![Page 1: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/1.jpg)
ICMP, September 1, 2013 / Brighton, England 1 Image Quality in X-ray Detectors
Image Quality Parameters
and their Measurement
John M. Boone, Ph.D., FAAPM, FSBI, FACR
Professor and Vice Chair of Radiology Professor of Biomedical Engineering
University of California, Davis
![Page 2: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/2.jpg)
2 Image Quality
![Page 3: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/3.jpg)
3 Image Quality
![Page 4: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/4.jpg)
4
Image Quality Parameters
and their Measurement Linear Systems / Stationary Behavior
Spatial Resolution
field measurements
scientific characterization
Contrast Resolution (noise)
field measurements
scientific characterization
Signal to Noise Ratio
field measurements
scientific characterization
Summary
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 5: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/5.jpg)
5
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 6: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/6.jpg)
6
Response of a 400 speed screen-film system (Fuji)
![Page 7: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/7.jpg)
7
Response of a Varian 4030CB Flat Panel Detector
high gain mode
low gain mode
![Page 8: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/8.jpg)
8
Response of a Fuji Computed Radiography (CR) System
![Page 9: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/9.jpg)
9
stationary
non-stationary
Image Quality
point spread functions (PSFs)
![Page 10: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/10.jpg)
10
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 11: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/11.jpg)
11
Field Measurement
![Page 12: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/12.jpg)
12
spatial resolution
line pair phantom
star pattern phantom
wedge
resolution test tools
Image Quality
![Page 13: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/13.jpg)
system input
system output
blurring
Spatial Resolution: Measurements in the Field
![Page 14: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/14.jpg)
modulation
14
![Page 15: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/15.jpg)
15
can resolve 4 bars?
yes
heck no
no
yes
yes
yes
limiting spatial resolution
![Page 16: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/16.jpg)
16
Limiting spatial resolution (~10% of MTF)
![Page 17: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/17.jpg)
17
Scientific Characterization
![Page 18: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/18.jpg)
Image Quality
The Point Spread Function (PSF)
18
![Page 19: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/19.jpg)
Image Quality
The Line Spread Function (LSF)
19
![Page 20: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/20.jpg)
The Edge Spread Function (ESF)
Image Quality 20
![Page 21: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/21.jpg)
Image Quality
21
![Page 22: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/22.jpg)
22
The Line Spread Function – lsf(x) – “slit image”
measures focal spot resolution measures detector resolution
slit
slit
focal spot
detector
{
![Page 23: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/23.jpg)
The Line Spread Function – lsf(x) – “slit image”
resulting image of slit measurement geometry
![Page 24: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/24.jpg)
The Line Spread Function – lsf(x) – “slit image”
resulting image of slit actual lsf(x)
position (mm)
gray
sca
le (d
n)
orthogonal trace thru slit
![Page 25: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/25.jpg)
Fourier Representation
25
![Page 26: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/26.jpg)
26
different phase
different amplitude
different frequency
sine waves 101
![Page 27: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/27.jpg)
27
2( )( )
( )
ifxlsf x e dxOTF f
lsf x dx
∞ − π
−∞∞
−∞
= ∫∫
( ) ( )MTF f OTF f=
2 2( ) [ ( )] [ ( )]MTF f OTF f OTF f= ℜ + ℑ
complex
real imaginary real
![Page 28: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/28.jpg)
28
lsf(x)
MTF(f)
2( )
( )
ifxlsf x e dx
lsf x dx
∞ − π
−∞∞
−∞
∫∫
normalizes area to 1.0
![Page 29: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/29.jpg)
modulation transfer function
29
![Page 30: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/30.jpg)
30
top view of detector array
side view
Direct Digital Detectors (eg Selenium)
a
aperture – lsf(x)
lsf(x) = RECT{x/a}
∆
pitch – Nyquist
FN = 1 2 ∆
![Page 31: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/31.jpg)
31
![Page 32: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/32.jpg)
32
other considerations
Pre-sampled LSF / ESF
Type of Image Processing
flat fielding OK
dead pixel correction OK
“for presentation” not OK
Paralax issues
![Page 33: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/33.jpg)
33
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 34: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/34.jpg)
34
Field Measurement
![Page 35: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/35.jpg)
35
