ge sensing & inspection technologies ultrasonic dm5e · this operating manual provides...
TRANSCRIPT
GESensing & Ultrasonic
P/N 1259551 Rev. 1February 2009
DM5Operatin
Inspection Technologies
Eg Manual
GESensingInspection.com
©2009 General Electric Company. All rights reserved.Technical content subject to change without notice.
DCo
Ope
P/N Febr
M5Errosion Thickness Gauge
rating Manual
1259551 Rev. 1uary 2009
to next page]
[no content intended for this page - proceedContents
DM5E Ope iii
Chapter
1.1 Supp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 Powe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.3 Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.4 Cont . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter
2.1 Keyp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.2 Inter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.3 Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
rating Manual
1. General Information
lying Power to the DM5E. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ring the Instrument ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Features of the DM5E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .DM5E Corrosion Thickness Measurement Base Instrument . . . . . . . .
ents of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Understanding the Keypad, Menu System & Displays
ad Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .preting the Display Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ing with the Configuration Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents
iv DM5E Operating Manual
Chapter
3.1 Conn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183.2 Conf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233.2.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 243.2.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3 Instr . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263.4 Setti . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303.5 B-SC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 313.6 Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Chapter
4.1 Selec . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 364.2 Norm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 394.3 Min . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404.4 Diffe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 414.5 B-SC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 424.6 Dual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3. Setting Up the DM5E
ecting a Probe and Loading a Setup File . . . . . . . . . . . . . . . . . . . . . . . . .iguring the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting the Instrument Gain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Setting the Update Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Specifying the Nominal Thickness . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ument Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ng the Maximum and Minimum Alarms . . . . . . . . . . . . . . . . . . . . . . . . . .AN Display Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ing and Unlocking Instrument the Controls . . . . . . . . . . . . . . . . . . . . . . .
4. Measuring Thickness
ting the Displayed View Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .al Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Scan and Max Scan Measurement View Modes . . . . . . . . . . . . . . . . . . . .rential/Rate-of-Reduction Measurement View Mode . . . . . . . . . . . . . . .AN Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Multi Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents
DM5E Ope v
Chapter
5.1 Crea . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 465.2 Reca . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 495.2.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3 Reco . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 505.3.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 515.3.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Chapter
6.1 Conn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Chapter
7.1 Instr . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 567.2 Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 607.3 DM5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
rating Manual
5. Using the Optional Data Recorder
ting a New Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .lling and Erasing Stored Data Recorder Files. . . . . . . . . . . . . . . . . . . . . .
Recalling a Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Erasing a Data Recorder File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
rding Thickness Measurements in Data Recorder Files . . . . . . . . . . . . . .Storing Minimum of B-Scan in Data Recorder Files . . . . . . . . . . . . . .Navigating Through Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . .
6. I/O Features
ecting the DM5E to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7. Specifications
ument Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recorder Option Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E Probe/Transducer Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents
vi DM5E Operating Manual
Chapter
8.1 Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 638.2 Prop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 638.3 Batte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Appendi
A.1 Rese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66A.2 Firm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67A.3 Web . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69A.4 EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70A.5 Man . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Appendi
B.1 Wast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74B.2 Batte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
B.2.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75B.2.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8. Maintenance
of the instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .er Cable Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
x A. Supplemental Information
tting the Operating Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ware Updates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Site Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ufacturing & Customer Service Locations . . . . . . . . . . . . . . . . . . . . . . . .
x B. Environmental Compliance
e Electrical and Electronic Equipment (WEEE) Directive . . . . . . . . . . . .ry Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
What do the Markings Mean? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .The Risks and Your Role in Reducing Them . . . . . . . . . . . . . . . . . . . .
DM5E Ope vii
Preface
Importa
The follo ng & Inspection Technologies ultrasonic thickness gauge. Failure to r other test results. Decisions based on erroneous results can, in tu
Genera
Proper us
• Selec
• Know
• Traini
This oper thickness gauge. There are, however, additional factors which aff hese additional factors is beyond the scope of this manual. The opera e detailed information.
rating Manual
nt Notice
wing information must be read and understood by any user of a GE Sensi follow these instructions can lead to errors in thickness measurements orn, lead to property damage, personal injury or death.
l Warnings
e of ultrasonic test equipment requires three essential elements:
tion of the correct test equipment
ledge of the specific “test application requirements”
ng on the part of the instrument operator
ating manual provides instruction in the basic setup and operation of theect the use of ultrasonic test equipment. Specific information regarding ttor should refer to textbooks on the subject of ultrasonic testing for mor
Preface
viii DM5E Operating Manual
Operat
Operators Operators must be trained in general ultrasonic testing procedure rs must understand:
• Sound
• Effect
• Behav
• Areas
More spe test specifications is available from various technical societies,
or Training
must receive adequate training before using ultrasonic test equipment. s and in the setup and performance required by a particular test. Operato
wave propagation theory
s of the velocity of sound of the test material
ior of the sound wave where two different materials are in contact
covered by the sound beam
cific information about operator training, qualification, certification, and industry groups, and government agencies.
DM5E Ope ix
Preface
Testing
In ultraso nd beam. Operators must exercise great caution in making inference le, when testing large materials it may be impossible or impractic
When a l own the specific areas to inspect. Inferences about the condition nly be attempted by personnel fully trained in applicable statistical corrosion, in which conditions can vary significantly in any given rs.
Sound be eometry and overlapped flaws or overlapped surfaces, thickness ack wall of the material. Operators must take steps to ensure th
rating Manual
Limitations
nic testing, information is obtained only from within the limits of the sous about the test material outside the limits of the sound beam. For exampal to inspect the entire test piece.
ess-than-complete inspection is to be performed, the operator must be sh of areas not inspected, based on data from the evaluated areas, should o and probability techniques. In particular, materials subject to erosion or area, should only be evaluated by fully trained and experienced operato
ams reflect from the first interior surface encountered. Because of part g gauges may measure the distance to an internal flaw rather than to the bat the entire thickness of the test material is being examined.
Preface
x DM5E Operating Manual
Ultraso res
The follo ickness gauges in order to minimize errors in test results.
1. Calib
The princ measures the time of flight of an ultrasonic pulse through the test p ickness measuring error is minimized by ensuring that the sound vel aterial being tested. Actual sound velocities in materials often vary st results are obtained if the instrument is calibrated on a velocity r should be flat and smooth and as thick as the maximum thickness
Operators material being tested; heat treating, for example, can cause sig ting the accuracy of the thickness provided by this instrumen ation should be checked after testing, to minimize testing errors.
2. Prob
Before pe t be performed first. If the zero procedure is not performed or if it is
nic Thickness Measurement Critical Operating Procedu
wing operating procedures must be observed by all users of ultrasonic th
ration of Sound Velocity
iple of operation of an ultrasonic thickness gauge is that the instrument iece and multiplies this time by the velocity of sound in the material. Thocity to which the instrument is calibrated is the sound velocity of the m significantly from the values found in published tables. In all cases, be
eference block made from the same material as the test piece; this block of the test piece.
should also be aware that the sound velocity may not be constant in thenificant changes in sound velocity. This must be considered when evaluat. Instruments should always be calibrated before testing, and the calibr
e Zero Procedure
rforming calibration with a contact probe, the probe zero procedure musperformed improperly, inaccurate thickness readings will result.
DM5E Ope xi
Preface
Ultraso res (cont.)
3. Effec
Temperat y lines and, therefore, calibrations. All calibrations should be perfor e, to minimize errors due to temperature variations.
4. Tran
The trans r of the front surface. Badly worn transducers will have a reduced e lude the complete range of thicknesses to be tested. The temperatu re range.
5. Use
Operators st be developed so that couplant is used and applied in a consisten test results. Calibration and actual testing should be performe nt and applying consistent pressure on the transducer.
rating Manual
nic Thickness Measurement Critical Operating Procedu
ts of Temperature on Calibration
ure variations change the sound velocity of materials and transducer delamed on-site and with test blocks at the same temperature as the test piec
sducer Selection
ducer used in testing must be in good condition, without noticeable weaffective measuring range. The specified range of the transducer must incre of the material to be tested must be within the transducer's temperatu
of Couplants
must be familiar with the use of ultrasonic couplants. Testing skills mut manner to minimize variations in couplant layer thickness and errors ind under similar coupling conditions, using a minimum amount of coupla
Preface
xii DM5E Operating Manual
Ultraso res (cont.)
