improve the chromatographic performance of your assay...
TRANSCRIPT
Improve the chromatographic performance of your assay without changing Improve the chromatographic performance of your assay without changing
method or HPLC equipment
L. Pereira, T. Edge, H. Ritchie, S. Luke
Thermo Fisher Scientific, Manor Park, Runcorn, UK, WA7 1TThermo Fisher Scientific, Manor Park, Runcorn, UK, WA7 1T
Overview ResultsOverview
Purpose: to compare the chromatographic performance of
solid core 4 µm particle packed columns with that of fully
porous 5 and 3 µm particle packed columns.
Efficiency and Pressure
FIGURE 1: Efficiency comparison using Van Deemter plots
for Accucore XL 4 µm and fully porous 5 and 3 µm.
Methods: Isocratic and gradient methods that allow
comparison of chromatographic parameters such as
efficiency, resolution, sensitivity, column backpressure and
for Accucore XL 4 µm and fully porous 5 and 3 µm.
14.0
16.0
18.0
analysis time.
Results: The solid core 4 µm particles provide significant
improvements over fully porous 5 and 3 µm in terms of
separation efficiency, resolution, and sensitivity, with only a
8.0
10.0
12.0
14.0
Pla
te h
eig
ht
(H,µ
m)
separation efficiency, resolution, and sensitivity, with only a
small increase or a decrease in column backpressure for the
5 and 3 µm respectively.
0.0
2.0
4.0
6.0
Pla
te h
eig
ht
(H,
Introduction
In the past decade there has been continuous drive to
0.0
0.0 0.5 1.0 1.5
Accucore XL 4µm
In the past decade there has been continuous drive to
develop chromatographic stationary phases to perform high
efficiency HPLC. One such area is the use of solid core
materials which has gained a lot of interest from
FIGURE 2: Comparison of column pressure for Accucore XL
4 µm and fully porous 5 and 3 µm (150x4.6 mm columns).materials which has gained a lot of interest from
manufacturers and also end users. This type of packing
material has a solid core surrounded by a porous outer layer,
which allows for more uniform particle sizing. When these
4 µm and fully porous 5 and 3 µm (150x4.6 mm columns).
200
250
which allows for more uniform particle sizing. When these
partially porous particles were introduced, the diameter would
typically be between 2 and 3 µm, and these particles provide
similar performance to smaller sub-2 µm fully porous 100
150
Pre
ssu
re (
ba
r)
particles at significantly lower column backpressures.
The very tight particle size distribution of solid core materials
results in columns with high permeability, and therefore for 0
50
Pre
ssu
re (
ba
r)
results in columns with high permeability, and therefore for
the same nominal pressure they give better separations than
fully porous materials. Often however, solid core particle
packed columns are narrow-bore and short which has
0
200 400 600 800
Accucore XL 4µmpacked columns are narrow-bore and short which has
implications in terms of the system set-up. System dead
volume and operating parameters have to be optimised in
order to get the best possible performance out of the column.
On average (across the flow rate range tested) Accucore XL
4 µm gives:order to get the best possible performance out of the column.
The work presented in this poster demonstrates that by using
a new solid core 4 µm particle, packed in conventional
column dimensions, significant improvements in the assay
75% more efficiency than fully porous 5
efficiency than fully porous 3
42% higher pressure than fully porous 5 column dimensions, significant improvements in the assay
performance can be achieved without the need to make
changes to the operating parameters or system set-up. With
the Thermo Scientific Accucore XL 4 µm columns it is FIGURE 3: Performance comparison of Accucore XL 4 µm
lower pressure than fully porous 3
the Thermo Scientific Accucore XL 4 µm columns it is
possible to dramatically improve separation efficiency,
resolution, sensitivity over those obtained with conventional
fully porous 5 and 3 µm particle packed columns. These
1.E+04
t0 (s)
FIGURE 3: Performance comparison of Accucore XL 4 µm
and fully porous 5 and 3 µm using kinetic plots: efficiency
per unit time.
fully porous 5 and 3 µm particle packed columns. These
improvements are obtained with only 40% increase in
backpressure over the 5 µm and a reduction over the 3 µm.
Some reduction in analysis time is also possible.
1.E+03
Some reduction in analysis time is also possible.
