Page 1: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Integrated

Management &

Educational

Consultancy

Services

  “Radiology Update Seminarfor The Dental Team”

  

 

Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email: [email protected]

www.imecs-cpd.co.uk

Page 2: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

What do dentists need to know?

• UK radiation dose distributions• Nature and effects of radiation• Clinical implications of IRR1999 and

IR(ME)R 2000• Patient safety and selection criteria• Optimisation of radiographic techniques• Quality assurance and record keeping

www.imecs-cpd.co.uk

Page 3: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

What do nurses need to know?

• Principle foundations of dental radiographic techniques

• Equipment used• Methods of processing films and the

practice of digital radiography• Hazards of exposure to ionising radiations• Have an understanding of the regulatory

requirements

www.imecs-cpd.co.uk

Page 4: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

What must nurses be able to do?

• Prepare the equipment and materials• Prepare and care for patients• Process radiographs• Properly mount radiographs• Undertake quality assurance measures in

processing

www.imecs-cpd.co.uk

Page 6: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

UK Dose Distribution

83%

17%

Comparison between dose sources in the UK

www.imecs-cpd.co.uk

Page 7: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Man Made Dose Distribution

90%

10%

• Medical dose compared with other man made sources

www.imecs-cpd.co.uk

Page 8: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Comparing Medical Doses

97%

3%

• General medical dose compared with dental

www.imecs-cpd.co.uk

Page 9: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Radiological Examinations in the UK

30%

70%

• Number of general medical examinations compared with dental

www.imecs-cpd.co.uk

Page 10: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Potential for Dose Saving

10%

45%

45%• Up to 50% of the medical dose delivered in

the UK is unnecessary

www.imecs-cpd.co.uk

Page 11: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Workshop

• List the characteristics of x-rays

e.g. travels in straight lines

www.imecs-cpd.co.uk

Page 12: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Nature of Ionising Radiation

• Travels in straight lines at the speed of light• Obeys the inverse square law• Penetrates, excites and ionises matter• Chemical and biological effect• Photographic and fluoroscopic effect• Unaffected by magnetic or electric fields

www.imecs-cpd.co.uk

Page 13: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Electromagnetic Spectrum

low high

Energy

radiowaves

visible light

cosmic rays

microwaves

X & gamma rays

www.imecs-cpd.co.uk

Page 14: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

X-Ray Tube

anodecathode

target

X-ray beam

electron beam

Page 15: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Workshop

• What units are used to measure radiation exposure and dose ?

www.imecs-cpd.co.uk

Page 16: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Units of Radiation Measurement

• Exposure = COULOMBS PER KILOGRAM Kinetic Energy Released per unit Mass (KERMA)

• Absorbed Dose = GRAY Joules per Kilogram

www.imecs-cpd.co.uk

Page 17: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Dose Reference Levels

• PA DRLs DOH 4Mgy

HPA 2mGy

Recommended 1 – 1.5mGy

OPG DRLs DOH 66mGy

HPA 20mGy

Recommended 10 – 15mGy

Page 18: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Units of Radiation Measurement

• Exposure = COULOMBS PER KILOGRAM Kinetic Energy Released per unit Mass (KERMA)

• Absorbed Dose = GRAY Joules per Kilogram

• Dose Equivalent = SIEVERT Gray x Q (x-rays=1, protons=10, alpha=20)

• Effective Dose = SIEVERT Biological Effect (dose equivalent x tissue sensitivity)

www.imecs-cpd.co.uk

Page 19: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Dose Reference Levels

• PA DRLs DOH 4Mgy

HPA 2mGy

Recommended 1 – 1.5mGy

(Effective dose 0.002mSv)

OPG DRLs DOH 66mGy

HPA 20mGy

Recommended 10 – 15mGy

(Effective dose 0.007mSv)

Page 20: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Units of Radiation Measurement

• Exposure = COULOMBS PER KILOGRAM Kinetic Energy Released per unit Mass (KERMA)

• Absorbed Dose = GRAY Joules per Kilogram

• Dose Equivalent = SIEVERT Gray x Q (x-rays=1, protons=10, alpha=20)

