g-exj-1030713 may 2012 cardiac mri diagnostic backgrounder note: these slides are for use in...

59
G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables from these slides are to be used for another purpose (e.g. in printed materials), it is the individual’s responsibility to apply for the relevant permission. Specific local use requires local approval

Upload: marisol-hush

Post on 14-Dec-2015

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

CARDIAC MRI

Diagnostic Backgrounder

NOTE: These slides are for use in educational oral presentations only. If any published figures/tables from these slides are to be used for another purpose (e.g. in printed materials), it is the individual’s responsibility to apply for the relevant permission. Specific local use requires local approval

Page 2: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

2G-EXJ-1030713May 2012

Outline

● Introduction to iron overload

● Assessing cardiac iron loading– echocardiography– cardiac MRI

● Cardiac MRI in practice– preparation of the patient– acquisition of the image– analysis of the data

• Excel spreadsheet• ThalassaemiaTools (CMRtools)• cmr42

• FerriScan• MRmap• MATLAB

● Summary

MRI = magnetic resonance imaging.

Page 3: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Introduction to iron overload

Page 4: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

4G-EXJ-1030713May 2012

Introduction to iron overload

● Iron overload is common in patients who require intermittent or regular blood transfusions to treat anaemia and associated conditions

– it may be exacerbated in some conditions by excess gastrointestinal absorption of iron

● Iron overload can lead to considerable morbidity and mortality1

● Excess iron is deposited in major organs, resulting in organ damage

– the organs that are at risk of damage due to iron overload include the liver, heart, pancreas, thyroid, pituitary gland, and other endocrine organs2,3

1Ladis V, et al. Ann NY Acad Sci. 2005;1054:445-50. 2Gabutti V, Piga A. Acta Haematol. 1996;95:26-36. 3Olivieri NF. N Engl J Med. 1999;341:99-109.

Page 5: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

5G-EXJ-1030713May 2012

Importance of analysing cardiac iron

● In β-thalassaemia major, cardiac failure and arrhythmia are risk factors for mortality1

– signs of myocardial damage due to iron overload: arrhythmia, cardiomegaly, heart failure, and pericarditis2

– heart failure has been a major cause of death in β-thalassaemia patients in the past (50–70%)1,3

● In MDS, the results of studies are less comprehensible

– the reported proportion of MDS patients with cardiac iron overload is inconsistent; from high to only a small proportion of MDS patients4–7

– cardiac iron overload occurs later than does liver iron overload4,7,8

– however, cardiac iron overload can have serious clinical consequences in MDS patients

1Borgna-Pignatti C, et al. Haematologica. 2004;89:1187-93. 2Gabutti V, Piga A. Acta Haematol. 1996;95:26-36. 3. Modell B, et al. Lancet. 2000;355:2051-2. 4Jensen PD, et al. Blood. 2003;101:4632-9. 5Chacko J, et al. Br J Haematol. 2007;138:587-93. 6Konen E, et al. Am J Hematol. 2007;82:1013-6. 7Di Tucci AA, et al. Haematologica. 2008;93:1385-8. 8Buja LM, Roberts WC. Am J Med. 1971;51:209-21.

Page 6: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

6G-EXJ-1030713May 2012

BaselineLatest follow-up

p < 0.001

p < 0.001

cT2* ≤ 20 ms cT2* < 10 ms

Pat

ien

ts (

%)

cT2* = cardiac T2*.

1Thomas AS, et al. Blood. 2010;116:[abstract 1011]. 2Modell B, et al. Lancet. 2000;355:2051-2.

Importance of analysing cardiac iron (cont.)

● In 2010, the overall mortality rate of β-thalassaemia major patients in the UK was substantially lower than a decade ago (1.65 vs 4.3 per 1,000 patient years)1,2

– due to improved monitoring and management of iron overload over the last decade, 77% of patients have normal cardiac T2*1

– cardiac iron overload is no longer the leading cause of death in this population1

60

1723

7

0

10

20

30

40

50

60

70

Page 7: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

7G-EXJ-1030713May 2012

Cardiac T2*: Overview of correlations with other measurements

1Wood JC, et al. Blood. 2004;103:1934-6. 2Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 3Tanner MA, et al. J Cardiovasc Magn Reson. 2006;8:543-7. 4Kirk P, et al. Circulation. 2009;120:1961-8.5Westwood MA, et al. J Magn Reson Imaging. 2005;22:229-33.

