70082 Instruction manual Transmitter Cond 7100 e/2(X)H

Upload: others

Post on 28-Aug-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

70082

Instruction manualTransmitter Cond 7100 e/2(X)H

Page 2: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

WarrantyDefects occurring within 1 year from delivery date shall be remedied freeof charge at our plant (carriage and insurance paid by sender).

Subject to change without notice.

Return of products under warantyPlease contact METTLER TOLEDO’s Customer Service Dept. before returninga defective device. Ship the cleaned device to the address you have beengiven. If the device has been in contact with process fluids, it must bedecontaminated/disinfected before shipment. In that case, please attach acorresponding certificate, for the health and safety of our service personnel.

Disposal (Directive 2002/96/EC of January 27, 2003)Please observe the applicable local or national regulationsconcerning the disposal of “waste electrical and electronicequipment”.

Mettler-Toledo GmbH, Process Analytics, Industrie Nord, CH-8902 Urdorf, Tel. +41 (01) 736 22 11 Fax +41 (01) 736 26 36 Subject to technical changes. Mettler-Toledo GmbH, 06/05. Printed in Germany.

Page 3: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7

Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8Conformity with FDA 21 CFR Part 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . .9EC-Type-Examination Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10

Overview of Transmitter Cond 7100 e/2(X)H . . . . . .13Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

Packing list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15Pipe mounting, panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

Installation and connection . . . . . . . . . . . . . . . . . . . . . . .18Information on installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18, 20Division 2 wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19Wiring examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22- 4-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22- 2-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23- with Mettler-Toledo 2-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . .24- with Mettler-Toledo 4-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . .25

User interface and display . . . . . . . . . . . . . . . . . . . . . . . .26Operation: Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28

Safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29Sensocheck, Sensoface sensor monitoring . . . . . . . . . . . . . . . . . . . . . . . .29GainCheck device self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29Automatic device self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30

Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31(Current output / Loop current, HART communication, Alarm) . . . . . . . .31

Passcodes (Factory settings) . . . . . . . . . . . . . . . . . . . . . . .33

Contents

3

Page 4: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H

Contents

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34Menu structure of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35Overview of configuration steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36Individual settings (for copy) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38Temperature compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54

Passcodes according to FDA 21 CFR Part 11 . . . . . . . . . .56

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58Calibration by entry of cell constant . . . . . . . . . . . . . . . . . . . . . . . . . . . .60Calibration with calibration solution . . . . . . . . . . . . . . . . . . . . . . . . . . . .62Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64Temp probe adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66

Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .67

Display of output currents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Display of calibration data (Cal nfo) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Display of last error message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Specify output current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69Error messages (error codes) . . . . . . . . . . . . . . . . . . . . . .70Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75Product line and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76Calibration solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82Concentration curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84FM Control Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92CSA Control Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94

Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

4

Page 5: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

5

Safety information

Be sure to read and observe the following instruc-tions!

The device has been designed using state of the art technologyand it complies with the applicable safety regulations.When operating the device, certain conditions may neverthe-less lead to danger for the operator or damage to the device.

Caution!Commissioning may only be carried out by trained experts.Whenever it is likely that protection has been impaired, thedevice shall be made inoperative and secured against unintend-ed operation.

The protection is likely to be impaired if, for example:• the device shows visible damage• the device fails to perform the intended measurements• after prolonged storage at temperatures above 70 °C• after severe transport stresses

Before recommissioning the device, a professional routine testin accordance with EN 61010-1 must be performed. This testshould be carried out by the manufacturer.

Caution!Before commissioning it must be proved that the device maybe connected with other equipment.

Page 6: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

6 Transmitter Cond 7100 e/2(X)H

Safety precautions for installation

• The stipulations of EN 60079-10 / EN 60079-14 must be observed during commissioning.

• The Transmitter Cond 7100 e/2H is approved for measure-ments in FM Class I Div II.

• The Transmitter Cond 7100 e/2XH is approved for operationin the following locations: ATEX, FM Zone 1 with measurement in Zone 0, and FM Class I Div 1.

Connection to supply units

• Transmitter Cond 7100 e/2H: Before connecting this device to a supply unit, make sure that its output voltage cannot exceed 30 V DC. Do not use alternating current or mains power supply!

• Transmitter Cond 7100 e/2XH: This device may only be connectedto an explosion-proof power supply unit (for input ratings refer to annex of EC-Type-Examination Certificate).Before commissioning it must be made sure that the connections to other equipment such as power supply unitand cables are intrinsically safe.

Terminals: suitable for single wires / flexible leads up to 2.5 mm2 (AWG 14)

Note for cleaning in a hazardous locationTo protect against electrostatic discharge, the device may onlybe cleaned with a damp cloth in hazardous locations.

Page 7: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

7

Intended use

The Transmitter Cond 7100 e/2(X)H is used for measurement ofelectrical conductivity and temperature in liquids. Fields of appli-cation are: biotechnology, chemical industry, environment, foodprocessing, water/waste-water treatment. The rugged moldedenclosure can be fixed into a control panel or mounted on a wallor at a post. The protective hood provides additional protectionagainst direct weather exposure and mechanical damage. TheTransmitter has been designed for 2- and 4-electrode sensors.

• The Transmitter Cond 7100 e/2H is approved for measure-ments in FM Class I Div II.

• The Transmitter Cond 7100 e/2XH is approved for operation in the following locations: ATEX, FM Zone 1 with measurement in Zone 0, and FM Class I Div 1.

Trademarks

The following names are registered trademarks. For practical rea-sons they are shown without trademark symbol in this manual.

InPro® is a registered trademark of Mettler-Toledo.HART® is a registered trademark of the HART CommunicationFoundation.

Page 8: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

8 Transmitter Cond 7100 e/2(X)H

EC Declaration of Conformity

Page 9: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

9

Conformity with FDA 21 CFR Part 11

In their directive “Title 21 Code of Federal Regulations, 21 CFRPart 11, Electronic Records; Electronic Signatures” the USAmerican health agency FDA (Food and Drug Administration)regulates the production and processing of electronic docu-ments for pharmaceutical development and production. Thisresults in requirements for measuring devices used for corre-sponding applications. The following features ensure that themeasuring devices of the Transmitter Cond 7100 e/2(X)H Seriesmeet the demands of FDA 21 CFR Part 11:

Electronic SignatureAccess to the device functions is regulated and limited by indi-vidually adjustable codes – “Passcodes” (for Passcode Editorsee Page 56, overview of factory settings see Page This prevents unauthorized modification of device settings ormanipulation of the measurement results. Appropriate use ofthese passcodes makes them suitable as electronic signature.

Audit TrailEvery (manual) change of device settings can be automaticallydocumented. For that purpose, each change is marked by a“Configuration Change Flag”, which can be interrogated anddocumented via HART communication. Then the changeddevice settings/parameters can also be retrieved and docu-mented via HART cmmunication.

33).

Page 10: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

10 Transmitter Cond 7100 e/2(X)H

EC-Type-Examination Certificate

Page 11: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

11

Page 12: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H12

Page 13: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

13

Overview of Transmitter Cond 7100 e/2(X)H

1

2

3

4

5

6

7

8

2/4-elcondinput

cond I hi

cond U hi

cond U lo

cond I lo

shield

7100 e/2H:not connected7100 e/2XH:equipotentialbonding

RTD

RTD

Tempinput

10

11

+ Output

– Output

EEx ia IIC(7100 e/2XH)

EEx ib IIC(7100 e/2XH)

4 ... 20 mA

n.c.

9

mA/HART

Page 14: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

14 Transmitter Cond 7100 e/2(X)H

Packing listCheck the shipment for transport damage and completeness.The package should contain:• Front unit of Transmitter• Lower case• Bag containing small parts• Instruction manual• Specific test report

Assembly

11

10

9

8

7 6 5 4

1

2

3

6 Sealing inserts (1 piece)7 Rubber reducer (1 piece)8 Cable glands (3 pieces)9 Filler plugs (3 pieces)10 Hexagon nuts (5 pieces)11 Sealing plugs (2 pieces),for sealing in case of wall mounting

1 Jumper (2 piece)2 Washer (1 piece), for conduit

mounting: place washer between enclosure and nut

3 Cable ties (3 pieces)4 Hinge pin (1 piece),

insertable from either side5 Enclosure screws (4 pieces)

Fig.: Assembling the enclosure

Page 15: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

15

Mounting plan

144

144

15 42

84

8032

21

43

105

27

726,

2

1 2

3

4

1 Cable gland (3 pieces)2 Breakthroughs for cable gland or

conduit 1/2“, ø 21.5 mm (2 breakthroughs)Conduits not included!