contrast resolution
1.2 µR/frame 1.8 µR/frame 3.8 µR/frame
![Page 36: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/36.jpg)
36
( )2
2
1
1 ( , )1
N
ii
I x y IN =
σ = −− ∑
2σ = σ
N = number of pixels in ROI hi(x,y): the data in the ROI
“noise”
First Order Determination of Noise
region on interest (ROI)
digital image
![Page 37: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/37.jpg)
37
Scientific Characterization
![Page 38: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/38.jpg)
σ σ
But “noise”, or the standard deviation σ, does not tell the whole story
38
![Page 39: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/39.jpg)
39
( )( )2 2 ( )
1
1( , ) ( , ) x yN
i f x f yx y ix y
iNPS f f I x y I e dxdy
N∞ ∞ − π +
=−∞ =−∞=
= −∑ ∫ ∫
The Noise Power Spectrum (NPS) or Wiener Spectrum
10 20 30 40 50 60
10
20
30
40
50
60
N regions on interest (ROIs)
10 20 30 40 50 60
10
20
30
40
50
60
10 20 30 40 50 60
10
20
30
40
50
60
10 20 30 40 50 60
10
20
30
40
50
60
average
analytical equation
![Page 40: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/40.jpg)
40
The Noise Power Spectrum (NPS) or Wiener Spectrum
10 20 30 40 50 60
10
20
30
40
50
60
N regions on interest (ROIs)
10 20 30 40 50 60
10
20
30
40
50
60
10 20 30 40 50 60
10
20
30
40
50
60
10 20 30 40 50 60
10
20
30
40
50
60
average
discrete implementation
![Page 41: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/41.jpg)
41
The Noise Power Spectrum (NPS) or Wiener Spectrum
10 20 30 40 50 60
10
20
30
40
50
60 radial average
spatial frequency (mm-1) N
PS (m
m2 )
![Page 42: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/42.jpg)
42
The Noise Power Spectrum (NPS) or Wiener Spectrum
10 20 30 40 50 60
10
20
30
40
50
60 radial average
spatial frequency (mm-1) N
PS (m
m2 )
![Page 43: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/43.jpg)
43
The Noise Power Spectrum (NPS) or Wiener Spectrum
spatial frequency (mm-1) N
PS (m
m2 )
σ2
2 ( )NPS f dfσ = ∫
![Page 44: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/44.jpg)
uncorrelated noise correlated noise
convolution
44
![Page 45: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/45.jpg)
45
![Page 46: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/46.jpg)
46
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 47: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/47.jpg)
47
Field Measurement
![Page 48: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/48.jpg)
48
Signal difference to noise ratio (SDNR)
Contrast to Noise Ratio (CNR)
signal
background
![Page 49: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/49.jpg)
detail detail detail
cont
rast
cont
rast
cont
rast
A B C
the contrast detail curve
Image Quality 49
![Page 50: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/50.jpg)
the contrast detail curve
interpretation
50
lower dose higher dose
![Page 51: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/51.jpg)
51
Scientific Characterization
![Page 52: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/52.jpg)
52
22 ( )( )
( )outMTF fSNR fNPS f
=
2 2 ( )( )( )
k MTF fNEQ fNPS f
=
The Noise Equivalent Quanta (NEQ)
![Page 53: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/53.jpg)
53
2 2 ( )( )( )
k MTF fDQE fq NPS f
=
( )( ) NEQ fDQE fq
=
2
2
( )( )( )
out
in
SNR fDQE fSNR f
=2 ( )inSNR f q=
![Page 54: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/54.jpg)
54
2 2 ( )( )( )
k MTF fDQE fq NPS f
=
k = mean GS value
q = mean photons/mm2 for this acquisition
Image acquired to compute NPS(f)
![Page 55: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/55.jpg)
55
detective quantum efficiency (DQE)
![Page 56: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/56.jpg)
56
2.94 x 107 photons / mm2
120 kV / 1.0 mGy air kerma incident
To determine q: …some spectral modeling required
![Page 57: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/57.jpg)
57 IEC values courtesy Nicole T. Ranger on line presentation
RQA3 50 kV 10 4.0 RQA5 70 kV 21 7.1 RQA7 90 kV 30 9.1 RQA9 120 kV 40 11.5
IEC: International Electrotechnical Commission (IEC 61267-1)
Radiation Quality
Designation Tube Voltage
~added Filtration (mm Al)
HVL (mm Al)
150
cm
DQE measurement geometry
![Page 58: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/58.jpg)
58
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary
![Page 59: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/59.jpg)
59
Field measurements of image quality are adequate
but are subjective and imprecise
Scientific measurements are useful in detector evaluation
but their quantitative nature is desirable
Digital radiography QC should convert to quantitative measures in time
Clinical Medical Physicists should understand MTF, NPS, NEQ, DQE
Summary
![Page 60: Image Quality Parameters and their MeasurementImage Quality in X -ray Detectors ICMP, September 1, 2013 / Brighton, England 1 Image Quality Parameters . and their Measurement . John](https://reader030.vdocuments.us/reader030/viewer/2022041122/5f371084c6f460693a2fea59/html5/thumbnails/60.jpg)
60
Image Quality Parameters
and their Measurement
Linear Systems / Stationary Behavior
Spatial Resolution
Contrast Resolution (noise)
Signal to Noise Ratio
Summary