†For thalassaemia, but not sickle cell.APFR = atrial peak filling rate; EPFR = early peak filling rate; LIC = liver iron concentration; SF = serum ferritin.

Weak or no correlation

Transfusion duration† ↑1

Ventricular dysfunction ↑1-3

Arrhythmia and heart failure ↑4

APFR↓ EPFR:APFR↑5

Need for cardiac medication↑1-2

T2*↓

SF and LIC1-3

Page 8: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

8G-EXJ-1030713May 2012

LVE

F (

%)

0

50

70

40

30

20

10

60

80

90

0 20 40 60 9080 10010 30 50 70Cardiac T2* (ms)

Cardiac T2* value of 37 ms in a normal heart

Cardiac T2* value of 4 ms in a significantly iron-overloaded heart

LVEF = left-ventricular ejection fraction. Anderson LJ, et al. Eur Heart J. 2001;22:2171-9.

Normal T2* range

Normal LVEF range

Cardiac T2*: Relationship with LVEF

Myocardial T2* values < 20 ms are associated with a progressive and significant decline in LVEF

Page 9: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

9G-EXJ-1030713May 2012

0.1

Cardiac T2*: Relationship with cardiac failure and arrhythmia

Kirk P, et al. Circulation. 2009;120:1961-8.

T2* < 10 ms: relative risk 159 (p < 0.001)T2* < 6 ms: relative risk 268 (p < 0.001)

Cardiac failure

Pro

po

rtio

n o

f p

atie

nts

d

evel

op

ing

car

dia

c fa

ilu

re

Follow-up time (days)

600 120 180 240 300 360

0.3

0.2

0

0.4

0.5

0.6< 6 ms

6–8 ms

8–10 ms

> 10 ms

Arrhythmia

600 120 180 240 300 360

0.15

0.10

0.05

0

0.20

0.25

0.30

< 10 ms

10–20 ms

> 20 ms

T2* < 20 ms: relative risk 4.6 (p < 0.001)T2* < 6 ms: relative risk 8.65 (p < 0.001)

Follow-up time (days)P

rop

ort

ion

of

pat

ien

ts w

ith

arr

hyt

hm

ia

Low myocardial T2* predicts a high risk of developing cardiac failure and arrhythmia

Page 10: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Assessing cardiac iron overload

Page 11: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

11G-EXJ-1030713May 2012

Assessing cardiac iron loading: Agenda

● Echocardiography

● Cardiac MRI

– advantages and disadvantages of cardiac MRI

– MRI: a non-invasive diagnostic tool

– T2* is the standard method for analysing cardiac iron

Page 12: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Echocardiography

Page 13: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

13G-EXJ-1030713May 2012

Assessing cardiac iron loading: Echocardiography

EF = ejection fraction.1Leonardi B, et al. JACC Cardiovasc Imaging. 2008;1:572-8. 2Hoffbrand AV. Eur Heart J. 2001;22:2140-1.

Pros Cons

• Readily available1

• Relatively inexpensive1

• Does not detect early damage2

• Echocardiographic diastolic function parameters correlate poorly with LVEF and T2*1

• Cannot directly or indirectly quantify cardiac iron levels

Page 14: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Cardiac MRI

Page 15: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

15G-EXJ-1030713May 2012

MRI: A non-invasive diagnostic tool

● Indirectly measures levels of iron in the heart

● MRI measures longitudinal (T1) and transverse (T2) relaxation times of the protons

– iron deposition disrupts the homogeneous magnetic field and shortens T1 and T2 times in a concentration-dependent manner

RF = radio-frequency.1Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96. 2Wood JC, et al. Circulation. 2005;112:535-43. 3Wang ZJ, et al. Radiology. 2005;234:749-55. 4Ghugre NR, et al. Magn Reson Med. 2006;56:681-6.

Protons

Magnetic field

RF/spin echo/gradient echo

Echo signal → T1, T2

Signal processing

Iron

Page 16: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

16G-EXJ-1030713May 2012

MRI: A non-invasive diagnostic tool (cont.)

● If a spin-echo sequence is used, the relaxation time is T2

● If a gradient-echo sequence is used, it is T2*

● Cardiac MRI methods

– gradient-echo T2* MRI: most used in clinical practice

– spin-echo T2 MRI: less useful (motion artefacts common due to characteristics of the heart)

TE = echo time.Adapted from Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96.