6. Dou
Ultrasoni are twice (or in some cases, three times) the actual material t can occur below the minimum specified range of the transduce s greater than the minimum of the specified range.
When usi specified range of the transducer may be a “doubled” reading, a se of other methods. If the transducer shows any sign of wear, dou y produce a readable signal. The instrument reading and apparent greater than twice the minimum of the specified range. This thick bination on reference blocks that represent the complete range of p rly important when a test piece is being ultrasonically measured t specimen is unknown.
nic Thickness Measurement Critical Operating Procedu
bling
c thickness gauges will, under certain conditions, display readings whichhickness being measured. This effect, commonly known as “doubling,” r. If the transducer being used is worn, doubling is possible at a thicknes
ng a new transducer, any reading which is less than twice the minimum nd the thickness of the material being tested should be verified by the ubling may occur at a second echo, or other echo signal combinations mathickness are up to about twice the actual value, resulting in a thickness ness should be determined by calibrating the instrument/transducer comossible thicknesses that may be encountered in testing. This is particula for the first time or in any case where the history of thickness of the tes
DM5E Ope xiii
Preface
Safety I
IMPORTA cal applications or other purposes is not allowed. The
IMPORTA eded. You should only use the products recommended by y recommended.
IMPORTA one in compliance with all applicable regulations. See
Softwa
Accordin rors. Before using any software-controlled test equipment, please ma bination.
rating Manual
nformation
NT: The DM5E is an instrument for materials testing. Any use for mediDM5E may only be used in industrial environments.
NT: For battery operation of the DM5E, two “AA” size batteries are neus for battery operation. Use of well-known battery brands is highl
NT: Disposal of all components of the DM5E and its batteries must be dAppendix B, Environmetal Compliance, for detailed information.
re
g to the current state of the art, software is never completely free from erke sure that the required functions operate perfectly in the intended com
Preface
xiv DM5E Operating Manual
Defects
If you ha ble, you must disconnect the instrument and secure it against un
Safe oper
• The in
• The in
• The in ons such as exceptional temperatures and/or especially high a
• The in
Service
Every eff ld service become necessary, GE Sensing & Inspection Technolo he location of the nearest facility, see the back cover of this manu
/Errors and Exceptional Stresses
ve reason to believe that safe operation of your DM5E is no longer possiintentional reconnection. Remove the batteries if necessary.
ation is no longer possible if conditions such as the following exist:
strument shows visible damages
strument no longer operates perfectly
strument has been subjected to prolonged storage under adverse conditiir humidity, or corrosive environmental conditions
strument has been subjected to heavy stresses during transportation
ort has been made to provide you with a reliable product. However, shougies, has established a number of Factory Trained Service Centers. For tal.
Chapter 1. General Information
DM5E Ope 1
Chapt
The DM5 ent is available in three versions:
• DM5E
• DM5E
• DM5E
This chap ins the keypad functions and display-menu contents. Closely r e of the more detailed information found in subsequent chapters o
1.1 Su
The instru ately 100 hours of service life. To install the batteries, remove th
IMPORTA
rating Manual
er 1. General Information
E is an ultrasonic corrosion thickness measurement device. The instrum
Basic
with Dual Multi functionality
DL with Dual Multi and Data Recorder
ter introduces the key features of the DM5E, and the next chapter explaeviewing the material in these two chapters will help you make better usf this manual.
pplying Power to the DM5E
ment is powered by two “AA” size batteries, providing up to approxime battery-compartment cover as shown in Figure 1 on page 2.
NT: You cannot charge or power the DM5E through its USB port.
Chapter 1
2 DM5E Operating Manual
1.1 Su
with battery cover removed)
R
I/O Device Connection(mini USB)
Probe Cable Connection
Install 2in comp
. General Information
pplying Power to the DM5E (cont.)
Figure 1: DM5E Connections and Batteries (shown
ear of Instrument
“AA” batteriesartment as shown
Top of Instrument
Chapter 1. General Information
DM5E Ope 3
1.1 Su
When the symbol , replace the batteries as soon as possible. The DM5E au tion. However, your settings are saved and restored when the instru spare batteries.
1.2 Po
Power the y activates. Press and hold while the instrument is ON to tur
1.3 Ke
• Large
• Alarm
• Stand
• Suppo
• Alpha
rating Manual
pplying Power to the DM5E (cont.)
battery indicator is in the last quarter, as indicated by the display-screentomatically shuts OFF when the batteries are too weak for reliable operament is turned ON again. When testing in remote locations, always carry
wering the Instrument ON and OFF
instrument ON by pressing and holding until the instrument displan it OFF.
y Features of the DM5E
hollow/filled thickness digits
bar graph
ard and custom parameter setups
rts new set of DA5xx probes
numeric naming of data files (DM5E DL)
Chapter 1
4 DM5E Operating Manual
1.3 Ke
• Locko
• Large
• Impor
• Appro
• Easy-
• Multi
• User-
• Norm easurement modes are standard
• Timed
• Optio
• Optio ts in data recorder file
• Field
• Comp
. General Information
y Features of the DM5E (cont.)
ut feature with password access
LCD display with backlight and adjustable contrast
ting of data recorder files directly into Microsoft Excel (DM5E DL)
ximately 100 hours of operation with 2 “AA alkaline batteries
to-use, single-level menu system
ple display screen languages
selectable measurement resolution to 0.001 inch (0.01 mm)
al (thickness), MinScan, MaxScan, and Differential/Rate of Reduction M
B-Scan
nal Dual Multi function (DM5E and DM5E DL)
nal data recorder (DM5E DL) stores up to 50,000 thickness measuremen
upgrades available
atible with UltraMATE© and UltraMATE LITE© software programs
Chapter 1. General Information
DM5E Ope 5
1.3.1 D nt
1.3.1a• DM5E
• 2 “AA
• Plasti
• Coup
• Upgra
• Opera
• Opera
• Certif
Note: A cable can be purchased separately for use with theDM
rating Manual
M5E Corrosion Thickness Measurement Base Instrume
Contents of the DM5E Base Instrument instrument
” batteries
c carrying case
lant sample
de Utility Program on CD-ROM (requires mini-USB to PC cable)
ting manual in CD-ROM
ting instruction card
icate of conformity
mini-USB to PC cable is only included with the DM5EDL version. This 5E Basic or DM5E version.
Chapter 1
6 DM5E Operating Manual
1.3.1b• DM5E
• DM5E
• DM5E
1.4 Co
This DM ll chapters apply to all instrument versions, except for two functions
• Data version
• Dual to the DM5E DL
Note: An
. General Information
Instrument Versions Basic
with Dual Multi functionality
DL with Dual Multi and Data Recorder
ntents of this Manual
5E Operating Manual is divided into eight chapters and one appendix. A:
Recorder, which is described in Chapter 5, applies only to the DM5E DL
Multi mode applies only to the DM5E with Dual Multi functionality and
y DM5E can be retrofitted to incorporate the above functions.
hapter 2. Understanding the Keypad, Menu System & Displays
DM5E Ope 7
Chapt & Displays
The DM5 use. In this chapter, you'll find a brief explanation of all display an more detailed information is available.
Note: Th ures installed and the operational settings selected. The di nstrument configurations.
2.1 Ke
The instru ction keys. The key functions, as illustrated in Figure 2 on page 8, ar
• Arrow een navigation.
• Funct irectly above the key.
• Cal/O n selections. Press and hold to power the instrumentON an
• Mode play modes that are installed.
C
rating Manual
er 2. Understanding the Keypad, Menu System
E's display, keypad, and functional commands are easy to interpret and d keypad features. You'll also find references to manual sections where
e DM5E's display screen contents varies depending on the optional featsplay screens shown in this chapter are typical of those found in all the i
ypad Features
ment's keypad includes a combination of dedicated keys and virtual fune as follows:
Keys - Use to adjust the value of the selected parameter and for on-scr
ion Keys - Use to activate any functional control shown in the display d
n Key - Press once to launch the calibration process or to make on screed OFF.
