1-800-235-1638 www.blancett.com Flow Meters 3-A Sanitary Turbine Flow Meter ® FloClean TM Sanitary Turbine Flow Meter FloClean TM Meets 3-A Sanitary Standard number 28-03 Accuracy of ±1.0% of Reading and Repeatability of ±0.1% 0.6 to 400 gpm flow rates Models available for CIP/SIP and COP/SOP environments Materials comply to FDA requirements

Upload: others

Post on 26-Apr-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Flow Meterscmeters.com/pdf/blancett/FloClean Data.pdf · 2013-05-07 · 1-800-235-1638 Flow Meters 3-A Sanitary Turbine Flow Meter ® FloCleanTM Sanitary Turbine Flow Meter FloCleanTM

1-800-235-1638 www.blancett.com

Flow Meters

3-A Sanitary Turbine Flow Meter

®

FloCleanTM Sanitary Turbine Flow Meter

FloCleanTM

Meets 3-A Sanitary Standard number 28-03

Accuracy of ±1.0% of Reading and Repeatability of ±0.1%

0.6 to 400 gpm flow rates

Models available for CIP/SIP and COP/SOP environments

Materials comply to FDA requirements

Toll Free: 1-800-235-1638 Tel: 1-262-639-6770 Fax: 1-262-639-2267

© 2005 Blancett Printed in USA 4/05 Form No. 1600

Blancett is a registered trademark of Racine Federated Inc. FloClean is a trademark of Racine Federated Inc.Teflon is a registered trademark of E.I. DuPont de Nemours & Co.

8635 Washington Avenue, Racine, WI 53406-3738 U.S.A.

Flow Meters®

E-mail: [email protected]

Division of Racine Federated Inc.

Related Blancett Products

Scales turbine meter output to desired engineering unitsAmplifies turbine meter pulse outputConverts frequency outputs into recognizable unitsfor PLCs and other devicesSwitch-selectable or programmable versions availableCSA approved

Microprocessor-basedLinearized output capabilityChoice of 4-20 mA or 0-5 VDC outputEnables integration with data acquisition devicesFrequency measurement accuracy ±1%CSA approved

K-Factor Scaler Frequency Divider

Microprocessor-based flow monitor and totalizerUse with Blancett turbine flow meters as well as otherflow meters with a frequency outputBattery (1.5 VDC) and loop-powered (4-20 mA) versionsMeter, remote, panel and swivel mounting optionsHand-held and explosion-proof models also availableCSA and CE approvedClass 1, Division 1 (Intrinsically Safe) certification

B2800 Flow Monitor

F to I / F to V Intelligent Converter

Page 2: Flow Meterscmeters.com/pdf/blancett/FloClean Data.pdf · 2013-05-07 · 1-800-235-1638 Flow Meters 3-A Sanitary Turbine Flow Meter ® FloCleanTM Sanitary Turbine Flow Meter FloCleanTM

Introduction

FloCleanTM

2

The FloClean Sanitary turbine flow meter was developed for use in the food, beverage and pharmaceutical industries. The 316L stainless steel construction provides a durable and cost efficient flow measurement system that offers excellent accuracy and repeatability.

The FloClean utilizes the most up-to-date polishing technology on all internal components and all materials comply with FDA requirements. FloClean meets the requirements of 3-A Sanitary Standard number 28-03 for use in clean-in-place (CIP) and sanitize-in-place (SIP) environments, as well as clean-out-of-place (COP) and sanitize-out-of-place (SOP) applications.

The FloClean output signal is a sine-wave that is proportional to volumetric flow. With optional Blancett electronics, FloClean provides local flow rate and volume totalization and will interface with most displays, PLCs and computers.

