in situ generation of ruthenium tetroxide and osmium

4
In situ Generation of Ruthenium Tetroxide and Osmium Tetroxide for the Physical Sciences and Their Reaction Indicators Paul Beauregard Chemist and Electron Microscopist [email protected] Introduction: Recently, there was a suggestion on the MSA listserver about the use of osmium tetroxide (OsOJ and how to handle it. One suggestion was that ampoules be scored, placed in a glass jar, and the ampoule smashed to release the contents. This seemed like a very unsafe way to use osmium tetroxide or ruthenium tetroxide. The purpose of this article is to suggest a way to generate smaller amounts of these compounds in a safer manner than smashing ampoules and wondering about what to do with the unused portion after staining or storing. Another purpose is to discuss a new reaction indicator for mainly osmium tetroxide. The use of a reaction specific indicator was mandatory for judging the level or degree to which staining had proceeded in thin sections for the transmission electron microscope (TEM). Tetroxide Backgrounds: The book Polymer Microscope, shows a reaction where ruthenium trichloride reactes with sodium hypochlorite to produce ruthenium tetroxide (RuO 4 ). This reaction could be used to generate in situ RuO 4 and its vapors at a fairly fast rate. The safest way to proceed is to minimize the amount of RuO 4 generated, but still generate enough to stain thin sections. A method used to do this in the past is outlined below. While using gloves, 1 to 2 mgs of RuCI 3 .xH 2 O were typically weighed out, to which were added a few mLs of - 5 % sodium hypochlorite solution. The generated RuO 4 gas was contained inside a nearly sealed glass container like a Petri dish or ground glass covered dish. This "minimal use" procedure limited accidental human exposure to a very minute amount of the dangerous fumes generated if anything went wrong. One to two milligrams (mgs) was enough RuO 4 material to stain most TEM grids in less than 5-10 minutes. Since this is a vapor-state staining process, one needs a reaction indicator and various materials that can be used, and are listed as reacting with RuO 4 in the book Polymer Microscopy*. After experimentation, 3M tape 665 was found to turn dark enough to indicate the degree of reaction. 665 does not work with OsO4. Looking at the periodic table of the elements, it seemed reasonable that OsO 4 could be generated by using an analogous reaction equivalent to the reaction of ruthenium trichloride hydrate with sodium hypochlorite or Clorox^ bleach. However, these two tetroxides did not react at the same rates. Various reactions for RuO 4 generation were listed in the book Advanced Inorganic Chemistry, A Comprehensive Tex? and could be reviewed. The use of heat and nitric acid were described. What was needed was an OsO 4 analog of the in situ RuO 4 generation at room temperature. Discussion of Results: During the listserver exchange mentioned above, it was suggested that sodium or potassium periodate be used for in situ generation of OsO 4 . That reaction was stated or hinted at in Advanced Inorganic Chemistry by Cotton & Wilkinson 2 . On the Internet there was a strange statement about osmium dioxide. Paraphrasing, "When you opened a bottle of osmium dioxide you could smell traces of 20 • IMCnOSCOPYTODfnf September/October 2002 OsO 4 ." This air oxidation statement suggested that in situ generation of OsO 4 should be possible with a more powerful oxidi2er. A review of microscopy textbooks yielded nothing of value about this statement, neither about in situ generation of OsO,,, nor about how to gauge the progress of the staining reaction with an indicator. This gauge was something that was needed. If one stained with fumes from a 2% or 4% solution, the reaction times would vary and one risked over- staining thin sections on TEM grids. Not only that, the concentration ofthe fumes would vary with the apparatus volume used. If one used an opened ampoule to suspend the grids in or over the OsO 4 , then air currents could dilute the vapors and give variable results. Clearly, any procedure needed a reaction indicator gauge for either tetroxide. Furthermore, a way was needed to partially control the way the tetroxides were produced and their concentration in a closed system like that mentioned above. The next step was to find a good reaction indicator for OsO,., An old dilute solution of osmium tetroxide in sulfuricacid from G. Fredrick Smith was used to conduct experiments to test for a good OsO,, reaction indicator. It was quickly discovered that OsO,, did not stain the 665 indicator tape because the tape did not have enough double bonds, if any. Having worked on rubber samples, some obvious indicator choices existed. One was a gum stock butadiene rubber known as Budene 15 1207 (Goodyear Tire & Rubber). It was relatively colorless and transparent. Natural rubber worked fine, but it was not pure and was colored depending on the grade used. The Budene 1207 was dissolved in cyclohexane as a thick saturated solution. This took about a week of occasional shaking. This solution could then be spread out on a Fisher Superfrost s microscope slide to form a thin indicator film. Sometimes a thicker localized spot of deposited rubber would also be created. This Budene indicator gradually turned darker with increased exposure to the GFS osmium tetroxide solution when placed in the Petri dish apparatus. This was a fairly good indicator but a better commercial indicator might be found by someone else. Osmium dioxide and osmium chloride were chosen forthe starting osmium materials. It was very clear that In situ generation of OsO 4 was not going to lead to a cheaper way to make OsO,.. None of this material was cheap. On the other hand, one only had to buy a small amount, such as f ve grams for testing. For oxidizers, I chose 30% hydrogen peroxide, sodium periodate and Clorox® bleach (sodium hypochlorite). Both osmium compounds produced a darkening of my indicator and natural rubber pieces in the Petri dish with selected oxidizers. The bad news was that the reaction did not proceed at a very fast rate. Instead of 1 to 2 minutes, the osmium compounds required 1 to 2 hours or longer to stain thin sections using 1 to 2 mgs. The good news was that it did stain the double bond indicator at room temperature. I did not try to heat any ofthe reactants to see if a quicker reaction occurred. I did not want to generate hypochlorous acid fumes by heating bleach, for example. OsO z reacted very slowly and could take hours or overnight to generate enough OsO,, to stain grids, OsO;, powder also caused hydrogen peroxide to almost instantly decompose very violently. DO NOT use 30% H ; O ; with OsO 3 . The other two oxidizers worked fine. OsO 2 had some drawbacks. One was that it never reacted totally, and thus one could be left with a dangerously fine powder of OsO2 at the end ofthe reaction to clean up. This could be messy, so use gloves, OsCl 3 reacted with all three oxidizers and did generate OsO 4 at a faster rate than OsO 2 , but at a much slower rate than RuCl 3 generated RuO r OsCl 3 did generate enough OsO 4 over a 1 to 3 hour time period to stain thin sections. One only needed a few milligrams to do the job, just like RuCl 3 . Using more OsCl 3 would Downloaded from https://www.cambridge.org/core. IP address: 65.21.228.167, on 12 Nov 2021 at 03:18:48, subject to the Cambridge Core terms of use, available at https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1551929500058314