Key - Press to activate the standard display modes and any optional dis
Chapter 2
8 DM5E Operating Manual
2.1 Ke
Functions
S
Sele
Thickn
Press to activate the functional control(when present) above the key.
Press to launch the calibration processor make on-screen selections. Press andhold to power the instrument ON or OFF.
Press to adjust the value of the selectedparameter and for on-screen navigation.
Press to activate standard display modes(shown here) and the optional displaymodes (if installed). Optional display modesinclude Data Recorder Contents.
. Understanding the Keypad, Menu System & Displays
ypad Features (cont.)
Figure 2: DM5E Keypad and Key
elect a probe setup
ct, adjust and confirmoperating settings
ess measurement display
hapter 2. Understanding the Keypad, Menu System & Displays
DM5E Ope 9
2.2 Int
This secti :
• Meas icons, displays data recorder location and file name (when this o ntative display screens. The display's appearance varies based . Additional measurement view modes are available when the D
• Probe nstrument setup, each of which is matched to a specific probe
• File D data recorder files. This feature is only available when the data r n installed and activated, the data recorder option allows users gh file contents.
• Confi screen. The controls listed on the configuration display screen contents of the Configuration display depends on the instru
C
rating Manual
erpreting the Display Screens
on describes the layout of the DM5E's primary display modes including
urement Display Mode - Indicates measured thickness, contains status ption is installed and activated). Figure 3 on page 10 shows four represe on the installed instrument options as well as instrument display settingsual Multi or Data Recorder options are installed.
Setup Display Mode - Allows the user to select a standard preloaded i model (see Figure 4 on page 11).
isplay Mode - Allows the user to create and store thickness readings in ecorder option is installed and activated (see Figure 5 on page 12). Wheto create data recorder files, store thickness readings, and navigate throu
guration Display Mode - Instrument controls are accessed through this vary depending on the installed options (see Figure 6 on page 13). The
ment's installed-option configuration.
Chapter 2
10 DM5E Operating Manual
2.2 In
y Mode
. Understanding the Keypad, Menu System & Displays
terpreting the Display Screens (cont.)
Figure 3: Measurement Displa
hapter 2. Understanding the Keypad, Menu System & Displays
DM5E Ope 11
2.2 In
y Mode
C
rating Manual
terpreting the Display Screens (cont.)
Figure 4: Probe Setup Displa
Chapter 2
12 DM5E Operating Manual
2.2 In
de
. Understanding the Keypad, Menu System & Displays
terpreting the Display Screens (cont.)
Figure 5: File Display Mo
hapter 2. Understanding the Keypad, Menu System & Displays
DM5E Ope 13
2.2 In
y Mode
C
rating Manual
terpreting the Display Screens (cont.)
Figure 6: Configuration Displa
Chapter 2
14 DM5E Operating Manual
2.3 W
By repeat n adjust the following DM5E settings:
• VIEW tion 4.1 on page 36 to select the view to be displayed.
• DATA arameter to ON activates the file recording capability and the da
• MINIM .4 on page 30).
• MAXI 3.4 on page 30).
• UNITS .2 on page 21).
• NOM calculate and the differential and rate of reduction measu n 3.2.3 on page 25).
• RESO ss measurement that is displayed and stored in a Data Recor
• UPDA e Section 3.2.2 on page 24).
. Understanding the Keypad, Menu System & Displays
orking with the Configuration Display
edly pressing , you can access the Configuration display and you ca
MODE - In base-model units this parameter offers five settings. See Sec
RECORDER - When the optional data recorder is installed, setting this pta recorder file display.
UM ALARM - Enables and sets the minimum alarm value (see Section 3
MUM ALARM - Enables and sets the maximum alarm value (see Section
- Sets the units of measurement to inches or millimeters (see Section 3
INAL THICKNESS - Sets the nominal thickness value that is then used to rement values when selected by the VIEW MODE parameter (see Sectio
LUTION - Controls the format (number of decimal places) of the thickneder file.
TE RATE - Sets the rate at which the measurement reading is updated (se
hapter 2. Understanding the Keypad, Menu System & Displays
DM5E Ope 15
2.3 W
• BACK O switches backlighting ON whenever a key is pressed or the pr
• CONT ay contrast level.
• LANG 1).
• BSCA is processing (see Section 3.2.1 on page 23).
• BSCA is processing (see Section 3.2.1 on page 23).
• DUAL ual Multi option is allowed (see Section 3.2 on page 21).
• READ be is uncoupled in measurement mode (see Section 3.2 on page
C
rating Manual
orking with the Configuration Display (cont.)
LIGHT - Turns the display backlight feature to ON, OFF, or AUTO (AUTobe is connected).
RAST - Select this control and then press or to adjust the displ
UAGE - Sets the instrument display language (see Section 3.2 on page 2
N MINIMUM - Sets the minimum value that can be drawn when BSCAN
N MAXIMUM - Sets the maximum value that can be drawn when BSCAN
MULTI - Used to switch between Dual Multi and IP test modes, when D
ING HOLD - Can be set to hold the last measurement value when the pro21).
Chapter 2
16 DM5E Operating Manual
2.3 W
• TIME
• DATE
• GAIN T) (see Section 3.2.1 on page 23).
• RADIX ction 3.2 on page 21).
• AUTO y powering the instrument OFF if no key presses or
measu ument is only powered OFF when is pressed and held.
. Understanding the Keypad, Menu System & Displays
orking with the Configuration Display (cont.)
- Used to set the current time (see Section 3.2 on page 21).
- Used to set the current date (see Section 3.2 on page 21).
- Used to select a different level of gain (LOW/HIGH/AUTO/DEFAUL
- Selects a period (.) or comma (,) to be used as a decimal point (see Se
POWER DOWN - When set to ON, this control conserves battery life b
rements occur for the set number of minutes. When set to OFF, the instr
Chapter 3. Setting Up the DM5E
DM5E Ope 17
Chapt
Prior to m explains the steps that must be taken prior to measurem urement mode match the desired configuration. The steps described y incorporate the Dual Multi or Data Recorder options.
Most fun ased on the Setup file selected by the user. Each instrumen robe is the first step in setting up the instrument.
In prepar
• Instal n 3.1 on page 18).
• Set up
• Set th
• Calib
• Set m
• Lock
rating Manual
er 3. Setting Up the DM5E
easuring thickness, the instrument must be properly setup. This chapterent, to ensure that the display appearance, ultrasonic settings, and meas
in this chapter apply to all instruments, regardless of whether or not the
damental instrument parameters are automatically set to a default value bt is supplied with standard Setup files, and naming these to match the p
ation for measuring thickness, this chapter explains how to:
l a probe and configure the instrument by loading a Setup file (see Sectio
the instrument display appearance (see Section 3.2 on page 21).
e instrument's Gain and Update Rate (see Section 3.2 on page 21).
rate the instrument/probe combination (see Section 3.3 on page 26).
aximum and minimum alarm values (see Section 3.4 on page 30).
and unlock the instrument controls (see Section 3.6 on page 32).
Chapter 3
18 DM5E Operating Manual
3.1 Co
Prior to m re 7 below) and select a Setup file that is compatible with the probe
ment
robe Cableonnection
. Setting Up the DM5E
nnecting a Probe and Loading a Setup File
easuring thickness, you must connect a probe to the instrument (see Figu.
Figure 7: Top View of Instru
PC
Chapter 3. Setting Up the DM5E
DM5E Ope 19
3.1 Co
After a pr obe Setup display, which is shown in Figure 8 below, allows th
File
rating Manual
nnecting a Probe and Loading a Setup File (cont.)
obe is connected, press to activate the Probe Setup display. The Pre user to select one of the standard Setup files.
