For Angular Stable Locking of Intramedullary Nails
Angular Stable Locking System (ASLS)Surgical Technique
Introduction
Surgical Technique
Product Information
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 1
Table of Contents
Angular Stable Locking System (ASLS) 2
AO Principles 4
Indications 5
Preparation 6
Targeted Locking 8
Freehand Locking 14
Screw Insertion 21
Implant Removal 27
Implants 28
ASLS Compatibility With Existing Nails 29
Instruments 30
Set Lists 33
MR Information The Angular Stable Locking System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. The safety of the Angular Stable Locking System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.
Image intensifier control
Angular Stable Locking System (ASLS)
Diameter 2: Expands sleeve, providing angular stability
Diameter 1: Provides purchase in reamed near cortex
Angular and axial stability– Improved fixation in poor quality bone
– Decreased risk of secondary loss of reduction
Compatibility with existing nails– Compatible with all DePuy Synthes
cannulated titanium nails (see page 29 for nail compatibility)
– A combination of standard and ASLS screws may be used in the same nail
– Intraoperative choice for ASLS
Titanium Angular Stable Locking Screws– Titanium alloy*
– Available lengths – 4.0 mm ASLS screw: 26 mm–80 mm
– 5.0 mm ASLS screw: 30 mm–100 mm
– 6.0 mm ASLS screw: 32 mm–125 mm
– Instruments are color coded to screws
4.0 mm ASLS (brown)
5.0 mm ASLS (light blue)
6.0 mm ASLS (dark purple)
*Titanium-6% aluminum-7% niobium alloy
2 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Angular Stable Locking System (ASLS)
Diameter 3: Holds unexpanded sleeve for screw insertion, provides purchase in far cortex
Technique overview
1. Drill both cortices
2. Ream near cortex
3. Insert unexpanded sleeve into nail
4. Expanded sleeve provides angular stability
Resorbable Sleeve– 70:30 poly (L /DL-lactide)– 4.0 mm, 5.0 mm and 6.0 mm sleeves
are used with corresponding ASLS screws
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 3
4 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
AO Principles
1
4
2
3
4_Priciples_03.pdf 1 05.07.12 12:08
4 DePuy Synthes Expert Lateral Femoral Nail Surgical Technique
AO PRINCIPLES
In 1958, the AO formulated four basic principles, which have become the guidelines for internal fixation1, 2.
1 Müller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991.
2 Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management. 2nd ed. Stuttgart, New York: Thieme. 2007.
Anatomic reductionFracture reduction and fixation to restore anatomical relationships.
Early, active mobilizationEarly and safe mobilization and rehabilitation of the injured part and the patient as a whole.
Stable fixationFracture fixation providing abso-lute or relative stability, as required by the patient, the injury, and the personality of the fracture.
Preservation of blood supplyPreservation of the blood supply to soft tissues and bone by gentle reduction techniques and careful handling.
In 1958, the AO formulated four basic principles, which have become the guidelines for internal fixation.1,2
Anatomic reductionFracture reduction and fixation to restore anatomical relationships.
Early, active mobilizationEarly and safe mobilization and rehabilitation of the injured part and the patient as a whole.
Stable fixationFracture fixation providing absolute or relative stability, as required by the patient, the injury, and the personality of the fracture.
Preservation of blood supplyPreservation of the blood supply to soft tissues and bone by gentle reduction techniques and careful handling.
1. Müller ME, Allgöwer M, Schneider R, Willenegger H. Manual of Internal Fixation. 3rd ed. Berlin, Heidelberg, New York: Springer-Verlag; 1991.
2. Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management. 2nd ed. Stuttgart New York: Thieme; 2007.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 5
Indications
The Synthes Angular Stable Locking System (ASLS), when used with Synthes cannulated titanium intramedullary nails, is intended to aid in the alignment and stabilization of tibial, humeral, femoral and ankle fractures.
6 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Preparation
The following pages describe techniques for locking with ASLS. All illustrations use the Tibial Nail-EX, as an example.
1Preoperative planning
Use the appropriate AO preoperative planner template to estimate nail diameter and length.
2Reduce fracture and insert nail
Reduce the fracture and insert the nail. Use a compatible DePuy Synthes titanium cannulated intramedullary nailing system according to the relevant technique guide.
