instruction manual hmp pdgpro en 07 2013
DESCRIPTION
Instruction Manual HMP PDGpro en 07 2013TRANSCRIPT
-
INSTRUCTION MANUAL
THE STATIC PLATE LOAD TESTER
www.hmp-online.com > information, video, animation
Calibration-
I N S T I T U T E
authorised by the
Federal Highway
Research Institute
click
HMP PDGpro
-
1
CONTENTS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
General Information
Safety instructions ...................................................................... 2Information for Users ........................................................... 2Symbols Used ..................................................................... 2Legal terms of reference ...................................................... 2
Intended Use .............................................................................. 3
Plate Load Tester Basics ......................................................................................... 4
Specification ........................................................................ 4General structure / Components ......................................... 5
Measuring Instrument HMP PDGpro ....................................... 7Operation ............................................................................. 7Power Supply ....................................................................... 8Charging of Rechargeable Batteries ................................... 8
Measuring
Prepare Measurements .............................................................. 9Install Loading mechanism .................................................. 9Installation of Measuring Frame (refer to Annex 1) ............. 9Connect Devices .................................................................. 9
Measuring Procedure ............................................................... 10Start Measurement ............................................................ 10Error Menus ....................................................................... 12Quit Measurement ............................................................. 12
Evaluation
Reading / Printing Measured Data ........................................... 13Transfer the Stored Measured Data ........................................ 13Delete Measuring Results ........................................................ 14
Options AP 1300 Miniprinter .................................................................. 15GPS Receiver ........................................................................... 18
Settings Settings Measuring Instrument ................................................ 20
Display ............................................................................... 20Device ................................................................................ 20Printer ................................................................................. 20Contact data ...................................................................... 20
Maintenance
Calibration ................................................................................ 21Hotline ...................................................................................... 21
Annex 1 Plate Load Tester spare parts .................................................. 22
Annex 2 Operating Instructions Hydraulic System ................................ 23
Parts of Hydraulic System ................................................. 23Using Hydraulic System .................................................... 23Maintenance ...................................................................... 23
Annex 3
EC Declaration of Conformity
-
2
GENERAL INFORMATION
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Safety instructions
Information for Users This instruction manual was prepared such that users can easily become familiar with the HMP PDG Plate Load Tester, and make use of the tester for intended applications. Users should carefully read this instruction manual and the safety instructions prior to using the Plate Load Tester. Users also have to read carefully through the operating instructions by company LUKAS for hand pump, hydraulic cylinder and quick coupler before using the Plate Load Tester. Follow the instructions of this instruction manual and the operating instructions of company LUKAS without exception.
Symbols Used Warnings and instructions are highlighted as described below:
Warning This symbol is used in conjunction with related text to draw user's attention to hazards and risks which may cause failure of tester components or adversely affect operating procedures, in case users do not take the corresponding precautions.
Note This symbol and the related text identify technical requirements and provide additional information to be taken into account by the operator to carry out the following operations effectively and safely.
Legal terms of reference The Plate Load Tester complies with the current state of the art and all applicable safety regulations. Construction and function of the Plate Load Tester meet the requirements laid down in DIN 18134, issue 2012 Soil Testing procedures and testing equipment Plate load test. The Plate Load Tester meets the basic safety requirements laid down in the EU Directives for Harmonisation referenced in the EC Declaration of Conformity.
-
3
GENERAL INFORMATION
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Intended Use The Plate Load Tester is exclusively intended for determining the soil bearing capacity and the compaction quality of the soil referred to DIN 18134, issue 2012 Soil Testing procedures and testing equipment Plate load test. Its intended use also includes: Compliance with the safety instructions and safety regulations
contained in this instruction manual and in the provided operating instructions from company LUKAS for the hydraulic components.
Compliance with the maintenance and servicing instructions contained in this instruction manual and in the provided operating instructions from company LUKAS for the hydraulic components.
Any other use or any use beyond this definition is not intended and may cause injury to people and damage to property. The manufacturer/supplier shall not be held liable for damages resulting from other than the intended use. The risk shall be borne solely by the user.
