Generation and Measurement of Surface Plasmon Coupled Emission
Kathleen Hamilton
Defense of the Masters ThesisUniversity of New Hampshire
June 15th, 2007
![Page 2: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/2.jpg)
•SPCE signals have been measured with novel material sets
•A compact apparatus has been constructed to measure these signals
•Depending on step size, a full 180º scan can be completed in < 5 minutes
•Reproducible scans can be made with resolutions up to 0.5˚
2
Main Results
![Page 3: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/3.jpg)
Outline
•Introduction
•Sample Preparation
•Apparatus Design and Construction
•Results
•Conclusions and Future Work
3
![Page 4: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/4.jpg)
Introduction
•Fluorescence
•Surface Plasmons
•Surface Plasmon Resonance (SPR)
•Surface Plasmon Coupled Emission (SPCE)
4
![Page 5: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/5.jpg)
Fluorescence
S0
S1
hν hν′
• Process of light emission from a molecule
• Repeatable
• Emission wavelength is characteristic of the dye
• Emission is isotropic
green: excitationred: emission
black: non-radiative processes5
![Page 6: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/6.jpg)
Surface Plasmons
•Oscillations of surface electrons in a metal
•Act as a wave propagating along interface
•Decay rapidly into the metal volume
•Can be used to transfer energy through thin films
6
![Page 7: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/7.jpg)
Surface Plasmon Resonance (SPR)
•Absorption of incident energy into surface plasmons
•Prism is rotated, reflectance is measured as a function of 7
Excitation of SPR through a glass prism, incident on a silver film (grey) and dielectric overlayer (pink)
![Page 8: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/8.jpg)
•Describe reflection/transmission of light at dielectric and metal surfaces
•Boundary Conditions:
•Tangential E continuous
•Tangential B continuous
•Wave phases are equal
•Used to identify SPR angle
Fresnel Equations
8
![Page 9: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/9.jpg)
Four phase Fresnel Equations
dm
d1
prism
air
•Affected by
•Thickness of silver films
•Thickness of dielectric overlayer
•Wavelength of incident light
9
m
2
1
0
![Page 10: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/10.jpg)
Calculated SPR curves for Ag films
Films of thickness: 0 nm (blue), 50 nm (black), 20 nm (red), 80 nm (green), 100 nm (lt. blue)
curves calculated using four-phase Fresnel programs from the research group of Robert M. Corn: http://unicorn.ps.uci.edu/calculations/fresnel/fcform.html 1
0
![Page 11: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/11.jpg)
Effect of Dielectric Layer Thickness on SPR Angle
SPR curves with dielectric overlayer thicknesses of: 5 nm (blue), 10 nm (black), 20 nm (red), 50 nm (green), 100 nm (lt. blue)
curves calculated using four-phase Fresnel programs from the research group of Robert M. Corn: http://unicorn.ps.uci.edu/calculations/fresnel/fcform.html 1
1
![Page 12: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/12.jpg)
Effects of Wavelength on SPR angle
Calculated SPR curves for wavelengths: 532 nm (black), 550 nm (blue), 575 nm (red), 600 nm (green), 660 nm (lt. blue)
curves calculated using four-phase Fresnel programs from the research group of Robert M. Corn: http://unicorn.ps.uci.edu/calculations/fresnel/fcform.html 1
2
![Page 13: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/13.jpg)
SPCE•“Inverse” process to SPR
•SPCE uses surface plasmons for emission at wavelength dependent angles
•Fluorescing molecules above a metal will induce plasmons in the metal
•Plasmons will couple to photons via glass interface
13
![Page 14: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/14.jpg)
Methods of Generation
SPCE
θSPCE θSPCE
θE
Left: Reverse Kretschmann configuration (RK)Right: Kretschmann configuration (KR)
14
![Page 15: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/15.jpg)
SPCE generation with the RK configuration
•Incident light excites fluorescence