• Effective Dose = SIEVERT Biological Effect (dose equivalent x tissue sensitivity)

www.imecs-cpd.co.uk

Page 21: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Round collimator at 6cms diameter = 29cm

Page 22: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

DRL = mGy/cm2

Page 23: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

DRL = mGy/cm2

Page 24: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Thyroid collars

Page 25: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Effective Dose and Kilo-voltage

• 60 kV 70kV

Page 26: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

Page 27: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

X-ray photon

Page 28: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

X-ray photon

Ionised electron

Page 29: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

Page 30: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

Page 31: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Excitation and Ionisation

Page 32: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Radiobiology

• Random interactions causing direct and indirect tissue damage

• Critical body macromolecules: protein, DNA (damage with incomplete or misrepair)

• H2O affected because of large size and content

• Ionised water becomes chemically reactive and can affect DNA by indirect action

• Formation of free radicals: HO2 (peroxyls) H2O2 (peroxides)

www.imecs-cpd.co.uk

Page 33: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Ionisation and free radicals

• RH + Radiation = RH+ + electron

R H

H2O HO H

Page 34: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

RadiobiologyThe exposure of a cell to radiation may result in:

• The premature death of the cell• Prevention or delay in cell division• Changes in cell structure and/or functions

which may be passed on to daughter cells

www.imecs-cpd.co.uk

Page 35: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

What is damaged?

• DNA

• Any of the cytoplasmic organelles

• I bet they say MITOCHONDRIA

www.imecs-cpd.co.uk

Page 36: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

What Type of Damage Can Occur?

• Stochastic• Deterministic

www.imecs-cpd.co.uk

Page 37: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Who is Affected?

• Somatic• Genetic

www.imecs-cpd.co.uk

Page 38: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Variations with Age at Exposure

Detriment vs Age at Exposure

0 to 5years

5 to 10years

10 to 20years

20 to 30years

30 to 40years

40 to 50years

50 to 60years

www.imecs-cpd.co.uk

Page 39: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Integrated

Management &

Educational

Consultancy

Services

  “Radiology Update Seminarfor The Dental Team”

  

 

Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email: [email protected]

www.imecs-cpd.co.uk

Page 40: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Legislative Framework

• Two sets of regulations have been introduced in the UK in the last 13 years

• Ionising Radiations Regulations 1999 (IRR 99)

• Ionising Radiations (Medical Exposures) Regulations 2000 (IR(ME)R 2000)

www.imecs-cpd.co.uk

Page 41: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

IRR 1999

• Is about:– Environment– Workers – Members of the public

www.imecs-cpd.co.uk

Page 42: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Public Protection

Page 43: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

IRR 1999

• Is about:– Environment– Workers – Members of the public– Management of radiationm protection

www.imecs-cpd.co.uk

Page 44: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

IRR 1999

• HSE notification • Risk assessment

www.imecs-cpd.co.uk

Page 45: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Risk Assessment

Window

Exposure Switch

Mains isolator

4M

Surgery 2

Waiting room

Page 46: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Staff Risk Assessment

Page 47: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Staff Risk Assessment

Page 48: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

IRR 1999

• HSE notification • Risk assessment• Contingency plans• Dose restriction/limitation• Training• RPS/RPA • Controlled areas

www.imecs-cpd.co.uk

Page 49: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Controlled areas

Page 50: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

?Fire regulations

Page 51: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

How MUCH Lead ?