Protons

Magnetic field

Most used in clinical practice:

Gradient echo

Image acquired at different TEs

Excel or software

T2* [ms}

R2* [Hz]=1,000/T2*

Spin echo

Image acquired at different TEs

Excel or software

T2* [ms}

R2* [Hz]=1,000/T2*

Page 17: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

17G-EXJ-1030713May 2012

Assessing cardiac iron loading: Cardiac MRI

Advantages of MRI Disadvantages of MRI

• Non- invasive• Rapidly assesses iron content in the

septum of the heart• Relative iron burden can be

reproducibly estimated • Functional parameters can be

examined concurrently (e.g. LVEF)• Iron status of liver and heart can be

assessed in parallel• Allows longitudinal follow-up• Good correlation with morbidity

and mortality outcomes

• Indirect measurement of cardiac iron

• Requires MRI imager with dedicated imaging method

• Relatively expensive and varied availability

Page 18: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

18G-EXJ-1030713May 2012

What are sequences?

Sequences are a set of radio-frequency and gradient pulses (slight tilts in the magnetization curves of the scanner) generated repeatedly during the scan, which produce echoes with varied amplitudes and shapes that will define the MR image

What is gradient echo?

A gradient-echo sequence is obtained after 2 gradient impulses are applied to the body, resulting in a signal echo that is read by the coils. In these sequences, the spins are not refocused and, therefore, are subject to local inhomogeneities, with a more rapid decay curve. For gradient-echo pulse sequences, the T2* relaxation times (which reflect these inhomogeneities) on the signal are more significant

1Image from Ridgway JP. J Cardiovasc Magn Reson. 2010;12:71.

FAQ: Cardiac MRI

Page 19: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

19G-EXJ-1030713May 2012

Gradient relaxometry (T2*, R2*) is the method for analysing cardiac iron levels

1Guo H, et al. J Magn Reson Imaging. 2009;30:394-400. 2Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 3Wood JC, Noetzli L. Ann N Y Acad Sci. 2010;1202:173-9.4Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96. 5Westwood M, et al. J Magn Reson Imaging. 2003;18:33-9.6Hoffbrand AV. Eur Heart J. 2001;22:2140-1. 7He T, et al. Magn Reson Med. 2008;60:1082-9.

T2* (gradient echo) T2 (spin echo)

Pros • Greater sensitivity to iron deposition2

• Shorter acquisition time1

• Less affected by motion artefacts3

• More readily available3

• Easier to perform4

• Good reproducibility5

• Less affected by susceptibility artefacts1, due to metal implants, air–tissue interfaces, proximity to cardiac veins

Cons • More sensitive to static magnetic field inhomogeneity1

• Noise, motion, and blood artefacts can complicate analysis (particularly in heavily iron-loaded hearts)7

• Lack of sensitivity6

• Motion artefacts6

• Poor signal-to-background noise ratios at longer TEs6

• Longer acquisition time1

Page 20: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

20G-EXJ-1030713May 2012

HIC = hepatic iron concentrationCarpenter JP, et al. J Cardiovasc Magn Reson. 2009;11 Suppl 1:P224.Hankins et al Blood. 2009;113:4853-4855.

30

25

20

15

10

5

0

0 200 400 600 800 1000

Liver R2* (Hz)H

IC (

mg

Fe/

g o

f d

ry w

eig

ht

live

r)

Hankins, et al.

Wood, et al.

Anderson, et al.

[Fe]

(m

g/g

dry

wt)

Cardiac R2* (Hz)

0

2

4

6

8

10

12

14

0 100 200 300 400

R2 = 0.82540

Liver MRICardiac MRI

Gradient relaxometry (T2*, R2*) can conveniently measure cardiac and liver iron

Cardiac and liver iron can be assessed together conveniently by gradient echo during the a single MRI measurement.

Page 21: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

21G-EXJ-1030713May 2012

Cardiac T2* MRI is usually measured in the septum of the heart

Heart with normal iron levels

Heart with severe iron overload

Images courtesy of Dr J. de Lara Fernandes.

