Page 1: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

Infant Resuscitation | Technical Manual

|

Page 2: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 1

TABLE OF CONTENTS

1. Specifications .....................................................................................................................................................................2

1.1 Warnings ..........................................................................................................................................................................2

1.2 PerformanceSpecifications ..............................................................................................................................................2

1.3 TechnicalSpecifications ...................................................................................................................................................2

1.4 SymbolDefinitions ...........................................................................................................................................................3

1.5 LabelIdentification ..........................................................................................................................................................3

2. Cleaning .............................................................................................................................................................................4

2.1 CleaningProductstoAvoid ...............................................................................................................................................4

2.2 CleaningoftheTestLung ..................................................................................................................................................4

2.2.1 Latex(black)TestLung(discontinuedMay2007) .................................................................................................4

2.2.2 Silicone(blue)TestLungRD020-01 .......................................................................................................................5

2.3 CleaningReusableT-Piece(discontinuedSeptember2008) ..............................................................................................5

2.4 Sterilization ......................................................................................................................................................................5

2.5 AfterCleaning ...................................................................................................................................................................5

3. Service Information .............................................................................................................................................................6

3.1 FunctionalSchematic .......................................................................................................................................................6

3.2 InstallationChecksandPreventativeMaintenance ...........................................................................................................6

3.2.1 InstallationChecks ................................................................................................................................................6

3.2.2PreventativeMaintenance ......................................................................................................................................6

3.3 TestingoftheF&PNeopuff™/Perivent™Performance(ManometerandValveSystem) .....................................................6

3.3.1 TestingoftheManometer ......................................................................................................................................7

3.3.2 TestingoftheValveSystem ...................................................................................................................................7

3.3.3 SettingMaxPressureReliefto40 cmH2O ...............................................................................................................7

3.3.4 ResettingtheManometertoZero ...........................................................................................................................7

3.3.5 ManometerReplacement ......................................................................................................................................7

3.3.6 ValveSystemReplacement ....................................................................................................................................7

3.4 Set-upandMaintenanceChecklist ...................................................................................................................................8

4. Assembly Diagrams .............................................................................................................................................................9

4.1 F&PNeopuff™/Perivent™InfantResuscitator(PostJune2010) ........................................................................................9

4.1.1 AssemblyDiagram ................................................................................................................................................9

4.1.2 PartsList ...............................................................................................................................................................9

4.1.3 SparesKit ...........................................................................................................................................................10

5. Mounting Options .............................................................................................................................................................11

5.1 900RD301Side-MountingBlock .....................................................................................................................................11

5.2 900MR088RailMountBracket .......................................................................................................................................11

5.3 900RD302 WallMountBracket .......................................................................................................................................11

5.4 RD050-01PoleandRailCentralMount ...........................................................................................................................12

Page 3: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 2

1.1 Warnings

1.3 Technical Specifications

Peak Inspiratory Pressure (PIP) Range

@ 5 L/min approx. 2 to 70 cmH2O [mbar]

@ 8 L/min approx.3 to 72 cmH2O [mbar]

@ 10 L/min approx.4 to 73 cmH2O [mbar]

@ 15 L/min approx.8 to 75 cmH2O [mbar]

Positive End Expiratory Pressure (PEEP) Range

@ 5 L/min approx.1 to 6 cmH2O [mbar]

@ 8 L/min approx.1 to 10 cmH2O [mbar]

@ 10 L/min approx. 2 to 15 cmH2O [mbar]

@ 15 L/min approx.4 to 17 cmH2O [mbar]

Input Gas Flow Range

Minimum 5L/min

Maximum 15L/min

Operating Time (400 L Cylinder)

5 L/min 80minutes

10 L/min 40minutes

15 L/min 26minutes

NOTE:Allperformancefigureslistedabovearerepresentativeonly. PEEPvaluesstatedarebasedontypicalclinicalPIPsettings. HigherPEEPvaluescanbeachievedifhigherPIPvaluesareset.

