HCA003300-00020R ELECTRICAL INSTRUMENT CO., LTD. 7F-2, NO.351, SEC.2, CHUNG SHAN ROAD., CHUNG HO CITY, TAIPEI HSIEN, TAIWAN TEL:886-2-2226-6789(REP.) FAX:886-2-2226-7979 E-mail:[email protected] http://www.deree.com.tw Thanks for purchasing K-type Wireless Thermometer DE-33. Please read the operation manual carefully before using and keep it properly for contingent use and reference. Thanks! Model No. DE-33 Instruction Manual K-Type Wireless Thermometer

Upload: others

Post on 03-Feb-2022

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Instruction Manual K-Type Wireless Thermometer

HCA003300-00020R

ELECTRICAL INSTRUMENT CO., LTD.

7F-2, NO.351, SEC.2, CHUNG SHAN ROAD.,

CHUNG HO CITY, TAIPEI HSIEN, TAIWAN

TEL : 886-2-2226-6789(REP.) FAX : 886-2-2226-7979

E-mail: [email protected]

http://www.deree.com.tw

Thanks for purchasing K-type Wireless Thermometer DE-33. Please read the operation manual carefully before using and keep it properly for contingent use and reference. Thanks!

Model No. DE-33

Instruction Manual

K-Type Wireless Thermometer

Page 2: Instruction Manual K-Type Wireless Thermometer

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .01~02

. . . . . . . . . .03~07

. . . . . . . . . . . . .03

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .04

. . . . . . . .05~07

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .08~13

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .08~11

. . . . . . . . . . . . . . . . . . . . . .09~11

. . . . . . . . . . . . . . . . . . . . . . . . . . . . .12~13

. . . . . . . . . . . . . . . . . . . . . . . . .14~25

. . . . . . . . . . . . . . . . . .14~15

. . . . . . . . . . . . . . . . . . . . . .16

. . . . . . . . . . . . . . . . . . . . . . . . .17~19

. . . . . . . . . . . . . . . . . . . . . . . . . .20

. . . . . . . . . . . . . . . . .21

. . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

. . .23

. . . . . . . . . . . . . . . . . . . . . .24

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

. . . . . . . . . .26~28

Catalogue

Ⅰ. Features

Ⅱ. Introduction of main unit and sensor unit

(A). Main unit – appearance and functions

(B). Main unit – display

(C). Sensor unit – appearance and placing

Ⅲ. Power on process

(A). Battery installation

(B). Set up I.D. (identity code)

(C). Set up time (clock)

Ⅳ. Application of thermometer

(A). Function and operation of key

(B). Measurement of temperature

(C). Set up alarm function

(D). Read temperature record

(E). Search temperature of sensor unit

(F). Max / Min temperature

(G). Low battery symbol of main unit and sensor unit

(H). Trouble shooting of main unit

(I). Function table

Ⅴ. Specification of main unit and sensor unit

Page 3: Instruction Manual K-Type Wireless Thermometer

● Main unit with big LCD display and easy for read.

● Simultaneous display 2 sets of temperature measured through sensor unit.

● Temperature data detected by sensor unit transmits to main unit placed at remote position from sensor unit (maximum distance is around 100M in open space). The thermometer is suitable to apply indoor or outdoor use by user’s need.

● 24 hours clock.

● Auto data logging for 12 hours with 24 sets data. Temperature and clock time are recorded every 30 minutes on the hour.

● Main unit automatically records and refreshes MAX (maximum) and MIN (minimum) temperatures with real time detected by sensor unit.

K-Type WIREL ESS THERMOMETERDE-33R

MODEOC LGR CALL

SEARCH MAX /MIN

RESET PUSH 2sec

DE-33T

K-Type

WIRELESS THERMOMETER TRANSMITTER

-01- -02-

Ⅰ. Features ● Sensor unit is equipped with 2 thermocouples to detecting the temperature change of environment, water or soil at the same time.

● Temperature alarm function to alert the temperature is beyond the range set by the user.

● K-type temperature range: - 200°C ~ +1370°C

● Main unit and sensor unit each has 8 identity codes (1 ~ 8 and I.D. for short). The I.D. provides the protection from interference of transmission where have two sets or more thermometer using at the same time in the same radius area.

