  • Programming manual

    H-2000-6021-0D-C

    Inspection software for lathes

    for Fanuc 0, 6, 1015, 16, 18, 21T and iT,

    and Yasnac LX3 controllers

  • 1998 2009 Renishaw plc. All rights reserved.

    This document may not be copied or reproduced in whole or in

    part, or transferred to any other media or language, by any means,

    without the prior written permission of Renishaw plc.

    The publication of material within this document does not imply

    freedom from the patent rights of Renishaw plc.

    Disclaimer

    RENISHAW HAS MADE CONSIDERABLE EFFORTS TO ENSURE

    THE CONTENT OF THIS DOCUMENT IS CORRECT AT THE

    DATE OF PUBLICATION BUT MAKES NO WARRANTIES OR

    REPRESENTATIONS REGARDING THE CONTENT. RENISHAW

    EXCLUDES LIABILITY, HOWSOEVER ARISING, FOR ANY

    INACCURACIES IN THIS DOCUMENT.

    Trademarks

    RENISHAW and the probe emblem used in the RENISHAW logo

    are registered trademarks of Renishaw plc in the UK and other

    countries.

    apply innovation is a trademark of Renishaw plc.

    All other brand names and product names used in this document

    are trade names, service marks, trademarks, or registered

    trademarks of their respective owners.

    Renishaw part no: H-2000-6021-0D-C

    Issued: 06.2009

  • IMPORTANT PLEASE READ CAREFULLY

    RENISHAW PRODUCT LICENCE

    Licensee: you, the person, firm or company accepting the terms of this Licence

    Renishaw: Renishaw plc, New Mills, Wotton-under-Edge, Gloucestershire, GL12 8JR, United Kingdom

    Product: the software, which is designed to operate on machine tool numeric controllers, supplied by Renishaw for use with Renishaws machine tool probing systems

    Licence to use: a non-exclusive licence to use the Product on a single machine tool only

    By installing and/or using the Product you indicate your acceptance of the terms of this Licence.

    Renishaw grants the Licensee a Licence to use the Product on condition the Licensee accepts the following terms and conditions:

    1. All rights in and title to the Product are and shall remain vested in Renishaw and its licensors.

    2. Renishaw shall replace or repair the Product if it does not materially perform to specification under proper use within 90 days of delivery. This warranty does not apply where the Product has been modified in any manner that is not specifically described in the Product or in the installation or programming manuals supplied with the Product, or where the Product is used with probing systems that have not been produced by Renishaw.Except as stated in this paragraph, all warranties, conditions and terms implied by law are excluded. In particular, no warranty is given that the Product is bug or error-free.

    3. NOTE - LIMITATION OF LIABILITY IN CONNECTION WITH USE OF THE PRODUCT

    Renishaw does not exclude liability for personal injury or death caused by Renishaws negligence.

    Renishaws liability is limited to (a) the warranty contained in paragraph 2 and (b) direct losses up to a maximum of 50,000.

    Renishaw has no liability to the Licensee for any indirect, consequential or economic loss (including, without limitation, loss of data, profits or goodwill).

    The Product has been designed for use with Renishaws machine tool probing systems. Renishaw hasno liability for the results of using the Product with another manufacturers machine tool probing systems.

    By accepting the terms of this Licence the Licensee agrees that this limitation of liability is reasonable.

    4. The Licensee may not make any copies of the Product except as provided in this Licence or as permitted by applicable law. The Licensee is authorised to make a backup copy of the Product for security purposes. The Licensee must not remove any licence and copyright notices, labels or marks contained in the original and shall ensure all copies contain such notices without modification.

    5. If the Product contains electronic manuals the Licensee may print out the manuals in part or in full, provided that the print outs or copies are not supplied to any third party that is not an employee or contractor for the Licensee without Renishaws written permission

    6. The Licensee shall not reverse engineer, decompile, or modify the Product or re-use any components separately from the Product unless permitted by a specific instruction contained in the Product or the programming or installation manuals supplied with the Product or by applicable law provided that in the latter case, Licensee has first contacted Renishaw to request any information required to interface with Licenseesother software.

    7. The Licensee shall not make the Product available to any third party in any manner whatsoever nor may this Licence and the Product be transferred to a third party without Renishaws prior written agreement. Any agreement by Renishaw is conditional on the permitted transferee agreeing to all terms of this Licence and the Licensee not retaining any copies of the Product. Where the Licensee is a reseller of Renishaws machine tool probing systems, Licensee may transfer the Product for ultimate use by an end user with Renishawsmachine tool probing systems.

    8. Renishaw shall have the right to terminate this Licence immediately if the Licensee fails to comply with any of these terms and conditions. The Licensee agrees upon receipt of notice of termination from Renishaw toimmediately return or destroy all copies of the Product in its possession or control.

    9. This Licence is governed by English law and the parties submit to the exclusive jurisdiction of the English courts.

    Renishaw Product Licence (EN) Issue 1: February 2007

  • Form 1

    EQUIPMENT REGISTRATION RECORD

    Please complete this form (and Form 2 overleaf if applicable) after the Renishaw equipment has been installed on your machine. Keep one copy yourself and return a copy to your local Renishaw Customer Support office (see www.renishaw.com/contact for the address and telephone number). The Renishaw Installation Engineer should normally complete these forms.

    MACHINE DETAILS

    Machine description..........................................................................................................................

    Machine type.................................................................................................................................

    Controller............................................................................................................................

    Special control options.................................................................................................................................

    .................................................................................................................................

    .................................................................................................................................

    RENISHAW HARDWARE

    Inspection probe type .......................................

    Interface type ..................................................

    Tool setting probe type ....................................

    Interface type ..................................................

    RENISHAW SOFTWARE

    Inspection disk(s).....................................................

    .....................................................................................

    ........................................................................................

    Tool setting disk(s) .....................................................

    .....................................................................................

    .........................................................................................

    SPECIAL SWITCHING M-CODES (OR OTHER) WHERE APPLICABLE

    Switch (Spin) probe on ...................................

    Switch (Spin) probe off ...................................

    Start/Error signal .............................................

    Dual systems only

    Switch on inspection probe .....................................

    Switch on tool setting ..............................................

    Other ..........................................................................

    ......................................................................................

    ADDITIONAL INFORMATION

    Tick box if Form 2 overleaf has been filled in.

    Customer's name..................................................................

    Customer's address..................................................................

    ......................................................................................................

    .........................................................................................................

    ......................................................................................................

    Customer's tel. no....................................................................

    Customer's contact name..................................................

    Date installed .....................................

    Installation engineer ............................

    Date of training........................................

  • Form 2

    SOFTWARE DEVIATION RECORD

    Standard Renishaw kit no. Software disk nos.

    Reason for deviation

    Software no. and macro no. Comments and corrections

    The software product for which these changes are authorised is subject to copyright.

    A copy of this deviation sheet will be retained by Renishaw plc.

    A copy of the software amendments must be retained by the customer they cannot be retained by Renishaw plc.

  • Cautions i

    Publication No. H-2000-6021

    Caution Software safety

    The software you have purchased is used to control the movements of a machine tool. It

    has been designed to cause the machine to operate in a specified manner under operator

    control, and has been configured for a particular combination of machine tool hardware

    and controller.

    Renishaw has no control over the exact program configuration of the controller with which

    the software is to be used, nor of the mechanical layout of the machine. Therefore, it is

    the responsibility of the person putting the software into operation to:

    ! ensure that all machine safety guards are in position and are correctly working

    before commencement of operation;

    ! ensure that any manual overrides are disabled before commencement of operation;

    ! verify that the program steps invoked by this software are compatible with the

    controller for which they are intended;

    ! ensure that any moves which the machine will be instructed to make under program

    control will not cause the machine to inflict damage upon itself or upon any person

    in the vicinity;

    ! be thoroughly familiar with the machine tool and its controller and know the location

    of all emergency stop switches.

    !