Height 250mm(9.8”)

Width 200mm(7.9”)

Depth 104mm(4.1”)

Weight 1.9kg(4.2lb)

Manometer Range -10to80 cmH2O [mbar]

Manometer Accuracy +/-2.0%ofFullScaleDeflection

Maximum Pressure Setting 65 to 80 cmH2O [mbar]

(dependentonflowrate)

900RD010 / 900RD110 Dead Space 6 ml

RD1300-10 Dead Space 3.3 ml

Storage Temperature Range -10to50°C(+14to+122°F),upto95%humidity

Operating Temperature Ranges

Humidified Circuit

Non-Humidified Circuit

+18to26°C(+64to+78°F),30 - 75%humidity

-18to50°C(-0.4to+122°F),upto95%humidity

Recommended Patient Body Weight

0 to 10kg(22lb)

Delivered Oxygen Concentration Upto100% depending on gas supply

Inputgasflowrate5 to15L/min.Recommended operating gas

flow rate 8 L/min.Inputflowrangesarecircuitspecific,referto

circuitUserInstructions.

Do not attempt to use a higher gas flow rate than 15 L/min.

Factorysettingofpressurelimitvalveisat40 cmH2O.Thissetting

isuseradjustableupto80 cmH2O. Do not attempt to set the

pressure relief valve above 80 cmH2O. Reset pressure relief valve

to 40 cmH2O after use (refer section 3.3.3 of this manual).

UseonlyrecommendedF&PNeopuff™/Perivent™Infant

Resuscitatoraccessories.

Use only a F&P gas supply line or approved equivalent.

Donotuseoil,greaseorothersubstancesthatareincompatible

withoxygenonanypartoftheF&PNeopuff/Perivent

T-PieceCircuit.

Incorrectusecanbehazardous.

Thedevicemustnotbeusedonunattendedpatients.

Tobeusedforresuscitationonly.

The operation of the F&P Neopuff / Perivent must be checked prior

to first use.

Refertosection 3.3“TestingoftheF&PNeopuff/Perivent

Performance”.

1.2 Performance Specifications

1. SPECIFICATIONS

Page 4: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 3

SYMBOL DEFINITION

Attention:ConsulttheOperatingInstructions

MAX-P SetstheMax Pressure Relief thatmay bedeliveredtothepatient (factorysetat40 cmH

2O)

PIP ControlsthePeak Inspiratory Pressure deliveredtothepatient

5-15 L/min Gas inletconnectionfromgassupply (5 to 15litresperminute)

Gas outletconnectiontopatient

1.4 Symbol Definitions

1. SPECIFICATIONS continued

1.5 Label Identification

REF - Thisdetailstheproductcodewhichidentifiestheoperating

instructionlanguageandfasciatype

LOT - Theproductiondate:(10) = 2010, (06) =June,

(15) =fifteenthday

SN- TheserialnumberisspecifictoaparticularF&PNeopuff/

PeriventInfantResuscitator.ThefirstsixdigitsmatchtheLOT

numberandlastsixdigitsidentifythespecificproductserial

number.

Thebarcodeconsistsofthefollowingpredefinedapplicationidentifiers:

01 -Notused

0-Sequencestart

94-Countryofmanufacture-NewZealand

200124-CompanyIdentifier-Fisher&PaykelHealthcareLtd

1093-ProductCode

1-CheckSum

REF - RD900XXX

RD900 XXX Operating Instruction Language Fascia Type

900 Series

T-Piece

Resuscitator

AEU English EnglishText

ADU Danish Symbols

OnlyAFU French

ALU Italian

AMU Finnish

ANU Dutch

ASU Spanish

ATU Portuguese

AVU Norwegian

AWU Swedish

AZU Chinese

AGU German(Perivent)

ThelabelpicturedistypicaloftheinformationcontainedonaF&PNeopuff™/Perivent™InfantResuscitator.

Page 5: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 4

2. CLEANING

• Complywithhospital,localandnationalguidelinesforproduct

cleaning frequencies.

• Ensurealloxygenandairsuppliesareturnedoffanddisconnected

fromtheF&PNeopuff™/Perivent™InfantResuscitatorbefore

performingcleaningprocedures.Explosionandfirehazardscan

existwhenperformingcleaningproceduresinanoxygen-enriched

environment.