● Audible alarm of main unit alerts 4 possible status – 1.Battery power is low at main unit or sensor unit. 2.Temperature is over the range set by the user. 3.Transmission is disconnected.4.In measuring mode but main unit displays “no”.

CH1 temperature (CH1 = A plug of sensor unit)

CH2 temperature (CH2 = B plug of sensor unit)

K-type thermocouple

Clock

main unit

sensor unit

Page 4: Instruction Manual K-Type Wireless Thermometer

K-Type WIREL ESS THERMOMETERDE-33R

MODEOC LGR CALL

SEARCH MAX /MIN

RESET PUSH 2sec

① ② ③ ④

① OC

MODE

LGR CALL

SEARCH

-03-

MAX /MIN

RESET PUSH 2sec

CH1

⑩⑪

Ⅱ. Introduction of main unit and sensor unit (A) Main unit – appearance and functions

(B) Main unit – display

MODE key: (function change key)a. CH1 & CH2 in measuring mode.b. Alarm set for CH1 & CH2 (initial set is 180°C at manufactory)c. Call out temperature data from memory.d. Clock setting.◎ Refer to page 14 ◎

SEARCH key:a. Search the transmission signal from sensor unit and display

data at main unit. b. Push SEARH key to search the signal from sensor unit if no

signal or interruption. Each search is lasting 5 minutes.◎ Refer to page 21 ◎

MAX/MIN key: (maximum / minimum temperature recording) a. Simultaneously record maximum (MAX) & minimum (MIN)

temperature from CH1 & CH2. b. Each memory comes with real time. c. Shift reading MAX and MIN value at CH1 & CH2◎ Refer to page 22 ◎

RS-232 socket (equipped PC use RS-232 adaptor socket) – optional

a. Connect RS-232 cable to link with PC and recording the temperature change and form in graphic.

b. Refer to Help in PC software operation for relevant operation.

● Main unit – key function

hanger hole beeper hole

battery door

stand rack

Attention: do not attempt to loose those 4 screws.

● Description of symbol on display:Upper layer : display CHI temperature and functions from sensor unit

Middle layer: display CH2 temperature and functions from sensor unit

Bottom layer: display clock (24 hours)

Total 8 I.D. codes for set (from 1 to 8)

Antenna symbol in moving means main unit is searching the transmission signal from sensor unit

Battery power at sensor unit is low and need to replace with new batteries

Recording symbol to indicate 24 records of temperature change at CH1 in past 12 hours.

Shift reading MAX & MIN temperature value measured at CH1

Display CH1 temperature

Indicate data received from CH1 (A plug) and CH2 (B plug) of sensor unit

Blinking and beeping indicates temperature at CH1/CH2 is over (higher or lower) the range set by user.

Battery power at main unit is low and need to replace with new batteries

-04-

Temperature

Page 5: Instruction Manual K-Type Wireless Thermometer

1. Check battery power of main unit and sensor unit before using. Do not place sensor unit and main unit from each other over the limit distance to prevent the fail transmission. And keep away from the interference objects or signals.

2. The limit distance between main unit and sensor unit is typically 100M (330 feet) in open space.

3. The distance between main unit and sensor unit is shorter than 100M while there are barriers between the set up positions of units. The optimum distance depends on the material of barrier or the construction of building.

-05- -06-

DE-33T

K-Type

WIRELESS THERMOMETER TRANSMITTER

A B1 2 3 4

O N SGM53

1 2 3 4 5 6 7 8

1 2 3

(C) Sensor unit – appearance and placing

⑳ CH2

Recording symbol to indicate 24 records of temperature change at CH2 in past 12 hours.

Indicate temperature range set at HI or LO at CH1 & CH2

Display CH2 temperature

Shift reading MAX & MIN temperature value measured at CH2

Temperature

K-type thermocouple

4. Do not put sensor unit into fridge or the metallic sealed equipment (enclosure) directly for measurement. Because the signal of sensor unit is blocked and can't transmit to main unit successfully.

5. For measuring fridge temperature only put the thermocouple wire tip in fridge.

6. A & B input socket (A → as CH1 on main unit display and B → as CH2) can take 2 K-type thermocouples for measurement. Apply the different temperature range of K-type thermocouple to work with the need. The applicable range is -200°C ~ +1370°C.