T2* = 22.8 ms or R2* = 43.9 Hz

T2* = 5.2 ms or R2* = 192 Hz

Page 22: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

22G-EXJ-1030713May 2012

Conversion from T2* to R2* is a simple mathematical calculation: R2* = 1,000/T2*

Level of cardiac iron overload T2*, ms R2*, Hz

Normal 201 < 50

Mild, moderate 10–201 50–100

Severe < 102 > 100

1Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 2Kirk P, et al. Circulation. 2009;120:1961-8.

These values are only applicable to 1.5 T scanners1

What is R2*?

Page 23: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

23G-EXJ-1030713May 2012

Why should the data be presented as R2* and not T2*?

● Seven whole hearts from patients with transfusion-dependent anaemias were assessed by histology and cardiac MRI

[Fe]

(m

g/g

dry

wt)

Cardiac T2* (ms)

[Fe]

(m

g/g

dry

wt)

0

2

4

6

8

10

12

14

0 10 20 30 40 50 60 70

R2 = 0.949

Cardiac R2* (Hz)

0

2

4

6

8

10

12

14

0 100 200 300 400

R2 = 0.82540

Carpenter JP, et al. J Cardiovasc Magn Reson. 2009;11 Suppl 1:P224.

R2* has a linear relationship with tissue iron concentration, which simplifies the interpretation of data

and allows comparison of changes over time

Page 24: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

24G-EXJ-1030713May 2012 Anderson LJ, et al. Eur Heart J. 2001;22:2171-9.

Hockey stick effect? Or a more gradual relationship?

The relationship between cardiac T2*/R2* and LVEF

Heart T2* (ms)

LVE

F (

%)

R2* (s–1)LV

EF

(%

)

90

80

70

60

50

40

30

20

10

0

10 20 30 40 50 60 70 80 90 1000

100

80

60

40

20

0

0 50 100 150 200 250

Why should the data be presented as R2* and not T2*? (cont.)

R2* allows demonstration of cardiac risk in a more gradual way

Page 25: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

25G-EXJ-1030713May 2012

Transform to R2*

Standard errors on a single measurement are approximately constant with R2*, but are non-uniform with T2*

Westwood M, et al. J Magn Reson Imaging. 2003;18:33-9.

60

50

40

30

20

10

00 10 20 30 40 50 60

T2* second measurement (ms)

T2*

fir

st m

easu

rem

ent

(ms)

120

100

80

60

40

20

00 20 40 60 80 100 120

R2* second measurement (s–1)

R2*

fir

st m

easu

rem

ent

(s–

1)

Why should the data be presented as R2* and not T2*? (cont.)

R2* has a constant standard error that makes assessment of the significance of changes easier

Page 26: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Cardiac T2* MRI in practice

Page 27: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

27G-EXJ-1030713May 2012

MRI scanners

● Manufacturers– Siemens Healthcare (Erlangen, Germany; www.siemensmedical.com)– GE Healthcare (Milwaukee, WI, USA; www.gemedicalsystems.com)– Philips Healthcare (Best, the Netherlands; www.medical.philips.com)

● Magnetic field– T2* varies with magnetic field strength1

– need 1.5 T for cutoff levels of 20 ms (iron overload) and 10 ms (severe iron overload)1,2

● Cardiac package– needs to be acquired separately from the manufacturers. The cost is

about USD 40,000. However, in most centres, this is available since MRI is frequently used in non-iron-related cardiovascular imaging

– includes all necessary for acquisition of the image – sequences are included in Siemens and Philips Healthcare cardiac

packages, but for GE Healthcare they need to be acquired separately (note: variations may exist between countries)

1Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 2Kirk P, et al. Circulation. 2009;120:1961-8.

Page 28: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

28G-EXJ-1030713May 2012

Cardiac T2* MRI in practice: The process

LIVER

TE ROI1.3 134

2.46 1143.62 994.78 815.94 70

7.1 598.26 499.42 40

10.58 3511.74 28

E 0.14925T2* 2.1 msR2* 476.1905 HzLIC 12.88801 mg/gLIC calculation according to: Hankins JS, et al. Blood. 2009;113:4853-5.

Normal >11.4 Light 3.8 - 11.4 Moderate 1.8-3.8 Severe <1.8 T2*Normal <88 Light 88-263 Moderate 263-555 Severe >555 R2*Normal <2 Light 2-7 Moderate 7-15 Severe >15 mg/g

y = 166.48552e-0.14925x

R² = 0.99845

0

20

40

60

80

100

120

140

160

0 2 4 6 8 10 12 14

Please insert the values of TE and ROI from an individual patient.

