FlashPro-CC Flash Programmer
for the CC series devices - Chipcon product from TI
Remote Control Programming User’s Guide
PM024A02 Rev.2
December-17-2007
Elprotronic Inc.
2
Elprotronic Inc.
16 Crossroads Drive
Richmond Hill,
Ontario, L4E-5C9
CANADA
Web site: www.elprotronic.com
E-mail: [email protected]
Fax: 905-780-2414
Voice: 905-780-5789
Copyright © Elprotronic Inc. All rights reserved.
Disclaimer:
No part of this document may be reproduced without the prior written consent of Elprotronic Inc.
The information in this document is subject to change without notice and does not represent a
commitment on any part of Elprotronic Inc. While the information contained herein is assumed to
be accurate, Elprotronic Inc. assumes no responsibility for any errors or omissions.
In no event shall Elprotronic Inc, its employees or authors of this document be liable for special,
direct, indirect, or consequential damage, losses, costs, charges, claims, demands, claims for lost
profits, fees, or expenses of any nature or kind.
The software described in this document is furnished under a licence and may only be used or copied
in accordance with the terms of such a licence.
Disclaimer of warranties: You agree that Elprotronic Inc. has made no express warranties to You
regarding the software, hardware, firmware and related documentation. The software, hardware,
firmware and related documentation being provided to You “AS IS” without warranty or support
of any kind. Elprotronic Inc. disclaims all warranties with regard to the software, express or implied,
including, without limitation, any implied warranties of fitness for a particular purpose,
merchantability, merchantable quality or noninfringement of third-party rights.
Limit of liability: In no event will Elprotronic Inc. be liable to you for any loss of use, interruption
of business, or any direct, indirect, special incidental or consequential damages of any kind
(including lost profits) regardless of the form of action whether in contract, tort (including
negligence), strict product liability or otherwise, even if Elprotronic Inc. has been advised of the
possibility of such damages.
3
END USER LICENSE AGREEMENT
PLEASE READ THIS DOCUMENT CAREFULLY BEFORE USING THE SOFTWARE AND
THE ASSOCIATED HARDWARE. ELPROTRONIC INC. AND/OR ITS SUBSIDIARIES
(“ELPROTRONIC”) IS WILLING TO LICENSE THE SOFTWARE TO YOU AS AN
INDIVIDUAL, THE COMPANY, OR LEGAL ENTITY THAT WILL BE USING THE
SOFTWARE (REFERENCED BELOW AS “YOU” OR “YOUR”) ONLY ON THE CONDITION
THAT YOU AGREE TO ALL TERMS OF THIS LICENSE AGREEMENT. THIS IS A LEGAL
AND ENFORCABLE CONTRACT BETWEEN YOU AND ELPROTRONIC. BY OPENING THIS
PACKAGE, BREAKING THE SEAL, CLICKING “I AGREE” BUTTON OR OTHERWISE
INDICATING ASSENT ELECTRONICALLY, OR LOADING THE SOFTWARE YOU AGREE
TO THE TERMS AND CONDITIONS OF THIS AGREEMENT. IF YOU DO NOT AGREE TO
THESE TERMS AND CONDITIONS, CLICK ON THE “I DO NOT AGREE” BUTTON OR
OTHERWISE INDICATE REFUSAL, MAKE NO FURTHER USE OF THE FULL PRODUCT
AND RETURN IT WITH THE PROOF OF PURCHASE TO THE DEALER FROM WHOM IT
WAS ACQUIRED WITHIN THIRTY (30) DAYS OF PURCHASE AND YOUR MONEY WILL
BE REFUNDED.
1. License.
The software, firmware and related documentation (collectively the “Product”) is the property of
Elprotronic or its licensors and is protected by copyright law. While Elprotronic continues to own
the Product, You will have certain rights to use the Product after Your acceptance of this license.
This license governs any releases, revisions, or enhancements to the Product that Elprotronic may
furnish to You. Your rights and obligations with respect to the use of this Product are as follows:
YOU MAY:
A. use this Product on many computers;
B. make one copy of the software for archival purposes, or copy the software onto the hard disk
of Your computer and retain the original for archival purposes;
C. use the software on a network
YOU MAY NOT:
A. sublicense, reverse engineer, decompile, disassemble, modify, translate, make any attempt
to discover the Source Code of the Product; or create derivative works from the Product;
B. redistribute, in whole or in part, any part of the software component of this Product;
4
C. use this software with a programming adapter (hardware) that is not a product of
Elprotronic Inc.
2. Copyright
All rights, title, and copyrights in and to the Product and any copies of the Product are owned by
Elprotronic. The Product is protected by copyright laws and international treaty provisions.
Therefore, you must treat the Product like any other copyrighted material.
3. Limitation of liability.
In no event shall Elprotronic be liable to you for any loss of use, interruption of business, or any
direct, indirect, special, incidental or consequential damages of any kind (including lost profits)
regardless of the form of action whether in contract, tort (including negligence), strict product
liability or otherwise, even if Elprotronic has been advised of the possibility of such damages.
4. DISCLAIMER OF WARRANTIES.
You agree that Elprotronic has made no express warranties to You regarding the software, hardware,
firmware and related documentation. The software, hardware, firmware and related documentation
being provided to You “AS IS” without warranty or support of any kind. Elprotronic disclaims all
warranties with regard to the software and hardware, express or implied, including, without
limitation, any implied warranties of fitness for a particular purpose, merchantability, merchantable
quality or noninfringement of third-party rights.
5
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital devices,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one
of more of the following measures:
* Reorient or relocate the receiving antenna
* Increase the separation between the equipment and receiver
* Connect the equipment into an outlet on a circuit different from that to which the receiver is connected
* Consult the dealer or an experienced radio/TV technician for help.
Warning: Changes or modifications not expressly approved by Elprotronic Inc. could void the user’s authority
to operate the equipment.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference and
(2) this device must accept any interference received,
including interference that may cause undesired
operation.
This Class B digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appereil numerique de la classe B respecte toutes les exigences du
Reglement sur le material brouilleur du Canada.
6
Table of Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2. Demo program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3. Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.1 Example with single FPA API DLL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2 Example with Multi-FPA API DLL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4. List of the DLL instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.1 Multi-FPA instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
F_Trace_ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
F_Trace_OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
F_OpenInstances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
F_CloseInstances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
F_OpenInstancesAndFPAs, F_OpenInstances_AndFPAs . . . . . . . . . . . . . . 29
F_API_DLL_Directory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
F_Set_FPA_index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
F_Get_FPA_index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
F_Disable_FPA_index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
F_Enable_FPA_index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
F_LastStatus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
F_Multi_DLLTypeVer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
F_Get_FPA_SN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2 Generic instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
F_Check_FPA_access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
F_DLLTypeVer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
F_Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
F_Close_All . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
F_GetSetup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
F_ConfigSetup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
F_SetConfig . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
F_GetConfig . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
F_DispSetup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7
F_ReportMessage, F_Report_Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
F_GetReportMessageChar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
F_ReadCodeFile, F_Read_CodeFile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
F_Get_CodeCS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
F_ConfigFileLoad, F_Config_FileLoad . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
F_Clr_Code_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
F_Put_Byte_to_Code_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
F_Get_Byte_from_Code_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
F_Put_IEEEAddr64_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
F_Put_IEEEAddr_Byte_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
F_Get_IEEEAddr64_from_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
F_Get_IEEEAddr_Byte_from_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
F_Power_Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
F_Reset_Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
F_Get_Targets_Vcc . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
4.3 Encapsulated instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
F_AutoProgram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
F_Verify_Lock_Bit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
F_Memory_Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
F_Memory_Blank_Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
F_Memory_Write . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
F_Memory_Verify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
F_Memory_Read . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
F_Write_IEEE_Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
F_Read_IEEE_Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
F_Write_Lock_Bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4.4 Sequential instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
F_Open_Target_Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
F_Close_Target_Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
F_Segment_Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
F_Sectors_Blank_Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
F_Write_Byte_to_XRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
F_Read_Byte_from_XRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
F_Write_Byte_to_direct_RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
F_Read_Byte_from_direct_RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
F_Copy_Buffer_to_Flash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
F_Copy_Flash_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
F_Copy_Buffer_to_XRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8
F_Copy_XRAM_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
F_Copy_Buffer_to_direct_RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
F_Copy_direct_RAM_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
F_Put_Byte_to_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
F_Get_Byte_from_Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
F_Set_PC_and_RUN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
F_Get_MCU_Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Appendix A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
FlashPro-CC Command Line interpreter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
9
1. Introduction
FlashPro-CC Flash Programmer (USB ) can be remotely controlled from other
software applications (Visual C++, Visual Basic etc.) via a DLL library. The Multi-FPA - allows
to remotely control simultaneously up to eight Flash Programming Adapters (FPAs)
significantly reducing programming speed in production.
Figure 1.1 shows the connections between PC and up to eight programming adapters. The
FPAs can be connected to PC USB ports directly or via USB-HUB. Direct connection to the PC
is faster but if the PC does not have required number of USB ports, then USB-HUB can be used.
The USB-HUB should be fast, otherwise speed degradation can be noticed. When the USB hub
is used, then the D-Link’s Model No: DUB-H7, P/N BDUBH7..A2 USB 2.0 HUB is
recommended.
Figure 1.1
10
Remote Control Programming User’s Guide PM024A02 Rev.2 11
Block diagram of the Multi-FPA application DLL is presented on the Figure 1.2.
To support the Multi-FPA API-DLL feature, the software package contains nine dll files
- the Multi-FPA API-DLL selector
- eight standard single FPAs API-DLLs
Figure 1.3 shows the logical connections between dll files.
Figure 1.2
Remote Control Programming User’s Guide PM024A02 Rev.2 12
The main FlashProCC-FPAsel.dll (Multi-FPA selector) allows to transfer API-DLL
functions coming from an application software to desired single application dll (FlashProCC-
FPA1.dll to FlashProCC-FPA8.dll).
Note: Software package contains one FlashProCC-FPA1.DLL. Files FlashProCC-
FPA2.DLL to FlashProCC-FPA8.DLL will be copied automatically if required.
The FlashProCC-FPAsel.dll is transparent for all API-DLL functions implemented in the
single API-DLLs functions. Desired destination FPA can be selected using the selector function
added to the Multi-FPA selector (FlashProCC-FPAsel.dll).
F_Set_FPA_index( fpa );
where the
Figure 1.3
Remote Control Programming User’s Guide PM024A02 Rev.2 13
fpa = 1 to 8 when the only one desired FPA required to be selected
or
fpa = 0 when ALL active FPAs should be selected.
The selected FPA index modified by the F_Set_FPA_index( fpa ) instruction can be
modified at any time. By default, the FPA index is 1 and if only one FPA is used then fpa index
does not need to be initialized or modified. When the fpa index 1 to 8 is used, then the result is
coming back to application software from the single API-DLL via transparent Multi-FPA
selector. When the fpa index is 0 (ALL-FPAs) and results are the same from all FPAs, then the
same result is passing back to application software. If results are not the same, then the Multi-
FPA selector DLL is returning value -1 (minus 1) and all recently received results can be read
individually using function
F_LastStatus( fpa )
Most of the implemented functions allows to use the determined fpa index 1 to 8 or 0
(ALL-FPAs). When functions return specific value, like read data etc, then only determined FPA
index can be used ( fpa index from 1 to 8). When the fpa index is 0 (ALL-FPAs) then almost all
functions are executed simultaneously. Less critical functions are executed sequentially from
FPA-1 up to FPA-8 but that process can not be seen from the application software.
When the inactive fpa index is selected, then return value from the selected function is -2
(minus 2). When all fpa has been selected (fpa index = 0) then only active FPAs will be serviced.
For example if only one FPA is active and fpa index=0, then only one FPA will be used. It is
save to prepare the universal application software that allows to remote control up to eight FPAs
and on the startup activate only desired number of FPAs.
It should be noticed, that all single API-DLLs (FlashProCC-FPA1.dll to FlashProCC-
FPA8.dll) used with the Multi-FPA DLL (FlashProCC-FPAsel.dll) are fully independent to each
other. From that point of view it is not required that transferred data to one FPA should be the
same as the transferred data to the others FPAs. For example code data downloaded to FPA-1 can
be different that the code data downloaded to the FPA-2, FPA-3 etc. But even in this case the
programming process can be done simultaneously. In this case the desired code should be read
from the code file and saved in the API-DLL-1, next code file data should be saved in the API-
DLL-2 etc. When it is done, then the F_AutoProgram can be executed simultaneously with
selected all active FPAs. All FPAs will be serviced by his own API-DLL and data packages
saved in these dlls.
Remote Control Programming User’s Guide PM024A02 Rev.2 14
The FlashPro-CC Flash Programmer software package contains all required files to
remotely control programmer from a software application. When software package is installed
then by default the DLL file, library file and header file are located in:
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\API-DLL
FlashProCC-FPAsel.dll - Multi-FPA selection / task manager API- DLL
FlashProCC-FPA1.dll - Single api-DLL for the UAB-FPA
FlashProCC-Dll.h - header file for C++
FlashProCC-Dos-Dll.h - header file for C++ (Borland) or DOS
FlashProCC-FPAsel.lib - lib file for C++
FlashProCC-FPAsel-BC.lib - lib file for C++ (Borland)
config.ini - default configuration file for the FPAs
FPAs-setup.ini - FPAs- vs USB ports configuration file
The FlashProCC-FPAsel.dll contains two groups of the same functions used in C++
application and Visual Basic (or similar) applications All procedure names used in Visual Basic
are starting from VB_xxxx, when the procedure names used in C++ are starting from F_xxxx.
All functions starting from F_xxxx using the _Cdecl declarations used in C++ . Function names
starting from VB_xxxx has the _stdcall calling declaration required in Visual Basic.
Reminding files listed above are required in run time - to initialize the flash programming
adapter (config.ini) and USB setup (FPAs-setup.ini).