7. Set up I.D. of sensor unit

※ 8. Install batteries -– always install main unit first and complete

I.D. set from panel keys and then install batteries into sensor unit and complete I.D. set from switch in battery slot. The I.D. code of main and sensor unit should be the same.

9. To ensure the data to be captured by the main unit the initial transmission from sensor unit acts aggressively by every 6 seconds once set up. After the 9th data, the transmission of sensor unit and the receiving of main unit is parallel corresponding to each other.

Attention: do not attempt to loose those 2 screws.

hanger

battery door

switch 4 is void

I.D. switch in battery slot

I.D. code (1 ~ 8)

switchin transmission the light is blinking

Page 6: Instruction Manual K-Type Wireless Thermometer

NNW

-07- -08-

DE-33T

K-Type

WIRELESS THERMOMETER TRANSMITTER

● application environment or barriers where main unit and sensor unit located. And the electronic interference will affect the transmission and signal and may display incorrect data on main unit. Therefore, please prevent those elements may affect the transmission.

● In extreme high or low temperature environment, the transmission may be slow or creeping.

● Besides battery door, please do not attempt to loose other screws in order to make sure the thermometer operates in good condition.

● Take the batteries out of the thermometer if it will be left idle for a long time.

Transmission and transmission distance is affected by

Attention:

Ⅲ. Power on process Demonstration of measurement

water temperature of aquarium

put thermocouple tip in fridge, water or soil for measurement

thermocouple tip

Simultaneously measuring and monitoring temperature of fridge and aquarium.

(A). Battery installation Always install main unit first and then sensor unit. Follow instruction of A & B as below to enable transmitting and receiving in good work.

1. Battery (alkaline) life time for main and sensor unit is one year round which mainly depends on the application in different temperature and environment. Use 2 AA size alkaline batteries each for main and sensor unit.

2. Battery life time depends on the working condition. Such as repeat search for signal (because of interruption) or frequent alarm will cut down the battery life time. Need to replace the new batteries while the low battery symbol (sensor unit) or (main unit) display.

3. Do not attempt to loose the screws marked on drawing. 4. Main unit has no power switch and can be power off from

removing batteries exclusively. 5. Contact the maker or deliver the thermometer back to the

maker for checking if it can't work after the batteries installation process is completed.

+

+

AA

(1.5

V)N

o.3

AA

(1.5

V)N

o.3

A. Battery installation of main unit

1. Open battery door and install batteries.

Attention: do not attempt to loose those 4 screws.

follow arrow direction to pull out the battery door

AA size battery (1.5V)install batteries with correct polarity marked in battery slot

sensor unit with thermocouple

Page 7: Instruction Manual K-Type Wireless Thermometer

-09- -10-

B. Battery installation of sensor unit

Follow the steps from dwg#1 to dwg# 7 to complete the installation.

(1) Loose 2 screws on battery door with Philips screw driver.

(2) Do not attempt to loose 4 crews marked on dwg#2.

(3) Choose I.D. by changing the switch in the battery slot before installing batteries. I.D. code from 1 to 8 (refer to page 6 (C)-7.). The I.D. should be the same as the main unit for transmitting and receiving recognition.

(4) Fasten screws on battery door after installing batteries. (1.5V AA size battery * 2) Be sure to put with the correct polarity as marked in the slot.

(5) After battery installation, the light on sensor unit is flashing to identify I.D. Then the light is flashing every 6 seconds and transmitting the data of each flashing. After the 9th data, the transmission of sensor unit and the receiving of main unit is corresponding to each other and go into regular cycle by transmission/receiving every 3 minutes. Each flash of light means transmission is working and main unit refresh display of CH1 & CH2 every 3 minutes as well.

(6) In 1 minute after battery installation, check main unit if display any reading of temperature from sensor unit. If not, please make sure if I.D. of main and sensor unit is the same. And check if batteries are well installation and light is flashing during transmission of data. If all those are positive but still can’t find any reading on main display, please contact the maker for problem remove or deliver the thermometer back to the maker for checking.

