Model 200 Flow Meters ® Positive Displacement Flow Meters B1750 B1750 Positive Displacement Flow Meter Accuracy of ±0.5% of reading and repeatability of ±0.1% No need for additional straight run piping Designed for fluids with a wide range of viscosities, as well as low flow rates Available in high strength aluminum or stainless steel housing

Upload: others

Post on 14-Apr-2020

20 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Flow Meters - RIME FLUID SYSTEMS...B1750 Positive Displacement Flow Meters B1750 Positive Displacement Flow Meter Accuracy of ±0.5% of reading and repeatability of ±0.1% No need

Model 200

Blancett is a registered trademark of Racine Federated Inc. Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Company

Flow Meters®

Flow Meters®

Division of Racine Federated Inc.

Positive Displacement Flow MetersB1750

B1750 Positive Displacement Flow Meter

Accuracy of ±0.5% of reading and repeatability of ±0.1%

No need for additional straight run piping

Designed for fluids with a wide range of viscosities, as well as low flow rates

Available in high strength aluminum or stainless steel housing

© 2009 Blancett Printed in USA 10/09 Form No. 1750

Toll Free: 800-235-1638 Tel: 262-639-6770 Fax: 262-417-1155

8635 Washington Avenue, Racine, WI 53406-3738 U.S.A.

E-mail: [email protected]

B1750 Dimensions - Inches (mm)

A20/S20 A60/S60 A80/S80

3.3(84.4)DIA

.47(12)

2.17(55)

1.73(44)

.40(10.2)

.40(10.2)

4.9(125)DIA

.91(23)

3.78(96)

1.61(41)

2.36(60)

1.25(32)2.63

(67)5.50(140)

.63(16)

7.85(199)DIA

5.50(140)

This variable reluctance magnetic pickup coil (B170109) is designed to exceed the performance and application criteria of any standard or modified pickups. Its modular design allows it to be used with the B1750 positive displacement meter, and interface with all Blancett electronics. If using another manufacturer’s electronics with this meter, a Blancett B170210 pre-amplifier may be needed to correctly interface the devices.Output: 30 mV – 3 VAC Sine Wave (B170109) 0-10 V – Square Wave (B170210) 4-20 mA - 3 wire 4-20 mA output (B175420)Connector: 2-pin gold plated type to mate with MS3106-10SL-4S or equivalent (B170109) 3-pin gold plated type to mate with MS3106A-10SL-3S or equivalent (B170210) Temperature: -100 °F to +325 °F (-73 °C to +162 °C) (B170109) -20 °F to +160 °F (-29 °C to +71 °C) (B170210) 0 °F to +185 °F (-18 °C to + 85 °C) (B175420)Materials of Construction: 303/304 stainless steel shell and bottom, solid epoxy encapsulation, standard commercial finish

Note: 80/90 Series utilize Quad-4 Sensor (B170180)

Optional Magnetic PickupUsed with 12, 20, 30 and 60 series only

.51(13)

2.64(67)

3.17(80.4)DIA

4.72(120)DIA

3.15(80)DIA

1.73(44)

3.32(84.4)DIA

A30/S30

1/2"-13 UNC(10 PL)

2.63(67)5.50(140)

1.25(32)

1.61(41)

A90/S90

7.00(178)

7.85(199)DIA

2.00(51)

1.50(38)

SENSOR HOLE(1/2-20 NF)

1/4 NPTPORTS

1.00(26)

.36(9.2)

1.40(36)

2.00(51)

1/4-20 NCMOUNTING

HOLES

2.63(67)

SENSORCOLLAR

.86(22)

A12/S12

1/2"-13 UNC(10 PL)

Racine Flow Meter Group

RAC

INE FEDERATED IN

C.

ISO 9001 : 2008 FILE N0. A68

56

Page 2: Flow Meters - RIME FLUID SYSTEMS...B1750 Positive Displacement Flow Meters B1750 Positive Displacement Flow Meter Accuracy of ±0.5% of reading and repeatability of ±0.1% No need

Introduction

Ordering Information

B1750

Accuracy: ±0.5% over the published flow range with fluids >100 cP; over a 10:1 turndown (from maximum flow) with fluids <30 cP Repeatability: ±0.1%Pressure Rating: 5000 psi [345 bar] maximumOperating Temperature: -20 °F to +185 °F [-29 °C to +85 °C] aluminum -20 °F to +400 °F [-29 °C to +204 °C] stainless steelConnections: Female NPT: ¼", ¾", or 1- ¼" (depending on meter size) Materials of Construction: Stainless steel (gears and bearings) O-Ring: Teflon®, Viton® (optional)Housing: 6061 - T6 aluminum or 303 stainless steel

Specifications

2 3

Operating Principle

K-factor

The Blancett Model B1750 positive displacement flow meter provides high measurement accuracy, trouble-free operation and long service life for fluids over a wide viscosity range. Applications include lubricants, fuels, chemicals and solvents, oils and greases. The B1750 meter is bi-directional (using appropriate electronics) and has an extremely large turndown ratio - up to 400:1 in some models - with full accuracy at extremely low flow rates.

