Page 1
Column Dimensions and
Carrier Gas Optimization
![Page 2: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/2.jpg)
Page 2
•Optimized Stationary Phase
•Longer Column Length
•Decrease Internal Diameter
•Carrier Gas: type and linear velocity
•Optimized Temperature Programming
Variables for Maximizing Resolution
![Page 3: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/3.jpg)
Resolution
N = (gas, L, rc)
k = (T, df, rc)
a = (T, phase)
RN k
ks =
4 1
1
a
a
Efficiency
Retention
Selectivity
L = Length
rc = column radius
df = film thickness
T = temperature
Page 3
![Page 4: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/4.jpg)
Agilent J&W has over 50 different stationary phase offerings
WCOT Column Types
Page 4
![Page 5: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/5.jpg)
FactorFourTM Phases
VF-1ms, VF-5ms, VF-5ht, VF-5ht UltiMetal™
VF-17ms, VF-17ms for PAH, VF-35ms,
VF-200ms,VF-Xms, VF-23ms, VF-624ms,
VF-DA, VF-1301ms, VF-Pesticides,
VF-1701ms, VF-WAXms
Page 5
![Page 6: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/6.jpg)
Column Typical Application
DB-624 EPA and USP volatiles
DB-VRX volatiles analysis
HP-VOC volatiles analysis
DB-502.2 EPA Method 502.2
DB-5.625 EPA semi-volatiles analysis
DB-608 EPA Method 608
DB-1701P EPA pesticides analysis
DB-MTBE total petroleum hydrocarbon (TPH)
HP-PONA petroleum hydrocarbon analysis
DB-HT SimDis hi-temp simulated distillation
DB-ALC1 & ALC2 blood alcohol analysis
HP-88 fatty acid methyl ester (FAME)
“Specialty phases” are columns that are optimized to perform
a specialized GC analysis.
20+ Different “Specialty Phases”
Page 6
![Page 7: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/7.jpg)
Select TM Column Examples
Environmental applications
CP-Sil 88 for dioxins, Select mineral oil, CP-Select 624 CB
Chiral applications
CP-Chirasil Val, CP-Chirasil-DEX CB
Chemical applications
CP-Volamine, CP-Select CB for MTBE, CP-PONA C8, CP-Propox,
Select Silanes, CP-SimDist UltiMetal TM , CP-Lowox TM
Food and Beverage applications
CB-Carbowax 400, Select FAME, CP-Sil 88 for FAME, CP-FFAP CB
Page 7
![Page 8: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/8.jpg)
PLOT columns are primarily, but not exclusively, used for the analysis of gases and low boiling
point solutes (i.e., boiling point of solute is at or below room temperature).
PLOT Column Types
• Agilent J&W PLOT columns begin
with the designation of
– GS (Gas Solid) or
– HP-PLOT followed by a specific name
– 10 stationary phases
• GS-OxyPLOT
• GS-Alumina
• HP-PLOT Al2O3 “M”
• HP-PLOT Al2O3 “S”
• HP-PLOT Al2O3 “KCl”
• HP-PLOT MoleSieve
• GS-CarbonPLOT
• HP-PLOT Q
• HP-PLOT U
• GS-GasPro
• GS-OxyPLOT: oxygenates
• HP-PLOT Molesieve: O2, N2, CO, Methane
• HP-PLOT Alumina and GS-Alumina: complex hydrocarbon gas matrices, ethylene and
propylene purity, 1,4-butadiene
• HP-PLOT Q: freons, sulfides
• HP-PLOT U: C1 to C7 hydrocarbons, CO2, Polar Hydrocarbons
• GS-GasPro: freons, sulfurs, inorganic gases
• GS-CarbonPLOT: inorganic and organic gases
Page 8
![Page 9: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/9.jpg)
PLOT GC Columns from Varian Line
Porous Polymers
• CP-PoraBOND Q
• CP-PoraBOND U
• CP-PoraPLOT Q
• CP-PoraPLOT U
• CP-PoraPLOT S
• CP-PoraPLOT Q-HT
• CP-PoraPLOT amines
Zeolites
• Molsieve 5A
• Molsieve 13x
Alumina
• KCL
• Na2SO4
• MAPD
Multi Layer
• CP-Lowox
Porous Silica
• SilicaPLOT
Graphatised Carbon
• CP-CarboPLOT P7
• CP-CarboBOND
SelectTM permanent gases
Page 9
![Page 10: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/10.jpg)
Page 10
Optimizing Selectivity
Match analyte polarity to stationary phase polarity
-like dissolves like (oil and water don’t mix)
Take advantage of unique interactions between analyte and
stationary phase functional groups
![Page 11: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/11.jpg)
Page 11
Stationary Phase Selection
Existing Information
Critical Separations
Selectivity/Polarity
Temperature Limits
Application Designed
Examples: DB-CLP1, DB-CLP2, DB-UI 8270D, DB-624UI,
DB-Select 624UI<467>, DB-VRX, HP-VOC, DB-MTBE,
Lowox, DB-TPH, DB-HTSimDis, DB-2887, DB-HT SimDis,
CP-Volamines, DB-ALC1, DB-ALC2, Select Phases, etc.