  • ii

    Publication No. H-2000-6021

    This page is intentionally left blank

  • Contents iii

    Publication No. H-2000-6021

    Contents

    Before you begin

    Before you begin................................................................................................................. 1

    Measurement values used in this manual .......................................................................... 2

    Associated publications ...................................................................................................... 2

    About the Inspection software ............................................................................................ 2

    Software kit no. A-4012-0541 ............................................................................................. 2

    File1: Probe horizontal cycles...................................................................................... 2

    File2: Probe vertical cycles.......................................................................................... 3

    File3: C-axis cycles...................................................................................................... 3

    Macro memory requirements.............................................................................................. 3

    Macro numbers and functions ............................................................................................ 4

    Measurement errors ........................................................................................................... 5

    Renishaw customer services .............................................................................................. 5

    Calling Renishaw......................................................................................................... 5

    Chapter 1 Installing the software

    Installing the software ...................................................................................................... 1-2

    Using the Wizard ...................................................................................................... 1-2

    Editing macros manually .......................................................................................... 1-2

    Setting and adjusting the software................................................................................... 1-2

    Chapter 2 Optional inputs

    Optional inputs ................................................................................................................. 2-2

    Chapter 3 Variable outputs

    Table 1 Variable outputs with probe vertical.................................................................... 3-2

    Table 2 Variable outputs with probe horizontal ............................................................... 3-2

    Chapter 4 Protected positioning cycles

    Probe protected positioning (probe vertical) O9610..................................................... 4-2

    Probe protected positioning (probe horizontal) O9810................................................. 4-4

    Chapter 5 Calibrating the probe

    Why calibrate a probe and stylus?................................................................................... 5-2

    Single-sided calibration.................................................................................................... 5-2

    Double-sided calibration .................................................................................................. 5-3

  • iv Contents

    Publication No. H-2000-6021

    Calibrating the probe length............................................................................................. 5-3

    Calibration cycles............................................................................................................. 5-3

    Calibrating the probe length (probe vertical) O9601..................................................... 5-5

    Calibrating the stylus Z tool offset double-sided (probe vertical) O9602..................... 5-7

    Calibrating the stylus ball radius double-sided (probe vertical) O9603 ....................... 5-9

    Calibrating the stylus single-sided (probe vertical) O9604 ......................................... 5-11

    Calibrating the probe length (probe horizontal) O9801 .............................................. 5-13

    Calibrating the stylus X tool offset double-sided (probe horizontal) O9802 ............... 5-15

    Calibrating the stylus ball radius double-sided (probe horizontal) O9803.................. 5-17

    Calibrating the stylus single-sided (probe horizontal) O9804..................................... 5-19

    Chapter 6 Standard measuring cycles

    X Z single surface measurement (probe vertical) O9611............................................. 6-2

    Web/pocket measurement (probe vertical) O9612....................................................... 6-5

    X Z single surface measurement (probe horizontal) O9811......................................... 6-8

    Web/pocket measurement (probe horizontal) O9812................................................. 6-11

    Chapter 7 Additional cycles

    Storing multi-stylus data (probe vertical) O9630 .......................................................... 7-2

    Loading multi-stylus data (probe vertical) O9631 ......................................................... 7-4

    Storing multi-stylus data (probe horizontal) O9830 ...................................................... 7-6

    Loading multi-stylus data (probe horizontal) O9831..................................................... 7-8

    Chapter 8 C-axis cycles

    C-axis find (probe vertical) O9650................................................................................ 8-2

    C-axis find (probe horizontal) O9850............................................................................ 8-4

    Chapter 9 Alarms and error messages

    General alarms ................................................................................................................ 9-2

    Chapter 10 Settings, macros and variables

    Customising the software .............................................................................................. 10-2

    Editing the settings macro (O9524 or O9724)........................................................ 10-2

    Setting #33.............................................................................................................. 10-3

    Variable base number setting (#111 and #112) ..................................................... 10-3

    Alarm settings (#110).............................................................................................. 10-3

    Process alarms ....................................................................................................... 10-4

    Work offset update tests ................................................................................................ 10-4

  • Contents v

    Publication No. H-2000-6021

    Use of macro variables .................................................................................................. 10-4

    Local variables........................................................................................................ 10-4

    Common variables.................................................................................................. 10-5

    Common retained variables.................................................................................... 10-6

    Chapter 11 General information

    Tolerances ..................................................................................................................... 11-2

    Experience values Ee .................................................................................................... 11-3

    Reason for using this option ................................................................................... 11-3

    Printing a macro output an example........................................................................... 11-3

    Appendix A Features, cycles and limitations

    Features of the Inspection software.................................................................................A-2

    Cycles ..............................................................................................................................A-2

    Limitations........................................................................................................................A-2

    Appendix B Co-ordinate systems used on lathes

    Example 1 Tool offset geometry .....................................................................................B-2

    Example 2 Using either G54 to G59 or G50 to define the component zero...................B-3

    G54 to G59 method ..................................................................................................B-3

    G50 method setting a program shift to the turret reference position........................B-4

    Example 3 Using G10 to define the component zero .....................................................B-5

    Example 4 Using either G92 or G50 to define the component zero-to-tool tip position.........................................................................................................................B-6

  • vi Contents

    Publication No. H-2000-6021

    This page is intentionally left blank.

  • Before you begin 1

    Publication No. H-2000-6021

    Before you begin

    This programming manual contains detailed information on using the Inspection software

    for programming, operating and controlling 2-axis lathes.

    Comprising eleven self-contained chapters and two appendices, the manual is structured

    to provide the information you require to use the Inspection software effectively.

    ! Chapter 1, Installing the software describes how to install the Inspection software

    on your machine.

    ! Chapter 2, Optional inputs describes the optional inputs that are required by many

    of the cycles.

    ! Chapter 3, Variable outputs lists the optional outputs that are produced by many of

    the cycles.

    ! Chapter 4, Protected positioning cycles describes how to use the protected

    positioning macros (O9610 and O9810). When correctly used, these macros prevent

    damage to the stylus in the event of the probe colliding with the workpiece.

    ! Chapter 5, Calibrating the probe describes how to use the macros that are

    provided for calibrating a probe. Cycles are provided for calibrating an inspection

    probe when it is vertically mounted and horizontally mounted.

    ! Chapter 6, Standard measuring cycles describes how to use the non-vector

    measuring cycles.

    ! Chapter 7, Additional cycles describes how to use the four macro cycles that are

    not described in previous chapters.

    ! Chapter 8, C-axis cycles describes how to use two C-axis measuring cycles.

    ! Chapter 9, Alarms and error messages describes the macro alarm numbers and

    messages that are displayed on the screen of the controller when an error occurs.

    An explanation of the meaning and possible cause of each alarm message is

    provided, together with typical actions you must take to correct the fault causing the

    message.

    ! Chapter 10, Settings, macros and variables describes how to customise the

    Inspection software on your machine. It also covers setting information and details

    about the macro variables used in the software.

    ! Chapter 11, General information, contains general information and reference

    material that is relevant to the Inspection software package.

  • 2 Before you begin

    Publication No. H-2000-6021

    Measurement values used in this manual

    Throughout this manual metric units of measurement (for example, millimetres) are used

    in the examples. Where appropriate, the equivalent imperial values (for example, inches)

    are shown in brackets.

    Associated publications

    When you are using the Inspection software, you may find it useful to refer to the following

    Renishaw publications:

    ! Probe systems Installation manual for machine tools (Renishaw part no.

    H-2000-6040).

    ! Probe software for machine tools data sheet (Renishaw part nos. H-2000-2289 and

    H-2000-2298).

    About the Inspection software

    For a comprehensive description of the facilities provided by the Inspection software and

    also the limitations of the software, refer to Appendix A, Features, cycles and limitations.

    Software kit no. A-4012-0541

    This kit comprises the following item:

    ! CD assembly (one CD) part no. A-4012-0480.

    The CD contains the following data:

    Probe horizontal cycles File1 (40121191)

    Probe vertical cycles File2 (40121192)

    C-axis cycles File3 (40121193)

    ReadMe (N-4012-1194)

    File1: Probe horizontal cycles

    O9721 O9724 O9726 O9730 O9732 O9801 O9802

    O9803 O9804 O9810 O9811 O9812 O9830 O9831

  • Before you begin 3

    Publication No. H-2000-6021

    File2: Probe vertical cycles

    O9521 O9524 O9526 O9530 O9532 O9601 O9602

    O9603 O9604 O9610 O9611 O9612 O9630 O9631

    File3: C-axis cycles

    O9550 O9551 O9650 O9850

    Macro memory requirements

    This section lists the amount of memory (in kilobytes) that is required by all macros

    contained on the software CD. Before you load macros, first calculate the total amount of

    memory required by the macros you wish to load. Next, check that the machine controller

    has sufficient memory capacity for these macros.

    If memory capacity of the controller is stated in metres of tape, use the following data to

    convert from kilobytes to length or vice-versa:

    Conversion: 1 KB = 2.5 m (8.2 ft) of software tape.

    8 KB = 20 m (65.6 ft) of software tape.

    File1: total amount of memory = 16 KB.

    File2: total amount of memory = 16 KB.

    File3: total amount of memory = 6 KB.

