYO
NL
PFTC
FWKN
RKIJ
QWEM
WS
XBBW
XWTTBY
YX2N
QOPT
T60TMT
BVFV4R
ADCN
JNDG
KC1G
E559
8CGRF2Z
2OMI8O8
PCGCQ
BPY
333P
7AAF
AT8SI888
A35P
ZPBM
NPI1
CKSN
PGG7SO8D
KU2B
UCO5R
DHB
ON4A
8JPG
OVB
ERS
PZ60
TMAR
DFXO
OOLO
XVXX
XXMUQPY
K1W
VSJ1
7ZJLKU
APAA
1AFLXW
44JQ
QQQZ
7CBV
XP5M
KKKW
9OXP
XM9U
J6IU
CZPR
CCH2U
F1T
MKK
FFY
MWBG
CGS12Q
5EU6U
XCE
Y2PT
PQUJJQFA
W88
A8P
KKCYOY
QLEZZZ
XTXX
6HNH4
NBU
6TMAIV
C8CH
VIFS2S09
ZB00
2KVIKH
IIZIIUIH
OEIEEC
K1TO
YRXB
NTS9
YLTLG9
SFUVO
JRU
2VD6
XVCT
KIDP
JL4D
WRH
BIA
3OT6ZK
43N3M
IAP
XA8I
PYGYE0
HPSK6
E6LMHP
MDE
1ABE
ERAR
RRHYFRM
YOTSLPP
TTEC
2LVM
CYPO
VXMWHA
GRV
GEG
UV6
OBH
OE3
YFV
KCFEK5
KR3ZKK
0D8I70
HQPTHH
LGA9
TV42
4PPI
X5PP
2AX9GLI9
ROOGOL
25A5
5I
‐7.3
‐7.0
‐5.3
‐4.1
6.8
6.44.5
4.9
4.4
Robust average = 7.2%UNODC value = 7.1%
z-scores can be interpreted by participants in line with ISO 13528:2005, section 7.4.2 and ISO/IEC guide43-1:1997(E) as follows:
|z| < 2 = satisfactory 2 ≤ |z| ≤ 3 = questionable
|z| > 3 = unsatisfactory
According to the recommendations in ISO 13528:2005, an unsatisfactory z-score is considered to give an action signal and a questionable z-score is considered to give a warning signal. A single action signal or warning signals in two successive rounds shall be taken that an anomaly has occurred that requires investigation.
82
The Laboratory and Scientific SectionUNODC, International CentreWagramer Strasse 5, P.O. Box 500A-1400, Vienna, AUSTRIAEmail: [email protected]
2014/2-SM Copyright (c) 2015 UNODC
INTERNATIONAL QUALITY ASSURANCE PROGRAMME (IQAP)INTERNATIONAL COLLABORATIVE EXERCISES (ICE)