implementation of tandem mass spectrometry: quality

46
Implementation of Tandem Mass Spectrometry: Quality Control/Quality Assurance of the New York State Newborn Screening Program. Mark A. Morrissey, PhD Wadsworth Center Department of Health State of New York

Upload: others

Post on 18-Mar-2022

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Implementation of Tandem Mass Spectrometry: Quality

Implementation of Tandem Mass Spectrometry: Quality Control/Quality

Assurance of the New York State Newborn Screening Program.

Mark A. Morrissey, PhDWadsworth CenterDepartment of Health

State of New York

Page 2: Implementation of Tandem Mass Spectrometry: Quality

Brand Names

Important:

The use of brand names in this presentations is for purposes of description only.

No endorsement is implied or intended.

Page 3: Implementation of Tandem Mass Spectrometry: Quality

Definitions

Quality Assurance is a comprehensive set of policies, procedures, and practices necessary to assure that the laboratory’s results are reliable.

Quality Control is a set of laboratory procedures designed to ensure that the test method is working properly.

From NYS DOH, Clinical Laboratory Evaluation Program (CLEP)

Page 4: Implementation of Tandem Mass Spectrometry: Quality

Outline

a. Administrative Preparation

b. Instrument Set-up/Method Validation

c. Routine/Daily Operations

Page 5: Implementation of Tandem Mass Spectrometry: Quality

a. Administrative Preparation

• Appropriate space

• Power and air conditioners

• Delineation of test panel

• Qualified staff

• Assay validation

• Lab certification

• Proficiency testing arrangements

Page 6: Implementation of Tandem Mass Spectrometry: Quality

More…

• Follow-up considerations

• Available medical specialists and diagnostic lab services

• Adequate notification of physician community

• Educational materials: lay and professional

• Advisory committee

Page 7: Implementation of Tandem Mass Spectrometry: Quality

Training Issues

• Instrument vendor site training- Instrument operation- Instrument troubleshooting

• Visit other MS/MS laboratories- 20 + MS/MS testing programs- spit equally between vendors

• APHL and NNSGRC sponsored courses- Baylor Institute of Metabolic Diseases- Duke Medical Center

• Develop training and competency logs

Page 8: Implementation of Tandem Mass Spectrometry: Quality

b. Instrument Set-up/Method Validation

Instrument Performance

Method Validation

Establishment of Cut-off values

Page 9: Implementation of Tandem Mass Spectrometry: Quality

Instrument Performance Issues

• Mass Calibration and Resolution– Establish operating range of instrument

• Probe rinses– Minimize Carryover

• Detection Optimization– Front End (Probe, cone, extractor, etc…) voltages.– Collision chamber, 2nd quadrapole, detector voltages

Page 10: Implementation of Tandem Mass Spectrometry: Quality

Instrument Performance Issues

• State files– Mass calibration

• Method files– Scan Parameters– Analytical run time (HPLC method)

• Data Reduction Software– Calibration Table

• Analyte & internal standard masses• Internal standard concentration• Blood spot volume

Page 11: Implementation of Tandem Mass Spectrometry: Quality

Instrument Performance Issues

• Mass Calibration

– Establish operating range of instrument

• Typical mass range 59amu to 1800 amu• We use 50 to 1100 amu, quarterly

– Materials used

• Polypropylene (PPG)• NaI / RbI solution

Page 12: Implementation of Tandem Mass Spectrometry: Quality

Method Validation Protocol

Page 13: Implementation of Tandem Mass Spectrometry: Quality

Method Validation

Requirements from CLIA or CLEP– Accuracy– Precision– Reportable Range– Specificity and Sensitivity– Reference Intervals– “Any other performance characteristics required for

test performance”

Page 14: Implementation of Tandem Mass Spectrometry: Quality

Method Validation

Establish Accuracy• Calculate recovery from analysis of fortified

samples (6 to 8 levels recommended)– [(S – C)/F] X 100

» Where: S is the observed concentrationC is the background concentrationF is the fortification concentration

• Acceptable recoveries: > 80%• Range is the demonstrated range of acceptable

recoveries (for example: C8 0.06 to 100 µmole/L)

Page 15: Implementation of Tandem Mass Spectrometry: Quality

Recovery (4 instruments, 4 Levels, 4 days)

Page 16: Implementation of Tandem Mass Spectrometry: Quality

Method Validation

• Intra run precision

– extract and prepare a set of blood spots.• Minimum of 10 replicate analyses

– Analyze in the same run on the same day.

– Calculate Coefficient of Variation (CV)

• Expected coefficient of variation: < 10%.