Fluid entering the meter drives two gears. A non-intrusive sensor detects the movement of the gears and produces a sine wave pulse for each gear tooth that passes the path of the sensor face. The resulting pulse frequency is proportional to the actual flow rate, and it provides a highly accurate representation of the fluid flow. The meter is relatively insensitive to changes in viscosity and there is no need for straight run piping.

The K-factor represents the number of output pulses transmitted per unit volume of liquid passing through the positive displace-ment meter. Each meter has a unique K-factor that is determined during factory calibration. The K-factor is very constant and linear over the published flow range when liquid viscosity is greater than 100 cP. When liquid viscosity is less than 100 cP positive displacement meters can experience “fluid slip” in the measuring chamber due to migration of liquid around the internal moving parts. As a result, the linear (constant K-factor portion) measuring range of the flow meter is reduced. At viscosities less than 30 cP, positive displacement meters maintain published linearity over a 10:1 turndown range from the maximum published flow rate.

B1750

Flow Rate vs. Pressure Drop

FLUID FLOW

(IN) (OUT)

Model B1750 Positive Displacement Flow Meters

B175-A12B175-A20B175-A30B175-A60B175-A804

B175-A904

B175-A12-VB175-A20-VB175-A30-VB175-A60-VB175-A80-V4

B175-A90-V4

¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT

0.003-0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.003- 0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

53,00015,9006,6001,800

1,600~800~

53,00015,9006,6001,800

1,600~800~

PartNumber1

SealMaterial

EndConnection

Flow Ranges3

Aluminum Housing - 185 °F Maximum Fluid TemperatureK-factor2

Pulses/GalGPM LPM

Teflon®

Standard

Viton®

B175-S12B175-S20B175-S30B175-S60B175-S804

B175-S904

B175-S12-VB175-S20-VB175-S30-VB175-S60-VB175-S80-V4

B175-S90-V4

¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT

0.003-0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.003- 0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

53,00015,9006,6001,800

1,600~800~

53,00015,9006,6001,800

1,600~800~

PartNumber1

SealMaterial

EndConnection

Flow Ranges3

303 Stainless Steel Housing - 400 °F Maximum Fluid TemperatureK-factor2

Pulses/GalGPM LPM

Teflon®

Standard

Viton®

¹ Does NOT include pickup - To order, see Optional Magnetic Pick-up, page 4² All K-factors are approximate ³ Accuracy: ±0.5% over the published flow range with fluids >100 cP; over a 10:1 turndown (from maximum flow) with fluids <30 cP 4 90° Flange fittings required for installation, see page 24 of the Blancett Price List~ Configured for Quad-4 sensor output (B170180)

0.005 0.02 0.13 0.26 0.39 0.52 0.66 0.79

80

70

60

50

40

30

20

10

0

105 cSt

80 cSt

63 cSt

30 cSt

15 cSt5 cSt

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

00.185

0 0.3700.555

0.7400.925

1.1101.294

1.4791.664

1.850

ΔPSI

ΔPSI

ΔPSI

3000 cSt 2000 cSt 1500 cSt

1000 cSt

800 cSt

500 cSt

400 cSt

300 cSt

200 cSt

100 cSt

30 cSt

GPM

GPM

GPM

B1750 - 20 Series Pressure Drop Chart

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

01.48

0 2.964.44

5.927.4 8.88

10.3611.83

13.3114.79

16.2717.75

ΔPSI3000 cSt 2000 cSt 1000 cSt

500 cSt

300 cSt

100 cSt

50 cSt

30 cSt

GPM

B1750 - 60 Series Pressure Drop Chart

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

00.53

0 1.061.58

2.112.64

3.173.7 4.23

4.755.28

5.816.34

ΔPSI3000 cSt 2000 cSt 1000 cSt

500 cSt

300 cSt

100 cSt

50 cSt

30 cSt

GPM

B1750 - 30 Series Pressure Drop Chart

120

100

80

60

40

20

0

6.6 13.2

19.8

26.4 33 39.6 45

B1750 - 80 Series Pressure Drop Chart

1000 cSt

300 cSt

150 cSt

40 cSt

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

013.2

0 26.439.6

52.866 79.2

92.4105.6

118.8132

∆PSI

1000 cSt

300 cSt

10 cSt

30 cSt

GPM

B1750 - 90 Series Pressure Drop Chart

B1750 - 12 Series Pressure Drop Chart

Page 3: Flow Meters - RIME FLUID SYSTEMS...B1750 Positive Displacement Flow Meters B1750 Positive Displacement Flow Meter Accuracy of ±0.5% of reading and repeatability of ±0.1% No need