Please insert the value from the graph, encircled green.

T2*, R2*

*Time to manually calculate T2*/R2* values in an Excel spreadsheet depends on the experience of the physician.

1. Patientpreparation

(5 min)

2. Acquisition of the MRI image

(approx. 5-20 min)

3. Analysis ofMRI data

(time depends on experience*)

Page 29: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

29G-EXJ-1030713May 2012

Cardiac T2* MRI in practice: The process (cont.)

● Preparation of the patient

● Acquisition of the image

● Analysis of the data (post-processing)

• Excel spreadsheet

• ThalassaemiaTools, CMRtools

• cmr42

• FerriScan

• MRmap

• MATLAB

Page 30: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Preparation of the patient

Page 31: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

31G-EXJ-1030713May 2012

Preparation of the patient

● Standard precautions need to be taken

● There is no need for peripheral vein access since no contrast agent is required

● Special care

– remove all infusion/medication pumps (e.g. with insulin, pain-relieving drugs)

– stop continuous i.v. application of ICT during the measurement

– ECG signal should be positioned according to scanner specifications

ECG = electrocardiography.

Page 32: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

32G-EXJ-1030713May 2012

Cardiac T2* MRI in practice: The process (cont.)

● Preparation of the patient

● Acquisition of the image

● Analysis of the data (post-processing)

• Excel spreadsheet

• ThalassaemiaTools, CMRtools

• cmr42

• FerriScan

• MRmap

• MATLAB

Page 33: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Acquisition of the image

Page 34: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

34G-EXJ-1030713May 2012

Acquisition of the image: MRI pulse sequences

● Pulse sequences

– are a preselected set of defined radio-frequency and gradient pulses

– are computer programs that control all hardware aspects of the scan

– determine the order, spacing, and type of radio-frequency pulses that produce magnetic resonance images according to changes in the gradients of the magnetic field

● Several different pulse sequences exist1

– a gradient-echo sequence generates T2*

1Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96.

Page 35: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

35G-EXJ-1030713May 2012

1Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 2Westwood M, et al. J Magn Reson Imaging. 2003;18:33-9. 3He T, et al. J Magn Reson Imaging. 2007;25:1205-9. 4He T, et al. Magn Reson Med. 2008;60:1082-9. 5Pepe A, et al. J Magn Reson Imaging. 2006;23:662-8.

Sequence GroupNumber of echoes per breath-hold

Heart regions

Pre-puls

e

RR intervals

TR

Bright blood(Anderson et al.)1

London(Pennell)

1 (but multiple breath-holds)

1 (septum) No 1 Variable

Novel bright blood(Westwood et al)2

London(Pennell)

Multiple 1 (septum) No 1 Fixed

Black blood(He et al)3-4

London(Pennell)

Multiple 1 (septum) Yes 2 Fixed

Multi-slice(Pepe et al)5

Pisa(Pepe)

MultipleMulti-region

No 1 Fixed

The most common commercially available T2* acquisition techniques

The various techniques give clinically comparable results.2-3, 5

Page 36: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

36G-EXJ-1030713May 2012

Acquisition of the image: TEs

● Images are taken at a minimum of 5 different TEs, normally 8‒121

● The choice of minimum TE determines the smallest measurable T21

• ideally, min TE 2 ms, max TE 17‒20 ms

● Different T2* acquisition techniques according to TE

• multiple breath-hold: acquire an image for each TE in separate breath-holds2

• single breath-hold multi-echo acquisition: acquire images for all TE during 1 breath-hold3

Mean R2* compared with true value in the case of synthetic images for different minimum TEs,

but same echo duration (18 ms)4

1Wood JC, Noetzli L. Ann N Y Acad Sci. 2010;1202:173-9. 2Anderson LJ, et al. Eur Heart J. 2001;22:2171-9. 3Westwood M, et al. J Magn Reson Imaging. 2003;18:33-9. 4Ghugre NR, et al. J Magn Reson Imaging. 2006;23:9-16.