Materials of Construction: Body and Internal Wetted Parts: 316L Stainless Steel Bearings: Standard - Nickel Bindery Tungsten Carbide Turbine: Nickel Plated CD4MCU Stainless Steel Shaft: Nickel Bindery Tungsten Carbide Sanitary Seals: Teflon® (CIP/SIP only)Flow Accuracy: ±1.0% of reading Repeatability: ±0.1%Calibration: Water (NIST Traceable Calibration)Pressure Rating: 1,000 psi (Rating based on Tri-clamp sanitary connection)Turbine Temperature: -150 °F to +300 °F (-100 °C to +149 °C)End Connections: Sanitary Clamp End Mag Pick-up: NEMA 6; -150 °F to +300 °F (-100 °C to +149 °C)

Specifications

FloClean Size Chart

Clean-In-Place Diagram Clean-Out-of-Place Diagram

FloClean 3-A (COP) Sanitary Meter with NEMA 6 pick-up

(P.N. B16A-110A-1BA)

FloClean Flow Meter Part Numbering Information

†Limited to combinations shown in the FloClean Size Chart below.*Not available with CIP/SIP

NOTE: Diagrams based on 1-1/2” clamp size with 1”: 5.00 - 50.00 GPM Meter

3-A Sanitary ratingA - COP / SOPB - CIP / SIP

Bearing MaterialA - Ni Bindery Tungsten Carbide

CalibrationA - Standard 5 PointB - 10 PointC - 20 point

Meter Body HubA - 1" NPT Hub*B - No HubPick-up Option

0 - NEMA 6 Magnetic1 - NEMA 6 Magnetic w/Pre-Amp2 - Magnetic (non-Nema 6)3 - Magnetic w/Pre-Amp (non-NEMA 6)4 - Active Sensor (non-NEMA 6) with 4-20 mA6 - High Temp (non-NEMA 6)8 - Active Sensor (non-NEMA 6) 0-5 VDC9 - None

Clamp Size†

0 - 3/4" Ferrule1 - 1-1/2" Ferrule2 - 2-1/2" Ferrule*

Meter Size: Flow Range†

03 - 3/8" : 0.6 - 30 GPM05 - 1/2" : 0.75 - 7.50 GPM07 - 3/4" : 2.00 - 15.00 GPM08 - 7/8" : 3.00 - 30.00 GPM10 - 1" : 5.00 - 50.00 GPM15 - 1-1/2" : 15.0 - 180.0 GPM*20 - 2" : 40.0 - 400.0 GPM*

Operating PrincipleFluid entering the meter first passes through an inlet flow straightener thatreduces its turbulent flow pattern. Fluid then passes through the turbine,causing the turbine to rotate at a speed proportional to fluid velocity. As eachturbine blade passes through the magnetic field generated by the meter’smagnetic pick-up, an AC voltage pulse is generated. These pulses providean output frequency that is proportional to volumetric flow.

K-FactorThe K-Factor represents the number of output pulsestransmitted per gallon of fluid passing through the turbine meter. Each turbine has a unique K-Factor. However, turbine meters are not functionally consistent throughout the full flow range of the meter.

There are several forms of “friction” inherent in turbine meters that retard the rotational movement of the turbine rotor. These frictional forces include: magnetic drag, created by electromagnetic force of pick-up transducers; mechanical drag, due to bearing friction; and viscous drag,produced by flowing fluid. See charts below.

As flow increases, the frictional forces are minimized andthe free-wheeling motion of the turbine rotor becomesmore linear (proportional to flow). The K-Factor becomesrelatively constant and linear throughout the balance of thelinear flow range. This is approximately a 10:1 turndownratio from the maximum flow rate down to the minimumflow rate.