When the C++ application is created, then following files should be copied to the source
application directory:
FlashProCC-Dll.h - header file for C++
FlashProCC-FPAsel.lib - lib file for C++ (Microscoft Visual C++)
or
FlashProCC-Dos-Dll.h - header file for C++
FlashProCC-FPAsel-BC.lib - lib file for C++ (Borland C++)
and to the release/debug application directory
FlashProCC-FPAsel.dll - Multi-FPA selection / task manager DLL
FlashProCC-FPA1.dll - API-DLL for the USB-FPA
config.ini - default configuration file for the FPAs
Remote Control Programming User’s Guide PM024A02 Rev.2 15
FPAs-setup.ini - FPAs- vs USB ports configuration file
Executable application software package in C++ or when application in Visual Basic is created,
then following files should be copied to the source or executable application directory:
FlashProCC-FPAsel.dll
FlashProCC-FPA1.dll
config.ini
FPAs-setup.ini
All these files ‘as is’ should be copied to destination location, where an application software
using the DLL library.
The config.ini file has default setup information. This file can be modified and taken directly
form the FlashPro-CC Flash Programmer application software. To create required config.ini file
the standard FlashPro-CC Flash programmer software should be open and required setup
(memory option, communication speed etc) should be created. When this is done, programming
software should be closed and the config.ini file with the latest saved configuration copied to
destination location. Note, that the configuration setup can be modified using DLL library
function.
Software package has a demo software written under Visual C++.net. All files and source
code are located in:
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\API-DLL-Demo\Cpp
Remote Control Programming User’s Guide PM024A02 Rev.2 16
2. Demo program
The demo program is a small GUI program with a lot of buttons allowing to separately
call functions using DLL library package software. Source code and all related project files are
located in the following directory:
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\API-DLL-Demo\Cpp
Program can be activated by selecting the FlashProCC-DLL-DemoCpp.exe located in the \release
subdirectory. This demo program can also be activated from the windows menu:
Start->Program->Elprotronic-Flash Programmer->(CCxx) USB FlashPro-CC->FlashProCC-DLL-Demo-
Cpp
Remote Control Programming User’s Guide PM024A02 Rev.2 17
When the demo program is activated, then the following dialogue screen is displayed (see figure
2.1). At the beginning the USB-FPA’s configuration file should be created, that contains list off
all FPAs used in the application. Using the Notepad editor the default FPA configuration file
‘FPAs-setup.ini’ should be opened by pressing the “Open FPAs-setup.ini” button in the
dialogue screen.
Take a serial numbers from the FPAs labels and write it on the desired FPAs locations FPA-1 up
to FPA-8. If for example two FPAs will be used with SN 20070031 and 20070032 then the
contents of the FPA’s configuration file will be as follows:
;===================================================================
; USB-FPA configuration setup *
; Elprotronic Inc. *
;-------------------------------------------------------------------
; up to eight FPA can be specified and connected via USB to PC *
Figure 2.1 Demo program dialogue screen using DLLs.
Remote Control Programming User’s Guide PM024A02 Rev.2 18
; syntax: *
; FPA-x Serial Number *
; where FPA-x can be FPA-1, FPA-2, FPA-3 .... up to FPA-8 *
; Serial number - get serial number from the desires FPA's label *
; Minimum one FPA's must be specified *
; FPA-x order - any *
; *
; e.g (without semicolon - comment) *
; *
;FPA-1 20050116 *
;FPA-3 20050199 *
;FPA-5 20050198 *
;===================================================================
FPA-1 20070031
FPA-2 20070032
Note, that only lines without comments (without semicolon on the front ) are used by
software. All lines with comment are ignored. The FPA’s serial numbers and FPA’s indexes can
be listed in any order and with gap like FPA-1, FPA-5 etc. without FPA-2, 3 etc. Minimum one
valid FPA with correct SN must be specified. Up to eight FPAs can be declared. When the FPA’s
configuration file is created then file should be saved using name starting from FPA and with
extention ini e.g FPAs-setup.ini.
Connect all required FPA’s to USB connectors and run the FlashProCC-
DLL-Demo-Cpp.exe demo software. First the DLL instances should be opened and all
connected FPA’s should be assigned to desired FPA’s indexes. It is recommended to press the
button ‘FPA assigment’ located inside the frame named ‘From File (recommended)’. When this
button is pressed, then the DLL function named
F_OpenInstancesAndFPAs( FileName );
is called. The list of defined FPA’s serial numbers are taken from the FPAs configuration file and
assigned all FPAs to desired FPA indexes (1 to 8). Number of instances to be opened is
calculated automatically, one per available and valid FPA. On described example with two FPAs
in the ‘FPAs selector’ will display two valid FPAs with list of used FPAs’ serial numbers. All,
others FPA-x fields will be disabled. In this example only two DLL instances becomes opened.
Valid FPA indexes becomes 1,3 and ALL.
Note: When one or more FPA adapters are connected to PC and the “FPAs-setup.ini”
does not contain valid FPA serial numbers, then the first detected FPA (and only one)
will be activated.
Remote Control Programming User’s Guide PM024A02 Rev.2 19
Other method (old method - not recommended) that allows to open required number of
instances uses ‘2. Open Instances’ button in ‘Default’ frame. First the number of the instances
should be defined in the ‘Total no of FPAs’ location. When the ‘2. Open Instances’ button is
pressed, then the DLL function named
F_OpenInstances( no );
called where ‘no’ - number of instances to be opened.. When the dll instances are opened, then it
is possible to check the access to the FPA connected to PC via USB ports. Pressing button ‘3.
FPA list’ ( function F_Check_FPA_access(index); called in loop for index = 1,2,3..8 ) allows
to check the access to these adapters. On the end the button ‘4. FPA assignment’ ( function
F_Check_FPA_access(index); with desired ‘fpa’ and USB indexes) allows to assign desired
FPA adapter to ‘fpa’ index. All these steps can be done automatically when the function
F_OpenInstancesAndFPAs( FileName ) described above is used (used button ‘FPA
assigment’ located inside the frame named ‘From File (recommended) ).
When the FPA(s)s has been assigned then all adapters should be activated by pressing the
‘1. Initialization’ button. This initialization calls the DLL function F_Initialization and
communication between programming adapter and PC is established. Report message is
displayed in the report window ( uses the F_ReportMessage function). By default the config.ini
file is empty and to make a required programmer setup the setup file should be downloaded to
programmer. It can be done by pressing the button ‘ 2.Setup File’, (executing
F_ConfigFileLoad DLL function). Setup file can be prepared first using standard FlashPro-CC
programming software with GUI. Also desired code file can be downloaded by pressing the
button ‘3. Open Code’ (executing F_ReadCodeFile DLL function).
There are seven buttons located on the right side of demo dialogue screen. Each of them
calls one encapsulated function from the following list - F_AutoProgram, F_Memory_Erase,
F_Memory_Blank_Check, F_Memory_Write, F_Memory_Verify and F_Memory_Read
When any of these button is pressed, then a function, exactly in the same way how it is
done in the standard FlashPro-CC Flash Programming software (GUI) is executed. Also buttons
Power ON/OFF, RESET has the same action as related buttons in standard programmer. Refer
to the FlashPro-CC Flash Flash Programmer for the CC series devices - User’s Manual for
details of these functions.
In the central part on dialogue screen there are buttons that can call the sequential DLL
functions.
Button Open Target - F_Open_Target_Device();
Remote Control Programming User’s Guide PM024A02 Rev.2 20
Button Erase Segment - F_Segment_Erase(....);
Button Blank Check Segm. - F_Sectors_Blank_Check(....);
Button Write Flash Block - F_Copy_Buffer_to_Flash(....);
Button Read Flash Block - F_Copy_Flash_to_Buffer(....)
Button Write XRAM Block - F_Copy_Buffer_to_XRAM(....);
Button Read XRAM Block - F_Copy_XRAM_to_Buffer(....);
Button Write direct RAM - F_Copy_Buffer_to_direct_RAM(....);
Button Read direct RAM - F_Copy_direct_RAM_to_Buffer(....);
Button Set PC and Run - F_Set_PC_and_RUN(....);
Button Close Target - F_Close_Target_Device();
When a sequential function is called then Open Target (calling F_Open_Target_Device
DLL function) must be pressed first. After that any button calling a sequential function can be
pressed - in any order and as many times as required. On the end of sequential communication
the button Close Traget (calling F_Close_Target_Device DLL function) should be pressed.
In the presented demo software all sequential functions have very small task to perform to
demonstrate how to use the DLL functions. See source code of the DLL-Demo program in the
software package in the ..\Demo-DLL subdirectory for details.
Erase Segment: Erase segment at location 0x1000 (segment size 1 or 2k )
Blank Check Segm. Segment blank check Erase at location 0x1000 to 0x107F
Write Flash Block Write 16 bytes to the flash memory at location 0x1020 to 0x102f using
function
F_Copy_Buffer_to_Flash( 0x1020, 16);
Following data are written no targets:
0x1020 -> 10 11 12 13 14 15 16 17 18 19 1A 1B 1C 1D 1E 1F
Read Flash Block Read 64 bytes from target device starting from the flash memory address
at 0x1000. On the report screen only 16 bytes from each target devices
taken from addresses 0x1020 to 0x102F are displayed.
Write XRAM Block Write 16 bytes to XRAM at location 0xF000 to 0xF00F using function
F_Copy_Buffer_to_XRAM( 0xF000, 16);
Following data are written no targets:
30 31 32 33 34 35 36 37 38 39 3A 3B 3C 3D 3E 3F
Remote Control Programming User’s Guide PM024A02 Rev.2 21
Read XRAM Block Read 16 bytes XRAM starting from address at 0xF000. On the report
screen 16 bytes from each target devices taken from addresses 0xF000 to
0xF00F are displayed
Write direct RAM Write 16 bytes to direct RAM at location 0x60 to 0x6F using function
F_Copy_Buffer_to_direct_RAM( 0x60, 16);
Following data are written no targets:
40 41 42 43 44 45 46 47 48 49 4A 4B 4C 4D 4E 4F
Read direct RAM Read 16 bytes direct RAM starting from address at 0x60. On the report
screen 16 bytes from each target devices taken from addresses 0x60 to
0x6F are displayed
Set PC and Run Small program is written and downloaded to each target devices.
Following program is saved in the XRAM at location stated from 0xF100.
//Downloaded code to RAM - address 0xF100
0xC2, 0xAF, // clr EA (Enable mask)
0xE4, // clr A
0x90, 0xF0, 0x00, // mov DPTR, #0xF000
0x79, 0x10, // mov R1, #0x10
loop:
0xF0, // movx @dptr,A
0x04, // inc A
0xA3, // inc dptr
0xD9,0xFB, // DJNZ R1,loop
//done, fake a breakpoint
0xA5, // DB 0xA5
0x00, // NOP
Written program is modifying contents of the XRAM at locations 0xF000
to 0xF00F. The new XRAM vales should be 0x00, 0x01, 0x02,..... 0x0F.
When the Set PC and Run program is pressed, then the contents described
above is downloaded to XRAM and function
Set_PC_and_RUN( 1, 0xF100 );
Remote Control Programming User’s Guide PM024A02 Rev.2 22
is executed. When program is finished, then XRAM at location should be
modified. XRAM contents can be read by pressing the Read XRAM Block
button.
To make a test do following steps (from the beginning):
1. FPA List
2. Initialization
3. Setup File
4. Power from FPA - Enable
5. Open Target
6. Write XRAM Block
7. Read XRAM Block ( remember XRAM contents)
8. Set PC and RUN
9. Read XRAM Block ( compare with the contents from point 7 ).
On the right part of the dialogue screen are located buttons with encapsulated functions
like Autoprogram, Erase Flash etc. Encapsulated functions are not requires to call the
F_Open_Target_Device() function. All functions, including “Open..”, “Close...” are build-in
inside the encapsulated functions (see chapter 4 for details).
Remote Control Programming User’s Guide PM024A02 Rev.2 23
3. Getting Started
The Multi-FPA API-DLL software package is the same for the application written under
Visual C++, Visual Basic, LabView etc. When the desired application is created then all files
from the Elprotronic directory
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\API-DLL
should be copied to the executable destination directory.
It is recommended to start the standard FlashPro-CC (GUI) programming software to
verify if the hardware and the drivers setup are correct. Using the FlashPro-CC programming
software the fully functional setup that satisfy desired requirements should be created. When it is
done then using the “File->Save Setup as..” option the configuration file should be saved. This
file can be used “as is” in the application uses Multi-FPA API-DLL. Copy and paste the required
configuration to your destination directory, where your application software is installed. It is
recommended to use the demo program and verify if the setup in your PC and destination
directory are done correctly. To do that copy the executable file
FlashProCC-DLL-DemoCpp.exe
from location
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\API-DLL-Demo\Cpp\release
to your destination location where your application software is installed. Run the demo program.
Follow instruction described in chapter 2 how to use the demo program.
3.1 Example with single FPA API DLL
The API-DLL always uses two API-DLL - selector DLL and AOI-DLL. When one FLA
is used only, then selector DLL always sould select the first FPA. Also it is recommended to use
the first detected FPA, since only one FPA is typically connected to PC. The application software
can be simplified in this case. All instructions related to single FPA are detailed described in the
chapters 4.2, 4.3 and 4.4. Instructions specific to Multi-FPA features described in the chapter 4.1.
Initialization opening procedure for the USB-FPA can be as follows:
Remote Control Programming User’s Guide PM024A02 Rev.2 24
F_OpenInstancesAndFPAs( “*# *” ); // DLL and FPA (one only) initialization
F_Set_FPA_index( 1 ); // select FPA 1 for
F_Initialization( ); // init FPA
Below is an example of the simplified (without error handling procedures) application
program written in C++ that allows to initialize one FPA, and run an autoprogram with the same
features like an autoprogram in the standard FlashPro-CC (GUI) software.
1. Download data to target device
F_OpenInstancesAndFPAs( “*# *” ); // DLL and FPA (one only) initialization
F_Set_FPA_index( 1 ); // select FPA 1
F_Initialization( ); // init FPA
F_ReadConfigFile( filename ); // read configuration data and save
// to API-DLL
F_ReadCodeFile( format, filename ); // read code data and save to DLL
do
{
status = F_AutoProgram(1); //start autoprogram
if ( status != TRUE )
{
status = F_LastStatus(1);
................................................
}
else
{
................................................
}
} while(1); //make an infinite loop until last target device programmed
.................................................................
F_CloseInstances();
3.2 Example with Multi-FPA API DLL
The code example described below uses Multi-FPA API-DLL. The Multi-FPA API-DLL
is a shell that allows to transfer incoming instructions from the application software to desired
FPA. All instructions related to single FPA are detailed described in the chapters 4.2, 4.3 and 4.4.