• Cleaningshallbeperformedatambientconditions.

• Beforecleaning,removeanddiscardalluseddisposableproducts

usingtherecommendedmethodofdisposal.

• Dustallsurfaceswithacleandampsoftcloth.

• Cleanallplasticsurfaceswithdetergent-basedsolution(maximum

2%inwater)ensuringthemanufacturer’sdirectionsforuseofthe

cleaningagentarefollowed.

• Thefollowingproprietarychemicalcleaningwipesare

recommendediftheF&PNeopuff/Periventfasciarequirescleaning

forinfectioncontrolpurposes.

Caviwipes(MetrexResearchCorporation)

Sani-clothHB(ProfessionalDisposables,Inc.)

Asepti-WipesII(Ecolab,Inc.)

Spartan’sTB-CideQuatWipes

VernacareTuffieWipesAlcoholFree

• Dryallsurfacesaftercleaningwithacleansoftclothor

papertowel.

CAUTION: EnsurenopartoftheF&PNeopuff/PeriventInfant

Resuscitatororrelatedaccessoriesisimmersedinanycleaningliquid

orcleaningsolution.ForTestLungcleaninginstructions,seesection

2.2.2.

CAUTION: DonotcleantheF&PNeopuff/PeriventInfantResuscitator

fasciawithproprietarycleaningproductscontainingeither

hydroxides,hypochlorites,peroxides,gluteraldehydeorcleaning

productswithagreaterthan30% alcohol base.

NOTE:Therecommendedchemicalcleaningwipeslistedabovehave

beencheckedforlong-termcompatibilitywiththeIW900SeriesInfant

WarmersandtheF&PNeopuff/PeriventInfantResuscitator.

2.1 Cleaning Products to Avoid

CAUTION: Thechemicalsusedintheseproprietarycleaningproducts

mayleadtodiscoloration,crazingandeventualcrackingofthefascia.

Examplesofproprietarycleaningproductswhichcontainsuch

cleaningchemicalsincludebutarenotlimitedto:

Asepti-Wipes(Ecolab,Inc)

Clorox(TheCloroxCompany)

Endbac256(JohnsonWaxProfessional)

Quat(3M)

Sporiciden(LibertyIndustries,Inc)

SporoxII(SultanHealthcare)

Steris(SterisCorporation)

Terralin(Schülke&Mayr)

Virtek(AVirtekCompany)

Virox(ViroxTechnologies)

CAUTION: Donotuseabrasivecleaningsolutions.

CAUTION: EnsureF&PNeopuff/PeriventInfantResuscitatorand

accessoriesarecheckedbeforereturningtheInfantResuscitatorto

service.

2.2 Cleaning of the Test Lung

Ifrequired,theTestLungcanbedisinfected.Therearetwodifferent

modelsofTestLungandeachhasspecificcleaningmethods.

• TheTestLungisaconsumableitem.Irrespectiveofmodel,the

TestLungshouldregularlybeinspectedforsignsofdamagesuch

asdiscoloration,perishingorcracking.ReplacetheTestLungif

damageisobserved.

2.2.1 Latex (black) Test Lung (discontinued May 2007)

WARNINGTheTestLungcontainsnaturalrubberlatexthatmay

causeallergicreactions.

CAUTION:Latexissusceptibletoattackbysolvents.Ensurenosolvents

areusedtocleantheTestLung.

ThisTestLungconsistsoftwoparts,theflexiblelatextestlungand

arigidblackplasticconnector.TheTestLungcanbedisassembled

fordisinfection.Itcanbedisinfectedusingethyleneoxidegasonly.

Somecarriergasescancausestresscrackingandarenotsuitable.If

indoubt,checkwiththechemicalsupplier.

Page 6: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 5

2.2.2 Silicone (blue) Test Lung RD020-01

TheTestLungmustbecleanedbeforeeveryuseorin

accordancewithhospitalcleaningprotocols.Thistestlungconsistsof

twoparts,theflexiblesiliconetestlungandarigidplasticconnector.