-
4
PLATE LOAD TESTER
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Basics
Specification Loading mechanism Hydraulic pump with 100 kN cylinder and 150 mm lift with high-pressure hose of 2 m in length. 1 set of hydraulic cylinder extension pieces (plug-type connection) 1 pressure plate with magnetic holder and upper ball-and-socket
joint Load plate with adjustable bubble level Diameter 300 mm Thickness of plate 25 mm Settlement measuring mechanism (measuring frame) Support frame on three-point bearing, equipped with insertable and turnable contact element (DIN 18134, Fig. 3a) and adjustable feet Dimensions: Length 2.320 mm Width 570 mm Height 420 mm Weight 13.2 kg Measuring of load Electrical load cell 50 kN, complete with thrust piece and adapter Measuring of settlement Electronic dial gauge
25 mm measuring range, 0.01 mm resolution, local display unit, IP 42
Automatic evaluation unit Measuring head
logging load and settlement data, interfering-immune digital transmission to measuring instrument
Measuring instrument data logger for load and settlement data with automatic processing and evaluation facilities as well as storage capability for 200 tests Comprising: - Measuring instrument for data logging, storage and evaluation
with comfortable user interface - LCD graphic display screen 65x65 mm - Accumulator for 12 hours of autonomous operation,
rechargeable from mains or vehicle adapter Miniprinter to print data and pressure settling lines on the site
(optional) Aluminium case for measuring instruments
Dimensions: 460 x 340 x 210 mm Weight: 8 kg
-
5
PLATE LOAD TESTER
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
General structure / Components Picture 1 (refer to Annex 1) Loading mechanism
1 Ball-and-socket joint with magnetic holder
2 Extension pieces
3 Hydraulic cylinder
4 Thrust piece
5 Load cell
6 Measuring tunnel
7 Load plate with bubble level
8 Hydraulic pump
Picture 2 Settlement measuring mechanism (measuring frame)
1 Extension arm with measuring finger
2 Traverse
3 Measuring frame
4 Measuring head
5 Dial gauge
6 Bubble level
Picture 3 Electronic dial gauge
1 Dial gauge
2 Dial gauge cable
Picture 1
Picture 3
Picture 2
-
6
PLATE LOAD TESTER
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Picture 4 Measuring case
1 Measuring instrument
2 Electrical load cell
3 Electronic dial gauge
4 GPS receiver (optional)
5 Measuring head
6 Printer (optional)
Picture 5 Measuring instrument
1 USB port
2 GPS port
3 Printer port
4 LCD graphic display
5 Measuring cable outlet
6 Function keys
Picture 5
Enter
Magdeburger Prfgertebau GmbHService: +49(0)391 2514666
www.hmp-online.com
HMP PDG
5
6
3 4
1
2
Picture 4
3
2
4
6
1 5
-
7
PLATE LOAD TESTER
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Measuring Instrument HMP PDGpro
Operation The measuring instrument HMP PDGpro can be operated easily and intuitively by means of the function keys. Key functions Switch on / off measuring instrument
Select forward / upward
Select backward / downward
Confirm selection / Start action Overview menu functions Measuring Taking measurement Measured data Read/Print USB Delete Back
Settings Display Date Time Language Contrast Illumination Back
Device Meas. range GPS Unit Graphics Calibr. date Back
Printer Print head Date/time Graphics Contact data Back
Back By confirming the menu item Back you always come back to the previous menu.
Enter
-
8
PLATE LOAD TESTER
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Power Supply The measuring instrument HMP PDGpro is provided by 4 NiMH rechargeable batteries, which are located within the battery compartment on the backside. These rechargeable batteries can be charged by means of the charger and the adapter cable of measurement instrument ( Charging of Rechargeable Batteries). In emergencies measuring instrument can also be provided by 4 AA batteries.
Charging of Rechargeable Batteries
In case that nonrechargeable batteries are used in the device, the charger must not be connected, because otherwise the device could be damaged.
In case that the rechargeable batteries voltage of the measuring instrument falls below the required minimum voltage, an information appears on the display and the rechargeable batteries have to be charged. The rechargeable batteries should be charged only by means of the supplied power pack charger and adapter cable of measuring instrument. The power pack charger can be connected to a 100 240 V / 50 60 Hz mains supply or, by means of an adapter cable, to a 12 24 V car battery. The adapter cables are included in the scope of supply, and are housed in the carrying case. The power pack charger is housed in the carrying case beneath the measuring instrument. Picture 6
1 Power pack charger
2 Cable of measuring instrument
3 Adapter cable of measuring instrument
4 Adapter cable for car battery
Connect the adapter cable of measuring instrument to the cable
of measuring instrument and to the cable of power pack charger. Connect the power pack charger to the mains supply. After completion of the charging procedure disconnect the
charger from the electricity grid. The supplied power pack charger is designed for two modes of operation indicated by LEDs of different colour, refer to passage Miniprinter AP 1300 / Charging of power pack on page 16.
