implementation of the laser scanning system used for dermatological treatments

70
國立彰化師範大學 國立彰化師範大學 國立彰化師範大學 國立彰化師範大學機電工程學系 機電工程學系 機電工程學系 機電工程學系 碩士論文 碩士論文 碩士論文 碩士論文 雷射掃瞄皮膚治療系統之研製 Implementation Of The Laser Scanning System Used For Dermatological Treatments 生:郭 指導教授:陳 教授 A Thesis Submitted to the Department of Mechatronics Engineering National Changhua University of Education in Partial Fulfillment of the Requirements For the Degree of Master of Science July, 2007

Upload: joey-kuo

Post on 11-Apr-2015

299 views

Category:

Documents


4 download

DESCRIPTION

This thesis discusses the implementation of laser scanning system used for dermatological treatments. To obtain the designated therapeutic area this system uses a photographic device and then transfers the image position to a driver device. Selecting the boundaries of the designated treatment area is done either manually or automatically. After calculating the chosen area, the system divides the area into many therapeutic “points” (the actual number of points depending on the laser spot size). In the meantime, the system also determines the coordinates of every therapeutic point. The brand-new method directs the laser beam to every designated point to perform the treatment process. The operator does not need to operate manually at all. Consequently, the flaws inherent with manual operations can be avoided, and the therapeutic area can be controlled with greater precision. Moreover, by utilizing the image-orientated control system, the energy of laser is distributed over the skin steadily and evenly. Another advantage of this system is that it avoids laser exposure to the operator’s eyes.

TRANSCRIPT

Page 1: Implementation Of The Laser Scanning System Used For Dermatological Treatments

國立彰化師範大學國立彰化師範大學國立彰化師範大學國立彰化師範大學機電工程學系機電工程學系機電工程學系機電工程學系

碩士論文碩士論文碩士論文碩士論文

雷射掃瞄皮膚治療系統之研製

Implementation Of The Laser Scanning System Used

For Dermatological Treatments

研 究 生:郭 奕 谷 撰

指導教授:陳 明 飛 教授

A Thesis Submitted to the Department of Mechatronics Engineering

National Changhua University of Education

in Partial Fulfillment of the Requirements

For the Degree of Master of Science

中 華 民 國 九 十 六 年 七 月

July, 2007

Page 2: Implementation Of The Laser Scanning System Used For Dermatological Treatments

ii

摘要摘要摘要摘要

本論文研發雷射光掃瞄系統來做皮膚方面的治療。此系統使用影像擷

取裝置來取得所欲治療範圍後,再轉換至驅動裝置定位。可以用手動或自

動方式來選擇治療區域的範圍。所選擇區域經過計算後,系統會將其分割

成許多的“點”(視雷射光點大小);並在此同時,系統亦決定每個治療點

的座標位置。此全新的治療方式將雷射光導引至每個被選擇的點來進行治

療。使用者完全不必用手操作治療病人;因此,在此文章內所提的手動治

療失誤將可以避免,治療區域也因此可以控制的更準確。此外,利用影像

導引控制系統,雷射能量可以穩定,平均地分配在皮膚上。另外一個好處

是使用者的眼睛可以避免曝露在雷射光之下。

Page 3: Implementation Of The Laser Scanning System Used For Dermatological Treatments

iii

Abstract

This thesis discusses the implementation of laser scanning system used

for dermatological treatments. To obtain the designated therapeutic area this

system uses a photographic device and then transfers the image position to a

driver device. Selecting the boundaries of the designated treatment area is

done either manually or automatically. After calculating the chosen area, the

system divides the area into many therapeutic “points” (the actual number of

points depending on the laser spot size). In the meantime, the system also

determines the coordinates of every therapeutic point. The brand-new method

directs the laser beam to every designated point to perform the treatment process.

The operator does not need to operate manually at all. Consequently, the flaws

inherent with manual operations can be avoided, and the therapeutic area can be

controlled with greater precision. Moreover, by utilizing the image-orientated

control system, the energy of laser is distributed over the skin steadily and

evenly. Another advantage of this system is that it avoids laser exposure to the

operator’s eyes.

Page 4: Implementation Of The Laser Scanning System Used For Dermatological Treatments

iv

CONTENTS

Abstract (Chinese) .............................................................................................. ii

Abstract (English) ............................................................................................. iii

Contents............................................................................................................. iv

List of Tables ....................................................................................................... vi

List of Illustrations ........................................................................................... vii

Chapter 1 Introduction .................................................................................. 1

1.1 Preface ................................................................................................ 1

1.2 Background ......................................................................................... 6

1.3 Patent review ........................................................................................ 8

1.4 Purpose of this study .......................................................................... 9

1.5 Structure of this study ....................................................................... 11

Chapter 2 The Laser System ....................................................................... 12

2.1 Laser source ...................................................................................... 12

2.2 Image system .................................................................................... 14

2.3 Optical compensation system ........................................................... 16

2.4 Scanning system ............................................................................... 22

Chapter 3 Galvo Scanner and Beam Alignment ........................................... 26

3.1 Galvanometer ................................................................................... 26

3.2 The mirrors ....................................................................................... 27

3.3 The driver board ............................................................................... 29

3.4 Alignment procedure ........................................................................ 29

Chapter 4 Assembly and Verification ......................................................... 33

4.1 Assembly concept ............................................................................. 33

4.2 Assembly procedure ......................................................................... 35

4.3 Verification ......................................................................................... 36

4.4 Target area control .............................................................................. 41

Chapter 5 Conclusion .................................................................................. 48

5.1 Feature of this device ........................................................................ 48

5.2 Future development .......................................................................... 49

Page 5: Implementation Of The Laser Scanning System Used For Dermatological Treatments

v

Appendix A Other dermatology laser patent analysis list ........................... 52

Appendix B Wavelight laser technology patents ........................................ 55

Appendix C InPro Innovations patents ....................................................... 57

References ...................................................................................................... 61

Page 6: Implementation Of The Laser Scanning System Used For Dermatological Treatments

vi

LIST OF TABLES

Table Page

1.1 Reference list of market product ............................................................ 2

1.2 Types of human skin .............................................................................. 5

3.1 Mirror manufacturer ............................................................................. 28

4.1 Parts description ................................................................................... 37

Page 7: Implementation Of The Laser Scanning System Used For Dermatological Treatments

vii

LIST OF ILLUSTRATIONS

Figures Page

1.1 Skin structure ........................................................................................... 4

1. 2 Wavelength absorption ............................................................................ 4

2.1 Optics compensation & scanning system .............................................. 18

2.2 The astigmatic aberration waist ties the position and the circular......... 21

2.3 Gauss light beam relations after lens function ..................................... 22

3.1 Beam alignment ................................................................................... 32

4.1 The 3-D structure of this new invention .............................................. 38

4.2 The 3-D structure of another direction ................................................ 39

4.3 The top view of movement situation 1 ................................................ 40

4.4 The top view of movement situation 2 ................................................ 40

4.5 The system schematic diagrams .......................................................... 41

4.6 Flow chart ............................................................................................ 42

Pictures Page

1.1 Treatment cabinet ................................................................................... 6

1.2 Lighting boxes ....................................................................................... 6

1.3 Hand probe ............................................................................................. 7

1.4 Mechanical arm ...................................................................................... 7

4.1 Auto-control scanning system – view 1 ............................................... 33

4.2 Auto-control scanning system – view 2 ............................................... 34

4.3 Small circle calibration ........................................................................ 43

4.4 Large circle calibration ........................................................................ 43

4.5 Cross calibration .................................................................................. 44

4.6 Treatment area calibration ................................................................... 44

4.7 Operation demo .................................................................................... 45

4.8 Colorful psoriasis – test 1 ...................................................................... 46

4.9 Colorful psoriasis – test 2 .................................................................... 46

Page 8: Implementation Of The Laser Scanning System Used For Dermatological Treatments

1

Chapter 1 Introduction

1.1 Preface

Lasers have been used in dermatology for decades since the 1960’s, and are

one of the benefits of high technology, but they still frighten many people.

This fear is mainly due to a lack of understanding of lasers, especially the

principles of laser treatment. Basically there are four interactions between light

energy and skin tissue: they are photoablation, thermal, electromechanical, and

photochemical [1]. Specifically, the treatment of psoriasis [2] mainly employs

photochemical interaction theory. It involves a biological effect that is directly

proportional to the exposure of the total energy dose over a specific area of skin,

no matter whether continuous or fractionated, and it’s light-induced reactions

are sensitive to wavelength.

In fact, a laser emits one kind of light energy. By lighting the skin surface,

the skin is affected by the reflection, scattering, penetration, and absorption of

that energy [3, 4]. By using a specific wavelength of light, a laser reaches the

designated tissue depth. Then the thermal energy, which is absorbed and also

transformed by the cells, destroys and activates the tissue to regrow and rebuild

the treated area. In this way, some level of therapy can be achieved.

Lasers have been used for medical dermatology applications such as the

removal of port wine stains, acne scars, tattoos, dark spots, and other blemishes

for over ten years. Lasers are also used for a growing number of cosmetic

procedures including teeth whitening, hair removal, and the treatment of

wrinkles.

