index [microdentsystem.com] · 2015-11-24 · the periodical offer of formation courses for...
TRANSCRIPT
INDEX
PRESENTATION
INTRODUCTION
IMPLANT DESIGN
RAW MATERIALS
MACRO-STRUCTURE
MICRO-STRUCTURE
TECHNICAL FEATURES OF THE PRODUCT
INTRODUCTION
FEATURES
IMPLANT DIMENSIONS
INSTRUCTIONS FOR USE
CONTRA-INDICATIONS
SURGICAL PROTOCOL
CONVENTIONAL SURGICAL PROTOCOL
SURGICAL PROTOCOL WITH THE BONE EXPANSION SYSTEM
MANUAL INSERTION OF THE IMPLANT
MECHANICAL INSERTION OF THE IMPLANT
FITTING THE COVER SCREW FOR FIRST SURGERY
IMPLANT PACKAGING
PACKAGE PRESENTATION
PACKAGE FEATURES
GUIDANCE ON THE OPENING OF THE CONTAINER
LABELING CODE SYMBOLOGY
INSTRUCTIONS FOR USE
MRI SAFETY INFORMATION
SURGICAL INSTRUMENTS
GUARANTEE PROGRAM
CHART OF MICRODENT GENIUS INSTRUMENTS
Pag. 05
Pag. 06
Pag. 07
Pag. 15
Pag. 17
Pag. 17
Pag. 18
Pag. 18
Pag. 12
Type of raw material
Composition table
Previous considerations
Recommended sequence
Technical features
Usage features
Relative contra-indications
Local contra-indications
Recommendations of the insertion sectors
Other recommendations
Surface
External shape features/characteristics (shape and thread type)
Connection
PRODUCT STORAGE PRECAUTIONS
PRESENTATION
Implant Microdent System S.L. is a company that, from its creation in 1983, has specialized in deve-loping, manufacturing and marketing components for odontological treatments, and, specially, oral implantology.
Always striving for customer satisfaction, Microdent offers all the guarantee, both of quality and service, backed up by its long and ascending professional history in the field of oral implantology.
The periodical offer of formation courses for professionals of the implantology field and the existence of a collaboration and assessment team allows Microdent to have products that verify the most strict quality parameters stablished, as well as to satisfy the needs of its clients.
Microdent has modern facilities equipped with the most cutting-edge technology, thus making Micro-dent into one of the pioneer companies in the research and manufacture of solutions for oral implantolo-gy. The patents obtained by Microdent products demonstrate the continued dedication of the company to the research work.
MICRODENT GENIUS SURGICAL MANUAL 5
INTRODUCTION
IMPLANT DESIGNThe internal connection Microdent Genius implant system has been designed and conceived to facilita-te and to solve the important medical work of practitioners, who require high quality materials.
RAW MATERIALS• Type of raw material
The raw material used in the manufacture of Microdent implants is Titanium grade 4, in compliance with the ISO 5832 standard, thus confirming its suitability for use in the manufacture of dental implants.
All the raw material received by Implant Microdent System is acquired from suppliers of established standing and proven experience. To ensure this level of requirement, all material received is accom-panied by the proper composition analysis and mechanical performance certificates issued by the manufacturer.
• Composition table
The material composition must be consistent with the ISO5832-2:1999 standard “Implants for surgery – Metallic materials – Part 2: Unalloyed Titanium”, as described in the following table:
Material type
Titanium Gr4
Limits in the material composition (% in relation to the mass)
Microdent Genius With ribbed cone internal connection
0.5 max 0.4 max 0.1 max 0.05 max 0.0125 max
MICRODENT GENIUS SURGICAL MANUAL6
MACRO-STRUCTURE • External shape features/characteristics (shape and thread type)
The threaded contour of the Microdent Genius implant is characterized by presenting auto-thread and a broad thread step.
• Connection
The Genius implant has a ribbed cone internal connection with 6 equidistant grooves.
MICRO-STRUCTURE • Surface
The Genius implant surface is treated with aluminum oxide (sandblasting).
It consists of two phases:
1. In the first phase, the implant is subjected to a projection of particles in a sealed room, with regulated and controlled pressure. This process increments the implant total outer surface.