Upload: others

Post on 12-Nov-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: In situ Generation of Ruthenium Tetroxide and Osmium

In situ Generation of RutheniumTetroxide and Osmium Tetroxide for

the Physical Sciences and TheirReaction Indicators

Paul BeauregardChemist and Electron Microscopist

[email protected]:

Recently, there was a suggestion on the MSA listserver aboutthe use of osmium tetroxide (OsOJ and how to handle it. Onesuggestion was that ampoules be scored, placed in a glass jar, andthe ampoule smashed to release the contents. This seemed like avery unsafe way to use osmium tetroxide or ruthenium tetroxide. Thepurpose of this article is to suggest a way to generate smaller amountsof these compounds in a safer manner than smashing ampoulesand wondering about what to do with the unused portion after stainingor storing. Another purpose is to discuss a new reaction indicator formainly osmium tetroxide. The use of a reaction specific indicator wasmandatory for judging the level or degree to which staining hadproceeded in thin sections for the transmission electron microscope(TEM).Tetroxide Backgrounds:

The book Polymer Microscope, shows a reaction whereruthenium trichloride reactes with sodium hypochlorite to produceruthenium tetroxide (RuO4). This reaction could be used to generatein situ RuO4 and its vapors at a fairly fast rate. The safest way toproceed is to minimize the amount of RuO4 generated, but stillgenerate enough to stain thin sections. A method used to do this inthe past is outlined below.

