instructionsforusefrankshospitalworkshop.com/equipment/documents... · orderinginformation flocare...

19
PT IT-CH FR-BE-CH-LU ES DE-AT-CH-LU BE-NL UK-IE INSTRUCTIONS FOR USE For enteral use onl Read this manual before using the pump ZMNL400336_AW.indd 2 3/18/10 3:47:27 PM

Upload: others

Post on 27-Apr-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

PTIT-CHFR-BE-CH-LUESDE-AT-CH-LUBE-NL UK-IE

INSTRUCTIONS FOR USEFor enteral use onl

Read this manual before using the pump

ZMNL400336_AW.indd 2 3/18/10 3:47:27 PM

Page 2: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

BE-NL • INHOUDAlgemene productinformatie 5Voorbereiding 5Service en garantie 5Gebruiksaanwizing 6Alarmfuncties enveiligheidsvoorzieningen 10Onderhoud 10Problemen en oplossingen 11Instelmodus 12Technische specificaties 14

DE-AT-CH-LU • INHALTAllgemeine Informationen 15Vorbereitung 16Service und Gewährleistung 17Bedienungsanweisung 17Alarmfunktionen undSicherheitstechnik 22Instandhaltung 22Fehlersuchtabelle 23Sstemeinstellungen 24Technische Daten 25Bestellinformation 26

ES • ÍNDICEIntroducción 27Preparación 27Servicio garantía 27Instrucciones de funcionamiento 28Funciones de alarma prestacionesde seguridad 32Mantenimiento 32Diagrama de solución deproblemas 33Modo de configuración 34Especificaciones técnicas 36

FR-BE-CH-LU • CONTENTIntroduction 37Précautions d’emploi 37Service après-vente et garantie 37Instructions d’utilisation 38Alarmes et sécurités 42

Entretien 42Résolutions des problèmes 43Mode de paramétrage 44Spécifications techniques 45

IT-CH • CONTENUTIIntroduzione 47Preparazione 47Servizio e garanzia 47Istruzioni per l’uso 48Allarmi e funzioni di sicurezza 52Manutenzione 52Risoluzione dei problemi 53Funzione di parametrizzazione 54Specifiche tecniche 55

PT • ÍNDICEIntrodução 57Preparação 57Serviço e garantia 57Instruções de funcionamento 58Funções do alarme e característicasde segurança 62Manutenção 62Quadro para soluções eproblemas 63Modo de configuração - Set up 64Especificações técnicas 65

UK-IE • CONTENTIntroduction 67Preparation 67Service and Warrant 67Operating instructions 68Alarm functions and safet features 72Maintenance 72Problem solver chart 73Set up mode 74Technical specifications 75

Appendix A:Guidance and manufacturer’s declaration -Electromagnetic compatibilit 78

ZMNL400336_AW.indd 3 3/18/10 3:47:27 PM

Page 3: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

Figure 4

Figure 5c Figure 5d

k

a

Figure 1

d

f c

gh

e

b

i j

Figure 3

Figure 2

Figure 5a Figure 5b

A

B

ZMNL400336_AW.indd 4 3/18/10 3:47:31 PM

Page 4: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

BE-NLPOMPSPECIFICATIES:Fig. 1:a Pompdeurb Scherm (LCD)c Toetsenbordd Rotore Stroomopwaartse druksensorf Stroomafwaartse druksensorg Vake voor pompseth Luchtsensor

Fig. 2:i Gebruiksaanwizingj Bevestigingspunt voor paalklemk Aansluiting voor adapter

Fig. 3: Paalklem

Fig. 4: Adapter

Fig. 5a-5d: De pompset vullen eninbrengen

DE-AT-CH-LUKOMPONENTENLISTE:Fig. 1:a Pumpentüreb Anzeigefeld (LCD)c Tastaturd Rotore Vorgeschalteter Drucksensorf Nachgeschalteter Drucksensorg Anschluss für das Pumpsegmenth Luftsensor

Fig. 2:i Bedienungshinweisej Anschluss für Schraubhalterungk Buchse für den Netzanschluss

Fig. 3: Schraubhalterung

Fig. 4: AC/DC Adapter/Ladegerät

Fig. 5a-5d: Einführung desPumpsegments

ESLISTA DE COMPONENTES:Fig. 1:a Compuerta de la bombab Panel indicador (LCD)c Tecladod Rotore Sensor de presión ascendentef Sensor de presión descendenteg Receptáculo para insertar el sis-

temah Sensor del aire

Fig. 2:i Instrucciones de usoj Receptáculo para el fiador del

polok Toma para la conexión a la uni-

dad cargadora de enchufe

Fig. 3: Fiador del polo

Fig. 4: Cargador

Fig. 5a-5d: Inserción del equipo debomba

FR-BE-CH-LULISTE DE COMPOSANTS:Fig. 1:a Porteb Ecran (LCD)c Clavierd Rotore Capteur de pression amontf Capteur de pression avalg Site insertion cassetteh Détecteur d’air