Instructions specific to Multi-FPA features described in the chapter 4.1.
Remote Control Programming User’s Guide PM024A02 Rev.2 25
Initialization opening procedure for the USB-FPA can be as follows: F_OpenInstancesAndFPAs( FPAs-setup.ini); // DLL and FPA initialization
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all FPA’s
F_Initialization( ); // init all FPA’s
In the example above number of the opened USB-FPAs are specified in the ‘FPAs-setup.ini’
Below is an example of the simplified (without error handling procedures) application
program written in C++ that allows to initialize all dlls and FPA, and run an autoprogram with
the same features like an autoprogram in the standard FlashPro-CC (GUI) software.
1. Download data to all target devices (uses USB-FPAs)
F_OpenInstancesAndFPAs( FPAs-setup.ini); // DLL and FPA initialization
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all FPA’s
F_Initialization( ); // init all FPA’s
F_ReadConfigFile( filename ); // read configuration data and save
// to all API-DLLs
F_ReadCodeFile( format, filename ); // read code data and save to all
// API-DLLs
do
{
status = F_AutoProgram( 1 );
//start autoprogram-to program all targets simultaneously with
//the same downloaded data to all target devices.
if ( status != TRUE )
{
if ( status == FPA_UNMACHED_RESULTS )
{
for (n=1; n<=MAX_FPA_INDEX; n++ ) status[n] = = F_LastStatus( n);
................................................
}
else
{
................................................
}
}
} while(1); //make an infinite loop until last target device programmed
.................................................................
F_CloseInstances();
Note, that all single API-DLL are independent from each others and it is not required that all
data and configuration should be the same for each API-DLLs (each FPAs, or target devices) .
For example - code data downloaded to the target devices connected to first FPA can be the same
Remote Control Programming User’s Guide PM024A02 Rev.2 26
(but it is not required) as code data downloaded to the target devices connected to second FPA
etc. In the example below the downloaded code to target devices are not the same .
2. Download independent data to target devices (uses USB-FPAs)
F_OpenInstancesAndFPAs( FPAs-setup.ini); // DLL and FPA initialization
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all FPA’s
F_Initialization( ); // init all FPA’s
.................................................................
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all FPA’s
F_ReadConfigFile( filename ); // read configuration data and save
// to all API-DLLs
F_Set_FPA_index( 1 ); // select FPA 1
F_ReadCodeFile( format, filename1 ); // read code data and save to
// API-DLL-1
F_Set_FPA_index( 2 ); // select FPA 2
F_ReadCodeFile( format, filename2 ); // read code data and save to
// API-DLL-2
................................................................
F_Set_FPA_index(7 ); // select FPA 7
F_ReadCodeFile( format, filename7 ); // read code data and save to
// API-DLL-7
F_Set_FPA_index( 8 ); // select FPA 8
F_ReadCodeFile( 8, format, filename8 ); // read code data and save to
// API-DLL-8
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all FPA’s
do
{
status = F_AutoProgram( 1 );
//start autoprogram - to program all targets simultaneously
//with the independent downloaded data to all target devices.
if ( status != TRUE )
{
if ( status == FPA_UNMACHED_RESULTS )
{
for (n=1; n<=MAX_FPA_INDEX; n++ ) status[n] = = F_LastStatus( n);
................................................
}
else
{
................................................
}
}
} while(1); //make an infinite loop until last target device programmed
.................................................................
F_CloseInstances();
Remote Control Programming User’s Guide PM024A02 Rev.2 27
4. List of the DLL instructions
All DLL instructions are divided to four groups - related to Multi-FPA selector, single
FPA generic, single FPA encapsulated and single FPA sequential instructions. Multi-FPA
specific instructions are related to the Multi-FPA DLL only. Generic instructions are related to
initialization programmer process, while encapsulated and sequential instructions are related to
target device’s function. Encapsulated and sequential instructions can write, read, and erase
contents of the target device’s flash memory.
Multi-FPA specific instructions are related to load and release the single-FPA dlls,
selection of the transparent path and sequential/simultaneous instructions transfer management.
All other instructions are related to single FPAs.
Generic instructions are related to initialization programmer process, configuration setup
and data preparation, Vcc and Reset to the target device. Generic instructions should be called
first, before encapsulated and sequential instruction.
Encapsulated instructions are fully independent executable instructions providing access
to the target device. Encapsulated instructions can be called at any time and in any order. When
called then all initialization communication with the target device is starting first, after that
requested function is executed and at the end communication with the target device is terminated
and target device is released from the programming adapter.
The encapsulated functions should be mainly used for programming target devices. These
functions perform most tasks required during programming in an easy to use format. These
functions use data provided in Code Files, which should be loaded before the encapsulated
functions are used. To augment the functionality of the encapsulated functions, sequential
functions can be executed immediately after to complete the programming process.
Sequential instructions allow access to the target device in a step-by-step fashion. For
example, a typical sequence of instructions used to read data from the target device would be to
open the target device, then read data and then close the target device. Sequential instruction
have access to the target device only when communication between target device and
programming adapter is initialized. This can be done when Open Target Device instruction is
called. When communication is established, then any number of sequential instruction can be
called. When the process is finished, then at the end Close Target Device instruction should be
called. When communication is terminated, then sequential instructions can not be executed.
Note: Inputs / outputs has been defined as INP_X defined as 4 bytes long (see header file )
#define INP_X _int32
Make sure that an application using the DLL file has the same length of desired data.
Remote Control Programming User’s Guide PM024A02 Rev.2 28
4.1 Multi-FPA instructions
The Multi-FPA API-DLL instructions are related to Multi-FPA selector only. These
instructions allows to initialize all single applications dlls and select the instruction patch
between application software and desired FPA and sequential/simultaneous instructions transfer
management Up to eight independent FPAs can be remotely controlled from the application
software. All instructions from application software can be transferred to one selected FPA or to
all FPAs at once. That feature allows to increase programming speed up to eight times and also
allows to have individual access to any FPA is required.
F_Trace_ON
F_Trace_ON - This function activate the tracing.
Syntax:
void MSPPRG_API F_Trace_ON( void );
The F_Trace_ON() opens the DLLtrace.txt file located in the current directory and records all
API-DLL instructions called from the application software. This feature is useful for debugging.
When debugging is not required then tracing should be disabled. Communication history
recorded in the in the last session can be viewed in the DLLtrace.txt located in the directory
where the API-DLL file is located. When the new session is established then the file
DLLtrace.txt is erased and new trace history is recorded.
Note: Tracing is slowing the time execution, because all information passed from application
software to API-DLL are recorded in the dlltrace.txt file.
F_Trace_OFF
F_Trace_OFF - Disable tracing, See F_Trace_ON for details.
Syntax:
void MSPPRG_API F_Trace_OFF( void );
F_OpenInstances
Remote Control Programming User’s Guide PM024A02 Rev.2 29
F_OpenInstances - API-DLL initialization in the PC.
Instruction must be called first - before all other instruction. Instead this function the
F_OpenInstancesAndFPAs can be used.
Important: It is not recommended to use this function. Function used only for compatible with
the old software. Use the F_OpenInstancesAndFPAs instead.
Do not use the F_OpenInstances or F_Check_FPA_access after using the
F_OpenInstancesAndFPAs. The F_OpenInstancesAndFPAs is assigning the FPAs to
USB ports and it is not recommended to reassign once again the USB port using the
F_Check_FPA_access function. To check the communication activity with FPA use the
F_Get_FPA_SN function that allows to check te communication with the FPA adapter
without modifying the USB ports assignment.
Syntax: INT_X MSPPRG_API F_OpenInstances ( BYTE no );
Parameters: no -> number of the single API-DLL to be open
no -> 1 to MAX_USB_DEV_NUMBER
where MAX_USB_DEV_NUMBER = 16
Return value:number of opened instances
F_CloseInstances
F_CloseInstances - Close all active API-DLLs and free system memory.
Syntax: INT_X MSPPRG_API F_CloseInstances ( void );
Parameters: void no -> 1 to MAX_USB_DEV_NUMBER
where MAX_USB_DEV_NUMBER = 16
Return value:TRUE
F_OpenInstancesAndFPAs, F_OpenInstances_AndFPAs
Remote Control Programming User’s Guide PM024A02 Rev.2 30
F_OpenInstancesAndFPAs - API-DLL initialization in the PC and FPA scan and
or F_OpenInstances_AndFPAs assignment to desired USB port according to contents of the
FPA’s list specified in the string or FPA’s configuration file.
Instruction must be called first - before all other instruction. Instead this function the
F_OpenInstances can be used. Function can be used only when the USB FPA are used. When the
USB-FPA is used, then the function F_OpenInstancesAndFPAs is recommended in the initialization
process. Function is very convenient - automatically is opening the number of the desired API-DLL
and assigning the desired FPA to available USB ports. Regardless of the USB port open sequence
and connection of the FPA to USB ports, the F_OpenInstancesAndFPAs instruction is reading the
FPA’s list, scanning all available FPAs connected to any USB ports and assigning the indexes to all
FPAs according to contents of the FPA list (from strinc or configuartion file). All FPAs not listed
in the FPA configuration file and connected to USB ports are ignored.
Important: Do not use the F_Check_FPA_access after using the F_OpenInstancesAndFPAs.
The F_OpenInstancesAndFPAs is assigning the FPAs to USB ports and it is not
recommended to reassign once again the USB port using the F_Check_FPA_access
function. To check the communication activity with FPA use the F_Get_FPA_SN
function that allows to check te communication with the FPA adapter without
modifying the USB ports assignment.
Syntax: INT_X MSPPRG_API F_OpenInstancesAndFPAs( char * List );
INT_X MSPPRG_API F_OpenInstances_AndFPAs( CString List );
Parameters:
1. When the first two characters in the List string are *#, then string reminding characters
contains list of desired FPAs serial numbers assigned to FPA-1, -2, ...-n indexes, eg.“*# 20060123, 20060234, 20060287"
2. When the first two characters in the List string are not *#, then string contains file name
or full path of the file with list of the FPA’s serial numbers, eg. “C:\Program Files\Elprotronic\FPAs-setup.ini”
Return value:number of opened instances
1. The FPA list in the string:
Remote Control Programming User’s Guide PM024A02 Rev.2 31
String -> “*# SN1, SN2, SN3, SN4, SN5...”
Where the
SN1- FPA’s serial number that should be assigned to FPA-1 index
SN2- FPA’s serial number that should be assigned to FPA-2 index
etc.
As a delimiter the comma ‘,’ or white space ‘ ’ can be used.
Example:
“*# 20060123, 0, 20060346, 20060222, 20060245"
or
“*# 20060123 0 20060346 20060222 20060245"
In example above the FPAs will be assigned as follows: FPA-1 20060123
FPA-2 0 //empty - FPA is not assigned
FPA-3 20060346
FPA-4 20060222
FPA-5 20060245
In the FPA list can be specified ONE adapter with any serial number when the character ‘*’ is used
instead the FPA’s serial number. Only one ‘*’ character can be specified in the FPA list and must
be located on the end of valid SN list. All other serial numbers specified after ‘*’ will be ignored.
This option allows to specify any FPA when the only one adapter is used eg.
“*# *"
FPA-1 -> Any FPA is one adapter is connected, or the first detected adapter, if more then one adapters are
connected.
or if more then one adapter is used
“*# 20060123 *" FPA-1 20060123
FPA-2 - first detected adapter (excluding already assigned adapters), if
more adapters are connected.
When the ‘*’ is inside the FPA list, eg.
“*# 20060123 * 20060137 20060166"
then the last two FPA’s SN will be ignored FPA-1 20060123
FPA-2 - first detected adapter (excluding already assigned adapters), if
more adapters are connected.
FPA-3 - not assigned
Initialization example:
1. F_OpenInstances_AndFPAs( “*# *” ); // only one FPA - any SN
or
Remote Control Programming User’s Guide PM024A02 Rev.2 32
2. F_OpenInstances_AndFPAs( snlist ); // hardcoded SN list
or
3. // scanned available FPA’s SN list
long SN[MAX_USB_DEV_NUMBER+1], Snr[MAX_USB_DEV_NUMBER+1];
CString snlist;
char * buf[20];
F_OpenInstances( 1 ); // DLL initialization - one instance
F_Set_FPA_index( 1 ); // select access to the first instance
n = 0; //number of detected FPAs
for( k=1; k<=MAX_USB_DEV_NUMBER ; k++ )
{
SN[k] = F_Check_FPA_access(k);
if ( SN[k] > 20000000 ) n++;
}
F_CloseInstances();
// write your own procedure .......
// remap available FPAs SN to desired FPAs order from SN[k] to Snr[p]
snlist = “*#”;
for( k=1; k<=n ; k++ )
{
sprintf( buf, “ %8.8li”, Snr[k] );
snlist += buf;
}
F_OpenInstances_AndFPAs( snlist );
2.The FPA list in the configuration file:
String -> “C:\Program Files\Elprotronic\FPAs-setup.ini”
Example of the FPA configuration file:
; -> semicolon - comment
; Syntax of the FPAs configuration specified
; FPA-x Serial Number
; where FPA-x can be FPA-1, FPA-2, FPA-3 .... up to FPA-8
; e.g
FPA-1 20050116
FPA-3 20050199
FPA-5 20050198
FPA-6 20050205
; FPA-x can be listed in any order and can contain gaps,
; like above without FPA-2, FPA-4
; When list like above is used, then following fpa can be valid
; fpa -> 1,3,5,6
Remote Control Programming User’s Guide PM024A02 Rev.2 33
; NotePad editor can be used to create the FPA configuration file.
When the ‘*’ is used instead FPA’s SN, then any FPA will be accepted. The ‘*’ can be used only
once and on the end of the FPA’s list eg.
FPA-1 20050116
FPA-3 20050199
FPA-5 *
or
FPA-1 *
when only one adapter (any adapter) is used.
Example:F_OpenInstancesAndFPAs( FPAs-setup.ini );
//DLL startup and FPA assignment
F_Set_FPA_index (ALL_ACTIVE_FPA);
//select all available FPAs
F_Initialization();
//init all FPAs
F_ReadConfigFile( filename );
//download the same configuration to all DLLs.
