Page 1: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Humidity Measurement in H2O2Bio-Decontamination

– Relative Saturation as the Key

Page 2: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Meet the Presenters

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2

Piritta MaunuLife Science Regulatory and Industry Expert at Vaisala with over 15 years of experience in biotechnology and life science applications.

Sanna LehtinenProduct Manager at Vaisala with 20 years of experience in life science applications and wide product management experience from leading international high tech companies.

Page 3: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Basics of bio-decontamination with vaporized H2O2

The importance of continuous measurement during bio-decontamination

Patented PEROXCAP® technology

The difference between Relative Humidity (RH%) and Relative Saturation (RS%)

Five most typical pitfalls in humidity measurement during bio-decontamination

Q&A session

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 3

Agenda

Page 4: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Basics of Bio-Decontamination with Vaporized H2O2

Page 5: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Why Use H2O2 for Bio-Decontamination

Easy to use

Destroys all biological contaminants

Works in low temperature processes

Processes can be validated

Compatible with a wide variety of materials

Environmentally friendly process

Leaves no real residues – only water vapor and oxygen

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 5

Page 6: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Isolators, RABS Transfer hatches, chambers Vapor generators

• Pharma / Manufacturing of pharmaceuticals:

• Aseptic filling• Sterility testing• Freeze dryers

• Pharmacy compounding• Blood and tissue banks

• Clenrooms• Healthcare (like hospitals)• Animal trials (GLP)• Service providers• HVAC• Transportation

• Ambulance, airplane, cruisers, trucks

• Army • Farming / Animal husbandry• Construction

• Hospitals• Pharma / Manufacturing of pharmaceuticals• Cleanrooms• Animal trials (food and supplies)• Army• Blood and tissue banks• Pharmacy compounding

Incubators• Pharma / GLP, GCP• Blood and tissue banks• Scientific research

Production lines• Processing plants• Filling lines• Milking machines

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 6

Page 7: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Bio-Decontamination Measurements

H2O2, ppm Humidity Temperature

Potential other parameters: Time Pressure (P) and differential pressure (dP) Airflow and velocity Airborne particles

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 9

Our main topic today

Page 8: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

The Importance of Continuous Measurement During Bio-Decontamination

Page 9: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Why Repeatable Online Measurements?

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 11

Provides continuous measurement data

Guarantees that a process works as planned

May decrease a number of biological, chemical or enzymatic indicatorsFrom monitoring to controlling

Page 10: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Indicators vs. Online MeasurementsChemical indicators (CI)

Biological indicators (BI)

Enzymatic indicators (EI)

Measurement sensor; Vaisala HPP272

What the product looks like?PROS Inexpensive

Easy to useQuantitative results Quantitative results

Instantaneous reactionContinuous, stable andrepeatable measurement

CONS Tells only ”±” resultNot quantitative results

Takes 7 days to get resultsNeeds qualified personnel and laboratory premises

Manual work neededDoesn’t give continuous measurement data

Doesn’t tell if micro-organisms are dead;combine with BIs/EIs

What does thisproduct measure?

Change in color; H2O2 concentration

Reduction of micro-organisms (SAL min. 10-6)

Reduction of micro-organisms (SAL min. 10-6)

H2O2 ppm concentration, RH/RS and temperature

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 12

Page 11: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Typical non-condensing bio-decontamination process in isolators

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 13

Example Bio-Decontamination Cycle

Page 12: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Patented PEROXCAP®

Technology

Page 13: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

The Vaisala PEROXCAP® sensor technology is based on two capacitive thin-film polymer sensors, building on the reliable HUMICAP® technology.

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 17

Sensor 1 Sensor 2

Page 14: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 18

H2O2molecules

H2O molecules

Catalytic layer on top of sensor 2

New Measurement Technology; PEROXCAP®

Page 15: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 19

H2O2

H2O

Page 16: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 20

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

Page 17: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 21

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

Page 18: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 22

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

Page 19: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 23

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

Page 20: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 24

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

ALGORITHMS

RELATIVE HUMIDITY, RH%

Page 21: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 25

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

ALGORITHMS

RELATIVE HUMIDITY, RH%

H2O2 CONCENTRATION,

ppm

Page 22: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

New Measurement Technology; PEROXCAP®

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 26

HUMIDITYSENSOR 1

HUMIDITYSENSOR 2

H2O2

H2O

CATALYTIC LAYER

ALGORITHMS

RELATIVE HUMIDITY, RH%

RELATIVE SATURATION, RS%

H2O2 CONCENTRATION,

ppm

Page 23: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

The Difference Between Relative Humidity (RH%) and Relative Saturation (RS%)

Page 24: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Relative Saturation vs. Relative Humidity

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 28

Only H2O vapor present H2O and H2O2 vapor present

Page 25: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

RELATIVE SATURATION %

100%

0%

Condensation

RS% value is the only parameter for controlling condensation when H2O2vapor is present.