Fig. 2:i Instructions d’utilisationj Site insertion supportk Connexion de l’adaptateur

Fig. 3: Support

Fig. 4: Chargeur

Fig. 5a-5d: Insertion de la tubulure

IT-CHELENCO DEI COMPONENTI:Fig. 1:a Sportello pompab Visualizzatore a cristalli liquidi

(LCD)c Tastierad Rotoree Sensore della pressione in

ingressof Sensore della pressione in uscitag Ricettacolo per l’inserto del seth Sensore dell’aria

Fig. 2:i Istruzioni per l’usoj Ricettacolo per il sistema di fis-

saggio alla piantana

k Connettore per l’alimentatore

Fig. 3: Sistema di fissaggio allapiantana

Fig. 4: Caricabatterie

Fig. 5a-5d: Inserimento del set

PTLISTA DE COMPONENTES:Fig. 1:a Portab Ecrã (LCD)c Tecladod Rotore Sensor de pressão ascendentef Sensor de pressão descendenteg Receptor para colocação do sistemah Sensor de ar

Fig. 2:i Instruções de utilizaçãoj Receptor para colocação do

suportek Conexão para o carregador

Fig. 3: Suporte

Fig. 4: Carregador

Fig. 5a-5d: Colocação do sistema dealimentação na bomba

UK-IELIST OF COMPONENTS:Fig. 1:a Pump doorb Displa panel (LCD)c Kepadd Rotore Upstream pressure sensorf Downstream pressure sensorg Receptacle for pump inserth Air sensor

Fig. 2:i Instructions for usej Receptacle for pole clampk Socket for connection to the plug

charger unit

Fig. 3: Pole clamp

Fig. 4: Charger

Fig. 5a-5d: Filling and Insertion of thepump set

ZMNL400336_AW.indd 5 3/18/10 3:47:32 PM

Page 5: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

ORDERING INFORMATION

Flocare® Infinit™ pump Western Europe corporate code 35676Flocare® Infinit™ ChargerPole clampInstructions for use

Optional accessoriesPlease refer to our local Nutricia contact for availabilit of Flocare® Infinit™ accessories, asthere are: feeding sets, carr bag, PDMS/data cable, nurse call, service suitcase and other.

MANUFACTURER

Nutricia Medical Devices b.v.Schiphol Boulevard 2611118 BH Schiphol AirportThe Netherlands

ZMNL400336_AW.indd 6 3/18/10 3:47:32 PM

Page 6: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

INTRODUCTION

• The Flocare® Infinit™ pump is a small, lightweight pump for both portable and bedside use. The pump is

based on the rotar peristaltic principle and is intended for enteral use onl.

PREPARATION

• Check the integrit of the pump. Do not use the pump if it is damaged. If a technical failure occurs or if

the pump is dropped, the pump should be checked b a qualified technician.

• Donot use thepump inareaswhere there is a riskof explosionse.g. in thepresenceof flammableanaesthetics.

• If the pump has been stored for an period of time, it should be plugged into the mains to recharge the bat-

ter before commencing enteral feeding. The batter will be completel charged after approximatel 6 hours.

• The Flocare® Infinit™ pump should onl be used in combination with the appropriate Flocare® Infinit™

pump set! (consult our local Nutricia sales representative for information on availbale pump sets).

• This pump operates in an orientation, making it ideal for ambulator use.

• Check the position of the feeding tube, as advised b our healthcare professional, before commencing

tube feeding.

• Pump fed patients should be regularl monitored and supervised. Specific patient groups require consistent

and controlled administration of enteral nutrition as well as simultaneous application of medication (e.g. insu-

lin administration). In these cases, regular and frequent checks, as determined b the attending healthcare

professional, should be carried out to ensure correct administration of nutrition throughout the therap period.

Using the Infinit™ pump’s DOSE function is recommended in these cases (see section “To set a Dose”).

• For bedside use, the multi-position pole clamp (figure 3) can be attached to the pump with the screw

provided. The pump can be fixed in an position (rotatable in 360°).

SERVICE AND WARRANTy

The manufacturer recommends an inspection of the pump at an authorised service centre ever 2 ears.

Onl authorised personnel should perform service work on Infinit™ pumps. Please contact our local Sales

Organisation / Nutricia Subsidiar for all service and repair of pumps (see address at the back of the booklet).

Limitations of warrantSolel for the benefit of the original buer/user, Nutricia Medical Devices B.V, warrants all new Flocare® Infinit™ pumps, ofits manufacture to be free from defects in material and workmanship, excluding normal wear and tear, and will replace orrepair, at its service facilit or other location designated b Nutricia Medical Devices B.V, an Flocare® Infinit™ pump returnedto it within thirt-six (36) months of original purchase b the buer/user. Such repair or replacement shall be free of charge.

Nutricia Medical Devices B.V warrants to the original buer/user, all repaired or replaced pumps to be free from defectsin material and workmanship and will replace or repair such products, at its service facilit or other location designa-ted b Nutricia Medical Devices B.V. Such repair or replacement shall carr a warrant of ninet (90) das from thedate of repair or replacement or the balance of the new pumps warrant as described above, whichever is greater.