ThetwopartsoftheTestLungmaybedisassembled

fordisinfection.Itcanbecleanedusingdisinfectantscontaining

eitherperaceticacid(e.g.Perasafe®)ororthopthaldehyde

(e.g.Cidex®OPA).Otherdisinfectionmethodscancausedamage

totheTestLungandarenotrecommended.

Pleasefollowthemanufacturer’sinstructionsforstorage,handling

andusageofthesechemicals.

WARNING EnsurethesiliconeTestLungandplasticconnectorare

bothremovedfromtheT-Piecebeforeuseonapatient.

2.3 Cleaning Reusable T-Piece (discontinued September 2008)

ThereusableT-Piece500RD104consistsoftwopartsandcanbe

disassembledfordisinfection(referdiagram).TheT-Pieceshould

bedisinfectedbyautoclavingatupto136°C,220kPafor4

minutes.Followingreassembly,theT-Pieceshouldbetestedprior

tousetoensurethatitisfunctioningcorrectly.RefertotheF&P

Neopuff™/Perivent™OperatingManual(Fisher&PaykelHealthcare

PartNo.185041726)forset-upinstructions.

UnscrewPEEPcapfordisinfection

2.4 Sterilization

CAUTION:

• Donotautoclaveorgas-sterilizeanypartofthe

F&PNeopuff™/Perivent™InfantResuscitator.

• DonotautoclaveanyoftheNeopuff/Periventaccessories,

e.g.TestLungandGasSupplyLine.

• Forcoldsterilization:Ensurethecoldsterilizationagentsaresafe

forusewiththerelevantsurfaces.

2.5 After Cleaning

CAUTION:

• EnsureallF&PNeopuff/PeriventInfantResuscitatorpartsand

accessoriesarecheckedbeforereturningthedevicetoservice.

PlasticConnectorSiliconeTestLung

2. CLEANING continued

Page 7: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 6

3.1 Functional Schematic

3.2 Installation Checks and Preventative Maintenance

WARNINGDroppingtheF&PNeopuff™/Perivent™Infant

Resuscitatororothersimilarformsofimpactmaycausedamage

resultinginincorrectoperationoftheunit.Ifyoususpectdamage

tohaveoccurred,pleaseperformchecksasoutlinedinsection3.3

beforeconnectiontoapatient.

3.2.1 Installation Checks

Priortofirstuse:

–Removemanometercover

–CompletePerformanceTesting(section 3.3)

3.2.2 Preventative Maintenance

• Theintegrityofthesystemandmanometershouldbecheckedprior

tofirstuse,annuallyandafterservicing,usingthe

“TestingoftheF&PNeopuff/PeriventPerformance”procedure

(section3.3).

• Theuseofamountingbrackettohelppreventthe

F&PNeopuff/Periventbeingdamagedordroppedisrecommended.

Mountingsolutionsavailablecanbefoundinsection5ofthis

manual.

• Allmaintenanceandserviceproceduresmustbeperformedby

qualifiedpersonnelusingonlyFisher&PaykelHealthcareparts.

• Alwaysensuregaspassagesarefreefromcontaminants,especially

hydrocarbons,oilsandgrease,priortoreassembly.

• PleasecontactanauthorizedFisher&PaykelHealthcare

representativeforfurtherassistancewithanyservicingor

maintenancerequirements.

• TheTestLungisaconsumableitem;itshouldbemonitoredfor

signsofwearandmaterialdegradationandreplacedasrequired.

3.3 Testing of the F&P Neopuff™ /Perivent™ Performance (Manometer and Valve System)

TheintegrityoftheF&PNeopuff/Periventmanometerandvalve

systemcanbetestedusingthefollowingguidelines.Theinletport

mustbeconnectedtoagassupplycapableofgeneratingconstant

flowat5,10 and 15L/min,andacertifiedgaugeandbleedvalve

shouldbeavailabletocheckthemanometeraccuracy.The

F&PNeopuff/PeriventresuscitationcircuitandT-Piececan

beusedinplaceofableedvalveasshownbelow.