1.E+02
Methods
Liquid Chromatography
1.E+01
Columns: Accucore XL C18 4 µm, 150x4.6 mm
Fully porous C18 5 µm, 150x4.6 mm
Fully porous C18 3 µm, 150x4.6 mm
1.E+00
1.E+04
Accucore XL 4µm
FIGURE 4: Performance comparison of Accucore XL 4 µm Fully porous C18 3 µm, 150x4.6 mm
1. Efficiency and pressure testing:
Mobile phase: water / ACN (50:50)
FIGURE 4: Performance comparison of Accucore XL 4 µm
and fully porous 5 and 3 µm using kinetic plots: column
impedance (E) relative to linear velocity (u).1.E+05Mobile phase: water / ACN (50:50)
Flow rates: 0.4 to 2.0 mL/min (every 0.1 mL/min)
Temperature: 30 °C
Imp
ed
an
ce
Detection: UV at 254nm (0.1s rise time, 20 Hz)
Injection volume: 1 µL
Test probe: o-Xylene (t0 marker – theophylline).
1.E+04
Test probe: o-Xylene (t0 marker – theophylline).
2. Resolution, sensitivity and productivity testing:
Mobile phase: A – water; B – ACN
Temperature: 30 °C
Detection: UV at 247 nm (0.1s rise time, 20 Hz)
1.E+03
0
Accucore XL 4µm
Injection volume: 5 µL
Sample: 1. Uracil, 2. Tebuthiuron; 3. Metoxuron; 4. Monuron;
5. Chlorotoluron; 6. Diuron; 7. Linuron
Impedance is a term that gives a true measure of the
performance of the column as it incorporates efficiency, time
and pressure, which are critical parameters of a separation.
Accucore XL 4µm
• Standard gradient: 35 to 60% B in 10 min
Flow rate: 1.0 mL/min
• Scaled gradient: 35 to 60% B in 7.5 min
and pressure, which are critical parameters of a separation.
Lower impedance values indicate faster and higher efficiency
separations performed at lower backpressures.
On average (across the flow rate range tested) Accucore XL • Scaled gradient: 35 to 60% B in 7.5 min
Flow rate: 1.3 mL/min
(optimum flow from van Deemter plot)
On average (across the flow rate range tested) Accucore XL
4 µm gives:
59% more efficiency per unit time than fully porous 5
and 53% more efficiency per unit time than fully porous 3 (optimum flow from van Deemter plot)
• Optimized gradient: 35 to 60% B in 4 min
Flow rate: 1.3 mL/min
and 53% more efficiency per unit time than fully porous 3
(Figure 3).
79% lower impedance than fully porous 5
lower impedance than fully porous 3 lower impedance than fully porous 3
Improve the chromatographic performance of your assay without changing Improve the chromatographic performance of your assay without changing
Thermo Fisher Scientific, Manor Park, Runcorn, UK, WA7 1TThermo Fisher Scientific, Manor Park, Runcorn, UK, WA7 1T
Efficiency comparison using Van Deemter plots
for Accucore XL 4 µm and fully porous 5 and 3 µm. FIGURE 5: Comparison of the resolution of a critical pair
(peaks 2 and 3) on fully porous 5 and 3 µm and Accucore
Resolution, Sensitivity and Productivity
for Accucore XL 4 µm and fully porous 5 and 3 µm. (peaks 2 and 3) on fully porous 5 and 3 µm and Accucore
XL 4 µm. Standard gradient conditions.
Fully porous 5 µmRs = 2.47
Fully porous 3 µmRs = 2.82
Accucore XL 4 µmRs = 3.13
1.5 2.0 2.5 3.0 3.5 4.0 4.5
Linear velocity u (mm/s)
Fully porous 5µm Fully porous 3µm Minutes0 1 2 3 4 5 6 7 8 9 10
Comparison of column pressure for Accucore XL
4 µm and fully porous 5 and 3 µm (150x4.6 mm columns).
FIGURE 6: Comparison of average signal-to-noise ratios
(S/N) on fully porous 5 and 3 µm and Accucore XL 4 µm
for 1 ng of each solute loaded on column. Scaled 4 µm and fully porous 5 and 3 µm (150x4.6 mm columns). for 1 ng of each solute loaded on column. Scaled
gradient conditions.