3 Breakthroughs for pipe mounting(4 breakthroughs)

4 Breakthroughs for wall mounting(2 breakthroughs)

Fig.: Mounting plan

Page 16: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

16 Transmitter Cond 7100 e/2(X)H

40 60132

1

2

34

5

Fig.: Pipe-mount kit

1 Protective hood (if required)2 Hose clamps with worm gear drive to DIN 3017 (2 pieces)3 Pipe-mount plate (1 piece)4 For vertical or horizontal posts or pipes5 Self-tapping screws (4 pieces)

1

132165

173

1

Fig.: Protective hood for wall and pipe mounting

Pipe mounting, panel mounting

Page 17: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

17

1

2

3

45

max. 2578 27

1...22

1 Screws (4 pieces)2 Gasket (1 piece)3 Panel4 Span pieces (4 pieces)5 Threaded sleeves (4 pieces)

Fig.: Panel-mount kit

Panel cutout 138 x 138 mm(DIN 43700)

Page 18: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

18 Transmitter Cond 7100 e/2(X)H

Installation and connection

Information on installation• Installation may only be carried out by trained experts in

accordance with this instruction manual and as per applicable local and national codes.

• Be sure to observe the technical specifications and input ratings.• Be sure not to notch the conductor when stripping the insulation.• When commissioning, a complete configuration must be

carried out by the system administrator.

Connection to supply units• Transmitter Cond 7100 e/2H: Before connecting this

device to a supply unit, make sure that its output voltage cannot exceed 30 V DC. Do not use alternating current or mains power supply!

• Transmitter Cond 7100 e/2XH: This device may only be connectedto an explosion-proof power supply unit (for input ratings refer to annex of EC-Type-Examination Certificate).

Terminals: suitable for single wires /flexible leads up to 2.5 mm2

(AWG 14)

Division 2 wiringThe connections to the Transmitter must be installedin accordance with the National Electric Code (ANSI-NFPA 70) Division 2 hazardous (classified) locationnon-incendive wiring techniques.FM Control Drawing: Refer to page 90.

Warning!Additional safety precautions have to be taken for applications inhazardous locations to CSA!(See Pg 93 et seq.)

Page 19: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

19

Terminal assignments

Fig.: Terminal assignments Transmitter Cond 7100 e/2H

Fig.: Terminal assignments Transmitter Cond 7100 e/2XH

Page 20: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

20 Transmitter Cond 7100 e/2(X)H

1 Recommended strippinglengths for multi-core cables

2 Pulling out the terminalsusing a screwdriver(also see 6)

3 Cable laying in the device

4 Connecting lines for loop current

5 Cover for electrode andtemperature probe terminals

6 Area for placing the screwdriver to pull out the terminals

7 Connection of handheld terminal

Fig.: Information on installation, rear side of device

1 2

3

4

65

7m

m

280

mm

7

Page 21: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

21

Page 22: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H22

Wiring examples

Cond measurement with 4-electrode sensor Ih

i

Vhi

Vlo

RTD

RTD

Ilo

shie

ld

Caution!Place jumper across terminals 4 and 5!When using a sensor with Solution Ground connection (SG) ora separate SG connection, the jumper is not required!

Cond 7100 e/2(X)H

Page 23: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

23

Cond measurement with 2-electrode sensor(coaxial electrodes)

Ihi

Vhi

Vlo

RTD

RTD

Ilo

shie

ld

Caution!Place jumpers: across terminals 1 and 2

across terminals 3 and 4across terminals 4 and 5

Jumper Jumper

Cond 7100 e/2(X)H

Page 24: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

24 Transmitter Cond 7100 e/2(X)H

Wiring examples

Cond measurement with Mettler-Toledo 2-elec-trode sensor via VP plug

Ihi

Vhi

Vlo

RTD

RTD

Ilo

shie

ld

Cond 7100 e/2(X)H

VP plug

wt

tran

sp.

rd gnwt/

bu

bu bk

Page 25: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

25

Cond measurement with Mettler-Toledo 4-electrodesensor via VP plug

Ihi

Vhi

Vlo

RTD

RTD

Ilo

shie

ld

Cond 7100 e/2(X)H

wt

tran

sp.

rd gn

VP plug

wt/

bu

bu bk

Page 26: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H

User interface and display

1 Display2 Mode indicators (no keys),

from left to right:- Measuring mode- Calibration mode- Alarm- Wash contact (Model 7100 Cond only)- Configuration mode

User interface

3 Keypad4 Coding5 Rating plate6 Model designation7 Alarm LED

26

Page 27: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

27

1 2 3 4 5 6 7 8 9 10

11

12

13

1617

20

18

19

15 14

1 Passcode entry2 Display of measured variable*3 Temperature4 Current output5 Limit values6 Alarm7 Sensocheck8 Calibration9 Interval/response time10 Wash contact*11 Measurement symbol12 Proceed with enter13 Bar for identifying the device

status, above mode indicatorsfrom left to right:- Measuring mode- Calibration mode- Alarm- Wash contact*- Configuration mode

14 Lower display15 Manual temp specification16 Hold mode active17 Waiting time running18 Electrode data19 Main display20 Sensoface

* Not in use

Display

Page 28: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

28 Transmitter Cond 7100 e/2(X)H

Operation: Keypad

Start, end calibration

Start, end configuration

Select digit position(selected position flashes)

Edit digit

• Calibration: Continue in program sequence• Configuration: Confirm entries,next configuration step• Measuring mode: Display output current

Cal Info, display of cell constant

Error Info: Display of last error message

Start GainCheck device self-test+

Page 29: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

29

Safety functions

Sensocheck, Sensoface sensor monitoringSensocheck continuously monitors the sensor and lines. Sensocheck can be switched off (Configuration, Pg 55).

Sensoface provides information on the conductivitysensor condition. Significant sensor polarization effectsor an excessive cable capacitance are indicated.

GainCheck device self testA display test is carried out, the software version is displayedand the memory and measured value transfer are checked.

Start GainCheck device self-test:

Automatic device self-testThe automatic device self-test checks the memory and meas-ured-value transfer. It runs automatically in the background atfixed intervals.

+

Page 30: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

30 Transmitter Cond 7100 e/2(X)H

Safety functions

Hold mode

Display:

The Hold mode is a safety state during configuration and cali-bration. The loop current is frozen (Last) or set to a fixed value(Fix).If the calibration or configuration mode is exited, theTransmitter remains in the Hold mode for safety reasons. Thisprevents undesirable reactions of the connected peripheralsdue to incorrect configuration or calibration. The measuredvalue and “HOLD” are displayed alternately. The Transmitteronly returns to measuring mode after enter is pressed and 20 seconds have passed.

Configuration mode is also exited automatically 20 minutes(timeout) after the last keystroke. The Transmitter returns tomeasuring mode.Timeout is not active during calibration.

Behaviour of output signal:

Last: The loop current is frozen at its last value.Recommended during short configuration procedures.The process should not change decisively during con-figuration. Changes are not noticed with this setting!

Fix: The loop current is set to a value that is noticeablydifferent from the process value in order to signal thecontrol system that the Transmitter is being worked at.

For configuration see Pg 51.

Page 31: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

31

Outputs

Current output / Loop currentThe loop current is controlled by the process variable selectedin the configuration.The current beginning and end can be set to represent anydesired value. To check connected peripherals (e.g. limitswitches, controllers), the loop current can be manually speci-fied (see Pg. 41).

HART communicationThe Transmitter Cond 7100 e/2(X)H can be remote-controlledvia HART communication. It can be configured using a hand-held terminal or from the control room. Measured values, mes-sages and device identification can be downloaded at any time.This allows easy integration also in fully automatic processcycles.