  • 4 Before you begin

    Publication No. H-2000-6021

    Macro numbers and functions

    Macro No. Function

    File 1

    O9721 X diameter move

    O9724 Settings macro

    O9726 X, Z basic move

    O9730 Print macro

    O9732 Offset update macro

    O9801 Probe length calibration

    O9802 Stylus offset (double-sided

    calibration)

    O9803 Stylus ball radius (double-

    sided calibration)

    O9804 Single-sided calibration

    O9810 Protected positioning

    O9811 XZ single surface measure

    O9812 Web / pocket measure

    O9830 Multi-stylus store

    O9831 Multi-stylus load

    Macro No. Function

    File 2

    O9521 Z diameter move

    O9524 Settings macro

    O9526 X, Z basic move

    O9530 Print macro

    O9532 Offset update macro

    O9601 Probe length calibration

    O9602 Stylus offset (double-sided

    calibration)

    O9603 Stylus ball radius (double-

    sided calibration)

    O9604 Single-sided calibration

    O9610 Protected positioning

    O9611 XZ single surface measure

    O9612 Web / pocket measure

    O9630 Multi-stylus store

    O9631 Multi-stylus load

    File 3

    O9550 C-axis basic move

    O9551 C-axis offset update

    O9650 C-axis measure vertical

    O9850 C-axis measure horizontal

  • Before you begin 5

    Publication No. H-2000-6021

    Measurement errors

    When you fit your Renishaw probe into the machines shank/holder, the probes stylus

    may not be in the correct theoretical position. An error may mean the stylus is either

    above or below the centre line. Take care to minimise this error otherwise measurements

    will be taken across a chord and will not be a true diametral or radial dimension.

    The measuring cycles do not mathematically calculate on-centre errors. However, these

    errors are usually small, particularly when the probe and stylus have been calibrated on a

    diameter that is similar in size to the feature being measured. For a description of how to

    calibrate a probe and stylus assembly, see Chapter 5, Calibrating the probe.

    Renishaw customer services

    Calling Renishaw

    If you have a question about the software, first consult the documentation and other

    printed information included with your product. If you cannot find a solution, you can

    receive information on how to obtain customer support by contacting the Renishaw

    company that serves your country (for worldwide contact details, see our main web site at

    www.renishaw.com/contact).

    When you call, it will help the Renishaw support staff if you have the appropriate product

    documentation at hand. Please be prepared to provide the following information (as

    applicable):

    ! The version of the product you are using (see the Equipment registration record

    form).

    ! The type of hardware that you are using (see the Equipment registration record

    form).

    ! The exact wording of any messages that appear on your screen.

    ! A description of what happened and what you were doing when the problem

    occurred.

    ! A description of how you tried to solve the problem.

  • 6 Before you begin

    Publication No. H-2000-6021

    This page is intentionally left blank

  • Installing the software 1-1

    Publication No. H-2000-6021

    Chapter 1

    Installing the software

    This chapter describes how to load the Inspection software. It supplements the

    information described in the Software installation section of the installation manual titled

    Probe systems for machine tools (Renishaw part no. H-2000-6040) and the Readme file

    on the CD.

    Contained in this chapter

    Installing the software ...................................................................................................... 1-2

    Using the Wizard....................................................................................................... 1-2

    Editing macros manually........................................................................................... 1-2

    Setting and adjusting the software................................................................................... 1-2

  • 1-2 Installing the software

    Publication No. H-2000-6021

    Installing the software

    Using the Wizard

    The Inspection software is supplied on CD. When the CD is inserted into the CD drive of

    the PC, it automatically launches a Wizard. The screen displays the following options:

    ! Programming manual

    ! Installation manual

    ! Readme file

    ! Generate macros.

    1. Click on Generate macros.

    2. Enter appropriate values in the fields. Information is displayed to help you.

    3. After you have completed all fields, click the Run button at the bottom of the screen.

    The system will generate the macros that are required for your machine set-up. The

    macros are stored on your PC in the folder and file that is shown above the Run

    button.

    The macros can now be loaded into the machine.

    Editing macros manually

    If, for any reason, the Wizard does not work, you will need to edit the macros manually to

    suit your machine. Do this as follows:

    1. From your PC, explore the CD and locate the folder named Macros.

    2. Open the Macros folder and locate a file named Macros.

    3. Select the files appropriate to your machine (probe vertical or probe horizontal

    and/or C-axis) and copy them to your PC.

    4. Either edit these files on your PC or load them into your machine and edit them

    there.

    Setting and adjusting the software

    After installing the software, you will need to customise the settings. Chapter 10, Settings,

    macros and variables describes general software settings, customising the software, and

    how to configure the option settings.

  • Optional inputs 2-1

    Publication No. H-2000-6021

    Chapter 2

    Optional inputs

    Many of the macros make use of standard optional inputs. Instead of describing them

    each time they are required, they are described once in this chapter. You will be referred

    to this chapter from other chapters whenever a standard optional input is required.

    Details of each non-standard optional input that is required by a macro is provided in the

    relevant macro description.

    Contained in this chapter

    Optional inputs ................................................................................................................. 2-2

  • 2-2 Optional inputs

    Publication No. H-2000-6021

    Optional inputs

    The examples given below assume that the controller has been configured for metric

    values (millimetres). The equivalent inch measurement values are shown in brackets.

    Ee e = Experience value.

    Specify the number of a spare tool offset where an adjustment value to

    the measured size is stored (see (see Chapter 11, General information).

    Example: E21. causes the experience value stored in tool offset 21 to be

    applied to the measured size.

    Ff f = This can be either one of the following:

    1. The percentage feedback that is used when updating a tool offset

    (see Chapter 10, Settings, macros and variables)

    Enter a value between 0 and 1 (0 % and 100 %).

    Default: 1 (100 %).

    2. The feedrate that is used in protected positioning macros O9610 and

    O9810 (see Chapter 4, Protected positioning cycles).

    Example: F3000 sets a feedrate of 3000 mm/min.

    (F120 sets a feedrate of 120 in/min.)

    Hh h = The tolerance value of a feature dimension being measured.

    Example: For dimension 50.0 mm +0.4 mm 0 mm, the nominal

    tolerance is 50.2 mm with H.2.

    (For dimension 1.968 in +0.016 in 0 in, the nominal

    tolerance is 1.976 in with H.008.)

    Mm m = The true position tolerance of a feature. This is a cylindrical zone about

    the theoretical position.

    Example: M.1 sets a true position tolerance of 0.1 mm.

    (M.004 sets a true position tolerance of 0.004 in.)

    Qq q = This is the probe overtravel distance for use when the default values are

    unsuitable. The probe will then travel beyond the expected position when

    it searches for a surface.

    Default: 4 mm (0.16 in) in the Z axis

    10 mm (0.394 in) in the X axis.

    Example: Q8. sets an overtravel distance of 8 mm.

    (Q.3 sets an overtravel distance of 0.3 in.)

  • Optional inputs 2-3

    Publication No. H-2000-6021

    Rr r = This is an incremental dimension that is used on external features, such

    as bosses and webs, to give a radial clearance from the nominal target

    surface prior to a Z-axis move.

    Default: 5 mm (0.200 in).

    Example: R10. sets a radial clearance of 10 mm.

    (R.4 sets a radial clearance of 0.4 in.)

    Rr r = This is similar to Rr, except that the clearance is applied in the opposite

    direction to force an internal boss or web cycle.

    Default: 5 mm (0.200 in).

    Example: R10.sets a radial clearance of 10 mm.

    (R.4 sets a radial clearance of 0.4 in.)

    Ss s = The number of the work offset to be updated.

    S0 The external work offset.

    S1 to S6 G54 to G59.

    S7 Update G10P0XZ.

    S8 The incremental G50W- - move will be active until another

    G50 is read.

    Example: S3 will set G56

    Tt t = The tool offset that is to be updated.

    T1000 A turning tool is to be updated.

    Example: T1 = update the X-wear or Z-wear compensation of a milling

    tool.

    T1001 = update the R-wear compensation of a turning tool.

    Uu u = Upper tolerance limit.

    If this value is exceeded, no tool offset or work offset is updated and the

    cycle stops with an alarm. Where applicable, this tolerance applies to both

    size and position.

    Example: U2. to set the upper tolerance limit to 2 mm.

    (U.08 to set the upper tolerance limit to 0.08 in.)

    Vv v = Null band.

    This is the tolerance zone where no tool offset adjustment occurs.

    Default: 0.

    Example: V.5 for a tolerance zone of 0.5 mm.

    (V.02 for a tolerance zone of 0.02 in.)

  • 2-4 Optional inputs

    Publication No. H-2000-6021

    Ww w = Print the output data.

    W1. Increment the feature number only.

    W2. Increment the component number and reset the feature number.

    Example: W1.

  • Variable outputs 3-1

    Publication No. H-2000-6021

    Chapter 3

    Variable outputs

    This chapter lists the variable outputs that may be produced by some of the macros. You

    will be referred to this chapter from other chapters when a variable output is produced.