F_ReadCodeFile( format, filename );
//download the same code file to all DLLs.
do
{
status = AutoProgram(1);
//start autoprogram to all FPAs simultaneously.
if( status != TRUE )
{
if( status == FPA_UNMATCHED_RESULTS )
{
// service software when results from FPAs are not the same
}
else
{
}
.......................
{
.......................
} while(1);
F_CloseInstances();
// release DLLs from memory
Remote Control Programming User’s Guide PM024A02 Rev.2 34
F_API_DLL_Directory
F_API_DLL_Directory - The DLL directory location.
VALID FPA index - irrelevant - the same directory location for all DLLs.
The F_API_DLL_Directory command can specify the directory path where the DLLs are
located. This command is not mandatory and usually is not required. But in some application
software (like in the LabVIEW) the default location of the DLLs is not transferred to the DLL. In
this case the related files with DLLs like MSPlist.ini located in the same directory where the
DLLs are located can not be find. To avoid this problem the full path of the directory where the
DLLs are located can be specified. The F_API_DLL_Directory must be used before
F_Initialization() function.
Syntax:MSPPRG_API void F_API_DLL_Directory( Cstring APIDLLpath );
or MSPPRG_API void F_APIDLL_Directory( char* APIDLLpath );
Example:
............................
F_API_DLL_Directory( “C:\\Program Files\\Test\\LabVIEW” );
// directory where the API-DLLs and MSPlist.ini are located
// C:\Program Files\Test\LabVIEW
If( F_Initialization() != TRUE ) //required API-Dll - initialization
{
// Initialization error
}
F_Set_FPA_index
F_Set_FPA_index - Select desired FPA index (desired DLL instance)
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs).
Syntax: INT_X MSPPRG_API F_Set_FPA_index ( BYTE fpa );
Parameters: fpa -> 1 to MAX_FPA_INDEX where MAX_FPA_INDEX = 8
or 0 -> ALL_ACTIVE_FPA
note: instead of ‘0' value it can be used global defined ALL_ACTIVE_FPA that is defined as
Remote Control Programming User’s Guide PM024A02 Rev.2 35
#define ALL_ACTIVE_FPA 0
in the header file
Return value:TRUE - if used fpa index is valid
FPA_INVALID_NO - if used fpa index is not activated or out of range
note: FPA_INVALID_NO -> -2 (minus 2)
F_Get_FPA_index
F_Get_FPA_index - Get current FPA index
Syntax: BYTE MSPPRG_API F_Get_FPA_index ( void );
Return value:current FPA index
F_Disable_FPA_index
F_Disable_FPA_index - Disable desired FPA index (desired DLL instance)
VALID FPA index - ( 1 to 8 )
Function allows to disable communication with selected FPA adapter. From application point of
view, all responses will be the same as from the not active FPA. Communication with target
devices connected to selected FPA will be stopped. When the F_Set_FPA_index(0) will be used,
then selected FPA will be ignored. Result will not be presented in the Status results (Status and
F_LastStatus(..)).
Syntax:
void MSPPRG_API F_Disable_FPA_index ( BYTE fpa );
Parameters: fpa -> 1 to MAX_FPA_INDEX where MAX_FPA_INDEX = 8
F_Enable_FPA_index
F_Enable_FPA_index - Enable desired FPA index (desired DLL instance)
VALID FPA index - ( 1 to 8 )
Remote Control Programming User’s Guide PM024A02 Rev.2 36
Function allows to enable communication with selected FPA adapter if the mentioned FPA has
been disabled using the function F_Disable_FPA_index(...). By default, all FPAs are enabled.
Syntax: void MSPPRG_API F_Enable_FPA_index ( BYTE fpa );
Parameters:
fpa -> 1 to MAX_FPA_INDEX where MAX_FPA_INDEX = 8
F_LastStatus
F_LastStatus - Get current FPA index
VALID FPA index - ( 1 to 8 )
Syntax: INT_X MSPPRG_API F_LastStatus ( BYTE fpa );
Parameters:fpa - FPA index of the desired status
fpa index -> 1..8
Return value:Last status from the desired FPAs
All F_xxx functions returns the same parameters (status) as the original API_DLL is returning.
When function is transferred to all API-DLLs (when the fpa=0) then returned parameter (status)
is the same as the returned value from the API-DLLs when the ALL returned values ARE THE
SAME. If not, then returned value is
FPA_UNMATCHED_RESULTS
(value of the FPA_UNMATCHED_RESULTS is minus 1).
To get the returned values from each FPAs, use the
For( n=1; n<=8; n++) status[n] = F_LastStatus( n);
where n -> desired FPA index
and get the last status data from FPA-1, 2, .. up to .8
F_Multi_DLLTypeVer
Remote Control Programming User’s Guide PM024A02 Rev.2 37
F_Multi_DLLTypeVer function returns integer number with DLL ID and software revision
version.
Syntax: MSPPRG_API INT_X F_Multi_DLLTypeVer( void );
Return value:VALUE = (DLL ID) | ( 0x0FFF & Version)
DLL ID = 0x1000 - Single DLL for the Parallel Port MSP430-FPA
DLL ID = 0x2000 - Single DLL for the USB MSP430-FPA (FlashPro430)
DLL ID = 0x3000 - Single API-DLL for the GangPro430
DLL ID = 0x4000 - Single API-DLL for the FlashPro-CC
DLL ID = 0x5000 - Single API-DLL for the GangPro-CC
DLL ID = 0x6000 - Multi-FPA API-DLL for the FlashPro430
DLL ID = 0x7000 - Multi-FPA API-DLL for the GangPro430
DLL ID = 0x8000 - Multi-FPA API-DLL for the FlashPro-CC
DLL ID = 0x9000 - Multi-FPA API-DLL for the GangPro-CC
Version = (0x0FFF & VALUE)
F_Get_FPA_SN
F_Get_FPA_SN - Get FPAs Serial number assigned to selected FPA-index (selected
DLL instance number).
Syntax: INT_X MSPPRG_API F_Get_FPA_SN ( BYTE fpa );
Parameters:fpa - FPA index of the desired status
fpa index -> 1..8
Return value:Serial number of the selected FPA
or FPA_INVALID_NO - if used fpa index is not activated or out of range.
note: FPA_INVALID_NO -> -2 (minus 2)
Remote Control Programming User’s Guide PM024A02 Rev.2 38
4.2 Generic instructions
Generic instructions are related to initialization programmer process, configuration setup and
preparation data, turning ON and OFF target’s DC and RESET target device. Any communication
with the target device is provided when any of the generic instruction is executed. Generic
instructions should be called before encapsulated and sequential instruction.
F_Check_FPA_access
F_Check_FPA_access - Check available Flash Programming Adapter connected to specified
USB drivers (USB driver index from 1 to 16)
Important: It is not recommended to use this function. Function used only for compatible with
the old software. Use the F_OpenInstancesAndFPAs instead.
Do not use the F_OpenInstances or F_Check_FPA_access after using the
F_OpenInstancesAndFPAs. The F_OpenInstancesAndFPAs is assigning the FPAs to
USB ports and it is not recommended to reassign once again the USB port using the
F_Check_FPA_access function. To check the communication activity with FPA use the
F_Get_FPA_SN function that allows to check te communication with the FPA adapter
without modifying the USB ports assignment.
VALID FPA index (DLL instance number) - ( 1 to 8 )
F_Check_FPA_access should be called as a first function when the *.dll is activated. Function
returns serial number of the detected flash programming adapter, or zero, if programming adapter
has not been detected with selected USB driver. Up to 16 USB drivers can be scanned.
To make a Multi-FPA software back compatible, the F_Check_FPA_access procedure is calling
the function F_OpenInstances if none of the instances has not been activated before. That
allows to use old application software without calling the new type of Multi-FPA functions.
Syntax:
MSPPRG_API INT_X F_Check_FPA_access ( INT_X USB_index );
Parameters:
Index: USB driver index from 1 to MAX_USB_DEV_NUMBER
where MAX_USB_DEV_NUMBER = 16
Remote Control Programming User’s Guide PM024A02 Rev.2 39
Return value:
0 - FALSE
>0 - Detected FPA’s Serial Number
Example:
long SN[MAX_USB_DEV_NUMBER+1];
F_OpenInstances( 1 ); // DLL initialization - one instance
F_Set_FPA_index( 1 ); // select access to the first instance
n = 0; //no of detected FPAs
for( k=1; k<=MAX_USB_DEV_NUMBER ; k++ )
{
SN[k] = F_Check_FPA_access(k);
if ( SN[k] > 0 ) n++;
}
F_CloseInstances(); // DLL initialization - one instance
F_OpenInstances( n ); // Open ‘n’ instances - one per FPA
// Find desired FPAs SN and assign the FPAs serial number every time to the
same // FPA-index.
// For example if the
// SN[1]= 20060123
// SN[2]= 20060147
// SN[3]= 0 - adapter not present
// SN[4]= 20060135
// and desired assignment
// FPA-1 20060123
// FPA-2 20060135
// FPA-3 20060147
// then following sequence instructions can be used
F_Set_FPA_index( 1 ); // select access to the first instance
F_Check_FPA_access( 1 ); //assign FPA SN[1] = 20060123 to FPA-1
F_Set_FPA_index( 2 ); // select access to the second instance
F_Check_FPA_access( 4 ); //assign FPA SN[4] = 20060135 to FPA-2
F_Set_FPA_index( 3 ); // select access to the third instance
F_Check_FPA_access( 2 ); //assign FPA SN[2] = 20060147 to FPA-3
F_Set_FPA_index( ALL_ACTIVE_FPA ); // select all active instances
F_Initialization() // All FPAs initialization
............................
Remote Control Programming User’s Guide PM024A02 Rev.2 40
F_DLLTypeVer
F_DLLTypeVer - Get information about DLL software type and software revision.
VALID FPA index - ( 1 to 8 )
F_DLLTypeVer function returns integer number with DLL ID and software revision version
and copying text message to report message buffer about DLL ID and software revision. Text
content can downloaded using one of the following functions
F_GetReportMessageChar( index )
or F_ReportMessage( text )
Syntax: MSPPRG_API INT_X F_DLLTypeVer( void );
Return value:VALUE = (DLL ID) | ( 0x0FFF & Version)
DLL ID = 0x1000 - Single DLL for the Parallel Port MSP430-FPA
DLL ID = 0x2000 - Single DLL for the USB MSP430-FPA (FlashPro430)
DLL ID = 0x3000 - Single API-DLL for the GangPro430
DLL ID = 0x4000 - Single API-DLL for the FlashPro-CC
DLL ID = 0x5000 - Single API-DLL for the GangPro-CC
DLL ID = 0x6000 - Multi-FPA API-DLL for the FlashPro430
DLL ID = 0x7000 - Multi-FPA API-DLL for the GangPro430
DLL ID = 0x8000 - Multi-FPA API-DLL for the FlashPro-CC
DLL ID = 0x9000 - Multi-FPA API-DLL for the GangPro-CC
Version = (0x0FFF & VALUE)
Example:INT_X id;
.......................
.......................
id = F_DLLTypeVer();
Disp_report_message();
//see F_ReportMessage or F_GetReportMessage for details
.......................
F_Initialization
F_Initialization - Programmer initialization.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Remote Control Programming User’s Guide PM024A02 Rev.2 41
F_Initialization function should be called after the communication with the FPA adapter is
established. To make a Multi-FPA software back compatible, the F_Initialization procedure is
calling the function F_OpenInstances if none of the instances has not been activated before.
That allows to use old application software without calling the new type of Multi-FPA functions.
In this case the F_Check_FPA_access function can be used to activate communication between
PC and Programming Adapter. When the F_Check_FPA_access is not called then by default
the USB driver number “1" is selected.
When the F_Initialization is called then:
- all internal data is cleared or set to the default value,
- initial configuration is downloaded from the config.ini file,
- USB driver is initialized if has not been initialized before (for the USB version
programmer) or Parallel Port becomes open (for the parallel port version programmer).
Programming adapter must be connected to the USB or Parallel Port to establish communication
between PC and programming adapter. Otherwise the F_Initialization will return FALSE result.
Syntax: MSPPRG_API INT_X F_Initialization( void );
Return value:0 - FALSE
1 - TRUE
4 - Programming adapter not detected.
Example: ............................
F_API_DLL_Directory( “.....” ) // optional - see F_API_DLL_Directory()
If( F_Initialization() != TRUE ) //required API-Dll - initialization
{
// Initialization error
}
............................
F_Close_All
F_Close_All - Close communication with the programming adapter and release
PC memory.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Remote Control Programming User’s Guide PM024A02 Rev.2 42
F_Close_All function should be called as the last one before *.dll is closed. When the
F_Close_All is called then communication port becomes closed and all internal dynamic data
will be released from the memory. To activate communication with the programmer when the
function F_Close_All has been used the F_Initialization function must be called first.
Syntax: MSPPRG_API INT_X F_Close_All( void );
Return value:0 - FALSE
1 - TRUE
Example:F_Initialization(); //required API-Dll - initialization
............................
............................
F_Close_All;
............................
F_GetSetup
F_GetSetup - Get configuration setup from the programmer.
VALID FPA index - ( 1 to 8 )
See F_ConfigSetup description for more details.