A list of the HART commands can be found in the “Transmitter Cond 7100 e/2(X)H Transmitter-Specific CommandSpecification”: www.mtpro.com/transmitters.

AlarmThe alarm delay is configurable.Error messages can also be signaled by a 22 mA loop current(see Configuration, Pg 55).

The alarm LED on the front panel can be configured as follows:HOLD off: Alarm: LED flashing HOLD on: Alarm: LED on. HOLD: LED flashing.

Page 32: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

32 Transmitter Cond 7100 e/2(X)H

Page 33: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

33

Passcodes (Factory settings)The passcodes allow fast access to the functions

Calibration

Key+passcode Description

Cal Info0000

1100

Product calibration1105

Temp probe adjustment1015

Configuration

Error InfoDisplay last error and erase

0000

Configuration

Sensor monitorDisplay resistance and temp

1200

2222

Key+passcode Description

Calibration: Calibration solution0110

5555 Current sourceSpecify output current

Calibration: Entry of cell constant

67

60

64

66

70

34

67

62

68

Page

Page

Passcode editor

Administrator passcodeChanging the passcodes

1989

Key+passcode Description

56

Page

Page 34: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

34 Transmitter Cond 7100 e/2(X)H

ConfigurationIn the Configuration mode you set the device parameters.

Activate with conf

Enter passcode “1200”*)

Edit parameter with 4 and 5,confirm/continue with enter.(End with conf, then enter.)

During configuration theTransmitter remains in the Holdmode for reasons of safety. Theloop current is frozen (at its lastvalue or at a preset fixed value,depending on the configuration),Sensoface is off, mode indicator“Configuration” is on.

Activate

Hold

Input errors The configuration parameters arechecked during the input. In the caseof an incorrect input ”Err” is dis-played for approx. 3 s. The incorrectparameters cannot be stored. Inputmust be repeated.

End End with conf. The measured valueand Hold are displayed alternately,“enter” flashes. End Hold mode withenter. The display shows the meas-ured value. The output currentremains frozen for another 20 s(HOLD icon on, “hourglass” flashes).

HOLD icon

*) Factory setting, for passcode editing see Pg 56

Page 35: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

35

Menu group Code

Current output o1.

Temperature compensation

tc.

Alarmsettings

AL.

DisplaySelect menu group

Previous menugroup:

Menu structure of configuration

Submenu 1

Submenu 2

Submenu ...

The configuration steps are assigned to different menu groups:• Current output (code: o1.)• Temperature compensation (code: tc.)• Alarm settings (code: AL.)With the arrow keys you can jump between the individual menu groups.Each menu group contains menu items for setting the parameters.

Select submenu

Pressing enter accesses the submenus. The values are edited using thearrow keys. Pressing enter confirms/stores the settings.Return to measurement: Press conf. Press enter to confirm safety prompt.After 20 sec the Transmitter will be in measuring mode again.

Example: “o1.” is displayed with all menu items of the“Current output” menu group.

Page 36: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

36 Transmitter Cond 7100 e/2(X)H

Overview of configuration steps

2-electrode, 4-electrode

µS, mS/cm, S/m, MΩ·cm, SAL, %, USP

LIN / LOG

tc.UnIT Select temperature unit °C / °F

Select temperature probe Pt100/Pt1000/NTC30/NTC8.55

Select temperature compensation(not for SAL)

OFF/LIN/NLF (natural waters)/-01- FCT (NaCl traces)-02- FCT (HCl traces)-03- FCT (NH3 traces)

Lin: Enter temperature coefficient xx.xx %/K (02.00 %/K)

tc Temperature compensation

MenuCode

out1 Current output

o1.CELL Sensor selection

Select measured variable

Select solution (Conc), see Pg 42Codes:

Characteristic linear / logarithmic(not for SAL, Conc, USP)

Selection / Default

ALrt Alarm settings

AL.SnSO Select Sensocheck ON / OFF

Enter alarm delay 0000 ... 0600 SEC (0010 SEC)LED in HOLD mode ON / OFF

(Factory setting bold print)

o1.4mA

o1.20mA

o1.4mA

o1.UnIT

o1.CoNC

o1.CHAR

o1.20mA

tc.rTD

tc.

tc.lin

AL.dLY

AL.LED

NaCl HCl NaOH H2SO4 HNO3

-01- -02- -03- -04- -05-

Fix:

Time constant of output filter xxxx SEC (0000 SEC)22 mA signal for error messages ON / OFFSignal behavior during HOLD Last / Fix

Enter fixed value xxx.x mA (021.0 mA)

xxxx mS (000.0 mS)

Enter current endLOG: Enter current beginning

Enter current end

xxxx mS (000.0 mS)

in decades: 0.001 ... 1000 mS(0.100 mS)

in decades: 0.001 ... 1000 mS(100.0 mS)

LIN: Enter current beginning

o1.FtME

o1.FAIL

o1.HoLD

o1.FIX

Page 37: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

37

Code Parameter Factory Individual setting setting

o1.CELL Sensor type 2-EL ____________

o1.UnIT Measurement unit mS/cm ____________

o1.CoNC Concentration NaCl ____________

o1.CHAR Characteristic (LIN/LOG) LIN ____________

o1.4mA Current start 000.0 mS ____________

o1.20mA Current end 100.0 mS ____________

o1.FtME Filter time 0000 SEC ____________

o1.FAIL 22mA signal OFF ____________

o1.HoLD Hold behavior LAST ____________

o1.FIX Fix current 021.0 mA ____________

tc.UnIT Unit °C / °F °C ____________

tc.rTD Temp probe Pt 100 ____________

tc. Temperaturecompensation OFF ____________

tc.LIN TC process medium 02.00 %/K ____________

AL.SnSO Sensocheck OFF ____________

AL.dLY Alarm delay 0010 SEC ____________

AL.LED LED in HOLD mode OFF ____________

Individual settings(Original for copy)

Page 38: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

38 Transmitter Cond 7100 e/2(X)H

ConfigurationCurrent output: Select sensor type.

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 39).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Select characteristic (LIN/LOG)

*) Factory setting, for passcode editing see Pg 56

Page 39: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

2-EL(2-El/ 4-El)

39

Note:: Characters represented in gray are flashing and can be edited.

Code Display Action Choices

o1. Select configuration(Press conf.)

Enter passcode “1200*)”(Select position with 4 key andedit number with 5 key. When the display reads “1200”,press enter to confirm.)

The Transmitter is in HOLDmode (HOLD icon is on).

Select sensor2-electrode sensor / 4-electrode sensorSelect with 4 arrow key Proceed with enter

After correct input a welcome text(CONF) is displayedfor approx. 3 sec.

*) Factory setting

Page 40: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

40 Transmitter Cond 7100 e/2(X)H

ConfigurationCurrent output: Select measured variable

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 41).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Select characteristic (LIN/LOG)

*) Factory setting

Page 41: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

41

Code Display Action Choices

o1. Select measured variable:

Select with 4 arrow keyProceed with enter

Conductivity:• 0.000 ... 9.999 µS/cm• 00.00 ... 99.99 µS/cm • 000.0 ... 999.9 µS/cm • 0000 ... 9999 µS/cm • 0.000 ... 9.999 mS/cm• 00.00 ... 99.99 mS/cm• 000.0 ... 999.9 mS/cm• 0.000 ... 9,999 S/m• 00.00 ... 99.99 S/m

Resistivity:• 00.00 ... 99.99 MΩ·cm

Salinity (SAL):• 0.0 ... 45.0 ‰ (0 ... 35 °C)

Concentration (Conc):• 0.00 ... 9.99 % by wt

USP:• 00.00 ... 99.99 µS/cm

000.0 mS

(0.000 µS00.00 µS 000.0 µS0000 µS0.000 mS00.00 mS000.0 mS

0.000 S/m00.00 S/m

00.00 MΩ

0.00 SAL

00.00 %

USP)

Note:: Characters represented in gray are flashing and can be edited.