    Contained in this chapter

    Table 1 Variable outputs with probe vertical .................................................................... 3-2

    Table 2 Variable outputs with probe horizontal ................................................................ 3-2

  • 3-2 Variable outputs

    Publication No. H-2000-6021

    Table 1 Variable outputs with Table 2 Variable outputs

    probe vertical with probe horizontal

    Singlesurface

    Web / pocket diameter

    C-axismeasure

    Single surface

    Web / pocket diameter

    C-axismeasure

    G65P9611 G65P9612 G65P9650 G65P9811 G65P9812 G65P9850

    # 135 X position from start

    X positionfrom start

    X position from start

    X positionfrom start

    # 136

    # 137 Z position from start

    Z position Z position from start

    Z position

    # 138 Size from start

    Size Size from start

    Size

    # 139 C position C position

    # 140 X error X error X error X error

    # 141

    # 142 Z error Z error Z error Z error

    # 143 Size error Size error Size error Size error

    # 144 C error C error

    # 145 True position error

    True position error

    True position error

    True position error

    # 146 Metal condition

    Metalcondition

    Metal condition

    Metalcondition

    # 147 Direction indicator

    Directionindicator

    Direction indicator

    Directionindicator

    # 148

    # 149

  • Protected positioning cycles 4-1

    Publication No. H-2000-6021

    Chapter 4

    Protected positioning cycles

    As the probe moves around the workpiece it is important that the stylus is protected

    against a collision with the workpiece. This chapter describes how to use the protected

    positioning cycles. After they are set correctly, the probe will stop moving in the event of a

    collision.

    Contained in this chapter

    Probe protected positioning (probe vertical) O9610 ..................................................... 4-2

    Probe protected positioning (probe horizontal) O9810 ................................................. 4-4

  • 4-2 Protected positioning cycles (probe vertical)

    Publication No. H-2000-6021

    Probe protected positioning (probe vertical) O9610

    X

    Z

    Figure 4.1 Probe protected positioning (probe vertical)

    Description

    It is important to protect the probes stylus against damage should it collide with an

    obstacle as the probe moves around the workpiece. When this cycle is used, the machine

    will stop in the event of a collision.

    Application

    The probe is selected and moved to a safe plane. At this point the probe is made active. It

    then moves to a measuring position using this macro call.

    In the event of a collision the machine will stop. Either a PATH OBSTRUCTED alarm is

    generated or error flag #148 is set (see the Mm input).

    Format

    G65 P9610 Xx and/or Zz [ Ff Mm ]

    where [ ] denote optional inputs.

    Example: G65 P9610 Z10. F3000. M1.

  • Protected positioning cycles (probe vertical) 4-3

    Publication No. H-2000-6021

    Compulsory inputs

    Xx x = The target diameter position for the probe positioning move.

    and/or

    Zz z = The target position for the probe positioning move.

    Optional inputs

    Ff f = The modal feedrate for all protected positioning moves.

    The feedrate will be modal to this macro and subsequent feedrate

    calls are unnecessary unless a change of feedrate is required. The

    maximum safe fast feedrate established during installation must not

    be exceeded.

    M1. This will set a probe trigger flag (but with no PATH OBSTRUCTED

    alarm).

    #148 = 0 No probe trigger.

    #148 = 7 Probe triggered.

    Example

    Proprietary G and M codes.

    T??? Select the probe.

    M?? Switch on the probe.

    G65P9610Z-10.X30.F3000 Protected positioning move.

    G65P9611X20.T3. Single surface measure.

    M?? Switch off the probe.

  • 4-4 Protected positioning cycles (probe horizontal)

    Publication No. H-2000-6021

    Probe protected positioning (probe horizontal) O9810

    X

    Z

    Figure 4.2 Probe protected positioning (probe horizontal)

    Description

    It is important to protect the probes stylus against damage should it collide with an

    obstacle as the probe moves around the workpiece. When this cycle is used, the machine

    will stop in the event of a collision.

    Application

    The probe is selected and moved to a safe plane. At this point the probe is made active. It

    then moves to a measuring position using this macro call.

    In the event of a collision, the machine will stop. Either a PATH OBSTRUCTED alarm is

    generated or error flag #148 is set (see the Mm input).

    Format

    G65 P9810 Xx and/or Zz [ Ff Mm ]

    where [ ] denote optional inputs.

    Example: G65 P9810 Z10. F3000. M1.

  • Protected positioning cycles (probe horizontal) 4-5

    Publication No. H-2000-6021

    Compulsory inputs

    Xx x = The target diameter position for the probe positioning move.

    and/or

    Zz z = The target position for the probe positioning move.

    Optional inputs

    Ff f = The modal feedrate for all protected positioning moves.

    The feedrate will be modal to this macro and subsequent feedrate

    calls are unnecessary unless a change of feedrate is required. The

    maximum safe fast feedrate established during installation must not

    be exceeded.

    M1. This will set a probe trigger flag (but with no PATH OBSTRUCTED

    alarm).

    #148 = 0 No probe trigger.

    #148 = 7 Probe triggered.

    Example

    Proprietary G and M codes.

    T??? Select the probe.

    M?? Switch on the probe.

    G65P9810X30.Z-10.F3000 Protected positioning move.

    G65P9811X20.T8 Single surface measure.

    M?? Switch off the probe.

  • 4-6 Protected positioning cycles (probe horizontal)

    Publication No. H-2000-6021

    This page is intentionally left blank.

  • Calibrating the probe 5-1

    Publication No. H-2000-6021

    Chapter 5

    Calibrating the probe

    Before a probe and stylus are used, they must be calibrated correctly. Only when they

    have been calibrated accurately can you achieve total quality control over your

    manufacturing process.

    This chapter explains why it is so important that the probe and stylus are calibrated and

    then describes how to use the macros that are provided to calibrate them.

    Contained in this chapter

    Why calibrate a probe and stylus?................................................................................... 5-2

    Single-sided calibration.................................................................................................... 5-2

    Double-sided calibration .................................................................................................. 5-3

    Calibrating the probe length............................................................................................. 5-3

    Calibration cycles............................................................................................................. 5-3

    Calibrating the probe length (probe vertical) O9601..................................................... 5-5

    Calibrating the stylus Z tool offset double-sided (probe vertical) O9602..................... 5-7

    Calibrating the stylus ball radius double-sided (probe vertical) O9603 ....................... 5-9

    Calibrating the stylus single-sided (probe vertical) O9604 ......................................... 5-11

    Calibrating the probe length (probe horizontal) O9801 .............................................. 5-13

    Calibrating the stylus X tool offset double-sided (probe horizontal) O9802 ............... 5-15

    Calibrating the stylus ball radius double-sided (probe horizontal) O9803.................. 5-17

    Calibrating the stylus single-sided (probe horizontal) O9804..................................... 5-19

  • 5-2 Calibrating the probe

    Publication No. H-2000-6021

    Why calibrate a probe and stylus?

    When you fit a probe into the machines shank/holder it is not necessary for the probes

    stylus to run true to the spindle centre-line. A small amount of run-out can be tolerated.

    However, it is good practice to mechanically adjust the stylus so that it is true to the

    virtual Y axis. This helps reduce the effects of cosine errors when measuring different

    sizes.

    Calibrating the probe and stylus ensures that run-out is automatically accounted for.

    Without calibration, run-out will lead to inaccurate results.

    It is important that you calibrate the stylus of a Renishaw inspection probe in the following

    circumstances:

    ! When a probe and stylus are to be used for the first time.

    ! When a new stylus is fitted to the probe.

    ! When it is suspected that the stylus has become distorted or that the probe has

    crashed.

    ! At regular intervals to compensate for mechanical changes of your lathe.

    Three different operations are used to calibrate a stylus. They are:

    ! Calibrating single-sided.

    ! Calibrating double-sided.

    ! Calibrating the probe length.

    Single-sided calibration

    Single-sided calibration determines the error value for the stylus ball in relation to a

    known surface. This value is stored in the tool wear offset and is used automatically in the

    measuring cycles.

    NOTE: Only one edge of the stylus is calibrated. Therefore, all subsequent measuring

    cycles must use only this calibrated edge.

  • Calibrating the probe 5-3

    Publication No. H-2000-6021

    Double-sided calibration

    Double-sided calibration determines the stylus offset and the radius of the stylus ball. The

    offset is stored in the tool wear offset while the radius is stored in a macro variable. Both

    values are used automatically in the measuring cycles.

    NOTE: Both edges of the stylus are calibrated. Therefore, measuring cycles that use

    both sides of the stylus can be used.

    Calibrating the probe length

    Calibrating a probe on a known reference surface determines the length of the probe,

    based on the electronic trigger point. The stored value for length is different from the

    physical length of the probe assembly.

    Additionally, the operation can automatically compensate for machine and fixture height

    errors by adjusting the length value of the probe that is stored.

    Calibration cycles

    Eight calibration cycles are provided with the Inspection software. Four are for calibrating

    a probe positioned vertically and four are for calibrating a probe positioned horizontally.

    They may be used in conjunction with one another for complete calibration of the probe.

    The purpose of each macro is summarised below.

    Macros O9601 and O9801 Used to establish the length of the probe in the tool

    shank.

    Macros O9602 and O9802 Used to establish the stylus offset (for double-sided

    calibration only).

    Macros O9603 and O9803 Used to establish the radius values of the stylus ball (for

    double-sided calibration only).

    Macros O9604 and O9804 Used to establish the error value of the stylus ball (for

    single-sided calibration only).

    For complete calibration of a probe system, use the macros as follows:

    ! For double-sided calibration of a vertically mounted probe, use macros O9601,

    O9602 and O9603.