Introduction

Ordering Information

B1750

Accuracy: ±0.5% over the published flow range with fluids >100 cP; over a 10:1 turndown (from maximum flow) with fluids <30 cP Repeatability: ±0.1%Pressure Rating: 5000 psi [345 bar] maximumOperating Temperature: -20 °F to +185 °F [-29 °C to +85 °C] aluminum -20 °F to +400 °F [-29 °C to +204 °C] stainless steelConnections: Female NPT: ¼", ¾", or 1- ¼" (depending on meter size) Materials of Construction: Stainless steel (gears and bearings) O-Ring: Teflon®, Viton® (optional)Housing: 6061 - T6 aluminum or 303 stainless steel

Specifications

2 3

Operating Principle

K-factor

The Blancett Model B1750 positive displacement flow meter provides high measurement accuracy, trouble-free operation and long service life for fluids over a wide viscosity range. Applications include lubricants, fuels, chemicals and solvents, oils and greases. The B1750 meter is bi-directional (using appropriate electronics) and has an extremely large turndown ratio - up to 400:1 in some models - with full accuracy at extremely low flow rates.

Fluid entering the meter drives two gears. A non-intrusive sensor detects the movement of the gears and produces a sine wave pulse for each gear tooth that passes the path of the sensor face. The resulting pulse frequency is proportional to the actual flow rate, and it provides a highly accurate representation of the fluid flow. The meter is relatively insensitive to changes in viscosity and there is no need for straight run piping.

The K-factor represents the number of output pulses transmitted per unit volume of liquid passing through the positive displace-ment meter. Each meter has a unique K-factor that is determined during factory calibration. The K-factor is very constant and linear over the published flow range when liquid viscosity is greater than 100 cP. When liquid viscosity is less than 100 cP positive displacement meters can experience “fluid slip” in the measuring chamber due to migration of liquid around the internal moving parts. As a result, the linear (constant K-factor portion) measuring range of the flow meter is reduced. At viscosities less than 30 cP, positive displacement meters maintain published linearity over a 10:1 turndown range from the maximum published flow rate.

B1750

Flow Rate vs. Pressure Drop

FLUID FLOW

(IN) (OUT)

Model B1750 Positive Displacement Flow Meters

B175-A12B175-A20B175-A30B175-A60B175-A804

B175-A904

B175-A12-VB175-A20-VB175-A30-VB175-A60-VB175-A80-V4

B175-A90-V4

¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT

0.003-0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.003- 0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

53,00015,9006,6001,800

1,600~800~

53,00015,9006,6001,800

1,600~800~

PartNumber1

SealMaterial

EndConnection

Flow Ranges3

Aluminum Housing - 185 °F Maximum Fluid TemperatureK-factor2

Pulses/GalGPM LPM

Teflon®

Standard

Viton®

B175-S12B175-S20B175-S30B175-S60B175-S804

B175-S904

B175-S12-VB175-S20-VB175-S30-VB175-S60-VB175-S80-V4

B175-S90-V4

¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT¼" Female NPT¼" Female NPT½" Female NPT¾" Female NPT

1-¼" Female NPT1-¼" Female NPT

0.003-0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.003- 0.800.01-2.000.03-7.000.05-20.00.5-60.01-120

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

0.011-3.030.04-7.60.11-26.50.19-75.71.9-2273.8-454

53,00015,9006,6001,800

1,600~800~

53,00015,9006,6001,800

1,600~800~

PartNumber1

SealMaterial

EndConnection

Flow Ranges3

303 Stainless Steel Housing - 400 °F Maximum Fluid TemperatureK-factor2

Pulses/GalGPM LPM

Teflon®

Standard

Viton®

¹ Does NOT include pickup - To order, see Optional Magnetic Pick-up, page 4² All K-factors are approximate ³ Accuracy: ±0.5% over the published flow range with fluids >100 cP; over a 10:1 turndown (from maximum flow) with fluids <30 cP 4 90° Flange fittings required for installation, see page 24 of the Blancett Price List~ Configured for Quad-4 sensor output (B170180)