Page 42: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

42 Transmitter Cond 7100 e/2(X)H

ConfigurationOutput 1Concentration measurement: Select process solutions

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 43).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Select characteristic (LIN/LOG)

*) Factory setting

Page 43: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

43

Code Display Action Choices

o1. Only with 000.0 % you canselect the process solution:

Select with 4 arrow key

-01-NaCl (0.00 ... 9.99 % by wt) (0 ... 100 °C)

-02-HCl (0.00 ... 9.99 % by wt)(0 ... 50 °C)

-03- NaOH (0.00 ... 9.99 % by wt)(0 ... 100 °C)

-04-H2SO4 (0.00 ... 9.99 % by wt)(0 ... 110 °C)

-05-HNO3 (0.00 ... 9.99 % by wt)(0 ... 50 °C)

Proceed with enter

-01-SOL

(-01-SOL-02-SOL-03-SOL-04-SOL-05-SOL)

Concentration measurementFor the solutions listed above, the Transmitter can determine the substanceconcentration from the measured conductivity and temperature values in% by wt. The measurement error is made up of the sum of measurementserrors during conductivity and temperature measurement and the accuracyof the concentration curves stored in the Transmitter, see Pg 84 et seq. We recommend to calibrate the Transmitter together with the sensor, pre-ferrably in the same conductivity range as measured later. For exact temper-ature measurement, you should perform a temperature probe adjustment.For measuring processes with rapid temperature changes, a separate tem-perature probe with fast response should be used.

Page 44: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

44 Transmitter Cond 7100 e/2(X)H

ConfigurationOutput current. LIN characteristic. Current start / end

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 45).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Select: LIN characteristic

*) Factory setting

Page 45: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

45

Code

LIN(LIN / LOG)

Display Action Choices

o1. Select output characteristicSelect with 4 arrow keyProceed with enter(Step omitted for % (Conc) orSAL

000.0 mS(xxx.x mS)

100.0 mS(xxx.x mS)

Conductivity

[mS/cm]

204

200

0

Assignment of measured values: Current beginning and current end

[mS/cm]

204

200

100[mA]

Example 1: Range 0...200 mS/cm Example 2: Range 100...200 mS/cmAdvantage: Higher resolution in rangeof interest

[mA]

Output current

Conductivity

With LIN selected:• Enter current start

(lower end of scale).Select with 4 key, edit number with 5 key, proceed with enter.• Enter current end

(upper end of scale). Proceed with enter

Output current

Page 46: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

46 Transmitter Cond 7100 e/2(X)H

ConfigurationOutput current. LOG characteristic. Current start / end

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 47).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Select: LOG characteristic

*) Factory setting

Page 47: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

47

Code Display Action Choices

o1. With LOG selected:• Enter lower end of scale

(=current start)Select with 4 key, edit number with 5 key, proceed with enter.

0.1 mS(0.1 mS1.0 mS10 mS100 mS1000 mS)

100 mS(0.1 mS1.0 mS10 mS100 mS1000 mS)

• Enter upper end of scale (=current end)

Select with 4 key, edit number with 5 key.

Proceed with enter

Example: Measurement range over 3 decades

Characteristic: LOG• 4 mA 0.1 mS/cm• 20 mA 100 mS/cm

[mA]

100

4.0

Output current

Y

X

Conductivity[mS/cm]

20.09.3 14.7

1

10

0.1

Page 48: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

48 Transmitter Cond 7100 e/2(X)H

ConfigurationOutput. Time constant of output filter

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 49).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Selection: LOG characteristic

*) Factory setting

Page 49: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

49

0 sec0 ... 120 sec

Code Display Action Choices

o1. Time constant of output filterDefault setting: 0 s (inactive).To specify a time constant:Select with 4 key, edit number with 5 key, proceed with enter

Time constant of output filter (attenuation)To smoothen the current output, a low-pass filter withadjustable filter time constant can be switched on. When thereis a jump at the input (100 %), the output level is 63 % afterthe time constant has been reached.The time constant can be set from 0 to 120 sec.If the time constant is set to 0 s, the current output follows theinput. Note:The filter only acts on the current output, not on the display!

Time constant 0 - 120 sec

Conductivity

Page 50: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

50 Transmitter Cond 7100 e/2(X)H

ConfigurationOutput. Output current during Error and HOLD.

Select sensor

Select measured variable

Enter current beginning

Enter current end

Set output filter

1. Press conf key.2. Enter passcode 1200*).3. Select Current output menu group

using arrow keys. All items of thismenu group are indicated by the code “o1.”

4. Press enter to select menu,edit with arrow keys (see Pg 51).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Current output:

22 mA in the case of error

Signal behavior during HOLD

1

5

Select solution (Conc)

Selection: LOG characteristic

*) Factory setting

Page 51: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

51

OFF (OFF / ON)

Code Display Action Choices

o1. 22 mA signal for error messageSelect with 4keyProceed with enter

Output signal during HOLDLAST: During HOLD the lastmeasured value is maintained atthe outputFIX: During HOLD a value (to beentered) is maintained at theoutputSelect with 4keyProceed with enter

LAST(LAST / FIX)

Only with FIX selected:Enter current which is to flow atthe output during HOLDSelect position with 4key andedit number with 5 key.Proceed with enter

021.0 mA(04.0 ...22.0 mA

Output signal during HOLD:(see Pg 30)Output current[mA]

Output signal HOLDSetting FIX = 21.0 mA

HOLD active

21

4

Output signal HOLDSetting LAST

HOLD active

Page 52: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

52 Transmitter Cond 7100 e/2(X)H

ConfigurationTemperature compensation

1. Press conf key.2. Enter passcode 1200*).3. Select Temperature compensation

menu group using arrow keys. All items of this menu group are indicated by the code “tc.”

4. Edit with arrow keys (see Pg 53).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

3

4

Temp compensation:

Selecttemperature compensation

1

5

Current output:

Select °C/°F

Select temperature probe

*) Factory setting

Page 53: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

53

OFF(OFFLINnLFnACLHCLnH3)

Code Display Action Choices

tc.

Only with linear temperature com-pensation (LIN) selected:Enter temperature coefficient*).Select position with 4key, edit num-ber with 5 key. Proceed with enter

02.00%/K(XX.XX %/K)

Temp compensation selection(not for USP, CONC, SAL)OFF: Temperature compensationswitched off. Select with 4 key,proceed with enterLIN: Linear temperature compen-sation with entry of temperaturecoefficient and reference tempera-ture.nLF: Temperature compensation fornatural waters to EN 27888

NaCl (nACL): Temperature com-pensation for ultrapure water withNaCl traces

HCl (HCL): Temperature compensation forultrapure water with HCl traces

NH3 (nH3):Temperature compensation forultrapure water with NH3 traces

°C(°F)

Specify temperature unitSelect with 4 arrow keyProceed with enter

Select temperature probeSelect with 4 arrow keyProceed with enter

Pt100(PT1000,NTC30, NTC8.55)

*) Reference temperature 25 °C

Page 54: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

54 Transmitter Cond 7100 e/2(X)H

ConfigurationAlarm settings

1. Press conf key.2. Enter passcode 1200*).3. Select Alarm settings menu group

using arrow keys. All items of this menu group are indicated by the code “AL.”

4. Press enterto select menu,edit with arrow keys (see Pg 55).Confirm (and proceed) with enter.

5. End: Press conf, then enter

2

Current output:

1

5

3

Alarm settings:

Sensocheck4

Delay

LED in HOLD mode

Temp compensation:

*) Factory setting

Page 55: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

55

Code Display Action Choices

OFF (ON / OFF)

AL. Select Sensocheck(Continuous monitoring of sen-sor properties)Select with 4 key. Proceed with enter

Alarm delaySelect with 4key, edit number with 5 key, proceed with enter

0010 s(xxxx s)

LED in HOLD modeSelect with 4 key, proceed withenter

OFF (ON / OFF)

Configuration Alarm HOLD

ON on flashes

OFF flashes off

LED in HOLD mode:

Page 56: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

56 Transmitter Cond 7100 e/2(X)H

Passcodes according to FDA 21 CFR Part 11

Access to the device functions can be protected with adjustablepasscodes if required. If such a protection is not required, you should use the presetpasscodes.

To call up passcode editor:Press conf key and enter Administrator passcode (Factory setting:1989).

