    Purpose

    The purpose of this procedure is to provide instruction for the calibration of digital

    multimeters (DMM).

    Scope

    All multimeters (multi-range instruments used to measure voltage, current, and

    resistance) are to be verified in calibration no less than once per year.

    Procedure

    Preliminary Instructions and Notes

    1. Read this entire procedure before beginning the calibration.

    2. Calibration shall be performed in an environment that conforms to Manufacturer

    Specifications.

    Download Calibration Control

    Email: *

    Name: *

    Company: *

    Day Phone:

    Sign In & Download Software

    sign in | register | contact us | cart

    phone: 559-440-6052

    site search

    pay your invoice or quote online

    site home| calibration procedures| digital multimeter

    Digital Multimeter Calibration

    Home Calibration Control Get Help Procedures

    http://www.apesoftware.com/default.aspxhttp://www.apesoftware.com/calibration-procedures/default.aspxhttp://www.apesoftware.com/calibration-procedures/digital-multimeter.aspxhttp://www.apesoftware.com/http://www.apesoftware.com/http://www.apesoftware.com/calibration-control/download.aspxhttp://www.apesoftware.com/calibration-procedures/http://www.apesoftware.com/calibration-control/help/http://www.apesoftware.com/calibration-controlhttp://www.apesoftware.com/http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspxhttp://www.apesoftware.com/calibration-procedures/default.aspxhttp://www.apesoftware.com/default.aspxhttp://www.apesoftware.com/invoice/http://www.apesoftware.com/search.aspxhttp://www.apesoftware.com/about/contact.aspxhttp://www.apesoftware.com/cart/http://www.apesoftware.com/about/contact.aspxhttp://__dopostback%28%27ctl00%24top1%24lbtnregister%27%2C%27%27%29/http://www.apesoftware.com/account/signin.aspx?returnurl=/calibration-procedures/digital-multimeter.aspx?http://www.apesoftware.com/calibration-control/download.aspxhttp://www.apesoftware.com/
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    2/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 2/12

    3. The digital multimeter will hereafter be referred to as the Instrument Under Test

    (IUT).

    4. Verify that the IUT is clean.

    5. Visually examine the IUT for any condition that could cause errors in the calibration.

    6. If any of the requirements cannot be met, refer to the applicable manufacturer

    manual.

    7. If a malfunction occurs or a defect is observed while calibration is in progress, the

    calibration shall be discontinued and necessary corrective action taken if corrective

    action affects a measurement function previously calibrated, the function shall be

    recalibrated before the remainder of the procedure implemented.

    Reference Material

    Applicable Manufacturers Manual or Brochures.

    Specifications

    The specifications of the IUT are determined by the applicable manufacturer's

    documentation. If the manufacturer's documentation is not available, then the

    specifications identified in this procedure are used.

    Equipment Required

    The Standards listed below should be selected on the basis of their higher accuracy level

    when compared to the unit under test. Equivalent Standards must be equal to or better

    than the Minimum-Use-Specification.

    http://www.apesoftware.com/calibration-control/help/thum.aspxhttp://www.apesoftware.com/calibration-control/help/permissions.aspx
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    3/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 3/12

    Minimum-Use-Specifications for Standards listed are 1/4 the accuracy required by the

    IUT.

    Rotek Model 2500 Calibrator or equivalent

    Diode Test Fixture

    Test Leads

    Continuity Test Fixture

    Capacitance Test Fixture

    Thermometer

    Hydrometer

    Set-Up

    Turn all power on (calibrator and the unit to be calibrated). Allow the instruments to

    stabilize for approximately 5 minutes. Conduct the tests in an ambient temperature of 25

    5C and a relative humidity of less than 80%.

    Note: If the IUT has an automatic ranging feature, then the function being checked (i.e.

    voltage, current, or resistance) must be verified in the variable range mode as well.

    Battery Replacement

    Replace the battery in the IUT and make sure the battery contacts are not dirty.

    Detailed Procedure

    DISPLAY AND SWITCH TEST

    http://www.apesoftware.com/calibration-control/help/calibration-event.aspxhttp://www.apesoftware.com/articles/barcode-scanner.aspx
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    4/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 4/12

    Turn the IUT on and verify that all LCD segments are working and not dim. Verify that the

    selector switch(s) is/are working properly. Make sure the Low Battery indicator is not

    showing.

    RESISTANCE CHECK

    Connect the calibrator between the V/ohm and common input terminals of the IUT. Use

    the following chart to verify the resistance values.