Picture 6
-
9
MEASURING
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Prepare Measurements
Install Loading mechanism Prepare the test site in compliance with the requirements of
DIN 18 134, issue 2012. Place the load plate on the test site. The opening of the measuring tunnel points into the direction
from which the measuring finger of the measuring frame is inserted.
Mount the load cell onto the pin of the load plate and the thrust piece onto the load cell.
Attach the hydraulic cylinder to the pin of the thrust piece. Mount the ball-and-socket joint with magnetic holder to the
loading vehicle. Use the extension pieces to adjust height differences.
Installation of Measuring Frame (refer to Annex 1) Mount the traverse (2) to the longitudinal beam of the measuring
frame (3) and align with height adjustment (star grips). Remove the extension arm (1) from the bracket, insert it into the
measuring frame and fix it by means of the 2 star grips. Insert the measuring finger into the measuring tunnel until the
relevant mark is reached, and align horizontally by using the circular bubble.
Place the dial gauge in the dial gauge bracket of the measuring frame and make settings it as appropriate (set to zero).
Connect Devices Position measuring head Mount the measuring head to the traverse (magnetic fixture). Connect the load cell Connect the load cell cable to the measuring head outlet labeled
Force. Connect the dial gauge (used for distance measuring) Connect the dial gauge cable to the dial gauge. Connect the dial gauge cable to the measuring head outlet
labeled Distance. Connect measuring instrument Connect the cable of measuring instrument to the measuring
head.
-
10
MEASURING
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Measuring Procedure For theoretical foundations of the measuring process please refer to DIN 18134, issue 2012.
Start Measurement Press the key to switch on the measuring instrument. On the LCD display appears the main menu (on the left) incl.
device type and serial number (xxxx), the individual menu items and the stress state of the measuring device batteries. Optionally appears a hint regarding the GPS device and/or charge state of the printer rechargeable batteries.
Before starting the measurement the desired measuring range has to be chosen within the measuring instrument ( page 20).
In case that the rechargeable batteries voltage of the measuring instrument falls below the required minimum voltage, the hint meas. dev. charge appears ( page 8).
If the note printer charge occurs on the display, the printer has to be charged before using to avoid a deep discharge of the power pack (Charging of power pack page 15).
Press the key to acknowledge the Measuring mode at the
basic menu. A note for logging GPS Data and/or the last calibration date
optionally appear on the LCD display (dependent on the settings of the measuring instrument).
.Acknowledge optionally the note for logging GPS data. It is not required to acknowledge the last calibration date. The menu shown on the left is displayed on the screen:
Number of the current measuring series and the current measuring range (xxx xxx) Number of the current measuring point Compression #.#### MN/m Settlement #.## mm These values result from weight of testing equipment.
Measuring is not possible in case of full memory. A corresponding note appears.
To continue measuring stored series may be removed (e.g. by transferring to the USB-Stick or the PC or printing) and subsequently deleted to provide the device with new storage capacity. It is recommended to transfer the test series and measurement results stored in the measuring instrument regularly onto the provided USB-Stick or to the PC ( page 13/13) and to delete afterwards the data in the measuring instrument ( Seite 14). In this way the transference time will be shortened and multiple data transfer avoided.
Enter
meas. point: 01
measurement XXX XXX
reset data
MN/m
mm
0.0049 0.52
meas. dev.
measuringmeasured data
settings
HMP PDGpro XXXX
printer
GPS ready
-
11
MEASURING
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Value 1
Acknowledge reset data by pressing the key to tare the displayed values. The values 0.0000 MN/m for compression and 0.00 mm for
settlement appear on the display. By that offset values, if any, are cleared.
Acknowledge reset gauge by pressing the key after
approaching the required preload, according to DIN 18134, issue 2012, to set the displayed value for the settlement to zero. The start value for the pressing and the zero value for the
settlement appear on the LCD display. These values are stored as first measured values. The local display on the dial gauge remains unaffected by this zero setting.