Page 52: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

IR(ME)R 2000

Is about:

• Patient protection

www.imecs-cpd.co.uk

Page 53: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Ionising Radiation (Medical Exposures) Regulations 2000

Summary

• Justification– Patient examination, identify clinical reason, written report

• Optimisation– Maximum yield with minimal dose

• Entitlement– Anyone can do anything as long as they are trained and it is agreed

and monitored

www.imecs-cpd.co.uk

Page 54: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Selection (Referral)Criteria

• SELECTIVE– Symptomatic to confirm or refute

• ROUTINE– Circumstantial evidence

• SCREENING– No evidence

www.imecs-cpd.co.uk

Page 55: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Optimisation

• Patient ID• Patient care• Check history, area• Admin• Justification• 70kV DC• Film holders

• Rectangular collimators• Film speed• Viewing• Maintenance & QA

– Equipment– Processor

• Audit• Training

www.imecs-cpd.co.uk

Page 56: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Digital Dental Radiography

• Direct– Charge Coupled Devices / CMOS – (wired systems)

• Indirect– Photo phosphor plates

www.imecs-cpd.co.uk

Page 57: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Digital Image Processing

• Analysis– Collect numeric information based on the image

• Enhancement– Improve appearance of image

• Encoding– Reduce information required to describe the image

www.imecs-cpd.co.uk

Page 58: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Ionising Radiation (Medical Exposures) Regulations 2000

Conceptual Framework• Clinical audit

– Radiographic and Radiological

• Expert advice– Medical Physics Expert

• Equipment inventory– Radiation Protection File

• Training– What, how and assessed

• Enforcement– Health & Safety @ Work

www.imecs-cpd.co.uk

Page 59: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Justification of an individual medical exposure

• No person shall carry out a medical exposure unless:

– It has been justified by a practitioner as showing sufficient net benefit.

– It has been authorised by a practitioner (where appropriate the operator)

– Has been authorised by the local research/ethics committee

– It is a part of an agreed medico-legal procedure

– In the case of a female of childbearing age, he has enquired whether she is pregnant, if relevant.

www.imecs-cpd.co.uk

Page 60: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Ionising Radiation (Medical Exposures) Regulations 2000

• Employer

• Referrer/Prescriber

• Practitioner

• Operator

www.imecs-cpd.co.uk

Page 61: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Duty of Role Holders

• Employer: to provide a framework for radiation protection of the patient through provision of standard operating procedures.

• Referrer/Prescriber: to provide sufficient relevant clinical information on the patient to enable the justification of the medical exposure.

• Practitioner: to justify each individual medical exposure.

• Operator: to undertake practical aspects of the medical exposure.

www.imecs-cpd.co.uk

Page 62: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Employer

• Any natural or legal person who, in the course of a trade, business or other undertaking, carries out (other than as an employee), or engages others to carry out, medical exposures or practical aspects, at a given radiological installation.

www.imecs-cpd.co.uk

Page 63: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Employer Procedures for Dentists

• Protocols and/or guidelines for:– Patient identification– Competencies of practitioners & operators– Medico-legal exposures– Women of childbearing age– Reduce accidental exposure to ALARP

• Expected doses and dose recording• Quality assurance

www.imecs-cpd.co.uk

Page 64: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Referrer

• A registered medical or dental practitioner, or other health care professional who is entitled in accordance with the employer’s procedures to refer individuals for a medical exposure to a practitioner.

www.imecs-cpd.co.uk

Page 65: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Practitioner

• A registered medical or dental practitioner, or other health professional who is entitled, in accordance with the employer’s procedures, to take responsibility for an individual medical exposure.

www.imecs-cpd.co.uk

Page 66: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Operator

• Any person who is entitled, in accordance with the employer’s procedures, to carry out practical aspects (including those to whom practical aspects have been allocated) except, when training, where they do so under the direct supervision of a person who is adequately trained.

www.imecs-cpd.co.uk

Page 67: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Adequate Training

• Practitioners and operators shall have successfully completed training, including theoretical knowledge and practical experience (in radiation protection and diagnostic or therapeutic techniques involving ionising radiation) appropriate to their specific area of practice.

www.imecs-cpd.co.uk

Page 68: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Quality AssurancePrimary FunctionsüConsistently high quality imagesüDoses to staff and patients are as low as reasonably practicable

Key Elementsü Darkroom, film and processingüX-ray equipmentü Working proceduresü Audits & Quality Targets

www.imecs-cpd.co.uk

Page 69: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Darkroom

• Film storage and stock rotation

• Light tight integrity (visual)

• Safelight (coin test)

www.imecs-cpd.co.uk

Page 70: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Quality Control:Equipment MonitoringWeekly clean with spirit based solution and checks for:

• Electrical: sockets, plugs, earthing, insulation, leads (mains, exposure), tube head (leaks).

www.imecs-cpd.co.uk

Page 71: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Quality Control:Equipment MonitoringWeekly clean with spirit based solution and checks for:

• Electrical: sockets, plugs, earthing, insulation, leads (mains, exposure), tube head (leaks).