0.005 0.02 0.13 0.26 0.39 0.52 0.66 0.79

80

70

60

50

40

30

20

10

0

105 cSt

80 cSt

63 cSt

30 cSt

15 cSt5 cSt

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

00.185

0 0.3700.555

0.7400.925

1.1101.294

1.4791.664

1.850

ΔPSI

ΔPSI

ΔPSI

3000 cSt 2000 cSt 1500 cSt

1000 cSt

800 cSt

500 cSt

400 cSt

300 cSt

200 cSt

100 cSt

30 cSt

GPM

GPM

GPM

B1750 - 20 Series Pressure Drop Chart

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

01.48

0 2.964.44

5.927.4 8.88

10.3611.83

13.3114.79

16.2717.75

ΔPSI3000 cSt 2000 cSt 1000 cSt

500 cSt

300 cSt

100 cSt

50 cSt

30 cSt

GPM

B1750 - 60 Series Pressure Drop Chart

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

00.53

0 1.061.58

2.112.64

3.173.7 4.23

4.755.28

5.816.34

ΔPSI3000 cSt 2000 cSt 1000 cSt

500 cSt

300 cSt

100 cSt

50 cSt

30 cSt

GPM

B1750 - 30 Series Pressure Drop Chart

120

100

80

60

40

20

0

6.6 13.2

19.8

26.4 33 39.6 45

B1750 - 80 Series Pressure Drop Chart

1000 cSt

300 cSt

150 cSt

40 cSt

217.5

203.0

188.5

174.0

159.5

145.0

130.5

116.0

101.5

87.0

72.5

58.0

43.5

29.0

14.5

013.2

0 26.439.6

52.866 79.2

92.4105.6

118.8132

∆PSI

1000 cSt

300 cSt

10 cSt

30 cSt

GPM

B1750 - 90 Series Pressure Drop Chart

B1750 - 12 Series Pressure Drop Chart

Page 4: Flow Meters - RIME FLUID SYSTEMS...B1750 Positive Displacement Flow Meters B1750 Positive Displacement Flow Meter Accuracy of ±0.5% of reading and repeatability of ±0.1% No need

Model 200

Blancett is a registered trademark of Racine Federated Inc. Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Company

Flow Meters®

Flow Meters®

Division of Racine Federated Inc.800-235-1638 www.blancett.com

Positive Displacement Flow MetersB1750

B1750 Positive Displacement Flow Meter

Accuracy of ±0.5% of reading and repeatability of ±0.1%

No need for additional straight run piping

Designed for fluids with a wide range of viscosities, as well as low flow rates

Available in high strength aluminum or stainless steel housing

© 2009 Blancett Printed in USA 10/09 Form No. 1750

B1750 Dimensions - Inches (mm)

A20/S20 A60/S60 A80/S80

3.3(84.4)DIA

.47(12)

2.17(55)

1.73(44)

.40(10.2)

.40(10.2)

4.9(125)DIA

.91(23)

3.78(96)

1.61(41)

2.36(60)

1.25(32)2.63

(67)5.50(140)

.63(16)

7.85(199)DIA

5.50(140)

This variable reluctance magnetic pickup coil (B170109) is designed to exceed the performance and application criteria of any standard or modified pickups. Its modular design allows it to be used with the B1750 positive displacement meter, and interface with all Blancett electronics. If using another manufacturer’s electronics with this meter, a Blancett B170210 pre-amplifier may be needed to correctly interface the devices.Output: 30 mV – 3 VAC Sine Wave (B170109) 0-10 V – Square Wave (B170210) 4-20 mA - 3 wire 4-20 mA output (B175420)Connector: 2-pin gold plated type to mate with MS3106-10SL-4S or equivalent (B170109) 3-pin gold plated type to mate with MS3106A-10SL-3S or equivalent (B170210) Temperature: -100 °F to +325 °F (-73 °C to +162 °C) (B170109) -20 °F to +160 °F (-29 °C to +71 °C) (B170210) 0 °F to +185 °F (-18 °C to + 85 °C) (B175420)Materials of Construction: 303/304 stainless steel shell and bottom, solid epoxy encapsulation, standard commercial finish

Note: 80/90 Series utilize Quad-4 Sensor (B170180)

Optional Magnetic PickupUsed with 12, 20, 30 and 60 series only

.51(13)

2.64(67)

3.17(80.4)DIA

4.72(120)DIA

3.15(80)DIA

1.73(44)

3.32(84.4)DIA

A30/S30

1/2"-13 UNC(10 PL)

2.63(67)5.50(140)

1.25(32)

1.61(41)

A90/S90

7.00(178)

7.85(199)DIA

2.00(51)

1.50(38)

SENSOR HOLE(1/2-20 NF)

1/4 NPTPORTS

1.00(26)

.36(9.2)

1.40(36)

2.00(51)

1/4-20 NCMOUNTING

HOLES

2.63(67)

SENSORCOLLAR

.86(22)

A12/S12

1/2"-13 UNC(10 PL)

Racine Flow Meter Group

RAC

INE FEDERATED IN

C.

ISO 9001 : 2008 FILE N0. A68

56