•Fluorescing molecules induce plasmons on metal surface
•Plasmons are coupled to photons at glass interface
•Resulting light is emitted at angles determined by plasmon wavenumber 1
5
![Page 16: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/16.jpg)
Outline
•Introduction
•Sample Preparation
•Apparatus Design and Construction
•Results
•Conclusions and Future Work
16
![Page 17: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/17.jpg)
Sample Preparation
Aga
b
glass
dye
a: adhesion layer b: passivation layer
•Two Components
•Thin films deposited by sputtering
•Thin fluorescent dye deposited by spin coating
17
![Page 18: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/18.jpg)
Sputter Deposition• Physical vapor deposition• Deposition under vacuum reduced impurities in films• Rotation of substrate creates films of uniform thickness• Reactive gas sputtering creates material for passivation layer1
8
![Page 19: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/19.jpg)
Thin Film StacksMaterial Material
SetSetAdhesion Adhesion
LayerLayer ThicknessThickness Ag Ag ThicknessThickness
PassivatioPassivation Layern Layer ThicknessThickness
AlSi/Ag/AlSi-N
AlSi 2 nm 34.1 nmAlSi-N
(15% N2)10 nm
AlSI/Ag/AlSi-N
AlSi 2 nm 50 nmAlSi-N
(15% N2)10 nm
Si/Ag/Si-O(x)
Si 1 nm 50 nmSi-O(x) (5% O2)
10 nm
Si/Ag/Si-O(x)
Si 1 nm 50 nmSi-O(x) (5% O2)
20 nm
Si/Ag/Si-O(x)
Si 1 nm 50 nmSi-O(x)
(10% O2)20 nm
19
![Page 20: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/20.jpg)
Spin Coating
•Small volume of dye deposited on substrate
•Rotated at high speeds (2200-3000 rpm)
•Result is thin, uniform coating
20
![Page 21: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/21.jpg)
Outline
•Introduction
•Sample Preparation
•Apparatus Design and Construction
•Results
•Conclusions and Future Work
21
![Page 22: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/22.jpg)
Apparatus Design and Construction
•Three Mechanical Components
•Excitation
•Rotation
•Detection
•Computer Interface and Control
22
![Page 23: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/23.jpg)
•Excitation with 5 mW laser ( = 532 nm)
•Driven by 2.85 V (DC)
•Beam divergence < 0.069º
23
![Page 24: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/24.jpg)
• Rotational motion is driven by a stepper motor
• Step resolution: 0.0281º
• Accuracy: <1º
• Repeatability: < 0.1º
24
![Page 25: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/25.jpg)
Fluorescence detection made by a photomultiplier
Wavelength filter
200 m aperture
PMT
amplifier
Amplifier is needed to convert PMT output from current to voltage
25
![Page 26: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/26.jpg)
•Data acquisition module (DAQ) converts analog PMT signal to digital
•DAQ signal is read by LabVIEW program
•DAQ also controls voltage applied to the PMT
Computer Control
26
![Page 27: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/27.jpg)
•Four part LabVIEW program
•PMT Voltage Specified
•Motor sent to home position
•Step size and number of steps specified
•Motor advances and DAQ signal is read on alternate iterations
•Angle and DAQ signal are written to file
27
![Page 28: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/28.jpg)
Data from LabVIEW is imported to Matlab, plotted and fitted with Gaussian curves
28
Theta (deg.)
![Page 29: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/29.jpg)
Outline
•Introduction
•Sample Preparation
•Apparatus Design and Construction
•Results
•Conclusions and Future Work
29
![Page 30: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/30.jpg)
•SPCE has been qualitatively observed from different material sets
•SPCE has been measured from different material sets
•SPCE signals have been measured with 4 different film stacks, made from 3 different material sets
•Depending on step size, a full 180º scan can be completed in < 5 minutes
Results
30
![Page 31: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/31.jpg)
SPCE Measurement
QuickTime™ and aMotion JPEG OpenDML decompressor
are needed to see this picture.