Choose the column phase that gives the best separation
but not at the cost of robustness or ruggedness.
![Page 12: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/12.jpg)
Complete Resolution & Fast CLP Pesticide Analysis
Page 12
![Page 13: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/13.jpg)
Agilent J&W DB-CLP1 and DB-CLP2
For 9 EPA Methods+ (More than any other CLP column pair!)
Page 13
![Page 14: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/14.jpg)
Selectivity Differences
DB-624
DB-VRX
Page 14
![Page 15: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/15.jpg)
Benzene and 1,2-Dichloroethane Separation
DB-Select 624UI<467>
30m x 0.53mm 30m x 0.32mm
DB-624
Co-elution
Page 15
![Page 16: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/16.jpg)
Benzene and 1,2-Dichloroethane Separation
Baseline
Resolved
R= 1.82
Vendor R
G43
Vendor P
G43
R= 1.59 R= 1.38
DB-Select 624UI<467>
All 30m x 0.32mm
![Page 17: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/17.jpg)
Resolution
N = (gas, L, rc)
k = (T, df, rc)
a = (T, phase)
RN k
ks =
4 1
1
a
a
Efficiency
Retention
Selectivity
L = Length
rc = column radius
df = film thickness
T = temperature
Page 17
![Page 18: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/18.jpg)
Column Diameter - Theoretical Efficiency
I.D. (mm) n/m
0.05 23,160
0.10 11,980
0.20 5830
0.25 4630
0.32 3760
0.45 2840
0.53 2060
0.18 6,660
k = 5
Page 18
![Page 19: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/19.jpg)
Different Column I. D. Equal Phase Ratios
Time (min)
0 5 10 15 20
Carrier: Oven: 65° C Injection: Split Detector: FID
Column : DB-624 30 m, 0.32 mm, 1.8 m
Helium, 40 (cm/sec)
Column : DB-624 30 m, 0.53 mm, 3 m
Page 19
![Page 20: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/20.jpg)
PHASE RATIO () Film Thickness
Column Dimensions Phase Ratio β
30 m x .53 mm x 3.0 m 44
30 m x .32 mm x 1.8 m 44
KC = k β
r
2df β =
Page 20
![Page 21: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/21.jpg)
High Resolution Megabore Same Resolution - Faster Analysis!
“Increasing Sample Throughput
With High-Speed Megabore”
Application note 5988-5271EN
Page 21
![Page 22: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/22.jpg)
Column Diameter and Capacity
Like Polarity
Phase/Solute
0.25 µm film thickness
I.D. (mm) Capacity (ng)
0.05 1-2
0.18 25-55
0.20 35-70
0.25 80-160
0.32 110-220
0.53 1000-2000
0.45 600-800
0.10 6-13
Page 22
![Page 23: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/23.jpg)
Column Length and Efficiency (Theoretical Plates)
0.25 mm ID
n/m = 4630 (for k = 5)
Length (m) N
15 69,450
30 138,900
60 277,800
More Meters = More Plates = More Resolution
Page 23
![Page 24: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/24.jpg)
Column Length and Resolution
Length X 4 = Resolution X 2
RaN aL
t a L
Upside = Cut a bunch off during routine inlet
maintenance and not lose a lot of Resolution
Page 24
![Page 25: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/25.jpg)
Column Length VS Resolution and Retention:
Isothermal
Double the plates, double the time
but not double the the resolution
15 m 60 m 30 m
R=0.84
2.29 min
R=1.68
8.73 min
R=1.16
4.82 min
Page 25
![Page 26: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/26.jpg)
Page 26
Column Length and Cost
15m 30m
60m
$ $ $ $ $ $ $
![Page 27: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/27.jpg)
Resolution
N = (gas, L, rc)
k = (T, df, rc)
a = (T, phase)
RN k
ks =
4 1
1
a
a
Efficiency
Retention
Selectivity
L = Length
rc = column radius
df = film thickness
T = temperature
Page 27
![Page 28: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/28.jpg)
Page 28
10 20 30 40 50 60
0.25
0.50
0.75
1.00
u (cm/sec)
h
He
H 2
Excessive Diffusion
Poor Mass Transfer
ū opt
Carrier Gas Maximum Resolution = Optimum Velocity = Slowest Velocity
Van Deemter Curves
N 2
![Page 29: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/29.jpg)
Carrier Gas
Type Velocity Range (uopt – OPGV)
Nitrogen 10-17 cm/sec
Helium 22-40 cm/sec
Hydrogen 35-55 cm/sec
Page 29
![Page 30: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/30.jpg)
Changes in Column Dimensions, Gas Type or
Velocity Require Changes in Temp Program Rates
Method Translation Software to the Rescue!