2. In the second phase, the implant is subjected to a washing and decontamination process.
Image 1: ATEC (Abrasive Treatment Extreme Cleaning) surface treatment
TECHNICAL FEATURES OF THE PRODUCT
INTRODUCTIONThe Microdent Genius dental implant is an internal connection endosseus sub-gingival implant. This implant is inserted at subcrestal or embedded position, along with the rest of the general rehabilitation.The connection design has a conical socket which ends in a cylindrical section.
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
7
Image 3: detail of a subcrestal insertion
Image 2: product technical features
FEATURES Technical features• Designed to adapt to different bone types.•1mm thread step.• Auto-thread screw.• Platform switching system (designed to be inserted at subcrestal level, see image 3).
Usage features • VersatileThe Genius system has a broad range of implants to answer to different implantological situations.
• SatisfactionMore than 30 years of experience guarantee implant osseointegration.
Platform switching
Self-tapping implant
Available in four different core diameters
Same internal connection allows for the connection of any abutment in the system
Surface treatment ATEC (Abrasive Treatment Extreme Cleaning)
Anatomical cutting grooves
MICRODENT GENIUS SURGICAL MANUAL8
IMPLANT DIMENSIONSThe Microdent Genius implant has a range of four platforms, 3.50, 4.00, 4.50 and 5.00mm. The implant lenght varies depending on the platform.
The different lengths available for each model are shown below:
8mm 10mm 12mm 14mm 16mm 18mmLength
Platform Ø 3.5 / Core Ø 3.5
GN35
08
GN35
10
GN35
12
GN35
14
GN35
16
GN35
18
18mm 16mm 14mm 12mm 10mm 8mm
GN40
18
GN40
16
GN40
14
GN40
12
GN40
10
GN40
08
Platform Ø 4.0 / Core Ø 4.0
8mm 10mm 12mm 14mm 16mm 18mmLength
Platform Ø 4.5 / Core Ø 4.5
GN45
08
GN45
10
GN45
12
GN45
14
GN45
16
GN45
18
16mm 14mm 12mm 10mm 8mm
GN50
16
GN50
14
GN50
12
GN50
10
GN50
08
Platform Ø 5.0 / Core Ø 5.0
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
9
Platform Ø 4.5
Indicated for single and multiple fixed rehabilitations on premo-lar and molar groups.Indicated for rehabilitation of edentulous patients.For multiple structures, it is recommended to use in combination with other larger diameter implants.
Platform Ø 4.0
Indicated for single and multiple fixed rehabilitations on canines and premolars.Indicated for rehabilitation of edentulous patients.For multiple structures, it is recommended to use in combination with other larger diameter implants.
Platform Ø 3.5
Indicated for single and multiple fixed rehabilitations of the previous group.Indicated for rehabilitation of edentulous patients.For multiple structures, it is recommended to use in combination with other larger diameter implants.
INSTRUCTIONS FOR USE
The Microdent dental implants are indicated to be surgically placed into the maxilla and/or mandible to provide support for the dental prosthe-sis. They are indicated for dental rehabilitation of partially or totally edentulous patients, for support and retention of single crowns, bridges or overdentures. The minimum age for placement of dental implants depends on the completion of an individual’s growth. When a person has completed their development, they will be able to enjoy the benefits of dental implants, it is not recommended before that point. There is no maximum age for dental implant treatments.
Platform Ø 5.0
Indicated for single and multiple fixed restorations on molar areas.Indicated for rehabilitation of edentulous patients.
Recommendations of the insertion sectors (Clearest zones)
MICRODENT GENIUS SURGICAL MANUAL10
Other recommendations
The ideal distance between an implant and a tooth is 1.5 mm or more. Between implants, the recomended distance is 3mm or more (see image 4).
Small diameter implants are indicated only for replacement of central and lateral incisors in the maxilla and mandible.
Small diameter implants and angled abutments are not recommended for the posterior region of the mouth.
CONTRA-INDICATIONS
As in any surgery, everything that endangers or might endanger the osseointegration of the implant has to be evaluated in order to minimize the implant placement risk.
