Baxter HemoconcentratorAnne Wespetal, Cody Felts, Marc Gruber
![Page 2: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/2.jpg)
Video
http://www.medivators.com/products-and-services/therapeutic-technologies/hemoconcentrators
![Page 3: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/3.jpg)
Device Description
The device is used as part of a heart and lung machine, which oxygenates patient's blood during surgery. ● Device measurements: 12.25”x3.125”● Device mass: 187.83g● Material: The product is composed of a clear
thermoset plastic exterior, with red and blue end caps, in which polysulfone filter fibers are housed.
![Page 4: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/4.jpg)
Gathering RequirementsUsers● Installed by Surgical teams(Surgical
Assistants and Nurses)● Users are professionals and highly skilled
in their particular field
Environment● Used in a Clinical setting ● Operating room/surgical settings● Product is used in Surgeries
![Page 5: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/5.jpg)
Gathering RequirementsUser Needs● Package needs to contain product● Product Shall be easy to remove
from packaging● Product shall maintain Sterility● Product Shall be properly Labeled● Easy to Store● Shall have a Shelf life of 2 Years
![Page 6: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/6.jpg)
Product Requirements● Product Shall withstand 48” Drop contained
with 4lbs user manual● Product is Light sensitive● Product shall have Graphic of Device on
Label● Product Shall be stored in an Omni Cell
![Page 7: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/7.jpg)
Product RequirementsMost Difficult Requirement● 48-Inch Drop Test with 4lb user manual● End ports are essential for proper function● The most fragile part of product
![Page 8: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/8.jpg)
Establishing SpecificationsThe product must be sterile upon use
The product must be intact and functional upon use
The barrier shall not be compromised within the distribution system or throughout the specified shelf life
The product shall be able to meet its intended function and be free from any apparent damage
![Page 9: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/9.jpg)
Establishing SpecificationsThe product must be protected from light.
The package must be easy to open
The packages shall protect the package from over 86 Lux.
Consumer shall be able to open in less than 10 seconds. Seal strength shall be between 1.0 lbs/in and 2.5 lbs./in
![Page 10: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/10.jpg)
Establishing SpecificationsThe package shall be optimally sized to fit in an Omni Cell
The product must be able to withstand severe handling
Package shall be stored hanging in a vertical position in the Omni Cell and have a designated area for a bar or QR code
The primary package shall protect the product from a 48” drop
![Page 11: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/11.jpg)
Establishing SpecificationsThe product will shipped in common order sizes
The package should show a graphic of the product
An instructional booklet will be included
The product will be offered in case of 4 and 12 units
The package shall feature a 4x2.5 inch graphic of the device
A 4 lb instructional booklet will be included in each case
![Page 12: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/12.jpg)
Specification Challenges● Determined the light sensitivity of the product. This was
compared with data that was gathered from light meters which packaged in the proposed packaging and ship tested.
● Attempted drop testing at variable temperatures with PC and PP clamshells.
● Opening time specification averaged 11.2 seconds for 20 test participants.