Figure 8: Selecting a Setup
Chapter 3
20 DM5E Operating Manual
3.1 Co
The follo is activated:
• View
• Gain
• Measu
• Meas
• Nomi
• Minim
• Maxim
. Setting Up the DM5E
nnecting a Probe and Loading a Setup File (cont.)
wing instrument settings are automatically adjusted when the Setup file
Mode
red Value display resolution
urement Mode
nal Thickness - when Differential, %RR mode are selected
um Alarm Setting
um Alarm Setting
Chapter 3. Setting Up the DM5E
DM5E Ope 21
3.2 Co
Before us ed at the Configuration display (see Figure 9) as follows:
1. PressConf
2. Press
3. Afterprog
4. Press
5. Repe
Afterretur
Figure 9: The Configuration Menu
rating Manual
nfiguring the Instrument
ing the DM5E, the parameters listed on the next page must be programm
as many times as necessary to access the iguration display.
or to select the desired parameter to modify.
the desired parameter is highlighted, press to ram that parameter.
or to adjust the parameter value, then press
to confirm the new setting.
at steps 2-4 until all remaining parameters have been set.
all of the settings have been configured, press to n to the measurement display.
Chapter 3
22 DM5E Operating Manual
3.2 Co
Using the meters:
• Langu
• Units
• Resolthickn
• Upda4 or 8
• Radixpoint
• Dual
• Readi
• Time
• Date -
N MINIMUM - BSCAN display of minimum value
N MAXIMUM - BSCAN display of maximum value
Power Down - Select ON and then choose the desired time omatically turn the instrument OFF 5/10/15/30 minutes he last key press (no data will be lost) or select OFF to the instrument to remain ON until manually powered
ast - Adjusts display contrast
ight - Adjusts the display backlighting. A setting of ON ses battery usage, a setting of AUTO causes the
ighting to illuminate and remain on for five seconds after ey press.
. Setting Up the DM5E
nfiguring the Instrument (cont.)
procedure described on the previous page, configure the following para
age - Set the language
- Set the units of measurement to inches or millimeters
ution - Sets the number of decimal places to which ess measurements are displayed
te Rate - Update the displayed measurement at a rate of Hz
- Select a period (.) or comma (,) to be used as a decimal
Multi - Select ON or OFF
ng Hold - Select HOLD or BLANK
- Set time in HH:MM format
Set date in Month-Date-Year format
• BSCA
• BSCA
• Auto to autafter tallowOFF.
• Contr
• Backlincreabackleach k
Chapter 3. Setting Up the DM5E
DM5E Ope 23
3.2.1
The instru ucing LOW setting. To select the instrument gain setting:
1. Press
2. Selec
3. Press value and use or to select the cursor position.
4. Press
rating Manual
Setting the Instrument Gain
ment gain level may set to AUTO, DEFAULT, HIGH, or to a noise-red
.
t the GAIN control.
to activate the control, then use or to adjust the selected
after the adjustment is complete.
Chapter 3
24 DM5E Operating Manual
3.2.2
Measurem the measurement update rate:
1. Press
2. Selec
3. Press
4. Press
. Setting Up the DM5E
Setting the Update Rate
ents are updated at a rate of 4 or 8 Hz, as selected by the user. To select
.
t the UPDATE RATE control.
to activate the control, then use or to adjust the setting.
after the adjustment is complete.
Chapter 3. Setting Up the DM5E
DM5E Ope 25
3.2.3
When the two differential values (in the instrument’s units and as a percentag e Section 4.4). To manually input or modify the nominal thickness
1. Press2. Selec
3. Press place value and or to select a different value.
4. Press
rating Manual
Specifying the Nominal Thickness
VIEW parameter is set to DIFF/RR%, the displayed thickness includese) that represent the variation from user specified nominal thickness (se:
.t the NOM THICKNESS control.
to activate the control, then use or to adjust the selected
after the adjustment is complete.
Chapter 3
26 DM5E Operating Manual
3.3 Ins
Before us ed using one or more standards of known thickness. It is importan hing the calibration process. Then, the user can select either 1pt he calibration, follow the instructions on your instrument's display sc t mode) or Figure 12 on page 29 (Auto mode).
Note: Pr he instrument should be calibrated if there is a change in th s changed, after performing any parameter adjustment, or at
. Setting Up the DM5E
trument Calibration
ing the DM5E, the instrument and the connected probe must be calibratt that the Setup file corresponding to the probe be selected prior to launc, 2pt, or AUTO calibration mode in the LOCKOUT menu. To complete treen and in Figure 10 on page 27 (1pt mode), Figure 11 on page 28 (2p
oper calibration is essential for accurate operation of the instrument. Te type or temperature of the material being tested, if the attached probe i intervals specified in your test procedures.
Chapter 3. Setting Up the DM5E
DM5E Ope 27
3.3 In
edure (1pt Mode)
rating Manual
strument Calibration (cont.)
Figure 10: Instrument Calibration Proc
Chapter 3
28 DM5E Operating Manual
3.3 In
edure (2pt Mode)
. Setting Up the DM5E
strument Calibration (cont.)
Figure 11: Instrument Calibration Proc
Chapter 3. Setting Up the DM5E
DM5E Ope 29
3.3 In
dure (Auto Mode)
rating Manual
strument Calibration (cont.)
Figure 12: Instrument Calibration Proce
Chapter 3
30 DM5E Operating Manual
3.4 Se
When the alarm setting, the measurement value alternates between solid and ears in B-SCAN mode, the measurement value flashes ON and OFF
To input a
1. Press
2. Press RM.
3. Whe
4. Press press or to adjust the value of the selected
posit ously press or .
5. Press
. Setting Up the DM5E
tting the Maximum and Minimum Alarms
thickness measurement value exceeds either the minimum or maximum hollow font type displays, except in B-SCAN mode. When an alarm appin 1/2 second intervals.
minimum or maximum alarm thickness:
to access the Configuration display (see Figure 9 on page 21).
or to select either MINIMUM ALARM or MAXIMUM ALA
n the desired parameter is highlighted, press to activate it.
or to change the selected position in alarm thickness and then
ion. To switch the alarm setting to OFF from any other value, simultane
after the alarm thickness has been set to the desired value.
Chapter 3. Setting Up the DM5E
DM5E Ope 31
3.5 B-
The defau the range according to the target, because setting an appropria the steps below.
Note: If s equal to the range limit.
1. Press
2. Press M.
3. Whe
4. Press press or to adjust the value of the selected
posit ously press or .
5. Press
rating Manual
SCAN Display Range
lt BSCAN display range is 0-20 inch. We highly recommend adjusting te range will display a clearer scan graph.To adjust the range, complete
the BSCAN value exceeds the preset range, that point will be graphed a
to access the Configuration display (see Figure 9 on page 21).
or to select either BSCAN MINIMUM or BSCAN MAXIMU
n the desired parameter is highlighted, press to activate it.
or to change the selected position in alarm thickness and then
ion. To switch the alarm setting to OFF from any other value, simultane
after the alarm thickness has been set to the desired value.
Chapter 3
32 DM5E Operating Manual
3.6 Lo
The DM5 play, which is only accessible from the measurement screen. F lockout (or enable) some or all of the function keys.
Note: Th re only available with advanced instrument versions. H se key.
In the Loc (CAL/ZERO MODE) by selecting 1pt, 2pt or AUTO. One of the fol
• REMI me from the last thickness calibration. The setting indica
• CAL R the SEND function the number of times indicated by the settin
• CAL P trument is powered ON when the value is set to ON.
. Setting Up the DM5E
cking and Unlocking Instrument the Controls
E allows operators to lockout specific functions by using the lockout disollow the instructions in Figure 13 on page 33 to access this display and
e DATA RECORDER and DUAL MULTI functions are optional and aowever, an upgrade of the basic version is possible by purchasing a licen
kout menu, the operator can set the probe zeroing and calibration mode lowing calibration reminder styles can also be selected:
ND TIME - The calibration reminder is displayed after a preset elapsed tites the preset time value.
EADINGS - The calibration reminder is displayed after the user presses g value.
OWER ON - The calibration reminder is displayed each time that the ins
Chapter 3. Setting Up the DM5E
DM5E Ope 33
3.6 Lo
Function Keys
rating Manual
cking and Unlocking Instrument the Controls (cont.)