THIS WARRANTy APPLIES ONLy TO FLOCARE® INFINITy™ PUMPS MANUFACTURED By NUTRICIA MEDICALDEVICES B.V AND IS THE ONLy WARRANTy GIVEN WITH RESPECT TO THE PUMPS. NO WARRANTIES IMPLIEDIN LAW, INCLUDING, BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITy AND FITNESS FORPARTICULAR PURPOSE, SHALL APPLy. NUTRICIA MEDICAL DEVICES B.V WILL BE LIABLE, IN ANy EVENT, ONLyFOR THE PURCHASE PRICE OF THE DEFECTIVE PRODUCT, BUT NOT FOR ANy CONSEQUENTIAL DAMAGES.

This Warrant ma not be modified, amended or otherwise changed, except b a written document properl executed b a cor-porate officer of Nutricia Medical Devices B.V

THE WARRANTy IS VOID IF THE FLOCARE® INFINITy™ PUMP IS SUBjECT TO ABUSE, ACCIDENT, ALTERATION, MODIFICATION,TAMPERING, MISUSE OR THE UNAUTHORIZED REPAIR OR SERVICE IN ANy WAy WITHOUT PRIOR AUTHORIZATION FROMNUTRICIA MEDICAL DEVICES B.V. IN ANy EVENT, NUTRICIA MEDICAL DEVICES B.V.’S LIABILITy SHALL NEVER EXCEEDTHE ORIGINAL PURCHASE PRICE OF THE PUMP AND SHALL NOT EXTEND TO ANy CONSEQUENTIAL LOSS OR DAMAGE.

-67-

UK-IE

ZMNL400336_AW.indd 69 3/18/10 3:47:42 PM

Page 7: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-68-

OPERATING INSTRUCTIONS

INSERTION OF THE FLOCARE® INFINITy™ PUMP SET

• Connect the set to the feed container as instructed on the packaging of the Flocare® Infinit™ pump set.

Remove the dust cap from the step connector.

• The Flocare® Infinit™ pump sets are equipped with an “automatic free-flow protection”, as a consequence

the set has no roller clamp.

Fill the pump set completel with feed b gentl pinching on the side of the cassette that is marked with a

drop (point A on figure 5a) or fill the pump set with help of the pump (see paragraph: filling the pump set).

• Open the pump door b pressing up on the lower wall and rotating the door upwards at the same time

(see figure 5b).

• Position the looped section of the silicone tubing around the rotor.

Stretching lightl, (see figure 5c) seat the cassette into the pump (see figure 5d).

• Close the pump door.

SWITCHING “ON”

For proper pump operation, make sure the pump door is closed prior to switching the

pump on. Press the “ON/OFF” ke for 2 seconds.

The pump beeps and carries out a short self test, showing the pump serial number

in 8 digits. Verif that all displa segments and smbols are active as shown in figure

6 (see next page).

The pump displas the total volume delivered since the memor was cleared and

switches in the hold mode read for programming.

SWITCHING “OFF”

Press the “ON/OFF” ke and keep it pressed for 2 seconds. A continuous alarm will be

heard and the pump switches off.

The feeding program (installed parameters) and total volume administered since the

last clearance will be retained in the pump’s memor.

If the pump shuts down due to low batter voltage, the memor will be retained for

24 hours.

“HOLD” MODE

To temporaril pause the pump, or switch into “hold” mode whilst operating, press the

“START/STOP” ke once. Three beeps are heard and the run smbol disappears.

The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume

(ml) are retained.

The “hold” mode is used to temporaril stop the flow of feed:

• to change the feeding program (installed parameters),

• to change the feed container,

ON/OFF

ON/OFF

START/STOP

UK-IE

ZMNL400336_AW.indd 70 3/18/10 3:47:42 PM

Page 8: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-69-

• to administer medication without switching the pump off,

• to silence an alarm and correct problems.

After 3 minutes a two tone audible alarm sounds and the message “PUSH STRT” appears in the displa.

Press “START/STOP” to stop the alarm and to extend the hold mode b a further 3 minutes or press “START/

STOP” twice (3 beeps will be heard) to resume programmed settings.

FILLING THE PUMP SET

The FILL SET function offered b the Infinit pump is used to fill a (new) feeding set with

nutrition (or water as the case ma be). When the FILL SET function runs all alarms

are deactivated helping avoid alarms being inadvertentl triggered (e.g. air alarm).

The FILL SET function should onl be used to fill an empt (air-filled) feeding set. Use

of the FILL SET function at an other time ma cause the pump to incorrectl calibrate.

When the FILL SET function is initiated the pump will calibrate to the giving set.