Fully porous 5 µmS/N = 4.9
Fully porous 3 µm
S/N = 10.6
S/N = 5.3
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00
Accucore XL 4 µmS/N = 10.6
1000 1200 1400 1600 1800 2000
Flow rate (µL/min)
Fully porous 5µm Fully porous 3µm
Minutes0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00
FIGURE 7: Productivity comparison for fully porous 5 µm
and Accucore XL 4 µm.
On average (across the flow rate range tested) Accucore XL
Fully porous 5 µm
Rs = 2.47 Rt = 8.62
Standard gradientRs = 3.13
75% more efficiency than fully porous 5 µm and 50% more
efficiency than fully porous 3 µm (Figure 1).
42% higher pressure than fully porous 5 µm and 13% Rs = 3.02
Standard gradientRs = 3.13
Rt = 6.56
Accucore XL 4 µm
Standard gradient
Performance comparison of Accucore XL 4 µm
lower pressure than fully porous 3 µm (Figure 2).
Accucore XL 4 µm
Rs = 2.99
Rs = 3.02
Rt = 5.07
Scaled gradient
Performance comparison of Accucore XL 4 µm
and fully porous 5 and 3 µm using kinetic plots: efficiency
Accucore XL 4 µm
Rs = 2.99Rt = 4.06
Optimized gradient
The higher efficiencies of the solid core 4 µm material over
the fully porous 5 and 3 µm materials results in improved
Minutes0 1 2 3 4 5 6 7 8 9 10
the fully porous 5 and 3 µm materials results in improved
resolution and sensitivity:
27 and 11% higher resolution of the critical pair (Figure 5)
119 and 100% higher signal/noise ratios (Figure 6)119 and 100% higher signal/noise ratios (Figure 6)
The higher efficiencies of the Accucore XL 4 µm over a wide
flow rate range can be used to reduce analysis time by
optimizing flow rate and gradient. In Figure 7, a reduction of Efficiency (/) optimizing flow rate and gradient. In Figure 7, a reduction of
run time by half is demonstrated, still with improved
resolution of the critical pair.
1.E+05
Fully porous 5µm Fully porous 3µm
Efficiency (/)
Performance comparison of Accucore XL 4 µm
Conclusion
The new Accucore XL 4 µm solid core columns provide
Performance comparison of Accucore XL 4 µm
and fully porous 5 and 3 µm using kinetic plots: column
impedance (E) relative to linear velocity (u).The new Accucore XL 4 µm solid core columns provide
improvements in efficiency in excess of 50% over fully porous
5 and 3 µm columns
The backpressure of the new Accucore XL 4 µm solid core η20
N
PtE
∆=
columns is moderately higher than fully porous 5 µm, but
lower than fully porous 3 µm
The new Accucore XL 4 µm solid core columns exhibit
η2N
significantly lower impedance than fully porous materials
Significant improvements in resolution and signal-to-noise
ratios were demonstrated with the Accucore XL 4 µm material
The Accucore XL 4 µm columns produce the improvements
listed above with no changes to methodology or HPLC
system set-up. The transfer to Accucore XL 4 µm is quick and
easy
3
Fully porous 5µm Fully porous 3µm
u0 (mm/s)
easy
Impedance is a term that gives a true measure of the
performance of the column as it incorporates efficiency, time
and pressure, which are critical parameters of a separation. Additional information
Fully porous 5µm Fully porous 3µm
and pressure, which are critical parameters of a separation.
Lower impedance values indicate faster and higher efficiency
separations performed at lower backpressures.
On average (across the flow rate range tested) Accucore XL
Additional information
For additional information, please visit our Chromatography Resource Centre which
can be found at: www.thermoscientific.com/CRC
©2012 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property On average (across the flow rate range tested) Accucore XL
59% more efficiency per unit time than fully porous 5 µm
and 53% more efficiency per unit time than fully porous 3 µm
of Thermo Fisher Scientific Inc. and its subsidiaries . Specifications, terms and pricing
are subject to change. Not all products are available in all countries. Please consult
your local sales representatives for details.
PSCCS 1512 0812and 53% more efficiency per unit time than fully porous 3 µm
79% lower impedance than fully porous 5 µm and 72%
lower impedance than fully porous 3 µm (Figure 4).lower impedance than fully porous 3 µm (Figure 4).