Please refer to DIN 18 134, issue 2012 for the required number of measurements and set values to perform. Set values according to DIN 18 134, issue 2012 are stored on the measuring device as orientation for the examiner, for load plates with diameter of 300 mm. If the measuring range 300 is selected, the specific set values are displayed on the screen.
Perform Measurement Approach the set point with the hydraulic pump.
(Operating Instructions Hydraulic System refer to Annex 2, page 23.) The measured value for compression and settlement appears
on the display. The number of the current measuring series and the current measuring range, the number of the current measuring point and also the nominal value for the pressing (setpoint) are displayed.
Accept the value after the required waiting time by pressing the key.
The Values by pressing key are frozen in the display. By means of the selection keys it can be switched
between the mode save, the mode delete and the mode quit?.
Press the key to acknowledge Store or Delete of the current (frozen) measuring value. The current measured values are displayed on the screen.
Follow the procedure described above for any further set values.
Enter
Enter
Enter
Enter
Enter
meas. point: 01
measurement XXX XXX
reset gauge
MN/m
mm
0.0100 0.00
meas. point: 02setpoint: 0,08
Wert halten
MN/m
mm
0.0831 1.16
measurement XXX XXX
-
12
MEASURING
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Error Menus The measuring instrument provides instructions to monitor the measuring procedure which pop up as an error report before measurement or during the measurement. The following error reports might appear before the measurement:
Error report Error cause
Force error no connection between load cell and measuring head
Gauge error no connection between dial gauge and measuring head or dial gauge is not switched on
no Communication
no connection between measuring instrument and measuring head
Check/establish the connections. Restart the measuring process by pressing the key, as soon
as the connections are correct. The following error reports might appear during the measurement:
Error report Error cause
Compression too small
The value for compression falls below the required minimum value for compression according to DIN 18134, issue 2012.
Press the key. Continue measurement. Quit Measurement All required measurements completed, the measuring procedure can be terminated. Quit measuring by pressing the key after storing of last
measuring value. The menu shown on the left is displayed on the screen.
In order to return to the last measuring value and to continue the measurement choose no and confirm.
In order to quit the measurement choose yes and confirm. Displaying results After acknowledgement of quit measuring, measuring series will be finished. The time of the start and the end of the measurement, the
Values of EV1, EV2 and the ratio EV2/EV1 are displayed on the screen.
After completing one measurement series it is possible to print out the current measurement row (only devices with printer) and also to display the curve data (pressure settling lines) and the measured data.
Enter
Enter
Enter
quit measurement?
no
measurement XXX XXX
measurement xxx23.04.2012 08:4723.04.2012 08:59
Ev1= 94.98MN/mEV2=241.99MN/mEV1/EV2= 2.55
print
measurement 1
print
-
13
EVALUATION
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Reading / Printing Measured Data The in the measuring instrument stored test series and -results can be displayed via menu item Read/Print on the LCD-screen (all types of measuring instrument) and they can be printed out (only devices with printer).
Prior to using the AP1300 miniprinter read the instruction manual furnished with the printer and follow the instructions given for putting into operation and for handling.
Choose and confirm the desired test series. The values of the chosen test series are displayed on the
LCD-screen. By pressing the keys, the following functions can be used (depending on the device settings):
Print: Printing the current test series Curve data: Displaying the settlement curves Measurements: Displaying the values for the individual
measuring points Evaluation: Displaying the test series number, the time of
the start and the end of the measurement, the values for EV1 and EV2 and also the ratio EV2/EV1
Back: Return to the previous menu
Transfer the Stored Measured Data The in the measuring instrument stored test series and results can be transferred via menu item USB to the supplied USB stick or to PC.
Data transfer measuring instrument USB stick Connect the USB stick to the measuring instrument. Choose transfer mode Device USB stick and confirm by
pressing . On the LCD display appears the inquiry USB stick
connected? Confirm this inquiry by pressing the key. The data are being copied to the USB stick. After completion of data transfer the measuring instrument
switches off automatically. To transfer the data from the USB stick to the PC see Protocol software for the Plate Load Tester HMP PDGpro.