Note: The ASLS is only compatible with cannulated nails. It is necessary for the sleeve to expand within the cannulation of the intramedullary nail to provide sufficient stability.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 7
3Choose appropriate locking screws
For each locking hole, preselect the optimal locking implant to stabilize the fracture. If using an ASLS screw, the diameter of the ASLS screw and instruments will match the diameter of the standard screw/bolt for the particular nail.
Note: A combination of standard locking screws/bolts and ASLS screws can be used in the same nail.
Caution: Do not exchange a standard locking screw/bolt for an ASLS screw or vice versa as this may reduce screw-to-bone interface stability.
Preparation
8 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Targeted Locking
1Attach aiming arm and insert trocar combination
Instruments
03.010.063† 12.0 mm/8.0 mm Protection Sleeve, 188 mm (for Expert nails)
03.025.040* 11.0 mm/8.0 mm Protection Sleeve, 188 mm, for ASLS
4.0 mm
03.010.064† 8.0 mm/3.2 mm Drill Sleeve
03.010.069† 3.2 mm Trocar
5.0 mm
03.010.065† 8.0 mm/4.2 mm Drill Sleeve
03.010.070† 4.2 mm Trocar
6.0 mm
03.010.066† 8.0 mm/5.0 mm Drill Sleeve
03.010.071† 5.0 mm Trocar
Once the nail is inserted, attach the aiming arm to the insertion handle. Confirm that the nail is tightly connected to the insertion handle.
Insert the appropriate three-part trocar combination (protection sleeve, drill sleeve and trocar) through the desired hole in the aiming arm. Make a skin incision and ensure that the dissection of the fascia is in line with the path of the protection sleeve. Advance the three-part trocar combination to the bone. Remove the trocar.
Note: Ensure that the drill sleeve is pressed firmly to the near cortex.
†Found in the EXPERT NAIL® Instrument Sets.*Also available.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 9
2Drill through both cortices
Instruments
4.0 mm
03.025.104 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 331 mm
5.0 mm
03.025.105 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 331 mm
6.0 mm
03.025.106 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 331 mm
Using the appropriate drill bit, drill through both cortices until the tip of the drill bit penetrates the far cortex.
If necessary, use image intensification to control the position of the drill bit.
Note: Follow all recommended techniques for screw/bolt insertion for the relevant nail system. When inserting a screw monocortically, stop drilling immediately after penetrating the near cortex.
Targeted Locking
11 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
3Determine ASLS screw length
Read the measurement from the calibrated drill bit at the back of the drill sleeve. Select the ASLS screw length accordingly. Remove the drill bit and drill sleeve.
Note: Follow all recommended techniques for screw/bolt measurement for the relevant nail system.
Precaution: Do not exchange a standard locking screw/bolt for an ASLS screw or vice versa as this may reduce screw-to-bone interface stability.
Targeted Locking
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 11
Alternative technique
Instrument
03.025.052 Depth Gauge, for ASLS screws
Remove the drill bit and the drill sleeve. Disassemble the depth gauge, separating the outer sleeve from the measuring device with hook. Insert the measuring device with hook into the protection sleeve.
Read the measurement from the back of the protectionsleeve and select the ASLS screw length accordingly.
Note: Ensure that the hook grasps the far cortex and that the protection sleeve is pressed firmly to the bone.
Targeted Locking
12 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Targeted Locking
4Ream near cortex
Instruments
4.0 mm
03.025.030 2.9 mm/5.0 mm Hand Reamer
5.0 mm
03.025.031 3.9 mm/6.0 mm Hand Reamer
6.0 mm
03.025.032 4.9 mm/7.0 mm Hand Reamer
Align the hand reamer with the locking hole axis. Ream the near cortex and clear the passage down to the nail. Remove the hand reamer.
If necessary, use image intensification to control the position of the hand reamer.
