  • PLT-110

    POINT LOAD TESTER

    OPERATING INSTRUCTIONS

    Copyright Geotechnical Consulting & Testing Systems1994-2012

  • PLT-110(Point Load Tester)

    i

    GCTS TESTING SYSTEMS

    PREFACE

    All Rights Reserved

    This document contains proprietary information. No part of this document may be copied,

    reproduced, transmitted, or translated into another language, in any form or by any method without

    the prior written consent of Geotechnical Consulting & Testing Systems, L.L.C. (GCTS).

    GCTS makes no warranties with respect to this documentation and disclaims any implied warranties

    of merchantability and fitness for a particular purpose. Information in this document is subject to

    change without notice. While GCTS makes every effort to ensure the accurateness of the information

    presented here, GCTS shall not be liable for errors that may appear in this document.

    Disclaimer

    Notwithstanding anything in this agreement to the contrary, GCTS shall have no liability whether

    arising in tort, or otherwise, for loss of capital, loss of product, loss of profit, loss of use, loss of

    power, power outages, cost of replacement power, or any indirect, special, incidental or

    consequential damages.

    GCTS shall not be liable for any direct, indirect, consequential or incidental damages resulting from

    the use or misuse of this product.

    GCTS does not authorize this equipment or any of its products and product components, for use in

    life support systems, or in any application where failure of the product can be reasonably expected to

    lead to personal injury and/or death.

    If the user does not wish to accept the terms of this disclaimer, the user may return this equipment

    (subject to the GCTS return policy, stated at www.gcts.com/terms.pdf) to GCTS within 30 days after

    purchase. If the user does not return the product within 30 days, this constitutes an acceptance by the

    user of the terms of this disclaimer.

    Limited Warranty

    GCTS warrants to the original Buyer only, that the equipment of its manufacture will be free from

    defects in material or workmanship for a period of 12 months after shipment from its factory or 12

    months from acceptance in the case of GCTS installation. In no case shall warranty be extended

    beyond a period of 15 months past shipment date (allowing a maximum of 3 months for installation

    and acceptance). GCTS shall, at its option, repair or replace free of charge within the warranty

    period any components or assemblies supplied by GCTS which prove to be defective in

    workmanship or materials, subject to the following:

    GCTS reserves the right to request the prepaid return of such defective items to its factory for inspection and evaluation.

    Expendable items (oil, seals, filters, etc.) and items subject to normal wear and/or replacement will not be covered under warranty if their failure is the result of normal wear.

    Items supplied by GCTS but not of its manufacture (computers, oscilloscopes, commercial software packages, etc.) carry the original manufacturers warranty which is passed on to the customer who is responsible to deal directly with the manufacturer on warranty issues.

    This warranty does not apply to damage due to accident, improper use, negligence, or modification of any part of the testing system.

    This warranty is extended only to the original Buyer of the equipment.

  • PLT-110(Point Load Tester)

    ii

    GCTS TESTING SYSTEMS

    This warranty is expressly in lieu of all other warranties expressed or implied, statutory or otherwise, including any implied warranty of merchantability or fitness or suitability for a specific

    purpose. No warranties are expressed or implied which extend beyond the description of the face

    hereof. In no event shall GCTS be held responsible or liable to the Buyer for collateral, indirect,

    incidental or consequential damages of any kind.

    GCTS assumes no liability for damages arising from the use of the equipment by the Buyer or any

    third party. By purchasing GCTS equipment, the Buyer assumes all liability for any damages of any

    kind which may result from its use or misuse by the Buyer, Buyer's employees, agents, contractors or

    any other third party unknown to GCTS or the Buyer, including damages due to failure of the

    equipment.

    Disclaimer

    The equipment warranty does not apply to damage due to accident, improper use, negligence, or

    modification of any part of the testing system!

    Geotechnical Consulting & Testing Systems, L.L.C shall not be liable for any direct, indirect,

    consequential or incidental damages resulting from the use of this product.

    Copyright Geotechnical Consulting & Testing Systems1994-2012

  • PLT-110(Point Load Tester)

    iii

    GCTS TESTING SYSTEMS

    Table of Contents

    PREFACE .......................................................................................................................................................... I All Rights Reserved ................................................................................................................................... i

    Disclaimer .................................................................................................................................................. i

    Limited Warranty ....................................................................................................................................... i

    Table of Contents ..................................................................................................................................... iii

    About this Manual ................................................................................................................................... iv

    User Responsibilities ................................................................................................................................ v

    Safety Practices ......................................................................................................................................... v

    Obtaining Technical Support .................................................................................................................. vii

    1.0 INTRODUCTION ............................................................................................................................................ 8 System Overview .................................................................................................................................... 10

    2.0 GENERAL .................................................................................................................................................... 11

    3.0 CALIBRATION PROCEDURES...................................................................................................................... 11 Gauge Setup ............................................................................................................................................ 11

    4.0 RUNNING A TEST ........................................................................................................................................ 12 4.1 Test Set-Up Procedure........................................................................................................................ 12 4.2 Performing a test ................................................................................................................................. 13

    5.0 SAFETY ....................................................................................................................................................... 14 Safety guideline while operating equipment: .......................................................................................... 14

    6.0 SOFTWARE ................................................................................................................................................. 14

    7.0 MAINTENANCE ........................................................................................................................................... 15

    8.0 GCTS DEVICES ORIGINAL ACCESSORIES ................................................................................................ 15 Appendix A : Wiring Schematics and Hardware ................................................................................ 17 Appendix B: Supplier Instruction Sheet ................................................................................................. 22 Appendix C: PLT-110 Shield Installation .............................................................................................. 29

  • PLT-110(Point Load Tester)

    iv

    GCTS TESTING SYSTEMS

    About this Manual

    Thank you for purchasing the PLT-110 (Point Load Tester). Please read this manual

    carefully before using the device. Should your device fail to operate correctly, contact

    Technical Support.

    DOCUMENT CONVENTIONS

    The following international symbols will be used in this document with the appropriate

    meanings:

    : This icon accompanies text dealing with hazards to personnel. When

    present, it indicates that a potential hazard to your personal safety exists

    if information stated within the WARNING paragraph is not adhered

    to or procedures are executed incorrectly.

    : This icon accompanies text dealing with potential damage to equipment

    when present, it indicates that there is a potential danger of equipment

    damage, software program failure or that a loss of data may occur if

    information stated within the CAUTION paragraph is not adhered to

    or procedures are executed incorrectly.

    NOTE : This identifier accompanies text dealing with potential situations which

    might cause inaccurate data to be gathered.

    WARNING

    CAUTION

  • PLT-110(Point Load Tester)

    v

    GCTS TESTING SYSTEMS

    User Responsibilities

    : Read the safety considerations before operating the hydraulic power

    supply. Failure to read and understand the operations manual and/or

    follow the recommended safety practices can result in serious personal

    injuries and/or damage to the equipment.

    : Never exceed the maximum equipment rating and capabilities!

    Safety Practices

    Review these guidelines to ensure that your current operating procedures do not result in

    hazardous situations. Although all hazards may not be able to be eliminated, the following

    guidelines can be used to identify hazards so that the proper training, operating procedures,

    and safety equipment can be implemented.