    ! For double-sided calibration of a horizontally-mounted probe, use macros O9801,

    O9802 and O9803.

  • 5-4 Calibrating the probe

    Publication No. H-2000-6021

    ! For single-sided calibration of a vertically mounted probe, use macros O9601 and

    O9604.

    ! For single-sided calibration of a horizontally-mounted probe, use macros O9801

    and O9804.

    The Renishaw calibration cycles are split into separate cycles for flexibility.

  • Calibrating the probe (probe vertical) 5-5

    Publication No. H-2000-6021

    Calibrating the probe length (probe vertical) O9601

    Tttool offset

    Xxreference

    size

    X

    Z

    Figure 5.1 Calibrating the probe length (probe vertical)

    Description

    The probe is positioned adjacent to a X-axis reference surface. When the calibration

    cycle is completed the active tool offset is adjusted to the reference surface.

    Application

    An approximate tool offset is loaded. The probe is positioned adjacent to the reference

    surface.

    When the cycle is run, the surface is measured and the tool offset is reset to a new value.

    The probe is then returned to the start position.

    Format

    G65 P9601 Xx Tt

    Example: G65 P9601 X50. T20

  • 5-6 Calibrating the probe (probe vertical)

    Publication No. H-2000-6021

    Compulsory inputs

    Tt t = The active tool offset number.

    Xx x = The size of the reference surface.

    Outputs

    The active tool offset is set.

    Example: Probe length calibration

    O0001

    Proprietary G and M codes

    T0101 Select the probe.

    M?? Switch on the probe.

    G65P9610X60.F3000 Protected positioning move.

    G65P9601X50.T1 Calibrate in the X direction and update tool offset 1.

    G65P9610X100. Protected positioning move.

    M?? Switch off the probe.

  • Calibrating the probe (probe vertical) 5-7

    Publication No. H-2000-6021

    Calibrating the stylus Z tool offset double-sided (probevertical) O9602

    X

    Z

    A

    X

    R

    Z

    B

    Z

    Figure 5.2 Calibrating the stylus Z tool offset double-sided (probe vertical)

    Description

    Before the cycle is run, the probe is positioned either above a pre-machined web (A) or

    inside a pre-machined groove (B). When the cycle is completed, the tool offset is set to

    the centre of the stylus in the Z axis.

    Application

    Pre-machine either a web or a groove with a suitable tool so that the exact centre of the

    feature is known. Position the probe to the centre of the web or groove.

    When the cycle is run, two measuring moves are made to determine the Z offset of the

    stylus. The probe is then returned to the start position.

    Format

    G65 P9602 Zz Tt [ Rr Xx ]

    where [ ] denote optional inputs.

    Example: G65 P9602 Z50.005 T8. X50.

    Compulsory inputs

    Tt t = The tool offset number that will be updated.

    Zz x = The nominal size of the feature.

  • 5-8 Calibrating the probe (probe vertical)

    Publication No. H-2000-6021

    Optional input

    Xx x = The absolute X axis measuring position when calibrating on an external

    feature. If this is omitted, a groove cycle is assumed.

    For the Rr optional input, see Chapter 2, "Optional inputs".

    Outputs

    The Z tool offset will be updated to the centre of the stylus.

    Example: Calibrating the stylus Z offset

    Run a complete positioning and calibration program as follows.

    Set the exact X and Z feature positions in a work offset (for example, using G54).

    O0002

    T0101 Select the probe.

    M?? Switch on the probe.

    G54 Select the work offset.

    G65P9610Z-20.F3000 Protected positioning move over the groove.

    G65P9610X20. Protected positioning move into the groove.

    G65P9602Z50.T1 Calibrate in a 50 mm (1.97 in) groove and update tool offset 1.

    G65P9610X100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe vertical) 5-9

    Publication No. H-2000-6021

    Calibrating the stylus ball radius double-sided (probevertical) O9603

    X

    Z

    A

    X

    R

    Z

    B

    Z

    Figure 5.3 Calibrating the stylus ball radius double-sided (probe vertical)

    Description

    Before the cycle is run, the probe is positioned either above a pre-machined web (A) or

    inside a pre-machined groove (B). When the cycle is completed, the radius values of the

    stylus ball are stored.

    Application

    The probe to be calibrated is positioned either above the web or inside the groove of

    known size. The position of the web or groove is not critical.

    When the cycle is run, two moves are made to determine the radius values of the stylus

    ball. The probe is then returned to the start position.

    Format

    G65 P9603 Zz [ Rr Xx ]

    where [ ] denote optional inputs.

    Example: G65 P9603 Z50.005 X50.

    Compulsory input

    Zz z = The nominal size of the reference web or groove.

  • 5-10 Calibrating the probe (probe vertical)

    Publication No. H-2000-6021

    Optional input

    Xx x = The absolute X axis measuring position when calibrating on an external

    feature. If this is omitted, a groove cycle is assumed.

    For the Rr optional input, see Chapter 2, "Optional inputs".

    Outputs

    The following data will be stored:

    #503 Stylus ball radius in Z.

    NOTE: The actual variable is dependent on #112 that is set in O9724.

    Example: Calibrating the radius of a stylus ball

    Run a complete positioning and calibration program as follows.

    Set the approximate X and Z feature positions in a work offset (for example, using G54).

    O0004

    T0101 Select the probe.

    G54 Select the work offset.

    M?? Switch on the probe.

    G65P9610Z-20.F3000 Protected positioning move over the groove.

    G65P9610X20. Protected positioning move into the groove.

    G65P9603Z15.001 Calibrate in a 15.001 mm (0.5905 in) groove.

    G65P9610X100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe vertical) 5-11

    Publication No. H-2000-6021

    Calibrating the stylus single-sided (probe vertical) O9604

    NOTE: When this method is used for calibrating a stylus, only the calibrated edge of the

    stylus must be used in subsequent measuring cycles.

    X

    Z

    A B

    Z

    OR

    Figure 5.4 Calibrating the stylus single-sided (probe vertical)

    Description

    Before the cycle is run, the probe is positioned adjacent to a Z axis reference surface.

    When the cycle is completed, the active probe tool offset is adjusted to the reference

    surface.

    Application

    An approximate tool offset is loaded. The probe is positioned adjacent to the reference

    surface.

    When the cycle is run, the surface is measured and the tool offset is reset to a new value.

    The probe is then returned to the start position.

    Format

    G65 P9604 Zz Tt

    Example: G65 P9604 Z0. T1.

    Compulsory inputs

    Tt t = The active tool offset number.

    Zz z = The position of the reference surface.

  • 5-12 Calibrating the probe (probe vertical)

    Publication No. H-2000-6021

    Outputs

    The Z tool offset will be updated to the edge of the stylus.

    Example: Calibrating a stylus edge

    Run a complete positioning and calibration program as follows.

    Set the exact X and Z feature positions in a work offset (for example, using G54).

    O0002

    T0101 Select the probe.

    M?? Switch on the probe.

    G54 Select the work offset.

    G65P9610Z10.F3000 Protected positioning move.

    G65P9610X20. Protected positioning move.

    G65P9604Z0.T1 Calibrate in the Z direction and update tool offset 1.

    G65P9610X100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe horizontal) 5-13

    Publication No. H-2000-6021

    Calibrating the probe length (probe horizontal) O9801

    Tttool offset

    Zz reference height

    X

    Z

    Figure 5.5 Calibrating the probe length (probe horizontal)

    Description

    The probe is positioned adjacent to a Z axis reference surface for calibration. When the

    cycle is completed, the active probe tool offset is adjusted to the reference surface.

    Application

    An approximate tool offset is loaded. The probe should be positioned adjacent to the

    reference surface.

    When the cycle is run, the surface is measured and the tool offset is reset to a new value.

    The probe is then returned to the start position.

    Format

    G65 P9801 Zz Tt

    Example: G65 P9801 Z10. T20

    Compulsory inputs

    Tt t = The active tool offset number.

    Zz z = The position of the reference surface.

  • 5-14 Calibrating the probe (probe horizontal)

    Publication No. H-2000-6021

    Outputs

    The active tool offset is set.

    Example: Probe length calibration

    O0001

    T0101 Select the probe.

    G54 Select the work offset.

    M?? Switch on the probe.

    G65P9810Z10.F3000 Protected positioning move.

    G65P9801Z-10.T1 Calibrate in the Z direction and update tool offset 1.

    G65P9810Z100. Protected positioning move.

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe horizontal) 5-15

    Publication No. H-2000-6021

    Calibrating the stylus X tool offset double-sided (probe horizontal) O9802

    X

    Z

    R

    Z

    X

    X

    X

    R

    A

    A

    B

    Z

    Figure 5.6 Calibrating the stylus X tool offset double-sided (probe horizontal)

    Description

    Before the cycle is run, the probe is positioned in front of the diameter (A) or inside the

    bore (B). When this cycle is completed, the centre of the stylus is updated to the tool

    offset.

    Application

    Pre-machine either a hole with a suitable boring bar or a diameter with a turning tool.