Page 9: Implementation Of The Laser Scanning System Used For Dermatological Treatments

2

In order to present a clear picture of laser therapy, and to get some idea of

its possible applications, it is helpful to mention products that are easily obtained

from the laser market. A reference list of marketed laser products is shown in

Table 1.1 with the manufacturer, the product name, specifications, a photo of the

product, product function, classification, and price provided.

Some parts of the skin’s structure (see Figure 1.1) can absorb the high

energy from the stable wavelength of a laser (see Figure 1.2) [5] and react in the

twinkling of an eye, so the designated treatment result can be obtained due to the

destruction of one specific element in the skin.

Table 1.1 Reference list of market products

Manufacturer Healiohealth LazrPulsr ETrans

Product name LTU 1000 LazrPulsr 4X Softlaser

Specification 670nm, 1mW 635nm, 20mW 670nm, 6mW

Photo

Function Pain healing Wounded recovering Rejuvenating

Classification Class II Class II Class II

Price(USD) 690 4,995 129

Page 10: Implementation Of The Laser Scanning System Used For Dermatological Treatments

3

Table 1.1 - Continued

Manufacturer Beauty Beauty Beauty

Product name SI-808 L-Dr.890 HFL850

Specification 808nm 890nm 850nm/LD,

660nm/LED

Photo

Function Hair removing Whitening Rejuvenating

Classification Class III R Class III R Class II

Price(USD) 200 210 --

Manufacturer Omega Beurer Win/Health

Product name Visible red probe SL30, VSL40 HairMax

Specification 660nm, 50mW 635-675nm, 6mW 655nm, 4mW

Photo

Function Hair removing Whitening Hair growing

Classification Class III B Class III A Class III R

Price(USD) -- 159 346

Page 11: Implementation Of The Laser Scanning System Used For Dermatological Treatments

4

Figure 1.1 Skin structure

Figure 1.2 Wavelength absorption

The laser lighting time is quite short; thus, the energy will not be spread to

Page 12: Implementation Of The Laser Scanning System Used For Dermatological Treatments

5

the surrounding tissue, protecting harmless skin tissue from destruction.

However, different skin colors have different results from the same energy.

Traditionally, for laser treatment reference, all human skins are classified into

six types (see Table 1.2) [6] from light to dark in color. The doctor chooses the

proper laser energy level or energy dose based on the patient’s skin type to

prevent over lighting or over lasing the skin.

Table 1.2 Fitzpatrick Classification Scale

Types Interaction with Sun burn Skin Eyes Hair

TYPE

I

Always burns in the sun. Never

gets tan.

Extremely

fair skin

Blue/green

eyes

Usually blonde

or red hair

TYPE

II

Sometimes burns, but it will

turn into a tan. Most

Caucasians

Fair skin Green/brown

eyes

Light brown to

brown hair

TYPE

III Usually tans, rarely burns.

Medium

skin Brown eyes

Light brown to

brown hair

TYPE

IV

Seldom burns.

Possibly Mediterranean, Asian,

or Hispanic descent.

Olive skin Black eyes Brown or black

Hair

TYPE

V

Almost never burns.

Possibly Mediterranean,

Hispanic, light skin

African-American, or other

descent.

Dark

brown skin Black eyes Black Hair

TYPE

VI

Never burns.

Possibly Indian, West Indian

descent, or African-American.

Black skin Black eyes Black Hair

Page 13: Implementation Of The Laser Scanning System Used For Dermatological Treatments

6

For psoriasis lesions, traditional treatment methods is using 311~313nm

wavelength lamps installed in one cabinet (in Picture 1.1) [7] or lighting boxes

(in Picture 1.2) [8]. Patients step into the cabinet exposing the psoriasis lesions

to the light [9, 10, 11, 12] after taking vitamin A acid, Retinoid; or wiping the

steroid Anthralin, Tar oil, on their skin [13]. However, most of the traditional

treatments have side effects and require about 25 to 30 procedures per treatment.

The worst part is that these treatment methods have to be changed after two or

three months in case the human body builds up an immunity to the medication

Picture 1.1 Treatment cabinet Picture 1.2 Lighting boxes

1.2 Background

While dermatologists are professionally trained, it is noted that current

308nm excimer laser skin therapeutic devices all need to use manual probes

(in Picture 1.3) [14] or mechanical arms (in Picture 1.4) [15]. The steps of a

manual probe or mechanical arm operation are the same. First, both the

Page 14: Implementation Of The Laser Scanning System Used For Dermatological Treatments

7

designated treatment area and its on-screen target area are determined by the

doctor’s naked eyes; second, the designated treatment area is covered by moving

the laser projecting probe by hand; and, finally, the doctor manually controls

when the laser is projected to perform the treatment process.

Based on these steps, it is easy to find that a lengthy, continuous surgery

would probably result in some inaccuracies and side effects from the doctor; for

example, torpor of hand movement, improper control of muscles, eye fatigue,

etc. These flaws might do unexpected harm on the healthy tissues and cells

around the desired treatment area. Moreover, re-treatment of the same area due

to unavoidable human forgetfulness, or miscalculation of laser projection times,

would both affect the treatment result.

Picture 1.3 Hand probe Picture 1.4 Mechanical Arm

Because manual probes and mechanical arm operations are not convenient

Page 15: Implementation Of The Laser Scanning System Used For Dermatological Treatments

8

or user friendly, and because operation times are too long for mild or medium

psoriasis lesions, these treatment methods are not efficient for either doctor or

patient. Therefore, an auto-controlled scanning method was created. X, Y

axes galvanometers with coated mirrors were installed to precisely position the

laser light. A CCD camera and touch screen panel are used by the doctor to

monitor the operation. A footswitch controls when the light is projected on the

psoriasis lesions. This method shortens operation time and also solves

problems due to human flaws from both the doctor and the patient’s skin.

1.3 Patent review

Because the auto-controlled laser scanning system is claimed as a

dermatology laser product invention, a patent review can be discussed from two

main competitors: Wavelight and PhotoMedex. It is noted the company InPro

Innovations is inactive in the medical device field.

There are 2,985 dermatology-related patents worldwide, including 33 cases

of dermatological laser devices and treatment methods (See Appendix A).

Most of the dermatology laser treatment device patents for psoriasis on the

market are from PhotoMedex and Wavelight, and a very small percentage are

from Inpro Innovations. Hence, this patent review focuses it’s analysis on

these three companies with the results presented below.

1. Wavelight Laser Technology has 22 patents (see Appendix B), but only

two relevant patents to dermatology devices: (1) a lamp used in the treatment of

skin (DE), and (2) a laser device for the treatment of patient skin and other

dermatological processes (DE). These patents are not relevant to the

Page 16: Implementation Of The Laser Scanning System Used For Dermatological Treatments

9

auto-controlled scanning system.

2. PhotoMedex has only two patents and seems focused on handheld optical

fibers to control laser energy: (1) controlled delivery doses of ultraviolet light

for treating skin disorders (WO02055149-2002-07-18), a patent that claims the

scope of using handheld fibers to control the treatment doses, and (2) rare

gas-halogen excimer lasers with baffles (EP1564851-2005-08-17). These

patents are not related to the auto-controlled scanning system.

3. InPro Innovations GmbH is in the laser welding industry and has no

patents concerning the auto-controlled scanning system (See Appendix C).

1.4 Purpose of this study

Psoriasis has been treated successfully with narrow-band UVB radiation

for twenty years [16, 17]. Now, with the excimer laser system, it is possible

to selectively apply ultraviolet radiation at a wavelength of 308nm [18]. This

concept of selectivity allows only the diseased skin to be exposed to the

radiation while protecting the healthy skin from contact, making it possible to

work with higher laser treatment doses at one time. As a result, it is possible

to achieve effective clearing of psoriasis with long periods of remission [19].

In comparison to conventional forms of treatment, the new method, using the

auto-controlled scanning system rather than a handheld optical fiber, offers

better results and requires fewer treatment sessions.

There is no real alternative to excimer laser therapy for the treatment of

Vitiligo dermatitis. Given its extensive nature, conventional forms of

Page 17: Implementation Of The Laser Scanning System Used For Dermatological Treatments

10

treatment demand an unusual amount of forbearance and discipline on the part

of patients. Often, treatments end with unsatisfactory results. On the other

hand, the excimer laser system gives the patient the opportunity to drastically

reduce the number of required treatment sessions. Re-pigmentation is

achieved via stimulation of the melanocytes with relatively high doses of

lasing energy. Here, the selective application made possible by the new

design system is a decided advantage.

Therefore, based on the area designated for destruction and the choice of

proper laser type and proper energy, the operator could destroy the target at will

without damaging other harmless or healthy tissues. That is to say, treatment

with a laser is quite sophisticated and safe under such conditions. Using the

selective excimer laser wavelength of 308nm is the ultimate treatment for

Psoriasis and Vitiligo dermatitis. However, 308nm excimer laser market

products use optical fibers to deliver laser energy onto the designated skin

lesions by either a handheld device or mechanical arm. The operator has some

inconvenience and the patient also suffers the risk of having the same area

unintentionally re-treated, or laser projection times that are too long. The

auto-controlled scanning system provides the best solution to these problems

and reduces human flaws to minimize the risk of excessive laser exposure.

