FNI Meeting Outlines: 1.Current status of FNI detector and results 2.New designs of neutron convertor 3.New material for collimator: Boron- loaded Polyethylene 4.Discussion: Next step test with Am-Be neutron source

Upload: imogene-patton

Post on 01-Jan-2016

11 views

Category:

Documents


0 download

DESCRIPTION

Outlines: Current status of FNI detector and results New designs of neutron convertor New material for collimator: Boron-loaded Polyethylene Discussion: Next step test with Am-Be neutron source. FNI Meeting. Current results. The Imaging are mainly come from X-rays. Current results. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: FNI Meeting

FNI Meeting

Outlines:1. Current status of FNI detector and

results2. New designs of neutron convertor3. New material for collimator:

Boron-loaded Polyethylene4. Discussion: Next step test with

Am-Be neutron source

Page 2: FNI Meeting

Current results

The Imaging are mainly come from X-rays

Page 3: FNI Meeting

Current results

The unclear imaging of the hole is caused by:1. very low detection efficiency of the detector for neutron;2. Cosmic muon’s influence.

• Thickness of paraffin: 5cm,• Diameter of hole: 7mm• X and gamma rays are shielded by

Lead(front and back).

Page 4: FNI Meeting

Current design

Neutron Events

X and gamm

a rays events

Page 5: FNI Meeting

Current Design -- Conclusion

1.Neutrons are detected by FNI detector;2.No clear neutron imaging, the reason is

caused by the low detection efficiency and cosmic muon’s influence;

3.Detection efficiency need to be improved;1.With 200 um polyethylene: 0.06%2.New design: 10% (estimated, in principle, if

the thicker structure used, the detection efficiency can be higher)

Page 6: FNI Meeting

2. New Convertor Design-- two ideas

Design 1: Polyethylene pillars (Blue). Top andbottom are metal mesh, which are served aselectrodes.

Design2: Polyethylene with hole. Top andbottom covered by metal/copper whichServed as electrodes.

Note: Huaya is doing the simulation of this new ideas and optimize the dimensions.

Page 7: FNI Meeting

2. New convertor design-- conversion efficiency

Lvxing An’s Simulation results:

1. Energy of incident neutron: 2 MeV2. Convertor: bulk Polyethylene

Conclusion: Based this simulation, the diameter of the pillar and the thickness of the wall in the new design should be chosen about 200 um for 2 MeV neutron. But for the 14 MeV neutron, it should be chosen more thicker than this value. The value will be obtained soon.

Page 8: FNI Meeting

3. New collimation material-- Boron-loaded polyethylene

Page 9: FNI Meeting

4. Next step test

1. Continue testing it with Am-Be neutron source2. Once it is done, test it with 14 MeV neutrons