Page 30
![Page 31: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/31.jpg)
Phenol’s – 30m x 0.25mm ID, 0.25μm, DB-5ms
Check Valleys
Page 31
![Page 32: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/32.jpg)
Phenol’s – 20m x 0.18mm ID, 0.18μm, DB-5ms
0.54 mL/min, H2
(~47cm/sec)
Page 32
![Page 33: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/33.jpg)
Slower than Best Efficiency? – must go faster!
(Sorry customer doesn’t want temperature program revealed)
Page 33
![Page 34: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/34.jpg)
Phenol’s – 20m x 0.18mm ID, 0.18μm, DB-5ms
1 mL/min, (rounded up?)
higher than optimum,
BUT still better Resolution!
Page 34
![Page 35: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/35.jpg)
2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0
1 e + 0 7
2 e + 0 7
3 e + 0 7
4 e + 0 7
5 e + 0 7
6 e + 0 7
7 e + 0 7
8 e + 0 7
9 e + 0 7
1 e + 0 8
1 . 1 e + 0 8
1 . 2 e + 0 8
1 . 3 e + 0 8
1 . 4 e + 0 8
1 . 5 e + 0 8
1 . 6 e + 0 8
1 . 7 e + 0 8
1 . 8 e + 0 8
1 . 9 e + 0 8
T im e
R e s p o n s e _
G C 3 -6 7 0 7 . D \ E C D 2 B
22
21
20
19
18
17
16
15 14
13
12
11
10
9
8
7
5,6
4
3
2
1
Column: DB-XLB
30m x 0.32mm i.d., 0.25µm
Carrier: He, constant flow, 38 cm/s at 120°C
Injector: Pulsed Splittless, 220 °C
Pulse pressure & time: 35psi for 1.15min
2µL, 50ppb
Oven: 120°C for 1.17min
120°C to 160°C at 25°/min
160°C to 260°C at 10°/min
260°C to 300°C (4min) at 15°/min
Detector: µ-ECD, 320°C
Ar/CH4 (P5) makeup gas at 60mL/min
<16 minutes
Original Method
Page 35
![Page 36: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/36.jpg)
Best Efficiency
New Velocity
(SLOWER)
New Temp.