Page 17: Implementation of Tandem Mass Spectrometry: Quality

Method Validation

• Inter run precision

– Multiple day analysis (4 days)

– Prepare 4 extracts for each spiked pool daily

– Calculate Coefficient of Variation (CV)

• Expected coefficient of variation (CV): 5 – 15%

Page 18: Implementation of Tandem Mass Spectrometry: Quality

Method Validation(intra/inter run precision)

Page 19: Implementation of Tandem Mass Spectrometry: Quality

Establishment of Cut-off Values

• Pilot Testing– Do a literature search– Contact existing MS/MS programs– Manufacture of instrument or reagents

• Routine Testing Cutoffs– Analyze several thousand normal specimens– Calculate mean and standard deviation– Recalculate after several thousand samples

Page 20: Implementation of Tandem Mass Spectrometry: Quality

Establishment of Cut-off Values

• Consult metabolic specialist/follow up staff.

• Typical cutoffs: 3 and 10 sd from mean

• Compare cutoffs with other MS/MS programs.

• A balance between false positives/negatives

• Consider separate ranges for age > 7 days.

Page 21: Implementation of Tandem Mass Spectrometry: Quality

Establishment of Cut-off Values

Page 22: Implementation of Tandem Mass Spectrometry: Quality

c. Routine Operations

a. Routine, but not daily considerationsb. Daily analysis of samples

i. Sample Preparation1. Derivatized2. Un-derivatized

ii. Documentationc. Daily Data Evaluation

i. Data Reductionii. Evaluation of QC resultsiii. Evaluation of abnormal samples

Page 23: Implementation of Tandem Mass Spectrometry: Quality

Standard Operating Procedure

Page 24: Implementation of Tandem Mass Spectrometry: Quality

Routine, but not daily considerations

Instrument Maintenance• Follow the manufacturers recommendations.• Most labs use a check-off sheet with signature and date• Activities may include: clean cone, ballast the rough pump

and check oil level, check chiller temp, etc…• Maintained in the instrument logbook• Ultimately up to the area supervisor and laboratory director

Page 25: Implementation of Tandem Mass Spectrometry: Quality

Routine, but not daily considerations

Non-routine maintenance• Documented with nature of the problem, who found the problem, resolution

and the person fixing the problem. • We include preventative maintenance from the manufacturer. They have their

own specification for function checks and mass calibration.

Tuning• Concentrated Standard Solution. May not include all the applicable analytes

(C14:1, C5OH, etc). • Collected material from a used plate (not documented or traceable). We do

this as needed if the QC results are out of range.

Page 26: Implementation of Tandem Mass Spectrometry: Quality

Reagent Preparation

• HPLC grade, or Reagent grade, or better• To filter, or not to filter? • Hazards

– Acetonitrile -- toxic, flammable – Methanol -- toxic, flammable – Butanol -- flammable, irritant– Acetyl Chloride -- corrosive, water reactive. Use a

chilled bath and mix with butanol under argon.

Page 27: Implementation of Tandem Mass Spectrometry: Quality

Internal Standard Preparation

• MMWR recommends a deuterated standard to correspond to each analyte.

• Pre-prepared Internal Standards– Available from Perkin-Elmer or Cambridge Isotope

Laboratories. – Concentration and Expiration are provided by the

supplier.

Page 28: Implementation of Tandem Mass Spectrometry: Quality

Internal Standard Preparation

• In-house Prepared Internal Standards– Available from Dr. H.J. ten Brink, Cambridge

Isotope Labs and CDN Isotopes Inc. – Cheaper, but requires more labor– I estimate that for 250,000 samples it requires:

• Approximately 80 hours to prepare the standard• 40 hours to go through a qualification

Page 29: Implementation of Tandem Mass Spectrometry: Quality

Internal Standard Preparation Procedure

• Weigh individual deuterated stds into a suitable container. • Dissolve in 0.01N HCl (d-C16 in MeOH) for an exact

concentration of 1.0 mg/mL or 0.1 mg/mL.• Add the appropriate amount of each to individual 15-mL

polypropylene centrifuge tubes. (Adds up to about 7.3 mL). We make about 40 tubes.

• Seal with Teflon tape and store at –70 C. Stable for at least 1-year

• Should check concentrations and adjust, if necessary

Page 30: Implementation of Tandem Mass Spectrometry: Quality

Internal Standard Preparation Procedure

• To prepare working internal standard dissolve the entire tube in 2-L of methanol. Store at –20C. Lasts about 10-days, (10,000 samples).

• 2-years worth of standards costs us approximately $10,000 (500,000 samples)

Page 31: Implementation of Tandem Mass Spectrometry: Quality

Preparation of Quality Control Samples

• Standards available from Life Science Resources, H.J. ten Brink (carnitines) and Sigma (amino acids).