60 7080

-20100 -10

20

30

4050

cmH2O+–

60 7080

100 -10

20

30

4050

cmH2O+–

Manometer

GasOutletPortGasInletPort

CircuitPressureManometer

Maximum Pressure ReliefValve

Peak Inspiratory PressureValve

PeakInspiratoryPressure Control

OutletPort

MaximumPressure ReliefCover

InletPort

Connectpressuregauge betweenF&PNeopuff/Perivent outletandresuscitationcircuit

UseT-Piecewith TestLungattached asableedvalve

CertifiedGauge

3. SERVICE INFORMATION

Page 8: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 7

3.3.1 Testing of the Manometer

1. LiftthecoverofftheMaxPressureReliefValveslightlyandturnoutoftheway.

2. DisconnectalldevicesfromtheF&PNeopuff™/Perivent™outletport.Checkthatthemanometerneedleiswithin+/-2 cmH

2O of

zeroonthemanometergauge.Ifthemanometerdoesnotreadzero,theresettingofthemanometertozeroprocedure(section3.3.4)shouldbefollowed.

3. ConnecttheoutletoftheF&PNeopuff/PeriventInfantResuscitatortoableedvalveandacertifiedgauge(e.g.MensorDigitalPressureGaugeSeries2400).Setthegassupplyto10L/min.Completelyclosethemaximumpressurelimitvalvebyturningtheleft-handknobcompletelyclockwise.Withthebleedvalveclosed,adjustthePeakInspiratoryPressureknobtosetthepressuresothatthecertifiedgaugereads10, 20 and 40 cmH

2O.Checkthatthe

manometerreadswithin+/-2 cmH2Oofthesevaluesateachset

point.

4. Withthepressuresetto40 cmH2O,openandclosethebleedvalve

threetimesandcheckthemanometerneedlerisesandfallssmoothly.If the F&P Neopuff / Perivent Infant Resuscitator fails any of these tests, the manometer should be regarded as inaccurate and replaced with a new manometer (Part No. 043040841).

Follow the manometer replacement guidelines in section 3.3.5 of this manual, or contact your Fisher & Paykel Healthcare service representative for further guidance.

3.3.2 Testing of the Valve System

1. Setthegassupplyto5L/min.CompletelyclosethePeakInspiratoryPressurecontrolandtheMaxPressurecontrolbyturningbothknobscompletelyclockwise.Closethebleedvalveandcheckthatthegaugereadsatleast60 cmH

2O.

2. Setthegassupplyto15L/min.Closethebleedvalveandcheckthatthegaugereadsnohigherthan80cmH

2O.

3. Setthegassupplyto10L/min.ClosethebleedvalveandturntheMaxPressureReliefvalveuntilthemanometerreads40 cmH

2O.

Checkthatthemanometerneedlerisesandfallssmoothly.GentlyrotatetheMaxPressureReliefcoverovertheMaxPressureReliefknob.

4. Resetthepeakinspiratorypressureto20 cmH2Oandturnoffthe

gasflow.Testingisnowcomplete.

If the F&P Neopuff / Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers). Follow the valve replacement guidelines in section 3.3.6 of this manual, or call your Fisher & Paykel Healthcare service representative for further information.

3.3.3 Setting Max Pressure Relief to 40 cmH2O

ThisisrequirediftheMaxPressureReliefhasbeenchanged.ThefactorysettingfortheMaxPressureReliefis40 cmH

2O.

AlternativesettingsfortheMaxPressureReliefshouldbemadeinaccordancewithhospitalprotocol.

1. Adjustgasflowto10L/min.

2. ClosethePeakInspiratoryPressurevalvebyturningtheknobfullyclockwise.

3. AdjusttheMaxPressureReliefknobclockwiseorcounterclockwiseuntilthemanometerreads40 cmH

2O.

4. TurnthePeakInspiratoryPressureknobcounterclockwisesothemanometerreads20 cmH

2Oandshutoffthegasflow.

3.3.4 Resetting the Manometer to Zero

Tosetthemanometertozero:

1. DisconnecttheF&PNeopuff™/Perivent™InfantResuscitatorfromanyotherequipment.

2. Removetheopaqueplasticpluginthelensofthemanometer.

3. Usingasuitableslotscrewdriver,carefullyadjustthescrewinthemanometerfaceclockwiseorcounterclockwisetoresetthemanometertozero.Caremustbetakenwhendoingthis,asover-rotationofthescrewcandamagethemanometerinternals.