500

450

400

350

300

250

200

150

100

50

00 100 200 300 400 500

True R2* (Hz)M

ea

n R

2*:

ra

mp

, d

ua

lto

ne

, &

u

nif

orm

(H

z)

Shortest TE = 2 msShortest TE = 1 msShortest TE = 4 msShortest TE = 5.5 msTrue

Page 37: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

37G-EXJ-1030713May 2012

How does the MRI data output looks like?

Data visualizationMRI data output

1Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96.

During a single breath hold the pulse sequence run several times at increasing echo time (TE), generating data points

corresponding to decreased signal intensity1

Frame TE (ms) Mean ST

0 1.9 89.5

1 3.6 83.6

2 5.3 76.8

3 7.0 70.6

4 8.7 64.5

5 10.4 59.2

6 12.2 54.9

7 13.9 50.2

8 15.6 45.8

9 17.3 42.4

Page 38: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

38G-EXJ-1030713May 2012

Which is recommended: single or multiple breath-hold technique?

Comparison of the 2 methods, single and multiple breath-hold, showed no significant skewing between T2* values in all patients with -thalassaemia major, regardless of their T2* value (see Bland-Altman plots)1

However, in cardiac MRI the most recommended technique is single breath-hold, because it allows quick acquisition of the information. This is especially important to avoid movement artefacts (heart beating, breathing) and assure the good quality of the MRI image

1Westwood M, et al. J Magn Reson Imaging. 2003;18:33-9.

Patients with T2* < 20 ms1 Patients with T2* 20 ms 1

FAQ: Acquisition technique

Page 39: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

39G-EXJ-1030713May 2012

Acquisition of the image

● Single breath-hold multi-echo acquisition

– take a short-axis slice of the ventricle (halfway between the base and the apex): orange line

– image acquisition should occur immediately after the R wave

– do not alter any settings that could alter TE (e.g. FOV)

Image courtesy of Dr J. de Lara Fernandes.

Page 40: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

40G-EXJ-1030713May 2012

Cardiac T2* MRI in practice: The process (cont.)

● Preparation of the patient

● Acquisition of the image

● Analysis of the data (post-processing)

• Excel spreadsheet

• ThalassaemiaTools, CMRtools

• cmr42

• FerriScan

• MRmap

• MATLAB

Page 41: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Analysis of the data (post-processing)

Page 42: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

42G-EXJ-1030713May 2012

How T2* is calculated from the MRI output?

Data visualization

1Wood JC, Ghugre N. Hemoglobin. 2008;32:85-96.

Curve Fitting

T2*

Noise level

T2* calculation is fitting a curve on the data points and calculating at what echo time no signal is left from iron (only noise)1

Page 43: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

43G-EXJ-1030713May 2012

Analysis of the data

● The data can be analysed manually or using post-processing software

Manually Post-processing software

•Excel spreadsheet •ThalassaemiaTools (CMRtools)•cmr42

•FerriScan•MRmap•MATLAB

Page 44: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

44G-EXJ-1030713May 2012

Analysis of the data (cont.)

Method Pros Cons

Excel spreadsheet • Low cost • Time-consuming• Tedious

ThalassaemiaTools (CMRtools)1

• Fast (1 min)2

• Easy to use• FDA approved

• GBP 3,000 per year

cmr42(3) • Easy to use• FDA approved3

• Can generate T2*/R2* and T2/R2 maps with same software

• Allows different forms of analysis• Generates pixel-wise fitting with

colour maps

• 40,000 USD first year costs• 12,000 USD per year after

FDA = Food and Drug Administration.1www.cmrtools.com/cmrweb/ThalassaemiaToolsIntroduction.htm. Accessed Dec 2010. 2Pennell DJ. JACC Cardiovasc Imaging. 2008;1:579-81.3www.circlecvi.com. Accessed Dec 2010.

Page 45: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

45G-EXJ-1030713May 2012

Method Pros Cons

FerriScan1 • Centralized analysis of locally acquired data (206 active sites across 25 countries)

• Easy set-up on most MRI machines• EU approved• Validated on GE, Philips, and Siemens scanners

• USD 100 per scan • Patients data are sent to

reference centre

MRmap2 • Uses IDL runtime, which is a commercial software (less expensive than cmr42/CMRtools)

• Can quantify T1 and T2 map with the same software

• Purely a research tool• Not intended for diagnostic or

clinical use

MATLAB3 • Low cost • Available only locally • Physicists or engineers need to

write a MATLAB program for display and T2* measurement

1www.resonancehealth.com/resonance/ferriscan. Accessed Dec 2010. 2www.cmr-berlin.org/forschung/ mrmapengl/index.html. Accessed Dec 2010. 3Wood JC, Noetzli L. Ann N Y Acad Sci. 2010;1202:173-9.