2. Once batteries are installed the full segment will display on LCD for 2 seconds.

CH1 temperature

CH2 temperature

clock

If display as above means no signal is received and clock is not set yet.

3. Push MODE key to shift the display from (1) to (3)

display (1) display (2) display (3)

If no reading on display of main unit after the batteries are installed completely which means sensor unit is not ready. (refer to page 10)

The bottom layer is clock (refer to page 12 for clock set). Check the battery installation is correct or not If can't shift the display from (1) to (3). If the installation is surely correct please contact the maker for problem remove or deliver the instrument back to the maker for checking if necessary.

(B). Set up I.D. (identity code) Choose I.D. of main unit:

Push MODE & MAX/MIN keys simultaneously for 3 sec. until

display and flashing. Push MAX/MIN key to choose

I.D. code from 1 to 8. Push MODE key to complete setting after chosen.

1

Page 8: Instruction Manual K-Type Wireless Thermometer

1 2 3 4

O N SGM53

1 2 3 4

O N SG M53

DE-33T

K-Type

WIRELESS THERMOMETER TRANSMITTER

-11- -12-

++

dwg#1

dwg#5

dwg#6 dwg#7

set I.D. (1~8)

Attention: Do not attempt to loose the screws marked.

light of transmission: one flashing to identify I.D. after batteries are installed and flashing every 6 seconds in transmission. After the 9th data the transmission become stable on regular basis with transmission / receiving every 3 minutes.

Main unit automatically searches signal from sensor unit (or push SEARCH key for manual search)

Main unit automatically searches signal from sensor unit in 5 minutes if no reading on display. Refer to page 5 Sensor unit appearance and placing

Set up time (refer to page 12)

(C). Set up time (clock) After power on must set up clock immediately enable to work with the relevant functions 1. Hold MODE key for 3 seconds to access time set function

as dwg#2.2. Push MAX/MIN key to adjust time and push SEARCH to

change position. Set hour and minute as dwg# 2 & 3 (clock starting from 00:00 with 24 hours).

3. Push MODE key to complete time set or if no pushing MODE key, it will automatically confirm the time set in 15 sec. as dwg#4.

4. Push MODE key to shift display and check if clock displays in good condition in different mode as dwg#5. If not, please operate from step 1 to 3 again.

dwg#1 dwg#2

dwg#3 dwg#4

dwg#5

no symbol display is in normal operation

push MODE key for 3 seconds until display as dwg#2 .

push SEARCH key to change position

push MAX/MIN key to adjust timepush MODE key to complete time set as dwg.#4

push MODE key to check clock if display in different mode.

push MODE key to shift display as dwg#5 to check clock display

hour minute

dwg#2

dwg#3 dwg#4

Page 9: Instruction Manual K-Type Wireless Thermometer

(A) Key function: MODE key shift 4 modes display 1. Simple measuring mode - CH1 and CH2 temperature

display (°C) (dwg#1)It's set as the initial display at manufactory. To operate any other function must go back to the simple measuring mode first.

2. Activate temperature alarm function ( ) (dwg#2)3. Read data in memory LGR symbol display (dwg#3) 4. Time setting (page 15 dwg#1 & 4)

-13- -14-

《ATTENTION》a. Time set can only operate in simple measuring mode as

dwg#1. In MAX/MIN, LGR or alarm mode can't operate time set as dwg.#6.

b. In case, accidentally push MAX/MIN key MAX, MIN symbol display on LCD and get into MAX/MIN mode. Hold MAX/MIN key for 2 sec. to cancel MAX/MIN mode as dwg#7.

c. Set sensor unit in position where is in the limit distance range from main unit. If main unit can't receive the data from sensor unit please check the distance if over the limit or having interference. Refer to page 5~6 II-(C) Sensor unit appearance and placing.

dwg#6 dwg#7

MAX, MIN symbol display can't operate time set.

Hold MAX/MIN key for 2 seconds to cancel MAX/MIN mode and MAX, MIN symbol disappear. (simple measuring mode).

if LGR symbol display

push MODE key one time to resume simple measuring mode

push MODE key twice to resume simple measuring mode

if symbol display

In MAX/MIN, LGR or alarm mode can't operate time set.