![Page 13: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/13.jpg)
Sterilization ● Gamma radiation will be used due to the closed
filter cavity of the device. ● Device will not discolor during this process● Gamma radiation allows deep penetration for
dense packages. ● 25-40 kGy
![Page 14: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/14.jpg)
DesignLock-in Plugs (2)
Live Hinge
![Page 15: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/15.jpg)
![Page 16: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/16.jpg)
![Page 17: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/17.jpg)
Materials Polycarbonate (PC) Advantages:● Excellent in Radiation● High impact resistance● Good dimensional stability● PC can undergo large plastic deformations without cracking or
breaking.Disadvantage: ● Subject to stress cracking● Expensive
![Page 18: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/18.jpg)
Materials
● PC clamshell● HDPE header bag (white)● B/C double wall RSC● Packaging Tape - Benchmark
![Page 19: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/19.jpg)
Design Justification● Ensures necessary product protection from 48
inch drop● Compatible with sterilization method● Protects product adequately from light without
over packaging ● May be more difficult and time consuming to
open
![Page 20: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/20.jpg)
Alternative Option 1● Two piece thermoform tray sealed with
Tyvek and placed in a paperboard carton● More standardized design which is more
intuitive for users● Standardized bottom for different products
in the same family● Waited space for some products in the
family● Snap fit may be difficult for manufacturing
control
![Page 21: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/21.jpg)
Alternative Option 2● Device in a poly bag, which is placed in a set
of corrugated end caps on a paperboard carton
● Low cost materials● Adequate drop protection ● Difficult to assemble● Lacks professional clean appearance
![Page 22: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/22.jpg)
Shelf Life● 2-year shelf life ● Common Group Tests: Seal Strength, Drop
Test, and Dye Penetration ● The Polysulfone membranes will expire
before our package design
![Page 23: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/23.jpg)
Equipment & ManufacturingRSC Double WallMastercase(4/case)Thermoformed PC TrayHDPE Header Bag
![Page 24: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/24.jpg)
Equipment & ManufacturingThermoformed Tray● PC durability Temp -100°C to
+130°C● Targeted Processing temperature is
191°C● Upper limit 210°C and Lower limit
180°C● The most difficult part of is keeping
the form consistent on the sharp edges
![Page 25: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/25.jpg)
Equipment & ManufacturingThermoforming Machinery
● Brown SR-Series ● Easy Changeover● Clam Shell oven with various
heater and zone configurations provide optimum material heat control
● PCL heat and ● Fast and Precise
![Page 26: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/26.jpg)
Equipment & ManufacturingHDPE Header bag
● Doboy CBS-DM(Bosch)● High Quality Sealer● Built specifically for Medical Packaging● Seal widths from 1/4" to 3/4" wide● Heating and cooling bars to ensure a
secure seal● 275 °F ● Seal temp of 275 °F Targeted Value +-
15°F
![Page 27: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/27.jpg)
Equipment & ManufacturingQuality Assurance● Equipment operator
inspections● Ensures quality product● Check critical aspects of
product(Seals and Edges)
![Page 28: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/28.jpg)
Design VerificationMethods of evaluation were driven by technical specificationsASTM F88 Drop Testing (48 inches) ASTM F1980 Accelerated AgingASTM D4169 Medical Level I Distribution TestingASTM Bubble EmissionsPracticed timed opening assessment Placement of products in Omni Cell and scan verificationMeasurement of Graphic
![Page 29: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/29.jpg)
Design VerificationTest Failure: PC clamshell had areas of stress in the ribs due to sharp cornersAdjustment: The ribs are now thicker and rounded offTest Failure: 10 second opening requirement was not meetAdjustment: Specification was changed to under 13 seconds
![Page 30: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/30.jpg)
Design ValidationDesign change to meet customer’s needs: Cut (0.15) inches off of the top shell to make opening the Clamshell easier
![Page 31: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/31.jpg)
Design ValidationThe validation includes approval of 18 out of 20 participants in the areas of: ● Ease of opening● Size (for omni cell)● Case counts● Graphic size and location● Intact and functional ● Sterility by visual inspection
❖ All customer needs were met
![Page 32: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/32.jpg)
Labeling
● Needed a product graphic● Would likely have to have
some content translated in multiple languages
● In house thermal label printing - variable information
![Page 33: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/33.jpg)
Post-commercialization● Complaint monitoring in is processes and data
is being complied ● Non-conforming occurrences will be classified
and processed according to protocol ● CAPA’s will be implemented if needed● Needed changes will be deemed as equivalent
if possible
![Page 34: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/34.jpg)
Question?
![Page 35: Hemoconcentrator](https://reader035.vdocuments.us/reader035/viewer/2022062420/55d39d4bbb61ebfd268b4733/html5/thumbnails/35.jpg)
Resources http://www.medivators.com/products-and-services/therapeutic-technologies/hemoconcentrators
https://www.lairdplastics.com/product/materials/petg/888-petg-polyethylene-terephthalate-glycol-modified
http://www2.ulprospector.com/
http://www.sterlitech.com/filters/membrane-disc-filters/polyethersulfone-pes-membrane-filters.html
ASTM and ISO Standards