Analysis of the data (cont.)

Page 46: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

46G-EXJ-1030713May 2012

What are the most common mistakes in analysing the data that could lead to a wrong interpretation of the T2* value?

Interpreting the data from cardiac MRI is usually quite straightforward; problems may arise when analysing data from patients with severe cardiac iron overload. In this case, the signal from heavily iron-loaded muscle will decay quickly and a single exponential decay curve does not fit the data well.1

Models exist that can help to solve this issue (see next slide):1. the offset model (Prof Wood and colleagues) 2. truncation of the data (Prof Pennell and colleagues)

Both models should give comparable results; the differences should not be clinically relevant

1Wood JC, Noetzli L. Ann N Y Acad Sci. 2010;1202:173-9. 2Ghugre NR, et al. J Magn Reson Imaging. 2006;23:9-16.

Signal decay curve from a patient with T2* ≈ 5 ms, showing that the data do not fit well2

FAQ: Mistakes in analysing the data

Page 47: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

47G-EXJ-1030713May 2012

What is truncation?

After the selection of the ROI, the signal decay can be fitted using different models. In the truncation model, the late points in the curve that form a plateau are subjectively discarded; the objective is to have a curve with an R2 > 0.995. A new single exponential curve is made by fitting the remaining signals.1 Generally, a truncation model should be used with the bright-blood technique to obtain more reproducible and more accurate T2* measurements1

What is an offset model?

The offset model consists of a single exponential with a constant offset. Using only the exponential model can underestimate the real T2* values (at quick signal loss at short TE, there is a plateau), while inclusion of the offset model into the fitting equation can improve this.2

Generally, the offset model is recommended to be used with the black-blood technique

1He T, et al. Magn Reson Med. 2008;60:1082-9. 2Ghugre NR, et al. J Magn Reson Imaging. 2006;23:9-16.

FAQ: Mistakes in analysing the data (cont.)

Page 48: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

48G-EXJ-1030713May 2012

How to start measuring cardiac iron loading in a hospital? What steps need to be taken?

To start assessing cardiac iron loading by MRI, these steps can be followed:1. Check MRI machine requirements

• 1.5 T• calibrated

2. Buy cardiac package from the manufacturer. It must include all that is necessary for acquisition of the data (the sequences are included with Siemens and Philips Healthcare cardiac packages, but for GE Healthcare they need to be acquired separately)

3. Optional: buy software for analysing the data (if not, Excel spreadsheet can be used)

4. Highly recommended: training of personnel for acquisition of cardiac MR images (e.g. functional analyses)

5. Highly recommended: training of personnel on how to analyse the data with the chosen software

FAQ: How to start measuring cardiac iron loading?

Page 49: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

49G-EXJ-1030713May 2012

Implementation of liver and cardiac MRI

1.5T MRI Scanner

Experienced radiologistExperienced radiologist

Cardiac acquisition package

Routine cardiac MR examsRoutine cardiac MR exams

Post-processing analysis

US$1.000.000

US$50.000US$40.000 or US$4.000/yor in-house or outsource

Yes

No

½ day training

1 day training

Yes

No

1-2 day training

4 day training

LiverAnalysis

LiverAnalysis

HeartAnalysis

HeartAnalysis

Slide presented at Global Iron Summit 2011 - With the permission of Juliano de Lara Fernandes

Page 50: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

Summary

Page 51: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

51G-EXJ-1030713May 2012

Summary

● Iron overload is common in patients who require intermittent or regular blood transfusions to treat anaemia and associated conditions

● Analysing cardiac iron levels is important

– in β-thalassaemia major, cardiac failure and arrhythmia are risk factors for mortality

– in MDS, cardiac iron overload can have serious clinical consequences

– due to improved monitoring and management of iron overload over the last decade, 77% of patients have normal cardiac T2*1

● MRI: the method to rapidly and effectively assess cardiac iron loading

– T2* allows specific assessment of cardiac iron levels. The use of this convenient, non-invasive procedure has had a significant impact on outcomes in patients with cardiac iron overload1

– R2* is a simple calculation from T2* and has a linear relationship with cardiac iron concentration

1Thomas AS, et al. Blood. 2010;116:[abstract 1011]. 2Modell B, et al. J Cardiovasc Magn Reson. 2008;10:42-9.