4. Replacetheplasticpluginthelensofmanometer.

5. Verifythatthemanometerneedleisnowwithin+/-2.0 cmH2O of

zero.Ifnot,repeattheaboveprocedure.

3.3.5 Manometer Replacement

ThemanometerisnotaserviceableitemandmustbereplacedbyManometerKit043040841.

1. Removethebackcover,fixedbyfourscrews.

2. Disconnectthetubefromthemanometer.

3. Removethemanometerbyunscrewingthetworetainingnuts.

4. Fitthenewmanometerintothefrontpanel,tightentheretainingnutsandreconnectthemanometertube.

5. Refitthefrontpaneltothebackcoverwiththefourscrews.

6. Carryoutthemanometerperformancetestaspersection3.3.1.Itisrecommendedtorecordthelotnumberfromtheboxlabelofthereplacementmanometeronthemaintenancechecklist.

3.3.6 Valve System Replacement

NOTE:Thevalvesareanintegralpartofthevalve,panelandmanifoldassemblyandarenotabletobeserviced.Pleasespecifythemodelnumberfromthepartslist(section4.1.2)whenorderingareplacementvalveassembly.

1. Removethebackcover,fixedbyfourscrews.

2. Disconnectthetubefrommanometer.

3. Removethemanometerbyunscrewingthetworetainingnuts.

4. Fittheexistingmanometerintothenewvalveassemblypanel,tightentheretainingnutsandreconnectthemanometertube.

5.Refitthefrontpaneltothebackcoverwiththefourscrewsfromstep1.

6.Carryoutthevalvesystemperformancetestaspersection3.2.2. ItisrecommendedtorecordthelotnumberofthenewvalveassemblyontotheSet-upandMaintenanceChecklist.

3. SERVICE INFORMATION continued

Page 9: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 8

Cust

omer

Det

ails

Name:

Address:

Country:

PhoneNumber:

Email:

F&P

Neo

puff

™/P

eriv

ent™

Perf

orm

ance

Tes

ting

Rec

ord TestDate:

SerialNumber:

For 900IW130only: InfantWarmerSerialNumber:

TestedBy:

NextTestDueDate:

Check Ref or x Comments

Manometer Checks:

ReplacementManometerSerialNumber:

Accuracy@0 cm (+/-2 cmH2O) 3.3.1-2

Accuracy@10 cm (+/-2 cmH2O) 3.3.1-3

Accuracy@20 cm (+/-2 cmH2O) 3.3.1-3

Accuracy@40 cm (+/-2 cmH2O) 3.3.1-3

Needlemovementsmooth? 3.3.1-4

Valve System Checks:

Pressuregreaterthan60 cmH2O? 3.3.2-1

Pressurelessthan80 cmH2O? 3.3.2-2

Maximumpressuresetto40 cmH2O 3.3.2-3

PIPresetto20 cmH2O 3.3.2-4

Additional Comments:

3.4 Set-up and Maintenance Checklist

Thefollowingtableisprovidedtorecordtheresultsoftheperformancetestsdescribedinsection3.3.Anycomponentsreplacedshouldbe

recordedalso,asappropriate.Thetablemaybephotocopiedorotherwisereproducedasrequired.

Upon receipt of the product and prior to use on a patient, please complete the tests and fill in the table below.

3. SERVICE INFORMATION continued

Page 10: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 9

4.1 F&P Neopuff™ / Perivent™ Infant Resuscitator (Post June 2010)

ThisF&PNeopuff/Periventmodulecontainsasealedvalveassembly

thatcannotbeservicedinthefield.Ifafaultyvalveoccurs,theentire

panelandvalveassemblymustbereplaced.