Enter
Enter
meas. dev.
read/printUSBdeleteback
HMP PDGpro XXXX
printer
No. date time Ev2
2 23.04./09:40 241 3 ./10:24 234 4 ./11:34 228 5 25.04./08:52 225 6 ./12:54 236 7 ./14:07 235 8 ./14:45 240 9 26.04./13:21 237 10 26.04./14:27 232page forwardpage backback
23.0423.04
25.0425.0425.04
1 08.04./09:29 233
meas. dev.
read/print
deleteback
USB
HMP PDGpro XXXX
printer
meas. dev.
device->USBstickdevice->PCback
HMP PDGpro XXXX
printer
-
14
EVALUATION
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Data transfer measuring instrument PC Connect measuring instrument and PC via the supplied USB
cable. Choose transfer mode Device PC and confirm by
pressing . On the LCD display appears the inquiry PC connected?
Confirm the inquiry by pressing the key. The measuring instrument works now just like removable
media. After completion of data transfer to PC remove the removable
media and switch off the measuring instrument. To transfer data from measuring instrument to PC see Protocol software for the Plate Load Tester HMP PDGpro.
Delete Measuring Results The test series and results stored in measuring instrument can be deleted via menu item Delete. Confirm menu item Delete by pressing key. On the LCD screen is displayed the choice
Delete all measurements resp. not delete. Confirm Delete all measurements by pressing the key. All measurements will be deleted.
Stored series cannot be deleted individually.
Enter
Enter
Enter
Enter
maes. dev.
device->USBStick
backdevice->PC
HMP PDGpro XXXX
printer
read/printUSB
back
delete
HMP PDGpro XXXX
meas. dev.printer
-
15
OPTIONS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
AP 1300 Miniprinter Included in the HMP PDGpro scope of supply is an AP 1300 miniprinter (optional).
Power supply The printer can be operated independent of a power supply unit, and is powered by a 1.8 Ah NiMH power pack housed in the printer. Thus, the printer can be carried from job to job.
Safety
The NiMH power pack is provided with an internal fuse unit. However, a short-circuit may occur when the NiMH power pack gets into contact with metallic items.
The power pack must not be opened; otherwise it may leak out or a short-circuit may occur.
Before you remove or replace the power pack, disconnect it from the external power pack charger.
The power pack has to be charged only by means of the supplied power pack charger. The power pack charger can be connected to a 100 240 V / 50 60 Hz mains supply or, by means of an adapter cable, to a 12 24 V car battery. The adapter cable is included in the scope of supply, and is housed in the carrying case. The AP1300 printer is shipped with a connected and fully loaded power pack.
When the printer is used for the first time after a lengthy period or has been standing idle for a lengthy period, recharge the power pack prior to use.
In the event of malfunction the printer may only be opened by authorised personnel.
Charging of power pack
For charging the power pack it is only allowed to use the supplied power pack charger.
Use the power pack charger only indoors. Disconnect the device from the mains if it is not used. Do not operate the device in case of damage to the housing or the mains plug.
Only charge nickel/metal hydride power packs; its use for charging other batteries may involve an explosion hazard.
Do not open the power pack charger.
-
16
OPTIONS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Connect the power pack charger to the Power Supply connection of the printer.
Connect the power pack charger to the mains supply. Disconnect the power pack charger from the mains supply when
charging of power pack has been finished. The supplied power pack charger is designed for two modes of operation indicated by LEDs of different colour. Red LED (flashing):
Indicates the proper contact of the power pack during the test phase (about 10 s);
Indicates that the power pack has not been properly connected to the correct pole;
Indicates a defect power pack or an incorrect number of cells; Red LED (steady):
Indicates the charging process of the power pack. Green LED (steady):
Power pack is charged.
When the power pack has to charge again, the power pack charger has to disconnect from the mains supply at first. New charging can start when the two LED (red and green) have been go out (after 10-15 seconds).
Front panel of printer 1 Paper feed Single-line paper feed:
Press the key for a short interval, and release.
Multi-line paper feed: Hold down this key until the desired length
of paper is reached. 2 LED Signals READY
LED off: The printer is in the power-saving mode. Power pack is discharged. Charging of power pack has been finished. Green LED (steady): Printer is active. Green flashing LED: Paper out. Green orange flashing LED: Power pack is charged (trickle charging). Orange flashing LED: Power pack is charged (fast charging). Red green flashing LED: Power pack voltage is too low.
3 Paper compartment opener
The miniprinter AP 1300 switches on and off automatically. It isnt necessary to switch it on or off manual.
-
17
OPTIONS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Insert paper roll Push the paper compartment opener to the front until the printer
lid opens (1). Unwind a few centimetres of the new roll and load the paper roll
into the compartment such that the paper will unwind from below (2).