    System operators should fully review the documentation supplied to gain an understanding

    of the system functions. Figures will be placed throughout the manual and should be

    reviewed.

    To ensure smooth system operation during the first execution of the control software, the

    user should be very familiar with this section and the specific test module to be used.

    After the documentation (for this device) has been reviewed, make modifications to your

    current operating procedures which will minimize hazards. It may be beneficial to make an

    adjustment checklist that corresponds to your specific hardware.

    OPERATING SAFETY CONDITIONS

    The following operating safety considerations are applicable to most testing systems. The

    user is required to read each item below and determine if it is applicable to the testing

    system for which this (device) will be used. The user is also required to obtain and review

    all safety instructions on specific testing equipment used in the system.

    WARNING

    CAUTION

  • PLT-110(Point Load Tester)

    vi

    GCTS TESTING SYSTEMS

    EMERGENCY STOPS

    Determine the location of system emergency stop buttons to allow for quick emergency

    stops. Refer to the individual product manuals for instructions on how to wire

    additional emergency stop circuits.

    POTENTIAL CRUSH POINTS

    Know where the potential loads units pinch and crush points are and take appropriate

    safety precautions.

    SYSTEM INTERLOCKS

    System interlock devices should always be used and properly adjusted. Test all

    interlock devices for proper operation immediately before a test. Because of the

    possibility of operator error, system misadjustment, or component failure, interlock

    devices cannot be relied on to protect personnel, test specimens, or test equipment.

    USE AND HANDLING

    Do not disturb sensors.

    Ensure secure cable connections.

    Keep bystanders away.

    Stay clear of moving equipment.

    Be aware of electrical hazards.

  • PLT-110(Point Load Tester)

    vii

    GCTS TESTING SYSTEMS

    Obtaining Technical Support

    If you have any question about a GCTS system or product, contact Sales or Technical

    Support:

    For Technical Support please include the Job number and as much detailed

    information as possible:

    Phone: 1-480-456-0110 ext. 3

    Fax: 1-480-456-4727

    E-mail:[email protected]

    For General Inquiries or Sales:

    Phone: 1-480-456-0110 ext. 2

    Fax: 1-480-456-4727

    E-mail: [email protected]

    For Shipping Products:

    Address: 6103 S. Maple Ave.

    Tempe, Arizona 85283

    U.S.A

    Mail: P.O. Box 23686

    Tempe, Arizona 85285

    USA

    Telephone: 1-480-456-0110

    Fax: 1- 480-456-4727

    Internet: Home Page http://www.gcts.com

  • PLT-110(Point Load Tester)

    8

    GCTS TESTING SYSTEMS

    1.0 INTRODUCTION

    The GCTS Point Load Tester is an easy to use apparatus for measuring the point load

    strength index of random and irregular shaped rock specimens. It incorporates digital

    technology to increase precision and ease of use while reducing size and weight. The

    introduction of a pressure sensor to indicate load provides a better accuracy at any load

    level eliminating the imprecision of traditional pressure gauges at low loads.

    The point load test does not require costly specimen preparation and is a quick and simple

    test. The test can be performed in the field on irregular pieces of rocks allowing a large

    number of test results. Typically, tests are performed on pieces of drill core at least 1.4

    times as long as the diameter. Shape errors should be avoided for platy or discoid

    specimens by applying the load along the shortest specimen dimension. The point load

    strength index, Is, is defined as:

    Where, P = Load at rupture, and

    D = Distance between the loading point.

    The point load strength index is corrected for a specimen with a 50 mm diameter, and the

    following formula gives the size corrected point load strength index:

    Franklin and Dusseault give a size correction factor when there is no data to interpolate or

    extrapolate. The correction factor, F, is calculated as:

    A typical frequently used correlation between size corrected point load strength index and

    Unconfined compression strength is:

    Where, Unconfined compression strength for cylinders with a length to diameter ratio

    of 2 to 1, and

    C = Conversion Factor from to .

  • PLT-110(Point Load Tester)

    9

    GCTS TESTING SYSTEMS

    The Conversion Factor C has traditionally been taken as 24 (Goodman, 1989, p.37),

    however since point load testing is an index testing it is the responsibility of the user to

    ensure that they are using the right conversion factor for their rock type. This can be done

    by either relying on past data to establish a correlation between to or by testing

    several rock cores to establish a correlation. Conversion Factor C may vary

    significantly. According to Franklin and Dusseault, while the conversion factor is usually

    between 20 and 25, it can vary between 15 and 50. Rusnak and Mark found that the low

    end can even be 12.6. In theory, each rock type should have its own conversion factor.

    The user is encouraged to pick a Conversion Factor C of 24 only if such a conversion

    factor is justifiable.

    The system includes a meter display to read the applied load calibrated in kN as default.

    Other load units can also be used but recalibration of the meter display is necessary. A 0 to

    2 VDC output is also provided to connect to a data acquisition device.

    An advanced option for the GCTS Point Load Tester is the ultrasonic platens for the

    measurement of P wave velocity. The P wave velocity is a more rational and better

    correlation to strength than the point load index. The major advantage of the ultrasonic

    velocity option is that, like the point load test, it does not require expensive specimen

    preparation. With this option the P wave velocity is measured during the point load test.

    This option also gives the advantage of being able to correlate the P wave velocity with the

    point load strength on the same specimen in a very short period of time.

    Other options include unconfined compression loading platens, triaxial cells, and frames for

    performing indirect tension, uniaxial, and triaxial tests on small specimens. Loads up to 100

    kN (22 kips) can be applied. These options include software to capture the complete stress-

    strain curve and automatically calculate strength parameters.

  • PLT-110(Point Load Tester)

    10

    GCTS TESTING SYSTEMS

    System Overview

    Safety Acrylic Shield

    Hydraulic Cylinder

    Gauge

    Base plate

    Upper crosshead

    Load Points

    Specimen Hydraulic Hand Pump

    Figure 1: PLT-110 Side View

  • PLT-110(Point Load Tester)

    11

    GCTS TESTING SYSTEMS

    2.0 GENERAL

    Specifications

    Table 1: PLT-110 Specifications

    Parameter Value

    Hydraulic Ram Area 1.445x10-3

    m2 (2.24 in

    2)

    Hydraulic Ram Stroke 5.4 mm (2.13 in)

    Maximum Load 100 kN (10 ton)

    Pressure Sensor Range 69,000 kPa (10,000 psi)

    Analog Output 0 2 VDC

    Display Units Load: kN (default factory setting)

    Power 3.6 V C-size lithium battery

    3.0 CALIBRATION PROCEDURES

    Gauge Setup

    It is advantageous to setup the Gauge to remember and display the peak/maximum load,

    especially if the PLT is not connected to the PntLoad software. The instruction manuals for

    these gauges can either be downloaded from the internet or you can request them from

    GCTS by sending an email to [email protected] and identifying your gauge type (PG-

    10000 or PG10).

    For the model PG-10000 (old) gauges, the following procedure can be used to enable the

    PEAK Hi feature (section 2.0 in the instructions):

    Select PEAK Hi mode by pressing and holding the MODE key until the reading starts

    flashing. Release the MODE key, then press MODE key once more, the display will

    read HOLD. Press ENTR key once, the display will read NO H. Using keys to

    select H Hi (Peak High Hold). Press ENTR key to lock your selection.