Note: To correctly insert the ASLS sleeve, ensure: – the near cortex is completely reamed; – the passage from the near cortex to the nail’s locking hole is cleared; – the drilled/reamed hole aligns with the nail’s locking hole.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 13
Alternative technique
Instruments
4.0 mm
03.025.134 2.9 mm/5.0 mm Drill-Reamer, quick coupling, 268 mm
5.0 mm
03.025.135 3.9 mm/6.0 mm Drill-Reamer, quick coupling, 288 mm
6.0 mm
03.025.136 4.9 mm/7.0 mm Drill-Reamer, quick coupling, 288 mm
Use a power-driven drill-reamer to open the near cortex. Align the drill-reamer with the locking hole axis and ream the near cortex. Once the large diameter of the drill-reamer passes through the near cortex, stop reaming immediately. Remove the drill-reamer.
If necessary, clear the passage down to the nail with the hand reamer.
If necessary, use image intensification to control the positionof the drill-reamer.
Precaution: Ensure that the drill-reamer does not damage the nail.
Insert screws in targeted locking holes (see Screw Insertion, page 21), before continuing with freehand locking.
Targeted Locking
14 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Freehand Locking
2Make incision
Using image intensification, place a long K-wire or alternative long instrument on the skin over the center of the hole to mark the incision point.
Make a skin incision with a scalpel blade at this location.
Note: Check the reduction, correct alignment of the fragments, and limb length before locking.
1Align image
The techniques for distal locking apply. Align the image intensifier with the desired locking hole until a perfect circle is visible in the center of the screen.
Note: Ensure that the image displayed on the image intensifier mirrors the actual image in all axes.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 15
3Drill through both cortices
Instruments
511.30* Radiolucent Drive
4.0 mm
03.025.124* 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for Radiolucent Drive
5.0 mm
03.025.125* 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for Radiolucent Drive
6.0 mm
03.025.126* 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for Radiolucent Drive
Attach the appropriate drill bit to the radiolucent drive. Using image intensification, insert the tip of the drill bit through the incision and to the bone.
Incline the drive to center the tip of the drill bit over the locking hole. The drill bit should nearly fill the circle of the locking hole. Hold the drill bit in this position and drill through both cortices. Stop drilling immediately after the tip of the drill penetrates the far cortex. Remove the drill bit.
Precaution: Follow all recommended techniques for screw/bolt insertion for the relevant nail system. When inserting locking screws monocortically, stop drilling immediately after penetrating the near cortex.
Freehand Locking
*Also available.
16 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
3Drill through both cortices continued
Alternative technique
Instruments
4.0 mm
03.025.082 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 150 mm
5.0 mm
03.025.083 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 150 mm
6.0 mm
03.025.084 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 150 mm
Standard freehand locking technique can be performed without the radiolucent drive, using the appropriate drill bit.
Freehand Locking
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 17
4Determine ASLS screw length
Instrument
03.025.052 Depth Gauge, for ASLS screws
Insert the depth gauge into the drilled hole. Read the measurement from the outer sleeve of the depth gauge and select the ASLS screw length accordingly.
Note: Follow all recommended techniques for screw/bolt measurement for the relevant nail system. Ensure that the hook grasps the far cortex and that the protection sleeve is pressed firmly to the bone.
Freehand Locking
18 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
4Determine ASLS screw length continued
Alternative technique for radiolucent drill bits
Instrument
03.010.106† Direct Measuring Device, for Locking Screws to 100 mm
Stop drilling immediately after the tip of the drill penetrates the far cortex.
Disassemble the drill bit from the radiolucent drive.
Place the direct measuring device onto the drill bit. Read the graduation of the measuring device at the end of the drill bit. This corresponds to the appropriate screw length.
Freehand Locking
†Found in the EXPERT NAIL Instrument Sets.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 19
5Ream near cortex
Instruments
4.0 mm
03.025.030 2.9 mm/5.0 mm Hand Reamer
5.0 mm
03.025.031 3.9 mm/6.0 mm Hand Reamer
6.0 mm
03.025.032 4.9 mm/7.0 mm Hand Reamer
Align the hand reamer with the locking hole axis. Ream the near cortex and clear the passage down to the nail. Remove the hand reamer.
If necessary, use image intensification to control the position of the hand reamer.
Note: To correctly insert the ASLS sleeve, ensure: – the near cortex is completely reamed; – the passage from the near cortex to the nail’s locking hole is cleared; – the drilled/reamed hole aligns with the nail’s locking hole.