Close the printer lid (3). Press the paper feed key to check the correct paper movement. Rapidly tear off excessive paper using the cutting edge. The AP1300 miniprinter is provided with sensors to detect lacking paper or opened paper compartment. A sensor activated, the printer switches to the storage mode; all data transmitted to the printer are preserved. Printing is continued immediately as soon as the defect has been removed.
It is recommendable to use original paper role for thermo printer only.
Malfunction Printer fails to start printing: Connection correct? Check connections. Has the printer automatically switched on and is the LED on? Is the power pack discharged? Charge the printer before use.
-
18
OPTIONS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
GPS Receiver The Plate Load Tester HMP PDGpro is optionally equipped with an external highly sensitive GPS receiver to determine the measuring position. The coordinates are captured fast and exactly, because of newest technology, are transferred to the measuring instrument HMP PDGpro, are associated to the corresponding test and saved in the device.
Basis for the determination of the measuring position with the GPS receiver is a measuring location, where this device has free reception.
Operation GPS Receiver 1 Power on/off 2 LED 3 Port measuring instrument The operation state is indicated by LED (2):
LED off: - GPS receiver is switched off. - Batteries empty.
Green LED steady: - GPS receiver is switched on. - Position not yet determined or no reception.
Green LED flashing: - GPS active (position is determined).
Green LED unsteady: - Batteries too low (not possible to determine the position).
In case the batteries of the GPS receiver are too low, the request GPS Batt. change is displayed.
Power supply / Changing Batteries The GPS receiver is powered by 2 pieces of LR03 batteries. The scope of delivery includes 2 spare batteries and a screwdriver suitable for battery changing. In case that the batteries of the GPS receiver are empty or too low, the batteries have to be replaced as follows: Loosen the screw of the battery box cover on the back side of
the device. Remove the battery box cover. Replace old batteries by new ones (pay attention to the polarity). Insert the battery box cover. Tighten the screw of the battery box cover.
HMP GPS
Service: +49(0)391 2514666
www.hm p-online.com
on
off
on
offactive
1
3
2
meas. dev.
measuringmeasured data
settings
HMP PDGpro XXXX
printer
GPS Batt. change
-
19
OPTIONS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Use GPS Receiver Should the GPS receiver be used, is to be proceeded as follows: Switch on the GPS receiver. Initialization and determination of the position
(lasts at least 1 minute). Wait, until the position is determined and the LED flashes green. Press the key to switch on the measuring instrument. At the bottom of the display the GPS status appears.
If the GPS status is not displayed, the GPS function can be activated in the menu settings under menu item device ().
Conduct the measurement corresponding to instruction manual. After acknowledgement the Measuring mode the coordinates are submitted from the GPS receiver to the measuring instrument and associated to the series of measurement. The coordinates are automatically saved when storing the measuring results.
For saving batteries energy turn off the GPS receiver during disuse or after transmission of the coordinates to the measuring instrument.
When printing out the measuring results via thermal printer, the position coordinates are automatically added. The measuring results and corresponding coordinates can be printed out immediately at the site or can be transferred to the PC for subsequent processing. If the GPS receiver is not being used, prior to measurement the GPS function of the measuring instrument has to be deactivated.
If in the measuring instrument the GPS function is activated and the GPS receiver is not connected or not ready, the note No GPS-Position appears. Now the measuring process can be aborted or continued without the acquisition of GPS-data.
HMP PDGpro XXXX
meas. range 300
unit MN/m
graphics calibr. date xback
GPS
-
20
SETTINGS
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Settings Measuring Instrument
Display Under the menu item Display the following settings can be carried out. Date set date Time set time Language choose language Contrast set contrast display Illumination set backlight display Back Return to previous menu
Device Under the menu item Device the following settings can be carried out for device configuration: Meas. range (300/600/ set measuring range
762/force) ( of loadplate or force) GPS ( / x) activate / deactivate GPS Unit (MN/m / MPa) set unit Graphics () show curve Calibr. date ( / x) show / do not show calibration
date on startup screen Back return to previous menu
Printer Under the menu item Printer the following settings can be carried out for printer configuration:
Print head () print out protocol head active. Date / Time ( / x) print out date / time Graphics ( / x) print out curve Contact data ( / x) print out company data (optional) Back return to previous menu
Contact data In particular, these are the company details of the device owner, which can optionally be printed out in the protocol head. The company details are not programmed by default. This requires a file with the desired data, which could be provided on request and would have to be installed on the measuring instrument.
datetimelanguage
contrastIllumination
back
HMP PDGpro XXXX
HMP PDGpro XXXX
meas. range 300GPS unit MN/m
graphics calibr. date xback
x
HMP PDGpro XXXX
print head date/time
graphics
back
contact data x
-
21
MAINTENANCE
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Calibration
In terms of DIN 18 134 re-calibration is required at least annually.