    Position the probe as described above.

    When the cycle is run, two measuring moves are made to determine the X offset of the

    stylus. The probe is then returned to the start position.

    Format

    G65 P9802 Xx Tt [ Rr Zz ]

    where [ ] denote optional inputs.

    Example: G65 P9802 X50.005 T8. Z50.

  • 5-16 Calibrating the probe (probe horizontal)

    Publication No. H-2000-6021

    Compulsory inputs

    Tt t = The tool offset number that will be updated.

    Xx x = The nominal size of the feature.

    Optional input

    Zz z = The absolute Z axis measuring position when calibrating on an external

    feature. If this is omitted, a bore cycle is assumed.

    For the Rr optional input, see Chapter 2, "Optional inputs".

    Outputs

    The X tool offset will be updated to the centre of the stylus.

    Example: Calibrating the stylus X offset

    Run a complete positioning and calibration program as follows.

    Set the exact X and Z feature positions in a work offset (for example, using G54).

    O0002

    T0101 Select the probe.

    M?? Switch on the probe.

    G65P9810X0.Z10.F3000 Protected positioning move over the hole.

    G65P9810Z5. Protected positioning move into the hole.

    G65P9802X50.T1 Calibrate in a 50 mm (1.97 in) diameter bored hole and

    update tool offset 1.

    G65P9810Z100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe horizontal) 5-17

    Publication No. H-2000-6021

    Calibrating the stylus ball radius double-sided (probe horizontal) O9803

    X

    Z

    R

    Z

    X

    X

    X

    R

    A

    A

    B

    Z

    Figure 5.7 Calibrating the stylus ball radius double-sided (probe horizontal)

    Description

    Before the cycle is run, the probe is positioned either in front of the diameter (A) or inside

    the bore (B). When the cycle is completed, the stylus centre is updated to the tool offset.

    Application

    Pre-machine either a hole with a suitable boring bar or a diameter with a turning tool.

    Position the probe as described above.

    When the cycle is run, two measuring moves are made to determine the X radius values

    of the stylus ball. The probe is then returned to the start position.

    Format

    G65 P9803 Xx [ Rr Zz ]

    where [ ] denote optional inputs.

    Example: G65 P9803 X50.005 Z50.

  • 5-18 Calibrating the probe (probe horizontal)

    Publication No. H-2000-6021

    Compulsory input

    Xx x = The nominal size of the feature.

    Optional input

    Zz z = The absolute Z axis measuring position when calibrating on an external

    feature. If this is omitted, a ring gauge cycle is assumed.

    For the Rr optional input, see Chapter 2, "Optional inputs".

    Outputs

    The following data will be stored:

    #500 Stylus ball radius in X (XRAD)

    NOTE: The actual variable is dependent on #111 that is set in O9724.

    Example: Calibrating the radius of a stylus ball

    Run a complete positioning and calibration program as follows.

    Set the approximate X and Z feature positions in a work offset (for example, using G54).

    O0004

    T0101 Select the probe.

    M?? Switch on the probe.

    G65P9810X0.Z10.F3000 Protected positioning move over the hole.

    G65P9810Z5. Protected positioning move into the hole.

    G65P9803X50.001 Calibrate in a 50.001 mm (1.9685 in) diameter ring gauge.

    G65P9810Z100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Calibrating the probe (probe horizontal) 5-19

    Publication No. H-2000-6021

    Calibrating the stylus single-sided (probe horizontal) O9804

    NOTE: When this method is used for calibrating a stylus, only the calibrated edge of the

    stylus must be used in subsequent measuring cycles.

    X

    Z

    A B

    OR

    Figure 5.8 Calibrating the stylus single-sided (probe horizontal)

    Description

    Before the cycle is run, the probe is positioned either above a diameter (A) or inside a

    diameter (B). When the cycle is completed, the edge of the stylus is updated to the tool

    offset.

    Application

    Pre-machine either a hole with a suitable boring bar or a diameter with a turning tool.

    Position the probe as described above and run the cycle.

    When the cycle is run, a measuring move is made to determine the X offset of the stylus.

    The probe is then returned to the start position.

    Format

    G65 P9804 Xx Tt

    where [ ] denote optional inputs.

    Example: G65 P9804 X50.005 T8.

    Compulsory inputs

    Tt t = The active tool offset number.

    Xx x = The nominal diameter of the feature.

  • 5-20 Calibrating the probe (probe horizontal)

    Publication No. H-2000-6021

    Outputs

    The X tool offset will be updated to the edge of the stylus.

    Example: Single-sided stylus calibration

    Run a complete positioning and calibration program as follows.

    Set the exact X and Z feature positions in a work offset (for example, using G54).

    O0002

    T0808 Select the probe.

    M?? Switch on the probe.

    G65P9810X50.Z-6.F3000 Protected positioning move over the hole.

    G65P9804X40.T8 Calibrate against a 40 mm (1.57 in) diameter and update

    tool offset 8.

    G65P9810Z100. Protected positioning move retract to 100 mm (3.94 in).

    M?? Switch off the probe.

    M30 End of the program.

  • Standard measuring cycles 6-1

    Publication No. H-2000-6021

    Chapter 6

    Standard measuring cycles

    This chapter describes how to use the non-vector measuring cycles with vertically-

    mounted and horizontally-mounted probes.

    Contained in this chapter

    X Z single surface measurement (probe vertical) O9611 ............................................. 6-2

    Web/pocket measurement (probe vertical) O9612 ....................................................... 6-5

    X Z single surface measurement (probe horizontal) O9811 ......................................... 6-8

    Web/pocket measurement (probe horizontal) O9812 ................................................. 6-11

  • 6-2 Standard measuring cycles (probe vertical)

    Publication No. H-2000-6021

    X Z single surface measurement (probe vertical) O9611

    X

    Z

    Figure 6.1 Measurement of a single surface (probe vertical)

    Description

    This cycle measures a surface to establish the size or position.

    Application

    The probe, with its tool offset active, should be positioned adjacent to the surface. The

    cycle measures the surface and returns to the start position.

    There are two possibilities, as follows:

    1. The surface can be treated as a size, where the tool offset is updated in conjunction

    with the Tt and the Hh input.

    2. The surface can be treated as a reference surface position, for the purpose of

    adjusting a work offset using the Ss and Mm inputs.

    Format

    G65 P9611 Xx or Zz [ Cc Ee Ff Hh Ii Qq Ss Tt Uu Vv Ww ]

    where [ ] denote optional inputs.

    Example: G65 P9611 X50. C1. E90. F0.8 H0.2 I0.5 Q10. S1. T20. U.5 V.5 W2.

    Compulsory inputs

    Xx x = The size of the target surface.

    or

    Zz z = The position of the target surface.

  • Standard measuring cycles (probe vertical) 6-3

    Publication No. H-2000-6021

    Optional inputs

    Inputs Cc and Ii are used with the reference touch option. This option eliminates thermal

    growth errors from within the probe results, helping to maintain measuring accuracy.

    Cc c = 1. Takes the error from the reference diameter and stores it in #134

    c = 2. Takes the error stored in #134 and adds it to the results of the measured

    diameter

    Ii i = This is the maximum tolerance used with the C input.

    For other optional inputs, see Chapter 2, "Optional inputs".

    X

    Z

    Chuck

    C = 1

    C = 2

    Figure 6.2 Reference touch option

    Example: X and Z single surface measurement

    X

    Z

    4

    3

    2

    17

    5

    6

    Figure 6.3 Probe movements

  • 6-4 Standard measuring cycles (probe vertical)

    Publication No. H-2000-6021

    T???? Select the probe.

    M?? Switch on the probe.

    1. G65P9610X50.Z10.F3000 Protected positioning move to the start position.

    2. G65P9610X15. Protected positioning move.

    3. G65P9611Z0.T2 Single surface measure.

    4. G65P9610X80. Protected positioning move.

    5. G65P9610Z6. Protected positioning move.

    6. G65P9611X40.T4 Single surface measure.

    7. G65P9610X100. Protected positioning move.

    M?? Switch off the probe.

    continue

    Tool offset 2 (Z wear) and tool offset 4 (X wear) will be updated.

  • Standard measuring cycles (probe vertical) 6-5

    Publication No. H-2000-6021

    Web/pocket measurement (probe vertical) O9612

    NOTE: This cycle must not be used unless double-sided calibration has taken place.

    X

    Z

    Z

    X

    X

    Z

    Z

    -R

    R

    Figure 6.4 Measurement of a web or pocket feature (probe vertical)

    Description

    This cycle measures a web or pocket feature. It uses two measuring moves along the

    Z axis.

    Application

    Position the probe to the expected centre line of the feature and at a suitable position in

    the X axis with the probe and probe offset active. Run the cycle with suitable inputs as

    described.

