Imaging Procedure 1
Final Revision
![Page 2: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/2.jpg)
Main Factors Affecting Recorded Image Detail
• kVp & mAs• Motion• Object Unsharpness• Focal Spot Size• SID (Source to Image Distance)• OID (Object to Image Distance)• Material Unsharpness
![Page 3: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/3.jpg)
![Page 4: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/4.jpg)
Size Distortion & OID• If source is kept constant, OID will affect
magnification• As OID , magnification • The farther the object is from the film, the
more magnification
![Page 5: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/5.jpg)
Image Distortion• When the part to be imaged – does not lay
parallel with the IR (cassette)
• If the Central Ray is not perpendicular to the part
![Page 6: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/6.jpg)
![Page 7: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/7.jpg)
FOCAL SPOT ANGLE
SMALLER ANGLE – SMALLER BEAM AT PATIENT
![Page 8: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/8.jpg)
Film Speed / Crystal size• Larger crystals or Thicker crystal layerFaster response= less detail, and less exposure (chest x-ray)• Finer crystals / thinner crystal layer
=Slower response, greater detail, more exposure (extremity)
![Page 9: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/9.jpg)
Axial Skelton- 80 bones• Cranium-8• Facial-14• Hyoid-1• Auditory ossicles-6• Cervical vert.-7• Thoracic vert.-12• Lumbar vert.-5• Sacrum-1• Coccyx-1• Sternum-1• Ribs-24 Total-80
![Page 10: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/10.jpg)
Anatomic Position• Upright, arms adducted,
palms forward, head and feet directed straight ahead
• Viewing Radiographs: Display x-rays so that the
patient is facing the viewer in anatomic position
R
![Page 11: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/11.jpg)
Body Surfaces and Parts
• Posterior or dorsal• Anterior or ventral• Plantar- sole of foot• Dorsal- top of anterior surface of foot, back or
posterior aspect of hand• Palmar- palm of hand or the anterior/ventral
surface
![Page 12: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/12.jpg)
Image Markers and Patient ID
• All films should have two markers:– Patient ID and date– Anatomic side markers
![Page 13: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/13.jpg)
Most Useful Applications for Plain X-Rays
• Chest• Musculoskeletal• Abdomen: limited usefulness
![Page 14: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/14.jpg)
Main Uses of Fluoroscopy
• Gastrointestinal Imaging• Genitourinary Imaging• Angiography• Other
– Intraoperative– Foreign body removal– Musculoskeletal
![Page 15: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/15.jpg)
Modified Barium Swallow
![Page 16: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/16.jpg)
Upper GI Exam
• Evaluates esophagus, stomach and duodenum• Double or Single Contrast• Can be combined with small bowel series• Largely replaced by endoscopy and cross-
sectional imaging• Fairly insensitive
![Page 17: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/17.jpg)
Cystogram
• Usually in adult patients• Looking for tear or intraluminal mass• Catheter placed and bladder filled with
contrast to capacity: usually 300-500 ml.• Spot films obtained when full• Post void film: usually overhead
![Page 18: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/18.jpg)
Cystogram with Intraperitoneal Rupture
Cystogram
![Page 19: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/19.jpg)
Techniques - Projection
•P-A (relation of x-ray beam to patient)
![Page 20: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/20.jpg)
Techniques - Projection (continued)
• Decubitus
![Page 21: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/21.jpg)
Rotation (continued)
![Page 22: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/22.jpg)
Heart
Right border: Edge of (r) Atrium
3. Left border: (l) Ventricle + Atrium
4. Posterior border: Reft Ventricle
5. Anterior border: Right Ventricle
![Page 23: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/23.jpg)
Heart (continued)
![Page 24: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/24.jpg)
Heart
•Size:
![Page 25: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/25.jpg)
Heart •Size of heart •Size of individual chambers of heart
•Size of pulmonary vessels
•Evidence of stents, clips, wires and valves
•Outline of aorta and IVC and SVC
![Page 26: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/26.jpg)
Procedure
• Apply proper compression to produce uniform breast thickness– Essential to high-quality mammograms