31
![Page 32: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/32.jpg)
First detected SPCE signal
32
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 33: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/33.jpg)
•Scans begin at 0˚ and traces out an angle
•Peaks are measured in the regions: < 90˚ and > 90˚
0 90
180
Geometry of an SPCE scan
![Page 34: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/34.jpg)
Repeated scans with increasing PMT voltage: 0.79 V (green), 0.81 V (blue), 0.84 V (black), 0.84
V (red)34
Sig
nal In
ten
sity
(V
)
Theta (deg.)
![Page 35: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/35.jpg)
5 repeated scans at 0.84 V
Run SPCE SPCE
01 57.8 ± 1.4
122.8 ± 0.3
02 58.3 ± 0.6
122.7 ± 0.2
03 56.6 ± 1.6
124 ± 0.9
04 58.0 ± 1.1
122.9 ± 0.8
05 57.7 ± 1.0
123.2 ± 0.3
First (red), second (blue), third (black), fourth (green), fifth (magenta)
35
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 36: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/36.jpg)
3 scans at 0.84 V
Scan SPCE SPCE
01 67.9 ± 11.9
118.7 ± 0.4
02 64.2 ± 6.2
118.7 ± 0.3
03 63.8 ± 5.3
118.3 ± 0.2
36
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 37: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/37.jpg)
•Angular Reproducibility
•High uncertainty in fits due to poor fitting of small peaks
•Variant peaks at same voltage
•Altering the voltage of consecutive scans led to uniformity
Repeated Scan Characteristics
37
![Page 38: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/38.jpg)
Tests of Si/Ag/Si-O(x) (5% O2)
Scans with PMT voltages: 0.75 V (blue), 0.79 V (green), 0.84 V (red), 0.84 V (black)
Run SPCE SPCE
0.75V 68.4 ± 0.2
127.1 ± 0.8
0.79V 69.5 ± 0.2
129.9 ± 0.6
0.83V 64.2 ± 0.2
124.9 ± 0.5
0.84V 66.2 ± 0.2
126.4 ± 0.4
38
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 39: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/39.jpg)
Tests of Si/Ag/Si-O(x) (10% O2)
Scan SPCE SPCE
0.75 V 47.4 ± 2.2 105.3 ± 5.4
0.81 V 53.3 ± 0.7 112.9 ± 2.3
0.84 V 55.2 ± 1.3 115.1 ± 2.1
PMT Voltages: 0.75 V (black), 0.81 V (red), 0.84 V (blue)
39
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 40: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/40.jpg)
Higher Resolution Scans
•Previous scans were made with step sizes of 1.0º
•Scans were made with smaller step sizes, 0.75º, 0.5º, 0.25º, 0.2º, 0.1º
•Step sizes > 0.5º show similar angular reproducibility as the 1.0º step size scans
•Step sizes < 0.5º show poor angular reproducibility
40
![Page 41: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/41.jpg)
Smaller Step Scans
Left: Step sizes of 1.0º (red), and 0.5º (black)Right: Step sizes of 0.1˚ (black), 0.15˚ (blue), and 0.2˚ (red)
41
Theta (deg.) Theta (deg.)
Sig
nal In
ten
sity
(V
)
Sig
nal In
ten
sity
(V
)
![Page 42: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/42.jpg)
Reverse Scans•Second LabVIEW program written
•Program will execute two scans in two directions
•First scan will be done in clockwise direction
•Second scan will be done in counter-clockwise direction
•Reversed direction scans were done to check angular reproducibility
42
![Page 43: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/43.jpg)
Clockwise and Counter-clockwise scans at 1.0º step size
Run SPCE SPCE
CW 67.3 ± 0.1 136 ± 0.3
CCW 68.2 ± 0.3 136 ± 0.3
Clockwise (red) and counterclockwise (black) SPCE scans with 1˚ step size.
43
Sig
nal In
ten
sity
(V
)
Theta (deg.)