4.1.1 Assembly Diagram

4. ASSEMBLY DIAGRAMS

UNITS MANUFACTURED POST JUNE 2010

# Description Part Number Reqd

1 HandleNeopuffSpare 043043976 1

2 Plug(setoffour) 693040706 2

3 ScrewM8x20 614040309 8

4 Endcap(upper) 043042565 1

5 Endcap(lower) 043042564 1

6 Backcover 641040816 1

7Modelspecificpaneland valveassembly (seeSparesKitsection4.1.3)

(seetableinsection4.1.3)

1

8 Manometerkit 043040841 1

9 CapValveNeopuffSpare(Blue) 043043977 1

10 Cover,maximumpressurereliefvalve 043041057 1

11 Foot 693041436 4

12 Screw#8x1”Csk 616050011 4

13ScrewM4x8 Pan hd (handleattachment)

614040117 2

NOTE:Whitevalvecap(pre-June2010)canbeorderedusing043042345(1/bag).

4.1.2 Parts List

Page 11: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 1 0

Model specific panel and valve assembly Blue (kit includes fascia)

043043595 Fascia&ValveAssyBlueEnglish

043043596 Fascia&ValveAssyBlueGerman

043043597 Fascia&ValveAssyBlueItalian

043043598 Fascia&ValveAssyBlueSpanish

043043599 Fascia&ValveAssyBlueFrench

043043600 Fascia&ValveAssyBlueNorwegian

043043601 Fascia&ValveAssyBlueDutch

043043602 Fascia&ValveAssyBlueSwedish

043043603 Fascia&ValveAssyBluePortuguese

043043604 Fascia&ValveAssyBlueDanish

043043605 Fascia&ValveAssyBlueFinnish

Model specific panel and valve assembly Purple (kit includes fascia)

043042347 Fascia&ValveAssyEnglish

043042348 Fascia&ValveAssyGerman

043042349 Fascia&ValveAssyItalian

043042350 Fascia&ValveAssySpanish

043042351 Fascia&ValveAssyFrench

043042353 Fascia&ValveAssyDutch

043042354 Fascia&ValveAssySwedish

043042355 Fascia&ValveAssyPortuguese

043042356 Fascia&ValveAssyDanish

043042357 Fascia&ValveAssyFinnish

043042358 Fascia&ValveAssyNorwegian

4. ASSEMBLY DIAGRAMS continued

NOTE:AllRD900unitsmanufacturedfromMay1999 can

accommodatetheabovespareskit.

4.1.3 Spares Kit

Page 12: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 1 1

ImpacttotheF&PNeopuff™/Perivent™InfantResuscitator,

causedbyroughhandlingortheunitbeingdropped,candamagethe

valvesystemandproduceirregularresuscitationpressures.

Tohelppreventanyimpacttothedevice,Fisher&PaykelHealthcare

recommendstheuseofoneofthemountingsystemsshownbelow.

5.1 900RD301 Side-Mounting Block, 900MR170 C-Clamp, 900MR030 Pole Bracket

The 900RD301SideMountingBlockfitsintothedovetailslotonthe

sideoftheInfantResuscitator.The900RD301maythenbeconnected

toa17 to 40mmpoleusingeitherthe900MR170C-Clampor

900MR030PoleBracket.

5-15 L/minGAS INLETGAS OUTLET

900RD301

900MR170 C-Clamp

900MR030PoleBracket

Mounting Option Parts to order

Quickdisconnect(C-Clamp) 900RD301 & 900MR170

PermanentPoleBracket 900RD301 & 900MR030

5.2 900MR088 Rail Bracket

FormountingtheF&PNeopuff/PeriventInfantResuscitatorcentrally

onstandardrails(2.5 to 5.5cmx1cm/0.98 to 2.17”x0.39”).

AlsorequiresRD050-01.

5. MOUNTING OPTIONS

5.3 900RD302 Wall Mount Bracket

The 900RD302WallMountBrackethastwotonguefittingsthat

maybescrewedtothebackoftheF&PNeopuff™/Perivent™Infant

Resuscitatorbyremovingthefourplasticplugsfromtherearpanel,

andaT-shapedbracketthatshouldbescrewedtothewall.

Thisallowstheinfantresuscitatortobequicklyandeasilyremoved

fromandreplacedonthebracketasrequired.