The purpose of this thesis is to show how to build the auto-controlled

scanning system, and to analyze and explain the concepts behind the system.

Page 18: Implementation Of The Laser Scanning System Used For Dermatological Treatments

11

1.5 Structure of this study

The information needed to show how to build the auto-controlled scanning

system, and to analyze and explain the concepts behind the system, are

presented in the following chapters.

Chapter 2 discusses the laser device used for the auto-controlled scanning

system. The structure of different components and explanations of each

function are given that: the laser source control and the laser power monitoring

system, the image system technology, the optical compensation system, and

the scanning system. Chapter 3 explains the galvo based scanner and

introduces its key components. The procedures and techniques of beam

alignment are then shown there after. In Chapter 4, the assembly and

verification methods used to prove that the auto-controlled scanning system is

constructed and performs according to the original design are provided.

Chapter 5 discusses the unique features of the auto-controlled laser

scanning system and possible future development trends, discussing improved

function and effectiveness from hardware and software upgrades. In the

mean time, the scanning system can also be used in other industries or for

other purposes.

Page 19: Implementation Of The Laser Scanning System Used For Dermatological Treatments

12

Chapter 2 The Laser Device

The laser system used in this thesis contains laser source control and a

laser power monitoring system, a laser spot scanning and optics compensation

system, a skin image identification and skin color information database

construction, a position mark to establish a correlation technology control

system, and a clinical experimental and treatment parameter construction. An

explanation of the research technique follows.

2.1 Laser source

Traditional dermatologists treated skin conditions like psoriasis and

Vitiligo dermatitis with medicines like tazarotene, calcipotriene, and so on.

The treatment effects, however, were not remarkable and there were obvious

side effects. Therefore, in modern times a method of treatment using UV

lamp lighting was developed. The principle was similar to treatment by

radioactive rays. Because the light illumination method was dependant on

the wavelength, the intensity, and the target area, control was not easy;

therefore, the treatment effect was not entirely positive. The treatment course

was excessively long and the reoccurrence probability was high.

Nearly a decade of research and development resulted in recognizing the

positive effects that UVB wave band 311 nanometer ultraviolet rays have in

the treatment of skin diseases such as psoriasis, particularly by enclosing the

target area. However, the delivery of the illumination was still too broad for

Page 20: Implementation Of The Laser Scanning System Used For Dermatological Treatments

13

precisely controlled treatments, and it was easy to harm the healthy

surrounding tissues. The 308nm excimer laser was found to have the single

wavelength with the most direct effect on skin tissue; therefore, starting from

1998, the United States focused its efforts on researching this wavelength and

developing 308nm laser treatment devices. Germany followed the research

and developed devices after the US results were published.

By directly and selectively destroying unhealthy tissues, the 308nm

excimer laser greatly improves the treatment course and the curative effect

over conventional UV lamp lighting technology, and by avoiding the UV

lamp’s large illumination area the probability of damaging healthy skin is

reduced [20]. Gradually, this laser has entered the mainstream for treating

skin psoriasis.

The development of the 308nm excimer laser for skin treatment is now a

priority project for several countries. Because this thesis utilizes an optical

delivery system, the laser needs a lower power than other kinds of laser market

products. Since the excimer laser has short-term stability, the laser power

monitoring system is used to maintain the laser power and test the laser power

stability. In this thesis, we use the feed-forward control concept. By using a

power meter to gauge laser power, the design makes it possible to control laser

power through an attenuator mirror, allowing the power to be maintained at a

set value. In addition to the power stabilizing control, this thesis will focus

on the amount of laser power adjustments needed for each kind of skin disease

to establish laser power reference values that help the doctor set up the

treatment parameters.

Page 21: Implementation Of The Laser Scanning System Used For Dermatological Treatments

14

The laser beam output delivery is of the pulse wave type. Generally

speaking, the shorter the laser pulse time, the more centralized the energy.

Therefore, for the same laser beam energy, a shorter pulse time means its

power is higher. This thesis will use a laser with a short pulse duration of

10ns (nanoseconds) to serve the function of selectively treating only the target

area.

In addition, part of this system develops a database of laser power

parameters for skin treatment. In this part, people of various skin types are

asked to accept a designed laser energy test, called the minimal erythema dose

(MED). The MED tests for the appearance of erythematous skin. The test

results will provide the doctor with a reference on which to base adjustments

of the laser dosage.

2.2 Image system

This thesis intends to develop a treatment method that picks up the image

of the skin disorder, recognizes the area of the disorder, and uses a localized

luminous spot with a base matrix to determine image position. This

information is then combined with a scanning system to precisely deliver laser

energy on the designated location. This part of the thesis will establish the

target acquisition technology, the conformity of the image with disorder shape

identification, the treatment zone scanning coordinate transformation and the

control system, establish the different skin color identification information

database, and so on.

1. The target acquisition technology maintains alignment between the

Page 22: Implementation Of The Laser Scanning System Used For Dermatological Treatments

15

laser’s scanning coordinates and the image coordinates from the treatment area.

The thesis’s proposed direction of development will use a standard grid with

the scanning treatment scope to align the central point of the treatment area

with a laser scanning reference point for precision alignment of the laser beam

with the target.

2. The conformity of the image of the skin with the disorder shape

identification and the image scanning coordinate transformation system

primarily establish the image center and the laser spot scanning system

reference point to a zero point localization. The image is strengthened by

using a filter, automatic binaryzation and disorder region edge detection.

Then the image scanning coordinates values are transformed to the laser beam

scanning system coordinate values in order to continue with the skin treatment.

Because the pick up image size is bigger than the laser treatment scope, a

working space is designed in the software to treat a square of the region.

When actual treatment starts, the region of disorder must be present in the

square to be identified. In addition, the doctor may also directly touch the

screen to revise the image detected region or use a mouse to select the

treatment.

For the laser scanning field depth position and the focusing image to

determine the position, this thesis plans to install an infrared range detection

sensor to measure the designated treatment region within the CCD distance.

This may further confirm the field depth position, to help achieve automation

and precise treatments.

Page 23: Implementation Of The Laser Scanning System Used For Dermatological Treatments

16

3. With regard to the different skin colors, it might be necessary to

establish a skin treatment image identification information database. This

thesis will obtain image parameter settings by collecting information about

different skin colors and pictures of the different skin diseases. A basic

parameter information database will be constructed to provide a reference to

the doctor for the laser treatment.

4. Most skin diseases are often distributed over disperse regions, and are

also distributed over three-dimensions. With regards to this aspect of

treatment, two kinds of techniques will be adopted. The first technique has

the doctor to mark a symbol on the center of the therapy region, and use the

image servo-control technology to move the platform to scan the image’s

central point until it matches the symbol alignment before the treatment starts.

The second technique divides the region of the skin disorder, using

medical-use asphalt wiped on the skin to make sections for alignment. This

might enhance skin disorder identification and might distinguish clearly the

treatment regions from the non-treatment regions. This sectional treatment

technique might also help avoid the spot from becoming elliptical in shape due

to slope deviations. which might cause healthy skin to be treated.

2.3 Optics compensation system

A diagram of the optics compensation part of the system is shown in

Figure 2.1. Laser light is sent by the laser source module to the optics

compensation system to maintain the spot shape and size after adjustments by

the power control system feedback. The laser light then enters the scanning

Page 24: Implementation Of The Laser Scanning System Used For Dermatological Treatments

17

system on its way to perform treatments on skin.

The design of the optics compensation system takes into consideration

scan speed, spot shape, size, power density, the two-dimensional scanning

mirror and the compensating mirror module, to handle a small spot with high

power density, and a large field depth illumination window (100mm x 100mm).

The purpose of two-dimensional scanning and the optics mirror module is to

eliminate aberrations.

Page 25: Implementation Of The Laser Scanning System Used For Dermatological Treatments

18

Figure 2.1 Optics compensation & scanning system

This thesis’s pulse wave excimer laser uses XeCl premix gas as the active

medium to generate a wavelength of 308nm. Based on this laser’s special

resonant cavity optical design, the laser output has a beam with an astigmatic

aberration. The divergence angle is 1mrad x 0.5mrad. The light beam spot is

an ellipse out of the pupil aperture. The beam sizes for the long and short

axis are 6mm x 3mm, and may be expressed as:

Image System

Laser source and

Laser power monitor

Optics

Compensation

System Scanning System

Projecting windows

Page 26: Implementation Of The Laser Scanning System Used For Dermatological Treatments

19

−×

×

++−×

−×

−×

×=

−−

2

0

1

2

0

1

22

2

2

2

221

00

0

2

1

2

1

22

yx

yX

yx

/

yx

W

)az(tanjexp

W

ztanjexp

R

y

R

xzjkexp

)z(W

yexp

)z(W

xexp

)z(W

W

)z(W

WE)z,y,x(E

πλ

πλ

+=

22

01z

WzR x

x λπ

; ( )( )

−+−=

22

01

az

WazR

y

y λ

π;

( )

+=

2

2

0

2

0

2 1x

xxW

zwzw

πλ

; ( ) ( )

−+=

2

2

0

2

0

2 1y

yyW

azwzw

πλ

;

x

xw0πλ

θ = , y

yw 0πλ

θ =

in which λ is the wavelength, w0x and w0y are the waist spot radii, θx and θy are

the x direction and y direction divergence angles, α is the astigmatic aberration

distance, corresponding in the x-z plane beam to the y-z plane beam distance.