When the pump is in the hold mode:

Press and hold the “FILL SET” ke for 2 seconds to activate the “FILL SET” maneuver. Release the button as

soon as the pump generates a beep and starts pumping at a flow rate of approx. 700ml/h. During this maneu-

ver the message “FILL SET” appears in the displa. The pump will automaticall stop when the Flocare® Pack

Infinit™ pump set is completel filled with feed. However this “FILL SET” maneuver can at antime be stopped

b pushing the “FILL SET” ke a second time. The pump will return to the hold mode when the “FILL SET”

maneuver is complete or stopped.

DISPLAy

The pump has a liquid crstal displa (L.C.D.) with large alphanumeric characters, smaller smbols, words

and a back light.

The following information can be found in the displa:

• Flow rate (ml/h), volume (ml), is displaed through the large characters.

Words below describe what the number relates to (rate, dose or volume).

The pump also displas messages, for example “end of dose” will appear when the pump has finished

delivering a single feed dose.

•The arcs around the run smbol rotate when the pump is

running.

•The remaining batter capacit (see paragraph: batter

operation).

•Thewallplugsmbol indicates that thecharger isplugged in.

The back light of the displa turns off 10 seconds after the last ke is pressed.

The back light will switch on for 10 seconds when the charger is connected to the mains.

Figure 6

ON/OFF

FILL SET

d

UK-IE

ZMNL400336_AW.indd 71 3/18/10 3:47:42 PM

Page 9: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-70-

PROGRAMMING THE PUMP

• The Flocare® Infinit™ pump can be used for continuous or dose feeding.

• Insert the pump set in the pump, switch the pump on and fill the pump set when necessar (see para-

graph: filling the pump set).

• The displa shows the total volume delivered since the pump was last cleared.

• If necessar clear the volume delivered b pressing the “CLR” ke.

• The last installed flow rate (ml/h) is now displaed.

• The pump is now in the “hold” mode and is read for programming.

CONTINUOUS FEEDING

• Adust the flow rate (ml/h) if required b using the “+” or “-” ke. Hold either ke down to change rapidl.

• If another parameter is indicated in the LCD, it will be necessar to first press the “ml/h” ke followed b the

“+” or “-” ke to set the flow rate.

• The flow rate ranges from 1 - 400 ml/h in 1ml increments.

• The flow rate slows down and stops shortl at 50 - 125 - 250 ml/h.

• Press the “DOSE=VOL” ke and make sure the value is put at 0 ml indicated b the word “CONT”. This

means the pump will run in a continuous mode.

• Start the pump b pressing “START/STOP”.

• The arcs around the word “run” start rotating.

• The flow rate (ml/h) is displaed whilst functioning.

TO SET A DOSE

From the hold mode, with the last flow rate (ml/h) displaed in the LCD:

• Set the flow rate b pressing the “+” or “-” ke.

If another parameter is displaed:

• Press first the “ml/h” ke followed b “+” or “-” ke to adust the flow rate. The flow rate can be set between

1 and 400 ml/h, with increments of 1ml.

• Press the “DOSE=VOL” ke and install the volume to be administered with the “+” and “-” ke. The range

goes from 1 - 4000 ml, with steps from 1ml.

• Start the pump b pressing “START/STOP”.

During functioning the following parameters can be found in the displa:

• B pressing the “ml/h” ke the “flow rate” is displaed.

• B pressing the “INFO” ke the total volume since the memor was cleared will be

visualized.

When the required volume or dose has been delivered, “END OF DOSE” will appear and the pump will beep

(or mute, depending upon the pumps’ configuration (see paragraph: set up mode).

ON/OFF d d d d dml/h START/STOP400 ml/h+1 ml/h-

4000 ml+1 ml-DOSE=VOL

ON/OFF d d dml/h400 ml/h+1 ml/h- START/STOP

INFO

UK-IE

ZMNL400336_AW.indd 72 3/18/10 3:47:43 PM

Page 10: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING

• Press the “START/STOP” ke to pause the pump.

• Change the program b pressing the required ke (ml/h or DOSE=VOL) and adust using the “+” or “-” kes.

• Restart the pump b pressing the “START/STOP” ke again.

CLEARING THE MEMORy

All parameters and values can be cleared one after the other in the following wa:

• Press “START/STOP” to pause the pump.

• Select the parameter that needs to be cleared.

• Press the “CLR” ke to clear the memor. The parameter or value returns to its default value:

ml/h = 0 ml/h

VOL = cont = No dose set. Pump will feed continuousl until feed container is empt or pump

is stopped

INFO = 0 ml = Volume delivered is cleared

In order to keep a clear overview of the dail amount of feed delivered, clear the total volume delivered as

each dail feeding schedule is started, as follows:

• Switch the pump “ON”

• Immediatel after the self-test the pump displas total volume delivered.

• Press the “CLR” ke. The total volume delivered returns to “0 ml”.

• The pump displas the previous programmed flow rate (ml/h).

• The pump is in the “HOLD” mode read for programming.

• When another feeding program needs to be started, or the pump will be used for another patient, all set-

tings (rate and dose) can be returned to default and the volume delivered can be set at zero b simpl

pressing the “CLR” ke and keeping it pressed for 2 seconds.