Page 13: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

N E O P U F F ™ T E C H N I C A L M A N U A L 1 2

5.4 RD050-01 Pole and Rail Central Mount

FormountingtheF&PNeopuff™/Perivent™InfantResuscitator

centrallyonapoleoronstandardrails(2.5 to 5.5cmx1cm/

0.98to2.17”x0.39”),acentralmountingblockcanbeaffixedto

thebackoftheunitbyremovingthefourplasticplugsfromtherear

panel.TheunitwillthenfittheF&Pbracketandclamps.

5. MOUNTING OPTIONS continued

Page 14: Infant Resuscitation | Technical ManualNE EN MNAL 3 SYMBOL DEFINITION Attention: Consult the Operating Instructions MAX-P Sets the Max Pressure Relief that may be delivered to the

www.fphcare.com

For more information please contact your local Fisher & Paykel Healthcare representative

1850

4159

7 REVO

EN

© 2

019Fisher&PaykelHealthcareLimited

Importer/ Distributor Australia (Sponsor) Fisher & Paykel Healthcare Pty Ltd, 19-31 King Street, Nunawading, Melbourne, Victoria 3131.

Tel: +61 3 9871 4900 Fax: +61 3 9871 4998

Austria Tel: 0800 29 31 23 Fax: 0800 29 31 22

Benelux Tel: +31 40 216 3555 Fax: +31 40 216 3554

Brazil Fisher & Paykel do Brasil, Rua Sampaio Viana, 277 cj 21, Paraíso, 04004-000, São Paulo – SP, Brazil

Tel: +55 11 2548 7002

China 代理人/售后服务机构: 费雪派克医疗保健(广州)有限公司, 广州高新技术产业开发区科学城科丰路31号G12栋301号

电话: +86 20 32053486 传真: +86 20 32052132

Denmark Tel: +45 70 26 37 70 Fax: +46 83 66 310

Finland Tel: +358 94 1590 355 Fax: +46 83 66 310

France Fisher & Paykel Healthcare SAS, 10 Av. du Québec, Bât  F5, BP 512, Villebon-sur-Yvette, 91946 Courtaboeuf Cedex, France

Tel: +33 1 6446 5201 Fax: +33 1 6446 5221 Email: [email protected]

Germany Fisher & Paykel Healthcare GmbH & Co. KG, Deutschland, Österreich, Schweiz, Wiesenstrasse 49, D 73614 Schorndorf, Germany

Tel: +49 7181 98599 0 Fax: +49 7181 98599 66

Hong Kong Tel: +852 2116 0032 Fax: +852 2116 0085

India Tel: +91 80 2309 6400

Ireland Tel: 1800 409 011 Fax: +44 1628 626 146

Italy Tel: +39 06 7839 2939 Fax: +39 06 7814 7709

Japan Tel: +81 3 5117 7110 Fax: +81 3 5117 7115

Korea Tel: +82 2 6205 6900 Fax: +82 2 6309 6901

Norway Tel: +47 21 60 13 53 Fax: +47 22 99 60 10

Russia Tel. and Fax: +7 495 782 21 50

Spain Tel: +34 902 013 346 Fax: +34 902 013 379

Sweden Tel: +46 8 564 76 680 Fax: +46 8 36 63 10

Switzerland Tel: 0800 83 47 63 Fax: 0800 83 47 54

Taiwan Tel: +886 2 8751 1739 Fax: +886 2 8751 5625

Turkey İthalatçı Firma: Fisher Paykel Sağlık Ürünleri Ticaret Limited Şirketi, İletişim Bilgileri: Ostim Mahallesi 1249. Cadde No:6, Yenimahalle, Ankara, Türkiye 06374,

Tel: +90 312 354 34 12 Fax: +90 312 354 31 01

UK Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, UK

Tel: 0800 132 189 Fax: +44 1628 626 146

USA/Canada Tel: 1800 446 3908 or +1 949 453 4000 Fax: +1 949 453 4001

Manufacturer Fisher & Paykel Healthcare Ltd, 15 Maurice Paykel Place, East Tamaki, Auckland 2013

PO Box 14 348 Panmure, Auckland 1741, New Zealand

Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: [email protected] Web: www.fphcare.com