Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea Fundamental and IVUS measurement Fundamental and IVUS measurement

Upload: others

Post on 07-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

Myeong-Ki Hong, M.D., Ph.DProfessor of Medicine

Department of Medicine,

University of Ulsan College of Medicine

Asan Medical Center, Seoul, Korea

Fundamental and IVUS measurementFundamental and

IVUS measurement

Page 2: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

We needed another methodWe needed another methodIndex procedure 1-year follow-up

Page 3: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

A

B

C

Cross Section RAO View LAO View

We needed another methodWe needed another method

Page 4: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

IVUS is a solution!IVUS is a solution!

Page 5: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 6: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 7: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Basic Physics of SoundBasic Physics of Sound

• Sound is a physical phenomenon that transfers energy from one point to another.

• Sound can pass only through matter.

• Ultrasound can be focused into small, well-defined beams that can probe the human body and interact with the tissue structures to form images

Page 8: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

PiezoelectricityPiezoelectricity• Piezo- : from piezein (Greek), sqeeze or press • Conversion electricity to sound and vice versa

When a piezoelectric crystal is placed in an electric field, or when charges are applied by external means to its faces, the crystal exhibits strain, i.e. the dimensions of the crystal change.

Page 9: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Ultrasound imaging processUltrasound imaging process

crystal

electric impulse

sound beam returning beam

electric impulse

Page 10: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Comparison of IVUS with other ultrasoundComparison of IVUS with other ultrasound

Technique Transducer size (cm)

Depth (cm)

Intervening tissues

Trans-thoracic > 2 ~3 – 20 Skin, fat, muscle

Trans-esophageal < 1.2 ~2 – 20 Esophagus,

atrium

Intravascular < 0.26 ~0.05 – 4 blood

Page 11: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

In early periodIn early period

Page 12: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

EquipmentEquipment

Mechanical systems Electronic systems

Page 13: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Image qualityImage quality

• Spatial resolution

- The ability to discriminate small objects

within the ultrasound image

- Axial : parallel to the beam

- Lateral : perpendicular to both the beam

and the catheter

Page 14: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 15: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Image Acquisitionpullback method

Image Acquisitionpullback method

• Manual pullback- to concentrate on specific regions of

interest- But, possible to skipping over and not to

perform precise measurement- not reproducible

• Motorized pullback- precise, reproducible- to reconstruct image (L-mode, 3D)

Page 16: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Image AcquisitionLongitudinal display (L-mode)Image Acquisition

Longitudinal display (L-mode)• For spatial orientation, assessment of length,

and distribution of plaque

Page 17: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

IVUS artifactsnon-uniform rotational distortion (NURD)

IVUS artifactsnon-uniform rotational distortion (NURD)

Full sector NURD Isolated sector NURD

Page 18: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

IVUS artifactsair bubble

IVUS artifactsair bubble

Page 19: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 20: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Histology and IVUSHistology and IVUS

Adventitia

IVUS catheterNormal intima

Ring-down artifact

Page 21: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Histology and IVUSHistology and IVUS

Lipid pool Echolucent intimal areaFibrous cap

Intimal thickening

Page 22: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

The Three-Layered Appearanceborder identification

The Three-Layered Appearanceborder identification

IntimaMediaAdventitia

Page 23: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

The Three-Layered Appearanceborder identification

The Three-Layered Appearanceborder identification

1. Adventitia : the outer covering of the artery

2. Media : the actual wall of the artery

3. Intima : a layer of endothelial and other cells that make direct contact with the blood inside the artery

4. Lumen : the actual open channel of the artery through which the blood flows.

Page 24: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 25: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementlumen measurements

Quantitative measurementlumen measurements

Page 26: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementlumen measurements

Quantitative measurementlumen measurements

Page 27: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementatheroma measurements

Quantitative measurementatheroma measurements

Page 28: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementatheroma measurements

Quantitative measurementatheroma measurements

Atheroma

= EEM

Page 29: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementatheroma measurements

Quantitative measurementatheroma measurements

Atheroma

= EEM - Lumen

Page 30: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementatheroma measurements

Quantitative measurementatheroma measurements

Atheroma eccentricity

= A – B / A

Page 31: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementatheroma measurements

Quantitative measurementatheroma measurements

Atheroma burden

= C – D / C

Page 32: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementstent measurements

Quantitative measurementstent measurements

Page 33: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementstent measurements

Quantitative measurementstent measurements

Page 34: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Dicotomous Classification of Remodeling

Positiveremodeling

Intermediateremodeling

Negativeremodeling

Proximal reference Lesion Distal reference

Page 35: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Lesion/Reference EEM CSA

% of lesionsInadequate arterial remodeling: 15%

Contribution of inadequate arterial remodeling to the development of focal coronary artery stenoses: an IVUS study

Mintz GS, et al. Circulation 1997;95: 1791-98

00

5

10

15

20

25

0.5 1.0 1.5 2.0

Page 36: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

EEM CSA: 9.0 mm2

Lumen CSA: 2.6 mm2

EEM CSA: 9.0 mm2

Lumen CSA: 2.6 mm2EEM CSA: 21.2 mm2

Lumen CSA: 14.2 mm2

EEM CSA: 21.2 mm2

Lumen CSA: 14.2 mm2STENT CSA: 10.9 mm2STENT CSA: 10.9 mm2

Negative remodelingNegative remodelingStenting aloneStenting alone

Page 37: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

EEM CSA: 23.9 mm2

Lumen CSA: 2.3 mm2

EEM CSA: 23.9 mm2

Lumen CSA: 2.3 mm2 STENT CSA 15.4 mm2STENT CSA 15.4 mm2EEM CSA: 18.6 mm2

Lumen CSA: 10.3 mm2

EEM CSA: 18.6 mm2

Lumen CSA: 10.3 mm2

Positive remodelingPositive remodeling

Stenting After DCAStenting After DCA

Page 38: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Quantitative measurementnegative remodeling

Quantitative measurementnegative remodeling

Page 39: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma MorphologyNormal

Atheroma MorphologyNormal

Page 40: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma MorphologySoft plaque

Atheroma MorphologySoft plaque

Page 41: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma Morphologyfibrotic plaque

Atheroma Morphologyfibrotic plaque

Page 42: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma MorphologyCalcium

Atheroma MorphologyCalcium

Page 43: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma Morphologyrupture

Atheroma Morphologyrupture

Page 44: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma Morphologylipid core

Atheroma Morphologylipid core

Page 45: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Atheroma Morphologythrombus

Atheroma Morphologythrombus

Page 46: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

ContentsContents

Basic Physics and Equipment

Image Acquisition and IVUS Artifacts

Histology

Quantitative and Qualitative Assessment

Reporting

Page 47: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Reporting of IVUS resultsReporting of IVUS results

Appropriate patient demographic information and date

Indication and brief description of procedure

Basic findings : MLD, minimum stent area, or plaque

burden…

Plaque features : dissection, calcium, or thrombus…

Changes of therapy by IVUS

IVUS-related complications and any consequent

therapy.ACC Clinical Expert Consensus. J Am Coll Cardiol. 2001

Page 48: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

In my opinion ….

When you meet the complex cases, IVUS will help you at any time and answer you clearly about difficult questions

Page 49: Fundamental and IVUS measurement · Myeong-Ki Hong, M.D., Ph.D Professor of Medicine Department of Medicine, University of Ulsan College of Medicine Asan Medical Center, Seoul, Korea

ASAN MEDICAL CENTER

Image qualityImage quality

• Contrast resolution

- the distribution of the gray scale of the

reflected signal and is often referred to as

dynamic range