BATTERy OPERATION

• The Flocare® Infinit™ pump is equipped with a Lithium-ion batter.

• The batter smbol in the displa is automaticall replaced b the plug smbol when the adapter is

plugged in. The “fuel gauge” then indicates that the batter is charging b displaing the segments in

an ascending low to high pattern starting with the leftmost segment.

This pattern continuousl repeats while the pump is charging.

• To check the status of the batter, disconnect the charger from the pump and turn the

pump on. The bars between E and F (E=Empt, F=Full) represent the “fuel gauge” of the

batter. Each bar is approximatel 1/4 of a full batter charge. If 2 bars appear the

batter is half full and remaining operating time is approximatel 12 hours at a flow rate

of 125 ml/h.

• In case of power failure the pump automaticall switches to batter powered operation.

• When during functioning the last block is gone, the batter smbol will blink to indicate that there is approxima-

tel 1 hour of charge left. The displa will flash “BATT” ever 3 seconds, alternating with the active displa and

the pump will beep ever 2 seconds to remind the user of the low batter charge condition. Plug in the charger

to continue to run and recharge the batter.

• As the batter level decreases the pump ma not be able to deliver tube feeds at rates greater than 300 ml/h,

even though some batter capacit remains. The pump will alarm and indicate “BATT”. If this happens, either

-71-

START/STOP

CLR

d

UK-IE

ZMNL400336_AW.indd 73 3/18/10 3:47:43 PM

Page 11: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-72-

recharge the batter or select a lower rate to complete the feed ccle.

• In case of batter failure, caused b, for example, an excessive temperature condition (which ma resolve

itself with time) or b a complete batter failure, the E, F and batter smbol will flash. Turn the pump “off” and

contact our healthcare provider or refer the pump to service.

CONNECTION TO AN EXTERNAL ALARM SySTEM

The power connector at the side of the pump offers the option of connecting the pump to an external alarm

sstem or Patient Data Monitoring Sstem (PDMS).

• The Flocare® Infinit™ Nurse Call (corporate code 35752) enables connection to an external alarm sstem.

This external alarm sstem ma be with open or closed contact. Follow the instructions for use of the

Flocare® Infinit™ Nurse Call to connect the pump.

• The Flocare® Infinit™ PDMS Cable (corporate code 35776) enables connection to an external PDMS

sstem. Follow the instructions for use of the Flocare® Infinit™ PDMS Cable to connect the pump.

ALARM FUNCTIONS AND SAFETy FEATURESIn case an of the problems listed on the problem solver chart occur, the pump delivers an audible and visual

alarm and stops working. The back light of the LCD automaticall switches on.

Exception to this is the low batter alarm “BATT”, in this situation the pump continues working.

Action in the event of an alarm:

Check the tpe of alarm displaed b the large displa characters.

• Press the “START/STOP” ke to stop the audible and visual alarm.

• Correct the cause of the alarm as described in the table.

• Start the pump again b pressing “START/STOP”.

The problem solver chart on the next page gives a clear explanation of the alarms.

MAINTENANCECleaning:

• Alwas unplug the pump prior to cleaning to avoid electric shock hazard.

• On a regular basis thoroughl clean all surfaces of the pump (including the sensors and rotor) with warm

soap water, a 5% bleach solution in water, or a multipurpose disinfectant cleaner.

• The Flocare® Infinit™ pump ma be rinsed b holding under a stream of warm, clean water. Do not sub-

merge the pump!

• Alwas maintain the rollers on the rotor in a clean state to ensure the spin freel.

• The adaptor normall does not require cleaning. When desired, a dr or slightl damp cloth ma be used

to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.

Safet notes:

• The manufacturer recommends an inspection of the pump at an authorised service centre ever 2 ears.

• If an fault occurs during use, or if the pump is dropped, it should be checked b authorised technical per-

sonnel prior to use.

• Replace the Flocare® Infinit™ pump set ever 24 hours to maintain deliver accurac and prevent the

growth of harmful bacteria. Dispose of Flocare® Infinit™ disposable sets properl, as required b local law.

• Do not use the Infinit alarm sstems to trigger actions related to secondar (electrical) medical devices

(e.g. a volumetric or sringe pump).

• Do not use pump functions (e.g. the fill set function) for an other purpose than described in this manual,

as this ma cause the pump to incorrectl calibrate.

UK-IE

ZMNL400336_AW.indd 74 3/18/10 3:47:43 PM

Page 12: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-73-

PROBLEM SOLVER CHART

OCC OUT(repeated)

NO SET

AIR

PUSH STRT

Condition

END OF DOSE

PROG

DOOR

BATTthe pump

remains working

Batter, “E” and “F”flashing

FILL SET

OCC IN

OCC OUT

•Calibration not et completedon the current feeding set

•The set is not fitted or wrongl fittedin the pump.

•The pressure sensor area is dirt

•The air sensor detected an emptpump set.

•The air sensor area is dirt.•The feeding set is not inserted cor-

rectl.