    Format

    G65 P9612 Zz [ Ee Ff Hh Mm Qq Rr Ss Tt Uu Vv Ww ]

    or

    G65 P9612 Xx Zz [ Ee Ff Hh Mm Qq Rr Ss Tt Uu Vv Ww ]

  • 6-6 Standard measuring cycles (probe vertical)

    Publication No. H-2000-6021

    where [ ] denote optional inputs.

    Example: G65 P9612 X50. Z100. E91 F0.8 H0.2 M.2 Q10. R10. S1. T20. U.5 V.5 W2.

    Compulsory inputs

    Xx x = The absolute X axis position when measuring a web feature. If this is

    omitted, a pocket cycle is assumed.

    Zz z = The nominal size of the feature when measured in the Z axis.

    Optional inputs

    See Chapter 2, "Optional inputs".

    Outputs

    See Chapter 3, "Variable outputs".

    Example 1: Measuring a web

    X

    Z

    4

    3

    2

    1

    3

    Figure 6.5 Probe movements

    G54

    T????? Select the probe.

    M?? Switch on the probe.

    1. G65P9610Z20.X60.F3000 Protected positioning move.

    2. G65P9610Z0. Protected positioning move.

    3. G65P9612X40.Z10.S2 Measure a 10.0 mm (1.968 in) wide web.

  • Standard measuring cycles (probe vertical) 6-7

    Publication No. H-2000-6021

    4. G65P9610X100. Protected positioning move.

    M?? Switch off the probe.

    continue

    The centre line of the feature in the X axis is stored in the work offset 02 (G55).

    Example 2: Measuring a pocket (referred datum)

    X

    Z

    4

    3

    2

    15

    Figure 6.6 Probe movements

    G54

    T????? Select the probe.

    M?? Switch on the probe.

    1. G65P9610X100.Z20.F3000 Protected positioning move.

    2. G65P9610Z20. Protected positioning move.

    3. G65P9610X40. Protected positioning move.

    4. G65P9612Z10.S2 Measure a 10.0 mm (1.181 in) wide pocket.

    5. G65P9610X100. Protected positioning move.

    M?? Switch off the probe.

    continue

    The error of the centre line is referred to the datum point Z0 and the revised Z0 position is

    set in work offset 02 (G55).

  • 6-8 Standard measuring cycles (probe horizontal)

    Publication No. H-2000-6021

    X Z single surface measurement (probe horizontal) O9811

    X

    Z

    Figure 6.7 Measurement of a single surface (probe horizontal)

    Description

    This cycle measures a surface to establish the size or position.

    Application

    The probe, with its tool offset active, should be positioned adjacent to the surface. The

    cycle measures the surface and returns to the start position.

    There are two possibilities, as follows:

    1. The surface can be treated as a size, where the tool offset is updated in conjunction

    with the Tt and the Hh input.

    2. The surface can be treated as a reference surface position, for the purpose of

    adjusting a work offset using the Ss and Mm inputs.

    Format

    G65 P9811 Xx or Zz [ Cc Ee Ff Hh Ii Qq Ss Tt Uu Vv Ww ]

    where [ ] denote optional inputs.

    Example: G65 P9811 X50. C1. E90 F0.8 H0.2 I0.5 Q10. S1. T20. U.5 V.5 W2.

    Compulsory inputs

    Xx x = The size of the target surface.

    or

    Zz z = The position of the target surface.

  • Standard measuring cycles (probe horizontal) 6-9

    Publication No. H-2000-6021

    Optional inputs

    Inputs Cc and Ii are used with the reference touch option. This option eliminates thermal

    growth errors from within the probe results, helping to maintain measuring accuracy.

    Cc c = 1. Takes the error from the reference diameter and stores it in #134.

    c = 2. Takes the error stored in #134 and adds it to the results of the measured

    diameter.

    Ii i = This is the maximum tolerance used with the C input.

    For other optional inputs, see Chapter 2, "Optional inputs".

    X

    Z

    Chuck

    C = 1

    C = 2

    Figure 6.8 Reference touch option

    Example: X and Z single surface measurement

    X

    Z

    4

    3

    2

    17

    5

    6

    Figure 6.9 Probe movements

    G54

    T????? Select the probe.

  • 6-10 Standard measuring cycles (probe horizontal)

    Publication No. H-2000-6021

    M?? Switch on the probe.

    1. G65P9810Z10.F3000 Protected positioning move.

    2. G65P9810X20. Protected positioning move to the start position.

    3. G65P9811Z0.T2 Single surface measure.

    4. G65P9810X80. Protected positioning move.

    5. G65P9810Z20. Protected positioning move.

    6. G65P9811X60.T4 Single surface measure.

    7. G65P9810X100. Protected positioning move.

    M?? Switch off the probe.

    continue

    Tool offset 2 (Z wear) and tool offset 4 (X wear) will be updated.

  • Standard measuring cycles (probe horizontal) 6-11

    Publication No. H-2000-6021

    Web/pocket measurement (probe horizontal) O9812

    NOTE: This cycle must not be used unless double-sided calibration has taken place.

    X

    Z

    XX

    X

    Z

    Z

    R

    -R

    Figure 6.10 Measurement of a web or pocket feature (probe horizontal)

    Description

    This cycle measures a web or pocket feature. It uses two measuring moves along the X

    axis.

    Application

    Position the probe to the expected centre line of the feature and at a suitable position in

    the Z axis with the probe and probe offset active. Run the cycle with suitable inputs as

    described.

    Format

    G65 P9812 Xx [ Ee Ff Hh Mm Qq Rr Ss Tt Uu Vv Ww ]

    or

    G65 P9812 Xx Zz [ Ee Ff Hh Mm Qq Rr Ss Tt Uu Vv Ww ]

    where [ ] denote optional inputs.

  • 6-12 Standard measuring cycles (probe horizontal)

    Publication No. H-2000-6021

    Example: G65 P9812 X50. Z100. E92 F0.8 H0.2 M.2 Q10. R10. S1. T20. U.5 V.5 W2.

    Compulsory inputs

    Xx x = The nominal size of the feature.

    or

    Zz z = The absolute Z axis position when measuring a web feature. If this is

    omitted, a pocket cycle is assumed.

    Optional inputs

    See Chapter 2, "Optional inputs".

    Outputs

    See Chapter 3, "Variable outputs".

    Example 1: Measuring a web

    X

    Z

    3

    2

    1

    7

    5

    6

    4

    Figure 6.11 Probe movements

    G54

    T????? Select the probe.

    M?? Switch on the probe.

    G65P9810.Z10.F3000 Protected positioning move.

    1. G65P9810X0. Protected positioning move.

    2 to 7. G65P9812X50.Z10.S2 Measure a 50.0 mm (1.968 in) wide web.

    G65P9810Z10. Protected positioning move.

  • Standard measuring cycles (probe horizontal) 6-13

    Publication No. H-2000-6021

    M?? Switch off the probe.

    continue

    The centre line of the feature in the X axis is stored in work offset 02 (G55).

    Example 2: Measuring a pocket

    X

    Z

    32 1

    Figure 6.12 Probe movements

    T????? Select the probe.

    M?? Switch on the probe.

    1. G65P9810Z10.F3000 Protected positioning move.

    2. G65P9810Z-10. Protected positioning move.

    3. G65P9812X30.S2 Measure a 30.0 mm (1.181 in) wide pocket.

    G65P9810Z10. Protected positioning move.

    M?? Switch off the probe.

    continue

    The centre line of the feature in the X axis is stored in work offset 02 (G55).

  • 6-14 Standard measuring cycles (probe horizontal)

    Publication No. H-2000-6021

    This page is intentionally left blank.

  • Additional cycles 7-1

    Publication No. H-2000-6021

    Chapter 7

    Additional cycles

    The Inspection software contains four macro cycles that cannot be described under the

    headings used in previous chapters (see chapters 4 to 6 inclusive). This chapter describes

    how to use these cycles.

    Contained in this chapter

    Storing multi-stylus data (probe vertical) O9630........................................................... 7-2

    Loading multi-stylus data (probe vertical) O9631.......................................................... 7-4

    Storing multi-stylus data (probe horizontal) O9830....................................................... 7-6

    Loading multi-stylus data (probe horizontal) O9831 ..................................................... 7-8

  • 7-2 Additional cycles (probe vertical)

    Publication No. H-2000-6021

    Storing multi-stylus data (probe vertical) O9630

    K1Calibration data

    K2Calibration data

    Probe 1 Probe 2

    Figure 7.1 Storing multi-stylus data (probe vertical)

    Description

    The macro is used to store the stylus calibration data that is established during the

    calibration cycles. The data is stored in a spare set of macro variables.

    Stored data for each probe can be recalled using macro O9631.

    Application

    Calibrate the stylus using macros O9602 and O9603. Run macro O9630 to store the

    calibration data #503. The values are recalled when macro O9631 is run.

    NOTE: All # numbers shown are with the base number set to 503 in O9524.

    Format

    G65 P9630 Kk

    Example: G65 P9630 K1.

  • Additional cycles (probe vertical) 7-3

    Publication No. H-2000-6021

    Compulsory input

    Kk k = This is the multi-stylus probe number that is used for storing calibration

    data.