Ⅳ. Application of thermometer

(1) CHI & CH2 temperature display ( C)°

simple measuring mode display

dwg#1

dwg#2

dwg#3

push MODE key one time

push MODE key

push MODE key (resume to simple measuring mode display).

(2) temperature alarm mode

(3) read data in memory (LGR CALL)

push SEARCH key to read the data at different time.

Page 10: Instruction Manual K-Type Wireless Thermometer

(B) Measurement of temperaturea. Change display to the simple measuring mode as dwg#1

below. b. Upper layer display of CH1 is for the temperature

measured by the thermocouple at A socket. Middle layer display of CH2 is for the temperature measured by the thermocouple at B socket. Temperature unit is Celsius. Bottom layer display is indicating clock.

c. The transmission from sensor unit is every 3 minutes and refresh the data on main unit display.

d. I.D. set is from 1 to 8 by option. Hold MODE & MAX/MIN key simultaneously for 3 seconds to enter I.D. setting. I.D. number is flashing and ready for set. Push MAX/MIN key to choose I.D. and push MODE key to confirm the setting or no push MODE key in 15 seconds will confirm the setting automatically.

e. is antenna symbol (dwg#1). “ ” is signal disconnection (dwg#2) and need to push SEARCH key to search signal. Check sensor unit location where has interference or not. Refer to page 5~7 II-(C) Sensor unit appearance and placing.

f. If push MAX/MIN key accidentally, the setting will disable. Hold MAX/MIN key for 2 seconds to cancel MAX/MIN mode and go back to the simple measuring mode. (dwg#1)

《ATTENTION》a. Before going ahead any function or setting, please always go

back to the simple measuring mode first. (dwg.#1)b. If push MAX/MIN key, MAX/MIN symbol display and enter

MAX/MIN mode. Hold MAX/MIN key for 2 seconds to cancel the function and back to the simple measuring mode. (dwg#1)

-15-

(1) CH1 & CH2 temperature display ( C)°

(4) time set

simple measuring mode display

dwg#1

dwg#4

refer to page 12 III-(C) SET up time (clock).

Hold MODE key >3 seconds

Ho

ld M

OD

E k

ey

>3

se

co

nd

s

Pu

sh

MO

DE

ke

y o

nce

ag

ain

(1) CH1 & CH2 simple measuring mode ( C)°

Quick note of MODE key operation

push MODE key once

(2) Activate alarm function ( )

push MODE key again

(3)Reading data in memory (LGR CALL)

(4)Clock Settting

I.D. number

dwg#1

dwg#2

upper layer display for CH1 temperature no symbol display is in the simple measuring mode

I.D and antenna symbol

middle layer display for CH2 termperature

bottom layer display is clock

signal disconnectionPush SEARCH key to search signal. Check sensor unit location where has interference or not or too far from main unit. Refer to page 21 IV-(E) search temperature of sensor unit

-16-

Page 11: Instruction Manual K-Type Wireless Thermometer

e. Set HI or LO temperature:

● Only one option to set alarm range at HI or LO temperature.

● Set temperature only from dwg#1 to dwg#2. HI or LO symbol is flashing and ready for setting. Push MAX/MIN key to choose HI or LO (ex. set at LO)

● Push SEARCH key to change position and push MAX/MIN key to choose number. (ex. 25°C). (available temperature range for setting : -199.9°C ~ 1372°C)

● Push MODE key or no push MODE key in 15 seconds will confirm the setting automatically.

-17- -18-

(C) Set up alarm function ( ) :(1) Set up temperature alarm of CH1 and CH2: in condition

of the temperature reaches and higher beyong ( ≧HI ) the temperature set at HI or lower below ( ≦LO ) the temperature set at LO, the alarm function activates and emit beep sound to alert the user.