  • PLT-110(Point Load Tester)

    12

    GCTS TESTING SYSTEMS

    NOTE: When the hold function is enabled, PEAK is displayed.

    For the model PG10 (new) gauges, the Peak-Hold feature is already enabled at the factory.

    To enable or disable it please see page 11 in the instructions. If the user does not like the

    Peak-Hold feature (because it only displays the peak/maximum at all times), the user can

    disable the Peak-Hold feature and use instead the Maximum/Minimum feature. This

    Maximum/Minimum feature does not need to be enabled, resetting it can be quite

    cumbersome, and the instructions to do that are given on page 8 in the instructions.

    4.0 RUNNING A TEST

    4.1 Test Set-Up Procedure

    For testing, the pump reservoir cap must be in the Vent position. The unit was shipped

    with the pump reservoir cap in the Closed position for transport, but during usage the

    pump reservoir cap must be in the Vent position!

    The operation of the GCTS Point Load Tester set up is simple and straightforward. First

    open the relief valve on the hand pump. If the meter displays an offset load use the TARE

    OUT or Zero feature.

    For the model PG-10000 (old) gauges, the following procedure can be used to TARE

    OUT or Zero (section 1.0 in the instructions): Press and hold the MODE key until the

    reading starts flashing. Release the MODE key. Press the ENTER key until the

    reading becomes 0.00.

    Place in the vent position

    when testing.

    Figure 2: hydraulic Pump reservoir cap set-up

  • PLT-110(Point Load Tester)

    13

    GCTS TESTING SYSTEMS

    For the model PG10 (new) gauges, the following procedure can be used to TARE OUT or

    Zero (page 4 in the instructions, Enter Button Secondary Function): Press and hold the

    Enter Button for approximately 3seconds. If it displays No Zero, the Tare Feature

    has to be enabled (see page 26 of the instructions).

    4.2 Performing a test

    Adjust the location of the upper crosshead so that the piston will not be fully extended

    during the test. It is best to have the loading piston operate within the lower region of its

    stroke to prevent alignment damage to the piston. This can occur when the piston is

    fully extended and loading against an uneven surface.

    Close the pump relief valve.

    Hold the specimen in between loading points, and begin pumping the handle until the

    load display indicates some initial contact.

    Read the displacement on the side ruler (or in the optional deformation meter display if

    available) and calculate the distance "D" between the point loads.

    Once "D" has been recorded, proceed pumping until the specimen fails (indicated by a

    sudden loss of the applied load).

    While the digital gage is in PEAK mode, the display will show the maximum load at all

    times. You must remember to reset the PEAK value before each test.

    To reset after capturing a Hi Peak:

    For the model PG-10000 (old) gauges, the following procedure can be used to TARE

    OUT or Zero (section 2.0 in the instructions): Press and hold MODE key until the

    reading starts flashing. Release the MODE key, then press ENTR, hold momentarily

    then release.

    For the model PG10 (new) gauges, the following procedure can be used to reset the Peak

    value (page 11 in the instructions): Press the (up arrow) key.

    Refer to the display unit instruction for further information if necessary

  • PLT-110(Point Load Tester)

    14

    GCTS TESTING SYSTEMS

    5.0 SAFETY

    Safety guideline while operating equipment:

    : Always use protective clothing and equipment to eliminate risk of personal

    injury. Do not operate this unit without goggles, gloves, and lab coat to

    protect your eyes, hands, and any other exposed body part. Specimen

    particles will explode outwards at failure.

    : Do not exceed the maximum pressure range for the installed sensor as

    indicated by the pressure gauge.

    IMPORTANT: Do not forget to reset the peak load on the meter before the next test is

    performed or the new failure load may not be recorded.

    6.0 SOFTWARE

    The GCTS PntLoad software for Windows has been designed to go along with the

    GCTS Point Load testing System in order to provide a complete package to test numerous

    specimens fast and with ease and to present the statistical results in a neat and organized

    format. The software can be configured to provide an automated system whereas very

    minimal interaction from the operator is needed as test can be started and stopped

    automatically based upon a threshold load value.

    Note: please refer to the PntLoad Software for the Point Load Strength Testing System

    Manual for a complete Guide and Reference.

    WARNING

    WARNING

  • PLT-110(Point Load Tester)

    15

    GCTS TESTING SYSTEMS

    7.0 MAINTENANCE

    Take care of your device

    Handle your device and accessories with care. The following suggestions help you protect

    your warranty coverage.

    Do not use or store the device in dusty areas. Moving parts and electronic components

    can be damaged.

    Do not attempt to open the device other than as instructed in the user guide.

    Unauthorized modifications may damage the device.

    8.0 GCTS DEVICES ORIGINAL ACCESSORIES

    For availability of approved accessories, check with customer support. An extensive range

    of accessories is available for your device.

    For more details contact the sales department at www.gcts.com

  • PLT-110(Point Load Tester)

    16

    GCTS TESTING SYSTEMS

    References

    Franklin, J. and Dusseault, M. (1989). Rock Engineering. NY: McGraw-Hill Publishing

    Company.

    Goodman, R. (1989). Introduction to Rock Mechanics, second edition. NY: John Wiley &

    Sons.

    Rusnak J. and Mark, C. (2000). Using the Point Load Test to Determine the Uniaxial

    Compressive Strength of Coal Measure Rock. NIOSHTIC-2 No. 20021059

  • PLT-110(Point Load Tester)

    17

    GCTS TESTING SYSTEMS

    Appendix A: Wiring Schematics and Hard

  • PLT-110(Point Load Tester)

    18

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    19

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    20

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    21

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    22

    GCTS TESTING SYSTEMS

    Appendix B: Supplier Instruction Sheet

  • PLT-110(Point Load Tester)

    23

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    24

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    25

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    26

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    27

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    28

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    29

    GCTS TESTING SYSTEMS

    Appendix C: PLT-110 Shield Installation

  • PLT-110(Point Load Tester)

    30

    GCTS TESTING SYSTEMS

    Operators Manual

    AUTOMAP RTO IDOU CNT S GROUP, INC. Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com E-mail: [email protected]

    PG10 Full Access DOC. 9003470 Rev A 9/10 PG10 Rev. A, 9/10

    2 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Table of Contents Warranty ................................................................................................... 3 Programming the PG10 .......................................................................4-32 Menu Flow Chart ................................................................................... 5 Mode Setting ......................................................................................... 6 Accessing/Exiting the Setup Menu ........................................................ 7 Maximum/Minimum Function ................................................................ 8 Units of Measure ................................................................................... 9 Using Custom Units ...................................................................... 10 Peak-Hold Feature ................................................................................. 11 Advanced Settings .........................................................................12-17 Auto-Off .......................................................................................... 12 Decimal Place ................................................................................ 13 Sample Rate .................................................................................. 14 Bar Graph ....................................................................................... 15 Range Adjustment .......................................................................... 16 Quick Calibration ............................................................................ 17 Datalogging ....................................................................................18-25

  • PLT-110(Point Load Tester)

    31

    GCTS TESTING SYSTEMS

    Date & Time ................................................................................... 19 Logging Sample Rate ..................................................................... 20 Event Logging ................................................................................ 22 Viewing Logged Readings .............................................................. 24 Enabling/Disabling the Clock ......................................................... 25 Tare Feature ........................................................................................ 26 Default ................................................................................................. 27 Outputs ...........................................................................................28-32 Analog Outputs ..........................................................................28-29 Trip Point Outputs .....................................................................30-32 Wiring ................................................................................................33-35 Specifi cations ....................................................................................36-37 Rev. A, 9/10 PG10

    3 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    APG warrants its products to be free from defects of material and workmanship

    and will, without charge, replace or repair any equipment found defective

    upon inspection at its factory, provided the equipment has been returned,

    transportation prepaid, within 24 months from date of shipment from factory.

    THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES

    ALL OTHER WARRANTIES NOT EXPRESSLY SET FORTH HEREIN,

    WHETHER EXPRESSED OR IMPLIED BY OPERATION OF LAW OR

    OTHERWISE INCLUDING BUT NOT LIMITED TO ANY IMPLIED

    WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A

    PARTICULAR PURPOSE.

    No representation or warranty, express or implied, made by any sales

    representative, distributor, or other agent or representative of APG which is

    not specifi cally set forth herein shall be binding upon APG. APG shall not be

    liable for any incidental or consequential damages, losses or expenses directly

    or indirectly arising from the sale, handling, improper application or use of the

    goods or from any other cause relating thereto and APGs liability hereunder, in any case, is expressly limited to the repair or replacement (at APGs option) of goods.

    Warranty is specifi cally at the factory. Any on site service will be provided at

    the sole expense of the Purchaser at standard fi eld service rates.

    All associated equipment must be protected by properly rated electronic/

    electrical protection devices. APG shall not be liable for any damage due

    to improper engineering or installation by the purchaser or third parties.

    Proper installation, operation and maintenance of the product becomes the

    responsibility of the user upon receipt of the product.

    Returns and allowances must be authorized by APG in advance. APG will

    assign a Return Material Authorization (RMA) number which must appear

  • PLT-110(Point Load Tester)

    32

    GCTS TESTING SYSTEMS

    on all related papers and the outside of the shipping carton. All returns are

    subject to the fi nal review by APG. Returns are subject to restocking charges as

    determined by APGs Credit Return Policy.

    Warranty and Warranty Restrictions PG10 Rev. A, 9/10

    4 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Mode Button

    Primary Function: Used to access the main setup menu or to return to

    the main setup menu from any of the submenus.

    Secondary Function: Used to cycle the gauge on and off by pressing and

    holding for approximately 2 seconds.

    Enter Button

    Primary Function: Used to accept the value displayed within each menu

    setting.

    Secondary Function: Used to Zero the gauge or Tare the gauge (when

    the tare feature is enabled) by pressing and holding for approximately 3

    seconds.

    Up & Down Arrows

    Primary Function: Used to cycle through all options and settings within

    the menu system.

    Secondary Function: Used to cycle between the Minimum, Maximum

    and the present pressure reading.

    Each of the 4 buttons on the PG10 performs dual functions; one within the setup

    menu system, and another when in standard operating mode.

    Programming the PG10 MODE

    ENTR Rev. A, 9/10 PG10

    5 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

  • PLT-110(Point Load Tester)

    33

    GCTS TESTING SYSTEMS

  • PLT-110(Point Load Tester)

    34

    GCTS TESTING SYSTEMS

    PG10 Rev. A, 9/10

    6 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Operating Modes: The PG10 has 4 standard operating modes to choose from:

    Mode 000: Full Access

    - Provides access to all menu settings. If no buttons are pushed for 1 minute,

    the gauge will revert back to Mode 003 (factory default).

    Mode 002: Limited Access

    - Menu is locked.

    - zeros the reading (press and hold for 3 seconds).

    - or cycles between the Max and Min readings.

    - On/Off functions only on battery powered gauges.

    Mode 003: Factory Default

    - Full access except Full Scale Calibration is locked.

    Mode 005: Locked Access

    - All buttons locked except will power on/off battery powered gauges.

    Accessing the Operating Mode Setting: Step 1: Simultaneously press and hold and for approximately

    3 seconds. This will bring up the 3 digit mode number.

    Step 2: Enter the desired mode number (see Operating Modes above) by

    using or to change the value of the fl ashing digit, and

    to accept the value and advance to the next digit. MODE

    ENTR ENTR ENTR MODE MODE

    Rev. A, 9/10 PG10

    7 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Accessing the Setup Menu: Step 1: Simultaneously press and hold and for approximately 3

    seconds. This will bring up the 3 digit mode number.

    Step 2: Using or to change the value of the fl ashing digit, and

    to advance to the next digit, change the mode number to 000.

    Step 3: Press to enter the setup menu, and press or to scroll

    through menu choices.

  • PLT-110(Point Load Tester)

    35

    GCTS TESTING SYSTEMS

    Exiting the Menu System: Step 1: While in the main setup menu, press or until EXIT is

    displayed.

    Step 2: Press to exit the menu system. MODE ENTR ENTR MODE

    ENTR

    PG10 Rev. A, 9/10

    8 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Maximum/Minimum Reset (MAXMIN): Pressing the or button while in operating mode will cycle between

    displaying the present pressure reading, the maximum pressure reading and

    the minimum pressure reading. The maximum and minimum readings will be

    stored until the gauge is powered down or the max/min readings are reset.

    Resetting the Max/Min readings: Step 1: Press once to enter the main setup menu.

    Step 2: Press or cycle through the options until MAXMIN is

    displayed.

    Step 3: Press to access the Max/Min reset options.

    Step 4: Press or to toggle between YES and NO until YES is

    displayed.

    Step 5: Press to reset the Max/Min readings and return to the main setup

    menu. MODE

    ENTR ENTR

    Rev. A, 9/10 PG10

    9 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Units of Measure (UNITS): Allows the user to select the unit of measure to be displayed as the pressure

    reading.

    Options:

    For gauges over 120 psi:

    PSI (pounds per square inch)

    bAR (bar)

    KPA (kilopascals)

    *CUSTOM (see Using Custom Units on next page) KGCM^2 (kilograms per cubic centimeter)

    MPA (megapascals)

    FTH20 (feet of water @ 60 F)

    cmHG (centimeters of mercury)

  • PLT-110(Point Load Tester)

    36

    GCTS TESTING SYSTEMS

    For gauges less than 120 psi:

    PSI (pounds per square inch)

    bAR (bar)

    KPA (kilopascals)

    *CUSTOM (see Using Custom Units on next page) mbAR (millibar)

    INHG (inches of mercury)

    INH20 (inches of water @ 60 F)

    mmHG (millimeters of mercury)

    Setting the Unit of Measure: Step 1: Press once to enter the main setup menu.

    Step 2: Press or to cycle through the options until UNITS is displayed.

    Step 3: Press to access the Units options.

    Step 4: Press or to cycle through setting options until the desired unit

    of measure is displayed.

    Step 5: Press to apply the setting and return to the main setup menu. MODE

    ENTR ENTR

    PG10 Rev. A, 9/10

    10 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Using Custom Units (CUSTOM): The Custom Units setting allows the user to display a volumetric weight by

    applying a conversion factor to the pressure reading.

    NOTE: The conversion factor must be calculated using PSI as the base unit of measure.