Display Action Remark

1. Press conf key.2. Enter Administrator passcode

(1989):Welcome text is displayed

This text is displayedfor approx. 3 s

“Cal Info”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:0000

“Cal - Input of cell constant”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:1100

“Product calibration”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:1105

“Temp probe adjustment”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:1015

“Cal - with cal solution”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:0110

Page 57: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

57

Caution!If you have lost theAdministrator pass-code, the PasscodeEditor cannot becalled up! Pleaseconsult our technicalsupport!

Display Action Remark

“Error Info”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:0000

“Configuration”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:1200

“Sensor monitor”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:2222

“Current source”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:5555

“Administrator passcode”Edit: Arrow keysProceed with: enterCancel: conf

Default setting:1989

• “NO” to cancel new Administrator passcode

Proceed with enter (old passcode)Cancel: conf (old passcode)

• “YES” to take over new Administrator passcode

Select “YES” with arrow keys.Accept with enter (new passcode)Cancel: conf (old passcode)

Page 58: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

58 Transmitter Cond 7100 e/2(X)H

CalibrationCalibration adjusts the device to the sensor.

Activate with cal

Enter passcode*):

• 1100 Entry of cell constant • 0110 With calibration solution • 1105 Product calibration • 1015 Temp probe adjustment Select with 4 key, edit number with5 key, proceed with enter key (End with cal + enter.)

The loop current is frozen (at its lastvalue or at a preset fixed value,depending on the configuration),Sensoface is off, mode indicator“Calibration” is on.

Activate

Hold

Input errors The calibration parameters arechecked during the input. In the caseof an incorrect input ”Err” is dis-played for approx. 3 s. The incorrectparameters cannot be stored. Inputmust be repeated.

End End with cal.Safety prompt:The measured value and Hold are dis-played alternately, “enter” flashes. Press enter to end the Hold mode. The measured value is displayed. The output current remains frozen foranother 20 sec (HOLD icon on,“hourglass” flashes).

HOLD icon

During calibra-tion theTransmitterremains in theHold mode.

*) Factory setting, for passcode editing see Pg 56

Page 59: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

59

Information on calibrationCalibration adapts the Transmitter to the conductivity sensor. Calibration can be performed by:• Input of cell constant (e.g. for ultrapure-water sensors)• Determining the cell constant with a known

calibration solution• Sampling (product calibration)• Temperature probe adjustment

Note:• All calibration procedures must be performed by trained

personnel. • During the calibration procedure the temperature must be

kept constant.• Incorrectly set parameters may go unnoticed, but change the

measuring properties.

Particularly with stray-field sensors the cell constant can strong-ly vary when the sensor is mounted in restricted space. In thatcase, the cell constant should be determined with the sensormounted using a calibration solution or by a reference meas-urement at the product.

Page 60: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

60 Transmitter Cond 7100 e/2(X)H

Display Action Remark

The lower displayshows the conduc-tivity value.

(When there hasnot been an entryfor 6 sec, the lowerdisplay alternatelyshows the conduc-tivity and tempera-ture value.)

Calibration by entry of cell constant

Input of cell constant with simultaneous display of the uncom-pensated conductivity value and the temperature

Press cal key, enter passcode1100*)

Select with 4 key, edit number with 5 key,proceed with enter

Enter cell constant ofconnected sensor:

Select with 4 key, edit numberwith 5key.

A change in the cell constantalso changes the conductivityvalue.

Press enter to confirm cellconstant.

Transmitter is inHold mode.If an invalid pass-code is entered, theTransmitter returnsto measuring mode.

Ready for calibration Display (3 s)

*) Factory setting

Page 61: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

61

Display Action Remark

The Transmitter now displaysthe conductivity and tempera-ture.

The measured value is shownin the main display alternatelywith “Hold”; “enter” flashes.End calibration withenter.

Safety prompt

After end of calibra-tion, the outputsremain in Holdmode for approx.20 sec.

Page 62: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

62 Transmitter Cond 7100 e/2(X)H

Display Action Remark

When there has notbeen an entry for6sec, the lowerdisplay alternatelyshows the cellconstant andtemperature value.

Calibration with calibration solution(not for measured variables: S/m, SAL, % (Conc), USP)

Input of temperature-corrected value of calibration solutionwith simultaneous display of cell constant

Press cal key, enter passcode 0110*)

Select with 4 key, edit number with 5 key, proceed with enter

Immerse sensor in calibrationsolution.

Determine the temperature-corrected conductivity value ofthe calibration solution fromthe corresponding table (see Pg 82).

Transmitter is inHold mode.If an invalid pass-code is entered,Transmitter returnsto measuring mode.

Ready for calibration

Dismount and clean sensor

Display (3 sec)

Enter value of calibrationsolution.

Select with 4 key, edit numberwith 5key.

Press enter to confirm thecalibration data.

The cell constantand temperatureare alternatelydisplayed in thelower displayduring the input.

*) Factory setting

Page 63: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

63

Display Action Remark

The determined cell constant isdisplayed.Confirm with enter.

Safety prompt

• Be sure to use known calibration solutions and the respectivetemperature-corrected conductivity values.(see “Calibration solutions” Pg 82 et seq.).

• During the calibration procedure the temperature must bekept constant.

• For a good mass transfer, the solution should be stirred.• The configured measuring range is automatically used when

this calibration routine is selected. If the measuring range(measured variable) is S/m, SAL, % (Conc), or USP, an errormessage is displayed for 3 s and the routine stopped.

Clean sensor and re-place it inthe process.The Transmitter now displaysthe conductivity and tempera-ture.

The measured value is shownin the main display alternatelywith “Hold”; “enter” flashes.End calibration withenter.

After end of cali-bration, the out-puts remain inHold mode forapprox. 20 sec.

Notes : (also see Pg 59)

Page 64: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

64 Transmitter Cond 7100 e/2(X)H

For product calibration the measured variable is used as configured:Conductivity (µS/cm, mS/cm, S/m), resistivity (MΩ·cm). During productcalibration the sensor remains in the process. The measurement is onlyinterrupted briefly.Calibration is without TC correction.

Procedure: During sampling the currently measured value is stored inthe Transmitter. The Transmitter immediately returns to measuringmode. The calibration mode indicator flashes and reminds you thatcalibration has not been terminated. The sample is measured in the lab or directly on the site using aportable meter. To ensure an exact calibration, the sample temperatureshould correspond to the measured process temperature. The samplevalue is then entered in the Transmitter. The new cell constant is calcu-lated from these two values. If the sample is invalid, you can take overthe value stored during sampling. In that case the old calibration val-ues are stored. Afterwards, you can start a new product calibration.

Product calibrationCalibration by sampling

Display Action Remark

Product calibration step 1:Press cal key.Enter passcode 1105*).(Press 4 key to select position,enter number using 5 key,confirm with enter)

Take sample and store value.Proceed with enter

If an invalid pass-code is entered, theTransmitter returnsto measuring mode.

The sample is meas-ured in the lab ordirectly on the site.

Display (approx. 3 sec)

*) Factory setting

Page 65: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

65

Display Action Remark

Measuring mode:

From the flashing CAL modeindicator you see that prod-uct calibration has not beenterminated.

Product calibration step 2:When the sample value hasbeen determined, call up theproduct calibration oncemore (cal, passcode 1105*)).

Display (approx. 3 sec)

Enter lab value. The new cellconstant is calculated.

While the samplevalue is determined,the Transmitter is inmeasuring mode.

The new cell constant is dis-played.

Confirm with enter.

The measured value is shownin the main display alternatelywith “Hold”; “enter” flashes.End with enter.

Safety prompt. After end of calibra-tion, the outputsremain in Holdmode for approx.20 sec.

New calibration:Press cal.

*) Factory setting

Page 66: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

66 Transmitter Cond 7100 e/2(X)H

Display Remark

Measurement

In the measuring mode the main display shows theconfigured process variable (conductivity, concentration,resistivity, salinity), the lower display shows the tempera-ture. During calibration you can return to measuringmode by pressing the cal key, during configuration bypressing conf and then enter (waiting time for meas-ured-value stabilization approx. 20 sec).