Freehand Locking
21 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
5Ream near cortex continued
Alternative technique
Instruments
4.0 mm
03.025.027 2.9 mm/5.0 mm Drill-Reamer, quick coupling, 145 mm
5.0 mm
03.025.028 3.9 mm/6.0 mm Drill-Reamer, quick coupling, 145 mm
6.0 mm
03.025.029 4.9 mm/7.0 mm Drill-Reamer, quick coupling, 145 mm
Use the appropriate drill-reamer to open the near cortex. Align the drill-reamer with the locking hole axis and ream the near cortex. Once the large diameter of the drill-reamer passes through the near cortex, stop reaming immediately. Remove the drill-reamer.
If necessary, clear the passage down to the nail with the hand reamer.
If necessary, use image intensification to control the position of the drill-reamer.
Precaution: Ensure that the drill-reamer does not damage the nail.
Freehand Locking
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 21
Screw Insertion
1Position sleeve
Select the appropriate resorbable sleeve.
Position the sleeve so that the sleeve lip is facing the screwhead. Thread the ASLS sleeve onto the ASLS screw.
Note: The sleeve is positioned correctly when gold is visible on either side of the sleeve and the screw tip protrudes 1 mm–2 mm beyond the sleeve.
Precautions: Do not use ASLS screws without sleeves. When attaching the sleeve, stop threading the sleeve once resistance increases. Resistance occurs when the sleeve expands as it moves onto the larger diameter of the screw. An expanded sleeve will prevent screw/sleeve insertion into a locking hole.
1–2 mm
Sleeve lip
22 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
1Position sleeve continued
Alternative technique
Instruments
4.0 mm/5.0 mm
03.025.067 Sleeve Positioner
6.0 mm
03.025.066 Sleeve Positioner
Place the sleeve into the appropriate side of the sleeve positioner. Rotate the screw into the sleeve until the sleeve is pushed out of the sleeve positioner.
Screw Insertion
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 23
2Insert screw/sleeve assembly into locking hole
Instruments
03.010.063† 12.0 mm/8.0 mm, Protection Sleeve, 188 mm (for Expert nails)
03.010.107† StarDrive Screwdriver, T25, self-retaining
03.025.040* 11.0 mm/8.0 mm, Protection Sleeve, 188 mm, for ASLS
4.0 mm
03.010.064† 8.0 mm/3.2 mm Drill Sleeve
03.010.069† 3.2 mm Trocar
5.0 mm
03.010.065† 8.0 mm/4.2 mm Drill Sleeve
03.010.070† 4.2 mm Trocar
6.0 mm
03.010.066† 8.0 mm/5.0 mm Drill Sleeve
03.010.071† 5.0 mm Trocar
Assemble the appropriate protection sleeve, drill sleeve and trocar and advance the assembly to the bone. Remove the drill sleeve and trocar. Attach the ASLS screw/sleeve assembly to the screwdriver.
Screw Insertion
†Found in the EXPERT NAIL Instrument Sets.*Also available.
24 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
2Insert screw/sleeve assembly into locking hole continued
Push the ASLS screw/sleeve assembly through the protection sleeve and near cortex, and into the nail’s locking hole. If necessary, lightly tap the screwdriver to advance the screw/sleeve into the locking hole.
Precaution: Do not rotate the screwdriver until the sleeve is seated in the locking hole.
Note: If the screw/sleeve assembly cannot be pushed to the nail stop, use the appropriate reamer to completely open the near cortex (see earlier step “Ream near cortex”).
Optional technique
Instrument
399.505 Synthetic Hammer
If necessary, lightly hammer to advance the screw/sleeve assembly into the nail’s locking hole.
Screw Insertion
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 25
3Screw into far cortex
Push the screwdriver to get the screw started. Rotate the screwdriver to advance the screw until the screw threads engage the far cortex.
Notes: The larger diameter core expands the sleeve in the nail, providing angular stability. In the final screw position, the largest diameter screw threads will engage the near cortex and the screw tip will protrude beyond the far cortex.
Screw Insertion
26 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
3Screw into far cortex continued
Note: To prevent overinsertion of the ASLS screw, use a protection sleeve during screw insertion. Ensure that the protection sleeve is pressed to the bone. Advance the screw until the engraved line on the screwdriver shaft meets the edge of the protection sleeve.