    Step Range Input Display

    1 200 ohm Short 00.0 to 00.5

    2 2 k ohm Short 0.000 to 0.001

    3 2 k ohm 1 k ohm .998 to 1.002

    4 20 k ohm 10 k ohm 9,98 to 10.02

    5 200 k ohm 100 k ohm 99.8 to 100.2

    6 2000 k ohm 1 M ohm 998 to 1002

    7 2000 M ohm Open 0.10 to 00.0

    CONTINUITY TEST

    1. Select the continuity check function on the calibrator.

    2. Connect the test leads to the V/ohm and common terminals of the IUT.

    3. Momentarily short the test leads together and observe that the tone sounds.

    http://www.apesoftware.com/calibration-control/help/measurement-templates.aspxhttp://www.apesoftware.com/calibration-control/help/grid-features.aspxhttp://www.apesoftware.com/calibration-control/help/calibration-event.aspx
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    5/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 5/12

    4. Connect the test leads to the continuity test fixture between the 100 ohm test

    points. No tone should be heard indicating non-terminating connection.

    5. Connect the test leads to the continuity test fixture between the 50 ohm set up. The

    test tone should indicate continuity between the test points.

    DC VOLTAGE TEST

    1. Use the calibrator to supply the correct DC voltage listed below.

    2. Set the calibrator for a zero volt input.

    3. Connect the calibrator output to the V/ohm and common input terminals of the IUT.

    Note: Connect the ground/common/low side of the Rotek 2500 to the common on

    the IUT.

    4. With reference to the table below, select the IUT voltage range and set the

    calibrator output to the corresponding IUT input voltage. Test per the following

    chart and verify that the display falls within the limits given.

    Step Range Input Display

    1 200mV 190mV 189.7 to 190.3

    2 200mV -190mV -189.7 to -190.3

    3 2V 0.0V -0.001 to 0.001

    4 2V 1.9V 1.897 to 1.903

    5 20V 19V 18.97 to 19.03

    http://www.apesoftware.com/calibration-control/help/status-change.aspxhttp://www.apesoftware.com/calibration-control/help/auto-notify.aspx
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    6/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 6/12

    6 200V 100V 99.84 to 100.16

    7 1000V 400V 399.3 to 400.6

    AC VOLTAGE TEST

    Use the calibrator to supply the correct AC voltage in the following test:

    1. Set the calibrator to 60 Hz.

    2. Connect the calibrator output to the V/ohm and common input terminals of the IUT.

    Connect the ground/common/low side of the calibrator to common on the IUT.

    3. With reference to the table below, select the IUT voltage range given in step 1 and

    set the calibrator output to the corresponding IUT input voltage. Verify that the

    display reading is within the limits shown.

    Step Range Input Display

    1 200mV 100mV 99.0 to 101.0

    2 2V 1V 0.990 to 1.010

    3 20V 10V 9.90

    4 200V 100V 99.0 to 100.1

    5 750V 400V 294 to 406

    DC CURRENT TEST

    Use the calibrator to supply the correct DC current listed in the table below:

    http://www.apesoftware.com/calibration-control/help/attachments-pictures.aspx
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    7/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 7/12

    1. Set the output of the calibrator to zero mA.

    2. Connect the output of the calibrator to the lowest current terminal (typically 200

    mA) and common input terminal on the IUT.

    Note: Do not exceed the maximum current rating of the meter. Only test for

    values the meter is capable of reading.

    3. With reference to the table below, select the IUT current range and set the

    calibration output to provide the corresponding IUT input current. Verify that the

    display reading is within the limits shown.

    Step Range Input Display

    1 200?A 190?A 189.7 to 192.3

    2 2mA 1.9mA 1.897 to 1.923

    3 20mA 19mA 18.97 to 19.23

    4 200mA 190mA 189.7 to 192.3

    4. Disconnect the calibrator and change the current terminal on the IUT to the highest

    setting (typically 10 A) leaving the common input terminal connected.

    5. Use the following table to verify that the meter is functioning within the specifiedparameters.

    Step Range Input Display

    1 10A 1A 0.98 to 1.02

    2 10A 5A 4.90 to 5.10

  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    8/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 8/12

    3 10A 9A 8.82 to 9.18

    AC CURRENT TEST

    Use the calibrator to supply the correct AC current listed in the table below:

    1. Use the following table to verify that the meter is functioning within the specifiedparameters.

    2. Set the output of the calibrator to zero mA.

    3. Connect the output of the calibrator to the low current input terminal (typically 200

    mA) and the common terminal of the unit under test.

    Note:Do not exceed the maximum current rating of the meter. Only test for

    values the meter is capable of reading.

    4. With reference to the table below, select the IUT current range and set the

    calibrator output to provide the corresponding IUT input current. Verify that the

    display reading is within the limits shown.

    Step Range Input Display

    1 200?A 190?A 18.27 to 19.28

    2 2mA 1.9mA 1.872 to 1.928

    3 20mA 19mA 18.72 to 19.28

    4 200mA 190mA 187.2 to 192.8

    5. Disconnect the calibrator and change the current terminal on the IUT to the highest

  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    9/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 9/12

    setting (typically 10 A) leaving the common input terminal connected.

    6. Use the following table to verify that the meter is functioning within the specified

    parameters.

    Step Range Input Display

    1 10A 1A 0.97 to 1.03

    2 10A 5A 4.87 to 5.13

    3 10A 9A 8.77 to 9.23

    LOGIC TEST

    Use this procedure to test weather the logic function is indicating the correct state.