•The pump has been untouched inhold mode for 3 minutes or more.

Cause

•The pump administered the installeddose = volume.

•No flow rate is installed.Flow rate = 0 ml/h

•The door is not correctl closed.

•The batter capacit is too low.The pump is not able to deliverhighl viscous fluids at high flowrates with the present charge levelof the batter.

•Batter failure.

•The pump is priming the set.

•The pump detected an upstreamocclusion between the pump andthe feeding bag.

•The pressure sensor area is dirt.

•The pump detected adownstream occlusion between thepump and the patient.

•The pressure sensor area is dirt.

•Stop the alarm b pressing the “START/STOP” ke.•Start the pump b pressing the “START/STOP” ke and

let it run onl briefl.•Stop the pump b pressing the “START/STOP” ke insu-

ring that there has been no occlusion out alarm.•Remove the feeding set from the pump and reinsert the

feeding set in the pump.•Restart the pump b pressing the “START/STOP” ke.

•Stop the alarm b pressing “START/STOP”• Insert the Flocare® Infinit™ cassette into the pump as

indicated on the blister packaging and close the door.•Restart the pump.•Clean the sensors, reinsert the cassette in the pump and

restart the pump.

•Replace the empt feeding reservoir and continue feeding.If necessar prime the set.

•Make sure the air sensor is clean.•Make sure the feeding set is properl inserted

in the pump.

•Stop the alarm and prolong the hold mode with another 3minutes b pressing “START/STOP” ke.

•Program the pump and start it b pressing the “START/STOP”.

Correction

•Turn the pump off b pressing the “ON/OFF” keand hold it during 2 seconds or

•Clear the memor of the total volume administered (seeparagraph: clearing the memor), reprogram a new fee-ding schedule and start the pump b pressing “START/STOP”.

•Make sure the correct flow rate is programmed.

•Make sure the door is properl closed prior tostarting a feeding program.

•Connect the adaptor to the pump and mains and chargethe pump for approx. 6 hours. During charging the pumpcan be used.

•Turn the pump “off”, contact our healthcare provider orrefer the pump to service.

•Press the “FILL SET” ke another time to stop the pumpand bring it back in the hold mode.

•Stop the alarm b pressing the “START/STOP” ke.•Remove the feeding set out of the pump and check the

permeabilit b flushing the line.•Re-insert the feeding set in the pump and restart.•Clean the sensors, reinsert the cassette in the pump and

restart the pump.

•Stop the alarm b pressing the “START/STOP” ke.•Check the permeabilit of the feeding tube b

aspirating liquid via the medication port.•Clean the sensors, reinsert the cassette in the pump and

restart the pump.

UK-IE

Alwas follow the instructions below in case a problem occurs.Using methods other than those described ma cause the pump to function incorrectl.

ZMNL400336_AW.indd 75 3/18/10 3:47:43 PM

Page 13: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-74-

SET UP MODE

The set up mode is used • to set the alarm level,

• to lock the kepad,

• to switch the audible alarm off when a dose is administered,

• to switch the light of the LCD permanentl on when connected to the mains,

• to enable connection with a Nurse call or Patient Data

Monitoring Sstem (PDMS).

This set up mode ma onl be entered b healthcare professionals and authorised personnel trained to use

this application of the pump.

The patient or his/her relatives without permission of the phsician, dietician, nurse or other licensed practi-

tioner, ma not change the settings of the set up mode.

TO SET THE ALARM LEVEL

Switch the pump “on” while holding the “+” ke down. The pump enters the set up mode.

First the alarm level can be set:

“BEEP HIGH” (+) or “BEEP LOW” (-)

Use the “+” or “-” ke to change the setting from “high” to “low” or from “low” to “high”.

TO LOCK THE KEyPAD

Press in the set up mode the “FILL SET” ke to switch to the next setting:

“UNLK” the kepad (-) or “LOCK” the kepad (+). In the “LOCK” mode the feeding program of the pump

can no longer be changed.

Use the “+” or “-” ke to change the setting from “lock” to “unlk” or vice versa.

ON/OFF & d Beep high+Beep low-

+2 sec 2 sec

FILL SET d LockUnlk-

+

ER01 - ER99

No plug smbol visi-ble, while the pumpis connected to the

mains.

Condition

LOCK

•The self test detected an electronicerror

•The wall outlet doesn’t work.•The adaptor is damaged.

Cause•Onl the current feeding

schedule is allowed for this patient.Another feeding program is notallowed b our healthcare profes-sional.

•Turn the pump “off”, make sure the pump door is closedand switch the pump back “on”. If the error persists,contact our healthcare provider or refer the pump toservice.

•Connect the pump to another wall outlet.•Contact our healthcare professional or

Nutricia subsidiar to replace the adaptor.

Correction•The programming feature is blocked in the set up mode

of the pump. Ask our healthcare professional to modifthis setting.