    Setting the Custom Units feature: Step 1: Calculate the conversion factor from psi to the desired unit of measure.

    Step 2: Press once to enter the main setup menu.

    Step 3: Press or to cycle through the options until UNITS is displayed.

    Step 4: Press to access the Units setting options.

    Step 5: Press or to cycle through the Units options until CUSTOM is

    displayed.

    Step 6: Press to access the Custom Units setting. A 5-digit conversion

    factor will appear with the fi rst digit fl ashing.

    Step 7: Press or to change the value of the fl ashing digit.

    Step 8: Press to accept the fl ashing digit and advance to the next

    digit......repeat steps 7 and 8 as necessary.

    Step 9: After the last digit is accepted by pressing , use or

    scroll through the custom units of measure; LBF, KN, LBS, KG,

    NEWTON, TON, TONNES.

    Step 10: Press to accept the custom unit.

  • PLT-110(Point Load Tester)

    37

    GCTS TESTING SYSTEMS

    Step 11: Press or to shift the decimal point position.

    Step 12: Press to accept the decimal point position and return to the main

    setup menu. MODE

    ENTR ENTR ENTR ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    11 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Peak-Hold (P HOLd): When the Peak-Hold is enabled, the gauge will display the peak or maximum pressure reading since the gauge was powered on or the Max/Min value was

    reset.

    NOTE 1: When the Peak-Hold feature is enabled, the icon will be displayed in the upper left area of the display.

    Options: Off or On

    Enabling the Peak-Hold feature: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until P HOLd is displayed.

    Step 3: Press to access the Peak-Hold setting options.

    Step 4: Press or to toggle between OFF and ON.

    Step 5: Press to apply the displayed setting and return to main setup menu.

    NOTE 2: The peak value can be reset by pressing MODE

    ENTR ENTR

    PEAK PG10 Rev. A, 9/10

    12 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Advanced Settings (AdVSET): The Advanced Settings menu is used to customize the LCD display and to setup

    any optional features, such as an analog output.

    Auto-Off (AUTO): This function is applicable to battery powered units only. The Auto-Off feature

    allows the user to designate the time of inactivity (no buttons pushed) until the

    gauge automatically powers down.

    Options: OFF, 2 MIN, 4 MIN, 8 MIN, 16 MIN, 32 MIN and LIGHT

    NOTE 1: Selecting OFF disables the Auto-Off feature; the gauge will then remain powered indefi nitely so long as suffi cient voltage is being supplied

    (>1.8V).

    NOTE 2: Selecting LIGHT will keep the gauge continuously powered as

  • PLT-110(Point Load Tester)

    38

    GCTS TESTING SYSTEMS

    long as the light sensor in the PG10 detects adequate light levels. When the

    environment is suffi ciently dark, the PG10 will automatically shut down until

    the lighting once again reaches the threshold of the light sensor.

    Setting the Auto-Off feature: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the menu options until AdVSET is

    displayed.

    Step 3: Press to enter the Advanced Settings menu.

    Step 4: Press or to cycle through the menu options until AUTO is

    displayed.

    Step 5: Press to access the Auto-Off setting options.

    Step 6: Press or to cycle through setting options until the desired

    setting is displayed.

    Step 7: Press to apply the setting and return to advanced setup menu.

    Step 8: To exit the advanced setup menu, press or until EXIT is

    displayed and press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    13 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Decimal Place (dEC PL): The reading can be set to display in High Resolution (HI RES), Medium

    Resolution (MEdRES) or Low Resolution (LO RES) mode. Switching between

    resolutions will shift the displayed reading by one decimal place position.

    NOTE: Gauges without a decimal place position will display a dummy zero (or zeros) when the resolution is changed to medium or low.

    Options: HI RES (high resolution), MEdRES (medium resolution) or LO RES

    (low resolution)

    Setting the Decimal Place feature: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until AdVSET is

    displayed.

    Step 3: Press to enter the Advanced Settings menu.

    Step 4: Press or to cycle through the options until dEC PL is

    displayed.

    Step 5: Press to access the Decimal Place setting options.

    Step 6: Press or to cycle through the resolution settings.

    Step 7: Press to apply the displayed setting and return to advanced setup

    menu.

    Step 8: To exit the advanced setup menu, press or until EXIT is

  • PLT-110(Point Load Tester)

    39

    GCTS TESTING SYSTEMS

    displayed and press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    14 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Sample Rate (SAMPLE): Adjusts the rate at which the gauge takes sample readings.

    NOTE: Setting the Sample Rate to SLOW will help preserve battery life (when applicable) and will also help to smooth rapidly fl uctuating

    readings.

    Options: SLOW (4x/second), MEdIUM (8x/second), FAST (16x/second)

    Setting the Sample Rate feature: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until AdVSET is

    displayed.

    Step 3: Press to access the Advanced Settings menu.

    Step 4: Press or to cycle through the options until SAMPLE is

    displayed.

    Step 5: Press to access the Sample Rate setting options.

    Step 6: Press or to cycle through the setting options until the desired

    setting is displayed.

    Step 7: Press to apply the displayed setting and return to advanced setup

    menu.

    Step 8: To exit the advanced setup menu, press or until EXIT is

    displayed and press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    15 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Bar Graph 0% (bAR 0) & Bar Graph 100% (bAR100) Settings: Allows the user to defi ne the reading values associated with 0% and 100% on

    the 2700 display bar graph. Bars will appear/disappear in 10% increments of the

    total span between the two values.

    NOTE: The 0% reference does not have to be the lower pressure setting; 0% can be set as the higher pressure setting, thereby causing the bar graph to

    increase as the pressure decreases. Negative pressure settings can also be used

    as either the 0% or 100% reference points.

    Setting the Display Bar Graph:

  • PLT-110(Point Load Tester)

    40

    GCTS TESTING SYSTEMS

    Step 1: Press once to enter the main setup menu.

    Step 2: Press or to cycle through the options until AdVSET is

    displayed.

    Step 3: Press to access the Advanced Settings menu.

    Step 4: Press or to cycle through the options until BAR 0 is displayed.

    Step 5: Press to access the Bar Graph 0% value. A 5-digit number will

    appear with the fi rst digit fl ashing.

    Step 6: Press or to change the value of the fi rst fl ashing digit.

    Step 7: Press to accept the value of the fl ashing digit and advance to the

    next digit......repeat steps 6 and 7 until the desired 0% reading is fully

    entered.

    Step 8: Press or to cycle through the options until BAR100 is displayed.

    Step 9: Repeat Steps 5-7 to enter the Bar Graph 100% value.

    Step 10: To exit the advanced setup menu, press or until EXIT is

    displayed and press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    16 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Full-Scale Range Adjust (RANGE): Allows the user to adjust the reading at full-scale pressure. The reading can be

    adjusted by +/-10% full-scale.

    NOTE: The pressure reading must be within 5% of the full-scale value in order to make Range adjustments. For example, a 1000 psi gauge would need

    to be reading between 950 psi and 1050 psi in order to adjust the Range

    feature. If the reading is not within 5% of full scale, NOAdJU (No

    Adjustment) will be displayed when trying to adjust the Range.

    Adjusting the Full-Scale Range: Step 1: Ensure that the pressure reading is within 5% of full-scale.