Traditionally, the astigmatism ellipse will be reshaped to a circular or

square shape by using one pair of anamorphic prism lenses, collimation optics,

and a cylinder mirror set of lens combinations. Moreover, to find the suitable

position to join the astigmatic aberration compensation part and the

collimation mirror set might also create a parallel light effect. Because of the

astigmatic aberration, the light beam in some position should be circular,

( ) ( )zwzwyx

=

Page 27: Implementation Of The Laser Scanning System Used For Dermatological Treatments

20

By substituting these in the above Gaussian light beam equation, it should

be possibly to get position z,

( )

−+±

−=

22

0

2

02

2

2

00

2

02

0

2

0

21

1yxyxx

yx

,wwawwaw

wwz

λπ

The above equation represents the two positions of the beam spot as

circular, the position of long and short axis is reciprocal, but the divergence

angle is invariable. Inserting a suitable astigmatic aberration optics part in

this position should make it possible to shape the light beam as shown in

Figure 2.2, transforming an elliptical Gaussian light beam into a full circle,

symmetrical Gaussian beam.

The excimer laser divergence angle of 1mrad x 2mrad is approximately in

parallel. It is possible to enhance the parallelism, but the price must then be

paid in optical width, and this is not really worth doing. But the field depth

length corresponds to the optics mirror group equivalent to the focal size;

therefore, the optical design must be within the field depth and between the

choice illumination spot sizes. The field depth ∆Ζ computation may be

decided by light beam radius wx and the Z relations.

Accordingly, the Gaussian light beam and the lens relationship may result

in the equation below by the ABCD matrix representation:

Page 28: Implementation Of The Laser Scanning System Used For Dermatological Treatments

21

W0x

W0y

z1z2

Figure 2.2 The astigmatic aberration waist ties the position and the circular

symmetry Gauss light beam position

( )

2 2

2 0 1

20 2 2

0 1

1

x x

x

x

x x

f WW

Wf z

πλ

=

− +

;

( )

22

2 0

1 0 1

22 2

0 1

1

x

x x x x

x

x

x x

Wf z W f

zW

f z

πλ

πλ

=

− +

As shown in Figure 2.3, this may estimate image point position Z2x and

the equivalent focal lens ƒχ, according to the above equation. It will be

slightly complex if this is directly by the Gaussian light beam and the ABCD

matrix transformation method. Not only is it not possible to design the optics

mirror set directly, but, also, there is no off-the-shelf analysis software

available.

If the light beam divergence angle is not too big, it may be thought of as

the near axis optics. Therefore, to design the optics compensation system it is

possible to use the available principal ray and marginal ray to solve the

problem. In this thesis, a conventional method of light tracing is used to

design the optics system. Generally speaking, using geometrical optics

software, and in view of three types of optics parts, it is possible to design and

optimize the performance within a short period of time.

Page 29: Implementation Of The Laser Scanning System Used For Dermatological Treatments

22

Z1x Z2x

W0x1 W0x2

Figure 2.3 Gauss light beam relations after lens function

2.4 Scanning system

At present the market laser products for the treatment of psoriasis lesions

primarily use optical fibers. Because the optical fiber projecting area and the

optical fiber delivery port is numerical aperture related, it has a bigger

divergence angle; therefore, the projecting spot size is large. In fact, human

error is unavoidable since the doctor must use manual methods, and the

operation stability, precision, response time and the projecting spot size choice

is determined completely by doctor's skill and experience. In order to

improve the operation stability, and the precision of the projecting spot size,

Germany's Wavelight Corporation used a method that combined mirror

projecting and a mechanical arm to treat the patient. It improved stability, but

the flexibility was poor, it still used optical fiber delivery, and the cost of the

system was high.

In addition, the optical fiber method has two important issues: one is the

insertion loss, another is the optical fiber loss. Because optical fiber delivery

uses one kind of multi-mold optical fiber, leakage is big. Besides, the optical

Page 30: Implementation Of The Laser Scanning System Used For Dermatological Treatments

23

fiber twists and curves during the operation resulting in massive energy loss.

Simultaneously, it changes the exposure intensity, causing the degree of

illumination to be very difficult to maintain constantly.

Normally, a glass optical fiber has a low absorption band around 1.3µm

and 1.5µm, but a 0.308µm or 308nm wavelength has a high absorption rate.

In order to revise these shortcomings, it is necessary to enhance laser power to

make up for the losses. But high power laser energy has a bad impact on the

plasma tube and the electrode, shortening the life of the laser. Therefore, the

cost is high. Furthermore, it is not easy to control in a fixed spot size when

delivering laser energy through an optical fiber, and during the operation the

optical fiber curves and sways, and is easily and frequently damaged. The

fiber becomes a kind of expensively consumable material.

Therefore, in this thesis, image processing and scanning technology are

unified by utilizing the auto-controlled treatment method. A linear scanning

spot delivered by one pair of stepping motors to actuate the lens angle change.

The two motors are actuated by using the controller to control the scanning

spot frequency at a level and upright position. In order to enable the scanning

precision to achieve a functional size (7mm x 7mm) of 100%, it is estimated

the stepping motor scanning angle with a working distance of 600mm needs to

be approximately 0.069∘. Because of the rotation angle θ of the stepping

motor, the spot will move at an angle of 2θ. Therefore the motor moves each

step approximately 0.035∘.

In order to guide the laser energy into the selected treatment region, that

region must first be recognized. The image has to be transformed into motor

Page 31: Implementation Of The Laser Scanning System Used For Dermatological Treatments

24

coordinates. The scanning optics coordinates are transformed from the image

coordinates to become geometrical positioning data. Then the geometrical

coordinates are sent to the motor to start or stop the direction of rotation. For

this fast treatment method, the X direction motor performs the back and forth

scanning, and the Y direction motor then performs the up and down scanning.

This combination of scanning enables the laser beam to be projected precisely

on the skin.

It is extremely important that the skin treatment is related to laser energy

control. This thesis uses the biggest average density (Fluence) of laser pulse

energy at less than 16mJ/cm², the repetition rate is 225Hz, the spot size is

designed to be 0.49cm² (7mm x 7mm); therefore, the maximum single pulse of

a laser energy dose treating the skin is 8mJ. The National Cheng Kung

University dermatology department provides the reference for treating skin

disease by using an UVB311 ultraviolet ray. The biggest treatment energy at

present is 2.5J/cm². By using UVB311 ultraviolet power, it must take about

five minutes at least to achieve the desired treatment, and the illumination

energy and the treatment area are also not easy to control. This thesis uses an

auto-controlled scanning system that may control the scanning region, and

control the motor scanning frequency and the scanning time. The fine,

accurate control treatment energy needed to reach 2.5J/cm² is only needed for

one second.

This thesis uses an excimer laser system. This is a gas laser that uses

specifically premixed XeCl gas to generate 308 nm light, an ultraviolet

wavelength used to perform skin surgery treatment. After careful data

Page 32: Implementation Of The Laser Scanning System Used For Dermatological Treatments

25

collection, analysis, and study, it was determined that the 308nm wavelength

can be effectively absorbed by the skin to control psoriasis. Also, from an

American Medical Association report, it is found this 308nm excimer laser

emission light can control psoriasis for a one year period [19] which is a great

improvement over the traditional Light Box treatment or various medicinal

treatments involving steroids and antibiotics. The 308nm wavelength has

also been approved by the FDA (Food and Drug Administration, U.S.

Department of Health and Human Services) as a safe and efficient method to

work on human skin and to improve the treatment of some dermatological

lesions over traditional treatments. This information provided the main

motivation to develop this psoriasis treatment device.

Page 33: Implementation Of The Laser Scanning System Used For Dermatological Treatments

26

Chapter 3 Galvo Scanner and Beam Alignment

This chapter focuses on the selection of the galvo scanner components in

order to build the scanner system and to perform the laser beam delivery job

through beam alignment. To make a galvanometer (shortened to “galvo”)

scanner function, three main components are needed: they are the galvo, the

mirror, and the driver board. To achieve the galvo scanning system’s ultimate

performance, it is extremely important in the design phase to properly select

the galvo type and the mirror size [21].

3.1 Galvanometer

When doing this research, the first concerns about the galvo emphasized

speed, accuracy, size and cost. As mentioned above, optical fiber has to be

operated manually and causes some inconveniences during the operation.