(2) Go to alarm temperature set and alarm: a. Hold MODE & SEARCH key simultaneously for 3

seconds to enter setting mode of CH1 as dwg#2: temperate alarm range display.

b. If no push any key or MODE key in 15 seconds, it will resume to the simple measuring mode automatically as dwg#1.

c. Hold MODE & SEARCH key simultaneously for 6 seconds to enter setting mode of CH2 as dwg#3: temperate alarm range display.

d. The initial temperature range is set 180°C at manufactory.

no symbol display is the simple measuring mode.

dwg#1

dwg#2 dwg#3

Hold MODE & SEARCH key simultaneously3 seconds. 6 seconds.

set HI or LO

push MODE key back to dwg#1 push MODE key back to dwg#1

dwg#1

dwg#2

no symbol display is in the simple measuring mode

set alarm at LO

set alarm temperature point at 25 C°

follow (2)-a. enter CH1 or CH2 setting mode, push SEARCH to change position and push MAX/MIN to choose LO or HI.

Page 12: Instruction Manual K-Type Wireless Thermometer

(D) Read temperature record (LGR CALL)(1) 12 hours auto data logging for CH1 and CH2

temperature corresponds with real time record. Recording is every 30 minutes with 24 sets maximum. Once capacity is full will start over the recording cycle from the beginning and the old record will be replaced. (Note: The record is made on every 30 minutes clock time. If initial clock time set at 13:52, the first data is at 14:00 and if the initial time is 12:17 the first data is 12:30 and the 2nd is at 13:00 and so on.).

(2) Push MODE key twice to switch from dwg#1 to dwg#2. The auto data logging function actuates once clock time is set.

(3) Push SEARCH key to read the previous data in memory. (dwg#2~4).

(4) Push MODE key back to the simple measuring mode.

f. Activate alarm function ( display):

● Push MODE key to switch dwg#1 to dwg#2

● Alarm function is available simultaneously at CH1 and CH2. Once activate alarm function (as dwg#2), if temperature at CH1 or CH2 is over the temperature alarm point (higher or lower) the audible beeper will be on and lasting 2 minutes. Beeper will stop in 2 minutes automatically or push SEARCH key by manual stop. Push MODE key twice to return the simple measuring mode as dwg#1.

● Once temperature at CH1 or CH2 meets the alarm condition will activate audible beeper and “ ” and “HI” or “LO” symbols are flashing. Even stop beeping (automatically or manually) , the symbols are remained flashing to alert the user the temperature ever exceeds the alarm point. If the next coming data is still over the alarm point the beeper will generate again to repeat beep for 2 minutes and symbols are flashing.

-19- -20-

dwg#1

CH1 temperature

Ch2 temperature

push MODE key to dwg#2

dwg#2

flashing in beeping

dwg#1 dwg#2

dwg#3 dwg#4

push MODE key twice to dwg#2 push SEARCH key to dwg#3

push SEARCH key to dwg#4 push MODE key to back to dwg#1

(5) In LGR mode, push MODE key to go back to the simple measuring mode. Or it will shift back to the simple measuring mode automatically if no operation in 5 minutes.

(6) If display “no” at CH1 or CH2 means the poor contact of the thermocouple, disconnection of thermocouple or temperature is over the maximum range 1372°C.

Page 13: Instruction Manual K-Type Wireless Thermometer

(E) Search temperature of sensor unit(1) Push SEARCH key for manual search of signal from

sensor unit. Every search is lasting 5 minutes. If 5 minutest search is failed the radius symbol ( ) disappears which means the transmission or receipt interrupts. Push SEARCH key once again to activate manual search function until the signal can be found.

(2) Once power on main unit and sensor unit, if no signal ( ) in transmission the auto search function will generate in 30 minutes and regenerate every 30 minutes if still can't find signal. Manual search is also available for option.

(3) Antenna radius symbol ( ) in flashing means main unit receives signal from sensor unit in good condition. If the radius symbol stops flashing means in searching for signal (search for 5 minutes).

(4) In normal operation if transmission is temporarily interrupted or change location of main unit or sensor unit to cause the interruption of signal, CH1 and CH2 temperature display becomes “---“. Need to check if have any interference while change the location of units and remove the problem.

-21- -22-

(F) MAX / MIN temperature(1) Maximum / Minimum temperature recording function

(set clock before entering MAX/MIN mode). Set MAX/MIN mode starting from dwg#1.

(2) Push MAX/MIN key, MAX/MIN symbol display and start recording maximum / minimum temperature.

(3) Push MAX/MIN key to shift reading MAX/MIN temperature at CH1 & CH2 respectively. (dwg# 3→6).