While using gloves, 1 to 2 mgs of RuCI3.xH2O were typicallyweighed out, to which were added a few mLs of - 5% sodiumhypochlorite solution. The generated RuO4 gas was contained insidea nearly sealed glass container like a Petri dish or ground glasscovered dish. This "minimal use" procedure limited accidental humanexposure to a very minute amount of the dangerous fumes generatedif anything went wrong. One to two milligrams (mgs) was enoughRuO4 material to stain most TEM grids in less than 5-10 minutes.Since this is a vapor-state staining process, one needs a reactionindicator and various materials that can be used, and are listed asreacting with RuO4 in the book Polymer Microscopy*. Afterexperimentation, 3M tape 665 was found to turn dark enough toindicate the degree of reaction. 665 does not work with OsO4.

Looking at the periodic table of the elements, it seemedreasonable that OsO4 could be generated by using an analogousreaction equivalent to the reaction of ruthenium trichloride hydratewith sodium hypochlorite or Clorox^ bleach. However, these twotetroxides did not react at the same rates.

Various reactions for RuO4 generation were listed in the bookAdvanced Inorganic Chemistry, A Comprehensive Tex? and couldbe reviewed. The use of heat and nitric acid were described. Whatwas needed was an OsO4 analog of the in situ RuO4 generation atroom temperature.Discussion of Results:

During the listserver exchange mentioned above, it wassuggested that sodium or potassium periodate be used for in situgeneration of OsO4. That reaction was stated or hinted at in AdvancedInorganic Chemistry by Cotton & Wilkinson2. On the Internet therewas a strange statement about osmium dioxide. Paraphrasing, "Whenyou opened a bottle of osmium dioxide you could smell traces of

20 • IMCnOSCOPYTODfnf September/October 2002

OsO4." This air oxidation statement suggested that in situ generationof OsO4 should be possible with a more powerful oxidi2er. A review ofmicroscopy textbooks yielded nothing of value about this statement,neither about in situ generation of OsO,,, nor about how to gauge theprogress of the staining reaction with an indicator. This gauge wassomething that was needed. If one stained with fumes from a 2% or4% solution, the reaction times would vary and one risked over-staining thin sections on TEM grids. Not only that, the concentrationofthe fumes would vary with the apparatus volume used. If one usedan opened ampoule to suspend the grids in or over the OsO4, thenair currents could dilute the vapors and give variable results. Clearly,any procedure needed a reaction indicator gauge for either tetroxide.Furthermore, a way was needed to partially control the way thetetroxides were produced and their concentration in a closed systemlike that mentioned above.

The next step was to find a good reaction indicator for OsO,., Anold dilute solution of osmium tetroxide in sulfuricacid from G. FredrickSmith was used to conduct experiments to test for a good OsO,,reaction indicator. It was quickly discovered that OsO,, did not stainthe 665 indicator tape because the tape did not have enough doublebonds, if any.

Having worked on rubber samples, some obvious indicator choicesexisted. One was a gum stock butadiene rubber known as Budene15

1207 (Goodyear Tire & Rubber). It was relatively colorless andtransparent. Natural rubber worked fine, but it was not pure and wascolored depending on the grade used. The Budene 1207 was dissolvedin cyclohexane as a thick saturated solution. This took about a week ofoccasional shaking. This solution could then be spread out on a FisherSuperfrosts microscope slide to form a thin indicator film. Sometimes athicker localized spot of deposited rubber would also be created. ThisBudene indicator gradually turned darker with increased exposure tothe GFS osmium tetroxide solution when placed in the Petri dishapparatus. This was a fairly good indicator but a better commercialindicator might be found by someone else.