    The active calibration data is stored in the macro variables shown in the

    following table. The actual variables used are dependent on the #112

    setting in O9524.

    Active data K1 K2

    Ball radius Z #503 #504 #505

    NOTE: Spare macro variable locations, in which the calibration data can be stored, are

    dependent on the machine options available and the variables that are already used by

    your programs. These can be moved, depending on the #112 setting in O9524.

    Example: Multi-stylus store K1

    O0003

    G54

    M?? Switch on the probe.

    G65P9610X5.F3000 Protected positioning move for X.

    G65P9603Z50.001 Calibrate in a reference feature (a ring gauge).

    G65P9630K1. Store calibration values for multi-stylus K1.

    G65P9610X100. Protected positioning move.

    M?? Switch off the probe.

    M30 End of the program.

  • 7-4 Additional cycles (probe vertical)

    Publication No. H-2000-6021

    Loading multi-stylus data (probe vertical) O9631

    K1Calibration data

    K2Calibration data

    Probe 1 Probe 2

    Figure 7.2 Loading multi-stylus data (probe vertical)

    Description

    This macro is used (in conjunction with O9630) to load stored calibration data for a

    specific probe/stylus configuration into the active calibration variable range. It must be run

    before using the probe and stylus.

    Application

    The program must be run immediately before a measuring cycle to select the correct

    probe calibration data. The Kk input determines the data that is to be loaded.

    NOTE: All # numbers shown are with the base number set to 503 in O9524.

    Format

    G65 P9631 Kk

    Example: G65 P9631 K1.

  • Additional cycles (probe vertical) 7-5

    Publication No. H-2000-6021

    Compulsory input

    Kk k = This is the multi-stylus probe number that is used for loading stored data.

    For a specific probe/stylus configuration, the Kk input number must be

    the same as that used when storing the data with O9630.

    The active calibration data is stored in the macro variables shown in the

    following table. The actual variables used are dependent on the #112

    setting in O9524.

    Active data K1 K2

    Ball radius Z #503 #504 #505

    NOTE: Spare macro variable locations, in which the calibration data can be stored, are

    dependent on the machine options available and the variables that are already used by

    your programs. These can be moved, depending on the #112 setting in O9524.

    Example: Multi-stylus load K1

    G54

    M?? Switch on the probe.

    G65P9610X10.F3000 Protected positioning move.

    G65P9631K1. Load calibration values for multi-stylus K1.

    G65P9612Z30.S2 Measure a 30.0 mm (1.181 in) diameter bore.

    G65P9610X10. Protected positioning move.

    G65P9610Z350. Protected positioning move.

    M?? Switch off the probe.

    continue

  • 7-6 Additional cycles (probe horizontal)

    Publication No. H-2000-6021

    Storing multi-stylus data (probe horizontal) O9830

    K1Calibration

    data

    K2Calibration

    data

    Probe 1

    Probe 2

    Figure 7.3 Storing multi-stylus data (probe horizontal)

    Description

    This macro is used to store stylus calibration data that is established during the calibration

    cycles. The data is stored in a spare set of macro variables.

    Stored data for each probe can be recalled using macro O9831.

    Application

    Calibrate the stylus using macro O9802 and O9803. Run macro O9830 to store the

    calibration data #500. The values are recalled when macro O9831 is run.

    NOTE: All # numbers shown are with the base number set to 500 in O9724.

    Format

    G65 P9830 Kk

    Example: G65 P9830 K1.

    Compulsory input

    Kk k = This is the multi-stylus probe number that is used for storing calibration

    data.

    The active calibration data is stored in the macro variables shown in the

    following table. The actual variables used are dependent on the #111

    setting in O9724.

  • Additional cycles (probe horizontal) 7-7

    Publication No. H-2000-6021

    Active data K1 K2

    Ball radius X #500 #501 #502

    NOTE: Spare macro variable locations, in which the calibration data can be stored, are

    dependent on the machine options available and the variables that are already used by

    your programs. These can be moved, depending on the #111 setting in O9724.

    Example: Multi-stylus store K1

    O0003

    G54

    M?? Switch on the probe.

    G65P9810Z-5.F3000 Protected positioning move for Z.

    G65P9803X50.001 Calibrate in a reference feature (a ring gauge).

    G65P9830K1. Store calibration values for multi-stylus K1.

    G65P9810Z100. Protected positioning move.

    M?? Switch off the probe.

    M30 End of the program.

  • 7-8 Additional cycles (probe horizontal)

    Publication No. H-2000-6021

    Loading multi-stylus data (probe horizontal) O9831

    K1Calibration

    data

    K2Calibration

    data

    Probe 1

    Probe 2

    Figure 7.4 Loading multi-stylus data (probe horizontal)

    Description

    This macro is used (in conjunction with O9830) to load stored calibration data for a

    specific probe/stylus configuration into the active calibration variable range. It must be run

    before using the probe and stylus.

    Application

    The program must be run immediately before any measuring cycle to select the correct

    probe calibration data. The Kk input determines the data that is to be loaded.

    NOTE: All # numbers shown are with the base number set to 500 in O9724.

    Format

    G65 P9831 Kk

    Example: G65 P9831 K1.

    Compulsory input

    Kk k = This is the multi-stylus probe number that is used for loading stored data.

    For a specific probe/stylus configuration, the Kk input number must be

    the same as that used when storing the data with O9830.

    The active calibration data is stored in the macro variables shown in the

    following table. The actual variables used are dependent on the #111

    setting in O9724.

  • Additional cycles (probe horizontal) 7-9

    Publication No. H-2000-6021

    Active data K1 K2

    Ball radius X #500 #501 #502

    NOTE: Spare macro variable locations, in which the calibration data can be stored, are

    dependent on the machine options available and the variables that are already used by

    your programs. These can be moved, depending on the #111 setting in O9724.

    Example: Multi-stylus load K1

    G54

    M?? Switch on the probe.

    G65P9810Z10.F3000 Protected positioning move.

    G65P9831K1. Load calibration values for multi-stylus K1.

    G65P9812X30.S2 Measure a 30.0 mm (1.181 in) diameter bore.

    G65P9810Z10. Protected positioning move.

    G65P9810X350. Protected positioning move.

    M?? Switch off the probe.

  • 7-10 Additional cycles (probe horizontal)

    Publication No. H-2000-6021

    This page is intentionally left blank

  • C-axis cycles 8-1

    Publication No. H-2000-6021

    Chapter 8

    C-axis cycles

    This chapter describes how to use the two C-axis macros that are supplied as part of the

    Inspection software.

    Contained in this chapter

    C-axis find (probe vertical) O9650 ................................................................................ 8-2

    C-axis find (probe horizontal) O9850 ............................................................................ 8-4

  • 8-2 C-axis cycles (probe vertical)

    Publication No. H-2000-6021

    C-axis find (probe vertical) O9650

    Figure 8.1 C-axis find (probe vertical)

    Description

    This cycle will take two readings by moving the C-axis. External features can also be

    measured by moving the X axis. The results can be used to update a work shift.

    Application

    The stylus must be programmed to a start position that is either above an external feature

    or inside an internal feature. The C-axis must be engaged before calling up the cycle.

    The inclusion of an Xx input indicates that an external feature is to be measured.

    Format

    G65 P9650 Dd [ Xx Qq Rr Hh Ss ]

    where [ ] denote optional inputs.

    Example: G65 P9650 D20. X10. Q10. R10. H0.2 S1.

    Compulsory input

    Dd d = The linear width of the feature.

    Optional inputs

    Hh h = Half the total positional tolerance.

    Qq q = The amount of scan past the nominal surface position and the clearance

    for external features.

    Default: 5 mm.

  • C-axis cycles (probe vertical) 8-3

    Publication No. H-2000-6021

    Rr r = The clearance move amount in C for an external feature only (angle

    value).

    Default: 5 mm plus the ball radius converted into an angle.

    Xx x = The absolute X position at which a reading is taken when measuring a

    external feature.

    For the Ss optional input, see Chapter 2, Optional inputs.

    Example: Bore

    Bore size = 20.0

    X position = X60.0

    Z position = Z40.0

    X100.0Z20.0C0 Move to the safe position.

    G65P9610Z40.0 Move over the bore.

    G65P9610X60.0 Move into the bore.

    G65P9650D20.0 Measure the bore.

    G65P9610X100.0 Move out of the bore.

    G65P9610Z20.0 Move to the start position.

  • 8-4 C-axis cycles (probe horizontal)

    Publication No. H-2000-6021

    C-axis find (probe horizontal) O9850

    Figure 8.2 C-axis find (probe horizontal)

    Description

    This cycle will take two readings by moving the C-axis. External features can also be

    measured by moving the Z axis. The results can be used to update a work shift.

    Application

    The stylus must be programmed to a start position that is either above an external feature

    or inside an internal feature. The C-axis must be engaged before calling up the cycle.

    The inclusion of an