    Step 2: Press to enter the main setup menu.

    Step 3: Press or to cycle through the options until AdVSET is

    displayed.

    Step 4: Press to access the Advanced Settings menu.

    Step 5: Press or to cycle through the options until RANGE is displayed.

    Step 6: Press to enter the Range adjust mode.

    Step 7: Press to increase the reading, or press to decrease the reading.

    Step 8: Press to accept the adjusted reading and return to the advanced

    setup menu.

    Step 9: To exit the Advanced Setup menu, press or until EXIT is

    displayed and press to exit to the main setup menu. MODE

  • PLT-110(Point Load Tester)

    41

    GCTS TESTING SYSTEMS

    ENTR ENTR ENTR MODE

    Rev. A, 9/10 PG10

    17 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Fast Calibration (CALIBR): Allows the user to perform a fast calibration of the zero and span.

    NOTE: Both zero pressure and full-scale pressure must be applied to the gauge in order to complete the quick calibration process.

    Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the menu options until AdVSET is

    displayed.

    Step 3: Press to enter the Advanced Settings menu.

    Step 4: Press or to cycle through the options until CALIBR is

    displayed.

    Step 5: Press to access the CALIBR mode. The gauge will display the

    word ZERO, prompting the user to perform the zero pressure quick

    calibration.

    Step 6: Ensure no pressure is applied to the gauge. Press to zero the gauge.

    The gauge will then display the word SPAN, prompting the user to

    perform the full-scale pressure quick calibration.

    Step 7: Apply full pressure to the gauge and press to complete the quick

    calibration and return to advanced setup menu.

    Step 9: To exit the advanced setup menu, press or until EXIT is

    displayed and press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    18 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Datalogging (dATLOG): When datalogging is enabled, the PG10 will log the pressure readings, along

    with the associated date/time stamps, at an interval determined by the Logging

    Sample Rate setting (see page 20).

    The PG10 can be confi gured to log continuously, based solely on the Logging

    Sample Rate setting, or to log only when user defi ned pressure conditions exist

    (event logging). The gauge will continue to log readings at the Logging Sample

    Rate interval for as long as the event conditions persist.

    The gauge can store a maximum of 60 records. Each record is logged in

    a FIFO (fi rst-in, fi rst-out) buffer. This means that once the buffer is full (60

  • PLT-110(Point Load Tester)

    42

    GCTS TESTING SYSTEMS

    records), the oldest record will be overwritten by the next logged reading. Rev. A, 9/10 PG10

    19 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Date (DATE) and Time (TIME): Used to set the date and time, which will be used as the reference for the time

    stamp of the logged readings.

    NOTE: Hours are set using the 24 hour system (e.g. 1:00 p.m. is 13:00). Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the menu options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu options.

    Step 4: Press or to cycle through the menu options until DATE is

    displayed.

    Step 5: Press to access the Date settings. A 2-digit month number will

    appear with the fi rst digit fl ashing.

    Step 6: Press or to cycle the fi rst digit, and to accept the value

    and advance to the second digit. Repeat the same process for

    the second digit and press to accept and advance to the day of the

    month setting, which will appear with the fi rst digit fl ashing.

    Repeat the process one last time by entering the year.

    Step 7: Press to apply the setting and return to Datalogging menu.

    Step 8: Press or to cycle through the menu options until TIME is

    displayed.

    Step 9: Press to access the time settings. Repeat the process used to set the

    date to set the time (hour, minute, seconds). MODE

    ENTR ENTR ENTR ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    20 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Logging Sample Rate (SAMPLE): Determine the interval between logged readings whenever logging is activated.

    Options: 1 second (1s), 30 seconds (30s), 1 minute (1m), 30 minutes (30m),

    1 hour (1h), and 1 day (1day).

    Setting the Logging Sample Rate: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

  • PLT-110(Point Load Tester)

    43

    GCTS TESTING SYSTEMS

    Step 4: Press or to cycle through the options until SAMPLE is

    displayed.

    Step 5: Press to access the Logging Sample Rate setting options.

    Step 6: Press or to cycle through the sample rate settings.

    Step 7: Press to apply the displayed setting and return to Datalogging

    menu. MODE

    ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    21 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Logging Control Menu (LOG): Used to enable/disable datalogging, and to clear all logged records.

    Setting the Logging Type (continuous or event): The PG10 can be set to continuously log readings or to only log readings based

    on the Trip Relay status.

    NOTE 1: Gauges not equipped with relay outputs will still use the Trip Relay settings to defi ne the pressure range(s) for the event logging (see pages 30-32

    programming details).

    NOTE 2: Whenever the PG10 is logging, whether in continuous or event mode, the interval between logged readings is determined by the Logging Sample Rate

    setting (see page 20).

    Enabling Continuous or Event Logging Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

    Step 4: Press or to cycle through the options until LOG is displayed.

    Step 5: Press to access the Logging Control options.

    Step 6: Press or to cycle through the setting options until CONTIN or

    EVENT is displayed.

    Step 7: Press to enable logging and return to logging control menu. MODE

    ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    22 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Setting Event Logging: When Event Logging is enabled, the gauge will log readings based on the status

    of one or both of the Trip Relays (see Trip Relay programming on pages 30-32).

    NOTE: Gauges not equipped with relay outputs will still use the Trip Relay

  • PLT-110(Point Load Tester)

    44

    GCTS TESTING SYSTEMS

    settings for event logging.

    Event logging can be set to initiate on one of the following conditions:

    When Trip 1 is On (T1ON)

    When Trip 2 is On (T2ON)

    When both Trip 1 & Trip 2 are On (T1T2ON)

    When both Trip 1 & Trip 2 are Off (T1T2OF)

    When Trip 1 is Off (T1OFF)

    When Trip 2 is Off (T2OFF)

    Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

    Step 4: Press or to cycle through the options until EVENT is displayed.

    Step 5: Press to access the Event options.

    Step 6: Press or to cycle through the options until the desired Event

    condition is displayed.

    Step 7: Press to accept the event condition and return to the Datalogging

    menu. MODE

    ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    23 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Disabling Logging (STOP) and Clearing Logged (CLEAR) Readings: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

    Step 4: Press or to cycle through the options until LOG is displayed.

    Step 5: Press to access the Logging Control options.

    Step 6: Press or to cycle through the setting options until STOP or

    CLEAR is displayed.

    Step 7: Press to apply the displayed option (stop or clear) and return to

    Datalogging menu. MODE

    ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    24 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Viewing Logged Readings (VIEW): The latest reading will be the fi rst record displayed when viewing the logged

  • PLT-110(Point Load Tester)

    45

    GCTS TESTING SYSTEMS

    readings.

    Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

    Step 4: Press or to cycle through the options until VIEW is displayed.

    Step 5: Press to access the logged readings.

    Step 6: Press to cycle back through progressively older readings. will

    cycle forward through progressively newer records.

    Step 7: Press to exit the logged readings.

    NOTE: When viewing a logged record, the time stamp of the record will toggle momentarily to display the date stamp once every 8 seconds. MODE

    ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    25 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Enabling and Disabling the Clock on the main display (CLOCK): When enabled, the time will be displayed on the lower line (menu line) of the

    main display, alternating every two seconds with the unit of measure.