Figure 13: Locking and Unlocking the
Chapter 3
34 DM5E Operating Manual
eed to next page]
. Setting Up the DM5E
[no content intended for this page - proc
Chapter 4. Measuring Thickness
DM5E Ope 35
Chapt
After sett can be made by selecting the type of measurement made and the w iguring and working with the BSCAN mode. The specific topics inc
• Selec
• Work
• Use th e 40)
• Interp .4 on page 41)
• Work
• Use th
Note: Lo e adjustments described in this chapter. Loading a Setup fil default value stored in the file.
rating Manual
er 4. Measuring Thickness
ing up the instrument as described in Chapter 3, additional adjustments ay in which it is displayed. You will also find information related to conflude how to:
t the measured value to view (see Section 4.1 on page 36)
with Normal measurement mode (see Section 4.2 on page 39)
e Min Scan and Max Scan measurement modes (see Section 4.3 on pag
ret the Differential/Rate-of-Reduction measurement mode (see Section 4
with the B-Scan measurement mode (see Section 4.5 on page 42)
e Dual Multi measurement mode (see Section 4.6 on page 43)
ad a Setup file corresponding to the connected probe prior to making the after these adjustments are made will force certain settings back to the
Chapter 4
36 DM5E Operating Manual
4.1 Se
Base mod isplay modes are available depending on the installed options. T
• NORM e center of the display.
• MIN S luate material thickness, provided the probe remains coupl the minimum material thickness is displayed. During the evalu imum observed thickness values and the calculated differ gins to fill as soon as the probe is uncoupled. Recoupling the pr he same evaluation period. After the evaluation period is comp ss observed during the evaluation period is displayed.
• MAX S luate material thickness, provided the probe remains coupl the maximum material thickness is displayed. During the evalu m observed thickness values and the calculated difference betwe l as soon as the probe is uncoupled. Recoupling the probe before valuation period. After the evaluation period is complete, as ind d during the evaluation period is displayed.
. Measuring Thickness
lecting the Displayed View Mode
el DM5E instruments offer four display modes. In addition, two more dhese display modes include:
AL (Thickness Only) - The thickness value appears as large digits in th
CAN (Minimum Thickness Scan) - Allows the user to continuously evaed or is uncoupled only briefly. After the evaluation period is complete, ation period, the thickness is displayed along with the minimum and maxence between the two. The display also includes a time-out bar, which beobe before the time-out period expires allows the user to continue with tlete, as indicated by the filling of the time-out bar, the minimum thickne
CAN (Maximum Thickness Scan) - Allows the user to continuously evaed or is uncoupled only briefly. After the evaluation period is complete, ation period, thickness is displayed along with the minimum and maximuen the two. The display also includes a time-out bar, which begins to fil the time-out period expires allows the user to continue with the same eicated by the filling of the time-out bar, the maximum thickness observe
Chapter 4. Measuring Thickness
DM5E Ope 37
4.1 Se
Note: D couplant may cause erroneous results. To eliminate the
ef under HOME. This restarts the scanning session w
• DIFF/ r-specified nominal thicknesses appear along the top of the di ominal values, expressed in percentage and dimensional terms
• B-SCA ss by graphing 1 point per second. During the evaluation period d maximum observed thickness values. The display has a hidde Recoupling the probe before the time-out period expires allow D key stores the minimum thickness value after the evalu
Note: Th Section 4.4 on page 41).
rating Manual
lecting the Displayed View Mode (cont.)
uring the placement of a very sensitive probe on the part surface, excess
fect of excessive couplant, couple the probe firmly to the part, then presshile the probe is properly in contact with the part.
RR% (Differential/Rate-of-Reduction) - The currently measured and usesplay, while the difference between these two values (measured minus n) appears in the middle of the display.
N (B Scan) - Allows the user to continuously evaluate material thickne, the measured thickness value is displayed along with the minimum an
n 5-second time-out bar, which starts as soon as the probe is uncoupled. s the user to continue with the same evaluation period. Pressing the SENation period is complete, as indicated by the filling of the time-out bar.
e displayed differences can have either positive or negative values (see
Chapter 4
38 DM5E Operating Manual
4.1 Se
The meas
1. Press
2. Press
3. Press
4. Press lect the highlighted mode.
. Measuring Thickness
lecting the Displayed View Mode (cont.)
urement mode displayed by the instrument may be modified as follows:
as many times as necessary to access the Configuration menu.
or to select the VIEW MODE parameter.
to activate this parameter.
or to scroll through the available modes, then press to se
Chapter 4. Measuring Thickness
DM5E Ope 39
4.2 No
When NO , as shown in Figure 14 below.
Note: If e displayed at the top left corner of the display, instead of th
View Mode
rating Manual
rmal Measurement View Mode
RMAL view mode is active, the display shows only a thickness reading
the Data Recorder file is active, the data location and stored value will be material velocity value.
Figure 14: NORMAL Measurement
Chapter 4
40 DM5E Operating Manual
4.3 Mi
The MIN material thickness then, after the evaluation period is complete ed. During the evaluation period, thickness is displayed along wit below). A time-out feature allows a momentarily uncouplin ent will continue to collect and compare readings for as long as th
t View Mode
. Measuring Thickness
n Scan and Max Scan Measurement View Modes
SCAN and MAX SCAN modes allow the user to continuously evaluate, display the extreme (minimum or maximum) material thickness observh the minimum and maximum observed thickness values (see Figure 15g of the probe during the evaluation period. In these modes, the instrume time-out period is not exceeded.
Figure 15: MAX SCAN Measuremen
Chapter 4. Measuring Thickness
DM5E Ope 41
4.4 Di
This view ness and a nominal thickness input by the user, along with the differ d in percentage and dimensional terms.
Note: Th
t View Mode
rating Manual
fferential/Rate-of-Reduction Measurement View Mode
mode (see Figure 16 below) displays both the currently measured thickence between these two values (“measured” minus “nominal”) expresse
e displayed differences can have either positive or negative values.
Figure 16: DIFF/RR% Measuremen
Chapter 4
42 DM5E Operating Manual
4.5 B-
The B-SC the material thickness in a graphical mode. During the evaluatio m and maximum thickness values. A time-out feature allows a m s mode, the instrument will continue to collect and compare SEND key will store the minimum value in the active file.
View Mode
. Measuring Thickness
SCAN Measurement View Mode
AN mode (see Figure 17 below) allows the user to continuously displayn period, the thickness measurement is displayed along with the minimu
omentarily uncoupling of the probe during the evaluation period. In thireadings for as long as the time-out period is not exceeded. Pressing the
Figure 17: B-SCAN Measurement
Chapter 4. Measuring Thickness
DM5E Ope 43
4.6 Du
Protectiv easurement of the underlying metal when using conventio ayers from the measurement by measuring between successiv be will vary depending upon the type and thickness of the coating, t g and the metal.
To activa on menu or press and hold for 1.5 seconds while in
measurem mode, while a dual-peak icon ( ) above the battery icon indic
Note: Th y specially-equipped DM5E and DM5E DL versions offer th
rating Manual
al Multi Measurement View Mode
e coatings, including paint, contribute significant error to the thickness mnal measurement techniques. The Dual-Multi mode eliminates coating le back wall reflections of the metal. The measurement range for each prohe thickness of the metal, and the quality of the bond between the coatin
te or deactivate the Dual-Multi mode, you can either use the Configurati
ent mode. A single-peak icon ( )above the battery icon indicates IPates Dual Multi mode (see Figure 18 on page 44).
e Basic version of the DM5E does not have the Dual Multi function. Onle Dual Multi function.
Chapter 4
44 DM5E Operating Manual
4.6 D
nt View Mode
Dual Multi mode deactivated
. Measuring Thickness
ual Multi Measurement View Mode (cont.)
Figure 18: DUAL MULTI Measureme
Dual Multi mode activated
Chapter 5. Using the Optional Data Recorder
DM5E Ope 45
Chapt
Note: Th Data Recorder.
When ins a Recorder files, store thickness readings (and the accompan s chapter you will learn how to:
• Creat
• Recal
• Store
• Navig 5.3.2 on page 51)
rating Manual
er 5. Using the Optional Data Recorder
e instructions in this chapter apply only to instruments equipped with a
talled and activated, the Data Recorder option allows users to create Datying B-SCAN, when desired), and navigate through file contents. In thi
e Data Recorder files (see Section 5.1 on page 46)
l and erase Data Recorder files (see Section 5.2 on page 49)
thickness readings in Data Recorder files (see Section 5.3.1 on page 51)
ate Data Recorder file locations to store or replace readings (see Section
Chapter 5
46 DM5E Operating Manual
5.1 Cr
To create es, however, you will wish to enter additional information. Paramete
• The n 20 on page 48).