After insertion of all screws, use image intensification to ensure that reduction is maintained, that the screwheads contact the bone, and that the screws are not overinserted.
Screw Insertion
†Found in the EXPERT NAIL Instrument Sets.
4Intraoperative screw exchange
Instruments
03.010.107† StarDrive Screwdriver, T25, self-retaining
03.010.112† Holding Sleeve, with Locking Device
Using standard technique, remove the ASLS screw with the screwdriver and, if necessary, by pulling with the holding sleeve.
Note: Since the sleeve remains in the nail, it does not need to be removed and exchanged.
Insert a new ASLS screw.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 27
Implant Removal
1Remove screws
Instruments
03.010.107† StarDrive Screwdriver, T25, self-retaining
03.010.112† Holding Sleeve, with Locking Device
Using standard technique, remove ASLS screws with the screwdriver and, if necessary, the holding sleeve.
2Remove nail
Remove the nail using standard technique for the relevant nail system.
Note: The sleeves remain in the nail if not fully resorbed and can be removed with the nail during nail removal.
†Found in the EXPERT NAIL Instrument Sets.
28 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Resorbable Sleeves for Angular Stable Locking Screws, sterile*
– 70:30 poly (L /DL-lactide)
– Inner thread for secure fit to ASLS screw
– Expands in nail’s locking hole to provide angular stability
– 4.0 mm, 5.0 mm, and 6.0 mm sleeves used with corresponding ASLS screws
– Sterile packaged (2 per package)
Implants
Titanium Angular Stable Locking Screws, with T25 StarDriveTM Recess, for intramedullary nails, sterile*
– Titanium alloy**
– Available lengths – 4.0 mm ASLS screw: 26 mm–80 mm – 5.0 mm ASLS screw: 30 mm–100 mm – 6.0 mm ASLS screw: 32 mm–125 mm
– Fully threaded shaft with 3 diameters – D1: Provides purchase in reamed near cortex – D2: Expands sleeve, providing angular stability – D3: Holds unexpanded sleeve for screw insertion,
provides purchase in far cortex
– T25 StarDrive Recess
– Sterile packaged
D1
D2
D3
*Also available. **Titanium-6% aluminum-7% niobium alloy
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 29
ASLS Compatibility With Existing Nails
The Angular Stable Locking System is compatible with all DePuy Synthes titanium cannulated nailing systems.
Angular Stable Locking Systems
Nail 4.0 mm 5.0 mm 6.0 mm Nailing System Diameters (mm) (brown) (light blue) (dark purple)
Humerus Titanium Cannulated Humeral Nail and Proximal Humeral Nail-EX
7, 9, 11
•
Femur Titanium Trochanteric Fixation Nail 10, 11, 12, 14 •
Titanium Cannulated Lateral Entry 9–13 • Femoral Nail-EX and Lateral Entry Femoral Recon Nail-EX 14–16 •
Titanium Cannulated Retrograde/ 9–13 • Antegrade Femoral Nail-EX 14, 15 •
Titanium Femoral Nail (cannulated) 10–15 •
Titanium Distal Femoral Nail (cannulated)
10–13
• •
Titanium Cannulated Adolescent Lateral Entry Femoral Nail-EX
8.2, 9, 10
•
Tibia Titanium Cannulated Tibial Nail-EX 8, 9 • and Tibial Nail-EX with Proximal Bend 10–13 •
Titanium Tibial Nail (cannulated) 10–13 •
Ankle Titanium Cannulated Hindfoot • •
Arthrodesis Nail-EX
10, 12, 13
31 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Instruments
03.025.040* 11.0 mm/8.0 mm, Protection Sleeve, 188 mm, for ASLS – to be used with non-Expert nails
03.010.107† StarDrive Screwdriver, T25
03.010.112† Holding Sleeve, with Locking Device
03.025.052 Depth Gauge, for ASLS screws
03.025.067 Sleeve Positioner, for 4.0 mm and 5.0 mm ASLS sleeves
03.025.066 Sleeve Positioner, for 6.0 mm ASLS sleeves
399.505 Synthetic Hammer (100 g)
General instruments