Display Action Remark

Measure the temperature ofthe process medium using anexternal thermometer. Entermeasured temperature value:Select with 4, edit numberwith 5, proceed with enter. End adjustment with enter.HOLD will be deactivatedafter 20 sec.

Temp probe adjustment

Activate calibration(Press cal.Enter passcode 1015*).)Select with 4key, edit number with 5key,proceed with enter.

Ready for calibration

Wrong settingschange the measure-ment properties! If an invalid passcodeis entered, theTransmitter returns tomeasuring mode.

Default:Value of secondarydisplay.

Transmitter is in Hold mode. Display for approx. 3 sec

*) Factory setting

Page 67: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

67

Display of output currentsPress enter while in measuring mode.For 5 sec, the secondary display shows the outputcurrent instead of the temperature

Diagnostics functions

Entry/display Remark

Display of calibration data (Cal Info)Press cal while in measuring mode and enter pass-code 0000*). The current cell constant is shown inthe main display. After 20 sec the Transmitterreturns to measuring mode (immediate return atpressing enter).

Sensor monitorfor validation of sensor and complete measured-value processing. Press conf while in measuringmode and enter passcode 2222*). The measuredresistance is shown in the main display, the measur-ing temperature in the lower display.Press enter to return to measurement.

Display of last error message (Error Info)Press conf while in measuring mode and enter passcode 0000*). The last error message is displayedfor approx. 20 sec. After that the message will be deleted (immediatereturn to measurement at pressing enter).

0000*)

2222*)

0000*)

*) Factory setting

Page 68: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

68 Transmitter Cond 7100 e/2(X)H

Specify output currentfor testing the connected peripherals• Press conf, enter passcode 5555*)

The output current indicated in the main displaycan be modified.Select with 4 key, edit number with 5 key, proceed with enter.The actually measured current is shown in thesecondary display. The Transmitter is in Hold mode.Press conf, then enter to return to measurement(Hold remains active for another 20 sec).

Diagnostics functions

Entry/display Action / Remarks

Cleaning

To remove dust, dirt and spots, the external surfaces of thedevice may be wiped with a damp, lint-free cloth. A mildhousehold cleaner may also be used if necessary.

*) Factory setting

5555*)

Page 69: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Operating state

69

Measurement

Operating states

Cal Info(cal) 0000

20 s

Error Info(conf) 0000

20 s

Calibration(cal) 1100

Temp adjustment(cal) 1015

Product cal 1(cal) 1105

Product cal 2(cal) 1105

Configuration(conf) 1200

20 min

Sensor monitor(conf) 2222

20 min

Current source(conf) 5555

20 min

LED

Tim

e ou

t

Out

Explanation: active

as configured (Last/Fix or Last/Off)

LED flashes during HOLD (configurable)

Page 70: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

70 Transmitter Cond 7100 e/2(X)H

Red

LED

Out

1 (

22 m

A)

Error messages (error codes)

Sensor• Wrong cell constant• Measurement range violation• SAL > 45 ‰• Sensor connection or cable defective• USP limit exceeded

Error DisplayProblemPossible causes

ERR 01 Measuredvalueflashes

xx

Unsuitable sensorConductance range > 3500 mS

ERR 02 Measuredvalueflashes

xx

System errorConfiguration or calibration datadefective. Completely reconfigureand recalibrate the device.Memory error in device program

ERR 98 “Conf”flashes

xx

Factory settingsEEPROM or RAM defectiveThis error message only occurs inthe case of a total defect. TheTransmitter must be repaired andrecalibrated at the factory.

ERR 99 “FAIL”flashes

xx

Temperature probeOpen or short circuitTemperature range exceeded

ERR 03 xx

Page 71: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

71

Sensocheck:Wrong or defective sensor /Polarization effects at the sensor /cable too long or defective /plug defective

Red

LED

Out

1 (2

2 m

A)

Current outputCurrent below 3.8 mA

ErrorSymbol(flashes)

ProblemPossible causes

ERR 11 xx

Current outputCurrent above 20.5 mA

ERR 12 xx

Current outputCurrent span too small / too large

ERR 13 xx

ERR 33 xx

Temperature outside conversiontables (TC, Conc, SAL)

independent ofSensoface

Sensoface activesee Pg 73

Page 72: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

72 Transmitter Cond 7100 e/2(X)H

SensocheckContinuously monitors the sensor and its wiring. Sensocheck can be switched off. Critical values make theSensoface “sad” and the corresponding icon flashes:

The Sensocheck message is also output as error messageErr 33. The red LED is lighted, the output current is set to22 mA (when configured correspondingly). Sensocheck canbe switched off during configuration (then Sensoface is alsodisabled). Exception: After a calibration a Smiley is alwaysdisplayed for confirmation.

Sensoface(Sensocheck must have been activated during configuration.)

The little smiley in the display (Sensoface) provides informationabout the sensor condition (defects, maintenance required,cable capacitance too high). It alerts to significant sensor polarization or excessive cablecapacitance e.g. caused by an unsuitable cable or a cable thatis too long. The permitted calibration ranges and the condi-tions for a friendly, neutral, or sad Sensoface are summarized inthe following chart. Additional icons refer to the error cause.

Note:The worsening of a Sensoface criterion leads to the devaluationof the Sensoface indicator (Smiley becomes “sad”). To reset the Sensoface indicator, the defect must be remediedand the Transmitter be calibrated.

Page 73: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

73

StatusDisplay Problem

Sensor defect Wrong or defective sensorSignificant polarization of sensorExcessive cable capacitance (also seeerror message Err 33, Pg 71).

Temperature outside range for TC,conc, SAL

Temperatureerror

Note:When very fast response times (t90) are required, e.g. whendetecting separation layers, Sensocheck should be switchedoff (see “Specifications” Pg 76).

Page 74: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

74 Transmitter Cond 7100 e/2(X)H

Page 75: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

75

Appendix

Product line and accessories

Devices Order No.

Transmitter Cond 7100 e/2H 52 120 903Transmitter Cond 7100 e/2XH 52 120 905

Mounting accessoriesPipe-mount kit 52 120 741Panel-mount kit 52 120 740Protective hood 52 120 739

SensorsMettler-Toledo GmbH, Process Analytics offers a wide range of 2-electrode and 4-electrode sensors for the following fields ofapplications:- Chemical process industry- Pharmaceutical industry- Food and beverage industry- Water/waste-water

For more information concerning our sensors and housingsprogram, please refer to http://www.mt.com.

Page 76: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

76 Transmitter Cond 7100 e/2(X)H

Conductivity input Input for 2-electrode/4-electrode sensorsEffective range Conductivity 4-EL 0.2 µS · c ... 1000 mS · c

Conductivity 2-EL 0.2 µS · c ... 200 mS · cRanges * Conductivity 0,000 ... 9.999 µS/cm

00.00 ... 99.99 µS/cm000.0 ... 999.9 µS/cm0000 ... 9999 µS/cm0.000 ... 9.999 mS/cm00.00 ... 99.99 mS/cm000.0 ... 999.9 mS/cm0.000 ... 9.999 S/m00.00 ... 99.99 S/m

Resistivity 00.00 ... 99.99 MΩ·cmConcentration 0.00 ... 9.99 % by wtSalinity 0.0 ... 45 ‰ (0 ... 35 °C)USP 00.00 ... 99.99 µS/cm

Measurement error 1,2,3) < 1 % meas. val. +0.4 µS · c

Concentration determinationOperating modes: * -01- NaCl 0,00 ... 9.99 % by wt (0 ... 100 °C)

-02- HCl 0,00 ... 9.99 % by wt (0 ... 50 °C)-03- NaOH 0,00 ... 9.99 % by wt (0 ... 100 °C)-04- H2SO4 0.00 ... 9.99 % by wt (0 ... 110 °C)-05- HNO3 0.00 ... 9.99 % by wt (0 ... 50 °C)See graphs in the Appendix Pg 84 and following

Sensor standardizationOperating modes • Input of cell constant with simultaneous

display of conductivity and temperature • Input of onductivity of calibration solution with

simultaneous display of cell constant and temperature

• Product calibration• Temperature probe adjustment

Adm. cell constant 00.0050 ... 19.9999 cm-1

Specifications

Page 77: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

77

Sensor monitoringSensocheck Polarization detection and monitoring of

cable capacitance

Sensoface Provides information on the sensor condition(Sensocheck)

