geant4 accuracy and benchmarking

5
Geant4 Accuracy and Benchmarking Bruce Faddegon, UCSF

Upload: cole-freeman

Post on 31-Dec-2015

26 views

Category:

Documents


0 download

DESCRIPTION

Geant4 Accuracy and Benchmarking. Bruce Faddegon, UCSF. Geant4 e-/x-ray RT benchmarks. SLAC Makoto Asai Tsukasa Aso Joseph Perl Jane Tinslay. Thick-target bremsstrahlung measurement at 10-30 MV. Subtract pile-up spectrum and bkg Add counts lost to pulse pile-up Unfold detector response - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Geant4 Accuracy and Benchmarking

Geant4 Accuracy and Benchmarking

Bruce Faddegon, UCSF

Page 2: Geant4 Accuracy and Benchmarking

Geant4 e-/x-ray RT benchmarks

• SLAC– Makoto Asai– Tsukasa Aso– Joseph Perl– Jane Tinslay

Page 3: Geant4 Accuracy and Benchmarking

Thick-target bremsstrahlung measurement at 10-30 MV

• Subtract pile-up spectrum and bkg• Add counts lost to pulse pile-up• Unfold detector response• Add counts lost to attenuation and

detector efficiency• Collimator effect

BA Faddegon, CK Ross, DWO Rogers, Medical Physics 17(5):773-785, 1990BA Faddegon, CK Ross, DWO Rogers, Medical Physics 18(4):727-739, 1991

Page 4: Geant4 Accuracy and Benchmarking
Page 5: Geant4 Accuracy and Benchmarking

No Bremsstrahlung SplittingGeant 4.7.1/EGSnrc

0

0.5

1

1.5

2

2.5

3

3.5

4

Be Al Pb

Relative Simulation Time

Bremsstrahlung Splitting x 100Geant 4.8.0/EGSnrc

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

Be Al Pb

Relative Simulation Time