Osmium dioxide and osmium chloride were chosen forthe startingosmium materials. It was very clear that In situ generation of OsO4 wasnot going to lead to a cheaper way to make OsO,.. None of this materialwas cheap. On the other hand, one only had to buy a small amount,such as f ve grams for testing.

For oxidizers, I chose 30% hydrogen peroxide, sodium periodateand Clorox® bleach (sodium hypochlorite). Both osmium compoundsproduced a darkening of my indicator and natural rubber pieces in thePetri dish with selected oxidizers. The bad news was that the reactiondid not proceed at a very fast rate. Instead of 1 to 2 minutes, the osmiumcompounds required 1 to 2 hours or longer to stain thin sections using1 to 2 mgs. The good news was that it did stain the double bond indicatorat room temperature. I did not try to heat any ofthe reactants to see if aquicker reaction occurred. I did not want to generate hypochlorous acidfumes by heating bleach, for example.

OsOz reacted very slowly and could take hours or overnight togenerate enough OsO,, to stain grids, OsO;, powder also causedhydrogen peroxide to almost instantly decompose very violently. DONOT use 30% H;O; with OsO3. The other two oxidizers worked fine.OsO2 had some drawbacks. One was that it never reacted totally, andthus one could be left with a dangerously fine powder of OsO2 at theend ofthe reaction to clean up. This could be messy, so use gloves,

OsCl3 reacted with all three oxidizers and did generateOsO4 at a faster rate than OsO2, but at a much slower rate thanRuCl3 generated RuOr OsCl3 did generate enough OsO4 over a 1to 3 hour time period to stain thin sections. One only needed a fewmilligrams to do the job, just like RuCl3. Using more OsCl3 would

Dow

nloaded from https://w

ww

.cambridge.org/core . IP address: 65.21.228.167 , on 12 N

ov 2021 at 03:18:48 , subject to the Cambridge Core term

s of use, available at https://ww

w.cam

bridge.org/core/terms . https://doi.org/10.1017/S1551929500058314

Page 2: In situ Generation of Ruthenium Tetroxide and Osmium

DON'T OPERATE

LINDFOLDED

failure analysis

— TEM sample preparation ;;•

: ::— Gross section investigation w

— Three dimensional structural examination

Use fhe LEO Crossbeam and stop working blindfolded!

Call 1-800-356-1090 today!

Dow

nloaded from https://w

ww

.cambridge.org/core . IP address: 65.21.228.167 , on 12 N

ov 2021 at 03:18:48 , subject to the Cambridge Core term

s of use, available at https://ww

w.cam

bridge.org/core/terms . https://doi.org/10.1017/S1551929500058314

Page 3: In situ Generation of Ruthenium Tetroxide and Osmium

only make dean up and waste disposal more complicated, but itwould speed up the staining. One should only use the amount thatthe apparatus would require to stain the desired grids in a reasonabletime period The Budene 1207 glass microscope slide was used togauge the degree of staining of the thin sections or polymer castfilms. Gloves were used at all times to avoid skin absorption andreaction. Different manufacturer's MSDS sheets were read to get abetter viewpoint of the toxic nature of OsO4, OsC!3 and OsO2. TheseMSDS sheets were a bit scary to read.

In summary, these osmium reactions were truly analogs of theRuOj in situ generation scheme. In situ generation of OsO,, was muchsafer to handle in my opinion than ampoules. Hypochlorite and OsCI3were the preferred materials for staining of thin sections by in situgeneration of OsO .̂

All MSDS sheets for OsOj, OsO2, OsCi3, RuO4, RuCljr etc., wereread. Gloves were required at all times. An unobstructed fume hoodairflow was required. Testing of the hood airflow was done withammonium hydroxide and hydrochloric acid fumes that were used tocreate white smoke at the hood inlet faceDisclaimers:

Osmium compounds are highly toxic, notjust hazardous. Extremecare should be used in handling osmium compounds or rutheniumcompounds. All the precautions and handling precautions for rutheniumand its compounds should apply to osmium and its compounds. Osmiumabsorbs more through the skin and travels further down into the skinthan does ruthenium, MSDS sheets suggest that osmium is morehazardous to use. Read the MSDS sheet(s) in detail. Think about theconsequences of using poor techniques. Nobody can be responsiblefor your handling of these compounds but you. You use them at yourown risk.