Sensor monitor Direct display of measured values from sensor for validation (resistance / temperature)

USP function Water monitoring in the pharmaceutical industry (USP)

Temperature input * Pt100 / Pt1000/ NTC 30 kΩ /NTC 8.55 kΩ (Betatherm)2-wire connection, adjustable

Ranges Pt100/Pt1000: -20 .. +200 °C(-4 ... 392 °F)

NTC 30 kΩ -20 ... +150 °C(-4 ... 302 °F)

NTC 8.55 kΩ -10 ... +130 °C(+14 ... 266 °F)

Resolution 0.1 °C / 1 °FMeasurement error 1,2,3) 0.5 K

(< 1K for Pt100; < 1K for NTC > 100°C)Temperature compensation *(Reference temp 25 °C)

(OFF) none(Lin) Linear characteristic 00.00 ... 19.99 %/K(NLF) Natural waters to EN 27888(nACL) Ultrapure water with NaCl traces (0...120°C)(HCL) Ultrapure water with HCl traces (0...120°C)(nH3) Ultrapure water with NH3 traces (0...120°C)

Page 78: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H

Loop current 4 ... 20 mA floating Supply voltage 14 ... 30 VMeasured variable * Conductivity, resistivity, concentration,

or salinityCharacteristic Linear or logarithmicOverrange * 22 mA in the case of error messagesOutput filter * Low-pass, filter time constant 0 ... 120 sec

Measurement error 1)

< 0.3 % current value + 0.05 mAStart/end of scale As desired within rangeMin. span LIN: 5 % of selected range

LOG: 1 decadeCurrent source function 3.8 mA ... 22 mA

78

Specifications

HART communication Digital communication by FSK modulation of loop current, reading of device identification,measured values, status and messages, reading and writing of parameters, start of product calibration, signaling of configuration changes according to FDA 21 CFR Part 11

Display LC display, 7-segment with iconsMain display Character height 17 mm, unit symbols 10 mmSecondary display Character height 10 mm, unit symbols 7 mmSensoface 3 status indicators (friendly, neutral, sad Smiley)

Status indication 4 mode indicators “meas”, “cal”, “alarm”, “config”18 further icons for configuration and messages

Alarm indication Red LED in case of alarm or HOLD, user defined

Keypad 5 keys: [cal] [conf] [4] [5] [enter]

* User-defined1) To IEC 746 Part 1, at nominal operating conditions2) ± 1 count3) Plus sensor error

Page 79: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

79

Service functionsCurrent source Loop current specifiable 3.8 ... 22.00 mADevice self-test Automatic memory test (RAM, FLASH, EEPROM)Display test Display of all segmentsLast Error Display of last error occurredSensor monitor Display of direct, uncorrected sensor signal

(resistance/temperature)Passcodes Modifiable according to FDA 21 CFR Part 11

“Electronic Signatures”

Data retention Parameters and calibration data > 10 years (EEPROM)

EMC EN 61326Emitted interference: Class B (residential area)

Class A Immunity to interference: Industry

Explosion protection7100 e/2XH: ATEX TÜV 99 ATEX 1433

II 2(1) G EEx ib [ia] IIC T6FM FMRC J.I. 300580

IS/I/1/ABCD/T4NI/I/2/ABCD/T4

CSA 1662790Cl I, Div 1, Gr ABC & D T4; Ex ib [ia] IIC T4Cl I, Div 2, Gr ABC & D T4; Ex nAL[L] IIC T4

7100 e/2H: FM FM 300580NI/I/2/ABCD/T4

Nominal operating conditionsAmbient temperature –20 ... +55 °CTransport/Storage temp –20 ... +70 °CSupply voltage 14... 30 V

Page 80: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

80 Transmitter Cond 7100 e/2(X)H

Enclosure Molded enclosure made of PBT(polybutylene terephtalate)

Color Bluish gray RAL 7031Assembly • Wall mounting

• Pipe mounting: Ø 40 ... 60 mm, 30 ... 45 mm

• Panel mounting, cutout to DIN 43 700Sealed against panel

Dimensions H 144 mm, B 144 mm, T 105 mmIngress protection IP 65/NEMA 4X

(USA, Canada: indoor use only)Cable glands 3 breakthroughs for cable glands

M20x1.5, 2 breakthroughs for NPT 1/2” or Rigid Metallic Conduit

Weight Approx. 1 kg

Specifications

Page 81: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

81

Page 82: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H82

Calibration solutionsPotassium chloride solutions(Conductivity in mS/cm)

Temperature Concentration *

[°C] 0.01 mol/l 0.1 mol/l 1 mol/l

051015161718192021222324252627282930313233343536

0.7760.8961.0201.1471.1731.1991.2251.2511.2781.3051.3321.3591.3861.4131.4411.4681.4961.5241.5521.5811.6091.6381.6671.696

7.158.229.3310.4810.7210.9511.1911.4311.6711.9112.1512.3912.6412.8813.1313.3713.6213.8714.1214.3714.6214.8815.1315.3915.64

65.4174.1483.1992.5294.4196.3198.22100.14102.07104.00105.94107.89109.84111.80113.77115.74

*) Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlenwerteund Funktionen ..., volume 2, part. volume 6

Page 83: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

*) Data source: Test solutions calculated according to DIN IEC 746-3**) Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein:Zahlenwerte

und Funktionen ..., volume 2, part. volume 6

Sodium chloride solutions(Conductivity in mS/cm)

0123456789

101112131415161718192021222324252627282930313233343536

134.5138.6142.7146.9151.2155.5159.9164.3168.8173.4177.9182.6187.2191.9196.7201.5206.3211.2216.1221.0226.0231.0236.1241.1246.2251.3256.5261.6266.9272.1277.4282.7288.0293.3298.7304.1309.5

5.7865.9656.1456.3276.5106.6956.8817.0687.2577.4477.6387.8318.0258.2218.4188.6178.8169.0189.2219.4259.6319.838

10.04710.25810.46910.68310.89811.11411.33211.55211.77311.99512.22012.44512.67312.90213.132

0.6310.6510.6710.6920.7120.7330.7540.7750.7960.8180.8390.8610.8830.9050.9270.9500.9720.9951.0181.0411.0641.0871.1111.1351.1591.1831.2071.2321.2561.2811.3061.3311.3571.3821.4081.4341.460

Temperature Concentration

[°C] 0.01 mol/l * 0.1 mol/l * saturated **

83

Page 84: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

84 Transmitter Cond 7100 e/2(X)H

400

300

200

100

00 5 10

60

46

35

20

10

0

c [Gew%]

χ [mS/cm] [°C]

82

100

Concentration curves

Conductivity in dependence on substance concentration andprocess temperature for sodium chloride solution (NaCl)

-01- Sodium chloride solution NaCl

c [% by wt]

Page 85: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

85

Conductivity in dependence on substance concentration andprocess temperature for hydrochloric acid (HCl)Source: Haase/Sauermann/Dücker; Z. phys. Chem. New Edition,Vol. 47 (1965)

1000

800

600

400

200

00 5 10

50

40

302520

10

0

c [Gew%]

χ [mS/cm] [°C]

-02- Hydrochloric acid HCl

c [% by wt]

Page 86: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Transmitter Cond 7100 e/2(X)H

Conductivity in dependence on substance concentration andprocess temperature for sodium hydroxide solution (NaOH)

1000

800

600

400

200

00 10c [Gew%]

χ [mS/cm] [°C]

50

40

302520100

60

70

80

90

100

-03- Sodium hydroxide solution NaOH

86

Concentration curves

c [% by wt]

Page 87: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

87

600

400

200

00 5 10

71,160,048,937,8

26,7

15,6

4,4

c [Gew%]

χ [mS/cm] [°C]

82,293,3

-04- Sulphuric acid H2SO4

Conductivity in dependence on substance concentration andprocess temperature for sulfuric acid (H2SO4)Source: Darling; Journal of Chemical and Engineering Data;Vol.9 No.3, July 1964

c [% by wt]

Page 88: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

88 Transmitter Cond 7100 e/2(X)H

800

600

400

200

00 5 10

50

40

3020

10

0

c [Gew%]

χ [mS/cm] [°C]

-05- Nitric acid HNO3

Conductivity in dependence on substance concentration andprocess temperature for nitric acid (HN03) Source: Haase/Sauermann/Dücker; Z. phys. Chem. New Edition, Vol. 47(1965)

Concentration curves

c [% by wt]

Page 89: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

89

Page 90: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

FM Control Drawing

90 Transmitter Cond 7100 e/2(X)H

Page 91: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

91

Page 92: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

92 Transmitter Cond 7100 e/2(X)H

Explosion protection

Page 93: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

93

Warnings and notes to ensure safe operation

Warning: Do not disconnect equipment unless power hasbeen switched off.

