Gaseous Micro-Emboli Removal during priming procedure using pulsatile flow with 4 different adult oxygenators with integrated arterial filter and open reservoir. D. Hella 1 , M.G. Lagny 1 , T. Amand 1 , J.N. Koch 1 , J. Goffoy 1 , F. Blaffart 1 , J.O. Defraigne 2 . 1. ECCP, Department of Cardiovascular and Thoracic Surgery (Prof JO Defraigne), University Hospital of Liège, Liège, Belgium. 2. MD, PhD, Head of Department of Cardiovascular and Thoracic Surgery, University Hospital of Liège, Liège, Belgium. 1

Upload: dodien

Post on 18-Feb-2019

216 views

Category:

Documents


0 download

TRANSCRIPT

Gaseous Micro-Emboli Removal during priming procedure using pulsatile flow with 4 different adult oxygenators with integrated arterial filter and open reservoir. D. Hella1, M.G. Lagny1, T. Amand1, J.N. Koch1, J. Goffoy 1 , F. Blaffart1, J.O. Defraigne2. 1. ECCP, Department of Cardiovascular and Thoracic Surgery (Prof JO Defraigne), University Hospital of Liège, Liège, Belgium. 2. MD, PhD, Head of Department of Cardiovascular and Thoracic Surgery, University Hospital of Liège, Liège, Belgium.

1

Dominique Hella, ECCP

The normal physiological way our heart pumps blood throughout our body is using Pulsatile Flow.

2

Dominique Hella, ECCP

Tests in vitro: one of each type of oxygenator to know their limits with Pulsatile Flow Optimal setting for PF with S5 = 45/75/35 Recorded pressures: 4 negative peaks/10sec

3

Dominique Hella, ECCP

Pulsatile Flow during CPB => High velocity => High amounts of energy => Some negative pressures => ECC circuit must be stressed before CPB during priming procedure => Protection against Fluid leakage Rupture of the membrane Abnormal appearance of air in the circuit

4

Dominique Hella, ECCP

Priming protocole: 2 Liters of Plasmalyte A 37°C 200mmHg Dynamic occlusivity Continuous flow (CF) - 2 min (Gampt) Pulsatile flow (PF) - 2 min (Gampt)

5

Dominique Hella, ECCP

Roller pumps Stöckert S5 - Liva Nova 10 Priming procedures/ oxygenator 4 modern adult oxygenators: Affinity Fusion®- Medtronic n = 10 Capiox FX25® - Terumo n = 10 Inspire 8F® - Liva Nova n = 10 Quadrox i®- Maquet n = 10

6

Dominique Hella, ECCP

The Gampt BCC200:

7

Dominique Hella, ECCP

Data recorded by the Gampt BCC200: QF NumRed % VolRed% Number Volume

8

60

65

70

75

80

85

90

95

100

NumRed (%) CF NumRed

(%) PF VolRed (%) CF VolRed (%)

PF

77,9

90,2

71,7

88,8

94,8 96,1

92,3 84,5

100 100 100 100

100 99,6 100 100

GME reduction by oxygenator and flow mode

Quadrox i n=10 Inspire 8F n=10 Terumo FX25 n=10 Fusion Affinity n=10

Dominique Hella, ECCP

Comparison between oxygenators: p-value < 0,001 for all variables

9

10

0 500

1000 1500 2000 2500

7,5 65,5

2080 2028

42 92 20,5 47

Unit

Number IN

0

0,2

0,4

0,6

0,8

0,03 0,13

0,23 0,18 0,05

0,72

0,02 0,08

µl

Volume IN

0

20

40

60

0 1

51,5 40

4,5 10

0 0

Unit

Number OUT

0

0,02

0,04

0,06

0

0,02 0,01

0,02

0,04 0,05

0 0

µl

Volume OUT

GME number and volume by oxygenator and flow mode

Dominique Hella, ECCP

Continuous Pulsatile Continuous Pulsatile

11

0

0,01

0,02

0,03

0,04

0,05

Continuous flow Pulsatile flow

0

0,02

0,01

0,02

0,04

0,05

0 0

GME volume downstream oxygenator

µL

Dominique Hella, ECCP

Dominique Hella, ECCP

Conclusion: In vitro ≠ In vivo Venous reservoir: the more sensitive piece of the circuit? Gampt: built to be used with continuous flow => Prospective and randomized study: to show the airhandling of 5 oxygenators during CPB comparing continuous and pulsatile flow

12

Thank you for your attention !

13

Dominique Hella, ECCP