High-pressure calibration- unique calibration facilities at FORCE Technology

Unique high-pressure calibration system

FORCE Technology provides high-pressure calibration of gas flow meters with natural gas as medium at our calibration facilities located in Vejen, Denmark. The high-pressure calibration facilities are built as closed loops with low pressure loss and operate at maximum flow and maximum pressure all the year round.

Our systemFORCE Technology has recently expanded the calibration facilities with the construction of a new system, which is internation-ally unique. Our high-pressure calibration facilities consist of:

• 12” (300 mm) high-pressure loop for calibration of gas meters• 4” primary twin Piston prover system • 24” (600 mm) high-pressure loop for calibration of gas meters

Technical information regarding the high-pressure calibration facilityies at FORCE Technology

Specification 12” facilities 24” facilities

Calibration pressure 0 - 50 bar / 0 - 725 PSI 0 - 65 bar / 0 - 945 PSI

Flow 8 - 6.500 m3/h (8 - 10.000 m3/h) /0.0678 - 5.51 Mmacfd (0.0678 - 8.48 Mmacfd)

10 - 32.000 m3/h (10 - 41.000 m3/h /0.0848 - 27,12 Mmacfd (0.0848 - 34.75 Mmacfd)

Power 310 KW 2,000 kW

Meter sizes 2” - 18” (50 mm - 450 mm) 4”- 30” (50”) (100 mm - 1.250 mm)

Maximum N flow 0.4 mill. Normal m3/h / 330 Mmscfd 2.5 mill. Normal m3/h / 2059 Mmscfd

Our customers and partners are offered to rent our high-pres-sure calibration facilities and trained staff for processes of de-velopment. We can perform calibration and verification of all types of gas and flow meters, including:

• Turbine meters• Swirl meters• Ultrasonic measuring instruments• Ultrasound meters• V-cone meters • Vortex meters• Coriolis meters• Thermal mass flow meters• Mass flow controllers

24” (600 mm) high-pressure loop for calibration of gas meters.

4” primary twin Piston prover system. 12” (300 mm) high-pressure loop for calibration of gas meters.

Industry turnover 2014

About usFORCE Technology is one of the leading technology, con-sulting and service companies on the international market.

We:• Are independent• Are unbiased• Have a development budget > MEUR 30

We transform highly specialised engineering knowledge into practical and value-creating solutions for a broad spec-trum of sectors and industries.

Business ConceptOur business revolves around the following areas:

• System & management development • Integrity management & maintenance optimization • Calibration, inspection, testing, verification & certifica-

tion • Design, product, process & production optimization • Training & mediation

within: • Process & Automation• Materials technology• Testing technology• Sensors & Metrology• Hydro- & Aerodynamics• Simulation technology• Energy & Environment

DenmarkBrøndby (HQ)

12 offices

Norway Hvalstad9 offices

Sweden Vësterås

23 offices

Singapore

ChinaBeijing

Our customersFORCE Technology’s business concept is internationally acknowledged, and industries and companies worldwide benefit from our services.

Amongst many industries, we service:• Oil & Gas industry• Energy & Environment industry• Trade & Service industry• Transport industry

UAEAbu Dhabi

About FORCE Technology

FORCE Technology offices 2016

010203040506070

0 10000 20000 30000 40000 50000

High Flow measurement at FORCE Technology.Pressure in bar; flow in m3/h.

12" High pressure;natural gas (2004)

24" High pressure;natural gas (2014)

air (1985-1992)0

20

40

60

80

100

120

0 100 200 300 400 500

Primary flow measurement at FORCE Technology.Pressure in bar; flow in m3/h.

Primary system; natural gas(2013)

Primary system, air (2012)

0

200

400

600

800

1000

0 10 20 30 40

High Flow measurement at FORCE Technology.Pressure in PSI; flow in MMacfd.

12" High pressure;natural gas (2004)

24" High pressure;natural gas (2014)

air (1985-1992)0

500

1000

1500

2000

0 0,1 0,2 0,3 0,4

Primary flow measurement at FORCE Technology.Pressure in PSI; flow in MMacfd.

Primary system; naturalgas (2013)

Primary system, air (2012)

FORCE TechnologyHeadquartersPark Allé 345

DK-2605 BrøndbyTlf. +45 43 25 00 00Fax +45 43 25 00 10

[email protected] forcetechnology.com

4609-5-da-en: Subject to change without notice

Historic Gas Flow Measurement at FORCE Technology

1985 Facilities for calibration of gas meters, atmospheric air - 4.000 m3/h / 3.390 Mmacfd.

1992 Facilities for calibration of gas meters, 0 - 8 bar air - 1.000 m3/h / 120 PSI air - 0.848 Mmacfd.

2003 Facilities for calibration of gas meters, 0 - 50 bar natural gas - 6.500 m3/h / 0 - 725 PSI natural gas - 5.51 Mmacfd.

2004 Primary system for generating of the unit m3/h. atmospheric air - 10 m3/h / 0.00848 Mmacfd.

2006 Facilities for calibration of gas meters, 0 - 35 bar natural gas - 10.000 m3/h / 0-510 PSI natural gas - 8.48 Mmacfd.

2012 Primary system for  generating of the unit m3/h. atmospheric air - 100 m3/h / 0.0848 Mmacfd.

2013 Primary system for high-pressure calibration, 0-100 bar natural gas - 400 m3/h / 0 - 1.450 PSI natural gas -

0.339 Mmacfd.

2014 Facilities for calibration of gas meters, 0 - 65 bar natural gas - 32.000 m3/h. (41.000 m3/h at lower pressure) /

0 - 945 PSI natural gas - 27.12 Mmacfd ( 34.745 Mmacfd at lower pressure).

FORCE TechnologyGas Laboratory

Navervej 1DK-6600 Vejen

Tlf. +45 76 96 16 00Fax +45 43 26 70 11

[email protected] forcetechnology.com

Further informationThomas Boy Jacobsen, Operations Manager Tel. +45 42 62 76 47 / E-mail: [email protected] Busk, Head of department, Tel. +45 22 69 76 20 / E-mail [email protected]