instalacion duramax

12
Technical Services Installation Instructions DuraMax Measuring Center

Upload: gabijona

Post on 03-Jan-2016

77 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Instalacion Duramax

Technical Services Installation Instructions

DuraMax

Measuring Center

Page 2: Instalacion Duramax
Page 3: Instalacion Duramax

1. Machine components, installation site and installation conditionsWhen selecting the installation site, it is essential for you to ensure that the requirements specified below either alreadyexist or will be created by you.

1.1 Components

The system comprises the following components:

1.2 Room height

This dimension equals the sum of the height of the CMM and the minimum installation clearance below the ceiling.

If your DuraMax is delivered without a base, the height of the machine table or measuring plate on which your DuraMaxis mounted must be used to calculate the required room height.

Your machine table or measuring plate must be designed for a weight load of ≥ 500 kg.

1.3 Floor space required

The dimensions are listed on Page 7.

If you draw up a floor plan on a CAD system, you can obtain a DXF and/or a DWG file with the dimensions. Please consultwith your contact persons at Carl Zeiss if necessary.

When drawing up a floor plan, please make sure that your coordinate measuring machine is readily accessible from allsides. Installation near transport lanes should be avoided at all costs.

Select the installation site so that the distance between the CMM and the wall or other machines is at least 500 mm.

1.4 Transport lanes

Check the floor and ceiling loading capacity of transport lanes leading to the installation site. The doorways and transportlanes must be at least 10 cm wider and higher that the CMM assemblies including the transport vehicles. The dimensionsand weights of the CMM assemblies are listed in this brochure.

1.5 Ambient conditions

Coordinate measuring machines are quality high-tech products. However, their specified quality can be attained only ifthe following points are observed.

Permissible storage temperature range

5°C to 40°C (only when left unopened in their original packaging)Avoid storage at temperatures below 5oC, especially in winter.

• CMM with controller • Control panel • Data station • Optional: base

Height of DuraMax with adjusting feet ....................................................... 1810 mm

Height of base (option)................................................................................. 670 mm

Minimum installation clearance .................................................................... 200 mm

Required minimum room height above floor or foundation .............. 2680 mm

Note:

Installation Instructions Page

3

Page 4: Instalacion Duramax

Ambient conditions for operational readiness

Permissible ambient temperature: +15°C to +40°C

Permissible air humidity: 40% - 70% without condensation

Altitude (above sea level): - at 100 - 125 V max. 3000 m- at 230 - 240 V max. 2000 m

Temperature conditions for operating the CMM:

The measuring uncertainty specified for the DuraMax is guaranteed under the following conditions:

1. Following installation, the CMM rests for 1 day so that it can adapt to room temperature.

2. A measurement correction program featuring input of the workpiece and CMM temperatures and the temperatureexpansion coefficient of the workpiece material is used.

3. The following temperature conditions are met:

Drafts caused by frequent opening and closing of doors should be avoided in the immediate vicinity.

Direct heat radiation of any type whatsoever also must be excluded.

1.6 Vibrations at the installation site

Vibrations caused by heavy machining centers, transport equipment (e.g. cranes and forklifts), presses and forging ma-chines located in adjacent rooms may impair the measuring accuracy of your CMM.

The DuraMax is equipped with an integrated damping system to prevent such impairments wherever possible.

If you wish, we can perform a vibration analysis for you to determine whether your ambient frequency and amplitude(acceleration) conditions are within tolerance. Please contact your nearest Carl Zeiss representative for more information.

1.7 Floor load

We recommend consulting a structural engineer who is familiar with your on-site subsoil and floor conditions for any floorload calculations and when determining whether reinforcement of the floor (foundation etc.) is required. The load exertedon the floor by the CMM is specified on Page 7 of this brochure.

CMM type

Temp. range(ºC)

Temperature gradient

per hour (ºK/h) per day (ºK/d) spatial (ºK/m)

DuraMax 18 - 30 2.0 5.0 1.0

Note:

Installation InstructionsPage

4

Page 5: Instalacion Duramax

2. Preparations for installation / Electric power supplyPlease make sure that the following preparations are made by the date of installation:

2.1 Preparation of floor surface

The floor surface must be absolutely clean.

Maximum difference in height between the 3 supporting surfaces: ≤ 5 mm

Max. incline per supporting surface (100x100mm): 1mm

The system requires two shock-proof wall outlets (max. 16A) and one 5-fold outlet strip (to be provided by the customer).The power supply of the DuraMax is provided by one of the wall outlets. The second wall outlet is required for connectingthe outlet strip to supply the peripheral devices with power. We recommend using a lightning arrester, e.g. Phoenix Flash-trab FLT 25-400, as a coarse protection against overvoltage at the installation end.

We recommend making the connection according to the following sketch. At the same time, it is also mandatory toobserve the relevant national regulations.

2.2 Electrical power supply parameters

To ensure trouble-free data exchange between the DuraMax and the data station, the power supply must have the fol-lowing parameters.