Serious internal medicine problems, bone metabolism disorders, uncontrolled coagulation disorders, uncooperative or unmotivated patients, alcohol or drug abuse, psychosis, long-time treat-ment-resistant functional disorders, xerostomia, weakened immune system, diseases with a regular use of steroids, titanium allergy, non-controllable endocrine diseases.
Relative Contra-indications:
Previously irradiated bone, diabetes, anticoagulant medication / hemorrhagic diabetes, bruxism, para-functional habits, unfavorable bone anatomy, smoking, uncontrolled periodontitis, jaw diseases and disorders of the oral mucosa susceptible of treatment, pregnancy, inadequate oral hygiene.
Local Contra-indications:
Insufficient available bone or inadequate bone quality, local root fragments.
PRODUCT STORAGE PRECAUTIONS
The dental implant, once packaged, does not require special storage conditions. The package only requires to keep its integrity and incorruptibility. It can be stored under normal conditions of temperature and humidity.
The packaging system guarantees, according to aging tests conducted in compliance with ASTM F1980, the sterile state of the implant for five years from the date of packaging.
Image 4: detail of implant-tooth and implant-implant distance
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
11
SURGICAL PROTOCOL
There are two different surgical protocols, the conventional protocol and the atraumatic expansion protocol. In the first, the alveolus preparation is performed through a drilling sequence, while in the second it is prepared through the employment of Microdent Expanders.
CONVENTIONAL SURGICAL PROTOCOLPrevious ConsiderationsAppropiate radiographic pre-operative analysis is necessary to determinate the position of any impor-tant anatomical structure.
Microdent Genius implants can be placed with the flap lifted to directly visualize the bone bed or without a mucoperisoteal incision by using circular scalpels.
The drilling sequence will be determined by the bone quality of the patient being different in a bone type 1 and 4. The amount of bone must be evaluated based on technical and radiological images used in implantology. The needed instruments for drilling are sharp instruments that require constant irrigation.
The Microdent Genius implant is designed for supracrestal placement. The nominal length of the implant is defined from the platform to the apical area of the implant.
Microdent surgical instruments for drilling are part of an overall concept. Their design has been adapted to products manufactured by Implant Microdent System S.L., therefore, the use of surgical instruments from other trademarks might cause a non-optimal performance of our products.
The surgical instruments for drilling have to be cleaned and sterilized before being used in another patient.
The surgical drills are indicated for 10-20 uses. To exceed these number of uses may compromise the suc-cess of the treatment.
When using drilling surgical instruments, it is mandatory to properly irrigate in order to avoid bone overheating since it could cause failure of the implant.
The marketed Implant Microdent drills have an irrigation system that diminishes bone heating during the surgical process.
Recommended sequenceThe drilling sequence varies according to the implant diameter. Below, the drilling protocols for each of the Genius implant series are shown.
Image 5: Microdent Genius implant with conical abutment
Implant 3.5 Platform Surgical Protocol
Reference
Drilling speed (RPM)
FP1812/ FP1817 F280/F300F250FC20
400-800400-800800-1200800-1200
MICRODENT GENIUS SURGICAL MANUAL12
Implant 4.0 Platform Surgical Protocol
Reference
Drilling speed (RPM)
FP1812/ FP1817 F320/F350F280FC20
400-800400-800800-1200800-1200
Implant 4.5 Platform Surgical Protocol
Reference
Drilling speed (RPM)
FP1812/ FP1817 F350/F380F320F250/F280FC20
300-500400-800400-800800-1200800-1200
Implant 5.0 Platform Surgical Protocol
Reference
Drilling speed (RPM)
FP1812/ FP1817 F380/F420F320F250/F280FC20
300-500400-800400-800800-1200800-1200
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
13
SURGICAL PROTOCOL WITH THE BONE EXPANSION SYSTEM
The alveolar osteotomy can also be prepared with Microdent atraumatic expanders, specially in cases of narrow bone crest or when a better use of the patient’s bone support is desired (see manual expansion at www.microdentsystem.com).