Conclusions:1) in situ generation of both RuO4 and OsO4 was possible and prac-

tical for staining things like polymer thin sections, etc..2) Sodium hypochlorite was the easiest oxidizer to use and was eas-

ily obtained.3) RuOa was more hazardous, more widely reactive, and was easier

to generate in situ.A) OsO4 was more toxic, hazardous, and was harder to generate in

situ.5} Use only the smallest amount needed to do the staining job in the

apparatus used.6) The use of a reaction indicator gauge was mandatory,7} In severe cases, like sterically hindered reactions, an indicator

made up of one's sample or its components would be needed forproper stain reaction gauging.

8) Samples were done in batches to avoid repeated exposure andtoxic accumulations.

9} Use these chemicals only if needed.10) It was easiest to generate RuO,, from RuCI3.xH2O. One could gen-

erate OsO,, vapors using OsO2 or OsCI3 but they were relativelyslow to react versus RuCI3.

11) Corn oil, Budene 1207, ethanol, DOP (dioctylphthalate) or someother appropriate tetroxide scavenger should be used at the endof the staining process to react with any excess tetroxides.

12) Safety was number one. Talking with a previous user was veryhelpful. These compounds are hazardous, toxic, poisonous andhad very low exposure limits. The TWA exposure was about 2000times lower than for phosgene gas that could kill. •

References:1) Sawyer Linda C, Grubb David, Polymer Microscopy, Chapman and Hall,

p105to 124, 1987.2) Cotton F. Albert, Wilkinson Geoffrey, Advanced Inorganic Chemistry A

Comprehensive Text, second edition, Interscience, p 1005, 1966,

STOP HASSLING WITHMULTIPLE SERVICE CONTRACTS!

START by putting all of your instruments under one service contract with MAS(regardless of make or model). Our expert EM SERVICE GROUP has the knowledge

and skills to keep your instrument working at its best.

TEMS/SEMS

HITACHI TOPCON

JEOL ISI

AMRAY CAMBRIDGE

PREP EQUIPMENT SPECIAL SERVICES

VACUUM COATERSPLASMA ASHERS

SPUTTER COATERSION MILLS

BEAM BLANKERS

CUSTOM DEVICES

NEW NEW NEW

Authorized service representativesfor Gatan preparation equipmentand Topcon TEM'S.

Contracts and On-DemandEmergency Service atReasonable Rates fromFactory Trained Specialists,

800-421-84513945 Lakefidd Conn Suwnntc, CicaraiaOTM 77O-866-.32UO f:AX'770-K66-3259

6l6Hullc>nSlree£Sijilel0l Raieigh. North Carolina 27606 y 19-S29-7(J4! FAX tflK!95ADVANCED ANALYTICAL PRODUCTS AND SERVICES

22 B (IllCROJCOPY TODJIY September/October 2002

Dow

nloaded from https://w

ww

.cambridge.org/core . IP address: 65.21.228.167 , on 12 N

ov 2021 at 03:18:48 , subject to the Cambridge Core term

s of use, available at https://ww

w.cam

bridge.org/core/terms . https://doi.org/10.1017/S1551929500058314

Page 4: In situ Generation of Ruthenium Tetroxide and Osmium

cflft*

3 Electronics Avenue * Danvers, MA 01923

978-774-5550 • www.amlimaging.com

Dow

nloaded from https://w

ww

.cambridge.org/core . IP address: 65.21.228.167 , on 12 N

ov 2021 at 03:18:48 , subject to the Cambridge Core term

s of use, available at https://ww

w.cam

bridge.org/core/terms . https://doi.org/10.1017/S1551929500058314