    1. Set the IUT to the logic test mode and insert the test leads into the V/ohm and

    common terminals.

    2. Short the leads together and verify that the meter indicates a low logic state.

    3. Set the calibrator to 0.400 VDC and connect the meter to the output terminals.

    Verify that the meter indicates a low logic state.

    4. Set the calibrator to 3.500 VDC and verify that the IUT indicates a high logic state.

    FREQUENCY TEST

    1. Connect an oscilloscope to a function generator to measure AC sine waves.

    2. Set the function generator to output 10 VAC and use the oscilloscope to set the

    frequencies from the following table.

  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    10/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 10/12

    3. Connect the IUT leads to the V/ohm and common terminals on the IUT and to the

    ground and output of the function generator.

    4. Verify the meter is displaying proper results from the table.

    Step Range Input Display

    1 Hz 60 Hz @ 10V 59.4 to 60.6

    2 Hz 1 kHz @ 10V .990 to 1.010k

    3 Hz 1 MHz @ 10V .990 to 1.010M

    CAPACITANCE TEST

    1. Set the IUT to measure capacitance, insert the test leads into the V/ohm and

    common test terminals.

    2. Connect the meter to the capacitance test fixture and use the following table to

    determine the ability of the IUT.

    Step Range Input Display

    1 Capacitance 10 micro F 9.85 to 10.15

    2 Capacitance 1 micro F .985 to 1.015

    3 Capacitance 100 micro F 98.5 to 101.5

    DIODE TEST

    1. Set the IUT to the diode check function and connect the test leads to the V/ohm

  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    11/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 11/12

    and common terminals.

    2. Connect the test leads to the diode test fixture and follow the steps in the

    proceeding table to verify diode test operation.

    Step Range Setup Polarity Display Tone

    1 Diode Check Short N/A .000 to .001 Yes

    2 Diode Check Si Forward .800 to .600 Yes

    3 Diode Check Si Reverse 1. No

    4 D iode C heck Ge Forw ard .300 to .500 Yes

    5 Diode Check Ge Reverse 1. No

    Note: Some meters are not designed to emit a tone.

    Reason for Reissue

    Revision A - First Release

    Software

    Calibration Control

    Download

    Version History

    Feature List

    Software

    Printers

    Scanner

    Minimum Requirements

    PDF Brochure

    Get Help

    Help Pages & Videos

    Remote Support

    Online Meeting

    On-Site Services

    Forum

    Forum Home

    Calibration Control

    Feedback

    Twitter

    About

    About Us

    Our Customers

    Important Links

    Site Map

    Contact Us

    Contact Us

    Email Sales

    Email Support

    Desktop Support

    http://www.apesoftware.com/about/teamviewer.aspxmailto:[email protected]?subject=Support%20Inquirymailto:[email protected]?subject=Sales%20Inquiryhttp://www.apesoftware.com/about/contact.aspxhttp://www.apesoftware.com/about/site-map.aspxhttp://www.apesoftware.com/about/links.aspxhttp://www.apesoftware.com/about/customers.aspxhttp://www.apesoftware.com/about/http://www.apesoftware.com/about/twitter.aspxhttp://www.apesoftware.com/forum/?f=6http://www.apesoftware.com/forum/?f=8http://www.apesoftware.com/forum/http://www.apesoftware.com/consulting/on-site-training-consulting.aspxhttp://www.apesoftware.com/about/meeting.aspxhttp://www.apesoftware.com/about/teamviewer.aspxhttp://www.apesoftware.com/calibration-control/help/http://www.apesoftware.com/calibration-control/brochure-calibration-control.pdfhttp://www.apesoftware.com/calibration-control/minimum-requirements.aspxhttp://www.apesoftware.com/articles/barcode-scanner.aspxhttp://www.apesoftware.com/calibration-control/help/label-printers.aspxhttp://www.apesoftware.com/calibration-control/program-features.aspxhttp://www.apesoftware.com/calibration-control/version-history.aspxhttp://www.apesoftware.com/calibration-control/download.aspxhttp://www.apesoftware.com/calibration-control/
  • 7/25/2019 Free Digital Multimeter Calibration Procedure

    12/12

    23/01/2016 Free Digital Multimeter Calibration Procedure

    http://www.apesoftware.com/calibration-procedures/digital-multimeter.aspx 12/12

    Pricing PDF Manual Multi-User Setup Privacy Policy

    Copyr ight 1992 to 2016 Ape Software all rights reserv ed | Home | Contact Phone: 559-440-6052

    http://www.apesoftware.com/about/contact.aspxhttp://www.apesoftware.com/about/contact.aspxhttp://www.apesoftware.com/http://www.apesoftware.com/about/privacy.aspxhttp://www.apesoftware.com/calibration-control/help/multiple-users.aspxhttp://www.apesoftware.com/calibration-control/calibration-control-manual-v83.pdfhttp://www.apesoftware.com/calibration-control/pricing.aspx