how to calibrate an ir thermometer, frank liebmann, 2013 05-07
TRANSCRIPT
![Page 1: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/1.jpg)
© 2013 Fluke Corporation.
Fluke Calibration Web Seminar Series
Real-world expertise and practical tips about electrical, flow, pressure, RF and temperature calibration
How to Calibrate an
Infrared Thermometer
![Page 2: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/2.jpg)
Today’s Web Seminar
How to Calibrate an IR Thermometer
May 7, 2013
How do you know that your infrared calibrations are accurate? Register for this free
webinar from Fluke Calibration to learn how to improve the accuracy of your infrared
thermometer calibrations. Topics discussed will include sources of uncertainty,
calibration equipment required, overview of the calibration procedure, review of a
sample uncertainty budget, and recommendations for reporting results.
2
![Page 3: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/3.jpg)
Your Presenter
Frank Liebmann
• B.S. Electrical Engineering University of Utah
• 9 years with Fluke/Hart Scientific
– Sr. Design Engineer
– Radiation Thermometry Engineer
• Chairman ASTM Subcommittee E20.02 on Radiation Thermometry
• 20 years U.S. Army
• Phone: 801.763.1700
3
![Page 4: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/4.jpg)
What we will be talking about…
• Sources of uncertainty
• Calibration equipment
• Traceability schemes
• Laboratory setup
• Calibration procedure
• Uncertainty analysis
• Reporting your results
• Questions
4
![Page 5: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/5.jpg)
Introduction
• IR thermometry calibration can be accurate
• Proper procedure must be followed
• ASTM E2847, “Standard Practice for Calibration and
Accuracy Verification of Wideband Infrared
Thermometers”
5
![Page 6: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/6.jpg)
Sources of Uncertainty
• Major
– Emissivity estimation of the calibration source
– Field-of-view of the infrared thermometer
– Temperature gradients on the radiation source
– Improper alignment of the infrared thermometer
– Calibration temperature of the radiation source
– Ambient temperature
– Reflected temperature
• Minor
– Source heat exchange
– Atmospheric absorption
– Noise
– Display Resolution
6
![Page 7: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/7.jpg)
Calibration Equipment
• Mandatory
– Thermal radiation source
– Transfer Standard
– Ambient Temperature Thermometer
– Mounting Device
– Distance Measuring Device
• Non-mandatory
7
![Page 8: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/8.jpg)
Thermal Radiation Source
• Calibrated source
– Flat Plate
– Cavity
• Provides thermal radiation
8
![Page 9: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/9.jpg)
Thermal Radiation Source
Size of the thermal radiation source
• For Fluke IR thermometer models
– 5” (~ 125 mm) diameter is enough
• Other IR thermometer manufacturers
– Consult manufacturer
– Testing
9
![Page 10: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/10.jpg)
Transfer Standard
• Contact Thermometry (Scheme I)
– PRT, thermistor, or thermocouple
• Non-contact Thermometry (Scheme II)
– Radiation thermometer
• More discussion under ‘Traceability Schemes’
• Implementation
– Internal
– Third-party laboratory
10
![Page 11: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/11.jpg)
Ambient Temperature Thermometer
• Monitor laboratory conditions
• Use to control uncertainties
– Reflected ambient radiation
– Ambient temperature
11
![Page 12: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/12.jpg)
Mounting Device
• Holds IR thermometer
• Maintains alignment and measuring distance
• May be any of the following
– A tripod
– A fixture
– A hand
12
![Page 13: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/13.jpg)
Distance Measuring Device
• Used to verify measuring distance
• Measuring distance: “Distance or distance range between
the radiation thermometer and the target (measured
object) for which the radiation thermometer is designed.”