03.010.063† 12.0 mm/8.0 mm, Protection Sleeve, 188 mm – to be used with Expert nails
†Found in the EXPERT NAIL Instrument Sets.*Also available.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 31
Instruments
Instruments for targeted locking
03.010.064† 8.0 mm/3.2 mm Drill Sleeve03.010.065† 8.0 mm/4.2 mm Drill Sleeve03.010.066† 8.0 mm/5.0 mm Drill Sleeve
03.010.069† 3.2 mm Trocar03.010.070† 4.2 mm Trocar03.010.071† 5.0 mm Trocar
03.025.104 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 331 mm, for 4.0 mm ASLS screws03.025.105 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 331 mm, for 5.0 mm ASLS screws 03.025.106 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 331 mm, for 6.0 mm ASLS screws
03.025.134 2.9 mm/5.0 mm Drill-Reamer, quick coupling, 268 mm, for 4.0 mm ASLS screws03.025.135 3.9 mm/6.0 mm Drill-Reamer, quick coupling, 288 mm, for 5.0 mm ASLS screws 03.025.136 4.9 mm/7.0 mm Drill-Reamer, quick coupling, 288 mm, for 6.0 mm ASLS screws
03.025.030 2.9 mm/5.0 mm Hand Reamer, for 4.0 mm ASLS screws03.025.031 3.9 mm/6.0 mm Hand Reamer, for 5.0 mm ASLS screws03.025.032 4.9 mm/7.0 mm Hand Reamer, for 6.0 mm ASLS screws
†Found in the EXPERT NAIL Instrument Sets.
32 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
03.025.030 2.9 mm/5.0 mm Hand Reamer, for 4.0 mm ASLS screws03.025.031 3.9 mm/6.0 mm Hand Reamer, for 5.0 mm ASLS screws03.025.032 4.9 mm/7.0 mm Hand Reamer, for 6.0 mm ASLS screws
Instruments
03.025.082 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 150 mm, for 4.0 mm ASLS screws03.025.083 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 150 mm, for 5.0 mm ASLS screws03.025.084 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 150 mm, for 6.0 mm ASLS screws
Instruments for freehand locking
03.025.027 2.9 mm/5.0 mm Drill-Reamer, quick coupling, 145 mm, for 4.0 mm ASLS screws03.025.028 3.9 mm/6.0 mm Drill-Reamer, quick coupling, 145 mm, for 5.0 mm ASLS screws03.025.029 4.9 mm/7.0 mm Drill-Reamer, quick coupling, 145 mm, for 6.0 mm ASLS screws
03.025.124* 2.9 mm/3.2 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for 4.0 mm ASLS screws, for Radiolucent Drive03.025.125* 3.9 mm/4.2 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for 5.0 mm ASLS screws, for Radiolucent Drive03.025.126* 4.9 mm/5.0 mm Three-Fluted Drill Bit, quick coupling, 145 mm, for 6.0 mm ASLS screws, for Radiolucent Drive
*Also available.
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 33
Drill-Reamers, quick coupling, 2 ea. 03.025.134 2.9 mm/5.0 mm, 268 mm, for 4.0 mm Angular Stable Locking Screws
03.025.135 3.9 mm/6.0 mm, 288 mm, for 5.0 mm Angular Stable Locking Screws
03.025.136 4.9 mm/7.0 mm, 288 mm, for 6.0 mm Angular Stable Locking Screws
Angular Stable Locking System Instrument Set (01.025.002)
60.025.002 Graphic Case for Angular Stable Locking System, for Cannulated Titanium Intramedullary Nails
Instruments399.505 Synthetic Hammer
Drill-Reamers, quick coupling, 145 mm, 2 ea. 03.025.027 2.9 mm/5.0 mm, for 4.0 mm Angular Stable Locking Screws
03.025.028 3.9 mm/6.0 mm, for 5.0 mm Angular Stable Locking Screws
03.025.029 4.9 mm/7.0 mm, for 6.0 mm Angular Stable Locking Screws
Hand Reamers 03.025.030 2.9 mm/5.0 mm, for 4.0 mm Angular Stable Locking Screws
03.025.031 3.9 mm/6.0 mm, for 5.0 mm Angular Stable Locking Screws
03.025.032 4.9 mm/7.0 mm, for 6.0 mm Angular Stable Locking Screws
03.025.052 Depth Gauge, for Angular Stable Locking System
03.025.066 Sleeve Positioner, for 6.0 mm Angular Stable Locking System Sleeves
03.025.067 Sleeve Positioner, for 4.0 mm and 5.0 mm Angular Stable Locking System Sleeves
Three-Fluted Drill Bits, quick coupling, needle point, 150 mm, 2 ea.