Therefore, avoiding the defects of the current laser products becomes a key

point, and the galvo scanner based auto-controlled system is the one that can

solve the problems. This system also offers improved features such as

flexibility of angle, high speed, accuracy, compact size and lower cost.

There are a few companies that offer different galvo selections on the

market such as General Scanning Inc., USA; Cambridge Technology Inc.,

USA; and one from Europe called LM. To choose an appropriate galvo for the

auto-controlled system, a list of specification requirements for the system was

made as listed below [22].

Page 34: Implementation Of The Laser Scanning System Used For Dermatological Treatments

27

Wavelength: 308nm Process Time: minimum 5 minute

Field Size: 10x10cm² Focused Spot Size: 0.49 cm²

Working Distance: 60 cm Accuracy: ±0.1cm

Lifetime: at least 1 year

After comparing different types of galvos, the General Scanning, Inc., the

G120D galvo was selected for the scanning system.

3.2 Mirrors

A mirror must be chosen based on several factors such as cost, stiffness,

weight, thickness, density and material. In order to match the galvo scanner

based system, the mirror must be very thin and light weight, speed and

accuracy in the layer image are also concerned to evaluate overall system

performance [22, 23]. The list for mirror selection specifications are as

follows:

Wavelength: 308nm

Thickness: 1mm

Material: fused silicon

Weight: as light as possible

A lightweight substrate of mirror-grade fused silicon is the best material

for repositioning the mirrors in fractions of a millisecond. Galvo mirrors can

be coated with a variety of metal/dielectric coatings to offer high reflectivity at

different wavelengths and increased laser power throughput. Most laser

Page 35: Implementation Of The Laser Scanning System Used For Dermatological Treatments

28

systems use coaxial HeNe lasers or visible laser diodes to aid with the setup

and alignment of the optical system. It is important to specify mirrors with

good reflection properties at both the primary wavelength and at that of the

alignment laser. This system is using a 308nm ultraviolet wavelength

excimer laser to treat the skin disorder; so the mirror coating selection is

limited. Because of this special requirement for the system, it is very difficult

to find the desired performance mirrors in off-the-shelf markets.

We have to search for companies that can make special design or

custom-made mirrors to fulfill system performance. There are few

companies can manufacture the thickness of the thin mirror, most of them need

to be custom designed. Table 3.1 provides a list of the companies searched.

Table 3.1 Mirror manufacturer

Online Catalog Company Headquarters

Online

Quote

Custom

Order

Custom Plasma Ireland Inc. (Manufacturer)

Ireland

Custom Axsys Technologies (Manuf., Distrib., Sole

Distrib. & Service)

Rochester

Hills, MI

Custom OPCO Laboratory, Inc. (Manuf. & Service)

Fitchburg, MA

Custom Qioptiq Polymer (Manuf. & Service)

La Verne, CA

Custom Tower Optical

Corporation (Manufacturer)

Boynton

Beach, FL

Each one of them has ability to design and manufacture the desired

Page 36: Implementation Of The Laser Scanning System Used For Dermatological Treatments

29

mirrors ; therefore, cost became the last factor used to select the company.

Plasma Ireland, Inc. willingly reduced costs and so became the current supplier

of mirrors for the system.

3.3 Driver board

This system is specially designed for dermatology treatments; but an

attempt was made to search for available driver boards on the market. Most

driver boards use some kind of combination of detected position, galvo drive

current, and velocity of angular and integral-of-error signals to make a

closed-loop system to be controlled at the desired positioning speed and

accuracy. Since the GSI galvo scanner has an available driver board call

“mini sax”, the best selection is their product in order to maintain consistency.

It isn’t necessary to worry about the driver board because the closed-loop

galvo offers the speed, accuracy, and low cost the system needs.

Although the galvo scanner system components are selected, there are still

many choices that can be substituted for the chosen components, and that may

have better or outstanding functions in comparison with them. If better

alternatives are found during the assembly process, they may be chosen to

improve system performance. At this point, it is important to focus on the

chosen parts in order to find out how well or not the system works performing

the following steps. To prove the system is at least currently functional, the

beam alignment techniques are done to build up the basic structures needed by

the system.

3-4 Alignment procedure

There are many kinds of beam alignment techniques. Some techniques

Page 37: Implementation Of The Laser Scanning System Used For Dermatological Treatments

30

use very simple methods to deliver the laser beam and others use high-tech

equipment to achieve the purpose of beam alignment. One of the easiest

alignment techniques, manual alignment, is used to align the auto-controlled

scanning system. No matter how simple a technique is used, the goal is to

deliver the beam precisely to the target and to match the design performance.

Before any alignment begins, the factors that affect laser alignment while

performing the alignment steps are summarized.

1. The laser beam must be parallel to each travel axis to achieve its maximum

accuracy.

2. The alignment should begin from the laser head and move out one

component at a time until the last component on an axis is aligned with the

laser beam target aperture.

3. The angle of laser beam can be aligned by moving the laser head or

adjusting the beam plane level.

4. The reflected laser beam can be aligned by adjusting a beam splitter.

5. The angular direction of the beam transmitted straight through a beam

splitter will not be changed by adjusting that component.

6. The retroreflector does not change the angular direction of the beam. The

laser beam remains parallel to its original path.

7. On two-dimensional X-Y coordinates, the first axis is the one that needs its

angular direction adjusted to receive the beam from laser head. When the

first axis and laser head are aligned, the rest of angular adjustment required

by the other axis is accomplished by rotating an optical component.

8. The laser beam path should be kept horizontal or vertical for ease of optical

layout and alignment.

Page 38: Implementation Of The Laser Scanning System Used For Dermatological Treatments

31

9. Make sure all components are set up on the plane and their direction is

parallel or perpendicular to the stage plane.

10. Before installing the optics, make sure all beam paths bend in an optical

square.

The principle of alignment is simple: it delivers a laser beam to a specific

point. The laser beam is one specific visible color or wavelength. Typically,

a red or green colored laser diode is used as the reference beam.

In this case, a manual method is used to perform the laser beam delivery

alignment, because the simplest is the best. Besides, while the

auto-controlled scanning system needs care concerning the galvo scanners, its

optics delivery method is not complicated; therefore, using a simple method

can make the system cost less. Figure 3.1 demonstates the simple way of beam

alignment [24].

Page 39: Implementation Of The Laser Scanning System Used For Dermatological Treatments

32

Figure 3.1 Beam alignment

Page 40: Implementation Of The Laser Scanning System Used For Dermatological Treatments

33

Chapter 4 Assembly and Verification

4-1 Assembly concept

In order to complete the auto-controlled scanning system (in Pictures 4.1

and 4.2), the final assembly steps are described. This system should at least

include an initial mirror to provide an entrance for the laser beam, a final mirror

which conducts and projects the laser beam onto the desired treatment area, and

a set of adjusting optics which are set between the first mirror and the final

mirror. Also, the adjusting mirror set could also make the first and final

mirrors adjust different angles, and correspond with each other from start to end.

Picture 4.1 Auto-controlled scanning system – view 1

Page 41: Implementation Of The Laser Scanning System Used For Dermatological Treatments

34

Picture 4.2 Auto-controlled scanning system – view 2

Between the initial mirror and the adjusting set, there are also the

attenuation lens and the distributing lens. Furthermore, a reflecting mirror is

set between the initial mirror and the adjusting set, to reflect the laser beam to

the dynamometer. In the corresponding route from where the laser beam

reflects from the final mirror, there’s a camera placed to photograph the desired

treatment area. Thus, through this brand-new system, when the first and final

mirrors adjust different angles and finally correspond with each other from start

to end, the laser beam can pass through the optical components, and finally be

conducted and projected precisely on the desired treatment area. That is to say,

the defects from manually controlled surgeries are avoidable and the goal of

automatically controlling the therapy is attainable.

Through the computer, a camera can not only provide the operator with a

convenient way to examine and choose the desired treatment area, but it can also

Page 42: Implementation Of The Laser Scanning System Used For Dermatological Treatments

35

transfer the image to the computer to get the coordinates of the treatment area.

This provides the controlling data for first and final mirrors to make angle

adjustments. The auto-controlled laser system can be more fully understood

through the following examples tied in with the attached figures.

4-2 Assembly procedure

This section starts the step by step process used to construct the designed

system. Please compare the steps with Figure 4.1 and Figure 4.2, which show

the positions of all components described below. The initial reflecting mirror 1

is set on the plate. Set in the path of the reflected route from the initial

reflecting mirror 1, the components are set up in the following order.

First is a pin hole 60, and its stand 6, through which the laser beam passes.

Second, behind pin hole 6, attenuator 4 is set up on a fixed head 42 and is driven

and rotated by a stepping motor 41. Third, the reflecting mirror 5 is fixed on

stand 51 and is rotated by motor 52. Fourth, a divergence lens 7 is set up

behind reflecting mirror 5, followed closely by another stand 6 with a pin hole

60. Fifth, an adjusting mirror set 3 is set right behind the second pin hole set;

this adjusting mirror set 3 is composed of two corresponding mirrors which are

mirror 31 and mirror 32, in that order. Mirror 31 and mirror 32 are

individually driven by galvo 33, and rotate to different angles. Sixth, and last,

a final reflecting mirror 2 is set up on the light reflecting route of second mirror

32; simultaneously, a camera 8 is set up on the reflecting route of the final

reflecting mirror 2; hence, camera 8 can photograph the object along the route of

reflecting mirror 2. Although not mentioned earlier, a power meter 9 is also set

on the reflecting route of reflecting mirror 5.