Page 26: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Maximum %RH

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 30

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm)

At every pointRS = 100% RS

Page 27: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 31

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm)

~ 53% RH

At every pointRS = 100% RS

Maximum %RH

Page 28: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 32

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm)

~ 53% RH

~ 72% RH

At every pointRS = 100% RS

Maximum %RH is Dependent on Temperature

Page 29: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 33

Rule:

Higher the °C

Higher the max. RH%

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm)

~ 53% RH

~ 72% RH

At every pointRS = 100% RS

Maximum %RH is Dependent on Temperature

Page 30: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 34

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm) ~ 40%

RH

~ 53% RH

At every pointRS = 100% RS

Maximum %RH is Dependent on ppm Concentration

Page 31: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 35

At every pointRS = 100% RS

Temperature (°C)

H2O

2 co

ncen

tratio

n (p

pm) ~ 40%

RHRule:

Higher the ppm

Lower the max. RH%

Bigger the difference RH% vs. RS%

~ 53% RH

Maximum %RH is Dependent on ppm Concentration

Page 32: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala 36Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28

H2O2 ppm as a Function of RS/RH Sensor Readings at 20.0˚C

Page 33: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala 37Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28

H2O2 ppm as a Function of RS/RH Sensor Readings at 20.0˚C

Page 34: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala 38Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28

~ 53% RH

H2O2 ppm as a Function of RS/RH Sensor Readings at 20.0˚C

Page 35: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala 39

%RS ~ 100%

Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28

~ 53% RH

H2O2 ppm as a Function of RS/RH Sensor Readings at 20.0˚C

Page 36: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

H2O2 ppm as a Function of RS/RH Sensor Readings at 5˚C

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 40

RS%

H2O2 ppm

RH%

Page 37: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 41

RS%

RH%

H2O2 ppm

Rule:

At 5°C

Max. ~ 550 ppm H2O2

H2O2 ppm as a Function of RS/RH Sensor Readings at 5˚C

Page 38: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 42

RS%

RH%

H2O2 ppm

H2O2 ppm as a Function of RS/RH Sensor Readings at 50˚C

Page 39: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 43

RS%

RH%

H2O2 ppm

Rule:

At 50°C

Max. ~12500 ppm H2O2

H2O2 ppm as a Function of RS/RH Sensor Readings at 50˚C

Page 40: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Five Most Typical Pitfalls in Humidity Measurement During Bio-Decontamination

Page 41: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Five Most Typical Pitfalls in Humidity Measurement During Bio-Decontamination

1. Understand how temperature affects RS%

2. Find out the correct measurement point

3. Control condensation with RS%

4. Select correct materials

5. Carefully plan onsite calibration

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 45

Page 42: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Pitfall 1: Understand How Temperature Affects RS%

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 46

Page 43: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Location: Inside a chamber

– Based on BI/EI/CI testing– Worst case location

In inlet In outlet

Method: In situ Pump & tubing & sample cell

Airflows: Does a sensor stand for airflows?

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 47

Pitfall 2: Find out the Correct Measurement Point

Page 44: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Pitfall 3: Control Condensation with RS%

RH not enough for condensation control

RH value varies with temperature and ppmH2O2

RS shows 100 %RS when the air mixture starts to condense

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 48

At every pointRS = 100% RS

Page 45: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Pitfall 3: Control Condensation with RS%

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 49

Water vapor only

RH readings of HPP272 and other brands’ sensors with catalytic filters

RH readings of other brands’ sensors without catalytic filters

HPP272 RS readings

NO CONDENSERelative saturation ~ 60 %RS

Time 4xHPP272 and several other brand’s sensors tested

A

B

C

Hum

idity

, RH

% o

r RS%

Page 46: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Material effects: Absorbtion Decomposition Desorbtion or

out-gassing

Good results with: PTFE LCP 316L (stainless

steel) Pure aluminium

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 50

PEEK+ABS Absorbtion +

decomposition(slow response time)

PEEK+ABS Desorbtion (slow response time)

PTFE + LCP Fast response time -

No decomposition

0 ppm of H2O2

160 ppm of H2O2

Aeration

Pitfall 4: Select Correct Materials

Page 47: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Pitfall 4: Select Correct Materials

Long out-gassing times with wrongly selected materials.

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 51

MaterialDesorption

maximum [ppm]PTFE 1.3PFA 1.3LCP 3.0PPS 28.5PC 57.2

PBT 67.2ABS 74.2IXEF 82.3

0102030405060708090

0 5 10 15 20 25

H2O

2 co

ncen

trat

ion

[ppm

]

Time elapsed [h]

Material desorption of various plastics

PTFEPFALCPPPSPCPBTABSIXEF

Page 48: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Challenge with H2O2 vapor Difficult to achieve stable conditions

Solution with PEROXCAP sensors Based on humidity sensors On-site calibration is easy to do with any humidity

chamber

For best measurement performance Traceable H2O2 factory calibration available at Vaisala

Service center With both humidity and H2O2 vapor

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 52

Pitfall 5: Carefully Plan On-site Calibration

Page 49: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Basics of bio-decontamination with vaporized H2O2

The importance of continuous measurement during bio-decontamination

Patented PEROXCAP® technology

The difference between Relative Humidity (RH%) and Relative Saturation (RS%)

Five most typical pitfalls in humidity measurement during bio-decontamintion

Q&A session

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 53

Conclusion

Page 50: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Q&A Session

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 55

Page 51: Humidity Measurement in H2O2 Bio-Decontamination...Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 2018-11-28 2 Piritta Maunu Life Science Regulatory

© Vaisala

Thank You for Attending!

PEROXCAP ® calibration technical note

HPP270 Series data sheet

Learn more about Bio-decontaminationBlogsVideosApplication Notes

2018-11-28Humidity Measurement in H2O2 Bio-Decontamination - Relative Saturation as the Key 56