![Page 44: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/44.jpg)
Clockwise and Counter-clockwise scans at 0.2º step size
Run SPCE
CW 48.2 ± 0.1
CCW 56.1 ± 0.1
Clockwise (red) and counterclockwise (black) scans with step size of 0.2˚44
Theta (deg.)
Sig
nal In
ten
sity
(V
)
![Page 45: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/45.jpg)
Outline
•Introduction
•Sample Preparation
•Apparatus Design and Construction
•Results
•Conclusions and Future Work
45
![Page 46: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/46.jpg)
Conclusions and Future Work
•SPCE can be generated with many different material sets
•Adhesion/Passivation materials may affect fluorescence intensities
•Scans with step sizes >0.5˚ are reproducible
•Fluorescent dye quality
•Symmetry of peaks46
![Page 47: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/47.jpg)
Acknowledgments • Thanks to the current and former members of Jim
Harper’s group: Amanda Brown, Don Carlson, Derya Deniz, Dana Filoti and Anne-Marie Shover
• Thanks to the current and former members of Tom Laue’s group: Brett Austin, Sue Chase, and Kari Hartmen for their help in early design of the apparatus and fluorescent dye preparation
• Thanks to Rudolf Seitz of the Chemistry Department for access to the spin coater
• Thanks to Rob Cinq-Mars of the UNH Instrumentation Center for help in LabVIEW programming and stepper motor operation
• Thanks to Ignacy and Zygmunt Gryczynski of the University of North Texas
• Thanks to the thesis committee members: Professors Tom Laue, Olof Echt and James Harper
47
![Page 48: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/48.jpg)
SupportThis thesis was financially supported in part by NIH Grant 1-R33CA14460-
01
48
![Page 49: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/49.jpg)
![Page 50: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/50.jpg)
Tests of AlSi/Ag/AlSi-N (15% N2)
Scan SPCE SPCE
0.75 V 50.8 ± 1.0 108 ± 1.8
0.79 V 48.3 ± 0.8 106.2 ± 1.9
0.81 V 54.4 ± 0.5 114.1 ± 1.1
0.83 V 50.0 ± 0.3 111.9 ± 0.9
0.85 V 54.3 ± 0.3 117.5 ± 1.7
Scans with PMT voltage: 0.75 V (red), 0.79 V (green), 0.81 V (grey),
0.84 V (blue), 0.84 V (black)50
![Page 51: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/51.jpg)
Scans of Si/Ag/Si-O(x) (5% O2), 20nm thick passivation layer
PMT voltages 0.75 V (green), 0.77 V (blue), 0.84 V (red) and 0.84 V (black).
Run SPCE SPCE
0.75 V 48.5 ± 0.5 109.4 ± 1.0
0.77 V 52.2 ± 0.4 116.9 ± 4.7
0.83 V 51.7 ± 1.4 110.5 ± 3.5
0.85 V 54.8 ± 1.4 113.7 ± 1.0
![Page 52: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/52.jpg)
Fluorescence Spectrum
plot from Molecular Probes: http://probes.invitrogen.com/servlets/spectra?fileid=6393ph8
![Page 53: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/53.jpg)
•Semi-transparent and highly reflective slides
•AlSi/Ag/AlSi-N
•Si/Ag/Si-O(x) (5% O2)
•Highly transparent and semi-reflective slides
•Si/Ag/Si-O(x) (10% O2)
53
![Page 54: Generation and Measurement of Surface Plasmon Coupled Emission Kathleen Hamilton Defense of the Masters Thesis University of New Hampshire June 15th, 2007](https://reader030.vdocuments.us/reader030/viewer/2022033108/56649d635503460f94a460d0/html5/thumbnails/54.jpg)
Theta (deg.)Theta (deg.)
Sig
nal In
ten
sity
(V
)
Sig
nal In
ten
sity
(V
)
Signal Attenuation as films erode
Left: Dye layer newly depositedRight: 18 hours later
Sample: Si/Ag/SiO(x) (5% oxygen)