high speed, high common mode immunity communication interface team may12-05 chendong yang mengfei xu...
TRANSCRIPT
![Page 1: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/1.jpg)
High Speed, high common mode immunity communication interface
Team May12-05
Chendong Yang
Mengfei Xu
Advisor: Nathan Neihart
Client: RBC Medical Development
![Page 2: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/2.jpg)
Statement of Purpose
A high speed serial communication interface that would be capable of a high common mode immunity and high speed transmission.
![Page 3: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/3.jpg)
Conceptual Sketch/ Block Diagram
Transmitter Comparator
V_pp = 300mVf = 40 MHZ
USB CableSerialLVCOMS
CLK CLK
ADC
![Page 4: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/4.jpg)
Functional Requirements
1. System is capable from 20MHz to1.12 GHz
2. The clock and associated circuitry will be electrically isolated from the rest of the system.
3.LVCMOS come in and same signal come out at the end of the system
4. at least 10 Volts Common mode voltage immunity
5. The system would also need to pass radiated emissions testing, tested at CISPR-11 Group 1 Class B.
6. The transmission distance should be at least 2 meter long.
![Page 5: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/5.jpg)
Non-Functional Requirements
1. Stability
2. Budget restriction
3. Easy configuration .
![Page 6: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/6.jpg)
Operating Environment and Constraints
The most important environment consideration is noise.
If the noise is bigger than the immunity voltage range, the output will be infected by the noise.
![Page 7: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/7.jpg)
Estimate Cost
Item Approx. Cost
Transmitter Board $200-$250
Receiver Board $200-$250
ADC Board $100-$200
Radiated Emission Test $200-$250
Unexpected Cost(i.e. damaged board, chip, electronic)
$200
Total Cost $1000 - $1500
![Page 8: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/8.jpg)
Schedule (This semester)
Define Project
Basic Block Diagram
Transmitter Receiver
Market Survey Market Survey
Schematic Test Schematic Test
PCB Design PCB Design
Order PCB Order PCB
Aug.
Sept.
Oct.– Nov.
Nov.-Dec.
![Page 9: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/9.jpg)
Functional Decomposition
Name Functional
Transmitter A device transform a signal line voltage input to double line Low voltage differential signal output
Receiver Comparator which compares two voltages or currents and switches its output to indicate which is larger.
USB cable transmit the data from transmitter board to receiver board
![Page 10: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/10.jpg)
Market Survey
Requirement:
1. High input toggle frequency, initially at least 40MHZ
2. Small falling/rising time, i.e. t<5ns
3. Affordable Price
Final Choose: DS90C031 from NationalSemiconductor
1. Rising/Falling time: 1.5 ns
2. Power supply : 5v
3. input switching frequency support: excess 77 MHZ
4. ~$3
5. ±350 mV differential signal
![Page 11: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/11.jpg)
Test Constraints
To keep their business secretes, companies don’t provide Pspice models of the transmitters we need. We have tried many ways to solve this problem such like using some other models, translating ibis model to Pspice model, and testing the sample transmitter to see if it is capable for our transmitter part. we are still not able to simulate the transmitters. In that case, we only test the receiver part with input signal generated by the software.
![Page 12: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/12.jpg)
Sample Test
![Page 13: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/13.jpg)
Test result output
when DS99R103 have finished transmitting the LVCOMS to LVDS, there are other 23bits signal in the LVDS
InputA square wave with Vpp=3v,frequency=1.5 Mhz,Vdd=5v, clock is as twice fast as the data
![Page 14: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/14.jpg)
Market Survey Requirement:
1. Capability of a high common mode immunity, at least 10V, i.e. power supply range should be greater than 10V
2. High input toggle frequency, initially at least 40MHZ
3. Small falling/rising time, i.e. t<5ns
4. Affordable Price Final Choose: LT 1711 from Linear Technology1. High common mode rejection : 65-75dB
2. Rising/Falling time: 2ns
3. Power supply range : 14 V
4. Maximum input toggle frequency : 100 MHZ
5. ~$5
![Page 15: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/15.jpg)
Schematic Test (Ideal)
• Test Interface: Pspice
• Test Condition:1. Square differential input Vpp = 300 mV, PW = 12.5n, f = 40MHZ2. Common mode input voltage = 10V3. V+ = 11V, V- = 0V4. No noise consideration
![Page 16: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/16.jpg)
Schematic
![Page 17: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/17.jpg)
Result 1st Graph: Vout+2nd Graph: Green: Vin+, Red: Vin-
![Page 18: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/18.jpg)
Add Noise
![Page 19: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/19.jpg)
Schematic Test (Improvement)• Test Condition:1. Square differential input Vpp = 300 mV, PW = 12.5n, f = 40MHZ2. Common mode input voltage = 10V3. V+ = 11V, V- = 0V4. Noise consideration: sine noise5. Add Hysteresis6. Line impedance = 100ohm
![Page 20: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/20.jpg)
Result
![Page 21: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/21.jpg)
USB Cable Test
Input signal:
• Square wave• f = 40 MHZ• Vpp = 300mV• High load impedence
Output
Conclusion:Normal USB cable can transmit 40MHZ, small differential voltage with tolerable distortion.
![Page 22: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/22.jpg)
PCB Design(In progress)
• Test Interface: Eagle PCB design software
• PCB designed size: 80mm * 70mm
![Page 23: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/23.jpg)
PCB Layout (Current Version)
![Page 24: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/24.jpg)
Current Project StatusDefine Project
Basic Block Diagram
Transmitter Receiver
Market Survey Market Survey
Schematic Test Schematic Test
PCB Design PCB Design
Order PCB Order PCB
Issue Left
1. Optimize the PCB layout
2. Order the electronic element need for our PCB.
![Page 25: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/25.jpg)
Individual Responsibility & ContributionMengfei Xu: • Team Leader• Transmitter Designer• Website design.
Chendong Yang• Communication Liaison • Receiver Designer• Document & File Arrangement
![Page 26: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/26.jpg)
Schedule (Next semester)
Transmitter PCB Setup Receiver PCB Setup
Verification Test Verification Test
Whole System Verification Test
Improvement
Radiated emission Test
Verification & Delivery
Jan.- Feb
Feb.
March.
April
Aug.- May.
![Page 27: High Speed, high common mode immunity communication interface Team May12-05 Chendong Yang Mengfei Xu Advisor: Nathan Neihart Client: RBC Medical Development](https://reader030.vdocuments.us/reader030/viewer/2022032600/56649d965503460f94a7f24d/html5/thumbnails/27.jpg)
Questions?