Repair Kits

3

2.0(50.8)

2.72(69.1)

Inches (mm) Inches (mm)

3.00(76.2)

4.00(101.6)

4.00(101.6)

Clamp Size MeterSize GPM LPM COP/SOP CIP/SIP

K-factorPulses/Gal

3/4"

3/4"

3/4"

1-1/2"

1-1/2"

1-1/2"

1-1/2"

1-1/2"

2-1/2"

3/8"

1/2"

3/4"

1/2"

3/4"

7/8"

1"

1-1/2"

2"

2.27 - 11.36

3.0 - 30

7.5 - 57

3.0 - 30

7.5 - 57

11 - 110

19 - 190

57 - 680

150 - 1,500

0.60 - 3.00

0.75 - 7.50

2.0 - 15.0

0.75 - 7.50

2.0 - 15.0

3.0 - 30

5.0 - 50.0

15 - 180

40 - 400

20,000

13,000

2,750

13,000

2,750

2,686

870

330

52

B16A-K03A

B16A-K05A

B16A-K07A

B16A-K05A

B16A-K07A

B16A-K08A

B16A-K10A

B16A-K15A

B16A-K20A

B16B-K03A

B16B-K05A

B16B-K07A

B16B-K05A

B16B-K07A

B16B-K08A

B16B-K10A

*

*

FLOW RATE REPAIR KIT PART NUMBER

3.00(76.2)

Typical K-Factor Curve (Pulses per gallon)

PRESSURE DROP

K-FACTOR (PULSES/GALLON)

K-FA

CTOR

(CYC

LES/

GALL

ON)

OUTP

UT FR

EQUE

NCY (

CYCL

ES/S

EC)

% O

F VAL

UE AT

FLOW

RAT

E

100%

50%

0% 10% 50% 100%

STANDARD TURNDOWN APPROX. 10:1

LINEA

RITY

(DES

IGN

DEVIA

TION)

+1%

+%

-%

-1%

AVG. K-FACTOR (K) K-FACTOR (PULSES/GALLON)

FULL FLOW RANGELINEAR FLOW RANGE

Factory calibrated replacement kits are available for field or factory service. Both of the FloClean models are designed to allow for quick, easy disassembly and replacement of the meter's internal components.

A clean-in-place (CIP) and sanitize-in-place (SIP) repair kit contains retaining screws, two rotor supports, one rotor assembly, two gaskets, and a K-factor tag. A clean-out-of-place (COP) and sanitize-out-of-place (SOP) repair kit contains two retaining rings, two rotor supports, one rotor assembly, and a K-factor tag.

For further details, contact the factory or refer to Form #4300.

FloCleanTM

Page 3: Flow Meterscmeters.com/pdf/blancett/FloClean Data.pdf · 2013-05-07 · 1-800-235-1638 Flow Meters 3-A Sanitary Turbine Flow Meter ® FloCleanTM Sanitary Turbine Flow Meter FloCleanTM

Introduction

FloCleanTM

2

The FloClean Sanitary turbine flow meter was developed for use in the food, beverage and pharmaceutical industries. The 316L stainless steel construction provides a durable and cost efficient flow measurement system that offers excellent accuracy and repeatability.

The FloClean utilizes the most up-to-date polishing technology on all internal components and all materials comply with FDA requirements. FloClean meets the requirements of 3-A Sanitary Standard number 28-03 for use in clean-in-place (CIP) and sanitize-in-place (SIP) environments, as well as clean-out-of-place (COP) and sanitize-out-of-place (SOP) applications.

The FloClean output signal is a sine-wave that is proportional to volumetric flow. With optional Blancett electronics, FloClean provides local flow rate and volume totalization and will interface with most displays, PLCs and computers.