Page 43: Implementation Of The Laser Scanning System Used For Dermatological Treatments

36

4-3 Verification

After explaining the general construction of the device, its function is

analyzed below (in Figures 4.2 and 4.3, and Table 4.1). First, the system turns

attenuator 4 and reflecting mirror 5 from vertical to horizontal. The laser beam

passes through initial reflecting mirror 1 and into pin hole 60. Second, the

beam then passes through attenuator 4 where the beam’s luminous intensity is

adjusted. The beam then hits reflecting mirror 5 and is reflected to power

meter 9 (see movement situation 1 in Figure 4.3), where its intensity is once

again adjusted after being measured by power meter 9. Third, after the

intensity is properly adjusted, mirror 5 is rotated back to its normal position so

that the beam projects onto distributing lens 7 (as Figure 4.2 and Figure 4.4,

movement situation 2 show), so the beam can be changed from centralizing to

distributing. Finally, after being distributed, the beam passes through another

pin hole 60 to start a series of reflections, from mirror 31, to mirror 32, to final

mirror 2, until finally the laser beam is projected onto the intended treatment

area [25].

In short, this new system calculates the size of the intended treatment region,

and sections the region into many coordinates for the entire therapy when it

receives the image photographed by camera 8. Then, it conveys the

coordinates to the drive units, and drives mirror 31 and mirror 32 to adjust

angles based on the calculated coordinates; thus, the beam can be projected on

every coordinate. The whole treatment is controlled automatically.

It’s not only safe, fast, and highly efficient, but also avoids lapses due to

manual operation; at the same time, the eye-damage caused by strong laser

luminosity could be minimized since the operator no longer has to look at the

Page 44: Implementation Of The Laser Scanning System Used For Dermatological Treatments

37

treatment area directly.

Description of the Drawings:

Figure 4.1: the 3-D structure of this new invention.

Figure 4.2: the 3-D structure of another direction.

Figure 4.3: the top view of movement situation one.

Figure 4.4: the top view of movement situation two.

Table 4.1 Parts description

Description of the Preferred Embodiment:

1. Initial reflecting

mirror

3. The set of adjusting

mirrors

31. First mirror

32. Second mirror 33. Galvo 4. Attenuator

41. Stepping motor 42. Fixed head 5. Reflecting mirror

51. Shutter stand 52. Motor 6. Pinhole stand

60. Pin hole 7. Divergence lens 8. Camera

9. Power meter 2. Final reflecting mirror none

Page 45: Implementation Of The Laser Scanning System Used For Dermatological Treatments

38

Figure 4.1 The 3-D structure of this new invention

Page 46: Implementation Of The Laser Scanning System Used For Dermatological Treatments

39

Figure 4.2 The 3-D structure from another direction

Page 47: Implementation Of The Laser Scanning System Used For Dermatological Treatments

40

Figure 4.3 The top view of movement situation 1

Figure 4.4 The top view of movement situation 2

Page 48: Implementation Of The Laser Scanning System Used For Dermatological Treatments

41

4-4 Target area control

After showing the system construction, we now take a look at how the

whole system works and the flow chart of the image selection. There are two

industrial uses PCs (IPC) in the system: one is a DOS based system; the other

is a Windows system. The DOS IPC is used to control the galvo scanner

system driver broad and communicate with the Windows IPC. The Windows

IPC is mainly used to control the laser source and the touch screen panel

monitor used to select and process the camera-grabbed image data. The two

IPCs work together to simultaneously operate the laser trigger and the driver

motions. The schematic diagram of the system is showing in Figure 4.5 for

reference.

Communication

Laser Head

Frame

Grabber

Touch

Panel

Driver

Laser

beam

Camera

Scanner

IPC IPC

Figure 4.5 Schematic diagram of the system

Page 49: Implementation Of The Laser Scanning System Used For Dermatological Treatments

42

The flow chart in Figure 4.6 shows the procedure as the camera grabs the

image and displays it on the screen. The user or operator selects the surgery

regions so that the computer can calculate its scanning area and treatment time.

The scanner performs the actual operation only while the operator is pushing

the surgery footswitch.

Grab/Display Image

Select Surgery Regions

Compute Scanner Locations

Control Scanner

Start Surgery

Figure 4.6 Flow chart

Before starting the treatment, the user calibrates the galvo position by the

small circle, in Picture 4.3, and large circle, in Picture 4.4; then test the center

cross calibration, in Picture 4.5, and each point of the treatment area

calibration, in Picture 4.6, to make sure the galvo is recognized and saved by

computer.

Page 50: Implementation Of The Laser Scanning System Used For Dermatological Treatments

43

Picture 4.3 Small circle calibration

Picture 4.4 Large circle calibration

Page 51: Implementation Of The Laser Scanning System Used For Dermatological Treatments

44

Picture 4.5 Cross calibration

Picture 4.6 Treatment area calibration

Page 52: Implementation Of The Laser Scanning System Used For Dermatological Treatments

45

A basic example is provided to illustrate how the concept works. The

letter “A” is input to the screen. The operator selects the “A” shape as the

surgical region. The computer calculates the area of the “A” shape and

determines the time needed to complete the surgery. The operator presses the

footswitch until the system automatically finishes the job and the letter “A”

appears on the target, as seen in Picture 4.7.

Picture 4.7 Operation demonstration

Following that example, a patient’s colorful psoriasis picture is placed in

Page 53: Implementation Of The Laser Scanning System Used For Dermatological Treatments

46

front of the working device. The image system is then used to manually

select the lesions to test its function. Pictures 4.8 and 4.9 present the results of

such tests.

Picture 4.8 Colorful psoriasis – test 1

Picture 4.9 Colorful psoriasis – test 2

Page 54: Implementation Of The Laser Scanning System Used For Dermatological Treatments

47

Chapter 5 Conclusion

5-1 Features

This auto-controlled laser scanning system for dermatology treatments has

already been completed, yet the whole system for human skin applications still

needs clinical studies to prove its value.

It is distinguishing the industrial purposes from the medical purposes of

the system. Once the system is permitted to perform surgeries, it will

perform features that the other competitors do not have, which are:

1. Delivers the laser energy specifically to the lesion site via a scanning spot

laser beam delivery system for automatic treatment.

2. The tiny laser spot size (0.49cm²) allows for precise and selective

treatment.

3. The treatment area image recognition system plus the flying spot laser

beam delivery system provide incomparably precise and selective

treatments.

4. MED tests and Doses options are installed in the system for effective

treatments.

Hopefully the system can assist doctors and patients for better solution.

Although the auto-controlled laser scanning system was originally designed

for dermatology purposes, the applications of the system are not supposed to

be limited in this medical field.

Page 55: Implementation Of The Laser Scanning System Used For Dermatological Treatments

48

5-2 Future development

In the future development of the system, there are some ideas that could

make the whole device more convenient or user friendly. These can be

discussed from both software and hardware viewpoints, and from the potential

applications based on the image system.

1. For the software part, combining the two IPCs (DOS and Windows) into

one computer with more efficient upgrades, could save on cost. It is possible

to replace the DOS by the Windows IPC, however, during the software test,

the DOS system triggered the laser with no time delay, but the Windows

system had a time delay. This would cause problems during real time

treatment when the laser beam delivery to the target could be delayed by a

couple of seconds. The auto-controlled scanner system is combined with the

image system which currently uses a touch panel screen with the user painting

and drawing the skin area.

The possibility of an automatic, computer-controlled system used to

isolate the treatment area by identifying the unhealthy lesions is being studied.

Unhealthy lesions can be identified as different from healthy skin tissue by

identifying changes in the skin color or skin texture. In other words, the

doctor’s only job in the future would be to provide confirmation from the

monitor screen after the computer selected the treatment region for one or

more lesions, select the energy level and doses, and then push the footswitch to

complete the treatment.

Page 56: Implementation Of The Laser Scanning System Used For Dermatological Treatments

49

2. For the hardware part, the auto-controlled laser scanning device is

installed and fixed in one place without moving the plane and controlling the

scanner angle to target. Although the device can only lift up and down so far,

but there is a possibility of letting the device shift right to left in front of the

target, so the patient can stand still or lie down and wait for treatment until

finish. Again, the problem faced is that the laser cannot move around.

The typical excimer laser is heavy and needs a gas tank to supply the gas

into the laser. The laser needs to cooperate with the scanner without using an

optical fiber which would diminish the energy, and it needs calibration

whenever the laser is moved.

The second thought is to replace a galvo with a stepping motor, because

the galvo rotation speed is higher than the system requires and the stepping

motor can handle a bigger mirror. The treatment area, therefore, may become

bigger and save more time during the surgery.