UK-IE

ZMNL400336_AW.indd 76 3/18/10 3:47:43 PM

Page 14: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

TO MUTE WHEN DOSE DONE

Press in the set up mode the “FILL SET” ke to switch to the next setting:

“BEEP WHEN DONE” (+): the pump will give an alarm when the dose is administered or “MUTE WHEN

DONE” (-): the pump will not give an alarm after administering the dose. Use the “+” or “-” ke to change

the setting from “beep when done” to “mute when done” or vice versa.

TO SWITCH LIGHT ON

Press in the set up mode the “FILL SET” ke to switch to the next setting:

“LITE ON” (+) the light of the LCD remains on when the pump is connected to the mains. “LITE OFF” (-) the

light of the LCD switches off after a few seconds even if connected to the mains. Use the “+” or “-” ke to change

the setting from “lite on” to “lite off” or vice versa.

TO CONNECT WITH NURSE CALL OR PDMS

Press in the set up mode the “FILL SET” ke to switch to the next setting:

“OUTP PDMS” (+) to connect with a Patient Data Monitoring Sstem. “OUTP NRSE” to connect with a NURSE

CALL sstem. Use the “+” or “-” ke to change to setting from “OUTP PDMS”, to “OUTP NRSE” or “OUTP

OFF” and vice versa.

Press the “ON/OFF” ke and hold it down for 2 seconds to exit the SET UP mode. The settings are auto-

maticall saved.

TECHNICAL SPECIFICATIONS

• This Flocare® Infinit™ pump in combination with the adapter and the Flocare® Infinit™ Nurse Call are designed

to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions

and immunit and are in compliance with Directive 93/42/EEC.

0344• IEC 601: Class II Medical Device Directive: Class IIa• Microprocessor controlled• BF Equipment

• Charger: input 100-240V AC / 50-60 Hz / 0.4A Max.

output 5 V DC 2.4 A• Batter: internal rechargeable Lithium ion batter, 3.6 V DC 2000 mAh

• Batter capacit: 24 hours at 125 ml/h• jet waterproof: IPX 5• IPX 5: Splash water proof: Water ets from an direction shall have no effect on the pump.

-75-

FILL SET d Beep when doneMute when done-

+

FILL SET d dOUTP PDMSOUTP NRSE-

+ ON/OFF2 sec

FILL SET d dLite onLite off-

+ ON/OFF2 sec

UK-IE

ZMNL400336_AW.indd 77 3/18/10 3:47:43 PM

Page 15: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-76-

• Dimensions: 140 x 95 x 35 mm

• Weight: approx. 392 g

• Accurac flow rate: ±5,0% with appropriate Flocare® Infinit™ pump set

• Humidit: Operation mode: 30% to 75% noncondensing

Storage: 10% to 95 % noncondensing

• Temperature: Operation mode: +5°C to +40°C

Storage and transportation: -20°C to +65°C

• Atmospheric pressure: Operation mode: 70-106 kPa

Storage and transportation: 50-106 kPa

• Occlusion detection pressure: Upstream occlusion: -34 kPa (tolerance 21 kPa)

Downstream occlusion: 83 kPa (tolerane 21 kPa)

• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies

from 0.5 to 1 ml which relates to an air bubble with an approximate length in the silicone pump segment of

6 - 13 cm.

• The Flocare® Infinit™ can safel be operated on commercial aircraft.

• The use of other accessories, adaptors and cables than listed within this manual ma result in increased

emissions or decreased immunit of the equipment of the Flocare® Infinit™ pump.

• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put

into service according to the EMC information provided in the accompaning documents.

• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equip-

ment. If the Flocare® Infinit™ pump is used adacent to or stacked with other equipment, the pump should

be observed to verif normal operation.

• In case of pump scrapping, alwas notif our Nutricia sales unit of the pumps serial number. Pump scrap-

ping should alwas occur according to local legislation.

UK-IE

ZMNL400336_AW.indd 78 3/18/10 3:47:43 PM

Page 16: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-77-

ZMNL400336_AW.indd 79 3/18/10 3:47:43 PM

Page 17: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-78-

Emissions test Compliance Electromagnetic environment – guidance

Guidance and manufacturer’s Declaration – Electromagnetic emissions

Immunit test IEC 60601 test level Electromagnetic environment – guidance

Guidance and manufacturer’s Declaration – Electromagnetic immunit

Compliance level

161 Vac for 0,5 s95 Vac for 100 ms0 Vac for 10 ms0 Vac for 5 s

161 Vac for 0,5 s95 Vac for 100 ms0 Vac for 10 ms0 Vac for 5 s

APPENDIX A: GUIDANCE AND MANUFACTURER’S DECLARATION -ELECTROMAGNETIC COMPATIBILITy

The Flocare® Infinit™ enteral feeding pump is intended for use in the electromagnetic environment specified below.The customer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such anenvironment.

RF EmissionsCISPR 11

Group 1 The Flocare® Infinit™ uses RF energ onl for its internalfunction. Therefore, its RF emissions are ver low and arenot likel to cause an interference in nearb electronicequipment.