• The A
• NOTE
Note: O cannot be edited. Only the NOTES contents can be mo he parameter.
Follow S file.
. Using the Optional Data Recorder
eating a New Data Recorder File
a new Data Recorder file, you need only specify a file name. In most casrs that can be adjusted during the file-naming process include:
umbers of ROWS and COLUMNS in the Data Recorder file (see Figure
DVANCE DIRECTION (see Figure 20 on page 48).
S related to the data file
nce created, the file name, size and ADVANCE DIRECTION parametersdified. To accomplish this, first select the file then activate and modify t
teps 1 through 5 in Figure 19 on page 47 to create a new Data Recorder
Chapter 5. Using the Optional Data Recorder
DM5E Ope 47
5.1 Cr
ecorder File
rating Manual
eating a New Data Recorder File (cont.)
Figure 19: Creating a New Data R
Chapter 5
48 DM5E Operating Manual
5.1 Cr
ion of a Data Recorder File
. Using the Optional Data Recorder
eating a New Data Recorder File (cont.)
Figure 20: Defining the Size and Advance Direct
Chapter 5. Using the Optional Data Recorder
DM5E Ope 49
5.2 Re
Stored Da ickness measurements can be stored in available file locations
5.2.1
1. Press d then set DATA RECORDER to ON.
2. Press
3. Activ will appear on the display.
4. Press to confirm the selection.
rating Manual
calling and Erasing Stored Data Recorder Files
ta Recorder files can be recalled or erased at any time. Once recalled, th, but the advance direction cannot be changed.
Recalling a Data Recorder File
as many times as necessary to access the Configuration display an
again to launch the File Display mode.
ate the file selection function by pressing below FILES. A file list
or to select the stored file you wish to recall, then press
Chapter 5
50 DM5E Operating Manual
5.2.2
1. Press
2. If ne
3. Press old below ERASE.
IMPORTA urements cannot be retrieved.
5.3 Re s
The Data lay. Once the Data Recorder is turned on, the contents of the active igure 20 on page 48.
To store a Remember that the units of the thickness value and the number o e UNITS and RESOLUTION settings as described in Section 3
. Using the Optional Data Recorder
Erasing a Data Recorder File
again to launch the File Display mode.
cessary, press below FILES to display a list of the stored files.
or to select the stored file you wish to erase, then press and h
NT: Once erased, the Data Recorder file and any stored thickness meas
cording Thickness Measurements in Data Recorder File
Recorder must be enabled (turned ON) via the CONFIGURATION disp Data Recorder file are displayed in the File Navigation Grid shown in F
thickness reading in the selected file position, press below SEND.f decimal places to which the thickness is recorded are determined by th.2 on page 21.
Chapter 5. Using the Optional Data Recorder
DM5E Ope 51
5.3.1
When in ata Recorder by pressing below SEND.
5.3.2
When the ile are displayed in the File Navigation Grid shown in
Figure 20 can be used to select any file position. A heavier grid box appe ed:
• The c
• A me
• A pre
rating Manual
Storing Minimum of B-Scan in Data Recorder Files
B-Scan mode, the minimum value is stored in the active position of the D
Navigating Through Data Recorder Files
Data Recorder is turned ON, the contents of the active Data Recorder f
on page 48. After the navigation grid is activated, , , and ars around the selected file position. Once a file position has been select
urrent thickness reading can be sent to an EMPTY file position.
asurement value can be stored by pressing below SEND.
viously-stored value can be overwritten by pressing below SEND.
Chapter 5
52 DM5E Operating Manual
eed to next page]
. Using the Optional Data Recorder
[no content intended for this page - proc
Chapter 6. I/O Features
DM5E Ope 53
ChaptNote: Th ent.
The DM5 mini-USB to PC cable is available for connection to this virtual CO ltraMATE Lite, or even a simple Excel Macro, a PC can import da
COM Port
rating Manual
er 6. I/O Featurese instructions in this chapter apply to all versions of the DM5E instrum
E can communicate with external devices through its built-in I/O port. AM port (see Figure 21 below for the location). With the UltraMATE, U
ta recorder files from the instrument.
Figure 21: Location of Mini USB
Chapter 6
54 DM5E Operating Manual
6.1 Co
The instru re to install the virtual COM port driver on the PC before making th the new hardware.
. I/O Features
nnecting the DM5E to a PC
ment’s I/O port can be connected to a PC using a mini-USB cable. Be sue connection. Then, follow the instructions on your PC screen to install
Chapter 7. Specifications
DM5E Ope 55
Chapt
This chap
• Instru
• Featu
• Probe
rating Manual
er 7. Specifications
ter lists the features and specifications of your DM5E including:
ment specifications (see Section 7.1 on page 56)
res of the optional Data Recorder (see Section 7.2 on page 60)
specifications (see Section 7.3 on page 61)
Chapter 7
56 DM5E Operating Manual
7.1 Ins
Operatin
Measurin to 1st BW Measurement Mode,ual-Multi Measurement Mode, 2.50mm (0in to 0.098in).
Measurin )
Material
Material
Units:
Calibrati -Point Cal
. Specifications
trument Specifications
g Principle: Ultrasonic, pulse echo measurement method
g Range: Depends on probe and material,0.60mm to 508mm (0.025in to 20.00in) in IP2.00mm to 127.0mm (0.079in to 5.00in) in Dthe coating thickness range shall be 0.3mm to
g Resolution: 0.001 inch default (selectable 0.001, 0.01 inch0.01 mm default (selectable 0.01, 0.1 mm)
Velocity Range: 0.0200 to 0.7362 in/msec0.508 to 18.699 mm/msec
Velocity Resolution: 0.0001 in/msec1 m/s
inch or millimeter
on: One-Point Cal, On-Block and Off-Block, Two
Chapter 7. Specifications
DM5E Ope 57
7.1 In
Pulser:Hz oscilloscope
Receiver:
Display T els,ble contrast
Update R
Thickness highigh
Display o or uncoupled condition
Setups:
rating Manual
strument Specifications (cont.)
Excitation Pulse: Spike PulserVoltage: 120 V into 50 ohm load, using 20 M
Bandwidth: 1.5 MHz to 12 MHz @ -3 dBGain: Automatic Gain Control
ype: High-resolution graphical LCD, 64 x 128 Pix53.0mm x 27.0mm with backlight and adjusta
ate: 4 Hz or 8 Hz, user selectable,24 Hz scan mode capture rate
Value Display: NORMAL view mode: 5 digit, 0.4" (10.6 mm)B-SCAN view mode: 5 digit, 0.1" (2.55 mm) h
f Last Reading: Solid filled or hollow digits indicate coupled
9 standard setups for probes
Chapter 7
58 DM5E Operating Manual
7.1 In
Alarm Se 508 mm)when alarms are enabled and violated
Power Re
Battery L
Instrumen 5, 10, 15, 30 minutes of inactivity
Language , Italian, Russian, Japanese and Chinese
I/O Conn
Temperat )
Weight:
. Specifications
strument Specifications (cont.)
ttings: Minimum and Maximum alarms,Range of 0.010 in to 20.00 in, 0. (0.25 mm toReading alternates between solid and hollow
quirements: 2 “AA” size batteries
ife/Operating Time: Approximately 100 hours
t Shut-0ff: Selectable ALWAYS ON or AUTO OFF after
: Selectable English, German, French, Spanish
ectors: Transducer: Dual Lemo 00 (coax)Mini-USB: MINI-B OTG to STD A
ure: Operating: +10°F to +120°F (-10°C to +50°CStorage: -10°F to +140°F (-20°C to +60°C)
0.597 lb (223 g) including batteries
Chapter 7. Specifications
DM5E Ope 59
7.1 In
Size:
Shock: 516.2 Procedure Ie applied 6 times per axis
Sealing: of as per IEC 529 specifications for IP54 classification
Note: In
rating Manual
strument Specifications (cont.)