    Options: ON or OFF

    Enabling or Disabling the Main Display Clock: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dATLOG is

    displayed.

    Step 3: Press to enter the Datalogging menu.

    Step 4: Press or to cycle through the options until CLOCK is

    displayed.

    Step 5: Press to access the Clock on/off option.

    Step 6: Press or to cycle between the options.

    Step 7: Press to apply the displayed option and return to Datalogging

    menu. MODE

    ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    26 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Tare (TARE): Enabling the Tare function will set the current pressure reading as the zero

    reference pressure in order to measure a net change in pressure as opposed to

    measuring the gross pressure.

  • PLT-110(Point Load Tester)

    46

    GCTS TESTING SYSTEMS

    Enabling the Tare feature: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until TARE is displayed.

    Step 3: Press to access the Tare setting options.

    Step 4: Press or to toggle between OFF and ON.

    Step 5: Press button to apply the displayed option and return to main setup

    menu.

    NOTE 1: When tare function is enabled, the icon will appear in the upper center of the display.

    WARNING: Do NOT disconnect the gauge from the pressure fi tting while the tare function is enabled. The gauge could still be under pressure even

    though the reading shows 0.

    NOTE 2: If the maximum gross full-scale pressure value is reached while the Tare feature is enabled, the PG10 will automatically disable the Tare feature

    and return to reading the gross pressure in order to help prevent the user from

    accidently overpressuring the gauge.

    NOTE 3: Pressing and holding with the tare function enabled will re-tare the gauge at the current pressure value. MODE

    ENTR ENTR

    TARE ENTR

    Rev. A, 9/10 PG10

    27 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Default (dEFAUL): Used to reset the gauge to the factory default settings.

    Resetting the gauge to factory default settings: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until dEFAUL is

    displayed.

    Step 3: Press to access the Default options.

    Step 4: Press or to cycle between NO and YES.

    Step 5: Press to apply the displayed setting and return to main setup menu. MODE

    ENTR ENTR

    PG10 Rev. A, 9/10

    28 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    OUTPUT: Used to confi gure any optional gauges outputs, such as an analog signal or trip

    point relays.

  • PLT-110(Point Load Tester)

    47

    GCTS TESTING SYSTEMS

    Analog Low (AL SET) & Analog High (AH SET) Setpoints: Allows the user to defi ne the reading values associated with the Low Analog

    signal (i.e. 4mA or 0V) and the High Analog signal (i.e. 20mA, 2V or 5V).

    NOTE: The analog setpoints must be entered using the gauges base unit of measure (typically PSI).

    Setting the Analog Signal Span: Step 1: Press to enter the main setup menu.

    Step 2: Press or to cycle through the options until OUTPUT is

    displayed.

    Step 3: Press to access the Output Settings menu.

    Step 4: Press or to cycle through the options until AL SET is displayed.

    Step 5: Press to access the Analog Low setpoint value. A 5-digit number

    will appear with the fi rst digit fl ashing.

    Step 6: Press or to change the value of the fi rst fl ashing digit.

    Step 7: Press to accept the value of the fl ashing digit and advance to the

    next digit......repeat steps 6 and 7 until the desired Analog Low reading

    is fully entered.

    Step 8: Press to reenter the Output Settings menu.

    Step 9: Press or to cycle through the options until AH SET is displayed.

    Step 10: Repeat Steps 5-7 to enter the Analog High setpoint value.

    Step 11: To exit the output menu, press or until EXIT is displayed and

    press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    29 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Analog Low (AL CAL) & Analog High (AH CAL) Calibration: Allows the user to calibrate or trim the endpoints of the analog signal output (i.e. 4mA & 20mA or 0V & 2V/5V)

    Calibrating the Analog Signal End-Points: Step 1: Use a calibrated meter to monitor the analog output signal.

    Step 2: Force a low analog output signal (i.e. 4mA or 0V) either by adjusting

    the applied pressure or by adjusting the Analog Setpoints (see Analog Setpointson page 18 for details). Step 3: Press to enter the main setup menu.

    Step 4: Press or to cycle through the options until Output is displayed.

    Step 5: Press to access the Output Settings menu.

    Step 6: Press or to cycle through the setup options until AL CAL is

    displayed.

    Step 7: Press to access the Analog Low Calibration value. A 5-digit

  • PLT-110(Point Load Tester)

    48

    GCTS TESTING SYSTEMS

    number will appear with the fi rst digit fl ashing.

    Step 8: Press or to change the value of the fl ashing digit.

    NOTE: Changing the digit farthest to the left will produce the coarsest

    adjustment, while each successive digit moving to the right will cause

    subsequently fi ner adjustments.

    Step 9: Press to accept the value of the fl ashing digit and advance to the

    next digit......repeat steps 7 and 8 until the desired Analog Output

    value is displayed on the meter.

    Step 10: Press to reenter the Output Settings menu.

    Step 11: Press or to cycle through the options until AH CAL is

    displayed.

    Step 12: Repeat Steps 7-9 to adjust the Analog High Calibration value.

    Step 13: To exit the output menu, press or until EXIT is displayed and

    press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR ENTR

    PG10 Rev. A, 9/10

    30 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Trip Points: The PG10 can be confi gured with dual relay trip-point outputs. The relays can either be solid-state or SPDT mechanical in form. Each trip point

    (T1 and T2) can be confi gured to perform one of six different logic functions as

    described below and on the chart on the next page.

    NOTE: The or icon will appear in the upper left area of the display whenever the associated trip point is active.

    Trip Relay Output Types (T1TYPE or T2TYPE): determines the logic function of the output relays (see chart on next page).

    Type 0 or Normally Closed setting will close the trip relay whenever the pressure is less than the Trip Pressure setting.

    Type 1 or Exclusive setting will close the trip relay whenever the pressure is less than the Trip Pressure or greater than the Trip Pressure + Trip Window.

    Type 2 or Normally Closed with Hysteresis setting will close the trip relay when the pressure drops below the Trip Pressure. Once closed, the trip relay

    will remain closed until the pressure becomes greater than the Trip Pressure +

    Trip Window, at which point the trip relay will open. Once open the trip relay

    will remain open until the pressure once again drops below the Trip Pressure

    setting.

    Type 3 or Normally Open setting will close the relay whenever the pressure is greater than the Trip Pressure setting.

    Type 4 or Inclusive setting will close the relay whenever the pressure is within the Trip Window pressure range (greater than the Trip Pressure, less than

  • PLT-110(Point Load Tester)

    49

    GCTS TESTING SYSTEMS

    the Trip Pressure + Trip Window).

    Type 5 or Normally Open with Hysteresis is used for pump up applications, or for a low level alarm with hysteresis function. This is the opposite function

    of the Type 2.

    Type 6 Trip Disable T1 T2 Rev. A, 9/10 PG10

    31 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Trip Pressure: determines the lower pressure point for the trip output function. Trip Window: determines the pressure range between the lower and upper trip pressures. Used for Types 1, 2, 4, 5 (see chart below). PG10 Rev. A, 9/10

    32 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Setting the Trip Point Outputs: Step 1: Press once to enter the main setup menu.

    Step 2: Press or to cycle through the options until OUTPUT is

    displayed.

    Step 3: Press to access the Output Settings menu.