Page 52: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

G-EXJ-1030713May 2012

GLOSSARY OF TERMS

Page 53: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

53G-EXJ-1030713May 2012

GLOSSARY

● AML = acute myeloid leukemia

● APFR = Atrialp peak filling rate

● BA = basilar artery

● ß-TM = Beta Thalassemia Major

● ß-TI = Beta Thalassemia Intermedia

● BM = bone marrow

● BTM = bone marrow transplantation

● BW = bandwidth

● CFU = colony-forming unit

● CMML = chronic myelomonocytic leukemia

● CT2 = cardiac T2*.

● DAPI = 4',6-diamidino-2-phenylindole

Page 54: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

54G-EXJ-1030713May 2012

GLOSSARY

● DFS = = disease-free survival.

● DysE = dyserythropoiesis

● ECG = electrocardiography

● EDV = end-diastolic velocity

● EF = ejection fraction

● EPFR = early peak filling rate

● FatSat = fat saturation

● FAQ = frequently asked questions

● FDA = Food and Drug Administration

● FISH = fluorescence in situ hybridization.

● FOV = field of view

● GBP = Currency, pound sterling (£)

Page 55: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

55G-EXJ-1030713May 2012

GLOSSARY

● Hb = hemoglobin

● HbE = hemoglobin E

● HbF = fetal hemoglobin

● HbS = sickle cell hemoglobin.

● HbSS = sickle cell anemia.

● HIC = hepatic iron concentration

● HU = hydroxyurea

● ICA = internal carotid artery.

● ICT = iron chelation therapy

● IDL = interface description language

● IPSS = International Prognostic Scoring System

● iso = isochromosome

Page 56: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

56G-EXJ-1030713May 2012

GLOSSARY

● LIC = liver iron concentration

● LVEF = left-ventricular ejection fraction

● MCA = middle cerebral artery

● MDS = Myelodysplastic syndromes

● MDS-U = myelodysplastic syndrome, unclassified

● MRA = magnetic resonance angiography

● MRI = magnetic resonance imaging

● MV = mean velocity.

● N = neutropenia

● NEX = number of excitations

● NIH = National Institute of Health

● OS = overall survival

Page 57: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

57G-EXJ-1030713May 2012

GLOSSARY

● pB = peripheral blood

● PI = pulsatility index

● PSV = peak systolic Velocity

● RA =refractory anemia

● RAEB = refractory anemia with excess blasts

● RAEB -T = refractory anemia with excess blasts in transformation

● RARS = refractory anemia with ringed sideroblasts

● RBC = red blood cells

● RF = radio-frequency

● RCMD = refractory cytopenia with multilineage dysplasia

● RCMD-RS = refractory cytopenia with multilineage dysplasia with ringed sideroblasts

● RCUD = refractory cytopenia with unilineage dysplasia

Page 58: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

58G-EXJ-1030713May 2012

GLOSSARY

● RN = refractory neutropenia

● ROI = region of interest

● RT = refractory thrombocytopenia

● SCD = sickle cell disease

● SD = standard deviation

● SI = signal intensity

● SIR = signal intensity ratio

● SF = serum ferritin

● SNP-a = single-nucleotide polymorphism

● SQUID = superconducting quantum interface device.

● STOP = = Stroke Prevention Trial in Sickle Cell Anemia

● STOP II = Optimizing Primary Stroke Prevention in Sickle Cell Anemia

Page 59: G-EXJ-1030713 May 2012 CARDIAC MRI Diagnostic Backgrounder NOTE: These slides are for use in educational oral presentations only. If any published figures/tables

59G-EXJ-1030713May 2012

GLOSSARY

● T = thrombocytopenia

● TAMMV = time-averaged mean of the maximum velocity.

● TCCS = transcranial colour-coded sonography

● TCD = transcranial doppler ultrasonography

● TCDI = duplex (imaging TCD)

● TE = echo time

● TR = repetition time

● WHO = World Health Organization

● WPSS = WHO classification-based Prognostic Scoring System