HMP Magdeburger Prfgertebau GmbH has calibrated instrumentation used to conduct force and distance calibrations. In addition, necessary repairs may be carried out.
Hotline HMP Magdeburger Prfgertebau GmbH Blstringer Strae 6 D-39126 Magdeburg Tel.: (03 91) 2 51 46 66 (03 91) 2 51 46 67 Fax: (03 91) 2 51 46 68 E-Mail: [email protected]
-
22
ANNEX 1
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Plate Load Tester spare parts
HM
PB
lstrin
ge
r St
r. 6
3912
6 M
ag
de
bur
gNa
me
Bene
nnun
gD
atu
m
Da
tum
Bea
rb.
Gep
r.
Buc
h-st
ab
e
Zeic
hnun
gsn
umm
er
zul
ss. A
bw
. fr
Ma
sse
ohn
e T
ole
ranz
ang
ab
eH
alb
zeug
/ W
erk
sto
ff
2004
nd
eru
ng
Ma
ssta
b
Hie
tsc
hold
Pla
ttend
ruc
kger
ae
t
PDG
100
0
25.0
8.
36
45
21
Nr.
Beze
ichn
ung
1A
usle
ge
r mit
Me
ssfin
ge
r
6H
and
pum
pe
5D
ruc
kzyl
ind
er
4La
stp
latte
3M
ess
ge
ste
ll
Tra
vers
e
7
7M
ess
koffe
r
-
23
ANNEX 2
Service +49(0)391 2514666 07/2013 Instruction Manual HMP PDGpro en
Operating Instructions Hydraulic System Parts of Hydraulic System Hydraulic system comprises of parts by LUKAS Company: Hand pump ZPH 1A/1 PN 500 Hydraulic cylinder LZM 10/160 PN 500 Hydraulic hose DN6PN700-2000RT Quick coupler The delivery includes operating instructions by LUKAS Company for hand pump, hydraulic cylinder and quick coupler.
Hydraulic system was installed complete and vented before delivery.
Using Hydraulic System
Before using the hydraulic system users have to read the operating instructions by LUKAS Company. Users must observe all safety instructions and warnings, to avoid bodily injuries and damages of the system.
The pump lever must be locked for transport.
Advancing cylinder Close the drain valve on the pump by turning the hand wheel
clockwise (to the right) to advance the cylinder. Unlock the pump lever and move it up and down to load
pressure.
Also read the operating instructions of hand pump and hydraulic cylinder by LUKAS Company to get more information about advancing and retracting cylinder.
Retracting cylinder Open the drain valve on the pump by turning the hand wheel
against clockwise (to the left) to release pressure or retract the cylinder. Turn the hand wheel slowly to have control over the load.
Release the pressure always slowly to have control over the load.
Maintenance Information and tips with regard to maintenance adding oil to the pump, oil recommendations venting the cylinder troubleshooting read in operating instructions by LUKAS Company.
-
Deckblatt Bedienungsanleitung HMP PDGpro E.pdfba_PDGpro_07_2013_neues_Layout_en.pdfCE EG-Konformittserklrung englisch LFG - 2013.pdf
/ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False
/CreateJDFFile false /Description > /Namespace [ (Adobe) (Common) (1.0) ] /OtherNamespaces [ > /FormElements false /GenerateStructure false /IncludeBookmarks false /IncludeHyperlinks false /IncludeInteractive false /IncludeLayers false /IncludeProfiles false /MultimediaHandling /UseObjectSettings /Namespace [ (Adobe) (CreativeSuite) (2.0) ] /PDFXOutputIntentProfileSelector /DocumentCMYK /PreserveEditing true /UntaggedCMYKHandling /LeaveUntagged /UntaggedRGBHandling /UseDocumentProfile /UseDocumentBleed false >> ]>> setdistillerparams> setpagedevice