Syntax:MSPPRG_API INT_X F_GetSetup( CFG_BLOCK *config );
Return value:0 - FALSE
1 - TRUE
F_ConfigSetup
F_ConfigSetup - Setup programmer’s configuration.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Remote Control Programming User’s Guide PM024A02 Rev.2 43
The F_ConfigSetup can modify configuration of the programmer. When the
F_ConfigSetup is called, then the structure data block is transferred from the software application
to the programmer software. Current programmer setup can be read using function setup
F_GetSetup. When data block is taken from the programmer, then part or all of the configuration
data can be modified and returned to programmer using F_ConfigSetup function. Configuration
data structure and available data for all listed items in this structure are defined below. Listed
name and indexes in the [] brackets are related to the F_SetConfig and F_GetConfig
instructions
See F_Set_Config(....) for detailed description of the all configuration data contents.
typedef struct
{
INT_X DeviceIndex;
INT_X PowerTargetEn;
INT_X CommSpeedIndex;
INT_X ResetTimeIndex;
INT_X CustomResetPulseTime;
INT_X CustomResetIdleTime;
INT_X RstVccOffTime;
INT_X ApplicationStartIndex;
INT_X ApplicationRunTime;
INT_X FlashEraseModeIndex;
INT_X FlashReadModeIndex;
INT_X FlashLockBits;
INT_X UnlockDebugBit;
INT_X LockBitsEn;
INT_X VerifyModeIndex;
INT_X IEEEAddeModeIndex;
INT_X ManIEEEAddeModeIndex;
INT_X BeepOKEn;
INT_X VccIndex;
INT_X TargetEnMask; //used in the GangPro-CC only
INT_X RetainDataEn;
INT_X RetainDataStartAddr;
INT_X RetainDataStopAddr;
INT_X EraseDefBlock1En;
INT_X EraseDefBlock1StartAddr;
INT_X EraseDefBlock1StopAddr;
INT_X EraseDefBlock2En;
INT_X EraseDefBlock2StartAddr;
INT_X EraseDefBlock2StopAddr;
INT_X EraseDefBlock3En;
INT_X EraseDefBlock3StartAddr;
INT_X EraseDefBlock3StopAddr;
INT_X EraseDefBlock4En;
INT_X EraseDefBlock4StartAddr;
Remote Control Programming User’s Guide PM024A02 Rev.2 44
INT_X EraseDefBlock4StopAddr;
INT_X ReadDefBlock1En;
INT_X ReadDefBlock1StartAddr;
INT_X ReadDefBlock1StopAddr;
INT_X ReadDefBlock2En;
INT_X ReadDefBlock2StartAddr;
INT_X ReadDefBlock2StopAddr;
INT_X ReadDefBlock3En;
INT_X ReadDefBlock3StartAddr;
INT_X ReadDefBlock3StopAddr;
INT_X ReadDefBlock4En;
INT_X ReadDefBlock4StartAddr;
INT_X ReadDefBlock4StopAddr;
INT_X Spare1;
INT_X Spare2;
INT_X Spare3;
INT_X Spare4;
INT_X Spare5;
INT_X Spare6;
INT_X Spare7;
INT_X Spare8;
INT_X Spare9;
INT_X Spare10;
INT_X Spare11;
INT_X Spare12;
INT_X Spare13;
INT_X Spare14;
INT_X Spare15;
INT_X Spare16;
} CFG_BLOCK;
Syntax: MSPPRG_API INT_X F_ConfigSetup( CFG_BLOCK config );
Return value:
0 - FALSE
1 - TRUE
Example:
Example below shows the method of modification of the programmers configuration setup. First
the current setup from the programmer is uploaded to the application, after that some of the
parameters have been modified and at the end the modified setup is returned back to the
programmer.
CFG_BLOCK config; //programmer’s configuration data
.......................
Remote Control Programming User’s Guide PM024A02 Rev.2 45
F_GetSetup( &config );
//API-DLL - get configuration from the programmer
config.CommSpeedIndex = SPEED_3MB_INDEX;
//select JTAG interface
config.FlashEraseModeIndex = ERASE_ALL_MEM_INDEX;
//select all memory erase option
F_ConfigSetup( config );
//API-DLL - setup configuration in the programmer
F_SetConfig
F_SetConfig - Setup one item of the programmer’s configuration.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Similar to the F_ConfigSetup, but only one selected item from the CFG_BLOCK structure
is modified.
Syntax: MSPPRG_API INT_X F_SetConfig( INT_X index, INT_X data );
Return value:0 - FALSE
1 - TRUE
Example: .......................
.......................
F_SetConfig( CFG_MICROCONTROLLER, config.uProcIndex );
or
F_SetConfig( CFG_MICROCONTROLLER, 7 );
Index’s list
CFG_MICROCONTROLLER 1
CFG_POWERTARGETEN 2
CFG_COMM_SPEED_INDEX 3
CFG_RESET_TIME_INDEX 4
CFG_RESET_PULSE_TIME 5
CFG_RESET_IDLE_TIME 6
CFG_RSTVCC_OFF_TIME 7
CFG_APPLSTARTEN 8
CFG_APPL_RUN_TIME 9
CFG_FLASHERASEMODE 10
CFG_FLASHREADMODE 11
CFG_FLASH_LOCKBIT 12
CFG_UNLOCK_DEBUG_BIT 13
CFG_LOCK_BITS_EN 14
Remote Control Programming User’s Guide PM024A02 Rev.2 46
CFG_VERIFYMODE 15
CFG_IEEE_ADDR_MODE 16
CFG_MAN_IEEE_ADDR_MODE 17
CFG_BEEP_OK_EN 18
CFG_VCCINDEX 19
CFG_TARGET_EN_MASK 20
CFG_RETAIN_DEF_DATA_EN 21
CFG_RETAIN_START_ADDR 22
CFG_RETAIN_STOP_ADDR 23
CFG_ERASE_DEFBLOCK1_EN 24
CFG_ERASE1_START_ADDR 25
CFG_ERASE1_STOP_ADDR 26
CFG_ERASE_DEFBLOCK2_EN 27
CFG_ERASE2_START_ADDR 28
CFG_ERASE2_STOP_ADDR 29
CFG_ERASE_DEFBLOCK3_EN 30
CFG_ERASE3_START_ADDR 31
CFG_ERASE3_STOP_ADDR 32
CFG_ERASE_DEFBLOCK4_EN 33
CFG_ERASE4_START_ADDR 34
CFG_ERASE4_STOP_ADDR 35
CFG_READ_DEFBLOCK1_EN 36
CFG_READ1_START_ADDR 37
CFG_READ1_STOP_ADDR 38
CFG_READ_DEFBLOCK2_EN 39
CFG_READ2_START_ADDR 40
CFG_READ2_STOP_ADDR 41
CFG_READ_DEFBLOCK3_EN 42
CFG_READ3_START_ADDR 43
CFG_READ3_STOP_ADDR 44
CFG_READ_DEFBLOCK4_EN 45
CFG_READ4_START_ADDR 46
CFG_READ4_STOP_ADDR 47
CFG_IEEE_ADDR_LOCATION 48
CFG_IEEE_ADDR_LOC_MODE 49
CFG_IEEE_ADDR_LSB_FIRST 50
// -------------- CONFIG_BLOCK - definitions ----
// CFG_MICROCONTROLLER 1
CC_ANY 0
// 1 - CC1110F8
// 2 - CC1110F16
// 3 - CC1110F32
// 4 - CC2430F32
// 5 - CC2430F64
Remote Control Programming User’s Guide PM024A02 Rev.2 47
// 6 - CC2430F128
// 7 - CC2431F32
// 8 - CC2431F64
// 9 - CC2431F128
// 10 - CC2510F8
// 11 - CC2510F16
// 12 - CC2510F32
// 13 - CC2511F8
// 14 - CC2511F16
// 15 - CC2511F32
// CFG_POWERTARGETEN 2
// 0 -> PowerTarget Disable
// 1 -> PowerTarget Enable
// CFG_COMM_SPEED_INDEX 3
SPEED_3MB_INDEX 1
SPEED_1MB_INDEX 2
// CFG_RESET_TIME_INDEX 4
RESET_10MS_INDEX 0
RESET_100MS_INDEX 1
RESET_200MS_INDEX 2
RESET_500MS_INDEX 3
RESET_CUSTOM_INDEX 4
RESET_TOGGLE_VCC_INDEX 5
// CFG_RESET_PULSE_TIME 5
time in ms
// CFG_RESET_IDLE_TIME 6
time in ms
// CFG_RSTVCC_OFF_TIME 7
time in ms
// CFG_APPLSTARTEN 8
APPLICATION_KEEP_RESET 0
APPLICATION_TOGGLE_RESET 1
APPLICATION_TOGGLE_VCC 2
APPLICATION_SOFT_RESET 3
// CFG_APPL_RUN_TIME 9
time in ms
// CFG_FLASHERASEMODE 10
ERASE_NONE_MEM_INDEX 0
ERASE_ALL_MEM_INDEX 1
Remote Control Programming User’s Guide PM024A02 Rev.2 48
ERASE_INFILE_MEM_INDEX 2
ERASE_DEF_CM_INDEX 3
// CFG_FLASHREADMODE 11
READ_ALL_MEM_INDEX 0
READ_PRGMEM_ONLY_INDEX 1
READ_INFOMEM_ONLY_INDEX 2
READ_DEF_MEM_INDEX 3
// CFG_FLASH_LOCKBIT 12
// CFG_UNLOCK_DEBUG_BIT 13
// CFG_LOCK_BITS_EN 14
// CFG_VERIFYMODE 15
VERIFY_NONE_INDEX 0
VERIFY_STD_INDEX 1
VERIFY_FAST_INDEX 2
// CFG_IEEE_ADDR_MODE 16
AP_IEEE_ADDR_DISABLE 0
AP_WR_NEW_IEEE_ADDR 1
AP_RETAIN_CODE_WR_IEEE 2
AP_IEEE_ADDR_BLANK 3
AP_RETAIN_IEEE_ADDR 4
AP_ASSIGN_WR_IEEE_ADDR 5
AP_WR_IEEE_ADDR_FROM_FILE 6
// CFG_MAN_IEEE_ADDR_MODE 17
AP_IEEE_ADDR_DISABLE 0
AP_WR_NEW_IEEE_ADDR 1
AP_RETAIN_CODE_WR_IEEE 2
// CFG_BEEP_OK_EN 18
// CFG_VCCINDEX 19
VCC_2V2_INDEX 0
VCC_2V4_INDEX 1
VCC_2V6_INDEX 2
VCC_2V8_INDEX 3
VCC_3V0_INDEX 4
VCC_3V2_INDEX 5
VCC_3V4_INDEX 6
VCC_3V6_INDEX 7
// CFG_RETAIN_DEF_DATA_EN 21
0-disable 1-enable
// CFG_RETAIN_START_ADDR 22
0x00000 to 0x1FFFF
Remote Control Programming User’s Guide PM024A02 Rev.2 49
// CFG_RETAIN_STOP_ADDR 23
0x00000 to 0x1FFFF
// CFG_ERASE_DEFBLOCK1_EN 24
0-disable 1-enable
// CFG_ERASE1_START_ADDR 25
0x00000 to 0x1FFFF
// CFG_ERASE1_STOP_ADDR 26
0x00000 to 0x1FFFF
// CFG_ERASE_DEFBLOCK2_EN 27
0-disable 1-enable
// CFG_ERASE2_START_ADDR 28
0x00000 to 0x1FFFF
// CFG_ERASE2_STOP_ADDR 29
0x00000 to 0x1FFFF
// CFG_ERASE_DEFBLOCK3_EN 30
0-disable 1-enable
// CFG_ERASE3_START_ADDR 31
0x00000 to 0x1FFFF
// CFG_ERASE3_STOP_ADDR 32
0x00000 to 0x1FFFF
// CFG_ERASE_DEFBLOCK4_EN 33
0-disable 1-enable
// CFG_ERASE4_START_ADDR 34
0x00000 to 0x1FFFF
// CFG_ERASE4_STOP_ADDR 35
0x00000 to 0x1FFFF
// CFG_READ_DEFBLOCK1_EN 36
0-disable 1-enable
// CFG_READ1_START_ADDR 37
0x00000 to 0x1FFFF
// CFG_READ1_STOP_ADDR 38
0x00000 to 0x1FFFF
// CFG_READ_DEFBLOCK2_EN 39
Remote Control Programming User’s Guide PM024A02 Rev.2 50
0-disable 1-enable
// CFG_READ2_START_ADDR 40
0x00000 to 0x1FFFF
// CFG_READ2_STOP_ADDR 41
0x00000 to 0x1FFFF
// CFG_READ_DEFBLOCK3_EN 42
0-disable 1-enable
// CFG_READ3_START_ADDR 43
0x00000 to 0x1FFFF
// CFG_READ3_STOP_ADDR 44
0x00000 to 0x1FFFF
// CFG_READ_DEFBLOCK4_EN 45
0-disable 1-enable
// CFG_READ4_START_ADDR 46
0x00000 to 0x1FFFF
// CFG_READ4_STOP_ADDR 47
0x00000 to 0x1FFFF
// CFG_IEEE_ADDR_LOCATION 48
0x00000 to 0x1FFF8
// CFG_IEEE_ADDR_LOC_MODE 49
0-default 1-defined
// CFG_IEEE_ADDR_LSB_FIRST 50
0-disable 1-enable
F_GetConfig
F_GetConfig - Get one item of the programmer’s configuration.
VALID FPA index - ( 1 to 8 )
Similar to the F_GetSetup, but only one item from the CFG_BLOCK structure is read.
Syntax: MSPPRG_API INT_X F_GetConfig( INT_X index );
Remote Control Programming User’s Guide PM024A02 Rev.2 51
Index’s list - see F_SetConfig
Return value:
Requested setup parameter;
Example: .......................
.......................
F_GetSetup( config );
DeviceIndex = config.DeviceIndex;
or directly
DeviceIndex = F_GetConfig( CFG_MICROCONTROLLER );
.......................
F_DispSetup
F_DispSetup - Copy programmer’s configuration to report message buffer in text form.
VALID FPA index - ( 1 to 8 )
Syntax: MSPPRG_API INT_X F_DispSetup( void );
Return value:
1 - TRUE;
Example: .......................
.......................
F_DispSetup();
Disp_report_message();
//see F_ReportMessage or F_GetReportMessage for details
.......................
F_ReportMessage, F_Report_Message
F_ReportMessage - Get the last report message from the programmer.
or F_Report_Message
Remote Control Programming User’s Guide PM024A02 Rev.2 52
VALID FPA index - ( 1 to 8 )
When any of the DLL functions is activated, a message is created and displayed on the
dynamically created programmer’s dialogue box. At the end of execution the dialogue box is
closed and function returns back to the application program. Reported message is closed as well.
The last report message can be read by application program using F_ReportMessage function.
When F_ReportMessage is called, then report message up to
REPORT_MESSAGE_MAX_SIZE 2000
characters is imported from the programmer software to the application software. Make sure to
declare characters string length no less then REPORT_MESSAGE_MAX_SIZE characters. When the
F_ReportMessage is called then at the end the internal report message buffer in the programmer
software is cleared. When F_ReportMessage is not called after every communication with the
target device, then the report message will collect all reported information up to
REPORT_MESSAGE_MAX_SIZE last characters.