• Mechanical: mountings, support, brakes, stability, fixtures and fittings (collimators).

• Radiation: output checks (annual), tube head (dents), mains isolation switch

www.imecs-cpd.co.uk

Page 72: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Trophy 70 Room 2 June 2010  5.6.10 12.6.10 19.6.10 26.6.10

Electrical        

Socket        

Plug        

Leads        

Tubes leaks        

Mechanical Security        

Wall support        

Tube arm        

Tube head        

Cone/collimators        

Radiation        

Tube dents        

Mains isolation        

Signature        

Date        

www.imecs-cpd.co.uk

Page 73: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Quality Control:Processor MonitoringWeekly clean with checks for:

• Electrical: sockets, plugs, leads, earthing and insulation.

• Lightproof: lid, feed cover, light screen (coin test)

• Transport: rollers and screens, daily cleaner.

• Image quality: temperature (heater, thermostat), activity (replenishment), wash, drainage, cleanliness, stepped wedge test.

www.imecs-cpd.co.uk

Page 74: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Velopex June 2006  5.6.06 12.6.06 19.6.06 26.6.06

Electrical        

Socket        

Plug        

Leads        

Lightproof        

Lid        

Sleeves        

Daily Cleaner        

Daily water drain        

Stepped wedge        

       

       

Signature        

Date        

www.imecs-cpd.co.uk

Page 75: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Stepped Wedge

This is what the wedge looks like

This is what a picture of it should look like

www.imecs-cpd.co.uk

Page 76: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

www.imecs-cpd.co.uk

Page 77: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Accessory Equipment

• Cassettes• Screens• Viewing boxes• Film holders

www.imecs-cpd.co.uk

Page 78: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Quality rating and quality standards for radiographs

Rating/Quality Basis

1. Excellent: No errors of exposure, positioning or processing. Not less than 70%.

2. Acceptable Some errors which do not detract from the diagnostic use. Not greater than 20%.

3. Unacceptable Errors which render the radiograph diagnostically unacceptable.

Not greater than 10%.www.imecs-cpd.co.uk

Page 79: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Guidelines for Image Critiques

• Area: film position and orientation

(axis, border, blip)

• Projection: beam centring and

angulation

www.imecs-cpd.co.uk

Page 80: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Critique• Projection: beam centring and angulation

• (no cone cut, crowns in profile/3mm border, no

interproximal overlap, bite block in profile)

3 mm

Page 81: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Guidelines for Image Critiques

• Area: film position and orientation (axis, border, blip)

• Projection: beam centring and angulation

• Density: level of blackness

• Contrast: difference in density for a range of structures

• Sharpness: level of detail

• Artifacts: something on the film that shouldn’t be there

www.imecs-cpd.co.uk

Page 82: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Orthopantomography

Page 83: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Plain Radiography

Page 84: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Scanography

Page 85: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Scanography

Page 86: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Plain Radiography

Page 87: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Tomography

Page 88: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Tomography

Page 89: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Focal Trough

Focal troughPivotal Plane

Page 90: I ntegrated M anagement & E ducational C onsultancy S ervices “Radiology Update Seminar for The Dental Team” Dr. Richard DeCann & Mr. Tim Reynolds IMECS

Principles of Radiation Protection

• Justification of the exposure• Positive identification of the patient• Reference to existing radiological records• Shielding: primary & secondary barriers• Distance• Beam restriction• Minimum number of personnel in the examination room• Quality assurance of all equipment and audit of activities• Choice of x-ray and imaging equipment

www.imecs-cpd.co.uk