heartmate ii left ventricular assist device (lvad)
DESCRIPTION
HeartMate II Left Ventricular Assist Device (LVAD). Courtesy University of Washington Medical Center. BACKGROUND. Ventricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patients - PowerPoint PPT PresentationTRANSCRIPT
HeartMate II Left Ventricular Assist Device (LVAD) Pivotal Trial
HeartMate II Left Ventricular Assist Device (LVAD)Courtesy University of Washington Medical Center
0BACKGROUNDVentricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patientsThe dramatic increase in the use of VADs has been unavoidable for those patients suffering end-stage heart failure due to the consistent shortage of donor organsAs mean support duration for VADs increases, more VAD patients will be living in the communityHeartMate LVAD is approved as Destination Therapy (DT)
1HM II CANDIDATESFDA Approved 4.21.08Bridge-to-Transplant (BTT)
FDA Approved 1.20.10Destination Therapy (DT)
2
Cored into LV
Outflow to aorta
Percutaneous tube
System Controller
Batteries HM II
3
Anatomic Placement
4FUNCTION & CAPABILITYAxial (continuous) flow: ?NO PULSE/ ?BPOne moving part: RotorSmall400 gm125 cc60% smaller than HM I (XVE)QuietFlows: 3 10 lpmAnticoagulation required
5ADVANTAGESSize: Potential BSA of .8 m2
Only one moving part
Blood lubricated bearings
Designed for long term support
Lower infection rates (smaller perc tube)
6
Pump HousingRotor
7PUMP ROTOR and STATORSInflow StatorInflow BearingsRotorOutflow BearinsOutflw StatorBLOOD FLOW
8
9Controller + Back-up ControllerSYSTEM CONTROLLER
10FACE OF SYSTEM CONTROLLER
11 POWER MODULESupplies main power to LVAD
Provides 30 minutes of backup powerRepeats alarms generated by the System Controller
Serves as the electrical interface between the System Controller and the Display Module
12BATTERY CHARGER
BATTERIES 14v Li-Ion 6 10 hours of support *Patients report up to 15 hrs of battery power Up to 4 hour recharge for fully discharged battery 3 years or 360 charges
14DISPLAY MODULE
Pump ModeFixedPower Saver
Pump Speed (rpm)
Pulse Index
Estimated Flow (lpm)Too low ---Too high +++
Power (watts)
Alarm Conditions Highest priority displayed
15AUTO MODE: only with Thoratec engineerEMERGENCY POWER PACK (EPP)Single use battery pack in a plastic carrying case with a shoulder strap
Provides battery power in the event of extended power outage
Approximately 12 hours of support
Must be replaced if used for a period exceeding three hours
16BATTERY ALARMSADVISORY ALARMSHAZARD ALARM
17BATTERY ALARMS BATTERY < 15 minutes of power BEEP Q SEC BATTERY < 5 minutes of power STEADY TONEDefaults to Power Saver ModePump defaults to Fixed Rate Mode of 8000 rpm, or fixed speed setpoint if lowerSystem will return to set speed once adequate power is restored
ACTION Replace batteries or switch to alternate power source
18
SYSTEM DRIVER CELL LOW VOLTAGESYSTEM CONTROLLER CELL LOW VOLTAGE Yellow cell symbol Beep every 4 seconds
ACTION Replace cell battery and perform System Controller self test
19POWER CABLE DISCONNECTED
and FLASHINGPOWER CABLE DISCONNECTEDFlashing green power symbol & battery power barsBeep every second
ACTIONCheck cable connections to power sourceCheck power leads for damage, replace if necessary
20 LOW FLOW; NO OPERATION or INCORRECT OPERATIONLOW FLOW < 2.5 lpmPump not operating or not operating correctlyDecreased preload (right heart failure, tamponade, hypovolemia, bleeding, etc)Obstruction of pump inflow or outflowSystemic hypertensionACTIONAssess patientMonitor
21 DRIVELINE DISCONNECTED FROM CONTROLLER Check connections
ACTIONReconnect driveline to controller
STEADY TONE andNO SYMBOLNO POWER TO PUMP
ACTIONCheck system driver connections to pumpCheck system driver power connections to power sourceIf persist, seek additional help immediately
23CPR SHOULD NOT BE PERFORMED ON HEARTMATE LVAD PATIENTSUNLESS DIRECTED
24VT or VFSTABLEPatient may feel funny light headed or differentPump speeds and flows are normal, low normal, or very lowConsider cardioversion after consultation with Mechanical Assist Device Coordinator
UNSTABLEPatient unresponsiveTreat as unstable VT/VF
25LVAD PATIENT TRANSPORTTransport to assigned facilitySpare batteries, PBU and the display module should be brought to the hospital with the patientPBU weighs 29 pounds without batteriesAll modes of emergency transportation are acceptableAviation electronics will NOT interfere with LVAD and visa versa