Syntax: MSPPRG_API void F_ReportMessage( char * text );
MSPPRG_API char* F_Report_Message( void );
note: F_Report_Message is available only with the Multi-FPA API-DLL.
Return value:none
Example:#include “FlashProCC-Dll.h”;
char text[REPORT_MESSAGE_MAX_SIZE];
.......................
.......................
F_ReportMessage( text );
.......................
Example below shows how to take a message and display it in the scrolling box. The Edit box
with the ID e.g. IDC_REPORT must be created first.
............................................
#include “FlashProCC-Dll.h”;
Cstring Message = "";
............................................
Remote Control Programming User’s Guide PM024A02 Rev.2 53
void CMspPrgDemoDlg::Disp_report_message()
{
char text[REPORT_MESSAGE_MAX_SIZE];
F_ReportMessage( text ); //API-Dll - get last report message
Message = text;
SetDlgItemText(IDC_REPORT, Message.GetBuffer(Message.GetLength()));
CEdit* pEdit = (CEdit*) GetDlgItem(IDC_REPORT);
pEdit->LineScroll(pEdit->GetLineCount(), 0);
UpdateWindow();
}
F_GetReportMessageChar
F_GetReportMessageChar - Get one character of the the last report message from the
programmer.
VALID FPA index - ( 1 to 8 )
See comment for the F_ReportMessage function.
F_GetReportMessageChar allows to get character by character from the report message buffer.
This function is useful in the Visual Basic application, where all message can not be transfered
via pointer like it is possible in the C++ application.
Syntax: MSPPRG_API char F_GetReportMessageChar( INT_X index );
Return value:Requested character from the Report Message buffer. 1 - TRUE
Example:#include “FlashProCC-Dll.h”;
char text[REPORT_MESSAGE_MAX_SIZE];
INT_X k;
.......................
.......................
for( k = 0; k< REPORT_MESSAGE_MAX_SIZE; k++ )
text[k] = F_GetReportMessageChar( k );
.......................
Example below shows how to take a message and display it in the scrolling box. The Edit box
with the ID e.g. IDC_REPORT must be created first.
Remote Control Programming User’s Guide PM024A02 Rev.2 54
............................................
#include “FlashProCC-Dll.h”;
Cstring Message = "";
............................................
void CMspPrgDemoDlg::Disp_report_message()
{
char text[REPORT_MESSAGE_MAX_SIZE];
INT_X k;
for( k = 0; k< REPORT_MESSAGE_MAX_SIZE; k++ )
text[k] = F_GetReportMessageChar( k );
Message = text;
SetDlgItemText(IDC_REPORT, Message.GetBuffer(Message.GetLength()));
CEdit* pEdit = (CEdit*) GetDlgItem(IDC_REPORT);
pEdit->LineScroll(pEdit->GetLineCount(), 0);
UpdateWindow();
}
F_ReadCodeFile, F_Read_CodeFile
F_ReadCodeFile - Read code data from the file and download it to internal buffer.
or F_Read_CodeFile
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Function F_ReadCodeFile downloads code from the file to internal memory buffer. Code file
format and file name and location path of the desired file must be specified. Three file formats
are supported - Texas Instruments text format, Motorola *.s19 format and Intel *.hex format.
When file is downloaded then contents of this file is analysed. Only code memory location valid
for the Ccxx device family will be downloaded to the internal memory buffer. Any code data
located outside memory space of the Ccxx device will be ignored and warning message will be
created.
Syntax: MSPPRG_API INT_X F_ReadCodeFile( int file_format, char * FileName );
MSPPRG_API INT_X F_Read_CodeFile( int file_format, CString FileName );
file_format:
FILE_TI_FORMAT (1) for TI (*.txt) format
FILE_MOTOROLA_FORMAT (2) for Motorola (*.s19, *.s28 or *.s37)
FILE_INTEL_FORMAT (3) for Intel (*.hex)
FileName: file name including path, file name and extention
Remote Control Programming User’s Guide PM024A02 Rev.2 55
Return value:(0xFFFe & info) | state
where state is defined as follows:
0 - FALSE
1 - TRUE
info is defined as follows:
warning -> CODE_IN_ROM
CODE_IN_RAM
CODE_OUT_OF_FLASH
CODE_OVERWRITTEN
error -> INVALID_CODE_FILE
OPEN_FILE_OR_READ_ERR
Example:
int st;
....................
st = F_ReadCodeFile( FILE_TI_FORMAT, “c:\test\demofile.txt” );
if(( st & 1 ) == TRUE )
{
....................
}
else
{
if ( st & CODE_IN_ROM ) {......}
if ( st & CODE_OUT_OF_FLASH ) {......}
if ( st & INVALID_CODE_FILE ) {......}
if ( st & OPEN_FILE_OR_READ_ERR ) {......}
....................
....................
}
F_Get_CodeCS
F_Get_CodeCS - Read code from internal buffer and calculate the check sum.
VALID FPA index - ( 1 to 8 ).
Syntax: MSPPRG_API INT_X F_Get_CodeCS( int index );
index - index of the desired code
Index = 1 - Calculate check sum of the code from internal code buffer.
Remote Control Programming User’s Guide PM024A02 Rev.2 56
Other Index values - reserved for the future option.
Check Sum is calculated as an arithmetic sum of the 16-bits unsigned words form the valid code
bytes. If the only one byte is present in the calculated word, then other byte is taken as a 0xFF.
Check Sum result is 32 bits.
For example from the following code
address data
0x0300 0xF2 0x12 0x23 0x34 0x78
Check Sum calculation
Word 1 0x12F2
Word 2 0x3423
Word 2 0xFF78
---------------------------------
CS = 0x0001468D
Return value:
Calculated check sum.
F_ConfigFileLoad, F_Config_FileLoad
F_ConfigFileLoad - Modify programmer’s configuration setup according to data taken
or F_Config_FileLoad from the specified configuration file.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
The F_ConfigFileLoad function can download the programmer setup from the external setup
file. Setup file can be created using standard FlashPro-CC (GUI) Flash Programmer software.
When the setup from the file is downloaded, then old configuration setup is overwritten. The
new setup can be modified using F_GetSetup and F_ConfigSetup functions.
Location path and file name of the config file must be specified.
Syntax:
MSPPRG_API INT_X F_ConfigFileLoad( char * filename );
MSPPRG_API INT_X F_ConfigFileLoad( Cstring filename );
Remote Control Programming User’s Guide PM024A02 Rev.2 57
filename - configuration file name including path, file name and extention
Return value:0 - FALSE
1 - TRUE
(0xFFFe & info) | state
where state is defined as follows:
0 - FALSE
1 - TRUE
info is defined as follows:
error -> OPEN_FILE_OR_READ_ERR
Configuration file is a standard text file with the parameters name and value.
Example:st = F_ConfigFileLoad( “c:\test\configfile.cfg” );
if(( st & 1 ) == TRUE )
{
....................
}
else
{
Info = st & 0xFFFE;
....................
}
F_Clr_Code_Buffer
F_Clr_Code_Buffer - Clear content of the code buffer.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax: MSPPRG_API INT_X F_Clr_Code_Buffer( void );
Return value:0 - FALSE
1 - TRUE
Example:.....................
F_Clr_Code_Buffer();
.....................
Remote Control Programming User’s Guide PM024A02 Rev.2 58
F_Put_Byte_to_Code_Buffer
F_Put_Byte_to_Code_Buffer - Write code data to code buffer.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Instruction allows to write contents of the code to code buffer instead using the F_ReadCodeFile
instruction. Contents of the downloaded code data can be modified or filled with the new data, if
code buffer has been cleared first (using F_Clr_Code_Buffer function).
Syntax: MSPPRG_API INT_X F_Put_Byte_to_Code_Buffer( INT_X address, BYTE data );
Parameters value:code address - 0x0 to 0x1FFFF
data - 0x00 to 0xFF
Return value:0 - FALSE
1 - TRUE
Example:
BYTE code[MAX_FLASH_SIZE];
.....................
F_Clr_Code_Buffer();
for( address = 0x0; address < MAX_FLASH_SIZE; address ++ )
{
F_Put_Byte_to_Code_Buffer( address, code[address]);
}
.....................
F_Get_Byte_from_Code_Buffer
F_Get_Byte_from_Code_Buffer - Read code data from code buffer.
VALID FPA index - ( 1 to 8 )
Instruction allows to read or verify contents of the code from code buffer
Syntax: MSPPRG_API INT_X F_Get_Byte_from_Code_Buffer( INT_X address );
Remote Control Programming User’s Guide PM024A02 Rev.2 59
Parameters value:code address - 0x0 to MAX_FLASH_SIZE-1 (0x1FFFF)
Return value:0x00 to 0xFF - valid code data
-1 (0xFFFF) - code data not initialized on particular address
F_Put_IEEEAddr64_to_Buffer
F_Put_IEEEAddr64_to_Buffer - Write IEEE address to buffer.
VALID FPA index - ( 1 to 8 ).
Instruction allows to write one unique IEEE address to buffer. Contents of the IEEE address from
the buffers will be saved to target device when the F_Autoprogram(0) is executed and when in
the configuration setup this option is enabled.
Syntax:MSPPRG_API void F_Put_IEEEAddr64_to_Buffer(ULONG64 data );
Parameters value: data - 64 bits unsigned long integer number 0 to 0xFFFFFFFFFFFFFFFF
F_Put_IEEEAddr_Byte_to_Buffer
F_Put_IEEEAddr_Byte_to_Buffer - Write one byte of the IEEE address to buffer.
VALID FPA index - ( 1 to 8 ).
Instruction allows to write one byte of the unique IEEE address to buffer. Contents of the IEEE
address from the buffers will be saved to target device when the F_Autoprogram(0) is executed
and when in the configuration setup this option is enabled.
Syntax:MSPPRG_API INT_X F_Put_IEEEAddr64_to_Buffer( BYTE no, BYTE data );
Remote Control Programming User’s Guide PM024A02 Rev.2 60
Instruction is functionally the same as the F_Put_IEEEAddr64_to_Buffer, but allows to transfer
byte by byte of the 64 IEEE address to buffer. Function is used when the Visual Basic 6 is used,
that not support the Int 64 bits data.
no -> 0 to 7.
Index no MUST started from 0 and finished on 7 to transfer whole IEEEAddr. When no = 0, the
lowest byte of the IEEEAddr must be transferred. When no = 7, the highest byte of the
IEEEAddr is transferred.
Parameters value: data - one byte from 64 bits IEEEAddr data
F_Get_IEEEAddr64_from_Buffer
F_Get_IEEEEAddr64_from_Buffer - Read IEEE address contents from the buffer
VALID FPA index - ( 1 to 8 )
Syntax: MSPPRG_API ULONG64 F_Get_IEEEAddr64_from_Buffer(BYTE no,);
Parameters value:none
Return value:64 bits unsigned long integer number 0 to 0xFFFFFFFFFFFFFFFF
F_Get_IEEEAddr_Byte_from_Buffer
F_Get_IEEEEAddr_Byte_from_Buffer - Read one byte of the IEEE address contents from
the buffer
VALID FPA index - ( 1 to 8 )
Syntax: MSPPRG_API BYTE F_Get_IEEEAddr_Byte_from_Buffer();
Instruction is functionally the same as the F_Get_IEEEAddr64_from_Buffer, but allows to
transfer byte by byte of the 64 IEEE address from buffer. Function is used when the Visual Basic
6 is used, that not support the Int 64 bits data.
no -> 0 to 7.
Remote Control Programming User’s Guide PM024A02 Rev.2 61
When no = 0, the lowest byte of the IEEEAddr is transferred. When no = 7, the highest byte of
the IEEEAddr is transferred.
Parameters value:no - byte number of the 64 bits IEEEAddr data
Return value:one byte of the IEEEAddr
F_Power_Target
F_Power_Target - Turn ON or OFF power from programming adapter to target
device.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Function F_Power_Target switches ON or OFF power from the programming adapter to the target
device.
Note: PowerTargetEn flag must be set to TRUE (1) in the configuration setup to switch the
power from the programming adapter ON.
Syntax: MSPPRG_API INT_X F_Power_Target( INT_X OnOff );
Return value:0 - FALSE
1 - TRUE
Example:.....................
F_Power_Target( 1 ); // Turn Power ON
.....................
F_Power_Target( 0 ); // Turn Power OFF
.....................
F_Reset_Target
F_Reset_Target - Generate short RESET pulse on the target’s device RESET line.
Remote Control Programming User’s Guide PM024A02 Rev.2 62
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Function F_Reset_Target resets target device and target device’s application program can start.
Length of the RESET pulse time is specified by ResetTimeIndex in configuration setup. See
F_ConfigSetup description for details.
Syntax: MSPPRG_API INT_X F_Reset_Target( void );
Return value:0 - FALSE
1 - TRUE
Example:.....................
F_Reset_Target( void );
.....................
F_Get_Targets_Vcc
F_Get_Targets_Vcc - Get Vcc in [mV] supplied target device.
VALID FPA index - ( 1 to 8 )
Syntax: MSPPRG_API INT_X F_Get_Targets_Vcc( void );
Return value:INT_X - Vcc in milivolts e.g 3000 -> 3.0 V
or (-1) if USB-FPA is not active.
Remote Control Programming User’s Guide PM024A02 Rev.2 63
4.3 Encapsulated instructions
Encapsulated functions are powerful and easy to use. When called then all device actions from the
beginning to the end are done automatically and final result is reported as TRUE or FALSE.
Required configuration should be set first using F_GetSetup and F_ConfigSetup functions.
Encapsulated function has following sequence:
- Power from the programming adapter becomes ON if PowerTargetEn in configuration
setup is enabled.
- Vcc is verified to be higher then 2.0V.
- Communication between programming adapter and target device is initialized.
- Selected encapsulated instruction is executed ( Autoprogram, Memory Erase etc. ).
- Communication between target device and programming adapter is terminated.
- Power from the programming adapter becomes OFF (if selected).
- Target device is released from the programming adapter.
F_AutoProgram
F_AutoProgram - Target device program with full sequence - erase, blank check,
program, verify and blow security fuse (if enabled).