(4) After reading 4 values will back to MAX/MIN recording mode (dwg#2). During reading MAX/MIN temperature, if no push MAX/MIN key for 15 seconds will go back to MAX/MIN recording mode automatically.

(5) Hold MAX/MIN key for 2 seconds to exit MAX/MIN mode and back to the simple measuring mode (dwg#1).

《ATTENTION》In MAX/MIN mode, other functions disable. To enable other modes or functions must exit MAX/MIN mode by holding MAX/MIN key for 2 seconds if in MAX/MIN mode.

dwg#1

dwg#2

(A) While1st disconnection once power on will have auto searching in 30 minutes.

(B) flashing ( ) is in good receiving of signal

(C) stop flashing is in search of signal from sensor unit (5 minutes).

no radius symbol moving is interruption

no temperature display

antenna symbol stop moving is in searching signal.

push SEARCH key to search signal from sensor unit and refer to page 6 II-(C) sensor unit-appearance and placing for problem remove.

dwg#1 dwg#2

dwg#3 dwg#4

upper layer display - CHI temperature middle layer display - CH2 temperaturbottom layer display - clock push MAX/MIN key

push MAX/MIN key back to dwg#2

push MAX/MIN key

CH1 maximum temperature CH1 minimum temperature

Page 14: Instruction Manual K-Type Wireless Thermometer

(H) Trouble shooting of main unit(1) Either in MAX/MIN mode, alarm mode or reading

temperature record (LGR) mode, can't operate temperate range alarm setting or clock set. All functions only enable at the simple measuring mode. (as below drawing status 1 ~ 3)

(2) Before operating other functions (ex. temperate alarm setting, clock, LGR) or go to MAX/MIN recording mode must go back to the simple measuring mode first.

(G) Low battery symbol of main unit and sensor unit(1) (sensor unit) or (main unit) symbol display

means battery power of sensor unit or main unit is low and constant audible beep sounds to alert the user to replace new batteries. To ensure operation in good condition, replace 2 batteries of unit at the same time.

(2) Apply 2 AAA batteries (1.5V).

-23- -24-

dwg#5 dwg#6

CH2 maximum temperature CH2 minimum temperature

push MAX/MIN key push MAX/MIN key

push MAX/MIN key back to dwg#2

low battery indication of sensor unit

low battery indication of main unit

status 1 status 2 status 3

status 4 status 5 status 6

simple measuring mode

hold MAX/MIN key to cancel MAX/MIN mode and MAX/MIN symbol disappear

in alarm mode, push MODE key twice to back simple measuring mode.

in LGR mode, push MODE key to back simple measuring mode.

ready for setting all kinds of function.

in simple measuring mode, push MAX/MIN key to enter MAX/MIN recording mode.

push MODE key to back simple measuring mode

push MODE key to back simple measuring mode

push MODE key to back simple measuring mode

Page 15: Instruction Manual K-Type Wireless Thermometer

-25- -26-

(I) Function table

Function Key Operation

Dwg#1CH1 & CH2 simple measuring mode

Dwg#2 alarm function activate

Dwg#3 LGR mode

Dwg#4 MAX/MIN temperature at CH1 & CH2

Dwg#5 Temperature alarm set at CH1

Dwg#6 Set clock

Initial setting of simple measuring mode to check CH1 & CH2 current temperature. To set dwg#2~6 must go back to dwg#1. Main unit display temperatures measured at two channels of sensor unit. Main unit is not equipped temperature detecting function.

At dwg#1, push MODE key twice to activate alarm function. Push SEARCH key to stop beeping. Push MODE key twice to back simple measuring mode (dwg#1)

At dwg#1, push MODE key to enter reading data mode (LGR). Push SEARCH to shift reading data in memory . After reading, push MODE key to back simple measuring mode (dwg#1)

At dwg#1, push MAX/MIN key to enter MAX/MIN recording mode. Push MAX/MIN key to shift reading maximum & minimum temperature at CH1 & CH2. Hold MAX/MIN key for 2 seconds to exit MAX/MIN mode and enable operating other functions.