Materials of Construction: Body and Internal Wetted Parts: 316L Stainless Steel Bearings: Standard - Nickel Bindery Tungsten Carbide Turbine: Nickel Plated CD4MCU Stainless Steel Shaft: Nickel Bindery Tungsten Carbide Sanitary Seals: Teflon® (CIP/SIP only)Flow Accuracy: ±1.0% of reading Repeatability: ±0.1%Calibration: Water (NIST Traceable Calibration)Pressure Rating: 1,000 psi (Rating based on Tri-clamp sanitary connection)Turbine Temperature: -150 °F to +300 °F (-100 °C to +149 °C)End Connections: Sanitary Clamp End Mag Pick-up: NEMA 6; -150 °F to +300 °F (-100 °C to +149 °C)

Specifications

FloClean Size Chart

Clean-In-Place Diagram Clean-Out-of-Place Diagram

FloClean 3-A (COP) Sanitary Meter with NEMA 6 pick-up

(P.N. B16A-110A-1BA)

FloClean Flow Meter Part Numbering Information

†Limited to combinations shown in the FloClean Size Chart below.*Not available with CIP/SIP

NOTE: Diagrams based on 1-1/2” clamp size with 1”: 5.00 - 50.00 GPM Meter

3-A Sanitary ratingA - COP / SOPB - CIP / SIP

Bearing MaterialA - Ni Bindery Tungsten Carbide

CalibrationA - Standard 5 PointB - 10 PointC - 20 point

Meter Body HubA - 1" NPT Hub*B - No HubPick-up Option

0 - NEMA 6 Magnetic1 - NEMA 6 Magnetic w/Pre-Amp2 - Magnetic (non-Nema 6)3 - Magnetic w/Pre-Amp (non-NEMA 6)4 - Active Sensor (non-NEMA 6) with 4-20 mA6 - High Temp (non-NEMA 6)8 - Active Sensor (non-NEMA 6) 0-5 VDC9 - None

Clamp Size†

0 - 3/4" Ferrule1 - 1-1/2" Ferrule2 - 2-1/2" Ferrule*

Meter Size: Flow Range†

03 - 3/8" : 0.6 - 3.0 GPM05 - 1/2" : 0.75 - 7.50 GPM07 - 3/4" : 2.00 - 15.00 GPM08 - 7/8" : 3.00 - 30.00 GPM10 - 1" : 5.00 - 50.00 GPM15 - 1-1/2" : 15.0 - 180.0 GPM*20 - 2" : 40.0 - 400.0 GPM*

Operating PrincipleFluid entering the meter first passes through an inlet flow straightener thatreduces its turbulent flow pattern. Fluid then passes through the turbine,causing the turbine to rotate at a speed proportional to fluid velocity. As eachturbine blade passes through the magnetic field generated by the meter’smagnetic pick-up, an AC voltage pulse is generated. These pulses providean output frequency that is proportional to volumetric flow.

K-FactorThe K-Factor represents the number of output pulsestransmitted per gallon of fluid passing through the turbine meter. Each turbine has a unique K-Factor. However, turbine meters are not functionally consistent throughout the full flow range of the meter.

There are several forms of “friction” inherent in turbine meters that retard the rotational movement of the turbine rotor. These frictional forces include: magnetic drag, created by electromagnetic force of pick-up transducers; mechanical drag, due to bearing friction; and viscous drag,produced by flowing fluid. See charts below.

As flow increases, the frictional forces are minimized andthe free-wheeling motion of the turbine rotor becomesmore linear (proportional to flow). The K-Factor becomesrelatively constant and linear throughout the balance of thelinear flow range. This is approximately a 10:1 turndownratio from the maximum flow rate down to the minimumflow rate.

Repair Kits

3

2.0(50.8)

2.72(69.1)

Inches (mm) Inches (mm)

3.00(76.2)

4.00(101.6)

4.00(101.6)

Clamp Size MeterSize GPM LPM COP/SOP CIP/SIP

K-factorPulses/Gal

3/4"

3/4"

3/4"

1-1/2"

1-1/2"

1-1/2"

1-1/2"

1-1/2"

2-1/2"

3/8"

1/2"

3/4"

1/2"

3/4"

7/8"

1"

1-1/2"

2"