• Blood obtained from the Red Cross.• Individual stock standards of amino acids and

acylcarnitines are prepared in saline. • Added individually to 50 mL aliquots of whole blood.

Stored frozen along with a desiccant. Lasts approximately six months. (24/day at each level)

Page 32: Implementation of Tandem Mass Spectrometry: Quality

Preparation of Quality Control Samples

• One level is approximately equal to the request for a repeat specimen level. The other level is approximately 2 to 3 times the concentration of the first.

• Determination of control limits.

• QC Samples available from CDC (bi-annually) for periodic use.

Page 33: Implementation of Tandem Mass Spectrometry: Quality

Sample Preparation and Analysis

• Underivatized - fewer steps, faster– Add Internal Standard (critical step)– Cover plates and extract for 20 –30 minutes– Transfer– Analyze by MS/MS

• Derivatized - published, more sensitive• Both methods require care and consistency in

sample preparation

Page 34: Implementation of Tandem Mass Spectrometry: Quality

Derivatized Method

1. Add Internal Standard (critical step) we use 12-channel pipetter. 2. Extract 20 minutes on a shaker (room temperature). 3. Transfer extract to a fresh plate. (We keep original plate). 4. Evaporate Solvent using warm air and gentle warming. 5. Add Butanolic HCl, cover (we use Teflon plates) and heat to

approximately 60 C for 20 minutes (ovens corrode over time). 6. Evaporate Butanolic HCl using warm air. 7. Add reconstitution solvent. 8. Cover with Aluminum foil and analyze by MS/MS.

Page 35: Implementation of Tandem Mass Spectrometry: Quality

Documentation -- Preparation

Page 36: Implementation of Tandem Mass Spectrometry: Quality

Documentation -- MS/MS Analysis

Page 37: Implementation of Tandem Mass Spectrometry: Quality

Derivatized Method, Challenges or Hints

0102030405060708090

1stQtr

2ndQtr

3rdQtr

4thQtr

EastWestNorth

High Quality Control C4

0.50

1.00

1.50

2.00

2.50

3.00

3.50

0 50 100 150 200

Day

Con

c (µ

mol

e/L)

9413

9412

9405

9406

LCL

UCL

• We have seen high C4 when samples were extracted in polystyrene plates.

• Overheating can cause high leucine results (QC results may appear normal)

• We have observed low response for Methionine internal standard. This is under investigation.

Page 38: Implementation of Tandem Mass Spectrometry: Quality

Data Reduction and Evaluation

• Objectives– Determine Concentration of Analytes– Incorporate Results into a Database of Samples– Compare Values to a set of Rules– Produce Reports of Normal and Abnormal

Samples

Page 39: Implementation of Tandem Mass Spectrometry: Quality

Data Reduction and Evaluation

• First Step –review all the “chromatograms.”– Missed injections– Failure of the pump program

• Data Conversion – Neolynx generates a tab-delimited file that can

be read by Excel or a database program. – Ratios can be calculated by the MS software,

Excel, or the database.

Page 40: Implementation of Tandem Mass Spectrometry: Quality

Evaluation of Quality Control Samples.

• Quality Control Plan• Documents quality control decisions

– Acceptable plate quality control• Control results within ± 3 sd• Allow some number outside ± 3 sd

• Latitude in decision making• Document decisions• Checked and Reviewed Daily• Reviewed for trends

Page 41: Implementation of Tandem Mass Spectrometry: Quality

Evaluation of Quality Control Samples.

Page 42: Implementation of Tandem Mass Spectrometry: Quality

Analysis of TrendsSample Average Phenylalanine

0.25

0.45

0.65

0.85

1.05

1.25

1.45

1.65

1.85

0 50 100 150 200 250 300 350

Day

Co

nc (

mg

/dL

)

Average High Quality Control Phenylalanine

2

3

4

5

6

7

8

9

10

0 50 100 150 200 250 300 350

Day

Co

nc (

mg

/dL

)

9413

9412

9405

9406

LCL

UCL

Page 43: Implementation of Tandem Mass Spectrometry: Quality

Evaluation of Abnormal Results

Page 44: Implementation of Tandem Mass Spectrometry: Quality

Evaluation of Results

Page 45: Implementation of Tandem Mass Spectrometry: Quality

References

• Morbidity and Mortality Weekly Report, April 13, 2001

• Clinical Laboratory Improvement Amendments• Clinical Laboratory Evaluation Program• Basic QC Practices, 2nd Edition, James O.

Westgard, Ph.D.

Page 46: Implementation of Tandem Mass Spectrometry: Quality

Summary

• Plan Ahead• Train & Educate• Rigorous Method Validation is Required• Start with well defined rules and written Standard

Operating Procedures• Be consistent• Follow long-term trends as well as the daily

results.