Program
(SLOWER)
Page 36
![Page 37: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/37.jpg)
Best Efficiency
23 minutes (Slower)
Better
Resolution
2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0
1 e + 0 7
2 e + 0 7
3 e + 0 7
4 e + 0 7
5 e + 0 7
6 e + 0 7
7 e + 0 7
8 e + 0 7
9 e + 0 7
1 e + 0 8
1 . 1 e + 0 8
1 . 2 e + 0 8
1 . 3 e + 0 8
1 . 4 e + 0 8
1 . 5 e + 0 8
1 . 6 e + 0 8
1 . 7 e + 0 8
1 . 8 e + 0 8
1 . 9 e + 0 8
T im e
R e s p o n s e _
G C 3 -6 7 0 7 . D \ E C D 2 B
Original NEW
![Page 38: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/38.jpg)
Resolution
N = (gas, L, rc)
k = (T, df, rc)
a = (T, phase)
RN k
ks =
4 1
1
a
a
Efficiency
Retention
Selectivity
L = Length
rc = column radius
df = film thickness
T = temperature
Temperature, the TRUMP card
Page 38
![Page 39: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/39.jpg)
Column Temperature Optimizing Temperature Programs
Most powerful variable
Changes Selectivity and Retention
Natural log (ln) relationship between retention and temperature
Most difficult to predict and develop
Often involves trial and error (Sorry)
Page 39
![Page 40: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/40.jpg)
Capillary Flow Technology Capabilities
• Backflushing with Purged Devices
• 2D GC – Deans Switch / Heart Cutting
• Splitter to Multiple Detectors
• Multiple Columns/Multiple Detectors
Page 40
![Page 41: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/41.jpg)
Purged Capillary Flow Devices
2-Way Splitter
with Makeup
3-Way Splitter
with Makeup Deans Switch
Page 41
Purged Union
(most recent)
![Page 42: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/42.jpg)
9.78 psi 11.14 psi
FID A
S/S Inlet
FID B
PCM
Restrictor
6.54 mL/min
4.54 mL/min
<< 1mL/min
2mL/min
purge
restrictor
8.54 mL/min
Off
6.54 mL/min
Column 1
Column 2
Heart Cutting 2-D GC – How It Works
Valve off, no heart cutting– inject sample, initial separation on column 1
Page 42
![Page 43: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/43.jpg)
FID A
S/S Inlet
FID B
PCM
On
6.54 mL/min
4.54 mL/min
6.54 mL/min
purge
restrictor
2mL/min 8.54 mL/min
<< 1mL/min
9.78 psi 11.14 psi
Restrictor
Column 1
Column 2
Heart Cutting 2-D GC – How It Works
Valve on – start heart cut from column 1 to column 2
Page 43
![Page 44: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/44.jpg)
FID A
S/S Inlet
FID B
PCM
6.54 mL/min
4.54 mL/min
<< 1mL/min
9.78 psi 11.14 psi
2mL/min
purge
restrictor
8.54 mL/min
Off
6.54 mL/min
BP>Benzene
Restrictor
Column 1
Column 2
Heart Cutting 2-D GC – How It Works
Valve off – end heart cut, perform 2nd separation on column 2
Page 44
![Page 45: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/45.jpg)
Calculator - Set Flows and Restrictor Size
Page 45
![Page 46: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/46.jpg)
Page 46
4 5 6 7 8 9 10 11 12
2.09 mg/kg thiophene
benzene
Need to measure 0.02 to 2.00 mg/kg thiophene in pure benzene
HP-Innowax Column
60m x 0.53mm x 0.5um
Trace Thiophene in Benzene
![Page 47: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/47.jpg)
Page 47
Analytical Precision
(15 runs over 5 days)
Avg Amt: 50 ppb
Std Dev: 0.95 ppb
RSD: 1.9%
Analysis of 20 ppb to 2 ppm using FID (2 uL splitless injection)
Trace Thiophene in Benzene
4 6 8 10 12 14 16 18 20
0.1
0.2
0.3
0.4
0.5
0.6
50 ppb thiophene
HP Plot Q
Column
4 6 8 10 12 14 16 18 20
0
100
200
300
400
500
Cut window
7.74 - 8.02 min
Innowax
Column
hydrocarbons
benzene
toluene
1,4-dioxane
![Page 48: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/48.jpg)
Agilent’s Ultimate Pesticide Analysis System
• FPD
• µECD
• MS SIM
• MS Scan
1 gas chromatograph
1 injection
Page 48
![Page 49: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/49.jpg)
Pesticides: Three Way Splitter with Makeup
1X method with 1:1:0.1 split FPD:MSD:ECD
3-Way
Splitter with
Makeup Auto-sampler
7890
GC
Column
Phosphorus FPD
30 m X 0.25 mm id X 0.25 um HP-5MS
5975C
MSD
uECD AUX EPC
3.8 psig
Page 49
![Page 50: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/50.jpg)
3-Way Splitter System
Deactivated splitter operates to 350 C.