138mm x 32mm x 75mm
IEC 68-2-27 Ea, as per Mil Std 810C Methodwith a 15g 11ms impulse half sinusoidal wav
IEC529 / IP54, Dust proof/dripping water pro
strument specifications are subject to change without prior notice.
Chapter 7
60 DM5E Operating Manual
7.2 Da
Capacity
File Struc
Number o
Number o
File Nam
Optional
Note: In
. Specifications
ta Recorder Option Features
: 50,000 readings
tures: Grid file
f Rows: 1 to 50,000
f Columns: 1 to 223
ing: Up to 24 character alphanumeric name
Software: UltraMATE and UltraMATE Lite
strument specifications are subject to change without prior notice.
Chapter 7. Specifications
DM5E Ope 61
7.3 DM
Model DA512 Model DA590
10 MHz 5 MHz
fingertip high-temperature
OperRa
-10 to 70°C -10 to 204°C
7.5 mm (.300”) 12.7 mm (.500”)
Me0.6 to 60 mm(.020 to 2.4”)
1 to 125 mm @20°C(.040 to 5” @68°F)
1.3 to 25.4 mm @204°C(.050 to 1” @400°F)
MinMe
2.0 to 10 mm N/A
rating Manual
5E Probe/Transducer Specifications
Model DA501 Model DA503
Frequency: 5 MHz 2 MHz
Probe Style: standard standard
ating Temperaturenge (continuous):
-10 to 70°C -10 to 70°C
Contact Diameter: 15 mm (.590”) 20 mm (.787”)
IP to Firstasurement Range:
1.0 to 200 mm(.040 to 8”)
5 to 300 mm(.200 to 12”)
imum Multi-Echoasurement Range:
3.0 to 25 mm N/A
Chapter 7
62 DM5E Operating Manual
eed to next page]
. Specifications
[no content intended for this page - proc
Chapter 8. Maintenance
DM5E Ope 63
Chapt
8.1 Ca
Clean the h water or a mild window-cleaning product. Never use harsh solv
8.2 Pr
• Avoid
• Grasp om the instrument.
8.3 Ba
Periodica s, remove and replace ALL the batteries. Take care to properly
rating Manual
er 8. Maintenance
re of the instrument
instrument housing and display using a soft cloth lightly dampened witents as they could cause the plastic to become brittle or damaged.
oper Cable Handling
twisting or knotting of the cable.
the cable only by the connectors when connecting or disconnecting it fr
tteries
lly inspect the batteries for signs of leakage or corrosion. If either occurdispose of the defective batteries.
Chapter 8
64 DM5E Operating Manual
eed to next page]
. Maintenance
[no content intended for this page - proc
Appendix A. Supplemental Information
DM5E Ope 65
Appen
This appe
• Reset
• Upgra
• Web s
• EMC
• Manu
rating Manual
dix A. Supplemental Information
ndix contains supplementary information, including:
ting the instrument’s operating software (see Section A.1 on page 66)
ding the instrument's operating software (see Section A.2 on page 67)
ite information (see Section A.3 on page 69)
documentation (see Section A.4 on page 70)
facturing and Service Center locations (see Section A.5 on page 71)
Appendix
66 DM5E Operating Manual
A.1 Re
IMPORTA ault values. Any files you wish to retain should be operating software. However, any Data Recorder files
To reset t
1. Turn
2. Press til the display indicates that the reset is complete.
3. A su briefly appears at the bottom center of the display screen.
A. Supplemental Information
setting the Operating Software
NT: The following procedure will set all parameters to their factory defdownloaded using the UltraMATE PC software before resetting thestored in the instrument will not be deleted during this procedure.
he instrument's operating software:
the instrument OFF.
and hold . Then press and hold for approximate 3 seconds, un
ccessful reset is acknowledged when the RESET COMPLETE message
Appendix A. Supplemental Information
DM5E Ope 67
A.2 Fir
The DM5 an upgrade utility program. Each instrument is shipped with a CD e, proceed as follows:
1. Insta ction.
2. Run typical MS Windows operating system, this is acco ting PROGRAMS and then selecting the DM5E Upgrade Utili
3. Selec l PC drive, or select the “Download new file” button to conn connected to the internet prior to this operation). The prog ersion that is on the FTP site. If the firmware version on the F code.
Note: If e and will exit the download routine.
4. Turn
5. Click
rating Manual
mware Updates
E operating software can be upgraded using an internet connection and containing the DM5E Upgrade Utility Program. To perform the updat
ll the Upgrade Utility Program on a PC that has an active internet conne
the Upgrade Utility Program from the Program Menu on your PC. For amplished by clicking the START button, clicking the RUN button, selecty (see Figure 22 on page 68).
t the “Browse local disk” button if the new firmware is stored on a locaect to the GE Sensing & Inspection Technologies FTP site (you must beram checks the firmware version you have on your machine against the vTP site is newer, the utility will download the updated DM5E operation
the firmware versions are the same, the utility will not download the cod
the DM5E OFF and connect it to the PC.
on the “Upgrade The Instrument” button.
Appendix
68 DM5E Operating Manual
A.2 Fi
r Settings
A. Supplemental Information
rmware Updates (cont.)
Figure 22: Changing Paramete
Appendix A. Supplemental Information
DM5E Ope 69
A.2 Fi
6. Whe M5E will enter re-programming mode. When the update has b
IMPORTA r, stored Data Recorder files will not be affected.
7. Rese
A.3 W
The follo gies web site at www.geinspectiontechnologies.com:
• Up-to
• Lates
• Speci
• New f
• Litera
• And m
rating Manual
rmware Updates (cont.)
n prompted, press and hold the and keys for 3 seconds. The Deen completed, the DM5E will automatically shut OFF.
NT: All stored custom settings will be lost during the next step. Howeve
t the operating software to purge the instrument’s memory.
eb Site Information
wing information is available on the GE Sensing & Inspection Technolo
-date information on operating software
t upgrade utility program
fications
eatures
ture
uch more!
Appendix
70 DM5E Operating Manual
A.4 EM
The follo
ness Gauge
Instrumen easuring materials using ultrasonic transducers operating at frequen by the measured time of flight in the material at the velocity p ortable thickness gauge in an industrial environment. The instrumen ess control system.
Complia
The DM5
• EN55
• EN61
A. Supplemental Information
C Documentation
wing is the DM5E Declaration of EMC Compliance:
DM5E Ultrasonic Corrosion Thick
t Description: The DM5E is an ultrasonic thickness gauge capable of mcies from 5 to 15 MHz. This instrument displays a thickness calculated rogrammed into the instrument. The instrument is intended for use as a pt is not intended to be part of a closed control loop in an automated proc
nce Statement:
E complies with the following EN standards:
011: 2007
000-6-2: 2005
Appendix A. Supplemental Information
DM5E Ope 71
A.5 Ma
The DM5 lity components. In-process testing and a quality managem conformance. The DM5E is manufactured by:
ologies
aoqiao Free Trade Zone,
Should yo t service page accessible at www.gein gies representative, or contact one of the Customer Support C
rating Manual
nufacturing & Customer Service Locations
E is manufactured according to state-of-the-art methods using high-quaent system certified to ISO 9001 ensure optimum quality of instrument
GE Sensing & Inspection TechnChina
Q Section, 4th Floor, #185, Tai-Gu Road, WaigShanghai, China
u experience a problem or require technical assistance, visit the producspectiontechnologies.com to locate your local GE Inspection Technoloenters listed on the back cover of this manual.
Appendix
72 DM5E Operating Manual
eed to next page]
A. Supplemental Information
[no content intended for this page - proc
Appendix B. Environmental Compliance
DM5E Ope 73
Appen
This appe
• WEE
• Batter
rating Manual
dix B. Environmental Compliance
ndix contains information on the following topics:
E Directive (see Section B.1 on page 74)
y disposal (see Section B.2 on page 75)
Appendix
74 DM5E Operating Manual
B.1 W ve
GE Sensi Electrical and Electronic Equipment (WEEE) take-back initiative,
and use of natural resources for its production. It may the environment.
ur environment and to diminish the pressure on the natural ck systems. Those systems will reuse or recycle most of
se systems.