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Auto Program button is the most frequently function when programming microcontrollers in
the production process. Auto Program function activates all required procedures to fully program and
verify the flash memory contents. Typically, when flash memory needs to be erased, Auto Program
executes the following procedures:
- initialization
- read retained data (if required)
- read IEEE address contents(if required)
- erase flash memory,
- memory blank check,
- flash programming,
- restoring or writing the new IEEE address (if required)
- restoring retained data (if required)
- flash memory verification (check sum verification of whole verification byte by byte),
- write lock protection bits (if required).
- switch-off Vcc from target device.
Remote Control Programming User’s Guide PM024A02 Rev.2 64
Syntax: MSPPRG_API INT_X F_AutoProgram( INT_X mode );
mode = 0;
mode = 1 and up - reserved
Return value:0 - FALSE
1 - TRUE
Example: ............................
if( F_Initialization() != TRUE ) //required API-Dll - initialization
{
// Initialization error
}
F_GetSetup( &config ); //API-DLL - get configuration from the programmer
............................ // modify configuration if required
F_ConfigSetup( config ); // download setup to programmer
int st = F_ConfigFileLoad( “c:\test\configfile.cfg” );
if(( st & 1 ) != TRUE )
{
Info = st & 0xFFFE;
....................
}
do{
.................... // prepare next microcontrollers
targets_mask = 0x3F //active all six target devices
F_SetConfig( CFG_TARGET_EN_INDEX, (INT_X)targets_mask );
if( F_AutoProgram(0) == targets_mask )
{
//all target devices programmed
}
else
{
//some targets has nod been programmed
}
.................... //exit if the last microcontrollers
// has been programmed
} while(1); ....................
F_Verify_Lock_Bits
Remote Control Programming User’s Guide PM024A02 Rev.2 65
F_Verify_Lock_Bits -Verify the Lock debug Bit. If debug access is disabled, then only
debug bit is verified. Other bits are not accessible when the debug bit
enable is clear.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Lock Protection bits
bit-4 0x10 - Boot Block Lock
0 - Page 0 is write protected
1 - Page 0 is writable, unless LSIZE is 000
bits3:1 LSIZE - Sets the size of the upper Flash area which is write-protected
bit 0 0x01 Debug lock bit
0 - Disable debug command
1 - Enable debug command
Syntax: MSPPRG_API INT_X F_Verify_Lock_Bits( void );
Return value:0x100 | 8 bits Lock Bits value (see above)
if debug bit is enabled or
0x100
if debug bit is disabled (access to other bits is locked)
F_Memory_Erase
F_Memory_Erase - Erase Target’s Flash Memory
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Erase flash size, or sector to be erased, should be specified in the configuration setup. When mode
erase flag is set to one, then all memory will be erased, regardless erase memory configuration setup
value.
Syntax: MSPPRG_API INT_X F_Memory_Erase( INT_X mode );
mode = 0 -> erase space specify by the FlashEraseModeIndex;
mode = 1 -> erase all Flash memory, regardless FlashEraseModeIndex;
Return value:
Remote Control Programming User’s Guide PM024A02 Rev.2 66
0 - FALSE
1 - TRUE
F_Memory_Blank_Check
F_Memory_Blank_Check - Check if the Target’s Flash Memory is blank.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Syntax: MSPPRG_API INT_X F_Memory_Blank_Check( void );
Return value:0 - FALSE
1 - TRUE
F_Memory_Write
F_Memory_Write - Write content taken from the Code file to the selected Target
Devices Flash Memory.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Syntax: MSPPRG_API INT_X F_Memory_Write( INT_X mode );
mode = 0;
mode = 1 and up - reserved
Return value:0 - FALSE
1 - TRUE
F_Memory_Verify
F_Memory_Verify - Verify contents of the selected Target Devices Flash Memory and
Code file.
Remote Control Programming User’s Guide PM024A02 Rev.2 67
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Note: During the verification process either all memory or just the selected part of the memory is
verified, depending on settings specified in the configuration setup FlashEraseModeIndex.
Only data taken from the Code file are compared with the target’s flash memory. If size of the
flash memory is bigger then code size then all reminding data in flash memory is ignored.
Syntax: MSPPRG_API INT_X F_Memory_Verify( INT_X mode );
mode = 0;
mode = 1 and up - reserved
Return value:0 - FALSE
1 - TRUE
F_Memory_Read
F_Memory_Read - Read Flash Memory from selected or all Target Devices.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Size of the read memory size is defined in the configuration setup
All data will be saved in the internal Read Buffer. Contents from the Read Buffer can be taken using
function
BYTE F_Get_Byte_from_Buffer( INT_X addr );
Syntax: MSPPRG_API INT_X F_Memory_Read( void );
Return value:0 - FALSE
1 - TRUE
Example: BYTE data[MAX_FLASH_SIZE];
unsigned int addr,n;
....................
st = F_Memory_Read();
if ( st != 0 )
Remote Control Programming User’s Guide PM024A02 Rev.2 68
{
for( addr=0x0; addr<MAX_FLASH_SIZE; addr++)
data[addr] = F_Get_Byte_from_Buffer( addr );
}
....................
F_Write_IEEE_Address
F_Write_IEEE_Address - Write content taken from the IEEE Address buffer to target
devices. Write IEEE address option should be enabled in the
configuration setup.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Syntax:INT_X MSPPRG_API F_Write_IEEE_Address( void );
Return value:0 - FALSE
1 - TRUE
Example:
F_Put_IEEEAddr64_to_Buffer( 0x0123456789ABCDE0 );
F_Write_IEEE_Address();
F_Read_IEEE_Address
F_Read_IEEE_Address - Read IEEE Addresses from target devices and save it in the
IEEE buffer
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
Syntax:INT_X MSPPRG_API F_Read_IEEE_Address( void );
Remote Control Programming User’s Guide PM024A02 Rev.2 69
Return value:0 - FALSE
1 - TRUE
Example:#define “FlashProCC-Dll.h”;
ULONG64 IEEE;
F_Read_IEEE_Address();
IEEE = F_Get_IEEEAddr64_from_Buffer();
F_Write_Lock_Bits
F_Write_Lock_Bits - Write lock bits to target devices. Contents of the lock bits should be
set first using configuration setup instructions.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
MSPPRG_API INT_X F_Write_Lock_Bits( void );
Return value:
0 - FALSE
1 - TRUE
Remote Control Programming User’s Guide PM024A02 Rev.2 70
4.4 Sequential instructions
Sequential instructions allow access to the target device in any combination of the small
instructions like erase, read, write sector, modify part of memory etc. Sequential instruction have an
access only when communication between target device and programming adapter is initialized. This
can be done when F_Open_Target_Device instruction is called. When communication is established,
then any of the sequential instruction can be called. When the process is finished, then at the end
F_Close_Target_Device instruction should be called. When communication is terminated, then
sequential instructions can not be executed.
Note: Erase/Write/Verify/Read configuration setup is not required when sequential instructions are
called. Also code file is not required to be downloaded. All data to be written, erased, and read is
specified as a parameter to the sequential functions. Data downloaded from the code file is ignored
in this case.
Very important:
The sequential functions allows to program words in the FLASH memory on any flash space location.
Also the same bytes / words can be programmed few times. Software is not be able to control how
many times the same location of the flash has been programmed between erasures. User should take
a full responsibility for programming the flash memory according to the CCxx specifications. See TI’s
data sheets and manuals for details.
The following flash programming limitation should be taken to consideration:
1. The same word or byte can not be programmed more then twice between erasures. Otherwise,
damage can occur.
2. In the CCxx flash device two or four bytes are programmed simultaneously. This means -
programmed bytes should be prepared first and flashed as a block with two or four bytes
length. Otherwise four independent bytes programmed separately will be programmed four
times - one time with required data and 3 times with 0xFF data.
Note: CCxx devices with max flash size up to 32 kB have two bytes size programming word
in the flash, while the CCxx with bigger flash memory size (up to 128 kB), have four bytes size
writing word.
Remote Control Programming User’s Guide PM024A02 Rev.2 71
F_Open_Target_Device
F_Open_Target_Device - Initialization communication with the target device.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed simultaneously.
When F_Open_Target_Device is executed, then
- Power from the programming adapter becomes ON if PowerTargetEn in configuration
setup is enabled.
- Vcc is verified to be higher then 2.0V.
- communication between programming adapter and target device is initialized.
Target device is ready to get other sequential instructions.
Syntax: MSPPRG_API INT_X F_Open_Target_Device( void );
Return value:0 - FALSE
1 - TRUE
Example: int st, mask;
long addr;
....................
mask = 0x3F; //enable all six target devices
F_SetConfig( CFG_TARGET_EN_INDEX, (INT_X)mask );
F_Open_Target_Device();
....................
F_Segment_Erase(0x1000);
st = F_Sectors_Blank_Check( 0x1000, 0x107f );
if( st != mask )
{ ................ }
for( addr = 0x1000; addr<0x1020; addr++ )
F_Put_Byte_to_Buffer( addr, data(addr) )
F_Copy_Buffer_to_Flash( 0x1000, 0x20 );
F_Segment_Erase(0x4000);
....................
F_Close_Target_Device();
....................
F_Close_Target_Device
Remote Control Programming User’s Guide PM024A02 Rev.2 72
F_Close_Target_Device - Termination communication between target device and programming
adapter.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Instruction should be called on the end of the sequential instructions. When F_Close_Target_Device
instruction is executed then:
- Communication between target device and programming adapter is terminated.
- Power from the programming adapter becomes OFF (if selected).
- Target device is released from the programming adapter.
Syntax: MSPPRG_API INT_X F_Close_Target_Device( void );
Return value:0 - FALSE
1 - TRUE
Example:
See example above (F_Open_Target_Device).
F_Segment_Erase
F_Segment_Erase - Erase any segment of the CCxx Flash memory.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Parameters:
segment address - Any address form the desired segment space addresses
To erase a memory segment specify an address within that memory segment. For example to erase
segment 0x2000-0x27FF any address from the range 0x2000 to 0x27FF can be specified. To erase
all memory segments, erase the memory segment by segment, or used the encapsulated instruction F_Memory_Erase(1);
Note: When encapsulated instruction is executed, then next access to the sequential instruction can
be accessed only when F_Open_Target_Device instruction is called again.
Syntax:
Remote Control Programming User’s Guide PM024A02 Rev.2 73
MSPPRG_API INT_X F_Segment_Erase( INT_X address );
Return value:0 - FALSE
1 - TRUE
Example: ....................
F_Segment_Erase(0x4000); // erase segment 0x4000 to 0x47FF
F_Segment_Erase(0x4100); // erase the same segment
....................
F_Sectors_Blank_Check
F_Sectors_Blank_Check - Blank check part or all Flash Memory. Start and stop address of the
tested memory should be specified.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Parameters:
start address - Even number from 0x0 to 0x1FFFE,
stop address - Odd number from 0x1 to 0x1FFFF,,
Syntax: MSPPRG_API INT_X F_Sectors_Blank_Check( INT_X start_addr,
INT_X stop_addr );
Return value:0 - FALSE
1 - TRUE
Example: ....................
F_Sectors_Blank_Check (0x1000, 0x107F) ; //INFO secto blank check
F_Sectors_Blank_Check (0x8000, 0xFFFF) ; //32 kB memory size blank check
F_Sectors_Blank_Check (0x1220, 0x123f) ; //part of sector blank check
....................
F_Write_Byte_to_XRAM
Remote Control Programming User’s Guide PM024A02 Rev.2 74
F_Write_Byte_to_XRAM - Write one byte to XRAM.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs).
Write one byte to any XRAM location of the target devices.
Parameters:
address - address where XRAM is located 0xDF00 to 0xFFFF,
data - one byte to be written to target device
Syntax: MSPPRG_API INT_X F_Write_Byte_to_XRAM( INT_X addr, BYTE data );
Return value:0 - FALSE
1 - TRUE
Example:F_Write_Byte_to_XRAM( 0xF010, 0x21 );
F_Read_Byte_from_XRAM
F_Read_Byte_from_XRAM - Reade one byte from XRAM.
VALID FPA index - ( 1 to 8 ).
Read one byte from any XRAM location of the target devices.
Parameters:
address - address where XRAM is located 0xDF00 to 0xFFFF,
data - one byte to be written to target device
Syntax: MSPPRG_API INT_X F_Read_Byte_from_XRAM( INT_X addr );
Return value:0x00 to 0xFF - valid code data
-1 (0xFFFF) - FALSE
Example:F_Read_Byte_from_XRAM( 0xF010 );
Remote Control Programming User’s Guide PM024A02 Rev.2 75
F_Write_Byte_to_direct_RAM
F_Write_Byte_to_direct_RAM - Write one byte to direct RAM.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Write one byte to any direct RAM location of the target devices.
Parameters:
address - address where RAM is located 0x00 to 0xFF,
data - one byte to be written to target device
Syntax: MSPPRG_API INT_X F_Write_Byte_to_direct_RAM( INT_X addr, BYTE data );
Return value:0 - FALSE
1 - TRUE
Example:F_Write_Byte_to_direct_RAM( 0x60, 0x33 );
F_Read_Byte_from_direct_RAM
F_Read_Byte_from_direct_RAM - Read one byte from direct RAM.
VALID FPA index - ( 1 to 8 ).
Read one byte from any direct RAM location of the target devices.
Parameters:
address - address where RAM is located 0x00 to 0xFF,
data - one byte to be written to target device
Syntax: MSPPRG_API INT_X F_Read_Byte_from_direct_RAM( INT_X addr );
Return value:
Remote Control Programming User’s Guide PM024A02 Rev.2 76
0x00 to 0xFF - valid code data
-1 (0xFFFF) - FALSE
Example:F_Read_Byte_from_direct_RAM( 0x60 );
F_Copy_Buffer_to_Flash
F_Copy_Buffer_to_Flash - Write data from the Buffer to target devices.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Parameters:
start address - Flash address from 0x0 to 0x1FFFF,
size - Size from 1 to MAX_FLASH_SIZE (0x20000) - block of data in bytes
to be written.
Syntax: MSPPRG_API INT_X F_Copy_Buffer_to_Flash( INT_X start_addr, INT_X size );
Return value:0 - FALSE
1 - TRUE
Example: long addr;
int mask;
....................
for( addr = 0x1000; addr<0x2100; addr++ )
F_Put_Byte_to_Buffer( addr, (BYTE)(0xFF & addr));
....................
mask = 0x3F; //enable all six target devices
F_SetConfig( CFG_TARGET_EN_INDEX, (INT_X)mask );
F_Open_Target_Device();
F_Copy_Buffer_to_Flash( 0x1000, 0x1100 );
....................
