Cuándo, a quién y cómo explanto un catéter de desfibrilador Sergio L. Pinski Cleveland Clinic Florida Weston, Florida, USA

Upload: sergio-pinski

Post on 02-Jun-2015

947 views

Category:

Health & Medicine


1 download

TRANSCRIPT

Page 1: Icd lead extraction ba nov 2012 final

Cuándo, a quién y cómo explanto un catéter de

desfibrilador

Sergio L. PinskiCleveland Clinic FloridaWeston, Florida, USA

Page 2: Icd lead extraction ba nov 2012 final

Simpler Classification of Simpler Classification of IndicationsIndications Infection Lead that represents a risk Need for venous conduit Superfluous lead during revision

or upgrade (abandoned or with failure)

Page 3: Icd lead extraction ba nov 2012 final
Page 4: Icd lead extraction ba nov 2012 final

Conductor fracture secondary to compression

Insulation defect and conductor fracture secondary to pinching

Gradau et al. PACE 2003;26:649

Page 5: Icd lead extraction ba nov 2012 final

Insulation defect due to abrasion

Insulation defect due to traction from lead migration

Gradau et al. PACE 2003;26:649

Page 6: Icd lead extraction ba nov 2012 final
Page 7: Icd lead extraction ba nov 2012 final

Figure 1

Hauser et al. Heart Rhythm 2011; 9:742

Page 8: Icd lead extraction ba nov 2012 final

Kaplan-Meier curve for all-cause lead removal or capping

Borleffs et al. Circ Arrhythm Electrophysiol 2009;2:411

Page 9: Icd lead extraction ba nov 2012 final

Kaplan-Meier curve for lead failure comparing all leads with the leads from Boston Scientific, Medtronic, and

St Jude Medical, grouped by lead diameter

Page 10: Icd lead extraction ba nov 2012 final

Strategies for ICD lead Strategies for ICD lead problemsproblems

Ignore it, reprogram around Repair lead Add pacing/sensing lead Add new defibrillation lead Extract and replace defibrillation

lead

Page 11: Icd lead extraction ba nov 2012 final

Wollman et al. PACE 2005;28:795

Problems with additional pectoral pace/sense lead

Observation/Complication Pectoral Leads

(n = 97)

No. (%)

Oversensing with inappropriate therapy 5 5.2

Oversensing only 4 4.1

Infection 2 2.1

High-voltage defect 5 5.2

Insulation defect

Sensing/pacing threshold/impedance value out of range 3 3.1

Fracture

Dislocation 1 1

Impossible upgrading to DDD because of vein thrombosis

Death in septicemia with vegetations on leads 1 1

Total 21 21.7

Page 12: Icd lead extraction ba nov 2012 final

Model 6949 High Voltage Conductor Survival after a Pace-Sense Conductor

Fracture

Months after a pace-sense conductor fracture

0 6 12 18 21

% survivalof HV conductors

100 98.7 89 80.3 77

Page 13: Icd lead extraction ba nov 2012 final

Potential Drawbacks of Potential Drawbacks of Abandoning Non-Infected Abandoning Non-Infected ICD LeadsICD Leads Lead “chatter” causing spurious shocks Insulation breach in pocket causing

electrical shorting during high-voltage shock and generator damage

Large diameter could promote venous obstruction (SVC syndrome) or tricuspid regurgitation

Extraction will be very difficult if needed down the road

Page 14: Icd lead extraction ba nov 2012 final

Epstein et al. Circulation 1998;98:1517

Page 15: Icd lead extraction ba nov 2012 final

Epstein et al. Circulation 1998;98:1517

Page 16: Icd lead extraction ba nov 2012 final

Epstein et al. Circulation 1998;98:1517

Page 17: Icd lead extraction ba nov 2012 final

Epstein et al. Circulation 1998;98:1517

Page 18: Icd lead extraction ba nov 2012 final

Epstein et al. Circulation 1998;98:1517

Page 19: Icd lead extraction ba nov 2012 final
Page 20: Icd lead extraction ba nov 2012 final
Page 21: Icd lead extraction ba nov 2012 final

Hauser et al. Europace 2010;12:395

Page 22: Icd lead extraction ba nov 2012 final
Page 23: Icd lead extraction ba nov 2012 final
Page 25: Icd lead extraction ba nov 2012 final

Wollman et al. JCE 2007;18:1172

Perioperative complications after ICD lead replacement

Additional ICD lead

(n = 33)

Replaced ICD lead

(n = 53)

% %

Lead dislodgement 3.1

Pneumothorax 6.3 5.7

Pericardial effusion 3.1 1.9

Pocket hematoma 3.1 3.6

Pleural effusion 3.1

Incomplete extraction 3.6

Total 18.2 15.1

Page 26: Icd lead extraction ba nov 2012 final

REPLACE: major complications by lead addition or revision.

Poole J E et al. Circulation 2010;122:1553-1561

Page 27: Icd lead extraction ba nov 2012 final

Factors to ConsiderFactors to Consider Patient age and gender (elderly women higher

risk) Overall health Previous cardiac surgery Number of leads in the intravascular space Duration of the implant Fragility, condition, and physical characteristics of

the lead Experience of the physician Desires of the patient

Page 28: Icd lead extraction ba nov 2012 final

My “Rules of Thumb”My “Rules of Thumb”

No more than 5 leads through SVC No more than 4 leads from one

side No more than one defibrillation

lead

Page 29: Icd lead extraction ba nov 2012 final
Page 30: Icd lead extraction ba nov 2012 final
Page 31: Icd lead extraction ba nov 2012 final

Figure 1

Subacute lead perforation

Danik et al. Heart Rhythm 2008;5:1667

Page 32: Icd lead extraction ba nov 2012 final
Page 33: Icd lead extraction ba nov 2012 final
Page 34: Icd lead extraction ba nov 2012 final

Figure 1

Generator exchange is associated with an increased rate of

Fidelis lead failure

Lovelock et al. Heart Rhythm 2012;9:16157

Page 35: Icd lead extraction ba nov 2012 final

What to do during elective What to do during elective replacement upgrade of ICD with replacement upgrade of ICD with not failed Fidelis lead?not failed Fidelis lead?

“Benign neglect” “The minimalist” New pace-sense lead “The next fool’s problem” New ICD lead “The all-the-way out” Extraction and

new lead “The switch” -LV lead in RV port “The downgrade” CRTP, no risk of

spurious shocks

Page 36: Icd lead extraction ba nov 2012 final

Programmable sensing vector in new MDT ICDs

Page 37: Icd lead extraction ba nov 2012 final

Patient “downgraded” from CRTD to CRTP 13 months ago. Fidelis lead programmed unipolar pace/sense. Lead fracture to the tip diagnosed during routine clinic check.

Page 38: Icd lead extraction ba nov 2012 final

BiV capture

Native LBBB

LV pacing with fusion

Page 39: Icd lead extraction ba nov 2012 final

Physicians who perform lead Physicians who perform lead extraction change their extraction change their implant techniquesimplant techniques

Use leads with excellent long-term track records.

Exclusive use of active-fixation leads Avoid dual-coil defibrillation leads Use leads “easier” to extract

Page 40: Icd lead extraction ba nov 2012 final
Page 41: Icd lead extraction ba nov 2012 final

Azygous vein coil for high defibrillation threshold

Page 42: Icd lead extraction ba nov 2012 final