Expansors description and main features
EXPANDER #1 EXPANDER #2 EXPANDER #3 EXPANDER #4 EXPANDER #5 EXPANDER #6
INITIAL SHORT
Ø TIP 1.00mm to 2.30mm
INITIAL LONG
Ø TIP 1.50mm to 3.00mm
EXPANDER
Ø TIP 1.90mm to 3.80mm
EXPANDER
Ø TIP 2.50mm to 4.50mm
EXPANDER
Ø TIP 3.00mm to 5.10mm
SINUS
STRAIGHT
Ø 3.50mm
LONG. 10mm LONG. 16mm
8mm=Ø 2.75mm 10mm=Ø 3.00mm
12mm=Ø 3.25mm 14mm=Ø 3.50mm
YELOW COLOR
RED COLOR
LIGHT BLUE COLOR
GREEN COLOR
DARK BLUE COLOR
8mm=Ø 3.40mm10mm=Ø 3.70mm12mm=Ø 4.00mm14mm=Ø 4.30mm
8mm=Ø 3.90mm10mm=Ø 4.20mm
FOR IMPLANTSØ 4mm
12mm=Ø 4.50mm14mm=Ø 4.80mm
Ø 4.20mm
ELEVATOR
MANUAL INSERTION OF THE IMPLANT
Insert the implant into the osteotomy and begin fixing the implant by turning it clockwise. If during the insertion of the implant in the osteotomy (on the first threads) it is necessary to apply more force than usual, it is recommended the removal of the implant from the alveolus and increasing its size with a larger burr. Forcing in the implant could cause damage.
Once the implant is 2/3 inserted in length, finish the insertion by using the extension adaptable to the metal implant holder r(eference ACC44 or ACL44) and if this is not possible due to particular space conditions, remove the metal implant holder with the aid of screwdriver (MH120) and finish the insertion process using the corresponding adaptor.
It is advised the use of an insertion torque not greater than 40 Ncm. To control the torque, the torque-measuring ratchet (reference LDR2070) is recommended.
MECHANICAL INSERTION OF THE IMPLANT
If the implant is to be inserted using a micromotor, insert the adaptor (reference LC44) that was pre-viously fitted to the hand piece, over the metal implant holder.
To extract the implant from the plastic support, press the support lightly in the direction of the opening where it is housed.
Move the implant to the osteotomy and proceed to insert it, regulating the speed of the micromotor to obtain maximum control over the process.
We recommend using an insertion torque not greater than 40 Ncm.
MICRODENT GENIUS SURGICAL MANUAL14
FITTING THE COVER SCREW FOR FIRST SURGERY
The cover screw is housed in the transport cap on the opposite side to the implant.Both implant and cover screw are presented in a sterile condition.
Once the implant has been inserted into the osteotomy, remove the cover screw from the plastic cap using screwdriver (MH120C) by turning it gently counterclockwise.
Carefully take it to the implant, insert it and screw in the accessory by turning it clockwise.It is advised the use of an insertion torque of 5 Ncm.Proceed to stitch the soft tissue.
IMPLANT PACKAGING
PACKAGE PRESENTATION
Microdent’s packaging system is designed to be easy and simple to use. All containers are decorated with the distinctive color of the corresponding implant system, being green the color chosen for the Microdent Genius implant system.
The rectangular design and the small size of the box allows an easy storage, minimizing the need for space.
Inside the package, two identification labels accompany the implant.In the secondary container, the web address to visit the instructions for use is indicated.
PACKAGE FEATURES
The implants from the Microdent Genius series are presented in a sterile condition.The implant, which carries a metal implant-holder, is mounted on a plastic support as well as the carrier cap. The conveyor cap hosts the cover screw on the opposite side of the implant.
Said plastic support is inside of a container that maintains its sterility by a metallic sealing cap. The protection of this sterile barrier system is achieved by a threaded cap with precint that closes the outer package and does not allow the manipulation of the content.
Both the outer packaging and the inner packaging have a label with the following information: batch number, size and model of the implant and expiration date.
To realize the correct monitoring of the obliged traceability of these products, the labels have to be included in the patient’s file.