F_Copy_Flash_to_Buffer
Remote Control Programming User’s Guide PM024A02 Rev.2 77
F_Copy_Flash_to_Buffer - Read specified in “size” number of bytes from Flash and save it in
the temporary buffer. Starting address is specified in the “start
address”.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax: MSPPRG_API INT_X F_Copy_Flash_to_Buffer( INT_X start_address,
INT_X size );
Parameters:start address
size
Return value:0 - FALSE
1 - TRUE
NOTE: Specified address in the temporary RAM/Flash buffer is the same as a physical RAM
address.
Example:
....................
....................
st = F_Copy_Flash_to_Buffer( 0x2220, 0xE0 );
if( st == TRUE )
{
for( addr = 0x2220; addr<0x2300; addr++ )
data[addr]= F_Get_Byte_from_Buffer( addr );
}
else
{
...................
}
....................
F_Copy_Buffer_to_XRAM
F_Copy_Buffer_to_XRAM - Write “size” number of bytes from the temporary
XRAM/Flash buffer to XRAM. Starting address is specified
in the “start address”.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
Remote Control Programming User’s Guide PM024A02 Rev.2 78
MSPPRG_API INT_X F_Copy_Buffer_to_XRAM( INT_X start_address, INT_X size );
Parameters:start address - physical XRAM address 0xDF00 to 0xFFFF
size - size in bytes
Return value:0 - FALSE
1 - TRUE
NOTE: Specified address in the temporary XRAM/Flash buffer is the same as a physical
XRAM address.
Example: ....................
....................
for( addr = 0xF220; addr<0xF300; addr++ )
st = F_Put_Byte_To_Buffer( addr, data[addr] );
st = F_Copy_Buffer_to_XRAM( 0xF220, 0xE0 );
....................
F_Copy_XRAM_to_Buffer
F_Copy_XRAM_to_Buffer - Read specified in “size” number of bytes from the XRAM and save
it in the temporary buffer. Starting address is specified in the “start
address”.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
MSPPRG_API INT_X F_Copy_XRAM_to_Buffer( INT_X start_address,
INT_X size );
Parameters:start address - physical XRAM address 0xDF00 to 0xFFFF
size - size in bytes
Return value:0 - FALSE
1 - TRUE
Remote Control Programming User’s Guide PM024A02 Rev.2 79
NOTE: Specified address in the temporary XRAM/Flash buffer is the same as a physical
XRAM address.
Example: ....................
....................
st = F_Copy_XRAM_to_Buffer( 0xF220, 0xE0 );
if( st == TRUE )
{
for( addr = 0xF220; addr<0xF300; addr++ )
data[addr]= F_Get_Byte_from_Buffer( addr );
}
else
{
...................
}
....................
F_Copy_Buffer_to_direct_RAM
F_Copy_Buffer_to_direct_RAM - Write “size” number of bytes from the temporary
XRAM/Flash buffer no ‘1' to direct RAM. Starting
address is specified in the “start address”.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
MSPPRG_API INT_X
F_Copy_Buffer_to_direct_RAM( INT_X start_address, INT_X size );
Parameters:
start address - direct RAM address 0x00 to 0xFF
size - size in bytes
Return value:
0 - FALSE
1 - TRUE
Example:
....................
....................
Remote Control Programming User’s Guide PM024A02 Rev.2 80
for( addr = 0x60; addr<0x7f; addr++ )
st = F_Put_Byte_To_Buffer( addr, data[addr] );
st = F_Copy_Buffer_to_direct_RAM( 0x60, 0x20 );
....................
F_Copy_direct_RAM_to_Buffer
F_Copy_direct_RAM_to_Buffer - Read specified in “size” number of bytes from the direct
RAM and save it in the temporary buffer. Starting address is
specified in the “start address”.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax: MSPPRG_API INT_X
F_Copy_direct_RAM_to_Buffer( INT_X start_address, INT_X size );
Parameters:start address - direct RAM address 0x00 to 0xFF
size - size in bytes
Return value:0 - FALSE
1 - TRUE
F_Put_Byte_to_Buffer
F_Put_Byte_to_Buffer - Write byte to temporary buffer.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax: MSPPRG_API INT_X F_Put_Byte_to_Buffer( INT_X address, BYTE data );
address: temporary buffer address equal the RAM/XRAM or Flash
destination address ( 0x0000 to 0x1FFFF )
data: Byte to be written.
Return value:TRUE if specified address is legal (0x0000 to 0x1FFFF) otherwise FALSE.
NOTE: Specified address in the temporary RAM or Flash buffer is the same as a physical
RAM/FLASH address.
Remote Control Programming User’s Guide PM024A02 Rev.2 81
NOTE: DLL contains two Buffers - one is dedicated to the data READ from target devices,
second one is dedicated to the data to be WRITE to the target devices. Contents of
the Buffers can not be verified by writing and read the same data to the Buffers e.g.
F_Put_Byte_to_Buffer (0x1000, 5);
data = F_Get_Byte_from_Buffer( 0x1000 );
Read data can be other that ‘5'.
Example: ....................
....................
for( addr = 0x1000; addr<0x1020; addr++ )
st = F_Put_Byte_to_Buffer( addr, data[addr][n] );
st = F_Copy_Buffer_to_Flash( 0x1000, 0x20 );
....................
F_Get_Byte_from_Buffer
F_Get_Byte_from_Buffer - Read one byte from the temporary RAM/Flash buffer.
VALID FPA index - ( 1 to 8 )
Syntax: MSPPRG_API BYTE F_Get_Byte_from_Buffer( INT_X address );
Return value:Requested byte from the specified address of the RAM/Flash temporary
buffer.
Example:
see F_Copy_All_Flash_To_Buffer.
NOTE: DLL contains two Buffers - one is dedicated to the data READ from target devices,
second one is dedicated to the data to be WRITE to the target devices. Contents of
the Buffers can not be verified by writing and read the same data to the Buffers.
F_Set_PC_and_RUN
Remote Control Programming User’s Guide PM024A02 Rev.2 82
F_Set_PC_and_RUN - Instructions allows to run program in microcontroller from specified
PC in the XRAM or Flash location. Program should be downloaded
first using the Write to Flash or XRAM procedures.
Note: The F_Open_Target_Device instruction is resetting the CPU. All internal registers states
are set to default value. The F_Synch_CPU_JTAG is synchronizing the CPU and JTAG on
fly. The CPU is stopped, but all registers have not been modified.
VALID FPA index - ( 1 to 8 ) or 0 (ALL FPAs) executed sequentially.
Syntax:
MSPPRG_API INT_X F_Set_PC_and_RUN( INT_X xram_en, INT_X PC_address );
Parameters:xram_en - 0 -> run program located in Flash
- 1 -> run program located in XRAM
address - set Program Counter to address and run
Return value:
0 - FALSE
1 - TRUE
F_Get_MCU_Data
F_Get_MCU_Data - Get the MCU status or device ID/silicon version
VALID FPA index - ( 1 to 8 ).
INT_X MSPPRG_API F_Get_MCU_Data( INT_X Type );
Parameters:Type: GET_MCU_ID ( 1 )
Result
Higher byte (bits 15:8) MCU ID
0x01 - CC1110F8
0x85 - CC2430F64
0x89 - CC2431F32
0x81 - CC2510F8
0x91 - CC2511F8
Lower byte (bits 15:8) Silicon version ID
GET_MCU_STATUS ( 2 )
Result one byte
0x80 - Chip Erase Done
Remote Control Programming User’s Guide PM024A02 Rev.2 83
0x40 - PCON Idle
0x20 - CPU halted
0x10 - Power Mode 0
0x08 - Halt Status
0x04 - Debug Locked
0x02 - Oscillator stable
0x01 - Stack overflow
Return value:data from the MCU buffer.
Remote Control Programming User’s Guide PM024A02 Rev.2 84
Appendix A
FlashPro-CC Command Line interpreter
The Multi-FPA API-DLL can be used with the command line interpreter shell. This shell allows to
use the standard Command Prompt windows or script file to execute the API-DLL functions. All
required files are located in the directory
C:\Program Files\Elprotronic\CCxx\USB FlashPro-CC\CMD-line
and contains
FP-CC-commandline.exe -> command line shell interpreter
FlashProCC-FPAsel.dll -> standard API-DLL files
FlashProCC-FPA1.dll -> ----,,,,,--------
All API-DLL files should be located in the same directory where the FP-CC-commandline.exe is
located. To start the command line interpreter, the FP-CC-commandline.exe should be executed.
Command Syntax:
instruction_name ( parameter1, parameter2, .... )
parameter:
1. string ( file name etc. ) - "filename"
2. numbers
integer decimal eg. 24
or integer hex eg. 0x18
Note: Spaces are ignored
Instructions are not case sensitive
F_OpenInstancesAndFPAs( "*# *" )
and f_openinstancesandfpas( "*# *" )
are the same.
Example-1:
Remote Control Programming User’s Guide PM024A02 Rev.2 85
Run the FP-CC-commandline.exe
Type:
F_OpenInstancesAndFPAs( "*# *" ) // open instances and find the first adapter (any SN)
Press ENTER - result ->1 (OK)
Type:
F_Initialization() //initialization with config taken from the config.ini
//setup taken from the FlashPro-CC - with defined CCxx type, code file etc.
Press ENTER - result ->1 (OK)
Type:
F_AutoProgram( 0 )
Press ENTER - result ->1 (OK )
Type:
F_Report_Message()
Press ENTER - result -> displayed the last report message (from the F_Autoprogram(0))
See figure A-1 for result:
Type quit() and press ENTER to close the FP-CC-commandline.exe program.
Figure A-1
Remote Control Programming User’s Guide PM024A02 Rev.2 86
Example-2:
Run the FP-CC-commandline.exe and type the following commands:
F_OpenInstancesAndFPAs( "*# *" ) // open instances and find the first adapter (any SN)
F_Initialization()
F_Report_Message()
F_ConfigFileLoad( "filename" ) //put vaild path and config file name
F_ReadCodeFile( 1, "FileName" ) //put vaild path and code file name (TI.txt format)
F_AutoProgram( 0 )
F_Report_Message()
........................
........................
F_Put_Byte_to_Buffer( 1, 0x8000, 0x11 )
F_Put_Byte_to_Buffer( 1, 0x8001, 0x21 )
........................
F_Put_Byte_to_Buffer( 1, 0x801F, 0xA6 )
F_Open_Target_Device()
F_Segment_Erase( 0x8000 )
F_Copy_Buffer_to_Flash( 0x8000, 0x20 )
F_Copy_Flash_to_Buffer( 0x8000, 0x20 )
F_Get_Byte_from_Buffer( 1, 0x8000 )
F_Get_Byte_from_Buffer( 2, 0x8000 )
F_Get_Byte_from_Buffer( 1, 0x8001 )
F_Get_Byte_from_Buffer( 2, 0x8001 )
........................
F_Get_Byte_from_Buffer( 1, 0x801F )
F_Get_Byte_from_Buffer( 2, 0x801F )
F_Close_Target_Device()
quit()
Remote Control Programming User’s Guide PM024A02 Rev.2 87
List of command line instructions
quit() ;close the command interpreter program
help() ;display list below
F_Trace_ON()
F_Trace_OFF()
F_OpenInstances( no )
F_CloseInstances()
F_OpenInstancesAndFPAs( "FileName" )
F_Set_FPA_index( fpa )
F_Get_FPA_index()
F_LastStatus( fpa )
F_DLLTypeVer()
F_Multi_DLLTypeVer()
F_Check_FPA_access(index )
F_Get_FPA_SN( fpa )
F_APIDLL_Directory( "APIDLLpath" )
F_Initialization()
F_DispSetup()
F_Close_All()
F_Power_Target( OnOff )
F_Reset_Target()
F_Report_Message()
F_ReadCodeFile( file_format, "FileName" )
F_Get_CodeCS( dest )
F_ConfigFileLoad( "filename" )
F_SetConfig( index, data )
F_GetConfig( index )
F_Put_Byte_to_Buffer( addr, data )
F_Get_Byte_from_Buffer( addr )
F_Clr_Code_Buffer()
F_Put_Byte_to_Code_Buffer( addr, data )
F_Get_Byte_from_Code_Buffer( addr )
F_Put_IEEEAddr64_to_Buffer("Hex data string")
F_Get_IEEEAddr64_from_Buffer( )
F_AutoProgram( 0 )
Remote Control Programming User’s Guide PM024A02 Rev.2 88
F_Verify_Lock_Bits()
F_Memory_Erase( mode )
F_Memory_Blank_Check()
F_Memory_Write( mode )
F_Memory_Verify( mode )
F_Memory_Read()
F_Write_IEEE_Address()
F_Read_IEEE_Address()
F_Write_Lock_Bits()
F_Open_Target_Device()
F_Close_Target_Device()
F_Segment_Erase( address )
F_Sectors_Blank_Check( start_addr, stop_addr )
F_Copy_Buffer_to_Flash( start_addr, size )
F_Flash_to_Buffer( start_addr, size )
F_Write_Byte_to_XRAM( addr, data )
F_Read_Byte_from_XRAM( addr )
F_Write_Byte_to_direct_RAM( addr, data )
F_Read_Byte_from_direct_RAM( addr )
F_Copy_Buffer_to_XRAM( start_addr, size )
F_Copy_XRAM_to_Buffer( start_addr, size )
F_Copy_Buffer_to_direct_RAM( start_addr, size )
F_Copy_direct_RAM_to_Buffer( start_addr, size )
F_Set_PC_and_RUN( xram_en, PC_addr )
F_Get_MCU_Data(type )
F_Get_Targets_Vcc()
F_Disable_FPA_index( fpa )
F_Enable_FPA_index( fpa )
See chapter 4 for detailed description of the instructions listed above.
Note: Not all instructions listed in the chapter 4 are implemented in the command line
interpreter. For example - all instructions uses pointers are not implemented, however
this is not limiting the access to all features of the API-DLLs, because all instructions
uses pointers are implemented also in the simpler way without pointers.