03.025.082 2.9 mm/3.2 mm, for 4.0 mm Angular Stable Locking Screws
03.025.083 3.9 mm/4.2 mm, for 5.0 mm Angular Stable Locking Screws
03.025.084 4.9 mm/5.0 mm, for 6.0 mm Angular Stable Locking Screws
Three-Fluted Drill Bits, quick coupling, needle point, 331 mm, 100 mm calibration, 2 ea.
03.025.104 2.9 mm/3.2 mm, for 4.0 mm Angular Stable Locking Screws
03.025.105 3.9 mm/4.2 mm, for 5.0 mm Angular Stable Locking Screws
03.025.106 4.9 mm/5.0 mm, for 6.0 mm Angular Stable Locking Screws
Note: For additional information, precautions and warnings, please see the package insert.
For detailed cleaning and sterilizationinstructions, please refer towww.synthes.com/hcp/cleaning-sterilization or sterilization instructions, if provided.
34 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Implant Sets
01.025.008 4.0 mm Angular Stable Locking System Screw Set
4.0 mm Titanium Angular Stable Locking Screws, with T25 StarDrive Recess, for intramedullary nails, sterile, 2 ea.
Length Length (mm) (mm)04.025.416S 26 04.025.444S 5404.025.418S 28 04.025.446S 5604.025.420S 30 04.025.448S 5804.025.422S 32 04.025.450S 6004.025.424S 34 04.025.452S 6204.025.426S 36 04.025.454S 6404.025.428S 38 04.025.456S 6604.025.430S 40 04.025.458S 6804.025.432S 42 04.025.460S 7004.025.434S 44 04.025.462S 7204.025.436S 46 04.025.464S 7404.025.438S 48 04.025.466S 7604.025.440S 50 04.025.468S 7804.025.442S 52 04.025.470S 80
01.025.009 5.0 mm Angular Stable Locking System Screw Set
5.0 mm Titanium Angular Stable Locking Screws, with T25 StarDrive Recess, for intramedullary nails, sterile, 2 ea.
Length Length (mm) (mm)04.025.520S 30 04.025.550S 6004.025.522S 32 04.025.552S 6204.025.524S 34 04.025.554S 6404.025.526S 36 04.025.556S 6604.025.528S 38 04.025.558S 6804.025.530S 40 04.025.560S 7004.025.532S 42 04.025.562S 7204.025.534S 44 04.025.564S 7404.025.536S 46 04.025.566S 7604.025.538S 48 04.025.568S 7804.025.540S 50 04.025.570S 8004.025.542S 52 04.025.575S 8504.025.544S 54 04.025.580S 9004.025.546S 56 04.025.585S 9504.025.548S 58 04.025.590S 100
ASLS Screws
Angular Stable Locking System (ASLS) Surgical Technique DePuy Synthes 35
Implant Sets
01.025.010 6.0 mm Angular Stable Locking System Screw Set
6.0 mm Titanium Angular Stable Locking Screws, with T25 StarDrive Recess, for intramedullary nails, sterile, 2 ea.
Length Length (mm) (mm)04.025.622S 32 04.025.656S 6604.025.624S 34 04.025.658S 6804.025.626S 36 04.025.660S 7004.025.628S 38 04.025.662S 7204.025.630S 40 04.025.664S 7404.025.632S 42 04.025.666S 7604.025.634S 44 04.025.668S 7804.025.636S 46 04.025.670S 8004.025.638S 48 04.025.675S 8504.025.640S 50 04.025.680S 9004.025.642S 52 04.025.685S 9504.025.644S 54 04.025.690S 10004.025.646S 56 04.025.695S 10504.025.648S 58 04.025.700S 11004.025.650S 60 04.025.705S 11504.025.652S 62 04.025.710S 12004.025.654S 64 04.025.715S 125
Alternative Set Configurations
01.025.001 Angular Stable Locking System Screw Set Consists of: 4.0 mm ASLS Screw Set (01.025.008) and 5.0 mm ASLS Screw Set (01.025.009)
ASLS Sleeves*
Resorbable Sleeve for Angular Stable Locking Screws, sterile, 2/pkg.