Images 6: primary packaging of a Microdent Genius implant
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
15
GUIDANCE ON THE OPENING OF THE CONTAINER
Before opening the package, perform a visual check to ensure that it has not been damaged.
Instructions for use packaging
Open the implant threaded cap counter-clockwise.
Remove the aluminum cap from the outer container by the flap.
Once the cap has been removed, extract the implant’s green carrier. The green plastic implant-holder contains the cover screw.
Once the small green carrier has been extracted from the outer container, use the green plastic implant-holder and set it on the metallic implant-holder. It adapts to the square of the implant-holder.
Remove the implant by vertically lifting upwards.
The implant can also be taken into the patient’s mouth with the handpiece by directly inserting it in the metallic implant-holder.
Do the same motion as in Fig. 5, remove by vertically lifting upwards.
Image 7: secondary packaging of a Microdent Genius implant
MICRODENT GENIUS SURGICAL MANUAL16
INSTRUCTIONS FOR USEIn accordance and compliance with legal requirements, all implants manufactured by Implant Microdent System have instructions for use. These instructions are available at www.microdentsystem.com
LABELING CODE SYMBOLOGY
MRI SAFETY INFORMATIONImplants and abutments have not been evaluated for safety and compatibility in the MR environment. Implants and abutments have not been tested for heating, migration, or image artifact in the MR environment. The safety of implants and abutments is unknown. Scanning a patient who has this device may result in patient injury.
Sterile by irradiation
Barcode
Endosteal diameter
Lot/batch number
Manufacturer
Do not use if package is damaged
Expiration date
Reference/article number
Caution: federal law restrics this device to sale by or on the order of a dental professional
Implant type
Do not re-use
Caution, consult accompanying documents
SURGICAL INSTRUMENTSFor a proper use and cleaning of surgical instruments, consult their instructions for use.
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
17
GUARANTEE PROGRAM Implant Microdent System provides guarantees for the products it distributes. The terms and conditions of these guarantees are indicated in the Microdent Implant System “Guarantee Program”, consult their instructions for use: www.microdentsystem.com
CHART OF MICRODENT GENIUS INSTRUMENTSMaterial instruments: Stainless steel
EXPANDERS
1025Expander #1
2938Expander #3
4452Expander #5
1830Expander #2
3345Expander #4
3535Sinus elevator
OTHER TOOLS
GNLLICContra-angle insertion key
DC120HLHex.1,20mm contra-
angle screwdriver
ACC44Extender (short)
DL120HLHex. 1,20mm
ratchet screwdriver
ACL44Extender (long)
LDR2070Torque measuring
ratchet
MH120CScrewdriver hex.1,20mm (short)
LC44Contra-angle key
MH120LScrewdriver hex.1,20mm (long)
X-ray templates (plastic film) also available. Part Number PPQEK
MICRODENT GENIUS SURGICAL MANUAL18
FC20CCortical dfrill (short)
F280CSurgical drill
Ø 2.80mm (short)
F350CSurgical drill
Ø3.50 mm (short)
FP1812Pilot drill (short)
F300CSurgical drill
Ø 3.00mm (short)
F380CSurgical drill
Ø 3.80mm (short)
F250CSurgical drill
Ø 2.50mm (short)
F320CSurgical drill
Ø 3.20mm (short)
F420CSurgical drill
Ø4.20mm (short)
FC20LCortical dfrill (long)
F280LSurgical drill
Ø 2.80mm (long)
F350LSurgical drill
Ø 3.50mm (long)
FP1817Pilot drill (long)
F300LSurgical drill
Ø 3.00mm (long)
F380LSurgical drill
Ø 3.80mm (long)
F250LSurgical drill
Ø 2.50mm (long)
F320LSurgical drill
Ø 3.20mm (long)
F420LSurgical drill
Ø4.20mm (long)
DRILLS
Drill stops also available
6mm
10mm
14mm
18mm
8mm
12mm
16mm
6mm
10mm
14mm
18mm
8mm
12mm
16mm
6mm
10mm
14mm
18mm
8mm
12mm
16mm
MICRODENT GENIUS SURGICAL MANUAL
ww
w.m
icro
dent
syst
em.c
om
19