RF EmissionsCISPR 11

Class B

Harmonic emissionsIEC 61000-3-2

Class A

Voltage fluctuations/Flicker emissionsIEC 61000-3-3

Complies

The Flocare® Infinit™ is suitable for use in allestablishments, including domestic establishments andthose directl connected to the public low-voltage powersuppl network that supplies buildings used for domesticpurposes.

The Flocare® Infinit™ enteral feeding pump is intended for use in the electromagnetic environment specified below. Thecustomer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such an environment.

Electrostaticdischarge (ESD)IEC 61000-4-2

≤ 6 KV contact

≤ 8 KV air

Floors should be wood, concrete or ceramic tile.If floors are covered with snthetic material, therelative humidit should be at least 30%.

≤ 6 KV contact

≤ 8 KV air

Electric fasttransient/burstIEC 61000-4-4

≤ 2 KV for power sup-pl lines≤ 1 KV for input/ out-put lines

Mains power qualit should be that of a tpicalcommercial or hospital environment.

≤ 2 KV for power sup-pl lines≤ 1 KV for input/ out-put lines

SurgeIEC 61000-4-5

≤ 1 KV differentialmode

Mains power qualit should be that of a tpicalcommercial or hospital environment.

≤ 1 KV differentialmode

Voltage dips, shortinterruptions andvoltage variations onpower suppl lines.IEC 61000-4-6

Mains power qualit should be that of a tpicalcommercial or hospital environment. If the userof the Flocare® Infinit™ requires continuedoperation during power mains interruptions, it isrecommended that the Flocare® Infinit™ enteralfeeding pump be powered from an uninterrupti-ble power suppl of batter.

ZMNL400336_AW.indd 80 3/18/10 3:47:44 PM

Page 18: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-79-

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres(m) can be estimated using the equation applicable to the frequenc of the transmitter, where P is the maximum outputpower rating of the transmitter in Watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequenc range applies.NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and

reflection from structures, obects and people.

0,01 0,01 0,01 0,020,1 0,04 0,04 0,071 0,12 0,12 0,2310 0,37 0,37 0,74100 1,17 1,17 2,33

The Flocare® Infinit™ enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF dis-turbances are controlled. The customer or the user of the Flocare® Infinit™ can help prevent electromagnetic interferenceb maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and theFlocare® Infinit™ as recommended below, according to the maximum output power of the communications equipment.

Rated Maximum outputpower of transmitter

(W)

Separation distance according to frequenc of transmitter (m)

Recommended separation distances between portable and mobile RF communications euipment and theFlocare® Infinit™ enteral feeding pump

150 KHz to 80 MHzd = 0,12CDP

80 MHz to 800 MHzd = 0,12CDP

800 MHz to 2,5 GHzd = 0,23CDP

Portable and mobile RF communications equi-pment should be used no closer to an part ofthe Flocare® Infinit™, including cables, than therecommended separation distance calculatedfrom the equation applicable toe the frequencof the transmitter:Recommended separation distance

Where P is the maximum output power ratingof the transmitter in watts (W) according to thetransmitter manufacturer and d is the recom-mended separation distance in metres (m).Field strengths from fixed RF transmitters, asdetermined b an electromagnetic site surve,a

should be less than the compliance level ineach frequenc range.b

Interference ma occurin the vicinit of equipmentmarked with the followingsmbol:

Conducted RFIEC 61000-4-6

3 Vrms150 KHz to 80 MHz

d = 1,17CDP3 Vrms150 KHz to 80 MHz

Radiated RFIEC 61000-4-3

3 V/m80 Mhz to 2,5 GHz

d = 0,12CDPd = 0,23CDP

80 MHz to 800 MHz800 MHz to 2,5 GHz

30 V/m80 Mhz to 2,5 GHz

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobileradios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoreticall with accurac.To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site surve should beconsidered. If the measurement field strength in the location in which the Flocare® Infinit™ is used exceeds the appli-cable RF compliance above, the Flocare® Infinit™ should be observed to verif normal operation. If abnormal perfor-mance is observed, additional measures ma be necessar, such as reorienting or relocating the Flocare® Infinit™.

b. Over the frequenc range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.

NOTE 1 At 80 Mhz and 800 MHz, the higher frequenc range applies.NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and

reflection from structures, obects and people.

The Flocare® Infinit™ enteral feeding pump is intended for use in the electromagnetic environment specified below. Thecustomer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such an environment.

Guidance and manufacturer’s Declaration – Electromagnetic immunit

Immunit test IEC 60601 test level Electromagnetic environment – guidanceCompliance level

ZMNL400336_AW.indd 81 3/18/10 3:47:44 PM

Page 19: INSTRUCTIONSFORUSEfrankshospitalworkshop.com/equipment/documents... · ORDERINGINFORMATION Flocare ® Infinit ™ pumpWesternEurope corporatecode35676 Flocare ® Infinit ™ Charger

-80-

ZMNL400336_AW.indd 82 3/18/10 3:47:44 PM