    Step 4: Press or to cycle through the options until T1TYPE (or

    T2TYPE) is displayed.

    Step 5: Press to access the Trip Type setting.

    Step 6: Press or to change the setting to the desired trip type function.

    Step 7: Press to accept the trip type function and return to the Output

    Settings menu.

    Step 8: Press or to cycle through the Output menu options until

    T1PRES (or T2PRES) is displayed.

    Step 9: Press to enter the Trip Pressure setting. A 5-digit number will

    appear with the fi rst digit fl ashing.

    Step 10: Press or to change the value of the fi rst fl ashing digit.

    Step 11: Press to accept the value of the fl ashing digit and advance to the

    next digit......repeat steps 10 and 11 until the desired Trip Pressure is

    fully entered.

    Step 12: Press or to cycle through the Output menu options until T1WIN

    (or T2WIN) is displayed.

    Step 13: Press to enter the Trip Window setting. A 5-digit number will

    appear with the fi rst digit fl ashing.

    Step 14: Press or to change the value of the fi rst fl ashing digit.

    Step 15: Press to accept the value of the fl ashing digit and advance to the

    next digit......repeat steps 14 and 15 until the desired Trip Window is

  • PLT-110(Point Load Tester)

    50

    GCTS TESTING SYSTEMS

    fully entered.

    Step 11: To exit the output menu, press or until EXIT is displayed and

    press to exit to the main setup menu. MODE

    ENTR ENTR ENTR ENTR ENTR ENTR ENTR ENTR

    Rev. A, 9/10 PG10

    33 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Wiring the PG10 Battery Replacement: Step 1: Untwist the display locking ring (on the face of the gauge) counterclockwise

    until the ring releases the display from the gauge.

    Step 2: Remove the front display to access the batteries.

    Step 3: Replace the front display ensuring proper alignment and secure into

    place using the locking ring. 3 5 2 8 6 4 7 1

    8_Pin Connector Wiring: PG10 Pin Out Table L1 L2 L4-C2 4-20 mA 0-2 VDC with Battery Power Power 2 SS Relays with External Power 8 Pin Connector 1 + Excitation n/a + Excitation 2

    - Excitation + Output - Excitation 3

    n/a - Output 4 n/a n/a 5 n/a n/a 6 7 8

    n/a n/a L4-C4 2 SPDT Relays with External Power

    + Excitation - Excitation

  • PLT-110(Point Load Tester)

    51

    GCTS TESTING SYSTEMS

    Relay 1 Com* Relay 1 NC* Relay 1 NO* Relay 2 Com* L5 RS-485 with External Power

    + Excitation - Excitation SS Relay 1* SS Relay 1* SS Relay 2* SS Relay 2* *Optional (C2 or C4) relay outputs n/a indicates not applicable. L3 0-5 VDC with External

    + Excitation + Output - Excitation n/a n/a n/a n/a Relay 2 NC* Relay 2 NO* RS-485 (A) RS-485 (B) SS Relay 1* SS Relay 1* SS Relay 2* SS Relay 2* n/a n/a SS Relay 1* SS Relay 1* SS Relay 2* SS Relay 2* - Output PG10 Rev. A, 9/10

    34 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected] 1 2 3 4 EXCITATION + EXCITATION Gauge Connector

    L1: 4-20 mA Circuit (C2 & C4 relay option not available) 4-20 mA CURRENT LOOP 5 6 7 8 Controller Process R L +Vin

  • PLT-110(Point Load Tester)

    52

    GCTS TESTING SYSTEMS

    Wiring (continued) Gauge

    L2: 0-2 V (battery powered) (C2 & C4 relay options not available) +Vin GROUND 1 2 3 4 5 6 7 8 Connector Controller Process Gauge Connector SS RELAY 1 SS RELAY 1 SS RELAY 2 SS RELAY 2 +Vin GROUND 1 2 3 4 5 6 7 8

    L3: 0-5 V (SS relays on C2 option only) + SUPPLY VOLTAGE Controller Process

    Rev. A, 9/10 PG10

    35 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected] + SUPPLY VOLTAGE EXCITATION Gauge Connector

    L4-C2: External Power with Dual SS Relays SS RELAY 1 SS RELAY 2 1 2 3 4 5 6 7 8 + SUPPLY VOLTAGE EXCITATION Gauge Connector

    L5: RS-485 SS RELAY 1 SS RELAY 2 1 2 3 4 5 6 7 8 RS-485 (B) Master Device

    RS-485 (A)

  • PLT-110(Point Load Tester)

    53

    GCTS TESTING SYSTEMS

    (SS relays on C2 option only) + SUPPLY VOLTAGE EXCITATION Gauge Connector

    L4-C4: External Power with Dual SPDT Relays Relay 1 Normally Closed Relay 1 Normally Open Relay 1 Common Relay 2 Normally Closed Relay 2 Normally Open Relay 2 Common 1 2 3 4 5 6 7 8

    Wiring (continued) PG10 Rev. A, 9/10

    36 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Specifi cations: Overpressure: 2x full scale Burst Pressure: 4x full scale or 20,000 psig, whichever is less. Accuracy (linearity & hysteresis): +/-0.25% (B.F.S.L.) Environmental: Housing: IP65 Compensated Temp: 20 to 1300F (-7 to 540C) Storage Temp: -40 to 1600F (-40 to 710C) Operating Temp: 0 to 1600F (-18 to 710C) Electrical: Batteries: (2) standard C cell External Power: 9-28 VDC Physical: Weight: 1.6 lb (0.73 kg) Case Material: Injection molded Phenolic Rev. A, 9/10 PG10

    37 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Output Specifi cations: 4-20 mA Output: Input Voltage (Excitation): 9 VDC min (no load) to 28 VDC max Input Current: 3-30 mA max Signal Variance: +/-0.16 mA at set points Output/Input: 2 wire loop powered Resolution: 14 bit

  • PLT-110(Point Load Tester)

    54

    GCTS TESTING SYSTEMS

    Protection: Reversed polarity 0-2 VDC Output: Input Voltage (excitation): Battery powered Output: Zero set point is +/-0.15 V with a 2 VDC span +/-0.02 VDC Output/Input: 2 wire Resolution: 14 bit 0-5 VDC Output: Input Voltage (Excitation): 9 to 28 VDC Input Current: 6 mA max Output: 0-5 VDC / +/-0.5 VDC at set points Output/Input: Non-isolated 3 wire Resolution: 14 bit Protection: Reversed polarity Solid State Relay Trip Points: Maximum Switched Voltage: 120 V AC/DC Maximum Switched Current: 120 mA SPDT Relay Trip Points: Maximum Switched Voltage: 125 VAC; 30 VDC Maximum Switched Current: 0.5 A Notes 38 PG5 Rev. A, 6/09

    Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    Rev. A, 6/09 PG5

    Notes 39 Automation Products Group, Inc. APG...Providing tailored solutions for measurement applications Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com [email protected]

    AUTOMATION APG...Providing tailored solutions for measurement applications To order additional copies of this manual, ask for APG part number PN9003470 Rev. A, 9/10

    P R O D U C T S GROUP, INC. Automation Products Group, Inc. Tel: 1/888/525-7300 1/435/753-7300 Fax: 1/435/753-7490