2.27 - 11.36

3.0 - 30

7.5 - 57

3.0 - 30

7.5 - 57

11 - 110

19 - 190

57 - 680

150 - 1,500

0.60 - 3.00

0.75 - 7.50

2.0 - 15.0

0.75 - 7.50

2.0 - 15.0

3.0 - 30

5.0 - 50.0

15 - 180

40 - 400

20,000

13,000

2,750

13,000

2,750

2,686

870

330

52

B16A-K03A

B16A-K05A

B16A-K07A

B16A-K05A

B16A-K07A

B16A-K08A

B16A-K10A

B16A-K15A

B16A-K20A

B16B-K03A

B16B-K05A

B16B-K07A

B16B-K05A

B16B-K07A

B16B-K08A

B16B-K10A

*

*

FLOW RATE REPAIR KIT PART NUMBER

3.00(76.2)

Typical K-Factor Curve (Pulses per gallon)

PRESSURE DROP

K-FACTOR (PULSES/GALLON)

K-FA

CTOR

(CYC

LES/

GALL

ON)

OUTP

UT FR

EQUE

NCY (

CYCL

ES/S

EC)

% O

F VAL

UE AT

FLOW

RAT

E

100%

50%

0% 10% 50% 100%

STANDARD TURNDOWN APPROX. 10:1

LINEA

RITY

(DES

IGN

DEVIA

TION)

+1%

+%

-%

-1%

AVG. K-FACTOR (K) K-FACTOR (PULSES/GALLON)

FULL FLOW RANGELINEAR FLOW RANGE

Factory calibrated replacement kits are available for field or factory service. Both of the FloClean models are designed to allow for quick, easy disassembly and replacement of the meter's internal components.

A clean-in-place (CIP) and sanitize-in-place (SIP) repair kit contains retaining screws, two rotor supports, one rotor assembly, two gaskets, and a K-factor tag. A clean-out-of-place (COP) and sanitize-out-of-place (SOP) repair kit contains two retaining rings, two rotor supports, one rotor assembly, and a K-factor tag.

For further details, contact the factory or refer to Form #4300.

FloCleanTM

Page 4: Flow Meterscmeters.com/pdf/blancett/FloClean Data.pdf · 2013-05-07 · 1-800-235-1638 Flow Meters 3-A Sanitary Turbine Flow Meter ® FloCleanTM Sanitary Turbine Flow Meter FloCleanTM

1-800-235-1638 www.blancett.com

3-A Sanitary Turbine Flow Meter

FloCleanTM Sanitary Turbine Flow Meter

FloCleanTM

Meets 3-A Sanitary Standard number 28-03

Accuracy of ±1.0% of Reading and Repeatability of ±0.1%

0.6 to 400 gpm flow rates

Models available for CIP/SIP and COP/SOP environments

Materials comply to FDA requirements

Toll Free: 1-800-235-1638 Tel: 1-262-639-6770 Fax: 1-262-639-2267

© 2005 Blancett Printed in USA 4/05 Form No. 1600

Blancett is a registered trademark of Racine Federated Inc. FloClean is a trademark of Racine Federated Inc.Teflon is a registered trademark of E.I. DuPont de Nemours & Co.

8635 Washington Avenue, Racine, WI 53406-3738 U.S.A.

E-mail: [email protected]

Related Blancett Products

Scales turbine meter output to desired engineering unitsAmplifies turbine meter pulse outputConverts frequency outputs into recognizable unitsfor PLCs and other devicesSwitch-selectable or programmable versions availableCSA approved

Microprocessor-basedLinearized output capabilityChoice of 4-20 mA or 0-5 VDC outputEnables integration with data acquisition devicesFrequency measurement accuracy ±1%CSA approved

K-Factor Scaler Frequency Divider

Microprocessor-based flow monitor and totalizerUse with Blancett turbine flow meters as well as otherflow meters with a frequency outputBattery (1.5 VDC) and loop-powered (4-20 mA) versionsMeter, remote, panel and swivel mounting optionsHand-held and explosion-proof models also availableCSA and CE approvedClass 1, Division 1 (Intrinsically Safe) certification