08.025.032S for 4.0 mm ASLS Screws08.025.044S for 5.0 mm ASLS Screws08.025.057S for 6.0 mm ASLS Screws
*Must be ordered separately.
36 DePuy Synthes Angular Stable Locking System (ASLS) Surgical Technique
Instruments Drill-Reamers, quick coupling, 145 mm, for Radiolucent Drive 03.025.037 2.9 mm/5.0 mm, for 4.0 mm Angular Stable Locking Screws
03.025.038 3.9 mm/6.0 mm, for 5.0 mm Angular Stable Locking Screws
03.025.039 4.9 mm/7.0 mm, for 6.0 mm Angular Stable Locking Screws
03.025.040 11.0 mm/8.0 mm Protection Sleeve, 188 mm, for Angular Stable Locking System
Three-Fluted Drill Bits, quick coupling, needle point, 145 mm, for Radiolucent Drive 03.025.124 2.9 mm / 3.2 mm, for 4.0 mm Angular Stable Locking Screws
03.025.125 3.9 mm/4.2 mm, for 5.0 mm Angular Stable Locking Screws
03.025.126 4.9 mm/5.0 mm, for 6.0 mm Angular Stable Locking Screws
Expert Nail Instruments†
03.010.063 12.0 mm/8.0 mm Protection Sleeve, 188 mm
Drill Sleeves 03.010.064 8.0 mm/3.2 mm, 200 mm
03.010.065 8.0 mm/4.2 mm, 200 mm
03.010.066 8.0 mm/5.0 mm, 200 mm
Trocars 03.010.069 3.2 mm, 210 mm
03.010.070 4.2 mm, 210 mm
03.010.071 5.0 mm, 210 mm
03.010.106 Direct Measuring Device, for Locking Screws to 100 mm, for IM Nails03.010.107 StarDrive Screwdriver, T25, self-retaining
03.010.112 Holding Sleeve, with Locking Device
03.010.150 Star/HexDrive Screwdriver, T25, 3.5 mm hex, self-retaining
Power Equipment105.955 Small Battery Drive Set
150.119 Battery Power Line Set
511.30 Radiolucent Drive
530.605 Battery Reamer/Drill
530.620 Battery, for Battery Power Line
530.680 Battery Casing, for Battery Power Line
530.741 Adaptor for Radiolucent Drive, for Battery Power Line
530.750 Quick Coupling for Drill Bits, for Battery Power Line
532.010 Small Battery Drive
532.013 Quick Coupling for Drill Bits
532.031 Adaptor for Radiolucent Drive
532.032 Small Battery Drive Battery Casing, for 14.4 V Battery
532.033 Small Battery Drive 14.4 V Battery
Also Available
†Found in the EXPERT NAIL Instrument Sets.
Limited Warranty and Disclaimer: DePuy Synthes products are sold with a limited warranty to the original purchaser against defects in workmanship and materials. Any other express or implied warranties, including warranties of merchantability or fitness, are hereby disclaimed.
Please also refer to the package insert(s) or other labeling associated with the devices identified in this surgical technique for additional information.
CAUTION: Federal Law restricts these devices to sale by or on the order of a physician.
Some devices listed in this surgical technique may not have been licensed in accordance with Canadian law and may not be for sale in Canada. Please contact your sales consultant for items approved for sale in Canada.
Not all products may currently be available in all markets.
© DePuy Synthes 2009–2017. All rights reserved.DSUS/TRM/1016/1125 6/17 DV
Synthes USA, LLC 1101 Synthes AvenueMonument, CO 80132
Manufactured or distributed by:Synthes USA Products, LLC 1302 Wrights Lane EastWest Chester, PA 19380
To order (USA): 800-523-0322 To order (Canada): 855-946-8999
Note: For recognized manufacturer, refer to the product label.
www.depuysynthes.com