3. The potential applications of the image system obtain image

identification of license plates, human face recognition, traffic condition

real-time feedback for driving safety, and the replacement of the laser source

from 308nm to various wavelengths for applications requiring different kinds

of laser treatment.

The monitor and CCD camera uses for crime tracking usually have a

problem with identifying license plates. If the image system can modify the

gray level image from 0 to 255 and find an easy way to identify the license

plate number, there might be a chance to catch the criminal quickly and

Page 57: Implementation Of The Laser Scanning System Used For Dermatological Treatments

50

prevent the next incident. This can also be applied to human face

identification for criminals or searching for missing people.

Driving safety is one of the major issues for human life with thousands of

people dying in car accidents every year. Some luxury sedans, like

Mercedes-Benz, Lexus, and BMW, have systems to prevent car crashes from

the rear, but the system function is limited to use during cruise mode, not all

the time.

The image system has an aiming beam sensor to detect the working

distance, and to make sure the laser power density and dosage can be precisely

delivered to the target. It can also detect if the distance is within safety limits

and feedback this information to the computer to make the decision whether to

trigger the laser or not. The image system has the potential to be modified as

a safety guard to prevent major car crashes.

For the medical laser device, the most interesting thing to do is change the

laser wavelength for applications involving different diseases or procedures.

Basically, the whole auto-controlled scanning system can be considered as

separate modules, as the system schematic diagram in Figure 4.5 shows. All

that needs to be done is put in a different laser source and adapt the optics to

the new wavelength. By doing it this way, a lot of research and development

time can be reduced and more useful laser treatment devices can be created.

Page 58: Implementation Of The Laser Scanning System Used For Dermatological Treatments

51

Appendix A

Other dermatology laser patent analysis list

Number Patent name Patent

number and

date

Patent scope

1 Multibeam laser for skin

treatment

US200608

4953

2006-04-20

For skin beauty,

not relevant

2 Improved hand-held laser

device for skin treatment.

MXPA0400

1537-

2004-10-27

Portable beauty

laser, not

relevant

3 Methods for treatment of human

skin damaged by laser treatment

or chemical peelings and

compositions useful in such

methods

US200510

0592 -

2005-05-12

Not relevant

4 A topical agent containing niacin

for application to the skin prior to

luminous treatment

WO20040

87093 -

2004-10-14

Not relevant

5 Laser therapy device for the

treatment of skin diseases

US200418

1267 -

2004-09-16

Optical fiber

treatment, no

infringement

6 Method and apparatus for skin

treatment using near infrared

laser radiation

US200501

5077 -

2005-01-20

Skin removal,

not relevant

7 Hand-held apparatus for skin

treatment with intensive light

WO20040

10884 -

2004-02-05

Portable beauty

laser, not

relevant

8 Combined device for cooling the

skin and suction removal of

smoke and ejected skin particles

generated during laser treatment

of the skin, has an open conical

design that ensures that both

functions are reliably fulfilled

DE103072

60 -

2003-08-21

Not relevant

9 Skin treatment device JP200421

5837 -

2004-08-05

Low

temperature

skin treatment,

no infringement

Page 59: Implementation Of The Laser Scanning System Used For Dermatological Treatments

52

Number Patent name Patent

number and

date

Patent scope

10 Skin diagnosis system, electronic

treatment apparatus

JP200414

1259 -

2004-05-20

Skin diagnosis,

not relevant

11 Methods for treatment of human

skin damaged by laser treatment

or chemical peelings and

compositions useful in such

methods

US200301

2762 -

2003-01-16

Not relevant

12 Laser skin treatment device with

control means dependent on a

sensed property of the skin to be

treated

GB238175

2 -

2003-05-14

Not relevant

13 Treatment of acne vulgaris skin

condition by irradiation with light

of specific wavelengths to target

specific chromophores &

stimulate collagen production

GB236802

0 -

2002-04-24

LED skin

treatment,not

relevant

14 Composite beauty skin treatment

device

JP200211

3116 -

2002-04-16

Not relevant

15 Probe for skin high frequency

treatment

US649770

2 -

2002-1

2-24

Not relevant

16 Automatic firing apparatus and

methods for laser skin treatment

over large areas

WO01265

73 -

2001-04-19

Not relevant

17 Laser skin treatment apparatus EP105745

4 -

2000-12-06

Optical fiber

treatment, it can

measure skin

temperature

18 Laser optics positioning element

for skin treatment

EP102387

4 -

2000-08-02

Not relevant

Page 60: Implementation Of The Laser Scanning System Used For Dermatological Treatments

53

Number Patent name Patent

number and

date

Patent scope

19 Device for hand and foot

treatment and care has a medical

laser for removal of skin or nail

material with such treatment

resulting in very little generation

of dust from skin or nail parts

and therefore being much more

hygienic

DE199520

45 -

2001-08-23

Not relevant

20 Laser for skin treatment WO00710

45 -

2000-11-30

Auto focus

laser, no

infringement

21 Skin treatment process using laser US603668

4 -

2000-03-14

Skin removal

laser, no

infringement

22 Method for non-invasive wrinkle

removal and skin treatment

US607729

4 -

2000-06-20

Not relevant

23 Enhanced laser skin treatment

mechanism

US598051

2 -

1999-11-09

Not relevant

24 Laser treatment/ablation of skin

tissue

US608658

0 -

2000-07-11

Not relevant

25 Laser dermal implants for the

treatment of facial skin

depressions

US581709

0 -

1998-10-06

Not relevant

26 Device for cooling skin during

laser treatment

EP082771

6 -

1998-03-11

Skin removal

laser, no

infringement

27 Laser application device for skin

treatment

JP110471

46 -

1999-02-23

Skin removal

laser, no

infringement

28 Skin treatment process using laser WO96415

79 -

1996-12-27

Not relevant

29 Method of treatment of skin

lymphomas

RU212853

3 -

1999-04-10

Not relevant

Page 61: Implementation Of The Laser Scanning System Used For Dermatological Treatments

54

Number Patent name Patent

number and

date

Patent scope

30 Skin treatment process using laser US581708

9 -

1998-10-06

Not relevant

31 Laser treatment method for

removing pigment containing

lesions from the skin of a living

human

US529027

3 -

1994-03-01

Optical fiber,

no infringement

32 Device for treatment of undesired

skin disfigurements

US532061

8 -

1994-06-14

Not relevant

33 Laser treatment method for

removing pigmentations, lesions,

and abnormalities from the skin

of a living human

US521745

5 -

1993-06-08

Not relevant, no

infringement

Page 62: Implementation Of The Laser Scanning System Used For Dermatological Treatments

55

Appendix B

Wavelight laser technology patents

Number Patent name Patent number and date

1 Fiber laser arrangement EP1650839 - 200

6-04-26

2 Laser system with soild laser heads WO2006034783

- 2006-04-06

3 Device for producing a white light ES2248534T - 20

06-03-16

4 Rapid wavefront measurement WO2006024504

- 2006-03-09

5 Device for ophthalmologically

treating the eye using a fixation light

beam

ES2245684T - 20

06-01-16

6 Laser system for corneal grafting ES2245384T - 20

06-01-01

7 Method of generating a control

program for a device for

photorefractive corneal surgery of

the eye

US2005187540 -

2005-08-25

8 Method for the minimal-to

non-invasive optical treatment of

tissues of the eye and for diagnosis

thereof and device for carrying out

said method

WO02076355 - 2

002-10-03

9 Method for producing a control

program for a device used for

performing corneal eye surgery

US2003105457 -

2003-06-05

10 Device for medical treatment of the

eye using laser radiation US6755817 - 20

04-06-29

11 Device used for the photorefractive

keratectomy of the eye using a

centering method

US2002128634 -

2002-09-12

12 Device for photorefractive cornea

surgery in higher-order visual

disorders

US6530917 - 20

03-03-11

13 Dispositivo para un treatmento

medico del ojo con radiacion laser. ES2231210T - 20

05-05-16

Page 63: Implementation Of The Laser Scanning System Used For Dermatological Treatments

56

Number Patent name Patent number and date

14 Device for treating bodily substances US6328732 - 20

01-12-11

15 Device for medical treatment with a

light source WO9962442 - 19

99-12-09

16 Surgical vitrectomy instrument has

aspiration duct containing radiation

conductor, chamber with at least

three openings

DE19842799 - 2

000-03-23

17 Device and method for the removal

of body substances US6027493 - 20

00-02-22

18 Laser device for treatment of patient

skin and other dermatological

processes

DE19811627 - 1

999-09-23

19 Arrangement for treating bodily

substances DE19734732 - 1

998-06-18

20 Surgery system for vitreous humour

of eye DE19720660 - 1

998-11-19

21 Intra-ocular cataract surgery device DE19702353 - 1

998-04-16

22 Material processing arrangement

with pulsed laser e.g. for eye surgery DE19702335 - 1

998-08-27

Page 64: Implementation Of The Laser Scanning System Used For Dermatological Treatments