At dwg#1, hold MODE & SEARCH keys simultaneously for 3 seconds (6 seconds for CH2) to enter alarm setting at CH1. After setting the temperature point, in 15 seconds will go back to simple measuring mode (dwg#1) automatically. (or push MODE to complete setting).

At dwg#1, hold MODE key for 3 seconds to enter clock set mode. In 15 seconds will go back to simple measuring mode (dwg#1) automatically. (or push MODE to complete setting).

dwg#1 dwg#2 dwg#3

dwg#4 dwg#5 dwg#6

Ⅴ. Specification of main unit and sensor unit

(1) LCD display: clock, CH1 & CH2 temperature (2) Transmission / Receiving interval time: every 3 minutes.(3) Clock: 24 hours(4) Features: MAX/MIN simultaneous recording with real time.

Auto data logging, Beeper function for over temperature range.

(5) RF frequency: approx. 433MHz (depends on frequency set at manufactory).

(6) Temperature range: -200°C~1370°C(7) Power: AA battery (1.5V) x 2 pcs(8) Dimension / Weight: 120(L) x 100(W) x 28.5(H)mm,

approx. 160g (without batteries)

K-Type WIR ELE SS THE RMOMETE RDE-33R

MODEOC LGR CALL

SEARCH MAX /MIN

RESET PUSH 2sec

DE-33T

K-Type

WIRELESS THERMOMETER TRANSMITTER

main unit

sensor unitmain unit

sensor unit

(1) Featurs: 2 channels, transmit data to main unit.(2) RF frequency: approx. 433MHz (depends on frequency set at

manufactory).(3) Transmitting distance: approx. 100M (330 feet) in open

space.(4) Power: AA battery (1.5V) x 2 pcs(5) Dimension / Weight: 179(L) x 72(W) x 32(H)mm,

approx. 155g (without batteries)

Page 16: Instruction Manual K-Type Wireless Thermometer

Instruction manual .......................................................................1AA batteries ................................................................................ 4

K-type thermocouple (NR-39B) ................................................... 2Optional accessories (equipped with RS-232 plug at main unit)RS-232 cable .............................................................................. 1PC software CD .......................................................................... 1

※ PC software operation-refer to operation on PC software CD.

(6) Accuracy (23 )

(7) K-type thermocouple: NR-39B temperature probe is applicable to -50°C~200°C (-58°F~392°F). For higher temperature measurement need to contact manufacturer for other option.

(8) Range: -199.9°C ~ 1370°C(9) Input limit: max. 60VDC or 24Vrms AC(10) Socket: applicable to standard thermocouple. (11) Operation temp./humidity: 0°C~50°C (32°F~122°F) /

below 80% R.H. (no condensation)(12) Storage temp./humidity: -10°C~60°C (-14°F~140°F) /

below 70% R.H. (no condensation)(13) Pre-caution:

To avoid danger and damage happened during operation, please read the instruction manual thoroughly before using the thermometer. Put the instruction manual with the thermometer together for reference in the future.

°C ± 5°C, 80% R.H. MAX.

-27- -28-

Type Measurement Ranges(°C)

Resolution(°C)

Accuracy(%rdg+°C)

-199.9~0.0

0~199.9

200~1370

0.1

0.1

1

±(0.5+1)

±(0.2+0.5)

±(0.2+1)

K

WARNING

TO AVOID ELECTRICAL SHOCK, DO NOT APPLY INPUT VOLTAGE OVER DC60V OR AC24V

WARNING

TO AVOID DAMAGE OR BURNS, DO NOT PUT THE THERMOMETER IN MICROWAVE OVENS OR STOVE F O R M E A S U R E M E N T O R A P P L Y F O R T H E MEASUREMENT OF ELECTRICAL USE.

Attention

The repeated sharp flexing can break the thermocouple leads. To prolong lead lifetime, avoid sharp bend in the leads, especially for the cable relief to connector.

ACCESSORIES

Maintance

Do not attempt to open case of sensor unit or main unit (except battery door) to avoid the inaccuracy of measurement. Keep dry hands while replacing batteries. Do not use organic solvent to clean the thermometer but with soft cloth if necessary.

Specifications and external appearance of the product may be revised for modification without prior notice

Please contact our agents or distributors for a variety of measuring instrument we produce under the strict quality control requirement of ISO 9001