Warning: Clean only with antistatic moistened cloth.Warning: Substitution of components may impair suitability

for hazardous locations.

• The equipment shall be installed and protected from mechanicalimpact and ultraviolet (UV) sources.

• Clean only with a moistened antistatic cloth as potential electrostatic hazard may exist. Service equipment only with conductive clothing, footwear and personal grounding devicesto prevent electrostatic accumulation.

• Internal grounding provisions shall be provided for field wiring. Bonding between conduit shall be provided during installation, and all exposed non-current carrying metallic parts shall be bonded and grounded.

• Installation in a Class I, Division 2 or Class I, Zone 2 hazardouslocation shall be in accordance with the Canadian ElectricalCode (CEC Part 1) Section 18 Division 2 wiring methods.

OBSERVE THE SPECIFICATIONS OF THE CONTROLDRAWING!

Page 94: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

94 Transmitter Cond 7100 e/2(X)H

CSA Control Drawing

Conductivity Measuring Loop

Terminals1, 2, 3, 4, 5, 6

Uo, Vsc Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 10V 143mA 357mW 3µF 1.3mHIIB (GRP C) 10V 143mA 357mW 9µF 5mHIIC (GRP D) 10V 143mA 357mW 24µF 10mH

Temperature Measuring Loop

Terminals7, 8

Uo, Vsc Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 5V 3mA 4mW 100µF 1HIIB (GRP C) 5V 3mA 4mW 300µF 1HIIC (GRP D) 5V 3mA 4mW 800µF 1H

All Combined Outputs

Terminals1, 2, 3, 4, 5, 6, 7, 8

Uo, Vsc Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 10V 146mA 365mW 3µF 1.3mHIIB (GRP C) 10V 146mA 365mW 9µF 5mHIIC (GRP D) 10V 146mA 365mW 24µF 10mH

Page 95: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

95

Conductivity Measuring Loop

Terminals1, 2, 3, 4, 5, 6

Uo, Vsc Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 10V 143mA 357mW 3µF 1.3mHIIB (GRP C) 10V 143mA 357mW 9µF 5mHIIC (GRP D) 10V 143mA 357mW 24µF 10mH

Temperature Measuring Loop

Terminals7, 8

Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 5V 3mA 4mW 100µF 1HIIB (GRP C) 5V 3mA 4mW 300µF 1HIIC (GRP D) 5V 3mA 4mW 800µF 1H

All Combined Outputs

Terminals1, 2, 3, 4, 5, 6, 7, 8

Uo, Vsc Io, Isc Po Co, Ca Lo, La

IIC (GRP A, B) 10V 146mA 365mW 3µF 1.3mHIIB (GRP C) 10V 146mA 365mW 9µF 5mHIIC (GRP D) 10V 146mA 365mW 24µF 10mH

Page 96: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

96 Transmitter Cond 7100 e/2(X)H

Page 97: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

97

Glossary

Conductivity Conductivity χ [S/cm] = G [S] · c [1/cm]

Conductivitysensor

Either 2- or 4-electrode sensors can be con-nected. The cell constant of the sensor inuse must be entered or be determined usinga calibration solution taking account of thetemperature. A special device variant (Transmitter CondInd 7100 e/2(X)H) is provided for electrode-less sensors.

Conductance G [S] = 1 / R [Ω]Conductance

Temperaturecoefficient

With temperature compensation activated,the measured value is calculated to the valueat the reference temperature (25 °C) usingthe temperature coefficient.

Temperaturecompensation

Calculates the measured conductivity valuefor a reference temperature.

Page 98: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

98 Transmitter Cond 7100 e/2(X)H

Index

22 mA signal for error message . . . . . . . . . . . . . . . . . . . . 51

AAlarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31, 54Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Audit Trail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Automatic device self-test . . . . . . . . . . . . . . . . . . . . . . . . . 29

CCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

by entry of cell constant . . . . . . . . . . . . . . . . . . . . . . . . 60Calibration solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . 82Display of calibration data . . . . . . . . . . . . . . . . . . . . . . 67Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64Temp probe adjustment . . . . . . . . . . . . . . . . . . . . . . . . 66with calibration solution . . . . . . . . . . . . . . . . . . . . . . . . 62

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Concentration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41, 43

Process solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Concentration curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

-01- Sodium chloride solution NaCl . . . . . . . . . . . . . . . 84-02- Hydrochloric acid HCl . . . . . . . . . . . . . . . . . . . . . . 85-03- Sodium hydroxide solution NaOH . . . . . . . . . . . . . 86-04- Sulphuric acid H2SO4 . . . . . . . . . . . . . . . . . . . . . . 87-05- Nitric acid HNO3 . . . . . . . . . . . . . . . . . . . . . . . . . . 88

Conductivity sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Sensor monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Page 99: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Individual settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37LIN characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44LOG characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Output current during Error and HOLD . . . . . . . . . . . . . 50Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Process solutions for concentration . . . . . . . . . . . . . . . 42Sensor type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Temperature compensation . . . . . . . . . . . . . . . . . . . . . 52Time constant of output filter . . . . . . . . . . . . . . . . . . . 48

Conformity with FDA 21 CFR Part 11 . . . . . . . . . . . . . . . . . 9Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6, 18, 22

via VP cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Control / Installation Drawing . . . . . . . . . . . . . . . . . . . 90, 94CSA Control Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Display of output currents . . . . . . . . . . . . . . . . . . . . . . 67

DDiagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67, 68Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

EEC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . 8EC-Type-Examination Certificate . . . . . . . . . . . . . . . . . . . . 10Electronic Signature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

99

Page 100: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Display of last error message . . . . . . . . . . . . . . . . . . . . 67

Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Cleaning in a hazardous location . . . . . . . . . . . . . . . . . . 6Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

FFM Control Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

GGainCheck device self test . . . . . . . . . . . . . . . . . . . . . . . . 29Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

HHART communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

LED in HOLD mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Output signal during HOLD . . . . . . . . . . . . . . . . . . . . . 51

IInstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

KKeypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

LLoop current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31, 78

MMeasurement range over 3 decades . . . . . . . . . . . . . . . . . 47Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

100 Transmitter Cond 7100 e/2(X)H

Index

Page 101: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

OOperating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69Output filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

PPacking list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Panel-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Passcodes according to FDA 21 CFR Part 11 . . . . . . . . . . . 56Pipe mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Pipe-mount kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6, 18Product line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75Protective hood. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

RReturn of products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

SSafety functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29, 30Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Salinity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Sensocheck. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29, 72

ON / OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29, 72Sensor monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

TTemperature compensation. . . . . . . . . . . . . . . . . . . . . . . . 52Temperature probe adjustment . . . . . . . . . . . . . . . . . . . . . 66

101

Page 102: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6, 18Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

UUser interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26USP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

WWarranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Wiring examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

with 2-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . 23with 4-electrode sensor . . . . . . . . . . . . . . . . . . . . . . . . 22with Mettler-Toledo 2-electrode sensor via VP plug . . . 24with Mettler-Toledo 4-electrode sensor via VP plug . . . 25

102 Transmitter Cond 7100 e/2(X)H

Index

Page 103: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical
Page 104: Instruction manual...3 4 5 Fig.: Pipe-mount kit 1 Protective hood (if required) 2 Hose clamps with worm gear drive to DIN 3017 (2 pieces) 3 Pipe-mount plate (1 piece) 4 For vertical

TA-194.200-MTE03 240605 Software version: 4.0