Using metal ferrules eliminates leaks and retightening
To µECD
Column in
Makeup gas from EPC
To MSD
To FPD
Page 50
![Page 51: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/51.jpg)
Full scan TIC
SIM
µECD
5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00
FPD(P)
Milk Extract (1 injection)
Page 51
![Page 52: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/52.jpg)
Front Inlet
Back Inlet
AUX 2 AUX 3AUX 1
MSD
Three Way
SplitterTwo Way
Splitter
FIDHP-5ms
Main Oven
DB-1ms
LTM
DB-17ms
LTM
R-1
R-2
R-3
R-4
HP-5ms 30m x 0.25 x 0.25 19091S-433 à Column 1
DB-1ms 15m x 0.25 x 0.25 122-0112 à Column 2
R1 = 0.30m x 0.1 restrictor à Column 4
R2 = 0.17m x 0.1 restrictor à Column 5
R3 = 0.50m x 0.25 restrictor à to FID
R4 = 1.00m x 0.18 restrictor à Column 3
HP-5ms 30m x 0.25 x 0.25 19091S-433 (Column 1)
DB-17ms LTM 15m x 0.25 x 0.25 122-4712LTM
DB-1ms LTM 15m x 0.25 x 0.25 122-0112LTM (Column 2)
Pre-Column = 1m x 0.25 uncoated FS
Pre
Column
Deans
Switch
LTM and Capillary Flow Technology, Possibilities! Metabolomics – Crops Research (Polar and Non-polar Analytes)
Page 52
![Page 53: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/53.jpg)
Finally the Ultra Inert Story
•All that resolution is worth nothing if the peaks start
tailing and/or disappearing!
•Ultra Inert Columns and Liners provide the
maximum inertness available on the market.
•Activity only gets worse once you start injecting
samples. Start as inert as you can.
Page 53
![Page 54: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/54.jpg)
Ultra Inert liners
• Touchless packaging
• Easy installation of new, clean liner
without risk of contamination from touching
• Includes non-stick plasma treated O-ring
Page 54
![Page 55: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/55.jpg)
1. 1-Propionic acid
2. 1-Octene
3. n-Octane
4 4-Picoline
5. n-Nonane
6. Trimethyl phosphate
7. 1,2-Pentanediol
8. n-Propylbenzene
9. 1-Heptanol
10. 3-Octanone
11. n-Decane
Ultra Inert Test Mix - DB-5ms Ultra Inert vs. competitors
1
2 3
4 5
6 7
8
9 10 11
min 0 2 4 6 8 10
pA
10
12.5
15
17.5
20
22.5
25
27.5
Agilent J&W DB-
5ms Ultra Inert 30m x 0.25mm x
0.25um
(P/N 122-5532UI)
1
2 3 4
5
6 7
8
9 10 11
min 0 2 4 6 8 10
pA
6
8
10
12
14
16
18
20
1
2 3 4
5
6 7
8
9 10 11
min 0 2 4 6 8 10
pA
6
8
10
12
14
16
18
20
1
2 3
4
5
7
8
9 10
11
6 min 0 2 4 6 8 10
pA
8
10
12
14
16
18
Competitor
Column
Competitor
Column
Page 55
![Page 56: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/56.jpg)
Semivolatiles Prone to Peak Tailing
(App. Note 5990-3416EN)
Page 56
![Page 57: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/57.jpg)
Semivolatiles Prone to Peak Tailing
(App. Note 5990-3416EN)
Page 57
![Page 58: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/58.jpg)
Resolution of benzo-b & k fluoranthene isomers
19.00 19.50
0
100000
500000
1000000
1500000
Time-->
Abundance
Benzo-b-fluoranthene
Benzo-k-fluoranthene
Resolution of 1.2
As good as it gets…it only gets worse!
Page 58
![Page 59: GC Resolution Do You See What I See? · 2015. 7. 28. · SelectTM permanent gases Page 9 . Page 10 Optimizing Selectivity ... Ar/CH4 (P5) makeup gas at 60mL/min](https://reader036.vdocuments.us/reader036/viewer/2022071002/5fbec041b9be3e4d85225f8a/html5/thumbnails/59.jpg)
Conclusions for Better Resolution in GC
• Application Specific Stationary Phase - tuned for max R
• Smaller Diameter - loss in capacity, so small changes
• Longer, More Plates - big changes needed since it has a
small effect and it increases analysis times
• Optimum Carrier Gas Velocity - slowest carrier gas velocity
• 2D GC - more “chromatographic space” but adds complexity
• Multiple columns, inlets, detectors - single system with LTM
and Capillary Flow Technology
• Sharp Peaks - tailing, broadening, etc. loses resolution
Page 59