57

Appendix C

InPro Innovations patents

Number Patent name Patent number and

date

1 Process for interior coating of hollow

bodies DE59912989D -

2006-02-02

2 Method for connecting pieces

accessible only from one side and

assembly of such pieces

DE50302005D -

2006-01-26

3 Method and device for determining

the resistance of sheet metal to

alternating bending loads

EP1577659 - 20

05-09-21

4 Method and device for forming of

undercuttings when joining

superimposed work pieces, in

particular coated and/or varnished

metal sheets

EP1249286 - 20

02-10-16

5 Process for welding thermoplastic

joining parts using laser diode

radiation

EP1238781 - 20

02-09-11

6 Teach-in generation of programs for

component 3-dimensional solid state

laser processing involves converting

laser diode radiation incidence point

image to bitmap with frame grabber

card

DE19961625 - 2

001-07-05

7 Device for plasma polymerizing

batches of hollow work pieces in

plural pieces processing

EP1054432 - 20

00-11-22

8 Process for making multifunctional

plasma polymerized layers on plastic

parts

EP1018532 - 20

00-07-12

9 Docking process for welded sheet

plates being deformed by internal

high pressure

DE19812884 - 1

999-09-23

10 Process for geometry recognition and

tracking during thermal treatment of

elements by means of laser beam

EP0904886 - 19

99-03-31

Page 65: Implementation Of The Laser Scanning System Used For Dermatological Treatments

58

Number Patent name Patent number and

date

11 Pump lamp exchange device for solid

laser DE19709660 - 1

998-09-24

12 Impermeable layer on inside wall of

vessel especially plastic fuel tank DE19700426 - 1

998-07-16

13 Method and arrangement for surface

treatment with temperature control,

particularly for superficial hardening

with laser radiation

EP0836905 - 19

98-04-22

14 Method of hardening the surface of a

work piece using a beam, particularly

a laser beam and device for executing

this method

EP0822027 - 19

98-02-04

15 Process for welding tinned metal

sheets, using a solid state laser EP0800888 - 19

97-10-15

16 Laser beam welding seam depth

control DE19605888 - 1

997-08-21

17 Joining motor vehicle body

components DE19604081 - 1

997-08-07

18 Monitoring protective glass of laser

welding optics DE19605018 - 1

997-08-07

19 Method for quality testing of

semi-products, modules and

components with ultrasounds

EP0770867 - 19

97-05-02

20 Process for plasma coating a plastic

object with multifunctional layers EP0739655 - 19

96-10-30

21 Method for creating low ohmic

contact areas on thermoplastic

articles filled with long steel fibers

EP0728571 - 19

96-08-28

22 Method for producing thermoplastic

objects with plasma-suitable surfaces EP0722823 - 19

96-07-24

23 Method for the decomposition and

separation of recyclable

three-material composite components

by type

EP1036596 - 20

00-09-20

24 Process and device for welding sheet

metal construction elements, in

particular tinned sheet metal elements

by means of laser beams

EP0687519 - 19

95-12-20

Page 66: Implementation Of The Laser Scanning System Used For Dermatological Treatments

59

Number Patent name Patent number and

date

25 Process for pre- or post-treatment of

components welding seam to be

executed resp. executed

EP0688627 - 19

95-12-27

26 Method and device for coating

electrostatically and/or pneumatically

a conductive substrate with a liquid

coating product

EP0695582 - 19

96-02-07

27 Method of manufacturing objects or

preforms using polymers as raw

materials, especially for plastic fuel

tanks.

EP0677366 - 19

95-10-18

28 Process and device for monitoring the

welding depth in work pieces during

laser beam welding.

EP0674965 - 19

95-10-04

29 Method of regulating the force

applied to sheet during deep drawing DE4339153 - 19

95-05-24

30 Method for determining reference

variables for process systems,

especially for a process controller

DE4303561 - 19

94-08-11

31 Method for making products by force

or pressure influenced drawing

processes.

EP0589066 - 19

94-03-30

32 Process and device for positioning and

controlling a high energy source, in

particular a laser relative to a work

piece.

EP0531558 - 19

93-03-17

33 Method of correcting regulation

parameters in a process control

system, especially to maintain the

dynamic range (regulating range) of

the process control system during the

course of the process and device for

implementing the process.

EP0489130 - 19

92-06-10

34 Process and device for automatic

determination of parameters for

process control systems with unknown

transfer behavior, in particular for

process control systems for resistance

spot welding.

EP0452440 - 19

91-10-23

Page 67: Implementation Of The Laser Scanning System Used For Dermatological Treatments

60

Number Patent name Patent number and

date

35 Cruciform planar specimen for

biaxial materials testing DE3914966 - 19

90-07-12

Page 68: Implementation Of The Laser Scanning System Used For Dermatological Treatments

61

References

[1] Carl R. (Rod) Nave, “HyperPhysics”, Department of Physics and

Astronomy, Georgia State University, Atlanta, Georgia, 2006.

[2] Randy Park, MD., “Psoriasis”, Department of Emergency Medicine, Denton

Regional Medical Center, May 1, 2007.

[3] Gladimir V. G. Baranoski and Aravind Krishnaswamy, “An Introduction to

Light Interaction with Human Skin”, RITA ; Vol. XI , N´umero 1 , 2004.

[4] Aravind Krishnaswamy and Gladimir V.G. Baranoski, “A

Biophysically-Based Spectral Model of Light Interaction with Human Skin”,

BioSpec; University of Waterloo, Ontario, Canada, 2005.

[5] Available at: www.shorelaser.com

[6] Fitzpatrick T.B., Arndt K.A., El Mofty A.M. and Pathak M.A.,

“Hydroquinone and Psoralens in the therapy of hypermelanosis and vitiligo”,

Arch Derm; Vol. 93, pp. 589 – 600, 1966.

[7] Available at: http://www.martinpletscher.ch/dermatologie/Lichttherapie.

html

[8] Available at: http://www.drhoenle.de/index.php?id=51

[9] Green C., Ferguson J., Lakshmipathi T., Johnson B.E., “313nm UVB

phototherapy: an effective treatment for psoriasis”, Br J. Dermato.; Vol.

119, pp. 691- 696, 1988.

[10] Van Weelden H., De la Faille B.H., Young E., Van der Leun J.C.

“Comparison of narrowband UV-B phototherapy and PUVA

photochemotherapy in the treatment of psoriasis”, Acta Dermatol Venereol

(Stockh); Vol. 70, pp. 212-215, 1990.

Page 69: Implementation Of The Laser Scanning System Used For Dermatological Treatments

62

[11] Van Weelden H., De la Faille B.H., Young E., Van der Leun J.C. “A new

development in UVB phototherapy of psoriasis”, Br J. Dermatol; Vol. 119,

pp. 11-19, 1988.

[12] Young A.R. “Carcinogenicity of UVB phototherapy assessed”, Lancet; Vol.

345, pp. 1431-1432, 1995.

[13] 王肇陽,“皮膚病的認識”,慈濟文化出版社,p. 161,民 94 年 11 月

[14] Available at: http://www.photomedex.com/pages/faq.html

[15] Available at: http://www.wavelight-laser.net/admin.tool/_cms/int/content/

na/Piy pyH8AAAEAAGP0YGoAAAAI/807,0.html

[16] Speight E.L. and Farr P.M. “Erythemal and therapeutic response of

psoriasis to PUVA using high-dose UVA”, Br J. Dermatol; Vol. 131, pp.

667-672, 1994.

[17] Parrish J.A and Jaenicke K.F., “Action spectrum for phototherapy of

psoriasis”, J. Invest Dermatol; Vol. 76, pp. 359-362, 1981.

[18] Bonis B., Kemeny L., Dobozy A., Bor Z., Szabo G. and lgnacz F. “308nm

UVB excimer laser for psoriasis (letter)”, Lancet; Vol. 350, p. 1522, 1997.

[19] Pravit Asawanonda, MD, R. Rox Anderson, MD, Yuchiao Chang, PhD, and

Charles R Taylor, MD., “308-nm Excimer Laser for the Treatment of

Psoriasis, A Dose-Response Study”, Arch Dermato,; Vol. 136, May 2000.

[20] Hee Choul Hwang, Chung Choo Chung, and Seung-Hi Lee, “Design of a

multi-rate controller and its application to galvanometer servo systems”,

American Control Conference; Proceedings of the 2001, Vol. 1, pp.

564-569, 2001.

[21] Rolland Zeleny, Nutfield Technology Inc, “Scanning the scene”, SPIE’s

Page 70: Implementation Of The Laser Scanning System Used For Dermatological Treatments

63

oemagazine; pp. 30-32, May 2004.

[22] Ricardo M. Lantican, PhD., “Serendipity: Accidental leads result in great

discoveries”, Manila Bulletin, November 26, 2006.

[23] Available at: http://www.alase.com/marking/engraving.htm

[24] Phillip V. Mitchell, Peter B. Griffith, and Donald K. Henderson,

“Fast-steering mirrors improve active beam stabilization”, Optoelectronics

World, October 2001.

[25] MARSHALL, G.F., “Scanning devices and systems.” In Applied Optics and

Optical Engineering 6, pp. 203-26, 1980.