IEC 62492-1
• Use:
– Tape measure
– Measuring rod
13
![Page 14: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/14.jpg)
Non-mandatory Equipment
Items that may be used on special calibrations
• Aperture
• Frost / dew prevention
14
![Page 15: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/15.jpg)
Traceability Schemes
• Scheme I
– Contact traceability
• Scheme II
– Radiometric traceability
15
![Page 16: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/16.jpg)
Traceability Schemes
4180 and 4181 scheme
16
SCHEME I
SCHEME II
SCHEME II
![Page 17: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/17.jpg)
Laboratory Setup
• Laboratory temperature limits
– Ambient temperature uncertainty
– Reflected temperature uncertainty
17
![Page 18: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/18.jpg)
Laboratory Setup - Incorrect
• A heat source facing a
flat-plate
• A flat-plate facing an
exterior wall
– Especially a window
• A flat-plate in the
vicinity of air drafts
– HVAC vent
– Commonly used traffic
way
18
![Page 19: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/19.jpg)
Laboratory Setup - Incorrect
• A heat source facing a
flat-plate
• A flat-plate facing an
exterior wall
– Especially a window
• A flat-plate in the
vicinity of air drafts
– HVAC vent
– Commonly used traffic
way
19
![Page 20: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/20.jpg)
Laboratory Setup - Incorrect
• A heat source facing a
flat-plate
• A flat-plate facing an
exterior wall
– Especially a window
• A flat-plate in the
vicinity of air drafts
– HVAC vent
– Commonly used traffic
way
20
![Page 21: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/21.jpg)
Laboratory Setup - Correct
• Well controlled reflected temperature
• Sufficient spacing from other instruments
• Isolated from air drafts in room
21
![Page 22: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/22.jpg)
Controlling Reflected Temperature
• Especially a concern for lower temperature calibrations
22
![Page 23: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/23.jpg)
Calibration Procedure
• Preparation
• Calibration Points
• Procedure
23
![Page 24: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/24.jpg)
Preparation
• Allow infrared thermometer time to reach room
temperature (at least 15 minutes)
• Lens cleaning
– Only clean if requested by customer
– Clean per manufacture's instructions
• Set up any special equipment
• Set radiation source to desired temperature
• Allow radiation source time to stabilize
24
![Page 25: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/25.jpg)
Calibration Points
What calibrations point to use?
• Determined by the customer
• Laboratory may offer advice
– Over wide range, minimum 3 points
– Narrow range, 1 or 2 points may be fine
• Best practice to perform calibration from lowest
cal point to highest
25
![Page 26: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/26.jpg)
Calibration Procedure
Procedure for each calibration point
• Set IR thermometer reflected temperature (if available)
• εIRT = εS
– εS is adjustable on a 4180 and 4181
• Align IR thermometer
– Set distance (Z-direction)
– Align side-to-side and up-and-down (X- and Y-directions)
– IR thermometer should be no more than 5° from normal to target
surface
• Make measurement
– Make measurement 10 times IR thermometer response time
– For instance: IR thermometer respose time 0.5 s, make measurement for 5 s
• Procedure < 15 s
26
![Page 27: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/27.jpg)
Procedure – Measuring Distance
27
![Page 28: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/28.jpg)
Procedure - Alignment
• Laser Alignment
• Maximizing Alignment
28
![Page 29: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/29.jpg)
Uncertainty Analysis
29
Uncertainty Desig. Type U(100 °C)
(°C)
So
urc
e
Calibration temperature U1 B 0.268
Source emissivity U2 B 0.128
Reflected ambient radiation U3 A 0.031
Source heat exchange U4 B 0.012
Ambient Conditions U5 B 0.001
Source uniformity U6 A 0.163
Infr
are
d
Th
erm
om
ete
r
Size-of-source effect U7 B 0.019
Ambient temperature U8 A 0.050
Atmospheric absorption U9 B 0.020
Noise U10 A 0.100
Display Resolution U11 A 0.058
Combined Expanded Uncertainty (k=2) 0.364
![Page 30: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/30.jpg)
Reporting Your Results
Every report of calibration should contain:
• Title
• Unique identification of the calibrated infrared thermometer
• Record of the person who performed the calibration
• Date of calibration
• Source temperature versus infrared thermometer readout
temperature
• Measuring distance
• Emissivity setting of the infrared thermometer
• Diameter of the source
• Ambient temperature
• Description of the aperture including aperture distance (if used)
• Measurement uncertainties
30
![Page 31: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/31.jpg)
Supplemental Information
• Description of the calibration procedure
• List of reference instruments
• Traceability statement
• Description of the uncertainty budget
31
![Page 32: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/32.jpg)
Summary
• Follow proper procedure
• Better calibration
• Happy customers
32
![Page 33: How to calibrate an ir thermometer, frank liebmann, 2013 05-07](https://reader031.vdocuments.us/reader031/viewer/2022012405/55abdba11a28abf62f8b45c8/html5/thumbnails/33.jpg)
© 2013 Fluke Corporation
Thank you.
For information about other web seminars in this series,
including previously recorded web seminars, visit:
www.flukecal.com/webseminars
Fluke also offers in-depth training courses in calibration
and metrology. For class descriptions, schedules, and
registration, visit:
www.flukecal.com/training
Be the first to know. Sign up for Fluke Calibration
e-news bulletins, and the quarterly Total Solutions in
Calibration newsletter:
www.flukecal.com/signmeup