Line power supply:

– Standard version:1/N/PE 100, 110, 120, 125, 230, 240V AC

– Voltage fluctuations: ±10%

– Frequency: 50 - 60 Hz

– Frequency fluctuations: ±3.5%

Line fuse protection

– For 230V or 240V: C 16A fuse

– For 100V to 125V: 15A fuse

Total power consumption:

– DuraMax: max. 700VA

– All peripherals: max. 1000VA

L

N

PE

Recommended connection (Customer installation (national voltage)

C16 A

Connection diagram

CMMSafetywall

5-fold safetysocket strip

Peripheralsmax. 1000 VA

Groundfault

circuitinterrupter

Safetywall

Connecting cablemin. 5 m

outlet

outlet

Installation Instructions Page

5

Page 6: Instalacion Duramax

2.3 EMC

EN 61326-1 Appendix A, Class A

Warning !

This is a class A device, and may therefore cause radio interference in the surrounding area. In such cases, the operatormay be required to take appropriate action and bear any costs thus incurred.

2.4 Stabilizing the line voltage

If out-of-tolerance r.m.s. fluctuations of the nominal voltage briefly occur, a line analysis must be performed. Please con-tact us if you have such a requirement. We will be glad to advise you and suggest suitable measures. In most cases, it issufficient to install a separate stub cable from the main distributor to the CMM.

2.5 Machine parameters

– Overvoltage category: ΙΙΙ

– Pollution degree: 2

– Protection class: 1

2.6 Teleservice

You will require an Internet connection for Teleservice. For more details, please refer to the "Teleservice" brochure.

2.7 Network topology

The networking of the system components is performed by the ZEISS service engineer according to the valid specificationsduring initial startup. These settings must not be modified by the customer. Any change could result in connection prob-lems.

2.8 Computer operating system

If the required computers are provided by the customer, the operating system used must not show any limitations in com-parison to the one used by ZEISS. Otherwise problems and additional costs may occur during startup.

2.9 Packaging/Transport vehicles

Disposable packaging: H= 2160 mm, W= 1270 mm, D= 1140 mm, total weight = approx. 450 kg (with DuraMax)

Please provide a forklift with a sufficient load carrying capacity, a fork length of 1000mm and an adjustable fork widthof 550mm (outer dimension) for transporting the packaged units to the storage and/or installation site. The required loadcarrying capacity of the transport vehicle is specified on the following pages. Be sure to keep all transport lanes free duringinstallation and startup to ensure smooth processing.

Packaging materials such as e.g. disposable packaging materials, plastic sheeting or "popcorn" (styrofoam chips)can be reused by Zeiss, provided that they are clean and no dirt is mixed in with them. Please hand over suchmaterials to the forwarding agent for return to their point of origin.

2.10 Moving the CMM to another location

To prevent damage to the CMM, it may be relocated only when properly secured (e.g. with transport braces)and under the supervision of a qualified system specialist.

Note:

Note:

Installation InstructionsPage

6

Page 7: Instalacion Duramax

905

2480

7321090

9101140

1330

➀ ➃➂

min. 500

min. 500

min. 200

= 2x safety wall sockets, 1x CMM, 1x socket strip(to be installed by customer/national voltage).

= 5-fold socket strip for peripherals.

= Internet connection for teleservice

= Network connection

= Cable: network connection, controller

= Cable: DuraMax computer station

= Cable: peripherals (e.g. computer, printer monitor)

= Control panel tray (removable)

= Side table (option) for peripherals

Net weight of CMM with base: 480 kg

Max. permissible workpiece weight: 100 kg

Average pressure on the 3 adjusting feet: 19.4 N/cm2

Calculated from the net weight of the machine with base and the max.workpiece weight distributed over the footprint of the three CMM feetmeasuring 100 cm2 ea.

Note:

For reasons of stability, the base must be fastened securelyto the floor (foundation) with screws and wall plugs. Twofastening holes are provided on the base for this purpose.The fastening materials must be provided by the customer.

3. Installation dimensions (with optional base) and sample layout (scale 1:30)

Installation Instructions Page

7

Page 8: Instalacion Duramax

60

1803

6701080

8351127

1330

230➊

3.1 Installation dimensions (without optional base/scale 1:30)

• An appropriate machine table or measuring plate is required to install the DuraMax ( ). It should have the recom-mended minimum table height of 700mm and a load carrying capacity of ≥ 500 kg.

• The required room height equals the sum of the following dimensions:- Height of machine table or measuring plate ( provided by the customer)

- Height of DuraMax without base: 1810 mm

- Required installation clearance below ceiling: 200 mm

• The average pressure exerted by the adjusting feet on the foundation is calculated from:- the net weight of CMM: 400 kg,

- the permissible max. workpiece weight: 100 kg

- the weight of machine table or measuring plate ( provided by the customer)

and

- the footprint of the adjusting feet of the machine table or measuring plate.

• For information on the distance from the wall, required connections, etc. please refer to the drawing on Page 7.

Installation InstructionsPage

8

Page 9: Instalacion Duramax

� 494

0

100±0,1100±0,1

100±0,1

100±0,1

200±0,1200±0,1

200±0,1

200±0,1

M10/20 tiefM10/20 deep(25X)

4. Measuring table with fastening holes

Installation Instructions Page

9

Page 10: Instalacion Duramax

Installation InstructionsPage

10

Page 11: Instalacion Duramax
Page 12: Instalacion Duramax

Carl Zeiss 73446 Oberkochen

600061-1123.020 Printed in Germany I-SD / 03.08