If you ne se contact your local or regional waste administration.
Visit ww information about this initiative.
B. Environmental Compliance
aste Electrical and Electronic Equipment (WEEE) Directi
ng & Inspection Technologies is an active participant in Europe’s Waste directive 2002/96/EC.
The equipment that you bought has required the extractioncontain hazardous substances that could impact health and
In order to avoid the dissemination of those substances in oresources, we encourage you to use the appropriate take-bathe materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use tho
ed more information on the collection, reuse and recycling systems, plea
w.ge.com/inspectiontechnologies for take-back instructions and more
Appendix B. Environmental Compliance
DM5E Ope 75
B.2 Ba
s unsorted municipal waste in the European Union. See . The battery is marked with this symbol, which may rcury (Hg). For proper recycling return the battery to your
B.2.1 W
Batteries or on its packaging, depending on size) with theseparate c mbols of specific levels of toxic metals as follows:
• Cadm
• Lead
• Mercu
rating Manual
ttery Disposal
This product contains a battery that cannot be disposed of athe product documentation for specific battery informationinclude lettering to indicate cadmium (Cd), lead (Pb), or mesupplier or to a designated collection point.
hat do the Markings Mean?
and accumulators must be marked (either on the battery or accumulator ollection symbol. In addition, the marking must include the chemical sy
ium (Cd) over 0.002%
(Pb) over 0.004%
ry (Hg) over 0.0005%
Appendix
76 DM5E Operating Manual
B.2.2 T
Your part eries and accumulators on the environment and on human health. Fo lators it contains to your supplier or to a designated collection
Some bat an health and to the environment. When required, the product m ls: Pb for lead, Hg for mercury, and Cd for cadmium.
• Cadm onic effects include kidney damage, pulmonary emph m may also cause anemia, discoloration of the teeth, and loss o
• Lead s significant. Ingestion and inhalation of lead can cause severe lnutrition, and sterility.
• Merc centrations of mercury vapor can cause a variety of severe symp ty change, nervousness, fever, and rashes.
B. Environmental Compliance
he Risks and Your Role in Reducing Them
icipation is an important part of the effort to minimize the impact of battr proper recycling you can return this product or the batteries or accumu point.
teries or accumulators contain toxic metals that pose serious risks to humarking includes chemical symbols that indicate the presence toxic meta
ium poisoning can result in cancer of the lungs and prostate gland. Chrysema, and bone diseases such as osteomalcia and osteoporosis. Cadmiuf smell (anosmia).
is poisonous in all forms. It accumulates in the body, so each exposure i damage to human health. Risks include brain damage, convulsions, ma
ury creates hazardous vapors at room temperature. Exposure to high contoms. Risks include chronic inflammation of mouth and gums, personali
Index
DM5E Ope 77
Alarms, S
BacklighBase InstBatteries
DispInstaLeveMainOperSize
B-ScanDispMeasStori
C
ections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53tenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63n
edure, 1pt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27edure, 2 pt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28edure, Auto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29n to Perform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26t
/O Portationsting Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21lay Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13lay Mode Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14u . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
eters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
rating Manual
Aetting Minimum and Maximum . . . . . . . . . . . . . . . . . . 30
Bt, Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22rument, Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
osal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75lling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1l Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3tenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63ating Life. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
lay Range, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31urement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42ng Minimum in Data Recorder File . . . . . . . . . . . . . . . 51
CablesConnI/O PMain
CalibratioProcProcProcWhe
COM Porsee I
ConfigurAdjuDispDispMenParam
Index
78 DM5E Operating Manual
ConnectiI/O DProbProb
Contrast,Controls,Customer
Data RecCreaNaviRecaSpecStoriStori
DateManSetti
Different
odeiguration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12urement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10ns. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
ries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75ronic Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3ad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7r Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1ring ON/OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
ng Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17ions Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 6ti Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Ecumentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
onsevice. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18e Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Locking & Unlocking. . . . . . . . . . . . . . . . . . . . . . . . . . 32 Support Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Dorderting a New File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46gating Through File. . . . . . . . . . . . . . . . . . . . . . . . . . . . 51lling & Erasing Files . . . . . . . . . . . . . . . . . . . . . . . . . . . 49ifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60ng B-Scan Minimum. . . . . . . . . . . . . . . . . . . . . . . . . . . 51ng Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
ual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22ial/Rate-of-Reduction Measurement Mode . . . . . . . . . . 41
Display MConfFile .MeasOptioProb
DisposalBatteElect
DM5EKey KeypPowePoweSettiVers
Dual Mul
EMC Do
Index
DM5E Ope 79
Features,File
DataDataDataDispSetup
FirmwareFunction
Gain, Set
I/O PortCablType
KeypadKeysPhot
L
ns. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58ng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Unlocking the Controls . . . . . . . . . . . . . . . . . . . . . . . 32M
nceries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63es . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63ument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63uring Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71ent Mode
an . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42rential/Rate-of-Reduction . . . . . . . . . . . . . . . . . . . . . . . 41 Multi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Scan & Max Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40al . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39ns. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36ting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
rating Manual
F Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Recorder, Creating . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Recorder, Navigating . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Recorder, Recalling & Erasing. . . . . . . . . . . . . . . . . . . 49lay Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12, Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, Updating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67Keys, Locking & Unlocking. . . . . . . . . . . . . . . . . . . . . 33
Gting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
I
e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 and Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
K
, Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7ograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
LanguageOptioSetti
Locking &
MaintenaBatteCablInstr
ManufactMeasurem
B-ScDiffeDualMin NormOptioSelec
Index
80 DM5E Operating Manual
MeasuremDispStoriThic
Min ScanMode
see M
Nominal Normal M
Part NumPC, ConnPower foPoweringProbe
ConnDispSetupSpec
R the Operating Software . . . . . . . . . . . . . . . . . . . . . . . . . 66
Sformation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiiiplay, Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
eing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
eters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20rument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59, Resetting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66tions Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60ument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56e/Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Tureating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58ge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
entslay Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10ng in a Data Recorder File . . . . . . . . . . . . . . . . . . . . . . 50kness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 & Max Scan Measurement Modes . . . . . . . . . . . . . . . 40
easurement Mode or Display ModeN
Thickness, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25easurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Pber, Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iecting to . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
r DM5E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
ecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 18lay Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
ifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Resetting
Safety InSetup DisSetup Fil
LoadParam
Size, InstSoftwareSpecifica
DataInstrProb
TemperatOperStora
Index
DM5E Ope 81
ThicknesMeasNom
Time, SeTraining,Transduc
see P
Units, SeUpdate RUpdatingUpgrade
Versions View Mo
see M
Warnings
sposalries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75ronic Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69irective . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74nstrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
rating Manual
suring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35inal, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25tting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viiierrobe
Utting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22ate, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 the Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67Utility Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
VAvailable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6deeasurement Mode
W, General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Waste DiBatteElect
Web SiteWEEE DWeight, I
Index
82 DM5E Operating Manual
eed to next page]
[no content intended for this page - proctiontechnologies.com
ric Company. All rights reserved.bject to change without notice.
rmeaux
216
0
lanta 4A, Nave 6
990
China5F, Building 1, No. 1 Huatuo Road,Zhangjiang High-Tech Park,Shanghai 201203ChinaTel: +86 800 915 9966 (toll-free)
+86 (0)21 3877 7888
Japan7F Medie Corp Bldg. 82-4-14-Kichijoji Honcho,Musashino-shiTokyo 180-0004JapanTel: +81 442 67 7067
www.geinspec
©2009 General ElectTechnical content suP/N 1259551 Rev 1
Customer Support Centers
North/South America50 Industrial Park RoadLewistown, PA 17044U.S.A.Tel: 866 243 2638 (toll-free)
717 242 0327
United Kingdom/Ireland892 Charter Avenue CanleyCoventry CV4 8AFEnglandTel: +44 845 130 3925
France68, Chemin des OLimonest 69760FranceTel: +33 47 217 9
GermanyRobert Bosch Str.50354 HürthGermanyTel: +49 2233 601
SpainSan Maximo,31, PMadrid 28041SpainTel: +34 195 005
E-mail: [email protected]