• Place ID markers
![Page 27: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/27.jpg)
Routine mammography projections
Craniocaudal (CC)
Mediolateral oblique (MLO)
![Page 28: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/28.jpg)
CT Protocolling
• Variables– Plain or contrast enhanced– Slice positioning– Slice thickness– Slice orientation– Slice spacing and overlap– Timing of imaging and contrast administration– Reconstruction algorhithm– Radiation dosimetry
![Page 29: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/29.jpg)
CT Terminology
• Window Width– Number of Hounsfield units from black to white
• Level or Center– Hounsfield unit approximating mid-gray
![Page 30: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/30.jpg)
CT Terminology
![Page 31: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/31.jpg)
![Page 32: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/32.jpg)
Clinical Applications
• Multisection CT aquires changes in the planning and staging of patient examinations. The scanning time is reduced for most examinations, requiring adjustments in the administration of intravenous contrast material. The amount of contrast material can be reduced, and different vascular phases can be better visualized. Thin-section scanning allows production of high-quality MPR images
![Page 33: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/33.jpg)
• Couinaud’s hepatic segments divide the liver into eight segments . The hepatic veins are the longitudinal boundaries . The transverse plane is defined by the right and left portal pedicles .
![Page 34: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/34.jpg)
Sonographic Evaluation of the Liver
• Evaluation of the hepatic structure is one of the most important procedures in sonography for many reasons. The normal , basiclly homogenerous parenchyma of the liver allows imaging of the neighboring anatomic structures in the upper abdomen.
![Page 35: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/35.jpg)
• The appropriate transducer depends on the patient’s body habitus and size
• The average adult abdomen usually requires a 3.5MHz
![Page 36: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/36.jpg)
Sagittal image of gallbladder
![Page 37: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/37.jpg)
• Doppler ultrasound is used primarily to identify and evaluate blood flow in vessels• velocity and waveform information can be used to evaluate stenoses, resistance and vessel patency
Doppler Frequency Shift (fD)
where f0 is the original frequency and f is the frequency of the returning echo(from RBC)
f0
f
(RBC)
• maximum shift is obtained when = 0°• minimum shift is obtained when = 90° (recall, imaging of strongest echoes)• the shift is comparatively small and typically is between 0 - 15 kHz (audible sound)• the shift is positive when the RBC is moving towards the transducer and is negative when the RBC is moving away from the transducer
doppler frequency shift (fD) = change in frequency = (f0 - f)
![Page 38: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/38.jpg)
Color Flow Doppler• Color Doppler is a hybrid that combines anatomicn obtained using B-mode system with flow information obtained using pulsed Doppler analysis• colors (blue and red) are assigned dependent on motion (toward or away) from the transducer• turbulence (i.e.: variations in flow direction) can vary between green and yellow• the depth of each color varies with the velocity of flow, stationary tissues appear gray• information is provided over a large area and superimposed on a gray scale image• color Doppler can detect flow in vessels too small to see by imaging alone• spectral analysis may also be obtained using commercial color Doppler systems• modern instruments incorporate both color Doppler and spectral Doppler
![Page 39: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/39.jpg)
• Oxygen, hydrogen, carbon, nitrogen elements
constitute 96 % of human body mass. • Oxygen is 65 % of body mass; carbon is 18.5
%, hydrogen 9.5 %, nitrogen 3.2 %.• Let us ignore all elements but Hydrogen.
![Page 40: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/40.jpg)
T1-weighted contrast
![Page 41: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/41.jpg)
T2-weighted contrast
![Page 42: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/42.jpg)
Tissue Appearance
WT FAT H2O MUSC LIG BONE
T1 B D I D D
Proton Density
I I I D D
T2 I B I D D
![Page 43: Imaging Procedure 1](https://reader035.vdocuments.us/reader035/viewer/2022062410/56816637550346895dd9a3d5/html5/thumbnails/43.jpg)